Microelettrovalvola per Bioprinting: prove e misure preliminari Candidato: Veronica Marafioti Docente tutore: Dott. Michele Conti Anno Accademico: 2014/2015 Laurea in Bioingegneria Università degli studi di Pavia
17

Microelettrovalvola per Bioprinting - unipv · Biomedical Engineering Università degli Studi di Pavia - Structural Mechanics Department Microelettrovalvola per Bioprinting: prove

Oct 08, 2019

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Microelettrovalvola per Bioprinting - unipv · Biomedical Engineering Università degli Studi di Pavia - Structural Mechanics Department Microelettrovalvola per Bioprinting: prove

Biomedical Engineering

Università degli Studi di Pavia - Structural Mechanics Department

Microelettrovalvola per Bioprinting: prove e misure preliminari

Candidato: Veronica Marafioti

Docente tutore: Dott. Michele Conti

Anno Accademico: 2014/2015

Laurea in Bioingegneria

Università degli studi di Pavia

Page 2: Microelettrovalvola per Bioprinting - unipv · Biomedical Engineering Università degli Studi di Pavia - Structural Mechanics Department Microelettrovalvola per Bioprinting: prove

Based on the three “B”:

Bioink:

Biopaper:

Bioprinter:

What is Bioprinting?

Bioprinting can be defined as the use of computer-aided transfer processes for

patterning and assembling living and non-living materials with a prescribed 2D or

3D organization in order to produce bio-engineered structures serving in

regenerative medicine, pharmacokinetic and basic cell biology studies. (International Conference about Bioprinting and Bio Manufacturing - Bordeaux 2009)

Page 3: Microelettrovalvola per Bioprinting - unipv · Biomedical Engineering Università degli Studi di Pavia - Structural Mechanics Department Microelettrovalvola per Bioprinting: prove

Bioprinting workflow

Page 4: Microelettrovalvola per Bioprinting - unipv · Biomedical Engineering Università degli Studi di Pavia - Structural Mechanics Department Microelettrovalvola per Bioprinting: prove

Bioprinters

NovoGen MMX 3D Bioplotter 3D Discovery

Not marketed $ 200000 $200000

$ 180000 £ 12000/18000 $ 10000 $ 5000/9000

BioScaffolder 2.1 Alpha and Omega BioBots Inkredible

Refs: https://replicatore.wordpress.com/2015/09/06/top-10-biostampanti-commerciali-2/

Page 5: Microelettrovalvola per Bioprinting - unipv · Biomedical Engineering Università degli Studi di Pavia - Structural Mechanics Department Microelettrovalvola per Bioprinting: prove

BioPrinting project – Lee’s bioplotter

Lee’s Bioplotter: modular tissue

printing platform

1. 4 syrynges as “cartridges” to

load cell suspensions and

hydrogel precursors

2. An array of 4-channel

dispensers

3. Target substrate

4. Horizontal stage

5. Vertical stage

6. Range finder

7. Vertical stage heater/cooler

8. Optional indipendent

heating/cooling for the

dispenser

BioPrinting project’s Goal:

Fig. Lee 2008, Multi-layered culture of human skin fibroblasts and keratinocytes through three-dimensional

freeform fabrication

Page 6: Microelettrovalvola per Bioprinting - unipv · Biomedical Engineering Università degli Studi di Pavia - Structural Mechanics Department Microelettrovalvola per Bioprinting: prove

BioPrinting project – Our approach

Develop a simple work plan, not

automated

Integrate the dispenser in a 3D

printer

printing proofs

(with cells)

Inspiring by Lee’s bioplotter, this is our purpose:

Page 7: Microelettrovalvola per Bioprinting - unipv · Biomedical Engineering Università degli Studi di Pavia - Structural Mechanics Department Microelettrovalvola per Bioprinting: prove

BioPrinting project

Global setup: the following material was necessary for the development of the

project

1. Acquisition of pressure signal whit 0Psi to 15Psi Gauge Honey-Well sensor, DAQ

National Instrument, computer

2. Acquisition of the piston position signal, to control the valve opening / closing

3. Air pressurization system

4. Control Box’s valve

5. Power supply

6. Microelectrovalve

6

Page 8: Microelettrovalvola per Bioprinting - unipv · Biomedical Engineering Università degli Studi di Pavia - Structural Mechanics Department Microelettrovalvola per Bioprinting: prove

Microelectrovalve - Features

Dolphin Fluidics’ DFD-Smart

Technical Data

Nozzle Diameter Ø 0.8 mm

Pressure Range 0 – 4.0 bar

Operating Temperature -10° C - +65° C

Current Range 150 mA – 240 mA

Response Time @ 0.6 W 180 ms

Holding Power 0.1 W

Implementation Power 0.4 – 0.6 W

Life-time Million of cycles

Control On/Off and Analogic

The DFD-Smart is a modular system with 2-way valves, total isolation, ideal for

controlling fluid flows at high hygienic nature and not be contaminated. Each valve can

be single, double or coupled in a fluidic block. Each channel can be controlled on-off or

proportional independently.

Page 9: Microelettrovalvola per Bioprinting - unipv · Biomedical Engineering Università degli Studi di Pavia - Structural Mechanics Department Microelettrovalvola per Bioprinting: prove

Microelectrovalve - Measurements

The purpose of the setup is to measure the performance of the first prototype

of the microelectrovalve DFD-Smart (Dolphin Fluidics) by drawing a graph

of the flow rate Q [ml/min] as a function of working pressure P [mmHg].

Steps of measurement process:

Pressure

signal P

[mmHg]

Valve

calibration

Voltage

Valve

opening time

T [s] Piston

position

signal

Dispensing

volume V [ml]

Plot of Flow

rate Q

[ml/min]

Page 10: Microelettrovalvola per Bioprinting - unipv · Biomedical Engineering Università degli Studi di Pavia - Structural Mechanics Department Microelettrovalvola per Bioprinting: prove

First setup - Results

Testing conditions:

Range pressure from 60 to 150 mmHg

Constant voltage of 3.3V (100% of the valve opening)

Conclusion:

1. Low accuracy and precision of the measurements

2. Leakage phenomenon for pressure under 150 mmHg

3. Channel 2 was clogged

y = 0,0611x + 0,2872 R² = 0,6687

3,00

4,00

5,00

6,00

7,00

8,00

9,00

10,00

50 70 90 110 130 150

Flo

w R

ate

[m

l/m

in]

Pressure [mmHg]

Channel 1

FlowRate [ml/min] VS Pressure [mmHg]

Lineare (FlowRate [ml/min] VS Pressure [mmHg])

Page 11: Microelettrovalvola per Bioprinting - unipv · Biomedical Engineering Università degli Studi di Pavia - Structural Mechanics Department Microelettrovalvola per Bioprinting: prove

Second setup - Results

Dolphin Fluidics

UniPV

@ P = 120 mmHg: plot of the Flow rate as a function of the valve opening

percentage

Conclusion:

1. The measures do not reproduce the sigmoid curve

2. Low repeatability

3. Both channels dispensed less than what we expected, because they were

clogged

0,0

5,0

10,0

15,0

20,0

25,0

30,0

35,0

10,0 20,0 30,0 40,0 50,0 60,0 70,0

Flo

w r

ate

[m

l/m

in]

% Opening

Channels’ characteristic

CH1

CH2

5,50

6,00

6,50

7,00

7,50

8,00

8,50

9,00

30,0 50,0 70,0 90,0 110,0

Flo

w r

ate

[m

l/m

in]

% Opening

CH1 flow rate [ml/min]

Misure 25/02/16

Misure 26/02/16 #1

Misure 26/02/16 #2

Page 12: Microelettrovalvola per Bioprinting - unipv · Biomedical Engineering Università degli Studi di Pavia - Structural Mechanics Department Microelettrovalvola per Bioprinting: prove

Third setup – Results (I)

0,0

5,0

10,0

15,0

20,0

25,0

95 105 115 125 135

Flo

w R

ate

[m

l/m

in]

Mean Pressure [mmHg]

Pressure Test

Channel 2 16/03/2016

Channel 1 16/03/2016

Channel 1 17/03/2016

Channel 2 17/03/2016

The flow rate was measured by pressurizing a 10 ml syringe, containing H2O

Conclusion:

1. Channel 1 presents more accuracy and repeatability than channel 2

2. Channel 2 dispensed less than channel 1

3. It’s often necessary to clean the channels

CH2 tends to clog easily

1 2

1. Microelectrovalve 2. Syringe

Page 13: Microelettrovalvola per Bioprinting - unipv · Biomedical Engineering Università degli Studi di Pavia - Structural Mechanics Department Microelettrovalvola per Bioprinting: prove

Third setup – Results (II)

Testing conditions for the characteristic of the channels:

Average working pressure: 120 mmHg

Valve opening time: 30 s

Variation of the voltage and of the valve opening percentage

Conclusion:

1. Both channels present a sigmoid curve

2. CH1 dispensed from 0.3V, while CH2 from 0.9V

3. CH1 dispensed more than CH2

0,0

2,0

4,0

6,0

8,0

10,0

12,0

14,0

16,0

18,0

0 0,2 0,4 0,6 0,8 1 1,2 1,4 1,6

Flo

w R

ate

[m

l/m

in]

Voltage[V]

CH2

CH1

Page 14: Microelettrovalvola per Bioprinting - unipv · Biomedical Engineering Università degli Studi di Pavia - Structural Mechanics Department Microelettrovalvola per Bioprinting: prove

Third setup – Results (III)

There is a real possibility of making printing tests

with silk-based solution

A step forward: the valve was tested with

glycerol solution to simulate silk hydrogel

Constant pressure of 120 mmHg

100% of the valve opening

Conclusion: 1. Flow rate decreases with increasing

viscosity

2. Channel 2 dispenses less than

channel 1

3. Both channels tend to become

clogged, so a frequently clean was

necessary

4. The valve is able to dispense up to

25 cP 0,0

5,0

10,0

15,0

20,0

25,0

30,0

35,0

0 10 20 30

Flo

w R

ate

[m

l/m

in]

Viscosity [cP]

Viscosity Test

Channel 2 18/03/2012

Channel 1 18/03/2016

Page 15: Microelettrovalvola per Bioprinting - unipv · Biomedical Engineering Università degli Studi di Pavia - Structural Mechanics Department Microelettrovalvola per Bioprinting: prove

Limitations and future developments

Improvements:

Digital control of the pressure

signal (constant pressure of 1 - 3

Psi)

Syringes washing system

New support for the syringes, to

minimize the distance between

syringes and the valve

Automated control of the

dispensing

Ackn.: Mr. Pierangelo Bergamaschi

Limitations:

The accuracy decreases under

100 mmHg of pressure

Low precision beacuse of the

manual control of the dispensing

Channels tend to clog easily

Page 16: Microelettrovalvola per Bioprinting - unipv · Biomedical Engineering Università degli Studi di Pavia - Structural Mechanics Department Microelettrovalvola per Bioprinting: prove

Conclusion

February 2016

April 2016 Video

Page 17: Microelettrovalvola per Bioprinting - unipv · Biomedical Engineering Università degli Studi di Pavia - Structural Mechanics Department Microelettrovalvola per Bioprinting: prove

GRAZIE PER L’ATTENZIONE