Klinik und Poliklinik für Augenheilkunde All authors have no financial interest in the subject matter of this poster. R. Khoramnia, J. P. Salgado, C.P. Lohmann, C. Winkler von Mohrenfels ASCRS 2011 Comparison Between Central Corneal Thickness Measurements in Healthy Eyes Using a New Device and 2 Other Methods
Comparison Between Central Corneal Thickness Measurements in Healthy Eyes Using a New Device and 2 Other Methods. Klinik und Poliklinik für Augenheilkunde. R. Khoramnia, J. P. Salgado, C.P. Lohmann, C. Winkler von Mohrenfels. - PowerPoint PPT Presentation
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Klinik und Poliklinik für Augenheilkunde
All authors have no financial interest in the subject matter of this poster.
R. Khoramnia, J. P. Salgado, C.P. Lohmann, C. Winkler von Mohrenfels
ASCRS 2011
Comparison Between Central Corneal Thickness Measurements in Healthy Eyes Using a New Device and 2 Other Methods
Purpose
Klinik und Poliklinik für Augenheilkunde
The TMS-5 (Tomey) is a new device for the examination of the anterior eye
segment combining Placido- and Scheimpflug-Technology. The purpose of this
study was to compare central corneal thickness (CCT) measurements using
TMS-5, LenStar/Allegro Biograph (Haag-Streit/WaveLight), and the gold
standard ultrasonic pachymetry (Pachymeter SP-3000, Tomey).
TMS-5
Foto: www.ophthalworld.de
Allegro Biograph/LenStar
Foto: www.wavelight.com
SP-3000
Foto: www.eyenovation.de
Material and methods
Klinik und Poliklinik für Augenheilkunde
48 eyes of 24 healthy volunteers were included in this comparative study. In all
eyes, CCT was measured with the TMS-5 (Tomey, Nagoya, Japan), LenStar