Providing Solutions for Critical Environments AccuTec Services, Inc*320 NW Capital Dr*Lee's Summit, MO 64086*Phone (816) 525-8080*FAX (816) 525-8081*www.atsiusa.com Prepared For Prepared By Project No.: Report Date: Fumehood Performance Test Report Labconco Corporation 6' Protector ECHO ASHRAE 110 TEST 142327 02-12-2015
26

Fumehood Performance Test Report - Labconco SF6... · 2017-08-25 · Labconco Corporation Project No: 142327 Fume Hood Performance Testing January 19, 2015 ASHRAE 110 Test 1.0 Introduction

May 31, 2020

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Fumehood Performance Test Report - Labconco SF6... · 2017-08-25 · Labconco Corporation Project No: 142327 Fume Hood Performance Testing January 19, 2015 ASHRAE 110 Test 1.0 Introduction

Providing Solutions for Critical Environments

AccuTec Services, Inc*320 NW Capital Dr*Lee's Summit, MO 64086*Phone (816) 525-8080*FAX (816) 525-8081*www.atsiusa.com

Prepared For

Prepared By

Project No.:

Report Date:

Fumehood Performance Test Report

Labconco Corporation

6' Protector ECHO ASHRAE 110 TEST

142327

02-12-2015

Page 2: Fumehood Performance Test Report - Labconco SF6... · 2017-08-25 · Labconco Corporation Project No: 142327 Fume Hood Performance Testing January 19, 2015 ASHRAE 110 Test 1.0 Introduction

Providing Solutions For Critical Environments www.atsiusa.com

Table of Contents

Title Section

Report Summary 1

Airflow Velocity Data 2

Airflow Visualization 3

Trace Gas Data 4

Certificates of Calibration 5

Services, Inc. AccuTec

Page 3: Fumehood Performance Test Report - Labconco SF6... · 2017-08-25 · Labconco Corporation Project No: 142327 Fume Hood Performance Testing January 19, 2015 ASHRAE 110 Test 1.0 Introduction

Section 1 Report Summary Services, Inc.

AccuTec

Page 4: Fumehood Performance Test Report - Labconco SF6... · 2017-08-25 · Labconco Corporation Project No: 142327 Fume Hood Performance Testing January 19, 2015 ASHRAE 110 Test 1.0 Introduction

Labconco Corporation Project No: 142327 Fume Hood Performance Testing January 19, 2015 ASHRAE 110 Test

1.0 Introduction

1.1 This report documents the work performed by AccuTec Services, Inc. for the Labconco Corporation, at the Labconco Manufacturing Plant located in Kansas City, MO.

1.2 The scope of the project is to test the as manufactured performance of the 6’ Protector ECHO Labconco Laboratory Fume Hood. The Protector ECHO series fume hoods are ductless bench top fume hood equipped with chemical filters.

1.3 The fume hood was tested in accordance with the ASHRAE-110 “Method of Testing Performance of Laboratory Fume Hoods”: testing protocol; to establish an “As Manufactured” rating for the equipment.

1.4 A canopy exhaust connection was used to capture and remove all tracer gas passing through the chemical filters. The canopy exhaust connection measured 750 CFM. This canopy was present throughout the testing process.

1.3.1 ASHRAE 110 testing protocol includes the following tests, utilizing the standard tracer gas sulfur hexafluoride.

1.3.1.1 Breathing zone left, center, right

1.3.1.2 Walk-By

1.3.1.3 Sash Movement Effect

1.3.1.4 Perimeter Scan

1.3.2 A mannequin, with a breathing zone 3” from the sash, is used to simulate lab personnel working in the hood.

1.3.3 Velocity measurements are taken in the center of no greater than 1ft² areas in the opening plane of the fume hood.

1.3.4 Large and small scale visible smoke pattern testing.

2.0 Summary

2.1 All test methods were performed in accordance with the guidelines established by the ASHRAE-110 “Method of Testing Performance of Laboratory Fume Hoods”

2.2 The Fume Hood airflow was balanced to within the design intake face velocity range and tested at 66 FPM.

2.3 Large and small scale smoke sources were used in the work area as well as around the front sash opening to confirm acceptable visible containment and airflow patterns.

Page 5: Fumehood Performance Test Report - Labconco SF6... · 2017-08-25 · Labconco Corporation Project No: 142327 Fume Hood Performance Testing January 19, 2015 ASHRAE 110 Test 1.0 Introduction

Labconco Corporation Project No: 142327 Fume Hood Performance Testing January 19, 2015 ASHRAE 110 Test

2.3.1 Airflow velocity data is located in Section 2 of this report. 2.3.2 Airflow visualization data is located in section 3 of this report 2.3.3 Trace gas test data is located in section 4 of this report. 2.3.4 Certificates of calibration are located in section 5 of this report

3.0 Fume Hood Ratings

3.1 The 6’ Protector ECHO ductless fume hood passed the acceptance criteria of no greater than AM-0.05 as listed in the specifications. The actual rating is shown in the table below.

IPA NIH Fume Hood Test

Fume Hood SN 140997697B Rating AM-0.000

4.0 Abbreviations

4.1 The following abbreviations may have been used throughout this report and are represented here for reference.

SME Sash Movement Effect AM As Manufactured FPM Feet per minute Sec. Seconds Cal. Calibration L/m Liters per minute ppm. Parts Per Million No. Number

Page 6: Fumehood Performance Test Report - Labconco SF6... · 2017-08-25 · Labconco Corporation Project No: 142327 Fume Hood Performance Testing January 19, 2015 ASHRAE 110 Test 1.0 Introduction

Labconco Corporation Project No: 142327 Fume Hood Performance Testing January 19, 2015 ASHRAE 110 Test

5.0 Standards Utilized

5.1 The following calibrated field equipment was used in the execution of this work. Current certificates of calibration are provided in Section 5 of this report.

Device Model Serial No. TSI Anemometer 966 P07460054 BIOS Flow Meter Definer 220 119896

USON 200700 2000513 6.0 References

5.1 ANSI/ASHRAE 110-1995 “Method of Testing Performance of Laboratory Fume Hoods”

5.2 Testing Protocol for ERLAB Ductless Fume Hood, Prepared by Exposure Control Technologies, Inc.

Page 7: Fumehood Performance Test Report - Labconco SF6... · 2017-08-25 · Labconco Corporation Project No: 142327 Fume Hood Performance Testing January 19, 2015 ASHRAE 110 Test 1.0 Introduction

Section 2 Airflow Velocity Data Services, Inc.

AccuTec

Page 8: Fumehood Performance Test Report - Labconco SF6... · 2017-08-25 · Labconco Corporation Project No: 142327 Fume Hood Performance Testing January 19, 2015 ASHRAE 110 Test 1.0 Introduction

Airflow Velocity Test Report

Project: Model:Location: Serial No.:Manufacturer: Type:Unit ID:

Position of Sash (Percent Open)Average Airflow Face Velocity (fpm)Highest Airflow Face Velocity (fpm)Lowest Airflow Face Velocity (fpm)

16" Actual Sash Height Notes: Tested with a canopy exhaust connection, 750 CFM.62.25" Actual Sash Width

16" Design Sash Height62.25" Design Sash WidthVertical Sash Configuration60 fpm Design Airflow Face Velocity

Acceptance Criteria: _ _X__ Has Been Met _____ Has Not Been Met

Standards Utilized: Cal. Due Date:

Technician: Test Date: 19-Jan-15

9565P1148030 12-Aug-15

6472

CLF

58

100%

Labconco Protector ECHO140997697BLabconco Test Lab

182610002Labconco Ductless Fume Hood

Page 1

Page 9: Fumehood Performance Test Report - Labconco SF6... · 2017-08-25 · Labconco Corporation Project No: 142327 Fume Hood Performance Testing January 19, 2015 ASHRAE 110 Test 1.0 Introduction

Airflow Velocity Test Report

Max Velocity: 77 fpm Max Velocity: 73 fpm Max Velocity: 70 fpm Max Velocity: 65 fpm Max Velocity: 73 fpm Max Velocity: 74 fpmMin Velocity: 55 fpm Min Velocity: 41 fpm Min Velocity: 57 fpm Min Velocity: 55 fpm Min Velocity: 57 fpm Min Velocity: 56 fpmAvg Velocity: 72 fpm Avg Velocity: 61 fpm Avg Velocity: 63 fpm Avg Velocity: 60 fpm Avg Velocity: 65 fpm Avg Velocity: 64 fpm

Max Velocity: 78 fpm Max Velocity: 84 fpm Max Velocity: 68 fpm Max Velocity: 73 fpm Max Velocity: 73 fpm Max Velocity: 78 fpmMin Velocity: 62 fpm Min Velocity: 60 fpm Min Velocity: 53 fpm Min Velocity: 52 fpm Min Velocity: 60 fpm Min Velocity: 60 fpm

304050607080

1 20

304050607080

1 20

304050607080

1 20

30405060708090

1 20

304050607080

1 20

304050607080

1 20

304050607080

1 20

304050607080

1 20

304050607080

1 20

304050607080

1 20

304050607080

1 20

304050607080

1 20

y p y p y p y p y p y pAvg Velocity: 67 fpm Avg Velocity: 68 fpm Avg Velocity: 58 fpm Avg Velocity: 61 fpm Avg Velocity: 67 fpm Avg Velocity: 68 fpm

Total:Avg Velocity: 64 fpmMax Velocity: 72 fpmMin Velocity: 58 fpm

Page 2

Page 10: Fumehood Performance Test Report - Labconco SF6... · 2017-08-25 · Labconco Corporation Project No: 142327 Fume Hood Performance Testing January 19, 2015 ASHRAE 110 Test 1.0 Introduction

Section 3 Airflow Visualization Services, Inc.

AccuTec

Page 11: Fumehood Performance Test Report - Labconco SF6... · 2017-08-25 · Labconco Corporation Project No: 142327 Fume Hood Performance Testing January 19, 2015 ASHRAE 110 Test 1.0 Introduction

Airflow Visualization Test Report(Local)

Project: Model:Location: Serial No.:Manufacturer: Type:Unit ID:

Challenge: _ X___ Local ___ Large Volume

16" Sash Height Narrative:62.25" Sash WidthVertical Sash Configuration

Smoke Pen Challenge Medium

Airfoil Test: _ _X__ Pass _____ Fail

Sash Perimeter Test: _ _X__ Pass _____ Fail

Sash Bottom Test: _ _X__ Pass _____ Fail

Work Surface Test: _ _X__ Pass _____ Fail

Hood Cavity Test: _ _X__ Pass _____ Fail

Sash Closed Perimeter/Interior Test: _ _X__ Pass _____ Fail

Acceptance Criteria: _ _X__ Has Been Met _____ Has Not Been Met

Notes:

Technician: Test Date:

Protector ECHO

19-Jan-15

182610002

CLF

LabconcoLabconco Test Lab

Labconco Ductless Fume Hood140997697B

Page 3

Page 12: Fumehood Performance Test Report - Labconco SF6... · 2017-08-25 · Labconco Corporation Project No: 142327 Fume Hood Performance Testing January 19, 2015 ASHRAE 110 Test 1.0 Introduction

Airflow Visualization Test Report(Large Volume)

Project: Model:Location: Serial No.:Manufacturer: Type:Unit ID:

Challenge: ____ Local _ _X__ Large Volume

16" Sash Height Narrative:62.25" Sash WidthVertical Sash Configuration

Glycol Fogger Challenge Medium

Airfoil Test: _ _X__ Pass _____ Fail

Sash Perimeter Test: _ _X__ Pass _____ Fail

Sash Bottom Test: _ _X__ Pass _____ Fail

Work Surface Test: _ _X__ Pass _____ Fail

Hood Cavity Test: _ _X__ Pass _____ Fail

Sash Closed Perimeter/Interior Test: _ _X__ Pass _____ Fail

Acceptance Criteria: _ _X__ Has Been Met _____ Has Not Been Met

Notes:

Technician: Test Date: 19-Jan-15

182610002

CLF

Labconco Protector ECHOLabconco Test Lab 140997697B

Labconco Ductless Fume Hood

Page 4

Page 13: Fumehood Performance Test Report - Labconco SF6... · 2017-08-25 · Labconco Corporation Project No: 142327 Fume Hood Performance Testing January 19, 2015 ASHRAE 110 Test 1.0 Introduction

Section 4 Trace Gas Data Services, Inc.

AccuTec

Page 14: Fumehood Performance Test Report - Labconco SF6... · 2017-08-25 · Labconco Corporation Project No: 142327 Fume Hood Performance Testing January 19, 2015 ASHRAE 110 Test 1.0 Introduction

Tracer Gas Containment Test Report

Project: Model:Location: Serial No.:Manufacturer: Type:Unit ID:

Left Center Right

Average Concentration: 0.000 0.000 0.000

Peak Concentration: 0.000 0.000 0.000

Static Mode Performance Rating: AM- 0.000

16" Sash Height Notes:62.25" Sash WidthVertical Sash Configuration

5.98 L/m Tracer Gas Release Rate

Acceptance Criteria: __X__ Has Been Met ____ Has Not Been Met

Notes:

Standards Utilized: Cal. Due Date:

Technician: Test Date:

Labconco Protector ECHOLabconco Test Lab 140997697B

Labconco Ductless Fume Hood182610002

USON 200700 13-Mar-15

19-Jan-15CLF

Page 6

Page 15: Fumehood Performance Test Report - Labconco SF6... · 2017-08-25 · Labconco Corporation Project No: 142327 Fume Hood Performance Testing January 19, 2015 ASHRAE 110 Test 1.0 Introduction

Tracer Gas Containment Test Report

Average: 0.000 ppmPeak: 0.000 ppm

0.000

0.500

1.000

0 50 100 150 200 250 300

Esca

pe (p

pm)

Seconds

Left

0.000

0.500

1.000

0 50 100 150 200 250 300

Esca

pe (p

pm)

Center

Average: 0.000 ppmPeak: 0.000 ppm

Average: 0.000 ppmPeak: 0.000 ppm

0 50 100 150 200 250 300Seconds

0.000

0.500

1.000

0 50 100 150 200 250 300

Esca

pe (p

pm)

Seconds

Right

Page 6

Page 16: Fumehood Performance Test Report - Labconco SF6... · 2017-08-25 · Labconco Corporation Project No: 142327 Fume Hood Performance Testing January 19, 2015 ASHRAE 110 Test 1.0 Introduction

Tracer Gas Containment ReportSash Movement Effect Report

Project: Model:Location: Serial No.:Manufacturer: Type:

Time Span Data Set No. Cycle 1 Open Cycle 2 Open Cycle 3 Open0-45 Secs 1 0.000 0.000 0.0021-46 Secs 2 0.000 0.000 0.0022-47 Secs 3 0.000 0.000 0.0023-48 Secs 4 0.000 0.000 0.0024-49 Secs 5 0.000 0.000 0.0025-50 Secs 6 0.000 0.000 0.0026-51 Secs 7 0.000 0.000 0.0027-52 Secs 8 0.000 0.000 0.0028-53 Secs 9 0.000 0.000 0.0029-54 Secs 10 0.000 0.000 0.00310-55 Secs 11 0.000 0.000 0.00311-56 Secs 12 0.000 0.000 0.00312-57 Secs 13 0.000 0.000 0.00313-58 Secs 14 0.000 0.000 0.00314-59 Secs 15 0.000 0.000 0.00315-60 Secs 16 0.000 0.000 0.003

Total 0.000 0.000 0.039Average 0.000 0.000 0.002

SME-AM 0.002

Sash Height Notes:Sash WidthSash ConfigurationTracer Gas Release Rate

Aceptance Criteria: _ _X__ Has Been Met _____ Has Not Been Met

Standards Utilized: Cal. Due Date:

Technician: Test Date:

Vertical

Labconco Protector ECHOLabconco Test Lab 140997697B

Labconco Ductless Fume HoodUnit ID: 182610002

Tracer Gas Concentration (ppm)

16"62.25"

CLF 19-Jan-15

5.98 L/m

USON 200700 13-Mar-15

Page 7

Page 17: Fumehood Performance Test Report - Labconco SF6... · 2017-08-25 · Labconco Corporation Project No: 142327 Fume Hood Performance Testing January 19, 2015 ASHRAE 110 Test 1.0 Introduction

Tracer Gas Containment ReportSash Movement Effect

Narrative:

-0.010

0.000

0.010

0.020

0.030

0.040

0.050

0.060

0 30 60 90 120 150 180 210 240 270 300

Esca

pe (p

pm)

Seconds

Sash Movement Effect

Sash Open

Sash ClosedSash Open Sash Open

Sash Closed

Notes: Spike detected during second closing. No known cause of detection events towardsend of test. Levels returned to zero.

Page 8

Page 18: Fumehood Performance Test Report - Labconco SF6... · 2017-08-25 · Labconco Corporation Project No: 142327 Fume Hood Performance Testing January 19, 2015 ASHRAE 110 Test 1.0 Introduction

Tracer Gas Containment ReportPerimeter Scan

Project: Model:Location: Serial No.:Manufacturer: Type:Unit ID:

Location and Concentration of Leakage:Indicate location of leaks and concentration below

28" Sash Height Notes:62.25" Sash WidthVertical Sash Configuration

182610002

Labconco Protector ECHOLabconco Test Lab 140997697B

Labconco Ductless Fume Hood

5.98 L/m Tracer Gas Release Rate

Acceptance Criteria: __X__ Has Been Met _____ Has Not Been Met

Standards Utilized: USON 200700 Cal. Due Date: 13-MAR-15

Technician: CLF Test Date: 25-Jul-12

Perimeter Scan Graph

0.00

0.05

0.10

0.15

0.20

0.25

0 50 100 150 200 250 300

Esca

pe (p

pm)

Seconds

Page 9

Page 19: Fumehood Performance Test Report - Labconco SF6... · 2017-08-25 · Labconco Corporation Project No: 142327 Fume Hood Performance Testing January 19, 2015 ASHRAE 110 Test 1.0 Introduction

Section 5 Certificates of Calibration Services, Inc.

AccuTec

Page 20: Fumehood Performance Test Report - Labconco SF6... · 2017-08-25 · Labconco Corporation Project No: 142327 Fume Hood Performance Testing January 19, 2015 ASHRAE 110 Test 1.0 Introduction
Page 21: Fumehood Performance Test Report - Labconco SF6... · 2017-08-25 · Labconco Corporation Project No: 142327 Fume Hood Performance Testing January 19, 2015 ASHRAE 110 Test 1.0 Introduction
Page 22: Fumehood Performance Test Report - Labconco SF6... · 2017-08-25 · Labconco Corporation Project No: 142327 Fume Hood Performance Testing January 19, 2015 ASHRAE 110 Test 1.0 Introduction
Page 23: Fumehood Performance Test Report - Labconco SF6... · 2017-08-25 · Labconco Corporation Project No: 142327 Fume Hood Performance Testing January 19, 2015 ASHRAE 110 Test 1.0 Introduction
Page 24: Fumehood Performance Test Report - Labconco SF6... · 2017-08-25 · Labconco Corporation Project No: 142327 Fume Hood Performance Testing January 19, 2015 ASHRAE 110 Test 1.0 Introduction
Page 25: Fumehood Performance Test Report - Labconco SF6... · 2017-08-25 · Labconco Corporation Project No: 142327 Fume Hood Performance Testing January 19, 2015 ASHRAE 110 Test 1.0 Introduction
Page 26: Fumehood Performance Test Report - Labconco SF6... · 2017-08-25 · Labconco Corporation Project No: 142327 Fume Hood Performance Testing January 19, 2015 ASHRAE 110 Test 1.0 Introduction