Page 1: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 1

ATOMIC SPECTROSCOPY

PRESENTED BY : GROUP 8

Page 2: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 2

ATOMIC SPECTROscopY

PRESENTED BY: HANEEN RASHID

Page 3: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 3

ATOMIC ABSORPTIONSPECTROSCOPY (AAS)

concerns the absorption of radiation by the atomised analyte element in the ground state.

• Only applicable for the detection of trace metals.

Page 4: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 4

ATOMIC EMISSION SPECTROSCOPY (AES)

In atomic emission spectrometry, atoms are thermally excited so that they emit light and the radiation emitted is measured.

• Only applicable to determination of alkali and alkaline earth metals.

Page 5: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 5

ATOMIC SPECTRA

PRESENTED BY: RUBINA AFZAL

Page 6: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 6

ATOMIC SPECTRA• In an atom, electrons have specific and discrete

energies in which electron are arranged in definite energy levels. When an electronic transitions (‘jumps’) from one energy level to another (by an electric arc ,temperature or flame), it emits or absorbs light – a photon – with a discrete, specific wavelength, the collection of all these specific wavelengths ( spectral lines) form the spectrum of the atom and it will be the characteristic of particular atom…so atomic spectra are the spectra of atoms.

Page 7: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 7

ATOMIC LINE SPECTRA ARE CHARACTERISTIC FOR EVERY ELEMENT

Page 8: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 8

TYPES OF ATOMIC SPECTRA

A. Atomic absorption spectra

B. Atomic emission spectra

Page 9: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 9

ATOMIC ABSORPTION SPECTRA

• When an electron is excited to a higher energy level it must absorbed energy.

• The energy absorbed as an electron jump from an orbit of low energy to one of the higher energy is characteristic of that transition.

• This mean that the excitation of electron in a particular element result in energy absorption at specific wavelength it will be the characteristic of particular atom thus in addition to emission spectrum every atom possess a characteristic absorption spectrum.

Page 10: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 10

For example

• In gaseous medium, sodium atoms are capable of absorbing radiation of wavelength characteristic of electronic transition from 3s state to higher excited states, sharp absorption peaks are observed experimentally.

Page 11: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 11

ATOMIC EMISSION SPECTRA

Page 12: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 12

For example

• When a sodium salt is heated in a flame the outer electron in the volatilized atoms are excited and returned to ground state with emission of energy, which appears as a yellow light (wavelength 589.5)

• The major line in the sodium emission spectrum is due to an electron falling from 3p excited state to 3s ground state.

• Common atom which give their bands in the emission spectrum are Ca, Ba , Na, Li, k.

Page 13: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 13

Continuous spectra:•It emitted by solids and are characterized

by the absence of any sharp lines as a function of wavelength.

Band spectra •It consists of group of lines. Each group is

characteristic of wavelength that becomes closed spaced as they approach the end of the band .The band also called molecular spectra band. Since radiation is emitted out by the excited molecules.

Line spectra:•These consist of sharply defined and often widely

irregularly spaced individual lines of a single wavelength. These spectra are characteristics of elements and are duo to the excitation of gaseous atoms or atomic ions. Hence, line spectra are also called atomic spectra

TYPES OF EMISSION SPECTRA

Page 14: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 14

• Continuous emission spectrum

• Band emission spectrum

• Line emission spectrum

Page 15: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 15

COMPARISON OF THE ABSORPTION(a) AND EMISSION

LINES (b)OF SODIUM

Page 16: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 16

PRINCIPLE OF ATOMIC SPECTROSCOPY

PRESENTED BY:MAHWISH MAQBOOL

Page 17: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 17

PRINCIPLE OF ATOMIC ABSORPTION SPECTROSCOPY

• The absorption of energy by ground state atoms in the gaseous state forms the basis of atomic absorption spectroscopy.

• By the help of atomic absorption

spectroscopy, one can determine the amount of light absorbed

Page 18: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 18

In atomic absorption spectroscopy:

• Atoms of a metal are volatilized in a flame

and their absorbance of a narrow band of

radiation produced by a hollow cathode

lamp, coated with the particular metal

being determined is measured.

• Absorption will be proportional to the density of atoms in flame.

Page 19: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 19

• Once absorption is known the concentration of the metallic element can also be known because absorption is proportional to concentration of atoms in the flame. Mathematically, the total amount of light absorbed is given by:

Page 20: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 20

PRINCIPLE OF ATOMIC EMISSION SPECTROSCOPY

• Atomic emission spectroscopy involves the measurement of electromagnetic radiation emitted from atoms.

• I n this, atoms are thermally excited so that that emit light and the radiation emitted is measured. Both qualitative and quantitative data can be obtained from this type of analysis.

Page 21: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 21

The process is as follows:• LIQUID STATE• FORMATION OF DROPLETS• FINE RESIDUE• FORMATION OF NEUTRAL ATOMS• EXCITATION OF ATOMS• EMISSSION OF RADIATION OF SPECIFIC

WAVELENGTH• WAVELENGTH AND INTESITY OF EMITTED

RADIATION

• The intensity of emitted radiation depends upon the proportion of thermally excited atoms, which in turns depends upon the temperature of flame.

Page 22: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 22

Fraction of free atoms thermally excited

• The wavelength of the radiation emitted is used to identify the element (Qualitative Analysis). The intensity of the radiation emitted depends upon the concentration of element analyzed (Quantitative Analysis)

Page 23: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 23

ATOMIC ABSORPTION SPECTROSCOPY

INSTRUMENTATION

PRESENTED BY: FARAH HUSSAIN

Page 24: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 24

BASIC COMPONENTS

• A typical atomic absorption spectrophotometer consists of following components.

Radiation sourceAtomic reservoirMonochromatorDetectorReadout device

Page 25: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 25

RADIATION SOURCE

• HOLLOW CATHODE LAMP

• ELECTRODELESS DISCHARGE LAMP

Page 26: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 26

HOLLOW CATHODE LAMP

CONSTRUCTION:• The lamp consists of thick walled glass envelope which

has a transparent window of glass and silica affixed to one end. It consists of one anode and cup shaped cathode which are both connected to tungsten wire. The tube is filled with highly pure inert gas at low pressure of 1 to 2mm. The gases generally used are neon ,argon or helium. Mica sheets are placed inside the lamp to limit the radiation to with in the cathode. The choice of window material depend upon the wavelength of the resonance lines of the element concerned.

Page 27: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 27

HOW IT WORKS:• A potential of about 500v is applied between the

electrodes and the current of 2-30mA is used. The filler gas becomes charged at the anode and the ions produced are attracted to the cathode and accelerated by the field. Bombardment of these ions on the inner surface of cathode causes metal atom to sputter out of the cathode cup. Further collisions excite these metal atoms and simple, intense characteristic spectrum of the metal is produced.

Page 28: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 28

HOLLOW CATHODE DISCHARGE LAMP

Quartz window

Pyrex body

Anode

Cathode

Cathode

Anode

Page 29: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 29

ELECTRODELESS DISCHARGE LAMP

CONSTRUCTION:• In this a small amount of the metal or salt of the

element for which the source is to be used is sealed inside the quartz bulb. This bulb is placed inside a small self contained RF generator or Driver

Page 30: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 30

HOW IT WORKS:

• When power is applied to the driver, an RF field is created. The coupled energy is vaporized and excite the atom inside the bulb, causing them to emit their characteristic spectrum. The emission from an EDL is higher than that from an HCL, and the line width is generally narrower, but EDLs need a separate power supply and might need a longer time to stabilize.

Page 31: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 31

ELECTRODELESS DISCHARGE LAMP

Page 32: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 32

ATOMIZERS

• The principle of atomic absorption requires light absorption by free atoms. However, elements in the sample are in a molecular form. The combination must be broken by some means to free the atoms. This is called ATOMISATION. The most popular method of atomization in AAS is flame excitation source.

Page 33: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 33

BASIC STEPS INVOLVED IN FLAME ATOMIZATION

Nebulisation: conversion of sample into droplets.

Desolvation: removal of solvents.Atomization: thermal or chemical

breakdown of solid particles.Condensation of reaction product:

removal of residue by exhaust flame gases.

Page 34: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 34

BURNERS

• Two major types of nebulizer burners are used in AAS are PREMIX NEBULISER BURNER and TOTAL CONSUMPTION BURNER.

• In premix type burner liquid is sprayed into mixing chamber where the droplets are mixed with combustion gas and are send to the burner.

• In the total consumption burner, nebulizer and burners are combined. This is also called TURBULENT FLOW BURNER

Page 35: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 35

MONOCHROMATORS Important device of the optical system of AAS. The function of this device is to separate the spectral line of

interest from the other spectral line of different wavelength emitted by the hollow cathode lamp. The desired spectral lines is chosen with the preferred wavelength and bandwidth with the help of grating.

GRATING Wavelength dispersion is accomplished with a grating, a reflective

surface ruled with many fine parallel lines very close together. Reflection from this ruled surface generates interference

phenomenon known as diffraction in which different wavelength of light diverge from grating at different angle.

Page 36: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 36

WORKING: Light from the source enters the monochromator

at the entrance slit and is directed to the grating where dispersion takes place. The diverging wavelength of light are directed towards exit slit. By adjusting the angle of grating, a selected emission line from the source can be allowed to pass through exit slit and fall onto the detector

Page 37: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 37

DETECTOR

• The most commonly used in AAS is Photomultiplier tube whose output is fed into readout system.

• A PMT is an electronic tube that is capable converting the photon current into electrical signal and of amplifying this signal.

.

Page 38: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 38

WORKING:• It consists of photo cathode and secondary electron

multiplier. The photons impact on the cathode surface and sputter electrons from its surface. These electrons are accelerated in an electric field and impact on other electrodes so called dynodes from the surface of each impacting electron sputter secondary electrons. This cascade effect results in significance increase in number of electron. At the end electron impact on an anode and flow off to the mass. The resulting current is measured.

Page 39: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 39

READOUT SYSTEM

It included meters, chart out device and digital display meter.

These days microprocessor controlled system are commercially available where every thing can be done by touch of button.

Page 40: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 40

INSTRUMENTATION OF AES

PRESENTED BY :

HAFIZA RABEEA NISAR

Page 41: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 41

ATOMIC EMISSION SPECTROMETER

Consists of three main parts: Emission source Optical system Detector

Page 42: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 42

EMISSION SOURCE

Atomization and excitation of sample is done by emission source:

Direct current plasma (DCP) Flame excitation source Inductively coupled plasma (ICP) Spark and arc Laser induced breakdown (LIB) Microwave induced plasma (MIP)

Page 43: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 43

1. Direct Current Plasma Excitation Source

DCP is created by an electrical discharge between two electrodes. A plasma support gas is necessary (Ar)

Sample deposited on one of the electrode.

Solid samples, near the discharge so that ionized gas atoms sputter the sample into the gas phase where the analyte atoms are excited.

Page 44: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 44

2. Flame Excitation Source

The sample solution is directly aspirated into the flame.

All desolvation, atomization, and excitation occurs in the flame.

Flame provides a high-temperature source for desolvating a sample.

Page 45: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 45

3. Inductively-Coupled Plasma Excitation Source

Very high temperature excitation source that efficiently desolvates, vaporizes, excites, and ionizes atoms.

Sample is nebulized and entrained in the flow of plasma support gas (Ar).

Page 46: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 46

The plasma torch consists of concentric quartz tubes, with the inner tube containing the sample aerosol and Ar support gas and the outer tube containing an Ar gas flow to cool the tubes.

An RF generator produces an oscillating current in an induction coil that wraps around tubes, this coil creates an oscillating magnetic field, which produces an oscillating magnetic field, which in turn sets up an oscillating current in the ions and electrons of the support gas.

These ions and electrons transfer energy to other atoms by collisions to create very high temperature plasma.

Page 47: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 47

4. Spark and Arc

Spark and arc excitation sources use a current pulse (spark) or a continuous Electrical discharge (arc) between two electrodes to vaporize and excite analyte atoms.

Samples are ground with graphite powder and placed into a cup-shaped lower electrode.

Page 48: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 48

5. Laser-Induced Breakdown

High-energy laser pulse is focused into a gas or liquid, or onto a solid surface, it creates hot plasma.

The energy of the laser-created plasma can atomize, excite, and ionize analyte species.

Page 49: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 49

6. Microwave Induced Plasma

Consists of a quartz tube surrounded by a microwave waveguide.

Microwaves produced by microwave generator fill the waveguide or cavity and cause the electrons in the plasma support gas to oscillate.

The oscillating electrons collide with other atoms to create and maintain a high-temperature plasma.

Page 50: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 50

OPTICAL SYSTEM

Once the sample has been introduced into the emission source, atomized, and excited, the emitted photons are diffracted by an optical system consisting of slits, mirrors, and gratings, etc.

It may be a Monochromators or Diffraction grating.

Page 51: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 51

1. Monochromators

A Monochromators is an optical device that transmits a mechanically selectable narrow band of wavelengths of light or radiation.

Two designs are usually employed.

1. Czerny–Turner Monochromators

2. Echelle Monochromators

Page 52: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 52

Czerny–Turner

• Two mirrors are used to reflect and focus the polychromatic and diffracted beams.

• As the grating rotates, a different wavelength is focused onto the exit slit.

Echelle Monochromators

• It’s made up of two dispersing elements. These elements are arranged in series.

• They have high resolution and dispersion property.

Page 53: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 53

2. Grating

Gratings are reflective surfaces containing parallel, equally spaced lines.

Their resolving power is proportional to the number of lines, which in turn depends on the line spacing.

Diffraction gratings are used in optical system to disperse the emitted radiation.

Page 54: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 54

DETECTORS

• Photomultiplier tube

READ OUT SYSTEM

Page 55: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 55

TYPES OF ATOMIC SPECTROPHOTOMETERS

PRESENTED BY:

ALMAS SHAMIM

Page 56: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 56

TYPES OF ATOMIC ABSORPTION SPECTROPHOTOMETER

• Single beam atomic absorption spectrophotometer

• Double beam atomic absorption spectrophotometer

Page 57: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 57

SINGLE BEAM ATOMIC ABSORPTION SPECTROPHOTOMETER

Page 58: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 58

DOUBLE BEAM ATOMIC ABSORPTION SPECTROPHOTOMETER

Page 59: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 59

TYPES OF ATOMIC EMISSION SPECTROPHOTOMETER

• Sequential spectrometers

• Simultaneous spectrometers

Page 60: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 60

• The sequential and simultaneous spectrometers are extensively used in the analytical laboratories.

• The sequential spectrometers are less expensive and more flexible but usually require a higher degree of operator skill and experience.

• On the other hand, the simultaneous spectrometers are more precise and accurate and are obviously more expensive.

Page 61: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 61

SEQUENTIAL SPECTROMETER

Page 62: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 62

SIMULTANEOUS SPECTROMETER

1. Polychromators

2. Solid state array based spectrometers

Page 63: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 63

POLYCHROMATORS

Page 64: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 64

SOLID STATE ARRAY BASED SPECTROMETERS

Page 65: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 65

INTERFERENCES

PRESENTED BY:

HANEEN RASHID

Page 66: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 66

INTERFERENCES IN ATOMIC

ABSORPTIONSPECTROPHOTOMETRY

Page 67: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 67

The presence of another atomic absorption line or a molecular absorption band close to the spectral line of the analyte element being monitored.

• This may be corrected by modulation of the radiation source and the detection system.

.

1. Spectral Interferences

Page 68: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 68

2. Chemical interferences

These include interferences due to ionisation, formation of low volatility compounds, etc

• Such interferences can be avoided by increasing the flame temperature.

Page 69: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 69

3. Physical interferences

These may be due to variations in the gas flow rates, changes in the solution viscosity which may finally change the atomic concentration in the flame.

• This can be avoided by matching the matrix and by performing frequent calibrations.

Page 70: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 70

INTERFERENCES IN ATOMIC EMISSION

SPECTROPHOTOMETRY

Page 71: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 71

1. Spectral Interferences

involves the overlap of the spectral lines of two or more elements in the matrix emitting radiation at the same wavelength.

• These can be minimized by using a high resolution dispersion system.

Page 72: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 72

2. Chemical interferences

The chemical interferences include molecular compound formation and ionization effects.

• These can be minimized by carefully controlling the operating conditions.

Page 73: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 73

3. Physical interferences

These interferences are associated with the processes of sample nebulisation and transport.

• The physical interferences can be reduced by diluting the sample.

Page 74: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 74

APPLICATIONS

PRESENTED BY KIRAN ILLAHI

Page 75: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 75

Atomic absorption spectroscopy is widely used in metallurgy, alloys and in inorganic analysis. Almost all important metals have been analyzed by this technique.

It is a sensitive means for the quantitative determination of more than 60 metals or metalloid elements.

Page 76: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 76

IN BIOLOGICAL ANALYSIS, a no of elements present in biological samples such as urine and blood sample can be analyzed by atomic absorption method. These include estimation of sodium, potassium, lead, zinc, mercury, cadmium, calcium, magnesium and iron.IN PHARMACEUTICAL ANALYSIS,

estimation of zinc in insulin preparation, oils, creams and in calamine, calcium in no of calcium salts has been reported.

Sodium, potassium and calcium in saline and ringer solution are estimated by this method.

Page 77: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 77

IN PETROLEUM INDUSTRY, metallic impurities in petrol, lubricating oils have been determined.

IN CEMENT INDUSTRY, estimation of sodium, potassium, magnesium, and calcium is carried out by this technique to determine the quantity of cement.

IN ENVIRONMENT ANALYSIS, trace metals and other metals in waters, soils, plants, composts and sludge.

Page 78: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 78

IN FORENSIC SCIENCE , gunshot powder analysis, toxicological examination( e. g, Thallium TI determination)

It is used in a number of limit tests for metallic impurities e. g, magnesium and strontium in calcium acetate and lead in bismuth sub gal late.

It is also used to assay metals in a number of other preparations like zinc in zinc insulin suspension and copper and iron in ascorbic acid.

It is also used in assay of Intra peritoneal dialysis fluid (for calcium, magnesium), activated charcoal (for zinc), cisplatin (for silver).

Estimation of elements in soil sample, water supply, effluents, ceramics, etc.

Page 79: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 79

OTHER BROAD AREAS WHERE ATOMIC

SPECTROSCOPY IS USED TO QUANTIFY ELEMENTS

INCLUDE:Agriculture, food,

geochemistry, forestry, oceanography, fertilizer etc.

Page 80: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 80

THE PRINCIPLE APPLICATION OF ATOMIC EMISSION SPECTROSCOPY IS:To determine the proportional quantity of a particular

element in a given sample. The various methods of atomic emission spectroscopy are utilized to examine different substances such as foods and drinks, motor oil and soil samples.

Atomic Emission Spectroscopy is predominantly utilized in space research labs by NASA and ESA. It is also used for aiding various military operations.

Page 81: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 81

Atomic absorption (AA) spectrometers analyze the

absorption spectra, whereas in atomic emission spectroscopy

(AES) the emission spectra of a sample is analyzed after

excitation by a flame, discharge or plasma source.(15)

Page 82: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 82

ADVANTAGES AND DISADVANTAGES

PRESENTED BY: MADIHA KHALID

Page 83: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 83

ADVANTAGES OF AAS The atomic absorption technique is specific

because the atom of a particular elements can only absorb radiation of their own characteristic wavelength i.e. the light of a particular frequency can easily be absorbed by the specific element to which it is characteristic.

Atomic absorption spectroscopy is independent of flame temperature.

Page 84: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 84

DISADVANTAGES OF AAS

A separate lamp for each element is required.

This technique is unsuccessful for estimation of elements like Al, Mo; vanadium because, these elements give rise to red metallic oxidizes in the flame.

In aqueous solution the signal is affected by predominant anion.

Page 85: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 85

ADVANTAGES OFAES

It is very specific; the unique character of the wavelength pattern produced by each element is the reason of its specificity.

The method is extremely sensitive, with this technique all metallic elements can be detected even if they are present in very low concentration (0.0001%). Even metalloids (arsenic, silicon, and selenium) have been identified.

The method can be used for qualitative analysis as well as to determine concentration as low as 1 ppm.

Page 86: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 86

The analysis has the advantage of providing results rapidly because there are situation (e.g. in industrial processes) where time is more important than the high levels of accuracy. If automated, time required is just 30 seconds to a minute.

A very small amount of sample (1-10mg) is sufficient for analysis.

Page 87: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 87

DISADVANTAGES OF AES

The equipment is costly and wide experience is required for its successful handling and interpretation of spectra.

The spectrograph is essentially a comparator, for quantitative analysis, standards (usually similar composition to the material under analysis) are necessary. For quantitative results, therefore, a problem is often posed for unknown samples.

The sample is destroyed in the process of analysis.

Page 88: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.

Page 88

This method is not recommended for elements present to a greater extent than 3 percent since for quantities greater than 2-5 percent of the method does not offer accuracy as in gravimetric, titrimetric and some spectrophotometric measurements.

The method is limited to the analysis of elements.

Page 89: Page 1 ATOMIC SPECTROSCOPY PRESENTED BY : GROUP 8.