FD159D Onyx ™ Liquid Embolic System Physician Orientation Review International
FD159D
Onyx™ Liquid Embolic System
Physician Orientation ReviewInternational
FD159D
Materials and Indications
• This presentation is intended for the purpose of didactic review with physicians regarding the use of the Onyx LD Embolic System– Model #’s 105-7000-060, 065, 080 (Onyx 18, 20, 34)
• Indications for Use: Embolization of lesions in the peripheral and neurovasculature, including arteriovenous malformations and hypervasculartumors.
FD159D
Onyx Development
• Ongoing research since 1993
• Human clinical studies commenced 1997
• Name change 1999: Embolyx to Onyx (LES)
• CE Mark received for AVM indication, July 1999– Regularly used since then in Europe
• Estimated worldwide procedures: 15,000+
FD159D
Onyx Liquid Embolic System
• Ethylene-vinyl alcohol copolymer– EVOH
• Dimethyl Sulfoxide Solvent– DMSO
• Micronized tantalum powder– Ta
FD159D
Onyx Embolization Process
• Onyx is a pre-mixed, radiopaque, injectable embolic fluid.
• Polymer precipitation occurs upon contact with aqueous solution.
• Contact with blood = “Precipitation”
• Solvent diffuses away
• Forms a spongy polymeric cast
• Forms a skin - solidifies from the outside in
• Liquid center continues to flow
FD159D
Characteristics
• Onyx delivers in a cohesive manner• Analogy of “lava” demonstrates characteristics of
behavior).
Onyx PrecipitateExternal Surface
Onyx PrecipitateInternal Surface
FD159D
Device Components
• 1.5 ml vial of Onyx– Onyx 18, 105-7000-060– Onyx 20, 105-7000-065– Onyx 34, 105-7000-080
• 1.5 ml vial of DMSO• MTI 1 ml luer-lock delivery syringes
– 2 white for Onyx, 1 yellow for DMSO
FD159D
Onyx Relative Viscosity
Onyx HD - Aneurysm Therapy
Onyx AVM v. HD for Aneurysms –
Entirely Different Viscosity
0
100
200
300
400
500
600
6% 6.50% 8% 20%Viscosity doubles between
Onyx 18 (6%) to Onyx 34 (8%)
Onyx LD - AVM Formulations
0
5
10
15
20
25
30
35
40
Water HumanBlood
25%nBCA
6% 6.5 8%
Scale = centistokes VS.
FD159D
DMSO (Dimethyl Sulfoxide Solvent)
FD159D
DMSO Study Review
• In ‘90 - ‘91 Taki and Terada successfully described embolization with EVAL dissolved in DMSO
• Safety concerns emerged
• Chaloupka and Viñuela conducted early UCLA study - showed severe vasospasm and angionecrosis
• Subsequent studies show safety of the Onyx delivery protocol– Dr. Chaloupka (1999)– UCLA (1998)
FD159D
Chaloupka et, al., UCLA, AJNR, June 1994
• 1st study: DMSO intra-arterial injections in volumes – 0.5 mL and 0.8 mL – Injected over a timeframe of 15 seconds or less
• Rete mirabile of 29 swine
• When injected over relatively rapid timeframes– DMSO was reported to cause moderate to severe
vasospasm– Sub arachnoid hemorrhage (SAH), stroke or death
FD159D
Murayama, (et al) UCLA, Neurosurgery, November 1998
• Embolyx E (now Onyx) and its solvent DMSO– 44 rete mirabile in 22 swine
• Study showed two important factors influencing vascular toxicity:– Contact time with arterial wall– Volume of DMSO
• Fast delivery (0.5 mL in 5-15 seconds) – Caused severe vasospasm and histological endothelial
necrosis• Slow delivery (0.5 mL in 30-120 seconds)
– Showed no vasospasm, minimal inflammation and no complications
FD159D
Chaloupka, (et al) AJNR, March 1999
• At MTI’s request, Dr. Chaloupka agreed to conduct a 2nd study reexamination DMSO angiotoxicity
• Purpose was to evaluate the effects of intra-arterial DMSO injections at slower rates of 30, 60 and 90 seconds
• Rete of 26 swine• No acute hemodynamic alterations• No infarction or SAH
FD159D
Chaloupka, (et al) AJNR, March 1999
• No significant changes in arterial blood pressure
• No alterations in heart rate or ECG
• Conclusion: – Found that lower dose rates of superselectively infused DMSO
are associated with a more acceptable safety and histotoxicity profile than previously reported
– Chaloupka confirmed Murayama’s results
FD159D
DMSO Overview
• Enters the bloodstream and is absorbed into tissues
• Metabolized to dimethyl sulfone (DMSO2) and dimethyl sulfide (DMS)– 80% of these metabolites are eliminated via the kidneys in
urine within a week
– Also eliminated via the skin or lungs which causes a garlic odor to the breath
• Complete elimination: 13-18 days
FD159D
Procedure Guidelines
Please refer to the Instructions for Use for full prescribing details
FD159D
Onyx Animation
Onyx AVM Animation available
FD159D
Shake Onyx Vials
• Place vials on mixer
• Set mixer to maximum setting
• Shake for a minimum of 20 minutes
• Maintain shaking of vials until ready to use
FD159D
Superselective Contrast Injection
• Confirm micro catheter placement
• Contrast agent per institutional procedure
FD159D
Flush Microcatheter
• Flush contrast from micro catheter
• IFU state 10ml of saline
FD159D
Prime Catheter with DMSO
• Aspirate 0.8 mL DMSO into the yellow MTI DMSO 1 mL syringe
• Inject DMSO to sufficiently fill dead space of delivery catheter– Marathon = 0.23mL – UltraFlow HPC = 0.26mL
FD159D
Prepare Onyx Delivery Syringe
• Remove prepared vial from shaker
• Aspirate 1 mL Onyx into the white MTI Onyx 1 mL syringe– Using an 18 or 20 gauge
needle
• Clear any air bubbles within the syringe and hub
FD159D
• When ready to deliver Onyx, remove DMSO syringe
• Hold the micro catheter hub in a vertical position
• Overfill hub of the catheter with the balance of DMSO
Overfill Hub with DMSO
FD159D
Onyx Delivery Syringe
• Connect Onyx syringe to the hub.
• For optimal fluoroscopic visualization, quickly point vertically to create an interface between DMSO and Onyx.
• While holding the syringe vertically, begin injection Onyx to displace DMSO.
• Inject Onyx at the recommended slow, steady rate of:0.25ml over 90 seconds, (approx. 0.1 mL/min)
IFU states: “Do not exceed a 0.3 mL/min”
FD159D
Continue Injecting Onyx
• Continue holding the syringe vertically until Onyx passes through the hub
• Once Onyx passes through the hub, the syringe may be held in a comfortable (horizontal) position
• Begin fluoroscopic imaging just prior to delivering catheter dead space:
– Marathon = 0.23mL – UltraFlow HPC = 0.26mL
FD159D
Detachment of Catheter
• After completion of Onyx injection, slightly aspirate syringe: Only 0.1ml is needed
• Gently, slowly and “incrementally” pull the catheter a few centimeters at a time.
• Tension will increase in catheter– Hold when sufficient tension is
reached– Gradually increase
• Catheter will release
FD159D
Onyx Preparation
1. Shake Vials 20 minutes
6. Displace DMSO 0.25ml / 90 secs
3. Load dead space with DMSO
4. Overwashcatheter hub
5. Connect Onyx Syringe vertically
2. Flush Catheter
7. Fluoro just prior to filling deadspace
FD159D
Onyx AVM
Technique
FD159D
“Plug and Push” Technique
• Create a “plug” around catheter tip
• Allow use of “Flow Arrest” to promote distal movement
• Use a “waiting technique” to create change in pressure
• Can help improve penetration
• Must balance with cautions and risks
FD159D
Hints for Plug and Push Technique
• Re- road map after each “injection” cycle (wait time)– This will help monitor deposition and movement of “new” material
• May be necessary to repeat cycle multiple times during an injection– When material first begins to reflux…– wait :30 seconds to break the motion…– upon re-injection again the material should move distally
– As a full circumferential plug forms…– wait up to 2:00min to solidify flow arrest…
• When beginning injection… SLOW gentle injection will help control new deposition
FD159D
Technique Decision Tree
Begin Slow OnyxInjection
Reflux (laminar/plug) ?
Pause
:30 SecondsMomentum Break
2:00 MinutesSolidify Plug
ContinueConstant Slow Rate
Forward Penetration
Re-InjectRepeat Process
FD159D
Potential Risks
• Using Plug and Push can create conditions that should be monitored closely.
• Waiting time– Catheter occlusion/pressure build up
• Possible reflux– Catheter Removal
FD159D
Fundamentals of AVM Embolization…
Imaging/Working Projection
Microcatheter Position
Microcatheter RetrievalPedicle Selection Size
Contribution / Flow
Onyx Technique
Are Critical to Onyx PerformanceAre Critical to Onyx Performance
Injection Speed Delivery Rate
Waiting Time / Pressures
FD159D
Imaging/Working Projection
• Always work with Reference Image and Working Projection
• A reference image becomes very useful during long Onyx injections – Catheter distal shaft – Nidal Angioarchitecture– Primary venous drainage
• Optimize working projection: – Separation of catheter tip, minimal foreshortening of catheter – Allows to recognize, measure and manage reflux– Instructions for Use recommends a maximum of 1cm reflux– Modifications based on anatomy
• Distal turn or bend in catheter tip• Select a point on catheter (a turn or bend) to use as a reference and
reflux limit
FD159D
Reference Images
FD159D
Feeder Selection
• Larger contributing feeders seem to promote penetration:
– Generally easier to reach nidal position– Generally provide a greater pressure
gradient– Therefore can capitalize on Onyx
characteristics for better penetration.
FD159D
Feeder Selection
• Small pedicles can have limited penetration
– Especially if proximal – because of change in pressure as feeder fills… can cause limited nidal penetration.
– Must inject very slowly, control reflux, and expect possibly limited nidal delivery from small feeders.
– Risk of delayed catheter retrieval can occur… because of potential spasm / reflux build up quickly.
FD159D
Microcatheter Position
• Intra-nidal (or wedged) position seems optimal– Or as close as possible to nidus – Especially because Onyx performs best when “pushed” into the
lesion
• Proximal positions:– May result in minimal nidal deposition and a “feeder” injection
– As Onyx is delivered into feeder, the pressure gradient changes.As feeder closes, the path of least resistance becomes retrograde reflux occurs.
– “Plug and Push” may not achieve same results as nidal position
FD159D
Injection Rate
• The Slower… The Better – Better Control and Better
penetration overall– More penetration before
reflux and more control when reflux occurs
• Rate must be constant and slow– Any change in injection rate
will create reflux– If rate is too fast, reflux will
occur quickly
• Using the “DMSO Displacement” rate as a reference– 0.25ml over 90 seconds
= 0.16ml / minute – Continue this rate (or
slower)– Case examples: 0.10 -
0.15 ml / minute
FD159D
Consider all the Consider all the factors that can factors that can affect affect difficult difficult catheter removal:catheter removal:
Length of Reflux
Distal 2cm (loop, hook)
Potential for Spasm
Time of Injection
Size of Feeder
Tortuosity
Difficult CatheterRetrieval
••Evaluating these parameters prior to injection can determine howEvaluating these parameters prior to injection can determine how much reflux will be acceptedmuch reflux will be accepted
Catheter Retrieval
FD159D
Retrieval Techniques: Two Paths
Two Techniques ProvideVersatile, Effective Use
Slow “Traction”Technique
Quick “Wrist Snap”Technique
FD159D
Slow “Traction” Technique
• Slowly, and incrementally (cm by cm), withdraw the catheter
• Tension will slowly increase creating retrieval force directly to the catheter tip
• Sustain “moderate” tension (hold for a moment)
• Catheter tip will “release” from Onyx cast
• Completely withdrawal catheter
FD159D
Quick “Wrist Snap” Technique
• Remove all “slack” from catheter. Withdraw catheter 3 – 5 cm.
• This will create a slight tension throughout the catheter
• Quickly move wrist from left to right (about 10 – 20cm)– No need to move the entire arm– This is NOT like a glue technique
• This provides a “spark” of energy through the catheter– Without stretching the catheter too much and risking breakage
FD159D
Step #2: Assess risk of spasm or vessel
damage
Is the catheter tip beginning to release
from Onyx cast?``
Is the vasculature straightening ?
Is the Onyx cast retracting?
Step #2a: Sustain Traction and Monitor Tip Release
Step #1: Apply gentle, incremental traction to catheter (cm by cm)
Step #3: Release Traction and Repeat Decision Tree
Applying Intermittent Traction As Necessary
Catheter Retrieval Management
FD159D
Quick Wrist Snap - Steps
Create slight tension
•Remove all “slack” from the catheter –•By pulling a few cm.
FD159D
Quick Wrist Snap - Steps
NOTE: It is not necessary to move the entire arm (such as a glue technique). This technique can risk catheter separation.
0cm 20cm
Quick but Short
Wrist Motion
•Quickly move wrist 10 – 15cm to the right•Making a “wrist snap” motion
FD159D
Onyx Delivery Microcatheters
FD159D
Onyx Delivery Systems
Rebar
Mirage
X-pedion, SilverSpeed
Guidewires
Marathon & Ultraflow Echelon 10 and 14
FD159D
Onyx Catheters
Onyx Delivery Catheter
Flow Directed CatheterOver The Wire
Benefits• Reinforced• High Tensile Strength• Burst Profile
Uses• Coils• Liquid Embolics• Particles
UltraFlow
Marathon
Apollo, 2009
Benefits • Soft Supple Tip• No Guidewire• Superior Navigation
UsesAVM embo (nBCA)
• Soft Supple Tip• Superior Navigation• High Tensile Strength• Burst Pressure
FD159D
Embolic Injection Pressure v. Burst
• Onyx injection requires very low pressure of approximately 20psi injected at a rate of (0.1 ml / min)
• Marathon provides a robust tolerance between Onyx injection pressure and Onyx burst pressure and is more than 2x in UltraFlow
Source: TR 03-106, UltraFlow TR01-116Test Method: TM0090, Onyx Plug & Push: 1 hour Onyx injections, repeated pressurization to intentional failure.Engineering Note: Test results show that when catheter is intentionally ruptured, the UltraFlow fails between the fuse joint and distal tip, while Marathon fails in the distal
7cm.
psi
0
50
100
150
200
250
300
350
400
Onyx 18Injection
UltraFlow Marathon
0.1 ml / min
18x tolerance
FD159D
Pressure to Re-inject v. Wait Time
0
20
40
60
80
100
120
140
160
0 30 secs 1min 1.5min 2min 3min 5min
PSI
Wait Time
• Recommendation: Limit “Wait” time less than 2 min.– Onyx mass can precipitate in 3 - 5 minutes.– The increased injection pressure (after a long wait) can approach this limit.
FD159D
Summary: Burst Performance
• Recognize occluded catheter if Onyx is not exiting catheter tip – Occlusion may occur after extended wait period during Onyx injection
• Do not interrupt Onyx injection more than 2 minutes
• If Onyx is not exiting tip, do not inject more than 0.05 ml into a potentially occluded catheter.
• Static Pressure to burst the UltraFlow or Marathon should be “recognizable”to identify resistance.
– If resistance is felt, stop Onyx injection.
FD159D
Detachment Force – from Onyx Cast
• Marathon design requires 25% less force to retrieve from a cast of Onyx (see test method) than UltraFlow
Source:, Marathon TR 03-106, UltraFlow TR01-116Test Method: TM 0090: Flow model, 3cm Onyx reflux, 1 hour dwell time, at body temperature, slow pull technique
0
10
20
30
40
50
60
70
UltraFlow Marathon
Retrieval Force
Gra
ms
FD159D
Distal Tensile Strength
0
100
200
300
400
500
600
Magic 1.2 Magic 1.5 Spinnaker UltraFlow Marathon 1.3
Distal Tensile Strength
• Marathon braiding provides significant tensile strength for excellent catheter retrieval
Source: UltraFlow TR01-116, Marathon TR 03-106, Other Catheters: TR02-088Test Method: TM0080 – test a 1” distal segment gauge length, stretch to break at 20” / min
Gra
ms
130%130%
71%
FD159D
FD Catheter Stretch Profile
• Recommended Traction Limit should be < 20cm– Note: If catheter is trapped at distal tip - recommendation offers safety
margin to minimize risk of catheter separation.
0
100
200
300
400
500
600
1cm 3cm 5cm 10cm 15cm 20cm 43cm
Recommended Traction Limit < 20cm
Distance of catheter ‘traction’ (cm)
Stretch Point to Breakage
FD159D
Access Summary
• Tip shaping: – It is not necessary to “overshape”
• Guidewire technique: – Some clinicians recommend keeping the guidewire placed just past the
distal tip marker during navigation. This helps maintain a patent lumen and minimize risk of catheter prolapse
• Angiographic technique – Some physicians suggest using a gentle, low volume “puff” of contrast to
confirm catheter patency.
• Less Magnification / Larger Field of View– Prior to embolic delivery, it is suggested to deliver a small volume contrast
injection – While viewing the entire distal catheter segment to verify catheter integrity.
FD159D
Onyx™ Liquid Embolic System
Physician Orientation ReviewInternational
CAUTION: Federal (USA) law restricts this device to sale by or on the order of a physician. A complete statement of indications, contraindications, warnings and instructions can be found in the product labeling supplied with each device. WARNINGS: Serious, including fatal, consequences could result with the use of the Onyx LES without adequate training. Contact your ev3 Sales Representative for information on training courses. Onyx is a registered trademark of ev3.