BL S811B User&Training Manual Intense Pulse Light
BLS811BUser&Training ManualIntense Pulse Light
BLS811BUser&Training ManualIntense Pulse Light
Intense Pulse Light Safety Guidelines
01
The Fo l l ow ing i n fo rma t i on i s p rov ided f o r t he co r rec t
ut i l izat ion of IPL Skin t reatment system.the Informat ion
i nc ludes no t on l y t he acc iden t p ro tec t i on regu la t i ons
the products comply with,but also the effective regarding
p roper use o f the p roduc ts .
1 )E lec t r i c sa fe ty regu la t ion
The EMC Di rec t ive 89/336/EEC amended by 92/31/EC
&93/68 /EECEN55015:2000+A1+A2/LUM200704025/
EMCEN61000-3-2 :2000+A2/LUM200704025/EMCEN
61000-3-3:1995+A1+A2/LUM200704025/EMCEN61547:
1997+A1/LUM200704025/EMC
2)Mater ia l ROHS tes t2002/95 /EC
Rest r i c t ion o f use o f cer ta in Hazardous
Substance in Electr ical&electronics Equipment/ROHS.
3 ) E l e c t r o m a g e t i c r a d i a t i o n s a f e t y r e g u l a t i o n L V D /
EN60601-1:2006/AGC10430904GZ05E3EMC/EN60601
-1-2 :2001+A1:2006/AGC10430904GZ05E1
4) ISO13485:2003 MEDICAL/33021200005459
5) ITC MEDICAL CE
IPL SKIN TREATMENT SYSTEM belongs to a according
to MDD direct ive,I t designed according to EN 980:2003;
EN 1041:1998; EN ISO 14971:2007; EN 60601-1-2:2007;
EN 61000-3-2:2006;EN 61000-3-3:1995+A1:2001+A2:
2006; EN 60601-2-22 :1996;
EN 60601-1-4-1996+A1-1999; EN 60601-1 :2006.
Certificate
02
BLS811BUser&Training ManualIntense Pulse Light
03
BLS811BUser&Training ManualIntense Pulse Light
Why is IPL be t te r than o ther t rea tments?
In addi t ion to t reat ing mul t ip le symptoms and locat ions
in each sess ion , IPL i s gen t le , non- invas ive and sa fe .
Because of this, there is no "down t ime." You see results
in the complexion and the skin texture of the enti re face.
You can re tu rn t o you r no rma l r ou t i nes and ac t i v i t i e s
immedia te ly . In fac t , many people even schedule the i r
s e s s i o n s d u r i n g t h e i r l u n c h b r e a k . A d d i t i o n a l l y, I P L
treatments are customized to your specif ic skin type and
co lo r , p rov id ing the bes t poss ib le resu l ts .
What a reas can be t rea ted w i th IPL sys tems?
T h e I P L S y s t e m c a n r e m o v e h a i r f r o m a l l p a r t s o f t h e
b o d y i n c l u d i n g t h e f a c e , l e g s , a r m s , u n d e r a r m s , a n d
back . I t can a l so t r ea t sens i t i ve a reas l i ke t he ches t ,
n ipp les and b ik in i l i ne .
What do IPL t rea tments fee l l i ke?
As the name implies, the IPL hand-piece cools and sooth
es as i t g l ides a long your sk in. When the pulse of l ights
del ivered, some pat ients exper ience a mi ld p inching or
stinging sensation. No local anesthesia or pain medication
is typical ly required, however some pat ients may prefer
t o u s e a t o p i c a l a n e s t h e t i c w h e n s e n s i t i v e a r e a s a r e
t rea ted .
How many t rea tments w i l l I need?
Most areas require 3-6 treatments, but the number of IPL
t r e a t m e n t s y o u m a y r e q u i r e f o r o p t i m a l l o n g - t e r m
benef i t s depends on var ious fac to rs such as your ha i r
t h i c k n e s s , t h e a r e a t r e a t e d , a n d y o u r h a i r g r o w t h
Question
04
BLS811BUser&Training ManualIntense Pulse Light
Question
cyc le . In te res t ing ly, no t a l l ha i rs a re ac t i ve ly g rowing
a t t h e s a m e t i m e , b u t a c t u a l l y g o t h r o u g h t h r e e
d is t i nc t phases ; g rowth , reg ress ion , and res t ing . IPL
sys tems wo rk by d i sab l i ng t he ha i r t ha t i s p r i nc i pa l l y
in the ac t i ve g rowth phase a t the t ime o f t rea tment .
S i n c e o t h e r h a i r s w i l l e n t e r t h i s a c t i v e g r o w t h p h a s e
a t d i f f e r e n t t i m e s , a d d i t i o n a l t r e a t m e n t s m a y b e
necessa ry to d i sab le a l l o f t he ha i r f o l l i c l es i n a g i ven
a rea .
How is IPL used to t rea t ben ign vascu la r les ions?
Ag ing , sun exposu re and o the r f ac to r s can cause t he
a p p e a r a n c e o f b r o k e n c a p i l l r i e s a n d b l o o d v e s s e l s
on the face known as te lang iectas ias , which appear as
sp ider ve ins . Somet imes broken cap i l la r ies can cause
a diffused redness in the skin, cal led erythema, a source
o f soc ia l embar rassment fo r many peop le . Rosacea is
a c o m m o n s k i n c o n d i t i o n t h a t i s c h a r a c t e r i z e d b y
d i f fused e ry thema. A l l o f these cond i t ions
1.Product Information
05
BLS811BUser&Training ManualIntense Pulse Light
1.PRODUCT INFORMATION
APPLICATION
*Hair removal
*Skin rejuvenat ion
*red face, red nose, couperosis, spider veins.
*age spots, sun spots, p igmentat ion troubles
*Acne
FEATURES
*1handle
*Lamp longevi ty:120000shots*Energy:15-50J
*Pulse Number:1-15adjust
*Pulse Width:1.0-20.0
*Pulse Delay:1-50
*Shot per iod:2-5second adjust
*Display:8.0 inch Touch display
*Voltage:110V or 220V
*Power:1200W*Cool ing system:Wind+water+semiconductor
*Cert i f icat ions:CE/ROHS
*Pakcge size:66*66*110cm
*Weight:56kg
*Free part : laser glasses,funnel
opt ional
Big handle:15*50mm with 6 f i l ter
640-1200nm for hair removal
590-1200nm for red face, red nose, couperosis, spider veins. 560-1200nm for Skin rejuvenat ion
430-1200nm for Acne opt ional
480-1200nm for age spots, sun spots, p igmentat ion troubles
750-1200nm for hair removal in Dark Skin opt ional
06
2.Installation Alignment and operation procedures
2.1 Open the Package and checking all parts.
2.2put the handle holder on machine by screw.then connect
well the handle in back of machine.
2.2 put the water to Input,when water is out of overflow means
the water tank is full
2.3 Put the filter in handle
BLS811BUser&Training ManualIntense Pulse Light
Crystal
Filter
Press the blue parts,then take out the water plug
07
3.Operate the Device
3.1 Auto interface
BLS811BUser&Training ManualIntense Pulse Light
AUTOMATIC MODEL
FUNCTION MENU
Skin Type
Pigmentation
Hair Thickness
Energy
AUTO MODEL BACK DATAOFF
Hair removal Skin Rejuvenation
Skin Type: - White Skin Dark Skin
Pigmentation:None Light Medium Intense
Hair Thickness:Thin Medium Thick
08
3.Operate the Device
3.2 Manual interface
BLS811BUser&Training ManualIntense Pulse Light
Pulse Number
Pulse Width
Pulse Delay
Release
Energy
TREATMENT PARAMETER
ON MANUAL MODEL SAVE DATA
15*50mm
15
20
50
5
50
Save:Save the parameter for current
Release:time for each shot distance
3.3 Data interface
BACK
Total Counter*
7 8 9
4 5 6
1 2 3
0 Del Enter
15*50mm
15*50mm:current shot account,press it will be 0
total couonter:all shot account,need input password for 0
09
3.Operate the Device
BLS811BUser&Training ManualIntense Pulse Light
3.3 Operating&setting
Choice one model language in Screen.The system will enter
Auto model.In Auto model,the system have memory right
parameter for many treatment.Only need chice right Skin
type,Pigment,Hair thichness,then press the READY,the
Machine can start to work.Manual model,User can input
r ight parameter to f ix sufferer treatment.Choice different
filter to make different treatment.User Data show the current
shot amount and total amount.
Delete total shot amount need password.
Pulse Width
Pulse Delay
3.4 Finish the treatment.
Frist need press the ON in screen,make the lamp off.Then
turn off the key switch+urgency button+main electric switch.
Do not let the lamp be light when the user leave for long time.
Before turn on machine,
must take out the water
Plug from machine back.
10
4.Treatment Parameter
BLS811BUser&Training ManualIntense Pulse Light
Dark hair by lower width
Light hair by higher width
thickness hair by higher delay and lower energy
thin hair by lower delay and higher energy
dark skin by higher delay and lower energy
light skin by lower delay and high energy
When add the width that can not reduce the energy
when reduce the width that can not add the energy
I - II
III - IV
none
light
medium
intense
thin
medium
thick
A
B
1
2
3
4
C
D
E
SKIN TYPE
PIGMENTATION
HAIR THICKNESS
Skin rejuvenation
A1
A2
A3
A4
B1
B2
B3
B4
1
1
1
1
2
2
2
2
9.7
9.2
8.8
8.4
9.2
8.8
8.4
8.2
0
0
0
0
10
10
15
15
28
27
26
25
26
25
24
23
Pulse Width Pulse DelayPulse Number Energy
11
4.Treatment Parameter
BLS811BUser&Training ManualIntense Pulse Light
9.5
8.5
7.5
8.5
7.5
6.5
7.5
6.5
6
7
6
5.5
9
8
7
8.5
7.5
6.5
7.5
6.5
5.5
6.5
6
5.5
Hair Removal
Pulse Width Pulse DelayPulse Number
A1C
A1D
A1E
A2C
A2D
A2E
A3C
A3D
A3E
A4C
A4D
A4E
B1C
B1D
B1E
B2C
B2D
B2E
B3C
B3D
B3E
B4C
B4D
B4E
1
1
1
1
1
1
2
2
2
3
3
3
3
3
3
3
3
3
3
3
3
5
5
5
36
35
33
34
33
32
32
31
30
30
29
28
35
34
33
33
32
31
31
30
29
29
28
27
Energy
0
0
0
0
0
0
10
10
10
15
15
15
5
5
5
10
10
10
15
15
15
17
17
17
12
5.Treatment Procedure
5.1 Shave the hair on skin and left hair be 1-2mm.
5.2 Put the Gel on skin,Gel Thickness is around 1mm.
5.3 Touch well the crystal on skin.
5.4 After treatment,clean the Skin by warm water.
BLS811BUser&Training ManualIntense Pulse Light
13
6.IPL Treatment General Knowledge
6.1.Curative effect
After frist one or two treatment,the hair can lost 80%.
After 4-5 treatment,the hair will lost 95%.then no hair grow
or less thin hair grow.(Acceptable)
6.2.treatment Caution And the principle
For treatment is that the patient should feel little Acceptable
pain.Keep the skin be cool.
6 . Taboos.3
a.To avoid using the machine on the abdomen of the pregnant
women or women in menses
b.To avoid using the machine on around the parts under which
metal,plastic or silicon wee buried.
c.To avoid using the machine on around the medical and
electric instrument,such as the hearing aid,heart starter,
mechanical heart,etc.
d.To avoid using the machine on the epileptic and those in
bad health
e.To avoid us ing the machine on the wound aroung the
treatment area and the areas with metals inside of the body.
f.To avoid using the machine on people who has high blood
pressure or diabetes or serious skin diseases or heart diseases,
or who is sensit ive to electr ic current,or who is reveiving
hormone treatment.
BLS811BUser&Training ManualIntense Pulse Light
Never shot to eyes
14
7.Service
BLS811BUser&Training ManualIntense Pulse Light
7.1.Screen
Message control cable
Connect to 12Voffer the power of screen
Foot SwitchNO.1 and NO.3 Pin
15
BLS811BUser&Training ManualIntense Pulse Light
7.2.IPL Power SupplyA1
A2
A3
A4
A5
A6
A1A2 to Key Switch and Urgent Button.
A3A4 220V or 110V AC in.
A5A6 220V or 110V AC out to 12V Power box.
A7 Message control to Screen board
A8 to CAP+,A9 to CAP-,A10 to lamp+,A11 to Lamp-
A7
A8
A9 A11
A10
7.Service
16
BLS811BUser&Training ManualIntense Pulse Light
7.3.Change Lamp
Glasses tubeGlasses tube
Water
Lamp
+-Black Blue
7.3.1.Disconnect the panel of handle
7.3.2.take out the screw of the wire connection
of lamp
(The insulating tape,protecting electricity
leakage)
7.3.3.Screw off these two plastic caps.
Take care inside tubber airproof ring,
Draw out the break lamp from inside.
7.3.4.Put the new lamp inside,do not break
each side cable of lamp.(Very important)
7.Service
17
8.Our IPL Device superiority
6.1.Whole ABS Plastic Mould.
6.2.Automatic program for treatment,the hair and skin data is from Europe and USA famous clinical.
6.3.Super cooling system for reduce pain.
6.4.It does not present any contagious risk.
6.5.Maximum security for the client, default programs
6.6.Super treatment effect
We Are Looking for Sale agent over the world.
BLS811BUser&Training ManualIntense Pulse Light