Effect of Electrode Posi/on on Electrophysiological and Psychophysical Parameters in CI Pa/ents with Lateral and Perimodiolar Electrode Arrays Chantal Degen, MD presented at: CI 2017 Hannover Medical School Department of Otorhinolaryngology Chairman: Thomas Lenarz MD, Ph.D.
18

Effect of Electrode Posion on Electrophysiological and ... · Effect of Electrode Posion on Electrophysiological and Psychophysical Parameters in CI Paents with Lateral and Perimodiolar

Oct 28, 2019

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Effect of Electrode Posion on Electrophysiological and ... · Effect of Electrode Posion on Electrophysiological and Psychophysical Parameters in CI Paents with Lateral and Perimodiolar

EffectofElectrodePosi/ononElectrophysiologicalandPsychophysicalParametersinCIPa/entswith

LateralandPerimodiolarElectrodeArrays

ChantalDegen,MDpresentedat:CI2017

HannoverMedicalSchoolDepartmentofOtorhinolaryngology

Chairman:ThomasLenarzMD,Ph.D.

Page 2: Effect of Electrode Posion on Electrophysiological and ... · Effect of Electrode Posion on Electrophysiological and Psychophysical Parameters in CI Paents with Lateral and Perimodiolar

Disclosures

•  FinancialSupportfromCochlearDeutschlandGmbH&Co.KG:– ConferenceFeeandTravelCosts

Page 3: Effect of Electrode Posion on Electrophysiological and ... · Effect of Electrode Posion on Electrophysiological and Psychophysical Parameters in CI Paents with Lateral and Perimodiolar

HighResolu/onCBCTImaging(Morita)

Slicethickness=125µm

Page 4: Effect of Electrode Posion on Electrophysiological and ... · Effect of Electrode Posion on Electrophysiological and Psychophysical Parameters in CI Paents with Lateral and Perimodiolar

Whythemedialwall?

MicroCTfromAvcietal.(2014)J.Comp.Neurol.

Page 5: Effect of Electrode Posion on Electrophysiological and ... · Effect of Electrode Posion on Electrophysiological and Psychophysical Parameters in CI Paents with Lateral and Perimodiolar

Cochlear Nucleus 522 „SRA“

Lateral artifacts

Cochlear Nucleus 532 „MRA“

Medial artifacts

Lateralwallvs.PerimodiolarArrays

Page 6: Effect of Electrode Posion on Electrophysiological and ... · Effect of Electrode Posion on Electrophysiological and Psychophysical Parameters in CI Paents with Lateral and Perimodiolar

CochleaMeasurementTool

CometProgram:Lexowetal.,2016.VisualizaTon,measurementandmodellingofthecochleausingrotaTngmidmodiolarsliceplanes.IntJComputAssistRadiolSurg.11(10):1855-69.doi:10.1007/s11548-016-1374-7.

Page 7: Effect of Electrode Posion on Electrophysiological and ... · Effect of Electrode Posion on Electrophysiological and Psychophysical Parameters in CI Paents with Lateral and Perimodiolar

Using3DReconstruc/ontoDeterminetheEMD

Page 8: Effect of Electrode Posion on Electrophysiological and ... · Effect of Electrode Posion on Electrophysiological and Psychophysical Parameters in CI Paents with Lateral and Perimodiolar

Electrodes

Perimodiolar Lateral wall

“MRA” 532

“Contour” 512 522

522

Preformed electrode array Straight electrode array

“SRA”

Page 9: Effect of Electrode Posion on Electrophysiological and ... · Effect of Electrode Posion on Electrophysiological and Psychophysical Parameters in CI Paents with Lateral and Perimodiolar

ElectrodePosi/on:Perimodiolarvs.LateralArraysNucleus 522

Nucleus 532

0.0

0.4

0.8

1.2

1.6

2.0

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22

EMD [mm]

Electrode No.

Electrode Modiolus Distance in Perimodiolar vs. Lateral Arrays

532 522 512

0.0

0.4

0.8

1.2

1.6

2.0

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22

EMD [mm]

Electrode No.

Electrode Modiolus Distance in Perimodiolar vs. Lateral Arrays

532 522 512

0.0

0.4

0.8

1.2

1.6

2.0

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22

EMD [mm]

Electrode No.

Electrode Modiolus Distance in Perimodiolar vs. Lateral Arrays

532 522 512

0.0

0.4

0.8

1.2

1.6

2.0

1 3 5 7 9 11 13 15 17 19 21

Electrode- Modiolus Distance

[mm]

Electrode number (basal to apical)

Nucleus 512

n= 10 per Group; Error bars: ± 1SD

Page 10: Effect of Electrode Posion on Electrophysiological and ... · Effect of Electrode Posion on Electrophysiological and Psychophysical Parameters in CI Paents with Lateral and Perimodiolar

Example:SpreadofExcita/onMeasurementCurveforElectrode15

0

20

40

60

80

100

120

140

160

180

200

10 11 12 13 14 15 16 18 19 20 21 22

ECAPAmplitudeatCurrentLevel200[µV]

MaskerElektrode

SOEWidthat½Amplitude

RecordingElectrodeProbeAcTveElectrode

Page 11: Effect of Electrode Posion on Electrophysiological and ... · Effect of Electrode Posion on Electrophysiological and Psychophysical Parameters in CI Paents with Lateral and Perimodiolar

Posi/veCorrela/onbetweenEMDandSpreadofExcita/onWidth

(Averages) Nucleus 532 (MRA: n=11) Nucleus 522 (SRA: n=10)

R²=0.41979

0.5

1.5

2.5

3.5

4.5

0.5 1 1.5

SOEWidth[mm]

Electrode-Modiolus-Distance[mm]

•  TheSOEbecomeswiderwithincreasingdistancetothemodiolus.

R²=0.89128

1

2

3

4

0.5 1 1.5

SOEWidth[mm]

Electrode-Modiolus-Distance[mm]

Page 12: Effect of Electrode Posion on Electrophysiological and ... · Effect of Electrode Posion on Electrophysiological and Psychophysical Parameters in CI Paents with Lateral and Perimodiolar

R²=0.1311

3

3.5

4

4.5

5

5.5

6

0.5 0.6 0.7 0.8 0.9 1

T-Level[nC]

Electrode-Modiolus-Distance[mm]

Nucleus 532 (MRA) Nucleus 522 (SRA)

Posi/veCorrela/onbetweenEMDandT-Levels(Averages)

•  PsychophysicalvaluesforMRA<PsychophysicalvaluesforSRA

R²=0.0383

3

3.5

4

4.5

5

5.5

6

0 0.5 1 1.5 2

T-Level[nC]

Electrode-Modiolus-Distance[mm]

Page 13: Effect of Electrode Posion on Electrophysiological and ... · Effect of Electrode Posion on Electrophysiological and Psychophysical Parameters in CI Paents with Lateral and Perimodiolar

Nucleus 532 (MRA)

Nucleus 522 (SRA)

Posi/veCorrela/onbetweenEMDandC-Levels(Averages)

R²=0.17254

8

8.5

9

9.5

10

10.5

11

11.5

12

12.5

13

0.5 0.8 1.1 1.4 1.7 2

C-Level[nC]

Electrode-Modiolus-Distance[mm]

•  PsychophysicalvaluesforMRA<PsychophysicalvaluesforSRA

R²=0.33071

8

8.5

9

9.5

10

10.5

11

11.5

12

12.5

13

0.5 0.6 0.7 0.8 0.9 1

C-Level[nC]

Electrode-Modiolus-Distance[mm]

Page 14: Effect of Electrode Posion on Electrophysiological and ... · Effect of Electrode Posion on Electrophysiological and Psychophysical Parameters in CI Paents with Lateral and Perimodiolar

0.00

4.00

8.00

12.00

16.00

1 3 5 7 9 11 13 15 17 19 21

T-/C-Level (Q) [nC]

Electrode No.

Psychophysical Thresholds

522 C-Level 512 C-Level 532 C-Level 522 T-Level 512 T-Level 532 T-Level

Page 15: Effect of Electrode Posion on Electrophysiological and ... · Effect of Electrode Posion on Electrophysiological and Psychophysical Parameters in CI Paents with Lateral and Perimodiolar

R²=0.1365

6

8

10

12

14

16

18

0.5 0.7 0.9 1.1

t-NRT[nC]

AverageEMD

•  TheECAPthresholdsbecomehigherwithincreasingdistancetothemodiolus.

Posi/veCorrela/onbetweenEMDandECAPThresholds(Averages)

Nucleus 532 (MRA) Nucleus 522 (SRA)

R²=0.07094

6

8

10

12

14

16

0.2 1 1.8

t-NRT[nC]

AverageEMD

Page 16: Effect of Electrode Posion on Electrophysiological and ... · Effect of Electrode Posion on Electrophysiological and Psychophysical Parameters in CI Paents with Lateral and Perimodiolar

SpeechPercep/onatthe1stFi]ng

0

10

20

30

40

50

60

70

80

90

100

Numbers Monosyllables HSM HSM+10dBnoise

[%]MRA(n=10)

Contour(n=10)

SRA(n=10)

No significant differences in speech perception in each test*.

*ANOVA

Page 17: Effect of Electrode Posion on Electrophysiological and ... · Effect of Electrode Posion on Electrophysiological and Psychophysical Parameters in CI Paents with Lateral and Perimodiolar

•  NewCBCTresoluTonallowsformeasurementofsingleelectrodeEMDinpostoperaTveCBCTscans

•  SOEWidthincreaseswithlargerdistancetothemodiolus

•  T-Levels,C-LevelsandECAPthresholdsarelowerinperimodiolararrays

Summary

Page 18: Effect of Electrode Posion on Electrophysiological and ... · Effect of Electrode Posion on Electrophysiological and Psychophysical Parameters in CI Paents with Lateral and Perimodiolar

Thankyouforyoura_en/on.