Page 1: Left Atrial Appendage   Occlusion Device, nicvd

7/27/2019 Left Atrial Appendage Occlusion Device, nicvd

http://slidepdf.com/reader/full/left-atrial-appendage-occlusion-device-nicvd 1/23

Drug Presentation

Presenter:DR. MD. ARIFUR RAHMAN

MD (Cardiology), Final Part StudentNICVD, DHAKA.

Moderator:

DR. MOHAMMAD ULLAH FIROZAssistant professor, Dept.of CardiologyNICVD, DHAKA.

Page 2: Left Atrial Appendage   Occlusion Device, nicvd

7/27/2019 Left Atrial Appendage Occlusion Device, nicvd

http://slidepdf.com/reader/full/left-atrial-appendage-occlusion-device-nicvd 2/23

Left AtrialAppendage

Occlusion Device 

Page 3: Left Atrial Appendage   Occlusion Device, nicvd

7/27/2019 Left Atrial Appendage Occlusion Device, nicvd

http://slidepdf.com/reader/full/left-atrial-appendage-occlusion-device-nicvd 3/23

Background

Atrial fibrillation (AF) is the most common sustainedcardiac rhythm disturbance, increasing in prevalence withage.

Stroke is the most feared consequence of AF and mostthrombi associated with AF arise within the left atrialappendage.

AF confers an increased relative risk of overall mortality(average 1.7 times controls) and is predominantly causedby stroke.

Stroke prevention is the major focus of management ofpatients with AF.

Page 4: Left Atrial Appendage   Occlusion Device, nicvd

7/27/2019 Left Atrial Appendage Occlusion Device, nicvd

http://slidepdf.com/reader/full/left-atrial-appendage-occlusion-device-nicvd 4/23

Background(cont.)

Anticoagulation is the treatment of choice to prevent

stroke in AF.

Anticoagulation has many side effects such as intracranialor gastrointestinal hemorrhage.

Closing the left atrial appendage (LAA)  might be analternative in patients who cannot take anticoagulationtreatment due to contraindications, intolerance or

conditions in which the hazard of hemorrhage is greaterthan the potential clinical benefits.

Page 5: Left Atrial Appendage   Occlusion Device, nicvd

7/27/2019 Left Atrial Appendage Occlusion Device, nicvd

http://slidepdf.com/reader/full/left-atrial-appendage-occlusion-device-nicvd 5/23

Left Atrial Appendage Occlusion Device  Percutaneous occlusion of the LAA is a new approach

used to prevent strokes in high-risk patients with AF.

This device is a permanent implant that blocks theconnection between the left atrium and the LAA, thus

prevents embolism.

Currently, two devices have been developed:

1.The PLAATO system ( Appriva Medical Inc)

2.The WATCHMAN filter system (Atritech, Inc) 

Page 6: Left Atrial Appendage   Occlusion Device, nicvd

7/27/2019 Left Atrial Appendage Occlusion Device, nicvd

http://slidepdf.com/reader/full/left-atrial-appendage-occlusion-device-nicvd 6/23

The PLAATO device

The PLAATO (Percutaneous Left Atrial AppendageTranscatheter Occlusion) device was the first successfulimplanted occluder in a human.

It was first introduced in August 2001.

The PLAATO framework consists of Nitinol in a basket-

type configuration with a tissueanchoring system on thestruts designed to maintain the correct position of thedevice.

Page 7: Left Atrial Appendage   Occlusion Device, nicvd

7/27/2019 Left Atrial Appendage Occlusion Device, nicvd

http://slidepdf.com/reader/full/left-atrial-appendage-occlusion-device-nicvd 7/23

A nonthrombogenic PTFE membrane covers thebasket structure and is designed to seal the orificeof the LAA and allow tissue adhesion.

The device is available in seven sizes ranging from15 to 32 mm in diameter.

The PLAATO device is introduced via a 14-Fr gaugetransseptal sheath system. 

The PLAATO device

Page 8: Left Atrial Appendage   Occlusion Device, nicvd

7/27/2019 Left Atrial Appendage Occlusion Device, nicvd

http://slidepdf.com/reader/full/left-atrial-appendage-occlusion-device-nicvd 8/23

The PLAATO device 

A.PLAATO deviceB. Angiographic view of the PLAATO device after its release

into the left atrial appendage from the delivery catheterC. The device in situ as seen on a plain radiograph

Page 9: Left Atrial Appendage   Occlusion Device, nicvd

7/27/2019 Left Atrial Appendage Occlusion Device, nicvd

http://slidepdf.com/reader/full/left-atrial-appendage-occlusion-device-nicvd 9/23

Angiogram of the LAA: (a) after venous and transseptal puncture a specially

designed transseptal sheath is advanced into the LAA,dye injection provides information on the diameter ofthe ostium and the morphology of the LAA

b) proximal and (c) distal dye injection after device’s expansion reveal the degree of occlusion.

The PLAATO device

Page 10: Left Atrial Appendage   Occlusion Device, nicvd

7/27/2019 Left Atrial Appendage Occlusion Device, nicvd

http://slidepdf.com/reader/full/left-atrial-appendage-occlusion-device-nicvd 10/23

Transesophageal echocardiography shows the LAA

before(left) and after (right) the PLAATO procedure

The PLAATO device

Page 11: Left Atrial Appendage   Occlusion Device, nicvd

7/27/2019 Left Atrial Appendage Occlusion Device, nicvd

http://slidepdf.com/reader/full/left-atrial-appendage-occlusion-device-nicvd 11/23

The WATCHMAN filter system

The Watchman implant is the newest of the LAA

occluding systems and was very recently introduced

into clinical studies (2005).

It is a self-expanding nitinol frame structure coveredwith a permeable (160-micron pores) polyester fabricthat faces the left atrial side of the device.

Page 12: Left Atrial Appendage   Occlusion Device, nicvd

7/27/2019 Left Atrial Appendage Occlusion Device, nicvd

http://slidepdf.com/reader/full/left-atrial-appendage-occlusion-device-nicvd 12/23

The WATCHMAN filter system

It has a row of fixation barbs to ensure the greaterstability of the device and is available in diametersranging from 21 to 33 mm.

The Watchman filter system consists of an integral

12-Fr delivery system preloaded with the device and a14-Fr transseptal access sheath. The filter is releaseddistal to the left atrial ostium.

Page 13: Left Atrial Appendage   Occlusion Device, nicvd

7/27/2019 Left Atrial Appendage Occlusion Device, nicvd

http://slidepdf.com/reader/full/left-atrial-appendage-occlusion-device-nicvd 13/23

The WATCHMAN filter system

Page 14: Left Atrial Appendage   Occlusion Device, nicvd

7/27/2019 Left Atrial Appendage Occlusion Device, nicvd

http://slidepdf.com/reader/full/left-atrial-appendage-occlusion-device-nicvd 14/23

Introduction Sheath With Double and

Single Curve for Different Anatomical Situations

Page 15: Left Atrial Appendage   Occlusion Device, nicvd

7/27/2019 Left Atrial Appendage Occlusion Device, nicvd

http://slidepdf.com/reader/full/left-atrial-appendage-occlusion-device-nicvd 15/23

The WATCHMAN® device (Atritech,Inc., North Plymouth, MN)

Syed TM and Halperin JL (2007) Left atrial appendage closure for stroke

prevention in atrial fibrillation: state of the art and current challenges Nat Clin Pract Cardiovasc Med 4: 428–435 doi:10.1038/ncpcardio0933

Page 16: Left Atrial Appendage   Occlusion Device, nicvd

7/27/2019 Left Atrial Appendage Occlusion Device, nicvd

http://slidepdf.com/reader/full/left-atrial-appendage-occlusion-device-nicvd 16/23

The WATCHMAN filter system 

Successful placement of the device is defined ifthe LAA is completely sealed with absence of flowor minimal flow (jet <3 mm) as measured by TEEat 45 days.

Page 17: Left Atrial Appendage   Occlusion Device, nicvd

7/27/2019 Left Atrial Appendage Occlusion Device, nicvd

http://slidepdf.com/reader/full/left-atrial-appendage-occlusion-device-nicvd 17/23

Page 18: Left Atrial Appendage   Occlusion Device, nicvd

7/27/2019 Left Atrial Appendage Occlusion Device, nicvd

http://slidepdf.com/reader/full/left-atrial-appendage-occlusion-device-nicvd 18/23

Clinical TrialsPROTECT AF Trial: The WATCHMAN filter system 

The effects of the Watchman device on the incidenceof stroke, TIA, systemic embolism, and cardiovasculardeath will be compared to conventional therapy withdose-adjusted warfarin in a multicenter prospective

randomized study in approximately 500 patients with AF.

If TEE performed at 45 days demonstratessatisfactory LAA occlusion, patients who received aWatchman device will discontinue therapy withwarfarin and will receive clopidogrel 75 mg for 6months followed by aspirin 325 mg for the duration ofthe trial (5 years). The study started in December2005.

Page 19: Left Atrial Appendage   Occlusion Device, nicvd

7/27/2019 Left Atrial Appendage Occlusion Device, nicvd

http://slidepdf.com/reader/full/left-atrial-appendage-occlusion-device-nicvd 19/23

The PLAATO device:

Within two prospective, multi-center trials, LAAocclusion was attempted in 111 patients (age 71-91years) with PLAATO device since August 2001. Allpatients had a contraindication for anticoagulationtherapy and at least one additional risk factor forstroke.

Among 111 patients, the complication rate was 6% to

7%; the annual rate for stroke or transient ischemicattacks, 5.5%; and the annual mortality, 6.6%.

Clinical Trials

Page 20: Left Atrial Appendage   Occlusion Device, nicvd

7/27/2019 Left Atrial Appendage Occlusion Device, nicvd

http://slidepdf.com/reader/full/left-atrial-appendage-occlusion-device-nicvd 20/23

Indication

Age> 18 years with high risk of stroke.

Paroxysmal, persistent or permanent AF.

Contraindications to long term anticoagulationtherapy.

Intolerance to anticoagulation therapy.

Bleeding complications.

S

Page 21: Left Atrial Appendage   Occlusion Device, nicvd

7/27/2019 Left Atrial Appendage Occlusion Device, nicvd

http://slidepdf.com/reader/full/left-atrial-appendage-occlusion-device-nicvd 21/23

Summery

The most significant risk associated with atrial

fibrillation (AF) is arterial thromboembolism andstroke.

Conventional anticoagulant therapy is highlyeffective but is not consistently applied to allthose at risk because of the inconvenience ofregular INR monitoring and fears relating to

bleeding complications.

Page 22: Left Atrial Appendage   Occlusion Device, nicvd

7/27/2019 Left Atrial Appendage Occlusion Device, nicvd

http://slidepdf.com/reader/full/left-atrial-appendage-occlusion-device-nicvd 22/23

A raft of new drugs and mechanically basedtherapies is now being actively investigated, and

results are promising so far.

It is very likely that the clinical approach to thethromboembolism risk management strategy inpatients with AF will change substantially in thevery near future with a bit of help from cardiacinterventions.

Summery(cont.)

Page 23: Left Atrial Appendage   Occlusion Device, nicvd

7/27/2019 Left Atrial Appendage Occlusion Device, nicvd

http://slidepdf.com/reader/full/left-atrial-appendage-occlusion-device-nicvd 23/23

THANK YOU ALL