In Vivo Imaging: Xenogen’s IVIS ® 200 Series and Living Image ® Software
In Vivo Imaging:
Xenogen’s IVIS® 200 Series
and Living Image® Software
2
What will be covered?
Introduction
Science of In Vivo Imaging
Xenogen IVIS® 200 Series Hardware Overview
Living Image® Software Overview
Fluorescence System
Training
Hands on Training
3
Science
Why Optical In Vivo Imaging?
Powerful labeling technique - gene expression results in
production of luciferase
Tracer Applications: Amount of light is proportional to number of cells
Functional Applications: Light is produced in response to a stimulus
Non-invasive – does not require subject to be euthanized
Relatively simple instrumentation. Users can run it
themselves – a lab instrument, not an imaging center.
4
Luciferase Emission Spectra and Tissue Transmission
Tissue is not Transparent - Light
Absorbance Depends on Wavelength
Science
5
CCD
Optics
Bioluminescent
Source
Photons Diffuse Through Tissue and the IVIS®
Views this Signal on the Surface of the Subject
Light traveling through tissue
scatters many times creating a "fuzzy" image at the surface of the animal
The IVIS® views the diffuse image on the surface of the subject
Science
6
IVIS® Imaging System 200 Series – Hardware
Customized for in vivo imaging
High sensitivity from 300-900 nm
Large dynamic range
Living Image® software
Hardware
7
IVIS® 200 Imaging System – Hardware
Chiller and
Camera controller
Lenses
CCD camera
Heated Sample
Stage
Electronics
Filter Wheels
Hardware
8 Hardware
Alignment Light Projector
Allows rapid and reproducible positioning of subjects.
Size changes with Field of View setting
9 Software
Living Image® Software
Controls all settings in the IVIS® system
Provides advanced cataloging and browsing tools
Provides analysis tools
Instrument settings are analogous to photography
Images are acquired in a two step process
10 Software – Acquisition
Standard Images are Composed of Two Images
Photographic + Luminescent = Overlay
11
Field of View
Emission Filters
Software
Camera and Lens Settings are Analogous
to Those Used in Standard Photography
Field of View (FOV) is dependent
on the distance from the lens to
the sample
Light collected is proportional
to how long the shutter is open
(exposure time)
Aperture (f/stop) controls the
amount of light collected
CCD resolution is referred
to as Binning
12
Setting Sensitivity – Luminescent Signal Level
The IVIS® CCD camera has a signal range of 0 to 65535
Analog to Digital Counts (216).
Adjust camera settings to obtain a signal level of 600 to
60,000 counts.
Settings that control signal level are:
Exposure time
Binning (CCD Resolution)
f/stop (Aperture)
Software
13
Living Image® Control Panel
Controls Sensitivity
Affects Sensitivity
Software
14
Field-of-View
FOV B: 6.5 cm FOV A: 4.0 cm FOV C: 13 cm
FOV E: 25 cm FOV D: 19.5 cm
Software
15
Exposure Time
Signal level is proportional to exposure time
Shorter exposure time improves throughput.
Longer exposure time increases signal Min exposure time = 0.5 second
Max exposure time= 5 minutes
2 sec f/1 small binning
~5000 counts peak
10 sec f/1 small binning
~25000 counts peak
Software – Acquisition
16
f/stop (lens aperture)
f/stop controls the amount of light
received by the CCD
f/1 is wide open
Max light collection,
default for luminescent
f/8 is smallest aperture
best resolution,
default for photo
f/1 f/8
Software – Acquisition
17
Pixel Binning (CCD Resolution)
Large Binning Medium Binning Small Binning
10 seconds f/2 10 seconds f/2 10 seconds f/2
Software – Acquisition
Binning refers to the grouping of
pixels into a larger super-pixel
Large Binning = High Sensitivity/
Lower resolution
Small Binning = High Resolution/
Lower Sensitivity
18
Summary of Basic Camera Settings
Controls Sensitivity
Affects Sensitivity
Software – Acquisition
19
Acquisition – Single Image
Single Image Acquisition
Overlay will automatically take Photo + Luminescent
Software – Acquisition
20
Acquisition – Sequential mode
Allows automatic acquisition of a series
of images separated by fixed time points.
Starts Sequential
Image Acquisition
Software – Acquisition
21
Series: 3 weeks, 6 weeks
Experiment: M626
Label: M626-34, M626-61
Comment: Dorsal View, Level B
Analysis Comment:
Software – Cataloging
Image Labeling
Good labeling practices are necessary
for effective data browsing
22
Standard Label Sets for Cataloging Images
Software – Cataloging
23
Image Cataloging & Browsing Tools
Software – Cataloging
24 Software – Analysis
Regions of Interest Tools
ROI’s available are:
Contour
Circle
Square
Grid
25
2 sec f/2 Small Binning
~5000 counts peak
2.82 x 108 photons/sec
10 sec f/2 Small Binning
~25000 counts peak
2.82 x 108 photons/sec
Software - Analysis
Calibrated Physical Units
Living Image® automatically compensates for device settings: Integration
Time, f/stop, Binning, and Field of View.
Calibrated units are Photons per Second, representing the flux radiating
omni-directionally from a user defined region.
26
Measurement Table
Measurement Table displays information about each
Region of Interest in the image.
Measurement Table is user configurable and can be
exported to a spreadsheet
Software - Analysis
27
Fluorescence
Fluorophore
ground
state
Excitation
Wavelength
Emission
Wavelength excited
state
Fluorescence
28
Twenty-two position computer
controlled Emission filter wheel
Twelve position computer
controlled Excitation filter wheel
150 Watt Tungsten/Halogen lamp
with computer controlled intensity
Low Auto Fluorescence
optics and fibers
Fluorescence
IVIS® Fluorescence System
29
Fluorescence Acquisition
Lamp level
High / Low Select Fluorescent
Imaging Mode
Locks Excitation and Emission filters together
Select filters
7
Fluorescence
30 Fluorescence
Excitation and Emission Filters
400 500 600 700 800
0.1
1.0
wavelength / nm
no
rma
lize
d in
ten
sity
normalized transmission of light
through 1cm of tissue
ICG
Cy 5.5
DsRed
GFP
0.1
1
0.1
1
0.1
1
0.1
0.1
0.1
0.1
normalized
intensity
31
Instrument Background Correction
Uncorrected Image
Instrument Measurement
Corrected Image
Fluorescence
32
Fluorescent Calibrated Units: Efficiency
Emitted Light (photons/sec)
Excitation Light (photons/sec)
Efficiency =
Excitation Light Pattern
GFP Well Plate Corrected: Efficiency
GFP Well Plate Uncorrected:
Photons or Counts
Fluorescence
33
Regular Rodent Food Alfalfa Free Rodent Food
Cy5.5 Cy5.5
p/sec/cm^2/sr p/sec/cm^2/sr
Fluorescence
Animal Diet and Autofluorescence
in Control Mice
34
Alfalfa Free Rodent Food
GFP DsRed Cy5.5 ICG
p/sec/cm2/sr p/sec/cm2/sr p/sec/cm2/sr p/sec/cm2/sr
Fluorescence
Autofluorescence of Control Mice
35
Autofluorescence Background Excitation Filters
Fluorescent
Image
Animal autofluorescence
Image using background
excitation filter
Corrected
Image
Fluorescence
36
Background Excitation Filters
400 500 600 700 800 900
0.1
1
wavelength / nm
no
rma
lize
d in
ten
sity
normalized transmission of light
through 1cm of tissue
GFPBkg, GFP
0.1
1DsRed Bkg, DsRed
0.1
1
Cy 5.5 BkG, Cy 5.5
0.1
1ICG BkG, ICG
normalized
intensity
Fluorescence
37
Click # XQA20050216121658_001Wed, Feb 16, 2005 12:18:01Em filter=560nmBin:M (8), FOV12.8, f1, 10sCamera: 23235, EEV
Series: 23235Experiment: in situ calib checkLabel: xls 58 2.89e08Comment: Analysis Comment:
8000
6000
4000
2000
ImageMin = -1550.7Max = 8722.2
counts
Color BarMin = 55
Max = 8509
bkg subflat-fieldedcosmic
Click # XQA20050216121658_002Wed, Feb 16, 2005 12:18:21Em filter=580nmBin:M (8), FOV12.8, f1, 1sCamera: 23235, EEV
Series: 23235Experiment: in situ calib checkLabel: xls 58 2.89e08Comment: Analysis Comment:
8000
6000
4000
2000
ImageMin = -1563.5Max = 5442.6
counts
Color BarMin = 56
Max = 5323
bkg subflat-fieldedcosmic
Click # XQA20050216121658_003Wed, Feb 16, 2005 12:18:33Em filter=600nmBin:M (8), FOV12.8, f1, 1sCamera: 23235, EEV
Series: 23235Experiment: in situ calib checkLabel: xls 58 2.89e08Comment: Analysis Comment:
8000
6000
4000
2000
ImageMin = -1568.9Max = 8222.8
counts
Color BarMin = 57
Max = 8023
bkg subflat-fieldedcosmic
Click # XQA20050216121658_004Wed, Feb 16, 2005 12:18:44Em filter=620nmBin:M (8), FOV12.8, f1, 1sCamera: 23235, EEV
Series: 23235Experiment: in situ calib checkLabel: xls 58 2.89e08Comment: Analysis Comment:
8000
6000
4000
2000
ImageMin = -1571
Max = 5198.2counts
Color BarMin = 56
Max = 5047
bkg subflat-fieldedcosmic
580 nm
600 nm
620 nm
640 nm
Fluorescence
Advanced Topics: Diffuse Luminescent
Imaging Tomography (DLITTM)
38
Software
For an In Depth Study – Software Manual
39
What we’ve covered . . .
Science
Light is scattered and absorbed by tissue dependant on wavelength of Calibrated physical units compensate for device settings
Hardware
Custom designed for in vivo bioluminescent imaging
Settings are analogous to photography
Software
Images are acquired in a two step process
Sensitivity is controlled by Integration Time, f/stop, and Binning
Living Image® controls IVIS® and provides image analysis tools
Fluorescence
Tissue Autofluorescence and Instrument Background can be subtracted
40