Lecture 4 Beam restricting Devices, their Use and Maintenance
• Topic 1 Importance of beam restricting devices
• Topic 2 Types of beam restricting devices
• Topic 3 Quality assurance tests of beam restricting devices
• Topic 4 Compression of area of interest
Topic 1 Importance of beam- restricting devices
Topic 1 Importance of beam restricting devices
• These devices restrict the primary x-ray beam to the area of interest.
• The larger the area covered by primary x-ray beam, the greater the scattered radiation produced.
• Scattered radiation must be minimized. They increase radiation dose and add to film darkening thus reducing visualization of details.
Topic 1 Importance of beam- restricting devices
Beam was not restricted to the chest – the area of interest. Entire abdomen and part of face of baby was exposed unnecessarily.
Topic 1 Importance of beam- restricting devices
Unrestricted beam-considerable scatter
Restricted beam – less scattered radiation
Topic 2 Types of beam-restrictingdevicesa. Attached to x-ray tube housing Aperture diaphragms - lead sheets with circular, square, or rectangular openings Cones – detachable metal tubes of different lengths, shapes, and sizes of opening
Variable aperture collimators – adjustable lead plates or shutters , contain "cross hairs", a light source, and a mirror to project the light
Beam- restricting devices
Removablemetal cone
Variable aperture collimator
Variable aperture, multiple shutter collimator
Diaphragm closer to window reduces off-focus x rays.
Smaller sized field reduces size of primary beam.
Multiple shutters
Multiple shutters
Topic 2 Types of beam - restricting devicesb. Not attached to x ray tube housing
Lead blockers – used to divide a
cassette when more than one
exposure is made on a single film;
reduces unnecessary exposure of
film thus improving image quality
Use of lead blockers• Place blocker on couch for lateral projections of thoracic,
lumbar, sacrum or
coccyx spine.• Position blocker next to patient's back to absorb radiation
that will not pass through spine.• Defective lead rubber aprons could be cut
into a range of sizes to serve as blockers.
Topic 3 Quality assurance tests of beam limiting devices• Collimation or limiting the x-ray beam to the area of
interest requires use of beam restricting devices. It is good practice to collimate such that the resulting image has collimated edges on all four sides of film
Topic 3 Quality assurance tests of beam- restricting devices
• Bring tube to rest on table top to check that all four edges of square light beam diaphragm touch the table top when using vertical beam. Do several times during the day.• Do the same test with cone inserted in front of the x ray tube window.
Caution: Cones sometimes become bent when dropped or bumped.
Topic 3 Quality Assurance tests of beam restricting devices• Do collimator-beam alignment check at least every six
months.
• Do check of alignment of centre of x ray
beam at least every six months.
Collimator-beam alignment test
• Using 100 cm focus-film distance (FFD), set collimator at 20 cm x 20 cm field size on surface of loaded cassette.
• Place metal coins or paper clips on top of cassette as shown in next slide.
• Expose the film at 60 kVp and 4 – 8 mAs.• Process the film.• Check whether the distance between outer edges of image and of paper clips/coins is
within +/- 2% of FFD.
Line diagram showing metal coins and field size
Coin to identify right upper corner.
Edge of light.
Cassette.
Collimator-beam alignment test result
Check of alignment of centre of x-ray beam• Place unexposed loaded cassette in centre of bucky tray
and centre tube to cassette.• Move tube to 100 cm FFD to bucky tray. Reduce
longitudinal collimators to a thin slit (e.g. 0.5 cm). Close lateral collimators. Expose using 60 kVp and 4 – 8 mAs.
• Do not remove cassette.
Check of alignment of centre of x ray beam (cont)
• Close longitudinal collimators and open lateral collimators to a thin slit. Expose film again. Process film.
• Bend film in half and check that exposed "cross" is in film centre.
• Acceptable deviation: 1 cm on either side of centre.
• Can also be done for non-bucky radiography. Just move tube so 100 cm FFD is to cassette top.
Centre of beam alignment test result
Topic 3 Summary
Beam restricting devices• may be attached or not attached to tube housing • require periodic checks• are necessary for radiation protection and contribute to
good image quality • minimize primary radiation• minimize scatter radiation
Topic 4 Compression of area of interest
• Reducing patient thickness by compression minimizes production of scattered radiation, thus reducing radiation dose.
• Device used : compression band.
Topic 4 Compression of area of interest
Compression band over abdomen reduces its volume. This reduces scatter radiation.
Lecture 4 Summary• Beam-restricting devices are important in
radiation protection.• These devices should be checked regularly.• Regular performance of quality assurance tests is in