SELECT TEST: AC{ REPORT NAME: ACP DATA MODE: ON BOARD TEST NAME: ACP CTRL INTERVAL: 1000 UNIT: U/L INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: SER/PL
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 0.0-9.0 - (2) 0.0 100 YEARS - - (3) 0.0
- - (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 0.0-35.0 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50. Rev.8-03
Pointe Scientific, Inc. Analyzer: Hitachi 912 Instrument Application Test: ALBUMIN
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: ALB REPORT NAME: ALBUMIN DATA MODE: ON BOARD TEST NAME: ALB CTRL INTERVAL: 1000 UNIT: G/DL INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: SER/PL
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 3.5-5.3 3.5-5.3 (2) 0.0 100 YEARS 3.5-5.3 3.5-5.3 (3) 0.0
3.5-5.3 3.5-5.3 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 0.5-6.0 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50. Rev.8-03
Pointe Scientific, Inc. Analyzer: Hitachi 912 Instrument Application Test: ALP
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: ALP REPORT NAME: ALP DATA MODE: ON BOARD TEST NAME: ALP CTRL INTERVAL: 1000 UNIT: U/L INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: SER/PL
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 35-123 35-123 (2) 0.0 100 YEARS 35-123 35-123 (3) 0.0
35-123 35-123 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 0-1000 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50. Rev.8-03
Pointe Scientific, Inc. Analyzer: Hitachi 912 Instrument Application Test: ALT
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: ALT REPORT NAME: ALT DATA MODE: ON BOARD TEST NAME: ALT CTRL INTERVAL: 1000 UNIT: U/L INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: SER/PL
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 4-36 4-36 (2) 0.0 100 YEARS 4-36 4-36 (3) 0.0
4-36 4-36 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 0-700 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50. Rev.8-03
Pointe Scientific, Inc. Analyzer: Hitachi 912 Instrument Application Test: AMMONIA
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: AMMON REPORT NAME: AMMONIA DATA MODE: ON BOARD TEST NAME: AMMON CTRL INTERVAL: 1000 UNIT: UMOL/L INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: SER/PL
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 11-35 11-35 (2) 0.0 100 YEARS 11-35 11-35 (3) 0.0
11-35 11-35 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 0-600 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Rev.8-03
Pointe Scientific, Inc. Analyzer: Hitachi 912 Instrument Application Test: AMYLASE
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: AMY REPORT NAME: AMYLASE DATA MODE: ON BOARD TEST NAME: AMY CTRL INTERVAL: 1000 UNIT: U/L INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: SER/PL
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 25-125 25-125 (2) 0.0 100 YEARS 25-125 25-125 (3) 0.0
25-125 25-125 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 0-2000 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50. Rev.8-03
Pointe Scientific, Inc. Analyzer: Hitachi 912 Instrument Application Test: ASO
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: ASO REPORT NAME: ASO DATA MODE: ON BOARD TEST NAME: ASO CTRL INTERVAL: 1000 UNIT: U/ML INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: SER/PL
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 0-156 0-156 (2) 0.0 100 YEARS 0-156 0-156 (3) 0.0
0-156 0-156 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 50-1000 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50. Rev.8-03
Pointe Scientific, Inc. Analyzer: Hitachi 912 Instrument Application Test: AST
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: AST REPORT NAME: AST DATA MODE: ON BOARD TEST NAME: AST CTRL INTERVAL: 1000 UNIT: U/L INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: SER/PL
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 5-34 5-34 (2) 0.0 100 YEARS 5-34 5-34 (3) 0.0
5-34 5-34 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 0-700 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50. Rev. 8-03
Pointe Scientific, Inc. Analyzer: Hitachi 912 Instrument Application Test: DIRECT BILIRUBIN
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: DBIL REPORT NAME: DIRECT BILIRUBIN DATA MODE: ON BOARD TEST NAME: DBIL CTRL INTERVAL: 1000 UNIT: MG/DL INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: SER/PL
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 0.0-0.5 0.0-0.5 (2) 0.0 100 YEARS 0.0-0.5 0.0-0.5 (3) 0.0
0.0-0.5 0.0-0.5 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 0.0-20.0 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50. Rev.8-03
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: UN REPORT NAME: UREA NITROGEN DATA MODE: ON BOARD TEST NAME: UN CTRL INTERVAL: 1000 UNIT: MG/DL INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: SER/PL
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 7-18 7-18 (2) 0.0 100 YEARS 7-18 7-18 (3) 0.0
7-18 7-18 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 0.0-150.0 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50. Rev.8-03
Pointe Scientific, Inc. Analyzer: Hitachi 912 Instrument Application Test: TOTAL BILIRUBIN
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: TBIL REPORT NAME: TOTAL BILIRUBIN DATA MODE: ON BOARD TEST NAME: TBIL CTRL INTERVAL: 1000 UNIT: MG/DL INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: SER/PL
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 0.2-1.2 0.2-1.2 (2) 0.0 100 YEARS 0.2-1.2 0.2-1.2 (3) 0.0
0.2-1.2 0.2-1.2 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 0.0-20.0 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50. Rev.8-03
Pointe Scientific, Inc. Analyzer: Hitachi 912 Instrument Application Test: CALCIUM
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: CA REPORT NAME: CALCIUM DATA MODE: ON BOARD TEST NAME: CA CTRL INTERVAL: 1000 UNIT: MG/DL INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: SER/PL
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 8.5-10.4 8.5-10.4 (2) 0.0 100 YEARS 8.5-10.4 8.5-10.4 (3) 0.0
8.5-10.4 8.5-10.4 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 0.0-15.0 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50. Rev.8-03
Pointe Scientific, Inc. Analyzer: Hitachi 912 Instrument Application Test: CHOLESTEROL
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: CHOL REPORT NAME: CHOLESTEROL DATA MODE: ON BOARD TEST NAME: CHOL CTRL INTERVAL: 1000 UNIT: MG/DL INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: SER/PL
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 0-200 0-200 (2) 0.0 100 YEARS 0-200 0-200 (3) 0.0
0-200 0-200 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 0-700 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50. Rev.8-03
Pointe Scientific, Inc. Analyzer: Hitachi 912 Instrument Application Test: CO2
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: CO2 REPORT NAME: CO2 DATA MODE: ON BOARD TEST NAME: CO2 CTRL INTERVAL: 1000 UNIT: MEQ/L INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: SER/PL
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 23-34 23-34 (2) 0.0 100 YEARS 23-34 23-34 (3) 0.0
23-34 23-34 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 0-40 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50. Rev.8-03
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: CK REPORT NAME: CREATINE KINASE DATA MODE: ON BOARD TEST NAME: CK CTRL INTERVAL: 1000 UNIT: U/L INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: SER/PL
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 0-160 0-130 (2) 0.0 100 YEARS 0-160 0-130 (3) 0.0
0-160 0-130 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 0-1500 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50. Rev.8-03
Pointe Scientific, Inc. Analyzer: Hitachi 912 Instrument Application Test: CREATININE
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: CREA REPORT NAME: CREATININE DATA MODE: ON BOARD TEST NAME: CREA CTRL INTERVAL: 1000 UNIT: MG/DL INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: SER/PL
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 0.4-1.4 0.4-1.4 (2) 0.0 100 YEARS 0.4-1.4 0.4-1.4 (3) 0.0
0.4-1.4 0.4-1.4 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 0.0-15.0 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50. Rev.8-03
Pointe Scientific, Inc. Analyzer: Hitachi 912 Instrument Application Test: CRP
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: CRP REPORT NAME: CRP DATA MODE: ON BOARD TEST NAME: CRP CTRL INTERVAL: 1000 UNIT: MG/DL INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: SER/PL
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 0-0.6 0-0.6 (2) 0.0 100 YEARS 0-0.6 0-0.6 (3) 0.0
0-0.6 0-0.6 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 0.1-25 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Rev.8-03
Pointe Scientific, Inc. Analyzer: Hitachi 912 Instrument Application Test: GGT
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: GGT REPORT NAME: GGT DATA MODE: ON BOARD TEST NAME: GGT CTRL INTERVAL: 1000 UNIT: U/L INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: SER/PL
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 9-54 8-35 (2) 0.0 100 YEARS 9-54 8-35 (3) 0.0
9-54 8-35 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 0-1000 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50. Rev.8-03
Pointe Scientific, Inc. Analyzer: Hitachi 912 Instrument Application Test: GLUCOSE
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: GLU REPORT NAME: GLUCOSE DATA MODE: ON BOARD TEST NAME: GLU CTRL INTERVAL: 1000 UNIT: MG/DL INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: SER/PL
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 65-110 65-110 (2) 0.0 100 YEARS 65-110 65-110 (3) 0.0
65-110 65-110 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 0-500 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50. Rev.8-03
Pointe Scientific, Inc. Analyzer: Hitachi 912 Instrument Application Test: GLUCOSE
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: GLU REPORT NAME: GLUCOSE DATA MODE: ON BOARD TEST NAME: GLU CTRL INTERVAL: 1000 UNIT: MG/DL INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: SER/PL
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 65-110 65-110 (2) 0.0 100 YEARS 65-110 65-110 (3) 0.0
65-110 65-110 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 0-500 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50. Rev.8-03
Pointe Scientific, Inc. Analyzer: Hitachi 912 Instrument Application Test: TOTAL IRON
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: IRON REPORT NAME: TOTAL IRON DATA MODE: ON BOARD TEST NAME: IRON CTRL INTERVAL: 1000 UNIT: UG/DL INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: SER/PL
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 60-150 60-150 (2) 0.0 100 YEARS 60-150 60-150 (3) 0.0
60-150 60-150 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 0-700 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50. Rev.8-03
Pointe Scientific, Inc. Analyzer: Hitachi 912 Instrument Application Test: LACTATE
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: LAC REPORT NAME: LACTATE DATA MODE: ON BOARD TEST NAME: LAC CTRL INTERVAL: 1000 UNIT: MMOL/L INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: SER/PL
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 0.5-2.2 0.5-2.2 (2) 0.0 100 YEARS 0.5-2.2 0.5-2.2 (3) 0.0
0.5-2.2 0.5-2.2 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 0-20 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50. Rev.8-03
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: LDL REPORT NAME: LDL CHOLESTEROL DATA MODE: ON BOARD TEST NAME: LDL CTRL INTERVAL: 1000 UNIT: MG/DL INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: SER/PL
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 0-130 0-130 (2) 0.0 100 YEARS 0-130 0-130 (3) 0.0
0-130 0-130 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 0-700 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50. Rev.8-03
Pointe Scientific, Inc. Analyzer: Hitachi 912 Instrument Application Test: LDH
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: LDH REPORT NAME: LDH DATA MODE: ON BOARD TEST NAME: LDH CTRL INTERVAL: 1000 UNIT: U/L INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: SER/PL
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 80-285 103-227 (2) 0.0 100 YEARS 80-285 103-227 (3) 0.0
80-285 103-227 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 0-1000 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50. Rev.8-03
Pointe Scientific, Inc. Analyzer: Hitachi 912 Instrument Application Test: LIPASE
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: LIP REPORT NAME: LIPASE DATA MODE: ON BOARD TEST NAME: LIP CTRL INTERVAL: 1000 UNIT: U/L INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: SER/PL
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 0-62 0-62 (2) 0.0 100 YEARS 0-62 0-62 (3) 0.0
0-62 0-62 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 0-650 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50. Rev.8-03
Pointe Scientific, Inc. Analyzer: Hitachi 912 Instrument Application Test: MAGNESIUM
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: MAG REPORT NAME: MAGNESIUM DATA MODE: ON BOARD TEST NAME: MAG CTRL INTERVAL: 1000 UNIT: MEQ/L INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: SER/PL
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 1.3-2.5 1.3-2.5 (2) 0.0 100 YEARS 1.3-2.5 1.3-2.5 (3) 0.0
1.3-2.5 1.3-2.5 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 0.0-4.0 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50. Rev.8-03
Pointe Scientific, Inc. Analyzer: Hitachi 912 Instrument Application Test: MICROALBUMIN
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: u-ALB REPORT NAME: MICROALBUMIN DATA MODE: ON BOARD TEST NAME: u-ALB CTRL INTERVAL: 1000 UNIT: MG/DL INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: URINE
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 0.0-2.1 0.0-2.1 (2) 0.0 100 YEARS 0.0-2.1 0.0-2.1 (3) 0.0
0.0-2.1 0.0-2.1 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 0.5-30.0 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Rev.8-03
Pointe Scientific, Inc. Analyzer: Hitachi 912 Instrument Application Test: MICROPROTEIN
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: MIPRO REPORT NAME: MICROPROTEIN DATA MODE: ON BOARD TEST NAME: MIPRO CTRL INTERVAL: 1000 UNIT: MG/DAY INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: URINE
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 28-141 28-141 (2) 0.0 100 YEARS 28-141 28-141 (3) 0.0
28-141 28-141 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 0-250 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Rev.8-03
Pointe Scientific, Inc. Analyzer: Hitachi 912 Instrument Application Test: PHOSPHORUS
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: PHOS REPORT NAME: PHOSPHORUS DATA MODE: ON BOARD TEST NAME: PHOS CTRL INTERVAL: 1000 UNIT: MG/DL INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: SER/PL
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 2.5-4.8 2.5-4.8 (2) 0.0 100 YEARS 2.5-4.8 2.5-4.8 (3) 0.0
2.5-4.8 2.5-4.8 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 0.0-12.0 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50. Rev.8-03
Pointe Scientific, Inc. Analyzer: Hitachi 912 Instrument Application Test: RF
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: RF REPORT NAME: RF DATA MODE: ON BOARD TEST NAME: RF CTRL INTERVAL: 1000 UNIT: U/ML INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: SER/PL
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 0-10 0-10 (2) 0.0 100 YEARS 0-10 0-10 (3) 0.0
0-10 0-10 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 0-300 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Rev.8-03
Pointe Scientific, Inc. Analyzer: Hitachi 912 Instrument Application Test: TOTAL PROTEIN
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: TP REPORT NAME: TOTAL PROTEIN DATA MODE: ON BOARD TEST NAME: TP CTRL INTERVAL: 1000 UNIT: G/DL INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: SER/PL
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 6.2-8.5 6.2-8.5 (2) 0.0 100 YEARS 6.2-8.5 6.2-8.5 (3) 0.0
6.2-8.5 6.2-8.5 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 1.0-15.0 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50. Rev.8-03
Pointe Scientific, Inc. Analyzer: Hitachi 912 Instrument Application Test: TRIGLYCERIDES
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: TRIG REPORT NAME: TRIGLYCERIDES DATA MODE: ON BOARD TEST NAME: TRIG CTRL INTERVAL: 1000 UNIT: MG/DL INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: SER/PL
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 44-148 44-148 (2) 0.0 100 YEARS 44-148 44-148 (3) 0.0
44-148 44-148 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 0-1000 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50. Rev.8-03
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: UA REPORT NAME: URIC ACID DATA MODE: ON BOARD TEST NAME: UA CTRL INTERVAL: 1000 UNIT: MG/DL INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: SER/PL
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 2.5-7.7 2.5-7.7 (2) 0.0 100 YEARS 2.5-7.7 2.5-7.7 (3) 0.0
2.5-7.7 2.5-7.7 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 0.0-20.0 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50. Rev.8-03
Pointe Scientific, Inc. Analyzer: Hitachi 912 Instrument Application Test: autoHDL
----------------------------------------------------APPLICATION SCREEN, RANGE WINDOW-------------------------------------------------------------------
SELECT TEST: HDL REPORT NAME: HDL DATA MODE: ON BOARD TEST NAME: HDL CTRL INTERVAL: 1000 UNIT: MG/DL INSTR. FACTOR: A = 1.0 B = 0.0 SELECT SAMPLE TYPE: SER/PL
EXPECTED RANGES: USE QUALITATIVE TABLE AGE MALE FEMALE (1) 2.999 <3.0 100 YEARS 35-55 45-65 (2) 0.0 100 YEARS 35-55 45-65 (3) 0.0
35-55 45-65 (4) 0.0 (5) 0.0
DEFAULT RANGE: OVER 100 YEARS (6) 0.0 DEFAULT SEX: MALE REPEAT RANGE: # - # CLASS 1 TECHNICAL RANGE: 0-150 CLASS 2 TECHNICAL RANGE: -99999 -99999
# DENOTES A USER-DEFINED PARAMETER. INFORMATION IS SPECIFIC TO THE CHANNEL BEING USED, OR IT IS DETERMINED BY THE LABORATORY’S OWN PREFERENCE FOR OPERATION. It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50. Rev.8-03