ZIV ZIVE Electrochemical Workstations Electrochemical Workstations ZIVE BP2A/BP2C ZIVE BP2A/BP2C ZIVE BP2A/BP2C ZIVE BP2A/BP2C Designing the Solution for Electrochemistry Potentiostat/Galvanostat І Battery Cycler І Fuel Cell Test Station +82-2-578-6516 І [email protected] www.wonatech.com І www.zivelab.com І www.fctest.com
24

ZIVE Electrochemical Workstations - 제목 없음 Electrochemical Workstations ... Current range 12 ranges (auto/manual setting) ... • Independent operation by DSP with FPGA

Mar 24, 2018

Download

Documents

ngotu
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: ZIVE Electrochemical Workstations - 제목 없음 Electrochemical Workstations ... Current range 12 ranges (auto/manual setting) ... • Independent operation by DSP with FPGA

ZIVZIVEE Electrochemical WorkstationsElectrochemical Workstations

ZIVE BP2A/BP2CZIVE BP2A/BP2CZIVE BP2A/BP2CZIVE BP2A/BP2C

Designing the Solution for ElectrochemistryPotentiostat/Galvanostat І Battery Cycler І Fuel Cell Test Station+82-2-578-6516 І [email protected] І www.zivelab.com І www.fctest.com

Page 2: ZIVE Electrochemical Workstations - 제목 없음 Electrochemical Workstations ... Current range 12 ranges (auto/manual setting) ... • Independent operation by DSP with FPGA

FeaturesZIVE BP2/BP2CZIVE BP2/BP2C

Page 3: ZIVE Electrochemical Workstations - 제목 없음 Electrochemical Workstations ... Current range 12 ranges (auto/manual setting) ... • Independent operation by DSP with FPGA

FeaturesZIVE BP2/BP2CZIVE BP2/BP2C

• 1MHz EIS capability potentiostat/galvanostat/impedance analyzerp y p /g / p y• Compact size with full functions• Wide current range(2A to 20pA) for various applications• Supports fourteen(14) kind of EIS techniques• Current interrupt IR measurement/IR compensation• Pulse charge/discharge test available• Bipolar pulse capability• High speed data sampling time• High speed data sampling time

- 50usec/sample in burst mode- 1msec/sample in normal mode- 2usec/sample in fast sweep modeI t l 350 000 d t i t t• Internal 350,000 data point storage

• Bi potentiostat application available (BP2C)• Full software packages as standard

- Energy software package (BAT)Energy software package (BAT)- EIS software package (EIS)- Electrochemical analysis software package(EAS)- Corrosion software package (COR)

• Free software package• Free software package

Page 4: ZIVE Electrochemical Workstations - 제목 없음 Electrochemical Workstations ... Current range 12 ranges (auto/manual setting) ... • Independent operation by DSP with FPGA

SpecificationZIVE BP2ZIVE BP2

System

Cell cable 1 meter shielded type(standard)working, reference, counter, working sense,

Control DSP with FPGA

Calibration Automatic

Data acquisitionADCDAC

2x16 bit ADCs(500kHZ) for voltage, current1x16bit ADCs(250kHz) for auxiliary reading2x16bit DAC(50MHz) for bias & scan

Filter selection 4ea (5Hz 1kHz 500kHz 5MHz)Filter selection 4ea (5Hz, 1kHz, 500kHz, 5MHz)

Scan rate 0~5000V/sec@10mV sampling interval

Max channel no 32 channels via USB connectionMax. channel no. 32 channels via USB connection

Internal data memory 350,000 points

Page 5: ZIVE Electrochemical Workstations - 제목 없음 Electrochemical Workstations ... Current range 12 ranges (auto/manual setting) ... • Independent operation by DSP with FPGA

SpecificationZIVE BP2ZIVE BP2

Power Amplifier(CE)

Power 24Watt (12V@2A)

Compliance voltage ±12VCompliance voltage ±12V

Max. current ±2A

Control speed selection 8eap

Bandwidth 5MHz

Slew rate 10V/usec

Page 6: ZIVE Electrochemical Workstations - 제목 없음 Electrochemical Workstations ... Current range 12 ranges (auto/manual setting) ... • Independent operation by DSP with FPGA

SpecificationZIVE BP2ZIVE BP2

Potentiostat Mode (Voltage Control)

Voltage Control

Control voltage range ±10V, ±1V, ±100mV

Voltage resolution 16 bit per each range

Voltage accuracy ±1mV, ±0.05% of setting(gain x1)

Max scan range ±10V vs ref EMax. scan range ±10V vs. ref. E

Current Measurement

Current range 12 ranges (auto/manual setting)20pA~2A20pA & 200pA with gain

Current resolution 16 bit60uA, 6uA, 600nA, 60nA, 6nA, 600pA,60uA, 6uA, 600nA, 60nA, 6nA, 600pA, 60pA, 6pA,600fA,60fA(6fA, 0.6fA with gain)

Current accuracy ±0.1% f.s. ±10pA (gain x1)>200nA f.s.

Page 7: ZIVE Electrochemical Workstations - 제목 없음 Electrochemical Workstations ... Current range 12 ranges (auto/manual setting) ... • Independent operation by DSP with FPGA

SpecificationZIVE BP2ZIVE BP2

Galvanostat Mode (Current Control)

Current Control

Control current range max. ±2A± full scale depending on selected range± full scale depending on selected range

Current resolution 16 bit60uA, 6uA, 600nA, 60nA, 6nA, 600pA, 60pA, 6pA,600fA,60fA(6fA 0 6fA ith i )(6fA, 0.6fA with gain)

Current accuracy ±0.1% f.s. ±10pA (gain x1)>200nA f.s.

Voltage Measurement

Voltage range 10V, 1V, 100mV

Voltage resolution 16 bit0.3mV, 30uV, 3uV0.3mV, 30uV, 3uV

Voltage accuracy ±1mV ±0.05% of reading(gain x1)

Page 8: ZIVE Electrochemical Workstations - 제목 없음 Electrochemical Workstations ... Current range 12 ranges (auto/manual setting) ... • Independent operation by DSP with FPGA

SpecificationZIVE BP2ZIVE BP2

EIS(Internal FRA) for System

Frequency range 10uHz ~ 1MHz

Frequency accuracy 0.01%

Frequency resolution 5000/decade

Amplitude 0.1mV ~ 5V rms (potentiostatic)0.1 ~ 70% f.s. (Galvanostatic)

Mode Static EIS:Potentiostatic GalvanostaticPotentiostatic, Galvanostatic,Pseudogalvanostatic, OCP

Dynamic EIS:Potentiodynamic, Galvanodynamic

Fixed Frequency Impedance:Fixed Frequency Impedance:Potentiostatic, Galvanostatic,Potentiodynamic, Galvanodynamic

Multisine EIS:Potentiostatic GalvanostaticPotentiostatic, Galvanostatic

Intermittent PEIS/GEIS

Page 9: ZIVE Electrochemical Workstations - 제목 없음 Electrochemical Workstations ... Current range 12 ranges (auto/manual setting) ... • Independent operation by DSP with FPGA

SpecificationZIVE BP2ZIVE BP2

Electrometer

M i t lt ±10VMax. input voltage ±10V

Input impedance 2x1013 Ω ιι 4.5pF

Bandwidth >22MHz

CMRR >114dB

Interface for System

Auxiliary port

Auxiliary voltage input(3channel) 3 analog input: ±10V

Auxiliary voltage output 1 analog output : ±10V

Zero resistance ammeter 2nA ~ 2A

DI/DO port DO: 3 DI: 2

External booster interface Via booster I/F cable

Peripheral communication I2C

Signal Generator output 1 analog output for FRA

Temperature input K type thermocoupleTemperature input K type thermocouple (default: max. 250°C)

Page 10: ZIVE Electrochemical Workstations - 제목 없음 Electrochemical Workstations ... Current range 12 ranges (auto/manual setting) ... • Independent operation by DSP with FPGA

SpecificationZIVE BP2ZIVE BP2

SoftwareSoftware

Max. step per experiment 1000

Shutdown safety limits voltage, current, temperature, AuxV, etc.

Max. sampling rate 20kHz(50usec) in burst mode500kHz(2usec) in fast sweep mode

Max. sampling time unlimitedp g

Sampling condition Time, dv/dt, dI/dt, temperature, etc.

Main Systemy

PC communication USB 2.0 high speed

Line voltage 100~240VAC, 50/60Hz, 1Amp

Size / Weight (8channel) 209x270x378mm(WxDxH) / 9.25kg

Max. output power 24Watt

Page 11: ZIVE Electrochemical Workstations - 제목 없음 Electrochemical Workstations ... Current range 12 ranges (auto/manual setting) ... • Independent operation by DSP with FPGA

SpecificationZIVE BP2ZIVE BP2

PC RequirementPC Requirement

Operating system Windows XP SP3/7/8/10(32bit/64bit OS)

PC ifi ti P ti 4 RAM 1GB hi hPC specification Pentium4, RAM 1GB or higher

Display 1600x900 high color or higher

USB High speed 2.0g p

General

Dummy cell One external dummy cell included

Impedance data analysis software ZMANTM software

DC data analysis software IVMANTM software

Page 12: ZIVE Electrochemical Workstations - 제목 없음 Electrochemical Workstations ... Current range 12 ranges (auto/manual setting) ... • Independent operation by DSP with FPGA

AppearanceFront ViewFront View

Internal FRA includedFPGA/DSP l

Aux port includes3 analog input (auxiliary voltage measurement)1 analog output

FPGA/DSP controlPlug-In type

1 analog output2 digital input3 digital output

Misc port includes ;I2C com port for external device control

Temperature measurement (K type) Thermocouple (Option) control

FRA port1 Sig generator output1 Voltage input1 Current inputCell port

Working, Counter,ZRA ,Working sense,

(Option)

gReference

Page 13: ZIVE Electrochemical Workstations - 제목 없음 Electrochemical Workstations ... Current range 12 ranges (auto/manual setting) ... • Independent operation by DSP with FPGA

AppearanceLCD Display – DC ModeLCD Display DC Mode

Page 14: ZIVE Electrochemical Workstations - 제목 없음 Electrochemical Workstations ... Current range 12 ranges (auto/manual setting) ... • Independent operation by DSP with FPGA

AppearanceLCD Display – EIS ModeLCD Display EIS Mode

Page 15: ZIVE Electrochemical Workstations - 제목 없음 Electrochemical Workstations ... Current range 12 ranges (auto/manual setting) ... • Independent operation by DSP with FPGA

AppearanceRear ViewRear View

Air flowing fan

Power switch

USB portp

Power connector

Page 16: ZIVE Electrochemical Workstations - 제목 없음 Electrochemical Workstations ... Current range 12 ranges (auto/manual setting) ... • Independent operation by DSP with FPGA

System Configuration

• System setup with multiple ZIVE BP2s• Connecting by USB hubg y

• System setup with one or more ZIVE product family• Connecting by USB hub

Page 17: ZIVE Electrochemical Workstations - 제목 없음 Electrochemical Workstations ... Current range 12 ranges (auto/manual setting) ... • Independent operation by DSP with FPGA

When communication fails…

• When communication error occurs between ZIVE BP2 and PC the data will• When communication error occurs between ZIVE BP2 and PC, the data will be stored on the device memory.

• The data will be transferred to the PC after PC is restored.

Page 18: ZIVE Electrochemical Workstations - 제목 없음 Electrochemical Workstations ... Current range 12 ranges (auto/manual setting) ... • Independent operation by DSP with FPGA

Strong Point (Hardware)

• Wide current rangeg: Max. 2 Amp: Suitable for various application

• 3 measurement/control voltage range &12 measurement/control current range12 measurement/control current range

• Independent operation by DSP with FPGA

• Built-in FRA• Built-in FRA : Impedance measurements available: No other device is needed.

• Compact size and light weightp g g: Small footprint

• Big instrument internal memory capacity: 350,000 data point can be stored.

• Three auxiliary voltage measurement available

• Temperature measurement available• Auxiliary voltage output• Auxiliary voltage output• DI/DO ports

Page 19: ZIVE Electrochemical Workstations - 제목 없음 Electrochemical Workstations ... Current range 12 ranges (auto/manual setting) ... • Independent operation by DSP with FPGA

FRA Function

FRA function is to control external electronic load or potentiostat for EIS measurement

•FRA potentiostatic EIS•FRA galvanostatic EIS•FRA potentiostatic HFR•FRA galvanostatic HFR

Page 20: ZIVE Electrochemical Workstations - 제목 없음 Electrochemical Workstations ... Current range 12 ranges (auto/manual setting) ... • Independent operation by DSP with FPGA

Multi Working Electrodes(BP2C)g ( )

The BP2C is designed for the test with multi working electrodes The BP2C is designed for the test with multi working electrodes sharing one reference electrode and one counter electrode.

BP2A BP2C

Page 21: ZIVE Electrochemical Workstations - 제목 없음 Electrochemical Workstations ... Current range 12 ranges (auto/manual setting) ... • Independent operation by DSP with FPGA

Applicable FieldMarket For the ProductMarket For the Product

Other Application- Coating evaluation- Heavy metal detection- Semiconductor- EDLC- Membrane- Bioelectrochemistry- Material testing etc.g

Page 22: ZIVE Electrochemical Workstations - 제목 없음 Electrochemical Workstations ... Current range 12 ranges (auto/manual setting) ... • Independent operation by DSP with FPGA

Ordering Guide

Zi BP2Zive BP2- Comes with following items as standard

a) Cell cable with alligator clips, 2eab) Dummy cell, 1ea) d) P d 1

Standard Systemc) d) Power cord, 1ead) USB cable, 1eae) Install CD, 1ea

Software Package- EIS software package (EIS)- Energy software package (BAT)- Corrosion software package (COR)- Electrochemical analysis software package (EAS)-- FRA function for external load’s and external potentiostat’s EIS

K type thermocouple Optional ItemsK type thermocoupleAux cable, ZRA cable, & FRA cableOptional Booster I/F Cable, 1.5m

Optional Items

Page 23: ZIVE Electrochemical Workstations - 제목 없음 Electrochemical Workstations ... Current range 12 ranges (auto/manual setting) ... • Independent operation by DSP with FPGA

Contact Information

+82-2-578-6516

[email protected]

+82-2-578-6516+82 2 578 6516

[email protected]

Page 24: ZIVE Electrochemical Workstations - 제목 없음 Electrochemical Workstations ... Current range 12 ranges (auto/manual setting) ... • Independent operation by DSP with FPGA

Thank o !Thank you !