Results Review of Microphone Prototype and LabView Detecting of Human Errors Algorithm Author: Minh Anh Nguyen Colorado State University School of Biomedical and Electrical Engineering Email: [email protected]
Feb 08, 2017
PowerPoint Presentation
Results Review of Microphone Prototype and LabView Detecting of Human Errors AlgorithmAuthor: Minh Anh NguyenColorado State University School of Biomedical and Electrical EngineeringEmail: [email protected]
1
Outline of the reviewUpdate on tasks from the previous meetingMicrophone prototypeLabView Results for detecting methods/algorithms for audio files
2
Update on labView Need: Vision and Motion package to analyze the images Issue: The Vision and Motion programming/function package is missing or not available for the Labview 2014 and NI myRio softwareThis issue is resolved
3
Image testSimulation modeMake sure image files are loaded into the workspace without any issues.Make sure a correct file is loadedPlot histogram of the imageVerify that the histogram value is correct.Positive test: load two image files which have 2 different images and verify that the test results show the difference.Negative test: load two image files which have 2 similar images and verify that the test results show the same thing or match.
4
Labview results for comparing two images which are identical
5
LabView results for comparing two images which are not identical
6
LabView results for comparing two images which are not identical
7
Audio testSimulation modeMake sure audio files are loaded into the workspace without any issues.Plot and listen to audio files; make sure a correct file is loaded.Verify that the FFT value is correct.Positive test: load two audio files which have 2 different signals and verify that the test results show the difference.Negative test: load two audio files which have 2 similar signals and verify that the test result show the same thing or match.
8
Microphone prototype for correct setting
9
LabView results for comparing two audio files which are identical
10
Microphone prototype for hex pin not popped out
11
Result of hex pin not popped out after executing the device for 1 min
12
LabView results for comparing two audio files (hex pin) which are not identical
13
Microphone prototype for upper arm not correctly set
14
LabView results for comparing two audio files (upper arm) which are not identical
15
Microphone prototype for upper arm not correctly set
16
LabView results for comparing two audio files (upper arm) which are not identical
17
Microphone prototype for lower bearing not correctly set
18
LabView results for comparing two audio files (lower bearing) which are not identical
19