Intense pulsed light (IPL) source measurement at NPL

Paul MillerNational Physical Laboratory

23rd May 2007NPL ORM Club

EU Physical Agents (Artificial) Optical Radiation Directive

Page 2: Intense pulsed light (IPL) source measurement at NPL

Contents

• What is an IPL?• Why measure IPLs?

– Optical radiation dosimetry needs of the UK medical/health sector

– EU Physical Agents (Artificial) Optical Radiation Directive

• Assemble & validate traceable measurement system• IPL measurements• Conclusions / recommendations

Page 3: Intense pulsed light (IPL) source measurement at NPL

What is an IPL?

• Emitting optical radiation in range 180 nm to 3,000 nm• Broadband • Non-coherent light• Intended for use on humans or animals for therapeutic,

medical diagnostic, cosmetic/aesthetic or veterinary applications

Page 4: Intense pulsed light (IPL) source measurement at NPL

Why measure IPLs?Optical Radiation Dosimetry Needs in Health Care

• UV Phototherapy

• UV Photodiagnosis & Phototesting

• Neonatal Phototherapy

• Photodynamic Therapy

• Laser Surgery and Intense Pulsed Light (IPL)

Page 5: Intense pulsed light (IPL) source measurement at NPL

Why measure IPLs?

• Rapidly advancing technology• Dramatic increase in the number of clinics • At present IPLs are very poorly characterised• Not subject to classification (unlike lasers)• Capable of causing considerable damage to skin and

possibly eyes

Page 6: Intense pulsed light (IPL) source measurement at NPL

Why measure IPLs?EU Physical Agents (Artificial) Optical Radiation Directive

• “The employer shall assess and, if necessary, measure and/or calculate the levels of exposure to optical radiation to which workers are likely to be exposed”

• “restrict exposure to the applicable limits”• “The methodology applied in assessment, measurement and/or

calculations shall follow international standards”• “Exposure not covered by standards…assessment, measurement

and/or calculations shall be carried out using available national or international science-based guidelines.

• “Assessment may take account of data provided by the manufacturers”

• “Particular attention to the following: the level, wavelength range and duration of exposure to artificial sources of optical radiation”

Page 7: Intense pulsed light (IPL) source measurement at NPL

TraceabilityTraceability

Page 8: Intense pulsed light (IPL) source measurement at NPL

Traceability

Page 9: Intense pulsed light (IPL) source measurement at NPL

Measurement - How?

• Diode arrays • Spectrographs• Broadband filters• Fast amplifiers• Oscilloscopes• Digital camcorders• Power / Energy meters

Page 10: Intense pulsed light (IPL) source measurement at NPL

Spectral Measurement

I/O Box

Trigger: Photodiode and Fast Amplifier

Page 11: Intense pulsed light (IPL) source measurement at NPL

Spectral change within pulse

500.00 550.00 600.00 650.00 700.00 750.00 800.00 850.00 900.00 950.00 1000.00

Wavelength (nm)

Normalised @ 800nm90027006300

Page 12: Intense pulsed light (IPL) source measurement at NPL

Spectral change pulse-to-pulse

650nm Treatment Handpiece IPL Settings: Fluence 24, Pulses 3, Delay 15

505 605 705 805 905 1005

Wavelength (nm)

Pulse 1

Pulse 2

Pulse 3

Page 13: Intense pulsed light (IPL) source measurement at NPL

Cut-off wavelength

585nm and 650nm Spectral Output

505 605 705 805 905 1005

Wavelength (nm)

585nm, Fluence 24

650nm, Fluence 24

Page 14: Intense pulsed light (IPL) source measurement at NPL

Temporal Measurement

ComputerNPLPhotometer

HeadVariable-Gain Low-Noise

High Speed Current Amplifier

Page 15: Intense pulsed light (IPL) source measurement at NPL

Temporal Results

650nm Treatment Head - Photometer MeasurementLumina Settings: Fluence 32, Pulses 4, Delay 15

0

0.2

0.4

0.6

0.8

1

1.2

0.0000 0.0100 0.0200 0.0300 0.0400 0.0500 0.0600 0.0700

Time (s)

Nor

mal

ised

Page 16: Intense pulsed light (IPL) source measurement at NPL

Conclusion

• Time resolved and time averaged spectral results.

• Pulsing shape, duration and delay between pulses.

• Drop in output during pulse train that is due to a fall in the lamp current.

• Lamp current affects the spectral distribution both between pulses and within a pulse.

Page 17: Intense pulsed light (IPL) source measurement at NPL

Conclusion (2)

• Good agreement between the two measurement systems and external data.

• System is limited in its spectral range because of1. Stray light2. Second order radiation3. CCD Quantum Efficiency.

• As yet there is no absolute calibration.

Page 18: Intense pulsed light (IPL) source measurement at NPL

Acknowledgements