Top Banner
Preliminary Experience in the Use of an Extracellular Matrix (CorMatrix) as a Tube Graft: Word of Caution Narutoshi Hibino, Patrick McConnell, Toshiharu Shinoka, Mahim Malik, Mark Galantowicz Department of Cardiothoracic Surgery, Nationwide Children’s Hospital
15

Preliminary Experience in the Use of an Extracellular Matrix (CorMatrix) as a Tube Graft: Word of Caution Narutoshi Hibino, Patrick McConnell, Toshiharu.

Jan 13, 2016

Download

Documents

Hector Russell
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Preliminary Experience in the Use of an Extracellular Matrix (CorMatrix) as a Tube Graft: Word of Caution Narutoshi Hibino, Patrick McConnell, Toshiharu.

Preliminary Experience in the Use of an Extracellular Matrix (CorMatrix) as a Tube Graft: Word of Caution

Narutoshi Hibino, Patrick McConnell, Toshiharu Shinoka, Mahim Malik,

Mark Galantowicz

Department of Cardiothoracic Surgery, Nationwide Children’s Hospital

Columbus, OH

Page 2: Preliminary Experience in the Use of an Extracellular Matrix (CorMatrix) as a Tube Graft: Word of Caution Narutoshi Hibino, Patrick McConnell, Toshiharu.

Disclosure

Patrick McConnellGrant/Research Support- Cormatrix Cardioavascular, Inc; Thoratec CorporationConsultant- Q test Laboratories; Clear Catheter Systems, Inc

Toshiharu ShinokaGrant/Research Support- Gunze, Inc

Page 3: Preliminary Experience in the Use of an Extracellular Matrix (CorMatrix) as a Tube Graft: Word of Caution Narutoshi Hibino, Patrick McConnell, Toshiharu.

• There are no ideal materials for the repair of congenital heart disease

• Decellularized extracellular matrix from porcine small intestinal submucosa (CorMatrix) has been developed and commercialized

Background

Page 4: Preliminary Experience in the Use of an Extracellular Matrix (CorMatrix) as a Tube Graft: Word of Caution Narutoshi Hibino, Patrick McConnell, Toshiharu.

Our Preliminary Experience using CorMatrix as a Tube Graft

Central PA reconstruction N=10

Aortic arch interposition N=3

Page 5: Preliminary Experience in the Use of an Extracellular Matrix (CorMatrix) as a Tube Graft: Word of Caution Narutoshi Hibino, Patrick McConnell, Toshiharu.

The CorMatrix sheet was wrapped around Hegar dilator of the required size and the edge was sutured with 7-0 prolene

Creation of CorMatrix Tube Graft

Page 6: Preliminary Experience in the Use of an Extracellular Matrix (CorMatrix) as a Tube Graft: Word of Caution Narutoshi Hibino, Patrick McConnell, Toshiharu.

Patient Diagnosis

Age at surgery (month)

Diameter of tube (mm) Intervention for tube Results

1 HLHS 9 9 Stent Fontan2 HLHS 6 7 Stent Fontan3 DILV, TGA, IAA 6 9 NA Sudden death4 HLHS 8 9 NA Fontan5 HLHS 6 9 Stent Fontan

6 DILV, TGA, Arch hypoplasia 6 8 Stent Waiting Fontan

7 HLHS 5 8 Stent Sudden death

8 HLHS 8 8Stent, Central PA reconstruction with homograft

Waiting Fontan

9 HLHS 4 8 Balloon Waiting Fontan10 HLHS 5 8 Stent Waiting Fontan

Interposition CorMatrix Tube Graft for Central PA Reconstruction in Comprehensive Stage II Surgery

Page 7: Preliminary Experience in the Use of an Extracellular Matrix (CorMatrix) as a Tube Graft: Word of Caution Narutoshi Hibino, Patrick McConnell, Toshiharu.

Pt 4

Pt 9

OP Follow UP

Patent Graft in Central PA (N=2 of 10)

Page 8: Preliminary Experience in the Use of an Extracellular Matrix (CorMatrix) as a Tube Graft: Word of Caution Narutoshi Hibino, Patrick McConnell, Toshiharu.

Pt 1

Pt 7

OP Follow Up

Stenosed Graft in Central PA (N=8 of 10)

Page 9: Preliminary Experience in the Use of an Extracellular Matrix (CorMatrix) as a Tube Graft: Word of Caution Narutoshi Hibino, Patrick McConnell, Toshiharu.

Graft Occlusion → Angioplasty, Stent →Stent Occlusion

1st Cath

2nd Cath

Page 10: Preliminary Experience in the Use of an Extracellular Matrix (CorMatrix) as a Tube Graft: Word of Caution Narutoshi Hibino, Patrick McConnell, Toshiharu.

Graft Occlusion → Stent Occlusion → Graft Replacement with PA Homograft

Explanted graft

Page 11: Preliminary Experience in the Use of an Extracellular Matrix (CorMatrix) as a Tube Graft: Word of Caution Narutoshi Hibino, Patrick McConnell, Toshiharu.

0

1

2

3

4

5

6

7

8

9

10

Dia

met

er (m

m)

OP

Diameter Change of CorMatrix Tube Graft for Central PA Reconstruction

1st cath 2nd cath 3rd cath

Page 12: Preliminary Experience in the Use of an Extracellular Matrix (CorMatrix) as a Tube Graft: Word of Caution Narutoshi Hibino, Patrick McConnell, Toshiharu.

Patient DiagnosisAge at

surgery SurgeryDiameter of tube (mm) Results

1 IAA,AP window,VSD,ASD 6 day Repair IAA/AP window/VSD/ASD 12

No stenosis/ dilatation

2 IAA,VSD,Dextrocardia,Hypoplastic right lung 16 day Repair IAA/VSD 11

No stenosis/ dilatation

3IAA,VSD,Brain trauma with

depressed skull fracture, s/p PDA stent/bilatereal PAB

5 monthRepair

IAA/VSD/ASD, Removal PDA stent/PA band

11No

stenosis/ dilatation

Interposition CorMatrix Tube Graft for Aortic Arch Repair for Interrupted Aortic Arch

Page 13: Preliminary Experience in the Use of an Extracellular Matrix (CorMatrix) as a Tube Graft: Word of Caution Narutoshi Hibino, Patrick McConnell, Toshiharu.

10

10.5

11

11.5

12

12.5

Dia

met

er (m

m)

Op FU

OP 3 mo 1 year

Page 14: Preliminary Experience in the Use of an Extracellular Matrix (CorMatrix) as a Tube Graft: Word of Caution Narutoshi Hibino, Patrick McConnell, Toshiharu.

While angioplasty using CorMatrix as an interposition tube vascular graft is feasible and safe in aorta, long follow up is required to assess the potential for growth of arterial conduit.

A high rate of occurrence of intimal hyperplasia formation with significant stenosis was found in reconstructed pulmonary arteries supplied by venous circulation. CorMatrix may not be the ideal material as a conduit in venous circulation to provide long-term durable outcomes.

Conclusion

Page 15: Preliminary Experience in the Use of an Extracellular Matrix (CorMatrix) as a Tube Graft: Word of Caution Narutoshi Hibino, Patrick McConnell, Toshiharu.

Presentation Number: 99Title: Preliminary experience in the use of an extracellular matrix (CorMatrix) as a tube graft: Word of cautionSession Name: Wednesday, April 29, 2015Session Room: 612Session Date: Congenital Heart Disease Simultaneous Scientific SessionPresentation Time: 8:45 AM - 8:56 AMPresentation Format: You will have an 5 minute presentation followed by 6 minutes of discussion.Discussant Name: David Morales