Top Banner
NIST Traceable Volumetric NIST Traceable Volumetric Flow Rate, Temperature and Flow Rate, Temperature and Pressure Calibration/Audit Pressure Calibration/Audit Devices for FRM and Devices for FRM and Continuous Ambient Monitors Continuous Ambient Monitors TAMS Continuous Particulate Training Workshop October 21 - 24, 2003 BGI Instruments
16

NIST Traceable Volumetric Flow Rate, Temperature and Pressure Calibration/Audit Devices for FRM and Continuous Ambient Monitors TAMS Continuous Particulate.

Dec 28, 2015

Download

Documents

Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: NIST Traceable Volumetric Flow Rate, Temperature and Pressure Calibration/Audit Devices for FRM and Continuous Ambient Monitors TAMS Continuous Particulate.

NIST Traceable Volumetric Flow Rate, NIST Traceable Volumetric Flow Rate, Temperature and Pressure Temperature and Pressure

Calibration/Audit Devices for FRM Calibration/Audit Devices for FRM and Continuous Ambient Monitorsand Continuous Ambient Monitors

TAMS

Continuous Particulate Training Workshop

October 21 - 24, 2003

BGI Instruments

Page 2: NIST Traceable Volumetric Flow Rate, Temperature and Pressure Calibration/Audit Devices for FRM and Continuous Ambient Monitors TAMS Continuous Particulate.

IntroductionIntroduction

Federal, State and Tribal AQ Agencies Require Flow Calibration and Audits on all Gravimetric Particulate Samplers and Continuous Monitors

Commercial Vendor Claim “Primary Flow” and “NIST Traceability” are Confusing

A NIST Traceable Volumetric Flow Rate, Ambient Temperature and Pressure Field Calibration/Audit Device Has Been Developed and Tested

Page 3: NIST Traceable Volumetric Flow Rate, Temperature and Pressure Calibration/Audit Devices for FRM and Continuous Ambient Monitors TAMS Continuous Particulate.

InvestigationInvestigation

NIST Primary Standard vs. Flow Transfer Standard

Requirements for a Field Calibrator Current Methods Available DeltaCal & TriCal NIST Traceable

Calibrators Meet Requirements

Page 4: NIST Traceable Volumetric Flow Rate, Temperature and Pressure Calibration/Audit Devices for FRM and Continuous Ambient Monitors TAMS Continuous Particulate.

NIST Primary & EPA Flow Transfer NIST Primary & EPA Flow Transfer Standards Standards

–NIST “PrimaryStandard” Flow Calibration FacilityNIST “PrimaryStandard” Flow Calibration Facility–Colorado Engineering Experiment Station (CEESI) Colorado Engineering Experiment Station (CEESI) –Titanium Vessels used for Gravimetric Analysis as Titanium Vessels used for Gravimetric Analysis as the NIST Primary Standardthe NIST Primary Standard–Pressurized, Evacuated and Weighed for Primary Pressurized, Evacuated and Weighed for Primary Volume.Volume.

–NIST “Transfer Standard”NIST “Transfer Standard”– Flow Device Traceable to NIST (CEESI) PrimaryFlow Device Traceable to NIST (CEESI) Primary

Page 5: NIST Traceable Volumetric Flow Rate, Temperature and Pressure Calibration/Audit Devices for FRM and Continuous Ambient Monitors TAMS Continuous Particulate.

CEESI - NIST Primary Calibration LabCEESI - NIST Primary Calibration Lab

Page 6: NIST Traceable Volumetric Flow Rate, Temperature and Pressure Calibration/Audit Devices for FRM and Continuous Ambient Monitors TAMS Continuous Particulate.

Priorities of a Field CalibratorPriorities of a Field Calibrator

– NIST Transfer Standard, Accuracy to <2%to Primary NIST Transfer Standard, Accuracy to <2%to Primary

– Calibrator must be Field Usable to EPA FRM SpecsCalibrator must be Field Usable to EPA FRM Specs

– Insensitive to Atmospheric ConditionsInsensitive to Atmospheric Conditions

– Calibrator should not add artificial pressure to Calibrator should not add artificial pressure to sampler to interfere with electronic flow controllersampler to interfere with electronic flow controller

– Portable & Battery Powered Portable & Battery Powered

– Direct Reading of Flow Rate, Ambient Temperature, Direct Reading of Flow Rate, Ambient Temperature, & Barometric Pressure & Barometric Pressure

– RS232 Calibration Report-Digital Calibration ReportRS232 Calibration Report-Digital Calibration Report

Page 7: NIST Traceable Volumetric Flow Rate, Temperature and Pressure Calibration/Audit Devices for FRM and Continuous Ambient Monitors TAMS Continuous Particulate.

““Flow Transfer Standard” CalibratorsFlow Transfer Standard” Calibrators

– Bubble MeterBubble Meter

– GilibratorGilibrator™ (Wet) - ™ (Wet) - (Excellent for Indoor Use)(Excellent for Indoor Use)

– Frictionless PistonFrictionless Piston - ( - (Not Field Hardened, Dirty Piston)Not Field Hardened, Dirty Piston)

– Bios/DryCal/DC-LiteBios/DryCal/DC-Lite™ (Dry)- 3% low bias ™ (Dry)- 3% low bias

– Orifice- Quartz TubeOrifice- Quartz Tube - - (Sensitive to Wind)(Sensitive to Wind)

– Streamline/ChinookStreamline/Chinook™ (Dry) – adds system ™ (Dry) – adds system pressure, double of a Venturipressure, double of a Venturi

– Venturi- Venturi- - - (Field Hardened with Gust Cap)(Field Hardened with Gust Cap)

– BGI DeltaCalBGI DeltaCal (Dry) – flow, temperature & pressure (Dry) – flow, temperature & pressure

Page 8: NIST Traceable Volumetric Flow Rate, Temperature and Pressure Calibration/Audit Devices for FRM and Continuous Ambient Monitors TAMS Continuous Particulate.

DeltaCal Calibration KitDeltaCal Calibration Kit

DeltaCal

Measurements:• Barometric Pressure• Ambient Temperature• Filter Temperature• Volumetric Flow Rate• Digital Output for

Recording Field Data• Software to Print

Summary report

Page 9: NIST Traceable Volumetric Flow Rate, Temperature and Pressure Calibration/Audit Devices for FRM and Continuous Ambient Monitors TAMS Continuous Particulate.

Schematic Diagram of DeltaCalSchematic Diagram of DeltaCal

Page 10: NIST Traceable Volumetric Flow Rate, Temperature and Pressure Calibration/Audit Devices for FRM and Continuous Ambient Monitors TAMS Continuous Particulate.

DeltaCal NIST Traceable SpecificationsDeltaCal NIST Traceable Specifications

– Operable Range –30 to 40°COperable Range –30 to 40°C

– 2 to 20 aLPM Model DC 12 to 20 aLPM Model DC 1

– 10 to 60 aLPM Model DC 10010 to 60 aLPM Model DC 100

– +/- 1% of NIST+/- 1% of NIST

– Temp Sensor Range – 40 to 50°CTemp Sensor Range – 40 to 50°C

– < +/- 0.5°C Difference to NIST< +/- 0.5°C Difference to NIST

– Pressure Range 400-800mmHgPressure Range 400-800mmHg

– < +/- 2mmHg of NIST< +/- 2mmHg of NIST

Page 11: NIST Traceable Volumetric Flow Rate, Temperature and Pressure Calibration/Audit Devices for FRM and Continuous Ambient Monitors TAMS Continuous Particulate.

Calibrator Software Screen DisplayCalibrator Software Screen Display

Historical Record

User Configurable

All Readings in one Display

Auto Resolution

ASCII File Export

Flow Statistics

Page 12: NIST Traceable Volumetric Flow Rate, Temperature and Pressure Calibration/Audit Devices for FRM and Continuous Ambient Monitors TAMS Continuous Particulate.

Summary Printed ReportSummary Printed Report

Page 13: NIST Traceable Volumetric Flow Rate, Temperature and Pressure Calibration/Audit Devices for FRM and Continuous Ambient Monitors TAMS Continuous Particulate.

TriCal CalibratorTriCal Calibrator

Operable Range -30 to 40°C

Flow Range 0.1 to 30.0Lpm

Venturi #1 0.1 - 1.2Lpm Venturi #2 1.2 -6.0Lpm Venturi #3 6.0 - 30Lpm

Barometric Pressure Ambient Temperature (Option) Filter Temperature

Page 14: NIST Traceable Volumetric Flow Rate, Temperature and Pressure Calibration/Audit Devices for FRM and Continuous Ambient Monitors TAMS Continuous Particulate.

Calibrating a Continuous PM MonitorCalibrating a Continuous PM Monitor

CPM Display

Page 15: NIST Traceable Volumetric Flow Rate, Temperature and Pressure Calibration/Audit Devices for FRM and Continuous Ambient Monitors TAMS Continuous Particulate.

BGI - NIST Certified Flow Calibration BGI - NIST Certified Flow Calibration LaboratoryLaboratory

– < Insert photo of BGI Test Bench >< Insert photo of BGI Test Bench >

Page 16: NIST Traceable Volumetric Flow Rate, Temperature and Pressure Calibration/Audit Devices for FRM and Continuous Ambient Monitors TAMS Continuous Particulate.

ConclusionsConclusions

BGI DeltaCal & TriCal are Flow Transfer Standards Directly Traceable to NIST Primary

Field Hardened Device with Low Pressure Drop and DeltaCal Includes Wind Gust Cap

Cost Includes 4 Measurements, Rather than just One $1695.

Affordable Re-Calibration by a Single Vendor $150.

US EPA Tested – Exclusive PEP PM2.5 Network & PM-Coarse Method Development Program