This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Optic nerve and retinal nerve fiber layer analyzers
Glaucoma is an optic neuropathy characterized by a typi-
cal pattern of visual field loss and optic nerve damage
resulting from retinal ganglion cell death caused by a
number of different disorders that affect the eye. Most,
but not all, of these disorders are associated with el-
evated intraocular pressure (IOP), which is the most im-
portant risk factor for glaucomatous damage. Although
clinical examination of the optic nerve head has been
considered to be the most sensitive test for detecting
glaucomatous damage, evidence suggests that examina-
tion of the retinal nerve fiber layer (RNFL) may provide
important diagnostic information [1-4]. Accurate and ob-
jective methods of detecting disc and RNFL abnormali-
ties, and their progression, would facilitate the diagnosis
and monitoring of glaucomatous optic neuropathy.
Clinical examination and photography of the RNFL is a
difficult technique in many patients, particularly older
individuals, those with small pupils, and subjects with
media opacities. It is subjective, qualitative, variably re-
producible, and often unreliable. In addition, optic nerve
head and RNFL photography is time consuming, opera-
tor dependent, has limited sensitivity and specificity,
and requires storage space. Recently, new technologies
have emerged which enable clinicians to perform accu-
rate, reproducible, objective, and quantitative measure-
ments of the retinal nerve fiber layer and optic nerve
head topography.
Confocal scanning laser ophthalmoscopy (CSLO), a
technology embodied in the Heidelberg Retinal
Tomograph (HRT, Heidelberg Engineering, Heidel-
berg, Germany), enables the operator to evaluate three-
dimensional characteristics of optic nerve head topogra-
phy quantitatively [5-8]. Thirty-two coronal sections of
the optic nerve head are acquired over a depth of ap-
proximately 3.5 millimeters, and a color-coded topo-
graphic map of the optic nerve head is generated.
Scanning laser polarimetry (SLP) is a technology embod-
ied in the GDx Nerve Fiber Analyzer (Laser Diagnostic
Technologies, Inc., San Diego, CA) employs a confocal
scanning laser ophthalmoscope and an integrated polar-
imeter. It evaluates the thickness of the RNFL by uti-
lizing the birefringent properties of the retinal ganglion
cell axons [9,10]. As polarized light passes through the
RNFL and is reflected back from the deeper layer, it
undergoes a phase shift. The change in polarization, re-
The Department of Ophthalmology, The University of Miami School of Medicine,Bascom Palmer Eye Institute, Miami, Florida, USA.
Correspondence to David S. Greenfield, MD, Bascom Palmer Eye Institute, 7108Fairway Drive, Suite 340, Palm Beach Gardens, FL, 33418; e-mail:[email protected]
Confocal scanning laser ophthalmoscopy employs a 670 nm
diode laser beam as a light source and scans the retina in
x- and y- directions [14,15]. Light originating from the
illuminated area passes through a diaphragm (pinhole) in
a plane optically conjugate to the retina. Planes unfo-
cused at the aperture are blocked by the diaphragm and
do not reach the detector. Each image contains 256 x 256
pixels (picture-elements); each pixel represents the reti-
nal height at that location relative to the focal plane of
the eye. Image acquisition and processing takes approxi-
mately 1.6 seconds. Thirty-two coronal sections are ob-
tained progressing from anterior to the optic nerve head
through the retrolaminar portion of the nerve head. The
axial distance between two adjacent sections is 50 to
75 m generating an axial range of 1.5 to 3.5 mm.
A standard reference plane is established parallel to the
peripapillary retinal surface and is located 50 microns
posterior to the retinal surface along a circle concentric
with the optic disc margin in a temporal segment be-
tween 350° and 356°. Neural rim is defined as tissue
within the optic disc margin and above the reference
plane. Optic cup is defined as tissue within the disc
margin and below the reference plane.
The optic disc margin is outlined and a color-coded
depth map is created from a mean topographic image
using a software algorithm (Fig. 1). Stereometric param-
eters of optic nerve head topography are generated rela-
tive to the reference plane including rim area and vol-
ume, cup area and volume, cup-disc area ratio, mean
retinal nerve fiber layer thickness, and retinal nerve fiber
cross-sectional area. Parameters independent of the ref-
erence plane include mean and maximum cup depth,
height variation contour, and cup-shape measure. A nor-
mal retinal height variation diagram demonstrates a
“double-hump” pattern corresponding to the thicker
retinal ganglion cell axons along the superior and inferior
portions of the optic nerve head.
Reproducibility
Various investigators have reported high levels of repro-
ducibility using this technology [5,15,16] Brigatti et al. [7]
found that topographic variability correlated with the
steepness of the corresponding region. Greater variabil-
ity was found at the edge of the optic disc cup and along
blood vessels. Weinreb et al. [14] have determined that
measurement reproducibility is improved from 35.5 µm
to 25.7 µm when a series of three examinations are ob-
tained instead of a single image analysis. Based upon
these data, acquiring three images per eye and creation
of a mean topographic image is recommended. Finally,
Zangwill et al. [17] have shown that image reproducibil-
ity is improved with pupillary dilation, particularly in
eyes with small pupils and cataract.
Sensitivity and specificity
Various investigators have reported topographic differ-
ences between normal, ocular hypertensive, and glauco-
matous eyes. It is essential to emphasize that the char-
acteristics of the study population will influence the
discriminating power involved in differentiating glauco-
matous from nonglaucomatous eyes. Determination of
sensitivity and specificity parameters is fundamentally
linked to the severity of glaucomatous damage among
the cohort studied. For any given technology, an instru-
ment will appear to be more sensitive if it is used to
separate eyes with advanced glaucoma from normal sub-
jects compared with eyes with mild glaucoma.
Heidelberg Retinal Tomograph employs software with
various statistical analyses to discriminate normal from
Figure 1. Confocal scanning laser ophthalmoscopy
topographic map
A patient with moderate normal-tension glaucoma shows loss of the inferiorneuroretinal rim (green) and associated stereometric parameters. There is a focaldepression in the double-hump pattern of the height variation diagramcorresponding to the decreased inferotemporal quadrant height (below).
Optic nerve and retinal nerve fiber layer analyzers in glaucoma Greenfield 69
abnormal optic discs. These include a multivariate dis-
criminant analysis based upon rim volume, height varia-
tion contour, and cup shape measure adjusted by age
[18], ranked-segment distribution curves [19,20], and re-
gression analysis using a normative database of 80 normal
eyes from 80 white subjects with a mean age of 57 years
[21]. The confidence interval limits derived from the
later are used commercially to generate the Moorfield’s
Regression Classification Score (normal, borderline, or
outside normal limits). Wollstein et al. [21] reported a
84.3% sensitivity and a 96.3% specificity for separating
normal and early glaucomatous eyes by taking into ac-
count the relation between optic disc size and the rim
area or cup-to-disc area ratio. In a different study, Woll-
stein et al. [22•] determined that by taking into account
the optic disc size, HRT image analysis was superior in
sensitivity (84.3%) for detection of early glaucoma com-
pared with expert assessment of stereoscopic optic disc
photographs (70.6%).
The sensitivity and specificity of various HRT param-
eters has been investigated and varies widely ranging
from 62% to 94% and 74% to 96%, respectively [18,23–
27]. Wide variability in discriminating power may be ex-
plained in part by variable sample size, definitions of
glaucoma, and varying degrees of glaucomatous optic
nerve damage. A recent study by Miglior et al. [28•]
found fair to poor agreement (� statistic 0.28-0.48) be-
tween visual field examinations and HRT classifications
among a population of 359 eyes (55 normal, 209 with
OHT, and 95 with moderate POAG, average visual field
mean defect –7.6 dB) The sensitivity and specificity of
the HRT examination were, respectively, 80% and 65%,
using the Mikelberg multivariate discriminant analysis
[18], and 31 to 53% and 90 to 92%, using ranked-segment
distribution curve analysis [19,20].
Using various HRT summary data including the reflec-
surements throughout the peripapillary region and along
the measurement ellipse. Average quadrantic measure-
ments, measurement ratios (eg, superior/nasal,
superior/temporal), symmetry measurements between
superior and inferior quadrants, and modulation param-
eters (an indication of the difference between the thick-
est and thinnest parts of the RNFL) are generated. A
neural network number is also calculated which is
thought to reflect the likelihood of glaucoma on a scale of
0 to 100.
Reproducibility
Intraoperator measurement reproducibility has been
shown by Weinreb et al. [10] (mean coefficient of varia-
tion (CV) of 4.5%) and Chi et al. [46] (CV ranging from
3.59–10.20% for both normal and glaucomatous sub-
jects). Swanson et al. [47] found significant interoperator
variability with the NFA I, among 4 operators all of
whom only scanned each of the 11 subjects twice. The
primary source of error was attributed to the variability in
the criterion used for establishing intensity setting. This
problem was subsequently reduced in the NFA II with a
hardware modification to the light system.
Retinal nerve fiber layer thickness measurements using
the NFA II have been reported to have high levels of
measurement reproducibility [40,48]. Hoh et al. [40] de-
scribed excellent intraoperator reproducibility and found
that variability between operators can be minimized by
using a single measurement ellipse acquired from the
original baseline image. As investigators have reported
high levels of measurement variability adjacent to retinal
blood vessels [49,50], an automated blood vessel removal
algorithm has been incorporated in the third generation
device, GDx.
Sensitivity and specificity
As described with CSLO, there is a wide range in RNFL
thickness values among normal individuals and consid-
erable measurement overlap between normal and glau-
comatous eyes may exist. Determination of sensitivity
and specificity parameters is fundamentally linked to the
Figure 2. Scanning laser polarimetry image
A patient with moderate primary open-angle glaucoma shows reducedretardation within the superior arcuate retinal nerve fiber layer bundle. Tworetardation parameters were classified as abnormal (outside 95% confidencelimits, illustrated in red) and four parameters were classified as borderline(outside 90% confidence limits, illustrated in yellow).
Optic nerve and retinal nerve fiber layer analyzers in glaucoma Greenfield 71
severity of glaucomatous damage among the cohort stud-
ied [24]. Sensitivity and specificity values will be greater
in studies involving eyes with advanced glaucoma than
in studies involving eyes with mild to moderate glau-
coma. Tjon-Fo-Song and Lemij [38] evaluated the sen-
sitivity and specificity of the first generation device,
NFA I, for detecting glaucoma among a diverse group of
200 eyes with early to advanced glaucoma (average visual
field mean deviation –10.33 decibels) compared with a
normal population. The sensitivity and specificity was
reported to be 96 and 93%, respectively. Weinreb et al.[51] reported a sensitivity of 74% and specificity of 92%
using a newer version of SLP with a linear discriminant
function to label glaucomatous damage among a popula-
tion with early to moderate glaucoma. Garcia-Sánchez
et al. [52] found the sensitivity and specificity of the GDx
to be 78% and 86%, respectively. The most sensitive and
specific parameters in their study were ellipse modula-
tion, superior/nasal ratio, and maximum modulation.
In a cross-sectional study comparing OCT and SLP, Hoh
et al. [53] found that structural information generated
from both technologies was significantly correlated with
visual function in glaucomatous eyes (average visual field
mean deviation –7.7 decibels). However, retardation pa-
rameters providing summary measures of RNFL thick-
ness (eg, average thickness and integral measurements)
had a weaker correlation with visual field mean defect
(R = 0.17 to 0.27) than with constructed retardation pa-
rameters (eg, modulation scores, ratio parameters, and
number; R = 0.36 to –0.51). Bowd et al. [54] recently
reported that constructed SLP parameters (modulation,
ratio, number, and linear discriminant function values)
have the greatest discriminating power. This is ex-
plained by recent evidence [44] suggesting that interin-
dividual variability in corneal birefringence has falsely
broadened the normative database of RNFL thick-
ness assessments, and reduced the sensitivity and speci-
ficity of this technology. Correction for corneal polariza-
tion axis has been shown to significantly increase
the correlation between RNFL structural damage and
visual function, and significantly improve the discri-
minating power of SLP for detection of mild to moder-
ate glaucoma.
Garcia-Sanchez et al. [29] evaluated the sensitivity and
specificity of the HRT, GDx, and OCT summary data for
detection of early to moderate glaucoma (average visual
field mean defect –5.0 dB) among three masked reviewers
(see Table 1). For the GDx, sensitivity and specificity
ranged from 72 to 82% and 56 to 82%, respectively.
Detection of progression
Scanning laser polarimetry strategies for change detec-
tion exist including evaluation of change in absolute val-
ues of retardation measurements, change in quadrantic
RNFL thickness measurements, change in double-
hump RNFL thickness profile, and color-coded map of
RNFL thickness change relative to baseline. However,
as with OCT, statistical units of change probability are
absent limiting the ability to differentiate change from
measurement variability, and there has been no prospec-
tive validation of this algorithm
Two published reports have described SLP evidence of
change detection in eyes with non-glaucomatous optic
neuropathy. Colen et al. [55] described a patient with
Peripapillary retinal nerve fiber layer (RNFL) retardation map (A) andcorresponding RNFL thickness plot (B) in the right eyes of six normal individualswith different corneal polarization axis values (18°, 27°, 37°, 52°, 59°, 76° nasallydownward from top left to bottom right). Upper and lower margins in (B)represent 95% confidence intervals. Note that peripapillary retardation andmeasured RNFL thickness increase with increasing corneal polarization axis.(Reprinted with permission: Greenfield DS, Knighton RW: Stability of cornealpolarization axis measurements for scanning laser polarimetry. Ophthalmology2001, 108:1065–1069. Figure 3).
Optic nerve and retinal nerve fiber layer analyzers in glaucoma Greenfield 73
commercially available device capable of performing
scan acquisition times in one second.
Published series of peripapillary retinal nerve fiber layer
measurement using optical coherence tomography have
sampled 100 evenly-distributed points on a 360 degree
peripapillary circular scan. Ozden et al. [64] evaluated
whether a four-fold increase in sampling density im-
proves the reproducibility of OCT measurement.
Twenty-two eyes of 22 patients (normal subjects, 3 eyes;
ocular hypertension, 2 eyes; glaucoma, 17 eyes) were
measurements and CV were calculated for the superior
and inferior quadrants for each sampling density tech-
nique. Normal eyes showed no difference between the
25 point/quadrant and 100 point/quadrant scans, respec-
tively. Among glaucomatous eyes, however, the CV in
25-point/quadrant scans (25.9%) was significantly higher
than that in 100-point/quadrant scans (11.9%, p = 0.01).
Sensitivity and specificity
Cross-sectional studies have compared OCT with CSLO
[65] and SLP [53] in normal, ocular hypertensive, and
glaucomatous eyes. OCT was capable of differentiating
glaucomatous from non-glaucomatous eyes, and RNFL
thickness measurements using OCT correlated with re-
tardation measurements using SLP and topographic
measurements using CSLO.
Bowd et al. [54] compared the discriminating powers of
SLP, OCT, short-wavelength automated perimetry
(SWAP), frequency-doubling technology perimetry
(FDT) for detection of early glaucoma (average visual
field mean defect –4.0 dB). The largest area under the
receiver operator characteristic (ROC) curve was found
for OCT inferior quadrant thickness, followed by the
FDT number of total deviation plot points </= 5%, SLP
linear discriminant function, and SWAP pattern SD.
Zangwill et al. [66• ] compared the ability of OCT, HRT,
and GDx to discriminate between normal eyes and eyes
with early to moderate glaucomatous visual field loss. No
significant differences were found between area under
the ROC curve and the best parameter from each instru-
ment: OCT inferior RNFL thickness, HRT mean height
contour in the inferior nasal position, and GDx linear
discriminant function).
Garcia-Sanchez et al. [29] evaluated the sensitivity and
specificity of the HRT, GDx, and OCT summary
data for detection of early to moderate glaucoma (aver-
age visual field mean defect –5.0 dB) among three
masked reviewers (see Table 1). For the OCT, sensi-
tivity and specificity ranged from 76 to 79% and 68 to
81%, respectively.
Detection of progression
Change analysis software has only recently been intro-
duced; therefore no reports have described longitudinal
change in patients with disease progression. As presently
configured, this algorithm generates a serial analysis of
RNFL thickness measurements among two OCT im-
ages, however statistical units of change probability are
not provided. Thus, true biological change cannot be
differentiated from test-retest variability.
Limitations
Currently, no statistical units of change probability are
absent from the change analysis software, therefore one
cannot differentiate biological change from measure-
ment variability by performing serial analysis of abso-
lute RNFL thickness values. Pupillary dilation is re-
quired to obtain acceptable peripapillary measurement
scans. Finally, sampling is limited to 25 A-scans per
quadrant, which may limit the ability to detect localized
change [64].
ConclusionsRecent advances in ocular imaging technology provide a
means to obtain accurate, objective, quantitative, and
reproducible structural measurements of optic disc to-
pography and RNFL thickness. Current imaging sys-
tems can differentiate between normal eyes and eyes
with mild to moderate glaucomatous optic neuropathy.
Although conflicting data exists, sensitivity and specific-
ity values approximate 70 to 80% depending upon
sample size, definition of glaucoma, and severity of glau-
comatous damage. Any one technology will have limited
usefulness as a single test to diagnose glaucoma and at
the present juncture should not be used as an indepen-
dent diagnostic screening test. However, these instru-
Figure 4. Optical coherence tomography image of a normal
eye obtained using a 3.4 mm peripapillary measurement scan
The anterior and posterior limits of the retinal nerve fiber layer (RNFL) aredemarcated using a computer algorithm (arrows) and clock hour and quadranticRNFL thickness measurements are obtained.
74 Glaucoma
ments have considerable potential for use as adjunctive
measures of glaucomatous damage along with careful
clinical and perimetric examination.
There is no uniform agreement regarding the most ap-
propriate technology for the evaluation of structural
damage in eyes with glaucomatous optic neuropathy.
Furthermore, among proponents of any given technol-
ogy, there is no consensus on the most appropriate sum-
mary measure to represent ganglion cell loss. It is im-
portant to recognize that the parameter or technology
most useful in the detection of glaucomatous damage
may vary from individual to individual and may differ
from the parameter or technology most useful for de-
tection of glaucomatous change. The most appropriate
measure(s) of disease detection will unlikely be the
most sensitive indicator of glaucomatous change. At
the present time, limited information exists regarding
the relation between glaucomatous progression and
RNFL/topographic measures.
Currently available imaging technologies hold consider-
able promise for detection of glaucomatous change.
Methods for change detection exist but have not been
prospectively validated in large populations. Moreover,
new strategies for detection of progressive structural
change need to be validated against accepted measures
of structural (stereoscopic disc photography) and func-
tional (psychophysical) change. Statistical units of
change probability are essential to differentiate true bio-
logical change from variability (eg, microsaccades during
fixation, vessel pulsations, instrument or operator-
induced variability). A significant challenge to the inves-
tigator has been the reality that technology improves
with time. Rapidly evolving hardware and software re-
sults in alteration of baseline measurements. This has
produced instability in longitudinal data collection and
has limited, in part, our ability to critically evaluate the
efficacy of these instruments to detect structural change
over time. Presently, it is unclear whether automated
detection of structural change meets or exceeds current
standard of care measures.
In summary, each ocular imaging technology has specific
advantages and disadvantages. One instrument may not
be best for all purposes and all patients, and different
analysis strategies may not agree. Because measure-
ment reproducibility is high, each technology holds
promise for improving our ability to detect glaucoma-
tous change. As with perimetry, it is not recommended
that isolated clinical decisions be based solely upon
ocular imaging results. Clinical correlation should be
performed and treatment recommendations should
be individualized.
Acknowledgments
Supported in part by the New York Community Trust, New York, New York; TheKessel Foundation, Bergenfield, New Jersey; The Boyer Foundation, Melbourne,FL; and NIH Grant R01-EY08684, Bethesda, Maryland. The author has no propri-etary interest in any of the products or techniques described in this manuscript.
References and recommended reading
Papers of particular interest, published within the annual period of review,have been highlighted as:• Of special interest•• Of outstanding interest
1 Sommer A, Miller NR, Pollack I, et al.: The nerve fiber layer in the diagnosis ofglaucoma. Arch Ophthalmol 1977, 95:2149.
2 Quigley HA, Dunkelberger GR, Green WR: Retinal ganglion cell atrophy cor-related with automated perimetry in human eyes with glaucoma. Am J Oph-thalmol 1989, 107:453–464.
4 Sommer A, Katz J, Quigley HA, et al.: Clinically detectable nerve fiber layeratrophy precedes the onset of glaucomatous field loss. Arch Ophthalmol1991, 109:77–83.
5 Lusky M, Bosem ME, Weinreb RN: Reproducibility of optic nerve head to-pography measurements in eyes with undilated pupils. J Glaucoma 1993,2:104–109.
6 Weinreb RN, Dreher AW, Bille JF: Quantitative assessment of the optic nervehead with the laser tomographic scanner. Int Ophthalmol 1989, 13:25.
7 Brigatti L, Weitzman M, Caprioli J: Regional test-retest variability of confocalscanning laser tomography. Am J Ophthalmol 1995, 120:433–440.
8 Zangwill L, Schakiba S, Caprioli J, et al.: Agreement between clinicians and aconfocal scanning laser ophthalmoscope in estimating cup-to-disc ratios. AmJ Ophthalmol 1995; 119:415–421.
9 Dreher AW, Reiter K, Weinreb RN: Spatially resolved birefringence of theretinal nerve fiber layer assessed with a retinal ellipsometer. Applied Optics1992, 31:3730–3749.
10 Weinreb RN, Shakiba S, Zangwill L: Scanning laser polarimetry to measurethe nerve fiber layer of normal and glaucomatous eyes. Am J Ophthalmol1995, 119:627–636.
11 Huang D, Swanson EA, Lin CP, et al.: Optical coherence tomography. Sci-ence 1991 254:1178–1181.
12 Izatt JA, Hee MR, Swanson EA, et al.: Micrometer-scale resolution imaging ofthe anterior eye in vivo with optical coherence tomography. Arch Ophthalmol1994, 112:1584–1589.
13 Hee MR, Izatt JA, Swanson EA, et al.: Optical coherence tomography of thehuman retina. Arch Ophthalmol 1995, 113:325–332.
14 Weinreb RN, Lusky M, Bartsch D, et al.: Effect of repetitive imaging on topo-graphic measurements of the optic nerve head. Arch Ophthalmol 1993,111:636–638.
15 Mikelberg FS, Wijsman K, Schulzer M: Reproducibility of topographic param-eters obtained with the Heidelberg Retina Tomograph. J Glaucoma 1993,2:101–103.
16 Dreher AW, Tso PC, Weinreb RN: Reproducibility of topographic measure-ments of the normal and glaucomatous optic nerve head with the laser tomo-graphic scanner. Am J Ophthalmol 1991, 111:221.
17 Zangwill L, Irak I, Berry CC, et al.: Effect of cataract and pupil size on imagequality with confocal scanning laser ophthalmoscopy. Arch Ophthalmol1997, 115:983–990.
18 Mikelberg FS, Parfitt CM, Swindale NV, et al.: Ability of the Heidelberg retinatomograph to detect early glaucomatous field loss. J Glaucoma 1995,4:242–247.
19 Bartz-Schmidt KU, Sengersdorf A, Esser P, et al.: The cumulative normalisedrim/disc area ratio curve. Graefes Arch Clin Exp Ophthalmol 1006, 234:227–231.
20 Asawaphureekorn S, Zangwill L, Weinreb RN: Ranked-segment distributioncurve for interpretation of optic nerve topography. J Glaucoma 1996,5:79–90.
21 Wollstein G, Garway-Heath DF, Hitchings RA: Identification of early glau-coma cases with the scanning laser ophthalmoscope. Ophthalmology 1998,105:1557–1563.
•22 Wollstein G, Garway-Heath DF, Fontana L, et al.: Identifying early glaucoma-
tous changes: comparison between expert clinical assessment of optic discphotographs and confocal scanning ophthalmoscopy. Ophthalmology 2000,107:2272–2277.
This study found the Heidelberg Retinal Tomograph to be superior to clinical evalu-ation of optic disc stereophotographs but was limited by the use of disparate ob-servers with limited levels of agreement.
23 Iester M, Mikelberg FS, Drance SM: The effect of optic disc size on diagnosticprecision with the Heidelberg retina tomograph. Ophthalmology 1997,104:545–548.
Optic nerve and retinal nerve fiber layer analyzers in glaucoma Greenfield 75
24 Bathija R, Zangwill L, Berry CB, et al.: Detection of early glaucomatous struc-tural damage with confocal scanning laser tomography. J Glaucoma 1998,7:121–127.
25 Caprioli J, Park HJ, Ugurlu S, et al.: Slope of the peripapillary nerve fiber layersurface in glaucoma. Invest Ophthalmol Vis Sci 1998, 39:2321–2328.
26 Gundersen KG, Asman P: Comparison of ranked segment analysis (RSA)and cup to disc ratio in computer-assisted optic disc evaluation. Acta Oph-thalmol Scand 2000, 78:137–141.
27 Swindale NV, Stjepanovic G, Chin A, et al.: Automated analysis of normal andglaucomatous optic nerve head topography images. Invest Ophthalmol VisSci 2000, 41:1730–1742.
•28 Miglior S, Casula M, Guareschi M, et al.: Clinical ability of Heidelberg Retinal
Tomograph examination to detect glaucomatous visual field changes. Oph-thalmology 2001, 2001:1621–1627.
In a broad clinical setting of 359 normal, ocular hypertensive, or glaucomatouseyes, this study found that the Heidelberg Retinal Tomograph lacked sensitivity andspecificity when using commercially available ranked-segment distribution curves,and Mikelberg’s multivariate discriminant analysis, respectively.
•29 Sanchez-Galeana C, Bowd C, Blumenthal EZ, et al.: Using optical imaging
summary data to detect glaucoma. Ophthalmology 2001, 108:1812–1818.This report evaluated the sensitivity and specificity of ocular imaging summary datausing three experienced observers with high levels of agreement and found similardiscriminating powers among HRT, GDx, and OCT.
30 Anderson DR, Chauhan B, Johnson C, et al.: Criteria for progression of glau-coma in clinical management and in outcome studies. Am J Ophthalmol 2000,130:827–829.
••31 Chauhan BC, Blanchard JW, Hamilton DC, et al.: Technique for detecting
serial topographic changes in the optic disc and peripapillary retina usingscanning laser tomography. Invest Ophthalmol Vis Sci 2000, 41:775–782.
This study describes a novel change analysis algorithm for differentiating test-retest variability from statistically significant change in optic disc topography and iscurrently employed within the latest version of HRT software.
32 Kamal DS, Viswanathan AC, Garway-Heath DF, et al.: Detection of optic discchange with the Heidelberg Retina Tomograph before confirmed visual fieldchange in ocular hypertensives converting to early glaucoma. Br J Ophthalmol1999, 83:290–294.
33 Lusky M, Morsman D, Weinreb RN: Effects of intraocular pressure on opticnerve head topography. Curr Opin Ophthalmol 1993, 4:40–44.
34 Chauhan BC, MacDonald CA: Influence of time separation on variability es-timates of topographic measurements with confocal scanning laser tomog-raphy. J Glaucoma 1995, 4:189–193.
35 Lesk MR, Spaeth GR, Azuara-Blanco A, et al.: Reversal of optic disc cuppingafter glaucoma surgery analyzed with a scanning laser tomograph. Ophthal-mology 1999, 106:1013–1018.
36 Irak I, Zangwill L, Garden V, et al.: Change in optic disk topography aftertrabeculectomy. Am J Ophthalmol 1996, 122:690–695.
37 Lee BL, Zangwill L, Weinreb RN: Change in optic disc topography associatedwith diurnal variation of intraocular pressure. J Glaucoma 1999; 8:221–223.
38 Tjon-Fo-Sang MJ, Lemij HG: The sensitivity and specificity of nerve fiber layermeasurements in glaucoma as determined with scanning laser polarimetry.Am J Ophthalmol 1997, 123:62–69.
39 Niessen AGJE, Van Den Berg TJTP, Langerhorst CT, et al.: Retinal nerve fiberlayer assessment by scanning laser polarimetry and standardized photogra-phy. Am J Ophthalmol 1996, 121:484–493.
41 Reiter K, Dreher A, inventors; Laser Diagnostic Technologies, Inc., assignee.Eye examination apparatus employing polarized light probe. 1998 Aug 4.United States Patent No. 5,787,890.
42 Dreher A, Reiter K, inventors; Laser Diagnostic Technologies, Inc., assignee.Retinal eye disease diagnostic system. 1994 April 19. United States PatentNo. 5,303,709.
43 Knighton RW, Huang W-R, Greenfield DS: Linear birefringence measured inthe central cornea of a normal population. Invest Ophthalmol Vis Sci 2002,43:82–86.
44 Greenfield DS, Knighton RW, Feuer W, et al.: Correction for corneal polar-ization axis improves the discriminating power of scanning laser polarimetry.Invest Ophthalmol Vis Sci 2001; 42:S314.
45 Hoh ST, Greenfield DS, Liebmann JM, et al.: Effect of pupillary dilation onretinal nerve fiber layer thickness measurement using scanning laser polarim-etry. J Glaucoma 1999, 8:159–163.
46 Chi QM, Tomita G, Inazumi K, et al.: Evaluation of the effect of aging on theretinal nerve fiber layer thickness using scanning laser polarimetry. J Glau-coma 1995, 4:406–413.
47 Swanson WH, Lynn JR, Fellman RL, et al.: Inter-operator variability in imagesobtained by laser polarimetry of the nerve fiber layer. J Glaucoma 1995,4:414–418.
48 Zangwill L, Berry CA, Garden VS, et al.: Reproducibility of retardation mea-surements with the Nerve Fiber Analyzer II. J Glaucoma 1997, 6:384–389.
49 Tjon-Fo-Sang MJH, van Strik R, de Vries J, et al.: Improved reproducibility ofmeasurements with the Nerve Fiber Analyzer. J Glaucoma 1997, 6:203–211.
50 Waldock A, Potts MJ, Sparrow JM, et al.: Clinical evaluation of scanning laserpolarimetry: Intraoperator reproducibility and design of a blood vessel re-moval algorithm. Br J Ophthalmol 1998, 82:252–259.
51 Weinreb RN, Zangwill L, Berry CC, et al.: Detection of glaucoma with scan-ning laser polarimetry. Arch Ophthalmol 1998, 116:1583–1589.
52 Garcia-Sanchez J, Garcia-Feijoo J, Arias-Puente A, et al.: Accuracy of theGDx system for the diagnosis of glaucoma. Invest Ophthalmol Vis Sci 1998,38:933.
53 Hoh ST, Greenfield DS, Mistlberger A, et al.: Optical coherence tomographyand scanning laser polarimetry in normal, ocular hypertensive, and glaucoma-tous eyes. Am J Ophthalmol 2000, 129:129–135.
54 Bowd CA, Zangwill LM, Berry CC, et al.: Detecting early glaucoma by as-sessment of retinal nerve fiber layer thickness and visual function. Invest Oph-thalmol Vis Sci 2001, 42:1993–2003.
55 Colen TP, Van Everdingen JAM, Lemij HG: Axonal loss in a patient with an-terior ischemic optic neuropathy as measured with scanning laser polarim-etry. Am J Ophthalmol 2000, 130:847–850.
56 Medeiros FA, Susanna R: Retinal nerve fiber layer loss after traumatic opticneuropathy detected by scanning laser polarimetry. Arch Ophthalmol 2001,119:920–921.
on assessment of retinal nerve fiber layer thickness by scanning laser polar-imetry. Am J Ophthalmol 2000, 129:715–722.
This report describes the effect of corneal birefringence upon RNFL thicknessdeterminations using the GDx nerve fiber analyzer and outlines the optical limita-tions of using a fixed corneal compensator to neutralize anterior segment polariza-tion. A novel method for estimating corneal birefringence is described using macu-lar birefringence characteristics.
60 Fercher AF, Hitzenberger CK, Drexler W, et al.: In vivo optical coherencetomography. Am J Ophthalmol 1993, 116:113–114.
61 Schuman JS, Pedut-Kloizman T, Hertzmark E, et al.: Reproducibility of nervefiber layer thickness measurements using optical coherence tomography.Ophthalmology 1996, 103:1889–1898.
62 Baumann M, Gentile RC, Liebmann JM, et al.: Reproducibility of retinal thick-ness measurements in normal eyes using optical coherence tomography.Ophthalmic Surg Lasers 1998, 29:280–285.
63 Blumenthal EZ, Williams JM, Weinreb RN, et al.: Reproducibility of nerve fiberlayer thickness measurements by use of optical coherence tomography. Oph-thalmology 2000, 107:2278–2282.
65 Mistlberger A, Liebmann JM, Greenfield DS, et al.: Heidelberg retina tomog-raphy and optical coherence tomography in normal, ocular hypertensive andglaucomatous eyes. Ophthalmology 1999, 106:2027–2032.
•66 Zangwill LM, Bowd C, Berry CC, et al.: Discriminating between normal and
glaucomatous eyes using the Heidelberg Retina Tomograph, GDx Nerve Fi-ber Analyzer, and Optical Coherence Tomograph. Arch Ophthalmol 2001,119:985–993.
This report provides a comparison of the ability of OCT, GDx, and HRT to discrimi-nate between healthy eyes and eyes with mild to moderate glaucoma and found nodifferences between the best parameter for each instrument using receiver opera-tor characteristic curves.
67 Kruse FE, Burk ROW, Volcker H-E, et al.: Reproducibility of topographic mea-surements of the optic nerve head with laser tomographic scanning. Ophthal-mology 1989, 96:1320–1324.
68 Choplin NT, Lundy DC, Dreher AW: Differentiating patients with glaucomafrom glaucoma suspects and normal subjects by nerve fiber layer assessmentwith scanning laser polarimetry. Ophthalmology 1998, 105:2068–2076.