THEMIS Instrument FM2/FM3 PER FGM 1 Teleconference 8/30/2005 THEMIS Instrument FM2/3 Pre-Environmental Review Fluxgate Magnetometer (FGM) Dr Ellen Taylor University of California - Berkeley
4

THEMIS Instrument FM2/FM3 PERFGM 1 Teleconference 8/30/2005 THEMIS Instrument FM2/3 Pre-Environmental Review Fluxgate Magnetometer (FGM) Dr Ellen Taylor.

Jan 14, 2016

Download

Documents

Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: THEMIS Instrument FM2/FM3 PERFGM 1 Teleconference 8/30/2005 THEMIS Instrument FM2/3 Pre-Environmental Review Fluxgate Magnetometer (FGM) Dr Ellen Taylor.

THEMIS Instrument FM2/FM3 PER FGM 1 Teleconference 8/30/2005

THEMIS

Instrument FM2/3

Pre-Environmental Review

Fluxgate Magnetometer (FGM)

Dr Ellen Taylor

University of California - Berkeley

Page 2: THEMIS Instrument FM2/FM3 PERFGM 1 Teleconference 8/30/2005 THEMIS Instrument FM2/3 Pre-Environmental Review Fluxgate Magnetometer (FGM) Dr Ellen Taylor.

THEMIS Instrument FM2/FM3 PER FGM 2 Teleconference 8/30/2005

Design Verification Design parameters same as FM1

FGE/FGS Calibration before delivery to UCB 8 test/calibration steps performed in Berlin, Braunschweig, Jeserigerhütten and

Graz

FGS Sensor Environmental Qualification FM2/FM3 (SN01/SN02) Vibration and TV test levels Parameters checked before and after tests

FGE Electronics Environmental Qualification FM2/FM3 (SN01/SN02) Vibration and TV testing with IDPU Suite CPT done before and after tests

No PFRs or modifications since FM1 PER

FGM FM2/FM3 Overview

Page 3: THEMIS Instrument FM2/FM3 PERFGM 1 Teleconference 8/30/2005 THEMIS Instrument FM2/3 Pre-Environmental Review Fluxgate Magnetometer (FGM) Dr Ellen Taylor.

THEMIS Instrument FM2/FM3 PER FGM 3 Teleconference 8/30/2005

Design Verification All requirements in MRD verified

Calibration Tests Digital functional test - check of board population, FPGA and current sources Analogue Adjustment of FGS and FGE - Test of excitation,input filter, phase

adjustment, sensitivity and noise Parameters vs. Electronics Temperature - Test of offset, sensitivity, inrush

current, phase in dependency on e-temp. Parameters vs. Sensor Temperature - Test of offset, sensitivity and noise in

dependency on sensor temperature Test of non-linearity - Test of DAC non linearity and filter characteristics Earth field registration - Test of resolution and stability by comparison of F4

and F6 measuring Earth field variations Determination of sensor axis direction - Direction of magnetic axis determined

with respect to mechanical ref. Frame Determination of scale value and orthogonality - Scale values and

orthogonality check vs. sensor temp. in coil system

FM2/FM3 Performance

Page 4: THEMIS Instrument FM2/FM3 PERFGM 1 Teleconference 8/30/2005 THEMIS Instrument FM2/3 Pre-Environmental Review Fluxgate Magnetometer (FGM) Dr Ellen Taylor.

THEMIS Instrument FM2/FM3 PER FGM 4 Teleconference 8/30/2005

Vibration Specs same as FM1 FM2/FM3 Vibration Test in Berlin: 6/28/05 to 7/2/04 Sine wave: up to 6.4g @ 50Hz Random noise: 11.7g RMS

TV-cycle same as FM1 FM2/FM3 TV Test Performed in Graz: 7/01/04 to 7/03/04 Two cycles between -120°C and 90°C

Parameter check (before, between and after tests) same as FM1 Resistance of all coils / Sensor output signals Coil constants and mechanical alignment of feedback coil system

FM2/FM3 FGS Envir.