WarrantyDefects occurring within 3 years from delivery date shall beremedied free of charge at our plant (carriage and insurancepaid by sender).Sensors, fittings, and accessories: 1 year.
Subject to change without notice.
KnickElekt ronische MessgerteGmbH & Co. KG
P.O. Box 37 04 15D-14134 Berlin
Phone: +49 (0)30 - 801 91 - 0Fax: +49 (0)30 - 801 91 - 200Internet: http://[email protected]
3
Safet y inf ormat ion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6Trademarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
EC Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . .7
Overview o f Strat os e 2402 CondI . . . . . . . . . . . . . . . . . . .9Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Package contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10Mounting plan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11Pipe mounting, panel mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12Information on installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14Terminal assignments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
Installat ion and connection . . . . . . . . . . . . . . . . . . . . . . .14Preparing the shield connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16Typical wiring SE 655 / SE 656 sensor . . . . . . . . . . . . . . . . . . . . . . . . . . .17Typical wiring SE 652 sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18Typical wiring SE 654 sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
Sensors from other manufacturers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19Protective wiring of relay outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
User int erface and display . . . . . . . . . . . . . . . . . . . . . . . .22Operation: Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
Safety f eat ures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25Sensocheck, Sensoface sensor monitoring . . . . . . . . . . . . . . . . . . . . . . . .25GainCheck device self test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25Automatic device self-test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25Hold mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26External activation of Hold mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
Configu rat ion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28Menu structure of configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
Overview of configuration steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30Output 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32Output 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48Temperature compensation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54Alarm settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56Limit function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .58Controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .62Controlling a rinsing probe or signaling parameter set 2 . . . . . . . . . . . .64
Contents
8/9/2019 2402 condi
3/65
4 Stratos e 2402 CondI
Selecting paramet er set 1/2 . . . . . . . . . . . . . . . . . . . . . . .66Manually or automatically via Control . . . . . . . . . . . . . . . . . . . . . . . . . .66External switchover of parameter sets . . . . . . . . . . . . . . . . . . . . . . . . . . .67Default settings of parameter sets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69Parameter set, individual settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70
Calibrat ion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .72
Calibration by input of cell factor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .74Calibration with calibration solution . . . . . . . . . . . . . . . . . . . . . . . . . . . .76Product calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .78Zero calibration in air . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .80Zero calibration with calibration solution . . . . . . . . . . . . . . . . . . . . . . . . .82Adjusting temp probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .84
Diagnostics functions . . . . . . . . . . . . . . . . . . . . . . . . . . . .85Display of output currents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .85Display of calibration data (Cal Info) . . . . . . . . . . . . . . . . . . . . . . . . . . . .85Sensor monitor for validation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .85Display of last error message (Error Info) . . . . . . . . . . . . . . . . . . . . . . . . .85Specify current at output 1 (current source 1) . . . . . . . . . . . . . . . . . . . . .86Specify current at output 2 (current source 2) . . . . . . . . . . . . . . . . . . . . .86
Relay test (manual test of contacts) . . . . . . . . . . . . . . . . . . . . . . . . . . . . .86Controller test (manual specification of controller output) . . . . . . . . . . . .87
Cont roller functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .86PID controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .88Pulse length / pulse frequency control ler . . . . . . . . . . . . . . . . . . . . . . . .90
Error m essages (error codes) . . . . . . . . . . . . . . . . . . . . . .92Operat ing sta tes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .94Sensoface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .96Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .99
Product line and accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .99Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .100
Explosion protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .106Warnings and notes to ensure safe operating . . . . . . . . . . . . . . . . . . . .107Control Drawing CSA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .108Calibration solutions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .112Concentration measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .114Concentration curves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .115
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .120
5
Safety informat ion
Be sure to read and observe the following instruc-tions!
The device has been designed in accordance with the state ofthe art and complying with the applicable safety regulations.When operating the device, certain conditions may neverthe-less lead to danger for the operator or damage to the device.
Caution!Commissioning may only be carried out by trained experts.Whenever it is likely that protection has been impaired, thedevice shall be made inoperative and secured against unintend-
ed operation.
The protection is likely to be impaired if, for example: the device shows visible damage the device fails to perform the intended measurements after prolonged storage at temperatures above 70 C after severe transport stresses
Before recommissioning the device, a professional routine testin accordance with EN 61010-1 must be performed. This test
should be carried out by the manufacturer.
Caution!Before commissioning it must be proved that the device maybe connected with other equipment.
8/9/2019 2402 condi
4/65
8/9/2019 2402 condi
5/65
8 Stratos e 2402 CondI 9
Overview of Stratos e 2402 CondI
1
2
3
4
5
C
D
E
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
CondIinput
Hold
input
Control
input
Output 1
Output 2
Alarm
Clean
PSet2
Power
receive hi
receive lo
drain
send lo
send hi
shield
RTD
RTD
hold
hold/control
control
+ output 1
- output 1/2
+ output 2
relay 1
relay 1/2
relay 2
alarm
alarm
clean
PSet2
power
power
R1
R2
Tempinput
cleanPset2
8/9/2019 2402 condi
6/65
10 Stratos e 2402 CondI
Package contentsCheck the shipment for transport damage and completeness.The package should contain: Front unit of Stratos Lower case
Bag containing small parts Instruction manual Specific test report
Assembly
11
10
9
8
7 6 5 4
1
2
3
6 Sealing inserts (1 piece)
7 Rubber reducer (1 piece)8 Cable glands (3 pieces)9 Filler plugs (3 pieces)10 Hexagon nuts (5 pieces)11 Sealing plugs (2 pieces),
for sealing in case of wall mounting
1 Jumper (2 piece)
2 Washer (1 piece), for conduitmounting: place washer betweenenclosure and nut
3 Cable ties (3 pieces)4 Hinge pin (1 piece), insertable from
either side5 Enclosure screws (4 pieces)
Fig.: Assembling the enclosure
11
M ounting plan
144
144
15
42
84
8032
21
43
105
27
72
6,
2
1 2
3
4
1 Cable gland (3 pieces)2 Breakthroughs for cable gland
or conduit 1/2 , 21.5 mm (2 breakthroughs)Conduits not included!
3 Breakthroughs for pipe mounting(4 breakthroughs)
4 Breakthroughs for wall mounting(2 breakthroughs)
Fig.: Mounting plan
8/9/2019 2402 condi
7/65
12 Stratos e 2402 CondI
40 60132
1
2
3
4
5
Fig.: ZU 0274 pipe-mount kit
1 ZU 0276 protective hood (if required)2 Hose clamps with worm gear drive to DIN 3017 (2 pieces)3 Pipe-mount plate (1 piece)4 For vertical or horizontal posts or pipes5 Self-tapping screws (4 pieces)
1
132165
173
1
Fig.: ZU 0276 protective hood for wall and pipe mounting
Pipe mount ing, panel mounting
13
1
2
3
45
max. 25
78 27
1...22
1 Screws (4 pieces)2 Gasket (1 piece)3 Panel4 Span pieces (4 pieces)5 Threaded sleeves (4 pieces)
Fig.: ZU 0275 panel-mount kit
8/9/2019 2402 condi
8/65
8/9/2019 2402 condi
9/65
8/9/2019 2402 condi
10/65
18 Stratos e 2402 CondI
Typical w iringSE 652 sensor
receivehi
receivelo
drain
shield
RTD
15
wt
16
bk(rcv)
14
bk
12
bu
11
rd
sendlo
sendhi
2402 CondI
Caution!Place jumper across terminals 3 and 4!
RTD
17
bk
13
bn(snd)
jumper!
19
Typical w iringSE 654 X sensor
receivehi
receivelo
drain
shield
RTD
r
d
t
ransparent
t
ransparent
w
t
b
n
b
u
sendlo
sendhi
2402 CondI
Sensors of ot her m anufacturers:For special applications (chemical resistance, type of mounting), youcan also connect sensors from other manufacturers.Permissible ranges for the Stratos e 2402 CondI as well as terminalassignments and factory settings for these sensors are available onrequest.
RTD
b
k
coaxwt
coaxbk
8/9/2019 2402 condi
11/65
20 Stratos e 2402 CondI
Protective w iring of switching contactsRelay contacts are subjected to electrical erosion. Especiallywith inductive and capacitive loads, the service life of the con-tacts will be reduced. For suppression of sparks and arcing,components such as RC combinations, nonlinear resistors,
series resistors and diodes should be used.
1
23
1 2
3
Typical AC applicationswit h inductive load
1 Load2 RC combination, e.g. RIFA PMR209
Typical RC combinationsfor 230 V AC:Capacitor 0.1F/ 630V,Resistor 100 Ohms / 1 W
3 Contact
Protective wiring of switching outputs
21
A: DC application with inductive load
B: AC/DC applications with capacitive loadC: Connection of incandescent lamps
A1 Inductive loadA2 Free-wheeling diode, e.g. 1N4007
(Observe polarity)A3 ContactB1 Capacitive loadB2 Resistor, e.g. 8 Ohms/1 W at 24 V / 0.3 AB3 Contact
C1 Incandescent lamp, max 60 W / 230 V, 30 W / 115 VC3 Contact
Warning!Make sure that the maximum ratings of the relaycontacts are not exceeded even during switching!
Typical protective w iring m easures
8/9/2019 2402 condi
12/65
8/9/2019 2402 condi
13/65
24 Stratos e 2402 CondI
Operation: Keypad
Start, end calibration
Start, end configuration
Select digit position
(selected position flashes)Edit digit
Calibration:Continue in program sequence
Configuration: Confirm entries,next configuration step
Measuring mode: Display output current
Cal Info: Display of cell factor and zero point
Error Info: Display of last error message
Start GainCheck device self-test+
25
Sensocheck, Sensoface sensor m onitoringSensocheck cont inuously monitors the sensor and its wiring.Sensocheck can be switched of f (Configuration, Pg 57).
Sensoface provides information on the conductivitysensor condition. The primary coil and its wirings arecontinuously monitored for short circuits, the secondarycoil and its wirings are checked for open circuits.The three Sensoface indicators inform of the sensorcondition.
GainCheck device self testA display test is carried out, the software version is displayed andthe memory and measured value transfer are checked.
Start GainCheck device self-test:
Automatic device self-testThe automatic device self-test checks the memory and meas-ured-value transfer. It runs automatically in the background atfixed intervals.
+
Safet y funct ions
8/9/2019 2402 condi
14/65
8/9/2019 2402 condi
15/65
28 Stratos e 2402 CondI
The configuration parameters arechecked during the input. In thecase of an incorrect input Err isdisplayed for approx. 3 s. Theincorrect parameters cannot bestored. Input must be repeated.
ConfigurationIn the Configuration mode you set the device parameters. The Stratos canstore two different parameter sets and switch between them. Sensor dataand Clean/Pset2 output are edited in parameter set 1 only. They are validfor both parameter sets.
Press conf.
Enter mode code 1200 :
Edit parameter set 1 with and, confirm/proceed with enter.
Enter mode code 1288 :Edit parameter set 2 with and
, confirm/proceed with enter.
The output current is frozen(at its last value or at a preset fixedvalue, depending on the configura-tion), limit and alarm contacts are
inactive. The controller is in the con-figured state, Sensoface is off, modeindicator Configuration is on.
Configuring
Hold
Input errors
End End with conf. The measured valueand Hold are displayed alternately,
enter flashes. Press enter to end the
Hold mode. The measured value is dis-
played. The output current remains
frozen for another 20 s (HOLD icon
on, hourglass flashes).
HOLD icon
During config-
uration theStratos remainsin the Holdmode.
Parameter set 1
Parameter set 2Configure: 88 appears inthe display.
29
Menu group Code DisplaySelect menu group
M enu structure of configuration
The configuration steps are assigned to dif ferent menu groups.With the arrow keys you can jump between the individual menugroups.Each menu group contains menu items for setting the parameters.
Pressing enter opens a menu item. The values are edited using thearrow keys. Pressing enter confirms/stores the settings.Return to measurement: Press conf .
Select menu item
Output 1 o1.
Output 2 o2.
Temperature
compensationtc.
Alarm
settingsAL.
Relay / Controller rL.
CLEAN contact /Signaling param-
eter set 2Cn.
Previousmenu
group:
Menu item 1
Menu item 2
Menu item ...
8/9/2019 2402 condi
16/65
30 Stratos e 2402 CondI
Overview of configuration steps
Selection / Default
SE652/SE654/SE655/SE656/other
xx.xxx c
8 kHz / 16 kHz
mS/cm, S/m, Conc, SAL
NaCl, HCl, NaOH, H2SO4, HNO3
(Code 01 ... 10, see Pg 114)
Select current range 0-20 mA / 4-20 mA
Characteristic (not for Conc, SAL) Linear LIN / Logarithmic LOG
Time constant of output filter xxxx SEC
out2 Output 2
o2. Select temperature unit C / F
Select current range 0-20 mA / 4-20 mA
Enter current beginning xxx.x
Enter current end xxx.x
Time constant of output filter xxxx SEC
ON / OFF
Signal behavior during HOLD Last / Fix
xxx.x mA
22 mA signal for error messages ON / OFF
MenuCode
out1 Output 1
o1. Sensor selection *
Enter transfer ratio
Select measured variable
Select solution (Conc)
Signal behavior during HOLD Last / Fix
Enter fixed value xxx.x mA
xxx.xx
Entry of cell factor
Meas. frequency selection
Select temperature probe Pt100/Pt1000/NTC100
LIN: xxxx mSEnter current beginning
Enter current end
LOG: Enter current beginning
Enter current end
xxxx mS
in decades: 0.001 ... 1000
in decades: 0.001 ... 1000
other * :
Enter fixed valueFix:
Fix:
22 mA signal in the case of temp error
31
tc Temperat ure compensation
tc. Temperature compensation selection OFF / Lin / nLF
xx.xx %/K
MenuCode Selection / Default
ALrt Alarm settings
AL.Select Sensocheck ON / OFF
Enter alarm delay xxxx s
LED in HOLD mode ON / OFF
rLAY
rL. Select limit function / controller LiMIT / CtROL
Select contact function Lo / HiL1.
Select contact response N/O / N/C
Enter switching point xxxx
Enter hysteresis xxxx
Enter delay xxxx SEC
Select contact function Lo / HiL2.
Enter controller setpoint xxxxCt.
Enter neutral zone xxxx
(P) Controller gain KP xxxx %
(I) Reset time TR xxxx SEC
(D) Rate time TD xxxx SEC
PLC: Pulse length xxxx SECPFC: Pulse frequency xxxx /min
Select HOLD behavior Y Last / Y Off
rinse / PSEt 2
Rinsing interval xxx.x hrinse
Rinse duration xxxx SEC
Contact response N/O / N/C
Select contact response N/O / N/C
Enter switching point xxxx
Enter hysteresis xxxx
Enter delay xxxx SEC
Relay 1/2: Limit values, controller
(Select Cleaning/Signal/Parameter set)*
Contact CLEAN / PSet2Cln
Cn.
* These parameters are only edited in parameter set 1.They are valid for both parameter sets.
Pulse length/frequency controller PLC / PFC
Lin: Input of temp. coefficient
C fi ti
8/9/2019 2402 condi
17/65
32 Stratos e 2402 CondI
ConfigurationOutput 1Select sensor
Sensor selection
Select measured variable
Select 0-20 / 4-20 mA
End:Press conf, then enter
Code
Output 1 o1.
DisplayMenu group Select menu item
Enter current beginning
Enter current end
Set output filter
22 mA in the case of error
Hold mode
Characteristic: LIN / LOG
Select solution (Conc)
33
Note: Characters represented in gray are flashing and can be edited.
Code Display Action Choices
o1. Select configurat ion(Press conf .)
The Stratos is in HOLD mode(HOLD icon is on).
Select sensor * : Select witharrow key. Proceed with enter.
Note:
After each sensor selection thenominal cell factor of the sensoris stored.To adjust the cell factor to thedevice, calibrate the sensorafterwards!
SE 655(SE 652SE 654SE 655SE 656other)
For parameter set 1:
Enter mode code 1200(Select position using arrow
key and edit number using .When the display reads 1200 ,press enter to confirm.)
For parameter set 2:Enter mode code 1288
(Select position using arrow
key and edit number using .When the display reads 1288 ,
press enter to confirm.)
After correct input a welcometext (CONF) isdisplayed forapprox. 3 s
After correct input a welcometext ( CONF) is displayed forapprox. 3 s
* These parameters are only edited in parameter set 1.They are valid for both parameter sets.
C
8/9/2019 2402 condi
18/65
34 Stratos e 2402 CondI
ConfigurationOutput 1Select sensor and temperature probe
Sensor selection
Select solution (Conc)
Select 0-20 / 4-20 mA
End:Pressconf, then enter
Code
Output 1 o1.
DisplayMenu group Select menu i tem
Enter current beginning
Enter current end
Set output filter
22 mA in the case of error
Hold mode
Characteristic: LIN / LOG
Select measured variable
35
Code Display Action Choices
o1.
Select measuring frequency
Select with arrow key
8 KHZ(8 KHZ16 KHZ)
100Pt(100Pt1000Pt100NTC)
Proceed with enter
Select temperature probe
Select with arrow keyProceed with enter
Note: Characters represented in gray are flashing and can be edited.
With other selected the sensorparameters are entered separate-ly:
Enter cell factor:(Select position using arrow
key and edit number using .)Proceed with enter
Enter transfer ratioProceed with enter.
NoteWhen other is called up oncemore, the last sensor parame-ters are displayed and can beedited.
C fi ti
8/9/2019 2402 condi
19/65
36 Stratos e 2402 CondI
ConfigurationOutput 1Select m easured variable
End:Pressconf, then enter
Code
Output 1 o1.
DisplayMenu group Select menu item
Sensor selection
Select measured variable
Select 0-20 / 4-20 mA
Enter current beginning
Enter current end
Set output filter
22 mA in the case of error
Hold mode
Characteristic: LIN / LOG
Select solution (Conc)
37
Code Display Action Choices
o1. Select measured variable:
Select with arrow keyProceed with enter
Conductivity: 0.000 ... 9.999 mS/cm 00.00 ... 99.99 mS/cm 000.0 ... 999.9 mS/cm 0000 ... 1999 mS/cm 0.000 ... 9.999 S/m 00.00 ... 99.99 S/m
Salinity (SAL): 0.0 ... 45
Concentration (Conc) 00.00 ... 99.99 % by wt
000.0 mS(0.000 mS00.00 mS000.0 mS
0000 mS0.000 S/m00.00 S/m00.00 SAL00.00 %(Conc))
Note: Characters represented in gray are flashing and can be edited.
C fi ti
8/9/2019 2402 condi
20/65
38 Stratos e 2402 CondI
ConfigurationOutput 1Concentration measurement: Select process solution
Sensor selection
Select measured variable
Select 0-20 / 4-20 mA
Enter current beginning
Enter current end
Set output filter
22 mA in the case of error
Hold mode
End:Pressconf, then enter
Code
Output 1 o1.
DisplayMenu group Select menu item
Characteristic: LIN / LOG
Select solution (Conc)
39
Code Display Action Choices
o1. Only with 00.00 % CoNCyou can select the process solution:
Select with arrow key
-01- NaCl (0 ... 26 % by wt)
-02- HCl (0 ...18 % by wt)
-03- NaOH (0 ... 14 % by wt)
-04- H2SO4 (0 ... 30 % by wt)
-05- HNO3 (0 ... 30 % by wt)
-06- H2SO4 (92 - 99 % by wt)
-07- HCl (22 - 29 % by wt)
-08- HNO3 (35 -36 % by wt)
-09- H2SO4 (32 - 84 % by wt)
-10- NaOH (18 ... 50 % by wt)Proceed with enter
-01-SOL
(-01-SOL-02-SOL-03-SOL
-04-SOL-05-SOL-06-SOL-07-SOL-08-SOL-09-SOL-10-SOL)
Concentration curves / rangessee Pg 114 and the following
The concentration curves of many sub-stances show a maximum. This means that
if the substance concentration continues toincrease and the temperature remains con-stant, the conductivity will drop. Therefore,a one-to-one correlation of values is onlypossible in defined ranges.These partial ranges must be selected asmeasurement ranges in the configuration.
Example:Measurement ranges forsulphuric acid
Configuration
8/9/2019 2402 condi
21/65
40 Stratos e 2402 CondI
ConfigurationOutput 1Output current range. LIN/LOG curveLIN curve: Current beginning / end
End:Pressconf, then enter
Code
Output 1 o1.
DisplayMenu group Select menu item
Sensor selection
Select measured variable
Select 0-20 / 4-20 mA
Enter current beginning
Enter current end
Set output filter22 mA in the case of error
Hold mode
Characteristic: LIN / LOG
Select solution (Conc)
41
Code
LIN(LIN / LOG)
Display Action Choices
o1.
Select output characteristicSelect with keyProceed with enter(Step omitted for % (Conc) and SAL))
000.0 mS(xxx.x mS)
100.0 mS(xxx.x mS)
Conductivity
[mS/cm]
204
200
0
Assignment of measured values:Current beginning and current end
[mS/cm]
204
200
100[mA]
Example 1: Range 0...200 mS/cm Example 2: Range 100...200 mS/cmAdvantage: Higher resolution in range ofinterest
[mA]
Output current
Conductivity
4-20 mA(0 - 20 mA4 - 20 mA)
Set output current range
Select with keyProceed with enter
With LIN selected: Enter current beginning
Enter lower end of scale
Select with key,
edit number with key,proceed with enter
Enter current endEnter upper end of scaleProceed with enter
Output current
Configuration
8/9/2019 2402 condi
22/65
42 Stratos e 2402 CondI
ConfigurationOutput 1Output current range. LOG curveCurrent beginning / end
End:Pressconf, then enter
Code
Output 1 o1.
DisplayMenu group Select menu item
Sensor selection
Select measured variable
Select 0-20 / 4-20 mA
Enter current beginning
Enter current end
Set output filter22 mA in the case of error
Hold mode
Characteristic: LOG
Select solution (Conc)
43
Code Display Action Choices
o1. With LOG selected: Enter current beginning
Enter lower end of scale
Select with key,
edit number with key,proceed with enter key.
0.1 mS(0.001 mS0.01 mS1.0 mS
10 mS100 mS1000 mS)
100 mS(0.001 mS0.01 mS1.0 mS10 mS100 mS1000 mS)
Enter current endEnter upper end of scale
Select with key,edit number with key.
Proceed with enter
Example: Measurement range over 3 decades
Selection: 0-20/4-20mA Current beginning: 4 mA Current end: 20 mA
Characteristic: LOG 4 mA 0.1 mS/cm
20 mA 100 mS/cm
[mA]
100
4.0
Output current
Y
X
Conductivity[mS/cm]
20.09.3 14.7
1
10
0.1
Configuration
8/9/2019 2402 condi
23/65
Stratos e 2402 CondI44
ConfigurationOutput 1Time constant of output filter
End:Pressconf, then enter
Code
Output 1 o1.
DisplayMenu group Select menu i tem
Sensor selection
Select measured variable
Select 0-20 / 4-20 mA
Enter current beginning
Enter current end
Set output filt er
22 mA in the case of error
Hold mode
Characteristic: LIN / LOG
Select solution (Conc)
45
0 s
0 ... 120 s
Code Display Action Choices
o1. Time constant of output filterDefault setting: 0 s (inactive).To specify a time constant:
Select with key,
edit number with key,proceed with enter key.
Time constant of output filter (attenuation)To smoothen the current output, a low-pass fil ter withadjustable filter time constant can be switched on. When thereis a jump at the input (100 %), the output level is 63 % afterthe time constant has been reached.
The time constant can be set from 0 to 120 s.If the time constant is set to 0 s, the current output follows theinput.Note:The filter only acts on the current output, not on the display, thelimit values, or the controller!
Leitfhigkeit
Time constant 0 to 120 s
Conductivity
Configuration
8/9/2019 2402 condi
24/65
Stratos e 2402 CondI46
ConfigurationOutput 1Output current during Error and HOLD.
End:Pressconf, then enter
Code
Output 1 o1.
DisplayMenu group Select menu item
Sensor selection
Select measured variable
Select 0-20 / 4-20 mA
Enter current beginning
Enter current end
Set output filter
22 mA in the case of error
Hold mode
Characteristic: LIN / LOG
Select solution (Conc)
47
OFF
(OFFON)
Code Display Action Choices
o1. 22 mA signal for error messageSelect with keyProceed with enter
LAST(LASTFIX)
Only with FIX selected:
Enter current which is to flow atthe output during HOLD
Select position with key,
edit number with key,proceed with enter key.
021.0 mA(00.0 ...
21.0 mA
Output signal f or HOLD:
Output current[mA]
Output signal HOLDSetting FIX = 21.0 mA
HOLD active
21
4
Output signal HOLDSetting LAST
HOLD active
Output signal during HOLDLAST: During HOLD the lastmeasured value is maintained atthe outputFIX: During HOLD a value (to beentered) is maintained at theoutput
Select with key.Proceed with enter
Configuration
8/9/2019 2402 condi
25/65
Stratos e 2402 CondI48
ConfigurationOutput 2Temperature unit and output current.
Select C/F
Select 0-20 / 4-20 mA
Enter current beginning
Enter current end
Set output filter
22 mA for temp error
Hold mode
End:Pressconf, then enter
Code DisplayMenu group Select menu i tem
Output 2 o2.
49
C(CF)
Code Display Action Choices
o2. Specify temperature unitSelect with keyProceed with enter
4-20 mA(4 - 20 mA0 - 20 mA)
Set output current rangeSelect with keyProceed with enter
Current beginning: Enter lowerend of scale.
Select with , edit number with
, proceed with enter.
000 .0 C(xxx.x C)
Current end: Enter upper end ofscale.
Select with , edit number with
, proceed with enter.
100 .0 C(xxx.x C)
Output current
Process temperature
204
100
0
Process temperature: Current beginning and current end
Process temperature
204
70
50[mA]
Example 1: Range 0 to 100 C Example 2: Range 50 to 70 C.Advantage: Higher resolution inrange of interest[C]
[C]
[mA]
Output current
Configuration
8/9/2019 2402 condi
26/65
Stratos e 2402 CondI50
Co gu at oOutput 2Time constant of output filter.
End:Pressconf, then enter
Output 2 o2.
Code DisplayMenu group Select menu item
Select C/F
Select 0-20 / 4-20 mA
Enter current beginning
Enter current end
Set output filt er
22 mA for temp error
Hold mode
51
0 s
(0 - 120 s)
Code Display Action Choices
o2. Time constant of output filterDefault setting: 0 s (inactive).To specify a time constant:
Select with key,
edit number with key,proceed with enter key.
Time constant of output filter (attenuation)To smoothen the current output 2, a low-pass filter withadjustable filter time constant can be switched on. When thereis a jump at the input (100 %), the output level is 63 % afterthe time constant has been reached.
The time constant can be set from 0 to 120 s.If the time constant is set to 0 s (default), the current outputfollows the input.Note:The filter only acts on the current output, not on the display!
Leitfhigkeit
Time constant 0 to 120 s
Conductivity
Configuration
8/9/2019 2402 condi
27/65
Stratos e 2402 CondI52
gOutput 2Temperature error. Output current during HOLD.
End:Pressconf, then enter
Output 2 o2.
Code DisplayMenu group Select menu item
Select C/F
Select 0-20 / 4-20 mA
Enter current beginning
Enter current end
Set output filter
22 mA for temp error
Hold mode
53
OFF(ON)
Code Display Action Choices
o2. 22 mA signal for error messageSelect with arrow keyProceed with enter
Output signal during HOLDLAST: During HOLD the lastmeasured value is maintained atthe outputFIX: During HOLD a value (to beentered) is maintained at theoutput
Select with arrow key.Proceed with enter
LAST(FIX)
Only with FIX selected:Enter current which is to flow atthe output during HOLD
Select position with key,
edit number with key,proceed with enter key.
21.0 mA(00.0 ...
21.0 mA
Output signal during HOLD:
Output current[mA]
Output signal HOLDSetting FIX = 21.0 mA
HOLD active
21
4
Output signal HOLDSetting LAST
HOLD acti ve
Configuration
8/9/2019 2402 condi
28/65
Stratos e 2402 CondI54
gTemperature compensationSelect temperature compensation
Selecttemperature compensation
End:Pressconf, then enter
Temperature
compensationtc.
Code DisplayMenu group Select menu item
55
OFF
(OFFLINnLF)
Code Display Action Choices
tc.
Only with manual temp detec-tion selected (MAN)Enter temperature.Select position with arrow keyand edit number with key.Proceed with enter
Only with linear temperaturecompensation (LIN) selected:Enter temperature coefficient.
Select position with key,edit number with key,proceed with enter key.
02.00%/K(xx.xx%/K)
Temperature compensationselection (not for SAL, CONC)
OFF: Temperature compensa-tion switched off
Select with key,proceed with enterLIN:Linear temperature compensa-tion with entry of temperaturecoefficient and reference tem-peratureNLF:Temperature compensation fornatural waters toEN 27888
Configuration
8/9/2019 2402 condi
29/65
Stratos e 2402 CondI56
Alarm sett ings
Select Sensocheck
Delay
LED in HOLD mode
Code DisplayMenu group Select menu i tem
Alarm
settingsAL.
End:Pressconf, then enter
15
16
Alarm Alarm contact
The alarm contact is closed during normal operation(N/C). It opens in the case of alarm or power outage.As a result, a failure message is provided even in thecase of line breakage (fail-safe behavior). For contact rat-ings, see Specifications.
Error messages can also be signaled by a 22 mA outputcurrent ( see Pg 47,
The operating behavior of the alarm contact is shown onPg 94.
The alarm delay acts on the LED, the 22 mA signaland the alarm contact.
53, 92).
57
Code Display Action Choices
OFF(ON / OFF)
AL. Select Sensocheck(Continuous monitoring ofsensor properties)
Select with , proceed with
enter
Alarm delay
Select with , edit number
with , proceed with enter.
0010 s
(xxxx s)
LED in HOLD modeSelect with , proceed withenter
OFF(ON / OFF)
Configurat ion Alarm HOLD
ON on flashes
OFF f lashes of f
LED state:
Configuration
8/9/2019 2402 condi
30/65
Stratos e 2402 CondI58
Limit functionRelay 1
Contact function
Contact response
Enter swit ching point
Enter hysteresis
Delay
Relay 2 menu group
Controller menu group
L1.
L2.
Ct.
Relay / Controller rL.
Code DisplayMenu group Select menu i tem
End:Pressconf, then enter
59
LiMIT(LiMitCtROL)
Code Display Action Choices
rL. Use of relays: Limit function (LiMIT) Controller (CtROL)
Select with .
Proceed with enterNote:Selecting CtROL leads toController menu group Ct.
Lo(Lo/Hi)
Limit 1 function , see Pg 61.
Select with .Proceed with enter
Limit 1 contact responseN/O: normally open contact
N/C: normally closed contactSelect with .Proceed with enter
N/O(NON/C)
Limit 1 switching point
Select with , edit number
with , proceed with enter.
000.0 mS(xxxx)
Limit 1 hysteresis
Select with , edit number
with , proceed with enter.
001.0 mS(xxxx)
Limit 1 delayThe contact is activated withdelay (deactivated withoutdelay)
Select with , edit number
with , proceed with enter.
0010 s(0 ...9999 s)
L1.
Configuration
8/9/2019 2402 condi
31/65
Stratos e 2402 CondI60
Limit functionRelay 2
Relay 1 menu group
Contact function
Contact response
Enter switching point
Enter hysteresis
Delay
Controller menu group
L1.
L2.
Ct.
Relay / Controller rL.
Code DisplayMenu group Select menu i tem
End:Pressconf, then enter
61
Code Display Action Choices
L2. Hi(Lo/Hi)
Select Limit 2, see Fig. below.
Select with .Proceed with enter
Limit 2 contact responseN/O: normally open contactN/C: normally closed contact
Select with .Proceed with enter
N/O(N/ON/C)
Limit 2 switching point
Select with , edit number with
, proceed with enter.
100.0 mS(xx.xx mS)
Limit 2 hysteresis
Select with , edit number with, proceed with enter.
001.0 mS(xx.xx mS)
Limit 2 delayThe contact is activated withdelay (deactivated withoutdelay)
Select with , edit number with, proceed with enter.
0010 s(0 ...9999 s)
Limit Lo
Hysteresis+
Switchingpoint
Contact
0
1
Signal
Limit Hi
Hysteresis-
Signal
Switchingpoint
Contact
0
1
ConfigurationC t ll (f d i ti P 88 d th f ll i )
8/9/2019 2402 condi
32/65
Stratos e 2402 CondI62
Controller (for description see Pg 88 and the follow ing)Setpoint. Neutral zone.
Relay 1 menu group
Relay 2 menu group
Controller setpoint
Enter neutral zone
(P) Controller gain
(I) Reset time TR
(D) Rate time TD
Pulse length/Pulse frequency
PLC: Pulse length
PFC: Pulse frequency
HOLD behavior
L1.
L2.
Ct.
Relay / Controller rL.
Code DisplayMenu group Select menu item
End:Pressconf, then enter
63
001.0 mS
(xxx.x mS)
050.0 mS(xxx.x mS)
Setpoint
Select with , edit number with, proceed with enter.
Neutral zone (dead band)
Select with , edit number with, proceed with enter.
Controller: P action
Select with , edit number with
, proceed with enter.
Code Display Action Choices
Ct.
0100 %(xxxx %)
Controller: I action (reset time)
Select with , edit number with
, proceed with enter.
0000 s(xxxx s)
Controller: D action (rate time)Select with , edit number with
, proceed with enter.
0000 s(xxxx s)
Pulse length /Pulse frequency
Select with .Proceed with enter
PLCPLC/(PFC)
PLC: Pulse length
Select with , edit number with, proceed with enter.
0010 s(xxxx s)
PFC: Pulse frequencySelect with , edit number with
, proceed with enter.
0060 /min(xxxx /min)
Behavior during HOLD
Select with .Proceed with enter.
Y Last(YOff/Y Last)
KonfigurationControlling a rinsing probe or
8/9/2019 2402 condi
33/65
64
Controlling a rinsing probe orsignaling parameter set 2
Select CLEAN/ PSEt2
Rinsing int erval
Rinse duration
Cont act response
Contact:CLEAN / PSEt2
Cln
Code DisplayMenu group Select menu item
Code Display Action (Rinsing probe) Choices
Cn. Control* of : Rinsing probe (rinse) Signaling parameter set 1/2
Select with .Proceed with enter
000.0 h(xxx.x h)
Rinsing interval*
Select with , edit number
with , proceed with enter.
Rinse duration*
Select with , edit numberwith , proceed with enter.
0060 s(xxxx s)
Contact response*N/O: normally open contactN/C: normally closed contact
Select with .Proceed with enter
N/O(N/ON/C)
rinse
rinse
(rinse / PSEt2)For PSEt2,
see next page
* These parameters are only edited in parameter set 1.They are valid for both parameter sets. 65
Signaling parameter set 2
17
18
CleanPSet2
Process control system(PCS)
Parameter set 1 selected
Parameter set 2 selected
Power supply
Stratos e 2402
Power supply:AC< 250 V / < 3 A / < 750 VADC< 30 V / < 3 A / < 90 W
17
18
Clean
e.g. rinse cleaning
Power supply
Controlling a rinsing probeThe Clean contact can be used to connect a simple rinsing probe.Rinse duration and rinsing interval are defined during configuration.Contact response can be set as N/O, N/C.
Depending on the selected parameter set, the relay
is active or inactive. The signal can be used forsuperordinated process control systems.Parameter set 2 is indicated by 88 in the upperleft corner of the display.
Select ing paramet er set (1/2)M anually or via a signal at the Control input
8/9/2019 2402 condi
34/65
Stratos e 2402 CondI66
M anually or via a signal at the Control input
Display Action Choices
Select: Parameter set 1 (MAN) Parameter set 2 (MAN) Automatic switchover via
Control input (Ctr-EXT)
Select with , proceed with enter
Select parameter setPress conf , enter code 7654
Select with , edit number with ,proceed with enter.
Wrong settings change the measure-ment properties!If an invalid code is entered, theStratos returns to measuring mode.
With -1- or -2- selected:Since the complete device configurationis changed in one step, there is a secu-rity prompt (No/Yes).
Note:
When pressing enter directly, theselection is not stored.
Activation of parameter set 2 is indi-cated by 88 in the upper left cornerof the display.
After correct input a welcome
text (CONF) isdisplayed forapprox. 3 s
-1-(-1- MAN-2-MANCtr-EXT)
Ctr-EXT:
see nextpage
67
With Control input Ctr-EXT selected:You can switch between the parame-ter sets by applying an external signalto the Control input:
Display Action Choices
External sw itchover of parameter sets
The parameter set can be selected from outside by sending asignal to the Control input (e.g. from the process control sys-tem). To do so, the Control input is set to Ctr-EXT during config-uration.
Note:
Parameter set 2 is indicated by 88 in the upper left corner of t hedisplay.
7
8
Control
input
0 ... 2 V AC/DC Parameter set 1 selected
10 ... 30 V AC/DC Parameter set 2 selected
Stratos e 2402 Process control system (PCS)
Power supply12...24 V AC/DC
8/9/2019 2402 condi
35/65
68 Stratos e 2402 CondI 69
Two complete parameter sets are stored in the EEPROM.As delivered, the two sets are identical but can be edited.Note:
Fill in your configuration data on the following pages.
Code. Parameter Default set t ing
o1. Sensor selection * SE655o1. Process variable 000.0 mSo1. Conc solut ion -01-o1. 0/4-20 mA 4-20 mAo1. Characteristic LINo1. Current start (LIN) 000.0 mSo1. Current end (LIN) 100.0 mS
o1. Current start (LOG) 0.1 mSo1. Current end (LOG) 100 mSo1. Filter time 0 so1. 22mA signal OFFo1. Hold behavior Lasto1. Fix current 021.0 mAo2. Unit C / F Co2. 0/4...20mA 4-20 mAo2. Current start 000.0 Co2. Current end 100.0 Co2. Filter time 0 so2. 22mA signal OFFo2. Hold behavior Last
o2. Fix current 021.0 mAtc. Temp compensation OFFtc. Temp coefficient 02.00%/K
Code. Parameter Default set t ing
AL. Sensocheck OFFAL. Alarm delay 0010 sAL. LED Hold OFFrL. Relay function LimitL1. Contact function LoL1. Contact response N/OL1. Switching point 000.0 mS
L1. Hysteresis 001.0 mSL1. Delay 0010 sL2. Contact function HiL2. Contact response N/OL2. Switching point 100.0 mSL2. Hysteresis 001.0 mSL2. Delay 0010 sCt. Setpoint 050.0 mSCt. Neutral zone 001.0 mSCt. Paction 0100 %Ct. I action 0000 sCt. D action 0000 sCt. PLC/PFC controller PLC
Ct. Pulse length 0010 sCt. Pulse frequency 0060 /minCt. Hold behavior LastCn. Rinse/Pset2 * rinseCn. Rinsing interval * 000.0 hCn. Rinse duration * 0060 sCn. Contact response* N/O
Default settings of parameter sets
* These parameters are only edited in parameter set 1.They are valid for both parameter sets.
Parameter set - user settings
8/9/2019 2402 condi
36/65
Stratos e 2402 CondI70
Code. Parameter Sett ing
P1 (conf 1200) P2 (conf 1288)
o1. Sensor selection ____________ *
o1. Process variable ____________ ____________
o1. Solut ion (Conc) ____________ ____________
o1. 0/4-20 mA ____________ ____________
o1. Characteristic (LIN/LOG) ____________ ____________
o1. Current start ____________ ____________
o1. Current end ____________ ____________
o1. Filter time ____________ ____________
o1. 22mA signal ____________ ____________
o1. Hold behavior ____________ ____________
o1. Fix current ____________ ____________
o2. Unit C / F ____________ ____________
o2. 0/4...20mA ____________ ____________
o2. Current start ____________ ____________
o2. Current end ____________ ____________
o2. Filter time ____________ ____________
o2. 22mA signal ____________ ____________
o2. Hold behavior ____________ ____________
o2. Fix current ____________ ____________
tc. Temp compensation ____________ ____________
tc. Temp coefficient ____________ ____________AL. Sensocheck ____________ ____________
AL. Alarm delay ____________ ____________
AL. LED Hold ____________ ____________
71
Code. Parameter Sett ingP1 (conf 1200) P2 (conf 1288)
rL. Relay funct ion ____________ ____________
L1. Contact function ____________ ____________
L1. Contact response ____________ ____________
L1. Switching point ____________ ____________
L1. Hysteresis ____________ ____________
L1. Delay ____________ ____________
L2. Contact function ____________ ____________
L2. Contact response ____________ ____________
L2. Switching point ____________ ____________
L2. Hysteresis ____________ ____________
L2. Delay ____________ ____________
Ct. Setpoint ____________ ____________
Ct. Neutral zone ____________ ____________
Ct. P action ____________ ____________
Ct. I action ____________ ____________
Ct. D action ____________ ____________
Ct. PLC/PFC controller ____________ ____________
Ct. Pulse length ____________ ____________
Ct. Pulse frequency ____________ ____________
Ct. Hold behavior ____________ ____________
Cn. Rinse/ PSet2 ____________ *Cn. Rinsing interval ____________ *
Cn. Rinse duration ____________ *
Cn. Contact response ____________ *
* These parameters are only edited in parameter set 1.They are valid for both parameter sets.
CalibrationCalibration adjusts the device to the sensor
8/9/2019 2402 condi
37/65
Stratos e 2402 CondI72
Calibration adjusts the device to the sensor.
Activate with cal
Enter mode code: Entry of cell factor 1100 With calibration solution 0110 Product calibration 1105 Zero point 1001
Temp probe adjustment 1015Select with , edit number with ,proceed with enter.(End with cal + enter.)
Output current is frozen (last valueor preset fixed value, depending onconfiguration), limit and alarm con-tacts are inactive. The controller isin the configured state, Sensofaceis off, Calibration mode indicatoris on.
Activate
Hold
Input errors The calibration parameters arechecked during the input. In thecase of an incorrect input Err isdisplayed for approx. 3 s. Theincorrect parameters cannot bestored. Input must be repeated.
End End with cal.
The measured value and Hold aredisplayed alternately, enter flashes.Pressenter to end the Hold mode.The measured value is displayed.The output current remains frozenfor another 20 s (HOLD icon on, hourglass flashes).
HOLD icon
During calibra-tion the Stratosremains in the
Hold mode forreasons of safe-
ty.
73
Informat ion on calibrationCalibration can be performed by: Entry of cell factor Determining the cell factor with a known calibration
solution taking account of the temperature Product calibration
Zero calibration in air or with calibration solution Temperature probe adjustment
Note:If measurements are taken in containers withA < 110 mm, be sure to choose the samedistance and the same container material(metal/plastic) for calibration.
Caution All calibration procedures must be performed by trained per-
sonnel. Incorrectly set parameters may go unnot iced, butchange the measuring properties.
When another sensor is used, its sensor data (cell factor,transfer ratio, measuring frequency, t emperature probe) must
be entered in the configuration menu before calibration. Each time a new sensor is connected, the Stratos must becalibrated.
Calibration by input of cell factor
8/9/2019 2402 condi
38/65
74 Stratos e 2402 CondI
Display Action Remark
Enter cell factor:Select with , edit number
with , proceed with enter.Conductivity and temperatureare alternately displayedduring the input(lower display).Confirm entry with enter.
Press cal key.Enter code 1100
Select with , edit number
with , proceed with enter.
Ready for calibration
Dismount and clean sensor
If an invalid code isentered, the Stratosreturns to measur-ing mode.
The entered cell factor andzero point are displayed.Confirm with enter.
Display (3 s)Stratos in Holdmode, measuredvalue frozen.
Sensoface inactive.
Input of cell factor with simultaneous display of conductivity andtemperature (without temperature compensation)
75
Display Act ion Remark
Conductivity and temperatureare displayed.
The measured value is shown
in the main display alternatelywith Hold ; enter flashes.End with enter.Press enter to end calibra-tion.
After end of cali-bration, the out-puts remain inHold mode for
approx. 20 sec.
Calibration w ith calibration solution
8/9/2019 2402 condi
39/65
Stratos e 2402 CondI76
Display Action Remark
Press cal key.Enter code 0110.
Select with , edit number
with , proceed with enter.
Ready for calibration
Dismount and clean sensor
If an invalid code isentered, the Stratosreturns to measur-ing mode.
Display (3 s)Stratos in Holdmode, measuredvalue frozen.
Sensoface inactive.
Be sure to use known calibration solutions and the respective tempera-ture-corrected conductivity values (see Calibration solutions Pg 112).During the calibration procedure the temperature should be kept con-stant.
Immerse sensor in calibrationsolution.Enter the temperature-corrected conductivity valueof the calibration solution:
Select with , edit numberwith , proceed with enter.
Cell factor and temperatureare alternately displayed in thelower display.Confirm entry with enter.
When there hasnot been an entryfor 6 sec, thelower display alter-nately shows the
conductivity andtemperature value.
77
Display Act ion Remark
The determined cell factor andzero point are displayed.Confirm cell factor with
enter.
Conductivity and temperatureare displayed.The measured value is shownin the main display alternatelywith Hold ; enter flashes.End with enter.Press enter to end calibra-tion.
After end of cali-bration, the out-puts remain inHold mode forapprox. 20 sec.
Product calibrat ionCalibration by sampling
8/9/2019 2402 condi
40/65
78 Stratos e 2402 CondI
The measurement process is only interrupted briefly. Duringproduct calibration the sensor remains in the process.Calibration is without TC correction!
Procedure: During sampling the currently measured value isstored in the Stratos. The Stratos immediately returns to meas-uring mode. The calibration mode indicator flashes andreminds you that calibration has not been terminated. Thesample is measured in the lab or directly on the site using aportable meter. The measured sample value is then entered inthe Stratos. The new cell factor is calculated from these twovalues.If the sample is invalid, you can take over the value stored dur-ing sampling. In that case the old calibration values are stored.Afterwards, you can start a new product calibration.
Display Action Remark
Product calibration step 1:Press cal key.Enter code 1105.
(Select position with key,edit number with key,proceed with enter key. )
Take sample and store value.Proceed with enter
If an invalid code isentered, the Stratosreturns to measur-ing mode.
Now the samplecan be measuredin the lab.
Display(approx. 3 sec)
79
Display Act ion Remark
Measuring mode:
From the flashing CAL modeindicator you see that samplecalibration has not been ter-minated.
Product calibration step 2:When the sample value hasbeen determined, call up theproduct calibration once more(cal, code 1105).
Display(approx. 3 sec)
Enter sample value. The newcell factor is calculated.
Confirm with enter.
While the samplevalue is deter-mined, the Stratosis in measuring
mode.
The new cell factor and zeropoint are displayed.Confirm with enter.
The measured value is shown
in the main display alternatelywith Hold ; enter flashes.End with enter.
After end of cali-
bration, the out-puts remain inHold mode forapprox. 20 sec.
New calibration:Press cal.
Zero calibration in air
8/9/2019 2402 condi
41/65
80 Stratos e 2402 CondI
Display Action Remark
When there has notbeen an entry for6 sec, the lower
display alternatelyshows the conduc-tivity and tempera-ture value.
Press cal key,enter code 1001
Select position with key,
edit number with key,
proceed with enter key.
Modify the zero point untilzero is displayed as conductiv-ity value in the lower display.
Select with .
Edit number with .
If required, change the sign ofthe zero point.
Press enter to confirm thezero point.
Stratos is in theHold mode.If an invalid code isentered, Stratos
returns to measur-ing mode.
Ready for calibration
Dismount and clean sensor.(Sensor must be dry!)
Display (3 s)
81
Display Action Remark
The cell factor and zero pointare displayed.Press enter to confirm thecalibration data.
Conductivity and temperatureare displayed.
Place sensor in process.
End calibration with
enter.
After end of calibra-tion, the outputsremain in Holdmode for approx.
20 sec.
Zero calibration with calibration solution
Calibration solution w ith low conductivity
8/9/2019 2402 condi
42/65
82 Stratos e 2402 CondI
Display Action Remark
When there has not
been an entry for6 sec, the lowerdisplay alternatelyshows the conduc-tivity and tempera-ture value.
Calibration solution w ith low conductivity
Press cal key,enter code 1001
Select position with key,
edit number with key,proceed with enter key.
Immerse sensor in calibration
solution.Modify the value until thelower display shows the con-ductivity value of the calibra-tion solution.Press enter to confirm cali-bration
Stratos is in theHold mode.If an invalid code isentered, Stratosreturns to measur-ing mode.
Ready for calibration
Dismount and clean sensor
Display (3 s)
The cell factor and zero pointare displayed.
Press enter to confirm thecalibration data.
83
Display Act ion Remark
Conductivity and temperatureare displayed.Remove the sensor from the
calibration solution and cleanit.Place sensor in process.
The measured value is shownin the main display alternatelywith Hold ; enter flashes.End with enter.End calibration with
enter.
After end of calibra-tion, the outputsremain in Holdmode for approx.20 sec.
Temperature probe adjustm ent Diagnostics functions
8/9/2019 2402 condi
43/65
Stratos e 2402 CondI84
Display Action Remark
Measure the temperature ofthe process medium using anexternal thermometer. Entermeasured temperature value:
Select with , edit number
with , proceed with enter.End adjustment with enter.HOLD will be deactivatedafter 20 sec.
Activate calibration(Press cal,enter 1015)
Select position with key,
edit number with key,proceed with enter key.
Ready for calibration
Wrong settingschange the meas-urement properties!If an invalid code is
entered, the Stratosreturns to measur-ing mode.
Default:Current value ofsecondary display.
Display (approx. 3 sec)Stratos is in the Holdmode.
Display Remark
Measurement
In the measuring mode the main display shows theconfigured process variable (conductivity [mS/cm,S/m] or resistivity [Mcm] or concentration [% bywt] or salinity [SAL]) and the lower display the tem-perature. During calibration you can return to meas-uring mode by pressing the cal key, during configu-ration by pressing conf .
85
Display of output currentsPress enter while in measuring mode.The current at output 1 is shown in the main dis-play, the current at output 2 in the secondary dis-
play.After 5 sec the Stratos returns to measuring mode.
Display Remark
Sensor monitor for validationof sensor and complete measured-value processing.Loop a defined sensing resistor (e.g. R = 100 )
through the sensor as shown in the figure.Press the conf key and enter code 2222.The sensor monitor displays the directly measuredresistance and the temperature.If t here is a signif icant dif ference between resistorvalue and display, the sensor and its transmissionbehavior should be checked.Press enter to return to measurement.Note:The Stratos does not automatically go to Holdmode.
Display of last error message(Error Inf o)Press conf while in measuring mode and confirmcode 0000. The last error message is displayed forapprox. 20 sec. After that the message will bedeleted. (immediate return to measurement atpressing enter).
Display of calibration data (Cal Info)Press cal while in measuring mode and confirmcode 0000. The current cell constant is shown inthe main display. After 20 sec the Stratos returns tomeasuring mode (immediate return at pressing
enter).
Diagnostics functionsThese functions are used for testing the connected peripherals.
8/9/2019 2402 condi
44/65
Stratos e 2402 CondI86
Relay t est ( manual test of contacts) Pressconf , enter code 5557The relays are frozen. This state is indicated in thedisplay. The 4 digits in the display correspond to the4 relays (as on terminal plate):1st digit: R1
2nd digit: R23rd digit: AL4th digit: CLNFunction test using arrow keys see left column.When exiting the function (enter), the relays are setcorresponding to the measured value.
Select arelay
Specify current for output 1(current source 1) Pressconf , enter code 5555The current indicated in the main display for output1 can be edited.
Select with key, edit number with key,proceed with enter key.The actually measured current is shown in the sec-ondary display. The Stratos is in Hold mode.Press enter to return to measurement(Hold remains active for another 20 sec).
Display Action / Remarks
Specify current at output 2(current source 2) Pressconf , enter code 5556The current indicated in the main display for output2 can be edited.
Select with key, edit number with key, proceedwith enter. The actually measured current is shownin the secondary display. The Stratos is in Hold mode.Press enter to return to measurement.
Test 0/1
Return tomeasurement
87
The arrows indi-cate which relay(valve) is active:
Controllercharacteristic
+100 %
-100 %
Relay 2 active(Meas. value >setpoint)
Relay 1 active(Meas. value 45 Sensor connection or cable defective
Errors DisplayProblem
Possib le causes
ERR 01 Measuredvalue
flashes
xxx
Unsuitable sensor Conductance range > 3000 mS
ERR 02 Measuredvalueflashes
xxx
System errorConfiguration or calibration datadefective; completely reconfigureand recalibrate the device.Memory error in device program
ERR 98 ConFflashes
xxx x
Factory sett ingsEEPROM or RAM defectiveThis error message only occurs inthe case of a complete defect.The Stratos must be repaired andrecalibrated at the factory.
ERR 99 FAILflashes
xxx x
Temperat ure probeOpen or short circuitTemperature range exceeded
ERR 03 xxx x
93
Sensocheck: Primary coil
Secondary coil
Alarmc
ontac
RedLED
Out1(22mA
Out2(22mA
Current output 1Current below 0 (3.8) mA
ErrorsSymbol(flashes)
Problem
Possib le causes
ERR 11 xxx
Current output 1Current above 20.5 mA
ERR 12 xxx
Current output 1Current span too small / too large
ERR 13 xxx
Current output 2Current below 0 (3.8) mA
ERR 21 xxx
Current output 2Current above 20.5 mA
ERR 22 xxx
Current output 2Current span too small / too largeERR 23xxx
ERR 33xxx
ERR 34Sensoface activesee Pg 97
Temperature outsideconversion tables(TC, conc, SAL)
Sensoface activesee Pg 97
Operating stat es
t
8/9/2019 2402 condi
48/65
Stratos e 2402 CondI
Conf par set 2(conf) 1288
Parameter set 1/2(conf) 7654
94
Alarmc
ontact
Cleaning
contact
LED
TimeoutOperating
state
Measurement
Rel.1/2
Controller
Rel.1/2
Limitvalue
Out2
Out1
Cal Info(cal) 0000
20 s
Error Info(conf) 0000
20 s
Calibration(cal) 1100
Temp adjustment(cal) 1015
Zero point(cal) 1001
Calibration(cal) 0110
Conf par set 1
(conf) 1200
Product cal 1(cal) 1105
Product cal 2(cal) 1105
20
min
20min
20min
95
Sensor monitor(conf) 2222
Current source 1(conf) 5555
Current source 2(conf) 5556
Relay test(conf) 5557
Manual controller(conf) 5559
Cleaningfunction
HOLD input
Alarmc
ontact
Cleaning
contact
LED
TimeoutOperating
state
Rel.1/2
Controller
Rel.1/2
Limitvalue
Out2
Out1
Explanation: active
as configured (Last/Fix or Last/Off)
20min
20min
20min
20min
20min
Sensoface
The little smiley in the display (Sensoface) alerts to sensor
8/9/2019 2402 condi
49/65
Stratos e 2402 CondI96
SensocheckContinuously monitors the primary coil and its lines for shortcircuits and the secondary coil and its lines for open circuits.Critical values make the Sensoface sad and the correspondingicon flashes:
The Sensocheck message is also output as error message Err 33.The alarm contact is active, the red LED is lighted, output cur-rent 1 is set to 22 mA (when configured correspondingly).Sensocheck can be switched off during configuration (thenSensoface is also disabled). Exception: After a calibration aSmiley is always displayed for confirmation.
The little smiley in the display (Sensoface) alerts to sensorproblems (defective sensor, defective cable).The conditions for a friendly, neutral, or sad Sensoface aresummarized in the following chart. Additional icons refer tothe error cause.
NoteThe worsening of a Sensoface criterion leads to the devaluationof the Sensoface indicator (Smiley becomes sad ). To reset theSensoface indicator, the defect must be remedied and theStratos be calibrated.
97
StatusDisplay Problem
Sensor defect Short circuit in primary coilOpen circuit in secondary coil(see also Error messagesErr 33 and Err 34, Page 93).
Temperature outside range for TC,conc, SAL
Temperatureerror
Appendix
Product line and accessories
8/9/2019 2402 condi
50/65
98 Stratos e 2402 CondI 99
Product line and accessories
Devices Order No.
Strat os e 2402 CondI 2402 CondI
M ountin g accessories
Pipe-mount kit ZU 0274Panel-mount kit ZU 0275Protective hood ZU 0276
SensorsSE 655 electrodeless conductivity sensor SE 655SE 656 electrodeless conductivity sensor SE 656
For more information concerning our sensors and fittings product line pleaserefer to our Sensors, Fitt ings, Accessories catalog:Download at http://www.knick.deor request catalog:Phone: +49 (0)30 - 801 91 - 290Fax: +49 (0)30 - 801 91 - 200Mail: [email protected]
Note:For special applications (chemical resistance, type of mounting), youcan also connect sensors from other manufacturers.Permissible ranges for the Stratos e 2402 CondI as well as terminalassignments and factory settings for these sensors are available onrequest.
Conductivity input Input for electrodeless conductivity sensors:
Specifications
Sensor standardizati on
8/9/2019 2402 condi
51/65
Stratos e 2402 CondI100
y p p ySE 652, SE 654 X, SE 655, SE 656
Display ranges Conductivity 0.000 ... 1999 mS/cm
Concentration 0.00 ... 100.0 % by wt
Salinity 0.0 ... 45 (0 ... 35 C)
Measurement ranges Conductivi ty 0.000 . .. 9.999 mS/cm00.00 ... 99.99 mS/cm000.0 ... 999.9 mS/cm0000 ... 1999 mS/cm0.000 ... 9.999 S/m00.00 ... 99.99 S/m
Concentration 00.00 ... 99.99 % by wt
Salinity 0.0 ... 45 (0 ... 35 C)
Response time (T90) Approx. 2/sec
Measurement error1,2,3)
< 1% meas.val. + 0.005 mS
Temperat ure compensation*
(Reference temp 25 C) (OFF) Without(Lin) Linear characteristic 00.00 ... 19.99 % /K(NLF) Natural waters to EN 27888 (0 ... 35C)
Concentration determinationOperating modes: * )
-01- NaCl 0 ... 26 % by wt (0 ... 60 C)-02- HCl 0 ... 18 % by wt (-20 ... 50 C)-03- NaOH 0 ... 14 % by wt (0 ... 100 C)-04- H2SO4 0 ... 30 % by wt (-17 ... 110 C)
-05- HNO3 0 ... 30 % by wt (-20 ... 50 C)
-06- H2SO4 92 ... 99 % by wt (-17 ... 115 C)-07- HCl 22 .. 39 % by wt (-20 ... 50 C)-08- HNO3 35 ... 96 % by wt (-20 ... 50 C)
-09- H2SO4 32 ... 84 % by wt (-17 ... 115 C)
-10- NaOH 18 ... 50 % by wt (0 ... 100 C)
See graphs in the Appendix Pg 115 and follow ing
101
Operating modes Entry of cell factor with simultaneous displayof conductivity and temperature
Input of onductivity of calibration solution withsimultaneous display of cell factor andtemperature
Product calibration
Zero point adjustment
Temperature probe adjustment
Adm. cell factor 00.100 ... 19.999
Adm. t ransfer rat io 01.00 ... 199.99
Adm. zero point deviation 0.5 mS/cm
Sensor monitoring Monitoring of primary and wiringSensocheck for short circuit
Monitoring of secondary and wiringfor open circuit
Sensoface Provides information on the sensor condition(evaluation of zero point, Sensocheck)
Sensor monitor Sensor monitor for validation of sensor andcomplete measured-value processing(Display: resistance / temperature)
Temperat ure input * Pt100 / Pt1000 / NTC 100 kOhms2-wire connection, adjustable
Ranges Pt100 / Pt1000: -20 .. +200 C(-4 ... +392 F)
NTC100 kOhms -20 ... +130 C
(-4 ... +266 F)Resolution 0.1 C / 1 F
Measurement error1,2,3)
0.5 K(
8/9/2019 2402 condi
52/65
Stratos e 2402 CondI102
Function Switches device to HOLD mode
Switching voltage 0 ... 2 V (AC/DC) Hold inact ive10 ... 30 V (AC/DC) Hold active
CONTROL input Galv. separated (OPTO coupler)Function Switch-over to second parameter set
Switching voltage 0 ... 2 V (AC/DC) Parameter set 1
10 ... 30 V (AC/DC) Parameter set 2
Output 1 0/4 to 20 mA, max. 10 V, floating(galv. connected to output 2)
Measured variable * Conductivity, concentration or salinity
Characteristic Linear or logarithmic
Overrange* 22 mA in the case of error messages
Output filter * (attenuation) Low-pass, filter time constant 0 ... 120 s
Measurement error 1) < 0.3 % current value + 0.05 mA
Start /end of scale As desired within range
Min. span LIN: 5 % of selected rangeLOG: 1 decade
Output 2 0/4 ... 20 mA, max. 10 V, floating(galv. connected to output 1)
Process variable Temperature
Overrange * 22 mA in the case of temp error messages
Output filter * Low-pass, filt er time constant 0 ... 120 s
Meas. error 1) < 0.3 % current value + 0.05 mA
Start/end of scale * -20 to +200 C / -4 ... +392 F
Adm. span 20 ... 320 K (36 to 608 F)
Alarm contact Relay contact, float ing
Contact rat ings AC< 250 V / < 3 A / < 750 VADC< 30 V / < 3 A / < 90 W
Contact response N/C (fail-safe type)Alarm delay 0000 ... 0600 s
103
(see PID process controller)
Contacts R1, R2 floating but inter-connected
Contact ratings * AC< 250 V / < 3 A / < 750 VADC< 30 V / < 3 A / < 90 W
Contact response * N/O or N/C
Delay * 0000 ... 9999 s
Switching points
*
As desired within rangeHysteresis * 0 ... 50 % full scale
PID process controller Output via relay contacts R1, R2(see limit values)
Setpoint * As desired within range
Neutral zone * As desired within range
Proport ional action * Controller gain KC: 0010 ... 9999 %
Integral action* Reset time TR: 0000 ... 9999 s
(0000 s = no integral action)
Derivative action * Rate time TD: 0000 ... 9999 s
(0000 s = no derivative action)
Controller type * Pulse length or pulse frequency controller
Pulse period * 0001 ... 0600 s, min. ON time 0.5 s(pulse length controller)
Max. pulse frequency * 0001 ... 0180 min-1
(pulse f requency controller)
Cleaning function / Parameter set 2 *
Clean / PSet2 Relay contact, floating,for controlling a rinsing probe orsignaling that 2nd parameter set is active
Contact ratings AC< 250 V / < 3 A / < 750 VADC< 30 V / < 3 A / < 90 W
Contact response N/O when signaling parameter set 2N/O or N/C when used as cleaning contact *
Rinsing interval * 000.0 ... 999.9 h(000.0 h = cleaning function switched off)
Rinse duration * 0000 ... 1999 s
Specifications
Display LC display, 7-segment with icons Protection against Protective separation of all extra-low-voltage circuits againstelectrical shock mains by double insulation as per EN 61010 1
8/9/2019 2402 condi
53/65
Stratos e 2402 CondI104
Main display Character height 17 mm, unit symbols 10 mm
Secondary display Character height 10 mm, unit symbols 7 mm
Sensoface 3 status indicators (friendly, neutral, sad Sensoface)
Mode indicators 5 status bars meas , cal, alarm , cleaning ,
config
18 further icons for configuration and messages
Alarm indication Red LED in case of alarm or HOLD, user defined
Keypad 5 keys: [cal] [conf] [ ] [ ] [enter]
Service f unctions
Current source Current specifiable for output 1 and 2 (00.00 to 22.00mA)
Manual controller Controller output entered directly (start of control process)
Device self-test Automatic memory test (RAM, FLASH, EEPROM)
Display test Display of all segment s
Last Error Display of last error occurred
Sensor monitor Display of direct sensor signal (resistance/temperature)
Relay test Manual cont rol of the four switching contacts
Parameter sets * Two selectable parameter sets for differentprocess phasesSwitchover via CONTROL input or manuallySignaling via relay contact PSet2
Data retention Parameters and calibration data > 10 years (EEPROM)
105
electrical shock mains by double insulation as per EN 61010-1
Pow er supply 24 (-15%) to 230 V AC/DC (+10%); approx. 5 VA, 2.5 WAC: 45 to 65 Hz; Overvoltage category II, Class II
Nominal operating conditionsAmbient temperature -20 to +55 CTransport/Storage temp -20 to +70 C
Relative humidity 10 to 95 % not condensingPower supply 24 (-15%) to 230 V AC/DC (+10%)Frequency for AC 45 to 65 Hz
EMC EN 61326Emitted interference Class B (residential environment)
Class A for mains supply > 60 V DCImmunity tointerference Industrial environment
Explosion prot ectionFM: NI Class I Div 2 Group A, B, C & D, T4 Ta = 55 C; Type 2
NI Class I Zone 2 Group IIC, T4 Ta = 55C; Type 2
CSA: Class I Div 2 Groups A, B, C and D, T4Ex nA IIC T4
Enclosure molded enclosure made of PBT (polybutylene terephtalate)Color Bluish gray RAL 7031Assembly Wall mounting
Pipe mounting: dia 40 to 60 mm, 30 to 45 mm Panel mounting, cutout to DIN 43 700
Sealed against panelDimensions H 144 mm, B 144 mm, T 105 mmIngress protection IP 65 / NEMA 4XCable glands 3 breakthroughs for cable glands M20x1.5
2 breakthroughs for NPT 1/2 orRigid Metallic Conduit
Weight Approx. 1 kg
* ) User-defined1) To IEC 746 Part 1, at nominal operati ng condit ions2) 1 count3) Plus sensor error
Explosion protectionWarnings and notesto ensure safe operation
8/9/2019 2402 condi
54/65
106 Stratos 2402 CondI 107
Warning: Do not disconnect equipment unless power hasbeen switched off.
Warning: Clean only with antistatic moistened cloth.Warning: Substitution of components may impair suitability
for hazardous locations. The equipment shall be installed and protected from mechanical
impact and ultraviolet (UV) sources. Clean only with a moistened antistatic cloth as potential
electrostatic hazard may exist. Service equipment only withconductive clothing, footwear and personal grounding devicesto prevent electrostatic accumulation.
Internal grounding provisions shall be provided for field wiring.Bonding between conduit shall be provided during installation,and all exposed non-current carrying metallic parts shall bebonded and grounded.
Installation in a Class I, Division 2 or Class I, Zone 2 hazardous
location shall be in accordance with the Canadian ElectricalCode (CEC Part 1) Section 18 Division 2 wiring methods.
The equipment shall have a switch or circuit breaker in thebuilding installation (that is in close proximity to the equipment)that is marked as the disconnect switch.
The enclosure Type 2 is only for indoor use. The mains supply voltage fluctuations should not exceed
-15/+10 percent of the nominal supply voltage. Do not use the equipment in a manner not specified in this
documentation.
Caution: Use supply wires suitable for 30 C above ambientand rated at least 250 V.
Caution: Use signal wires suitable for at least 250 V.
SPECIAL INSTRUCTIONS FOR FIELD REPRESENTATIVESNone
8/9/2019 2402 condi
55/65
108 Stratos 2402 CondI 109
8/9/2019 2402 condi
56/65
110 Stratos 2402 CondI 111
Calibration solutions
Potassium chloride solutions (conductivity in mS/cm) Sodium chloride solutions (conductivity in mS/cm)
8/9/2019 2402 condi
57/65
Stratos 2402 CondI112
Tempera ture Concentration
[C] 0.01 mol/l 0.1 mol/l 1 mol/l
0510
151617181920212223242526
27282930313233343536
0.7760.8961.020
1.1471.1731.1991.2251.2511.2781.3051.3321.3591.3861.4131.441
1.4681.4961.5241.5521.5811.6091.6381.6671.696
7.158.229.33
10.4810.7210.9511.1911.4311.6711.9112.1512.3912.6412.8813.13
13.3713.6213.8714.1214.3714.6214.8815.1315.3915.64
65.4174.1483.19
92.5294.4196.3198.22100.14102.07104.00105.94107.89109.84111.80113.77
115.74
Data source: K. H. Hellwege (Editor), H. Landolt, R. Brnstein: Zahlenwerte undFunktionen ..., volume 2, part. volume 6
113
012
3456789
1011121314151617
181920212223242526272829303132
33343536
134.5138.6142.7
146.9151.2155.5159.9164.3168.8173.4177.9182.6187.2191.9196.7201.5206.3211.2
216.1221.0226.0231.0236.1241.1246.2251.3256.5261.6266.9272.1277.4282.7288.0
293.3298.7304.1309.5
5.7865.9656.145
6.3276.5106.6956.8817.0687.2577.4477.6387.8318.0258.2218.4188.6178.8169.018
9.2219.4259.6319.838
10.04710.25810.46910.68310.89811.11411.33211.55211.77311.99512.220
12.44512.67312.90213.132
0.6310.6510.671
0.6920.7120.7330.7540.7750.7960.8180.8390.8610.8830.9050.9270.9500.9720.995
1.0181.0411.0641.0871.1111.1351.1591.1831.2071.2321.2561.2811.3061.3311.357
1.3821.4081.4341.460
Temperat ure Concentrat ion
[C] 0.01 mol/l * 0.1 mol/l * saturated **
* Data source: Test solutions calculated according to DIN IEC 746-3* * Data source: K. H. Hellwege (Editor), H. Landolt, R. Brnstein:
Zahlenwerte und Funktionen ..., volume 2, part. volume 6
8/9/2019 2402 condi
58/65
-02- Hydrochloric acid solution HCl-07-
-03- Sodium hydroxide solution NaOH-10-
8/9/2019 2402 condi
59/65
Stratos e 2402 CondI116
Conductivity in dependence on substance concentration and processtemperature for hydrochloric acid (HCl)Source: Haase/Sauermann/Dcker; Z. phys. Chem. New Edition, Vol. 47(1965)
07
-02- -07-
Concentrat ion measurement not possible in this range.c [% by wt]
117
Conductivity in dependence on substance concentration andprocess temperature for sodium hydroxide solution (NaOH)
-10-
Concentration measurement not possible in this range.
-03-
c [% by wt]
-10-
-04- Sulfuric acid H2SO4-06-
-05- Nitric acid HNO308
8/9/2019 2402 condi
60/65
118 Stratos e 2402 CondI
Conductivity in dependence on substance concentration and processtemperature for sulfuric acid (H2SO4)
Source: Darling; Journal of Chemical and Engineering Data;Vol. 9 No. 3, July 1964
-06--09-
-04- -09- -06-
Concentration measurement not possible in this range.
c [% by wt]
119
Conductivity in dependence on substance concentration and processtemperature for nitric acid (HN03)
Source: Haase/Sauermann/Dcker; Z. phys. Chem. New Edition, Vol. 46(1965)
-08-
-05- -08-
Concentration measurement not possible in this range. c [% by wt]
22 mA signal for error message . . . . . . . . . . . . . . . . . 47, 53Alarm settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Index
Use of relays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .59Configuration: Output 1 . . . . . . . . . . . . . . . . . . . . . . . . . . 32
8/9/2019 2402 condi
61/65
gAlarm contact . . . . . . . . . . . . . . . . . . . . . . . . . . .56, 102Alarm delay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .92Operating states . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .94
Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Cal Info . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Calibration by input of cell factor . . . . . . . . . . . . . . . . .74Calibration with calibration solution . . . . . . . . . . . . . . .76Product calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . .78Temperature probe adjustment . . . . . . . . . . . . . . . . . . .84Zero calibration in air . . . . . . . . . . . . . . . . . . . . . . . . . .80Zero calibration with calibration solution . . . . . . . . . . .82
Calibration solutions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112Concentration curves . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Hydrochloric acid solut ion . . . . . . . . . . . . . . . . . . . . .116Nitric acid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .119Sodium chloride solut ion . . . . . . . . . . . . . . . . . . . . . .115Sodium hydroxide solut ion . . . . . . . . . . . . . . . . . . . . .117Sulfuric acid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .118
Concentration measurement . . . . . . . . . . . . . . . . . . . . . . 114Concentration curves . . . . . . . . . . . . . . . . . . . . . . . . .115Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38Ranges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39, 114
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28Configuration steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30Configuration: Alarm settings . . . . . . . . . . . . . . . . . . . . . . 56Configuration: Controller . . . . . . . . . . . . . . . . . . . . . . . . . 62Configuration: Limit function . . . . . . . . . . . . . . . . . . . . . . 58
Sett ings for relay 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . .59Sett ings for relay 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . .61
120 Stratos e 2402 CondI
g pConcentration measurement . . . . . . . . . . . . . . . . . . . .38LIN curve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40LOG curve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .42Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . .36Output current during Error . . . . . . . . . . . . . . . . . . . . .46
Output current range . . . . . . . . . . . . . . . . . . . . . . .40, 42Output signal during HOLD . . . . . . . . . . . . . . . . . . . . .47Select sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32Select t emp probe . . . . . . . . . . . . . . . . . . . . . . . . . . . .34Time constant of output filter . . . . . . . . . . . . . . . . . . .44
Configuration: Output 2 . . . . . . . . . . . . . . . . . . . . . . . . . . 48Output current during HOLD . . . . . . . . . . . . . . . . . . . .52Output current range . . . . . . . . . . . . . . . . . . . . . . . . . .49Temperature error . . . . . . . . . . . . . . . . . . . . . . . . . . . . .52Temperature unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48
Time constant of output filter . . . . . . . . . . . . . . . . . . .50Configuration: Output Clean/PSEt2 . . . . . . . . . . . . . . . . . . 64Control of rinsing probe . . . . . . . . . . . . . . . . . . . . . . . .64Signaling parameter set 2 . . . . . . . . . . . . . . . . . . . . . . .64
Configuration: Temperature compensation . . . . . . . . . . . . 54Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Rinsing probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .65Special cable for SE 655 / SE 656 sensors . . . . . . . . . . .16Typical wirings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
Controller characteristic . . . . . . . . . . . . . . . . . . . . . . . . . . 88Controller configuration . . . . . . . . . . . . . . . . . . . . . . . . . . 62Controller equations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89Controller test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87Controller types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88Controlling a rinsing probe . . . . . . . . . . . . . . . . . . . . . . . . 65CSA Certif icate of Compliance . . . . . . . . . . . . . . . . . . . . 106
121
Current source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86Diagnostics functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Switchover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66User settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70
8/9/2019 2402 condi
62/65
Controller test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .87Display of calibration data . . . . . . . . . . . . . . . . . . . . . .85Display of last error message . . . . . . . . . . . . . . . . . . . .85Display of output currents . . . . . . . . . . . . . . . . . . . . . .85Relay test (manual test of contacts) . . . . . . . . . . . . . . .86
Sensor monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .85Specify current for output 1/2 . . . . . . . . . . . . . . . . . . .86
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23Division 2 wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15EC Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . . 7Err . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28Error Info . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92Hold mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
External activation . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
LED in HOLD mode . . . . . . . . . . . . . . . . . . . . . . . . . . .57Output signal for HOLD . . . . . . . . . . . . . . . . . . . . .47, 53Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24Limit function (LiMIT) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Mount ing plan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Operating states . . . . . . . . . . . . . . . . . . . . . . . .