8/12/2019 CT00-000.814.32.02.02.pdf
1/246
Siemens, 2009All documents may only be used for renderingservices on Siemens Healthcare Products. Anydocument in electronic form may be printedonce. Copy and distribution of electronic docu-ments and hardcopies is prohibited. Offenderswill be liable for damages. All other rights are re-
served.
Print No.:
SiemensWalz
SOMATOM Definition Flash
System
Installation and Startup
CT
2009
CT00-000.814.32.02.02
Replaces: CT00-000.814.32.01.02
EnglishDoc. Gen. Date: 10.09
n.a.H WSCSSD CV
Hardcopy Camera Information with SOMARIS/7
VA30 Siemens,2009Alldocuments may only be used for rendering services on SiemensHealthcareProducts. Any document in electronic formmaybe printed once.Copy and distributionof electronic documents and hardcopies is prohibited.Offenderswillbe liable for damages.Allother rights arereserved. Siemens,200910430603
8/12/2019 CT00-000.814.32.02.02.pdf
2/246
2 Copyright / Version / Disclaimer
SOMATOM Definition Flash CT00-000.814.32.02.02 Siemens, 200910.09 H WS CS SD CV
Page 2 of 246
1Copyright/ Version/ Disclaimer
Copyright
Siemens, 2009 refers to the copyright of a Siemens entity such as Siemens Aktienge-sellschaft - Germany, Siemens Shenzhen Magnetic Resonance Ltd. - China, SiemensShanghai Medical Equipment Ltd. - China, Siemens Medical Solutions USA Inc. - USA
and/or Siemens Healthcare Diagnostics Inc. - USA.
Document Version
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents corre-spond to the version at the time of system delivery and/or printout. Versions to hardcopydocumentation are not automatically distributed. Please contact your local Siemens officeto order current version or refer to our website http://www.healthcare.siemens.com.
DisclaimerSiemens provides this documentation as is without the assumption of any liability underany theory of law.
The installation and service of equipment described herein requires superior understand-ing of our equipment and may only be performed by qualified personnel who are speciallytrained for such installation and/or service.
8/12/2019 CT00-000.814.32.02.02.pdf
3/246
Siemens, 2009 CT00-000.814.32.02.02 SOMATOM Definition Flash10.09 H WS CS SD CV
Table of Contents 3
Page 3 of 246
0TableofContents
1 _______ General Hardcopy Camera Information______________________________ 7
HCC Information for SOMATOM Definition Flash. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Scope of applicability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Modality. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Image quality and optical density check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17General image quality information (LUT) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17Density verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17Reference films . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Description of Correction LUT Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20Additional and unnecessary LUT files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
2 _______ Specific Camera Information - Agfa________________________________ 23
Agfa Drystar AXYS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23SCU / modality software configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24SCP / hardcopy camera configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . . 26Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Agfa Drystar 5503. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 34Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Agfa Drystar 5500. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 42Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Agfa Drystar 5302. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 50Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Agfa Drystar 5300. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 56Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Agfa Drystar 3000. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 63Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Agfa LR5200 / LR DICOM Controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 69
8/12/2019 CT00-000.814.32.02.02.pdf
4/246
4 Table of Contents
SOMATOM Definition Flash CT00-000.814.32.02.02 Siemens, 200910.09 H WS CS SD CV
Page 4 of 246
Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
3 _______ Specific Camera Information - Codonics____________________________ 74
Codonics Horizon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 77Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
4 _______ Specific Camera Information - Fuji_________________________________ 81
Fuji Drypix 7000 / 5000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 84Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Fuji Drypix 4000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 92Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Fuji Drypix 3000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 99Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
Fuji Drypix 2000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103General Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 106Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
Fuji FM DPL/FN-PS551 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 115Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
5 _______ Specific Camera Information - CARESTREAM ______________________ 121
CARESTREAM DryView 5850 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121SCU / modality software configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 124Image quality and optical density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
Kodak DryView 6800. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129General Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129SCU / modality software configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130SCP / hardcopy camera configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 132Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135
Kodak DryView 5800. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137
8/12/2019 CT00-000.814.32.02.02.pdf
5/246
Siemens, 2009 CT00-000.814.32.02.02 SOMATOM Definition Flash10.09 H WS CS SD CV
Table of Contents 5
Page 5 of 246
General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137SCU / modality software configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138SCP / hardcopy camera configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 140Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144
Kodak DryView 8900 (small border). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 149Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151
Kodak DryView 8700-Plus / PACS Link MIM 200. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 152General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 152SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 153SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 154Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 158
Kodak DryView 8200 / PACS Link MIM 200 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 159
General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 159SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 160SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 161Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 165
Kodak DryView 8100 / PACS Link MIM 200 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 166General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 166SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 167SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 168Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 172
Kodak DryView 8150 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174
SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 176Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 179
6 _______ Specific Camera Information - Konica Minolta ______________________ 181
KONICA MINOLTA DryPro 873 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 181General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 181SCU / modality software configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 182SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 184Image quality and optical density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 189
Konica Minolta DryPro 832. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 191General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 191SCU / modality software configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 192SCP / hardcopy camera configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 194Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 199
Konica Minolta DryPro 793. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 201General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 201SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 202SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 204Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 207
Konica Minolta DryPro 771. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 209
General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 209SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 210
8/12/2019 CT00-000.814.32.02.02.pdf
6/246
6 Table of Contents
SOMATOM Definition Flash CT00-000.814.32.02.02 Siemens, 200910.09 H WS CS SD CV
Page 6 of 246
SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 211Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 216
Konica DryPro 751/752 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 217General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 217SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 218SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 219Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 222
7 _______ Specific Camera Information Sony________________________________ 224
SONY UP-DF 750 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 224General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 224SCU / modality software configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 225SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 227Image quality and optical density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 231
Sony UP-DF 550 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 233
General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 233SCU / Modality Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 234SCP / Hardcopy Camera Configuration (provided by OEM) . . . . . . . . . . . . . . . . . . 236Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 239
Sony UP-DF 500 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 241General Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 241SCU / modality software configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 242SCP / hardcopy camera configuration (provided by OEM) . . . . . . . . . . . . . . . . . . . 243Image Quality and Optical Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 245
8 _______ Changes to Previous Version ____________________________________ 246
Changes to previous version. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 246
8/12/2019 CT00-000.814.32.02.02.pdf
7/246
Siemens, 2009 CT00-000.814.32.02.02 SOMATOM Definition Flash10.09 H WS CS SD CV
General Hardcopy Camera Information 7
Page 7 of 246
1-1GeneralHardcopyCamera Information
HCC Information for SOMATOM Definition Flash 0
General information 0
Acronyms and abbreviations
Block Diagram SOMATOM Definition Flash
Fig. 1:
System references
Refer to Siemens Medical Solutions Intranet, UPTIME Services
AE Title Application entity title
BU Business Unit (Modality)
DCS DICOM Conformance Statement
DICOM Digital Imaging and Communication in Medicine
HW/SW Hardware/Software
HCC Hardcopy Camera
LAN Local Area Network
LUT Lookup Table
OEM Original Equipment Manufacturer
SCP Service Class Provider
SCU Service Class User
TCP/IP Transfer Control Protocol/Internet Protocol
8/12/2019 CT00-000.814.32.02.02.pdf
8/246
8 General Hardcopy Camera Information
SOMATOM Definition Flash CT00-000.814.32.02.02 Siemens, 200910.09 H WS CS SD CV
Page 8 of 246
Product-specific technical documentation
Hardware components
Ethernet switch ATI-FS708 (part no.: 30 79 332)
Cable, straight-through, S/UTP RJ45 (5m.; part no.: 30 74 515)
Cable, straight-through, S/UTP RJ45 (30 m.; part no.: 30 74 523)
or
Refer to Siemens Medical Solutions Intranet, UPTIME Services
Product-specific technical documentation
Restrictions and known effects
NOTE Sending film sizes that are not available at the customer site
may cause problems at the camera site. For this reason, it isrecommended that you delete these film sizes for portraitand landscape formats.
NOTE The Grayscale Standard Display Function (GSDF) is imple-mented for printing and imaging. No Presentation LUTs areavailable. For this reason, the DICOM Presentation LUT is notreleased.
NOTE The use of the presentation LUT has to be disabled by localservice for configuring LUT files.
NOTE The density values for the image quality were measured witha calibrated densitometer X-Rite 301. The result of the mea-surement from other densitometers used may differ from theabsolute density values as described in the tables.
NOTE Print the 17 graystep test image to check the modality andcamera adjustment according to the description. Verify the
results using the reference values for max. density 2.80 listedin the table. This is also part of the acceptance test.
Errors and warning messages
Only certain errors and warnings have been tested for each hardcopy camera.
These errors and warnings are described in the corresponding camera information.
.
8/12/2019 CT00-000.814.32.02.02.pdf
9/246
Siemens, 2009 CT00-000.814.32.02.02 SOMATOM Definition Flash10.09 H WS CS SD CV
General Hardcopy Camera Information 9
Page 9 of 246
Scope of applicability 0
Modality/system
Hardcopy Cameras
This table includes all hardcopy cameras using DICOM basic print.
Device type Modality (SCU)
Product name SOMATOM Definition
Modality host name/console CT host name
ICS (syngo Acquisition Workplace)
IES (syngo CT Workplace; optional)
IRS (Image Reconstruction System)
Manufacturer Siemens Med CT
SW revision Somaris 7 VA30 (syngo-based VE32D)
Conformance statement 10392318 EKL YK0 01
8/12/2019 CT00-000.814.32.02.02.pdf
10/246
10 General Hardcopy Camera Information
SOMATOM Definition Flash CT00-000.814.32.02.02 Siemens, 200910.09 H WS CS SD CV
Page 10 of 246
Manufacturer HCC nameSoftware validity
VA30 V.... V.... V....
AGFA Drystar AXYS X
AGFA Drystar 5503 X
AGFA Drystar 5500 X
AGFA Drystar 5302 X
AGFA Drystar 5300 X
AGFA Drystar 3000 X
AGFA LR5200 / LR DICOM Controller X
CODONICS Horizon X
FUJI Drypix 7000/5000 X
FUJI Drypix 4000 X
FUJI Drypix 3000 X
FUJI Drypix 2000 X
FUJI FM DPL / FN-PS 551 X
CARESTREAM DryView 5850 X
CARESTREAM [1] DryView 6800 X
CARESTREAM [1] DryView 5800 X
CARESTREAM [1] DryView 8900 (small border) X
CARESTREAM [1] DryView 8700-Plus /
PACS Link MIM 200X
CARESTREAM [1] DryView 8200 /
PACS Link MIM 200X
CARESTREAM [1] DryView 8150 X
CARESTREAM [1] DryView 8100 /
PACS Link MIM 200X
KONICA MINOLTA DryPro 873 X
KONICA MINOLTA DryPro 832 X
KONICA MINOLTA DryPro 793 X
KONICA MINOLTA DryPro 771 X
KONICA DryPro 751/752 X
SONY UP-DF750 X
8/12/2019 CT00-000.814.32.02.02.pdf
11/246
Siemens, 2009 CT00-000.814.32.02.02 SOMATOM Definition Flash10.09 H WS CS SD CV
General Hardcopy Camera Information 11
Page 11 of 246
[1] = formerly Kodak
X = hardcopy camera is released
N/A = hardcopy camera is not released
NOTE The use of non-released hardcopy cameras in conjunctionwith BU products is prohibited and does not meet therequirements specified.
DICOM associations
A DICOM association is a network connection between 2 DICOM applications. The asso-ciation is established and released by the SCU and may be aborted by the SCU or SCP.The maximum number of associations for each HCC device is defined in the DCS. Themaximum number of associations depends on the configuration and may be extended byincreasing the number of gateways connected to the printer device.
In a DICOM network environment, there are two ways to connect a modality (SCU) to anHCC device (SCP).
Case 1: The association will always be opened. Assuming that this behavior applies to allmodalities (connected to the same HCC device), the number of modalities has to be equalto or less than the maximum number of supported associations.
Case 2: The association will be opened (and closed) on demand. Assuming that thisbehavior applies to all modalities (connected to the same HCC device), the number ofmodalities may be greater than the maximum number of supported associations. This num-ber will be shared by the modalities.
The association open-close behavior is unknown to other non-SIEMENS modalities. Underoptimum circumstances, the number of modalities is equal to or less than the maximumnumber of supported associations.
The association is always opened by the SCU (case 1)
If the associations are always opened by the SCU for each configured HCC device, theassociation is reserved for this SCU only.
If the maximum number of associations = 1, and another modality never releases the asso-ciation, this HCC device is always reserved for this modality.
SONY UP-DF550 X
SONY UP-DF500 X
Manufacturer HCC nameSoftware validity
VA30 V.... V.... V....
8/12/2019 CT00-000.814.32.02.02.pdf
12/246
12 General Hardcopy Camera Information
SOMATOM Definition Flash CT00-000.814.32.02.02 Siemens, 200910.09 H WS CS SD CV
Page 12 of 246
The association is opened and closed on demand (case 2)
The associations will be opened (and closed) on demand by the SCU for each configuredHCC device via the product name.
Special cases (compare also to case 1):
If the number of modalities configured in the HCC device is greater than the maximum num-ber of supported associations, some modalities will not be able to print. The printing func-tion depends on the access time and release mechanism of each connected modality (e.g.,the maximum number of supported associations = 1, the number of modalities = 2. Duringaccess by the first modality, access by the second modality is denied and connectivity fails,and vice versa).
If other modalities do not release the associations, the shareable number of associationsdecreases.
Modality 0
Hardware and software prerequisites
All hardware and software prerequisites are included in the product deliveryvolume.
Parameters
For software configuration, HCC data/parameters and possibly additional film matrix prop-erties (syngo: filming properties) are required for software configuration.
Modality data and configuration parameters
HCC Data and Configuration Parameters1
1 The values for these parameters are part of the specific HCC information
Example: Kodak DryView 8300
Properties Example Supplier
Port number 104 Default entry
AE title PR_Hostname Default entry
TCP/IP address xxx.x.x.xxx Network administrator
Application Entity
PropertiesDisplay Supplier
AE TitleK8300
Network administrator orHCC service
Port number 5040 HCC Service
8/12/2019 CT00-000.814.32.02.02.pdf
13/246
Siemens, 2009 CT00-000.814.32.02.02 SOMATOM Definition Flash10.09 H WS CS SD CV
General Hardcopy Camera Information 13
Page 13 of 246
Film matrix properties
The settings for filming properties depend on the class selected (DICOM camera). Apull-down menu is provided for DICOM-class cameras. Select your DICOM camera type.A specific file for this type is loaded and the "filming properties" are shown. Changing the"filming properties" will affect the camera site and may not meet the requirements. Do notmake any changes, unless otherwise instructed.
NOTE Filming properties are "pixel size" and "media sheet size";this information is required for image scaling and cannot bechanged.
Software configuration
The following information is required:
- Host Name
HC Device /
General settingsDisplay Supplier
Type DICOM printer Default entry
Class (printer spec filename)
kodak8300 Default entry
DICOM Node dv8300 Default entry
Filming properties Display Supplier
Hold printed Film Jobs 10 Default entry
Min. density 20 Default entry
Pixel Size [1/1000 mm] 85.0 x 85.0 Default entry
Film Sheet Formats,
Portrait
Number of Pixels
[columns x rows]
8" x 10"
2256 x 2676Default entry
Film Sheet Formats,
Landscape
Number of Pixels
[columns x rows]
8" x 10"
2736 x 2136Default entry
Medium type Blue film Default entry
Film destination Processor Default entry
Color appearance Grayscale 8 bit Default entryBackground Black Default entry
Transformation No magnification Default entry
8/12/2019 CT00-000.814.32.02.02.pdf
14/246
14 General Hardcopy Camera Information
SOMATOM Definition Flash CT00-000.814.32.02.02 Siemens, 200910.09 H WS CS SD CV
Page 14 of 246
- TCP/IP address
- Application entity title
- Port number of the DICOM remote host (DICOM laser camera)
The AE title, port number and camera type have to be provided by the local camera
service.
syngo software configuration and laser camera
1. Select options in the menu line
2. Navigate to local service
The Siemens Service Software Authentication platform is displayed
3. Enter the password and click OK
The Home Menu Platform of the syngo Service Software is displayed
4. Click Configuration The Local Service Configuration List of system options is displayed
5. Click DICOM General
6. Click >
The local Application Entity Title platform is displayed
7. Provide the Local Camera Service with the information in the field Print (SCU) AE TitlePR_Hostname
8. Click Save; >; Finish
The Configuration Setup platform is displayed
9. Click Local Host TCP/IP LAN
The local Service Configuration TCP/IP LAN platform is displayed
10. Provide the Local Camera Service with the information in the IP address field (scrolldown menu for IP address but do not select it)
11.Click Finish
The Configuration Setup platform is displayed
12. Select DICOM Print Devices
The Local Service Hardcopy Device platform is displayed
13. Enter the host name and TCP/IP address of the laser camera (provided by the cameramanufacturer)
14. Click Test to test IP communication (ping)
15.Click Save
Host properties successfully saved
16.Click OK
17.Click >
The General Note Properties Dialog platform is displayed
18. Under logical name: Enter the logical name of the laser camera
8/12/2019 CT00-000.814.32.02.02.pdf
15/246
Siemens, 2009 CT00-000.814.32.02.02 SOMATOM Definition Flash10.09 H WS CS SD CV
General Hardcopy Camera Information 15
Page 15 of 246
19. Select the host and choose the previously defined host name in the hardcopy deviceplatform
20. Under application entity properties: Enter the AE title in the edit AE title field (providedby the camera manufacturer)
21. Enter the Port Number (provided by the camera manufacturer)22.Click Add
23.Click Save
DICOM node properties successfully saved
24.Click OK
25. Select the logical name again and click Verification to test the DICOM communication(mc3echo)
26.Click ">"
The camera-specific parameter platform is displayed.
27. HC device: Enter the camera name
28. General settings: /type: DICOM printer
29. /class: e.g., agfadrystar3000, or refer to specific HCC information
30. Filming properties:
- Hold printed film job, 10 = default; (max. = 10) (films can be printed again)
- Min. density 20 = default; (can be changed in increments of 1)
- Pixel size (changes not allowed)
- Film Sheet Formats Portrait (do not modify and/or add film sizes, select and deletefilm sizes that are not available at the customer site)
- Medium type, blue film = default; clear film; paper
- Film destination, processor = default; magazine
- Color appearance, monochrome = default; (select color for printers that support anduse color media only, e.g., Drystar2000; Codonics)
- Background, black =default; white
- Transformation no magnification = default;
do not change this value; except for Kodak DV8900; replicate;
(do not change any other values under filming properties)
31.Click "Save"
Caution: Please wait for the message "device properties successfully saved"
32.Click OK
33.Click Finish
The Configuration Setup platform is displayed
34. Click > to proceed and select the next marked DICOM HC Overview
35.Click Finish
The Configuration Setup platform is displayed
36.Click LUT files
The LUT platform is displayed.
8/12/2019 CT00-000.814.32.02.02.pdf
16/246
16 General Hardcopy Camera Information
SOMATOM Definition Flash CT00-000.814.32.02.02 Siemens, 200910.09 H WS CS SD CV
Page 16 of 246
NOTE If new LUT files for one or more modalities are available on aremovable medium, e.g., floppy or CD, these files can beselected on the source drive and copied to the dependentLUT directories on the local host. Unnecessary LUT files can
be removed from the LUT directories on the local host.
37.Click ">"
Usage of Presentation LUT platform is displayed
38.Select disable
39. Modifications are necessary for additional image impressions only.
NOTE Only the topic CT image and SC image are relevant. Depend-ing on the camera, specific entries have to be made. If noth-ing else is mentioned, set the CT image and SC image to
"CTLUT" and "2.80" for maximum density. The entries for BUimage and SC image have to be identical.
40. Overall maximum density:
- 3.00 = default (depends on the max. possible value of the max. density of all imagetypes)
41. Interpolation for printing:
- Bilinear mode = default; #Smoothing Type
- Replicate #Smoothing Type
- Cubic 0, 2, 3, 4 = concerns cubic mode only42. When changes are made, click "Save"
43.Click "Finish"
44. Click "Home"; the system automatically restarts the application
NOTE This may take some time. Confirm warnings with "OK."
8/12/2019 CT00-000.814.32.02.02.pdf
17/246
Siemens, 2009 CT00-000.814.32.02.02 SOMATOM Definition Flash10.09 H WS CS SD CV
General Hardcopy Camera Information 17
Page 17 of 246
Image quality and optical density check 1.1
General image quality information (LUT) 0
Only standard LUT files are loaded. For importing/removing additional/unnecessary LUTfiles (see section Additional/unnecessary LUT files) (LINK).
The image data from the modality to the HCC is transformed/corrected by the LUTs, oncein the modality and once in the HCC. As as result, the LUT has to be set on both sides:
Refer to the Siemens Medical Solutions Intranet, UPTIME Services
Product-Specific Technical Documentation
Correction LUT is selected
Density verification 0
There are two ways to check the LUT via the film density:
The modality is configured to a specific correction LUT and the HCC is configured to linearcurves. As an alternative, the modality is configured to a linear LUT and the HCC is config-ured to linear curves.
Modality Image Type Correction LUT Max. density
CT Image CT LUT 2.80
SC Image CT LUT 2.80
HCCRefer to the specific Camera Information "DICOM printer settings".
The HCC Service performs the setting "LINEAR LUT"
8/12/2019 CT00-000.814.32.02.02.pdf
18/246
18 General Hardcopy Camera Information
SOMATOM Definition Flash CT00-000.814.32.02.02 Siemens, 200910.09 H WS CS SD CV
Page 18 of 246
Density verification for CT
Since these two checks are equivalent; it is sufficient to check the specified density.
Checking the HCC linearity may be necessary if the desired density verification (1) showsdeviations and the HCC setting needs to be changed.
NOTE Even if the HCC service does not require a test image of themodality for the HCC setting, it is easier to check this setting- and possibly further HCC corrections - with the HCC linear-ity verification (2) than with the desired density verification(1). Only the HCC linearity LUT is effective. A graphical dis-play would now show the current HCC behavior only. Thiscould be helpful for HCC Service.
Density values for CT
Measure the density values in 17 gray levels in the IEC test image (Fig. 2 / p. 18)(foreach film size).
W:256; C:0;
Fig. 2: Test image
Checkno.
Verification for Modality adjustmentReference
valueCT
Application Image Type
Correction
LUT
Max.
Density
1Desired density CT Image CT LUT 2.80 Refer to the
table forCTLUT
2HCC linearity CT Image Linear 2.80 Refer to the
table for lin-ear LUT
8/12/2019 CT00-000.814.32.02.02.pdf
19/246
Siemens, 2009 CT00-000.814.32.02.02 SOMATOM Definition Flash10.09 H WS CS SD CV
General Hardcopy Camera Information 19
Page 19 of 246
The density difference of the BU LUT (Ref. LUT) between steps 1, 2, and 3 has to be aminimum density value of 0.02. Test image 3 (SMPTE image) can be used for visual veri-fication (the 95% gray square has to be visible).
Reference films 0
Reference film
To create a reference film with thedesiredLUT adjustment, proceed as follows:
1. The layout is configurable under Option/Configuration/Filming layout
2. Use format 3.2; default center/width, for the film sheet
Step
CT LUT Linear LUT
Ref. LUTMaximumDensity
MinimumDensity
Ref. LUTMaximumdensity
Minimumdensity
1 0.20 0.23 0.17 0.20 0.23 0.17
2 0.23 0.26 0.20 0.36 0.40 0.33
3 0.28 0.32 0.24 0.53 0.56 0.49
4 0.35 0.39 0.31 0.69 0.73 0.64
5 0.43 0.48 0.38 0.85 0.90 0.80
6 0.52 0.57 0.47 1.01 1.06 0.96
7 0.63 0.69 0.57 1.18 1.23 1.12
8 0.75 0.81 0.69 1.34 1.40 1.28
9 0.88 0.95 0.82 1.50 1.57 1.44
10 1.04 1.11 0.97 1.66 1.73 1.59
11 1.21 1.28 1.14 1.83 1.90 1.75
12 1.41 1.49 1.33 1.90 2.07 1.91
13 1.64 1.72 1.56 2.15 2.23 2.07
14 1.88 1.97 1.79 2.31 2.40 2.23
15 2.13 2.22 2.04 2.48 2.57 2.38
16 2.44 2.54 2.34 2.64 2.73 2.54
17 2.80 2.90 2.70 2.80 2.90 2.70
8/12/2019 CT00-000.814.32.02.02.pdf
20/246
20 General Hardcopy Camera Information
SOMATOM Definition Flash CT00-000.814.32.02.02 Siemens, 200910.09 H WS CS SD CV
Page 20 of 246
3. Select the "Viewing" tab, select Patient Browser, navigate to the folder "CT ReferenceImages" and view the test images:
Use test images (technical images) without text and the three CT Medical Ref images(cerebrum, pancreas, (LWS) lumbar spine) with text and print the film sheet. (the
DICOM command to the printer is always "magnification none")4. Measure IEC test image 1 (Fig. 2 / p. 18)and verify the CTLUT using the table "Density
values for modality CT
Segment layout:
5. Repeat this procedure with the linearLUT adjustment
6. If the reference images are not available, please load the reference images from
H:\SITE_DATA\Reference-images\* in the storage device (import from offline)
Description of Correction LUT Files 0
Description of Correction LUT Files for CT
1 2
3 4
5 6
Segment Image Window Center
1 IEC test image 1 (17 gray levels) 256 0
2 IEC test image 2 (50% gray field) 256 0
3 Test image 3 (SMPTE test pattern) 100 0
4 Pancreas (Abdomen) 520 104
5 Cerebrum (Head) 80 53
6 LWS (Spine) 440 56
Correction LUT LUT files with Dmin.0.2 and Dmax.2.80
Linear Linear LUT
CT LUT Standard LUT; default for all cameras; adapts Dmin.of 0.2
CT02 vs. CT LUTDark area = less contrast; darker; softer; white area = more con-trast
CT03 vs. CT02Dark area = less contrast; darker; softer; white area = more con-trast
CT04 vs. CT LUT
Dark area = more contrast; darker; sharper; white area: no
changes
8/12/2019 CT00-000.814.32.02.02.pdf
21/246
Siemens, 2009 CT00-000.814.32.02.02 SOMATOM Definition Flash10.09 H WS CS SD CV
General Hardcopy Camera Information 21
Page 21 of 246
Additional and unnecessary LUT files 0
If new, released LUT files are available on a removable medium (e.g., floppy or CD). Thesefiles may be selected on the source drive and copied to the dependent LUT directories onthe local system. Unnecessary LUT files may be removed from LUT directories on the localsystem.
Importing LUT files
1. In Local Service Configuration platform, select LUT files
2. Select the (source) drive and the path where the appropriate LUT files are located
3. Select the destination LUT directory (correction LUT directory) which depends on theimage type (CTImage).
4. By pushing the copy button, the selected LUT file will be transferred to CTImage type.
5. Repeat this procedure using the appropriate LUT files for SC Images.
6. Select the destination LUT directory (correction LUT directory) which depends on the
image type (SCImage).7. Pushing the copy button transfers the selected LUT file to SCImage type.
8. Confirm the message "Files successfully copied" by clicking "OK"
Removing LUT files
Unnecessary LUT files can be removed from LUT directories on the local host by selectingthe appropriate LUT file and clicking the "Remove" button.
Click ">"
CT05 vs. CT LUTBrighter; no changes in contrast for large window width; morecontrast for head images (small window width)
CT06 vs. CT LUT Brighter than CT5; white area = less contrast
cthead1 LUT for head images, images smoother
cthead2 vs. cthead1 Images smoother
CT015Adapts Dminof 0.15 (for cameras which are not supported Dmin.command)
CT017Adapts Dmin.of 0.17 (for cameras which are not supported Dmin.command)
CT LUTdryLUT for dry cameras; adapts Dmin.of 0.22 (5% visible in SMPTEtest pattern)
CT Lung LUT for lung images
Correction LUT LUT files with Dmin.0.2 and Dmax.2.80
8/12/2019 CT00-000.814.32.02.02.pdf
22/246
22 General Hardcopy Camera Information
SOMATOM Definition Flash CT00-000.814.32.02.02 Siemens, 200910.09 H WS CS SD CV
Page 22 of 246
Real Size Scaling
This procedure should be performed after system installation and when the HardcopyCamera is ready to print. The real size behavior has to be performed with each connectedHardcopy Camera.
Real size scaling verification
1. Load the test image (water image phantom) into the viewer (e.g. 20 cm water phantom,generated with the constancy procedure by CT).
2. Insert Crosshair (Viewing Card: Tools) and copy the image to the film sheet (copyimage to film sheet button).
3. Select film sheet format 1 X 1.
4. Select "Original Image" in the filming card and print the image.
5. Use a calibrated ruler to measure the crosshairs of the printed image.
The results should correspond to the printed value underneath the scale with atolerance of 1% in both directions.
Refer to the specific camera information.
8/12/2019 CT00-000.814.32.02.02.pdf
23/246
Siemens, 2009 CT00-000.814.32.02.02 SOMATOM Definition Flash10.09 H WS CS SD CV
Specific Camera Information - Agfa 23
Page 23 of 246
2-2Specific CameraInformation - Agfa
Agfa Drystar AXYS 0
General Information 0
Block Diagram
Fig. 3:
Available Film Sizes
Performance and Connections
Associations
Maximum number of associations supported: 10
Format Release status Film type Medium type
14" x 17"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
14" x 14"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
11" x 14"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
10 x 12Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
8" x 10"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
Other n.a. n.a. n.a.
Released: Yes - Film size is available and released
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
8/12/2019 CT00-000.814.32.02.02.pdf
24/246
24 Specific Camera Information - Agfa
SOMATOM Definition Flash CT00-000.814.32.02.02 Siemens, 200910.09 H WS CS SD CV
Page 24 of 246
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image,4x5 format, Copies=1, default center/width, magnification type none.
The optimum connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
NOTE European EMC Classification is Class A.
NOTE Slight inhomogeneity visible in gray field
NOTE Typical slight line structure of the thermo head is visible intechnical image.
NOTE Two edges of the diagnostic area are cut (approx 1 mm) for8x10, 10x12, 11x14,14x17.
NOTE The background command does not work properly. The back-ground is always white.
Known effects
NOTE The built in densitometer has to be set to X-Rite 301x toachieve the required image quality.
SCU / modality software configuration 0
Time 74 sec
Application Entity
PropertiesDisplay Changes allowed
AE Title AXYS Yes
Port number 104 No
8/12/2019 CT00-000.814.32.02.02.pdf
25/246
Siemens, 2009 CT00-000.814.32.02.02 SOMATOM Definition Flash10.09 H WS CS SD CV
Specific Camera Information - Agfa 25
Page 25 of 246
HC Device /
General settingsDisplay Changes allowed
Type DICOMprinter No
Class
(Printer spec file name)
agfadrystar5503_AXYS
No
DICOM Node DSAXYS Yes
Filming properties Display Changes allowed
Hold printed film jobs 10 Yes
Min. density 20 Yes
Pixel Size [1/1000 mm] 50.0 x 50.0 No
Film sheet formats,
Portrait
Number of pixels
[columns x rows]
14" x 17"
6922 x 8368
14" x 14"
6882 x 6882
11" x 14"
5376 x 6922
10" x 12"
4880 x 5860
8" x 10"
3852 x 4880
In define new
Do not modify/add
Select only
Delete only
Film sheet formats,
Landscape
Number of pixels
[columns x rows]
14" x 17"
8368 x 6922
14" x 14"
6882 x 6882
11" x 14"
6922 x 5376
10" x 12"
5860 x 4880
8" x 10"
4880 x 3852
Medium type Blue film Yes
Film destination Processor No
Color appearance Gray scale 8 bit No
8/12/2019 CT00-000.814.32.02.02.pdf
26/246
26 Specific Camera Information - Agfa
SOMATOM Definition Flash CT00-000.814.32.02.02 Siemens, 200910.09 H WS CS SD CV
Page 26 of 246
NOTE Do not modify or add film sizes.
Do not modify pixel sizes.
Delete film sizes not available at the customer site.
SCP / hardcopy camera configuration (provided by OEM) 0
Identification
Printer Settings
Background Black Yes
Transformation No magnification No
Filming properties Display Changes allowed
Device type: DICOM printer
Device name: DRYSTAR AXYS
Manufacturer: AGFA
SW Revision: R 1.60C1
Physical interface: Ethernet 10/100 Base-Tx
Network protocol: TCP/IP DICOM
DICOM conformance statement: Document No.: 001150 Revision 1.0
Image memory: 65 GByte
Hard disk: 70 GByte
Selected Variable name Setting
n.a. Nickname CT hostname
n.a. Use this profile only for incoming call-ing AE Title
PR_hostname
n.a. Use this profile only for local SCPserver
DEFAULT_SCP
n.a. Host profile type GENERIC-GENERIC (customized)
No Number of Copies 1
No Print priority Low
No Polarity Normal
8/12/2019 CT00-000.814.32.02.02.pdf
27/246
Siemens, 2009 CT00-000.814.32.02.02 SOMATOM Definition Flash10.09 H WS CS SD CV
Specific Camera Information - Agfa 27
Page 27 of 246
No Film orientation Portrait
No Trim No
No Film Size ID (empty)
No Medium type (empty)
n.a. Print even if Film Size ID / MediumType is not supported
No
n.a. Print even if Application Type is notsupported
No
No Initial memory size allocated per printjob (Mbytes):
100
No Kernel Site, Edit Custom:
Interpolation type;
Cubic high res. -smoothing factor -0.7
No Magnification Magn. none
No Decimate/Crop behavior: Disabled
No Pixel Size ratio 1.0 / 1.0
No Perception LUT Linear
n.a. Kanamori-like value 0n.a. Custom Value empty
No Illumination (cd/m) 2000
No Reflected ambient light (cd/m2) 10
No Border density (OD x 100) Black
No Empty image density (OD x 100) Black
No Minimum density empty
No Maximum density Site 320
n.a. Header annotation Contents: Edit custom:
None
Courier 12
n.a. Footer annotation Contents: Edit custom:
None
Courier 12
n.a. Window (%) Empty
n.a. Level (%) Empty
Selected Variable name Setting
8/12/2019 CT00-000.814.32.02.02.pdf
28/246
8/12/2019 CT00-000.814.32.02.02.pdf
29/246
Siemens, 2009 CT00-000.814.32.02.02 SOMATOM Definition Flash10.09 H WS CS SD CV
Specific Camera Information - Agfa 29
Page 29 of 246
Image Quality and Optical Density 0
No specific parameter available
Real Size Scaling
Send print job to
camera if the camerais in service mode
PRINTER OFFLINE PRINTER OFFLINE Camera offline
Sent print job to cam-era with empty inputtray
SUPPLY EMPTY SUPPLY EMPTY Camera out of film
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
Camera warning orerror messages
Printer message Message displayedon SCU (*2)
Defined DCS (*1) Sent by SCP
Film size / Direction Desired (mm)Measured
(mm)
Deviationobtained fromdesired value
in %
14" x 17"Portrait
X 330.0 330.0 0.00
Y 390.0 391.0 0.26
14" x 17"
Landscape
X 410.0 411.0 0.24
Y 320.0 320.0 0.00
14" x 14"
Portrait
X 330.0 331.0 0.30
Y 320.0 320.0 0.00
14" x 14"
Landscape
X 330.0 330.0 0.00
Y 330.0 331.0 0.30
11" x 14"
Portrait
X 260.0 261.0 0.38
Y 330.0 330.5 0.15
11" x 14"
Landscape
X 330.0 330.0 0.00
Y 250.0 251.0 0.40
10 x 12
Portrait
X 230.0 230.0 0.00
Y 270.0 271.0 0.37
8/12/2019 CT00-000.814.32.02.02.pdf
30/246
30 Specific Camera Information - Agfa
SOMATOM Definition Flash CT00-000.814.32.02.02 Siemens, 200910.09 H WS CS SD CV
Page 30 of 246
10 x 12
Landscape
X 280.0 280.5 0.18
Y 220.0 220.0 0.00
8" x 10"
Portrait
X 180.0 180.0 0.00
Y 220.0 220.0 0.00
8" x 10"
Landscape
X 230.0 230.0 0.00
Y 170.0 170.0 0.00
Film size / Direction Desired (mm)Measured
(mm)
Deviationobtained fromdesired value
in %
8/12/2019 CT00-000.814.32.02.02.pdf
31/246
Siemens, 2009 CT00-000.814.32.02.02 SOMATOM Definition Flash10.09 H WS CS SD CV
Specific Camera Information - Agfa 31
Page 31 of 246
Agfa Drystar 5503 2.1
General information 0
Block diagram
Fig. 4:
Available film sizes
Performance and Connections
Associations
Maximum number of associations supported: 10
Format Release status Film Type Medium type
14" x 17"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
14" x 14"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
11" x 14"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
10 x 12Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
8" x 10"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
Other n.a. n.a. n.a.
Released: Yes - Film size is available and released
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
8/12/2019 CT00-000.814.32.02.02.pdf
32/246
32 Specific Camera Information - Agfa
SOMATOM Definition Flash CT00-000.814.32.02.02 Siemens, 200910.09 H WS CS SD CV
Page 32 of 246
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image,format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
NOTE The background command does not work properly. The back-ground is always white.
NOTE European EMC Classification is Class A.
NOTE All four edges of the diagnostic area are cut (approx 1 mm)for 8x10, 10x12, 11x14,14x17.
NOTE Typical line structure of the thermo head is visible.
Known effects
NOTE The built in densitometer has to be set to X-Rite 301x toachieve the required image quality.
SCU / Modality Software Configuration 0
Time 70 sec
Application EntityProperties
Display Changes allowed
AE Title DRY5503 Yes
Port number 104 No
8/12/2019 CT00-000.814.32.02.02.pdf
33/246
Siemens, 2009 CT00-000.814.32.02.02 SOMATOM Definition Flash10.09 H WS CS SD CV
Specific Camera Information - Agfa 33
Page 33 of 246
HC Device /
General settingsDisplay Changes allowed
Type DICOMprinter No
Class
(Printer spec file name)
agfadrystar5503_AXYS
No
DICOM Node DRY5503 Yes
Filming properties Display Changes allowed
Hold printed Film Jobs 10 Yes
Min. density 20 Yes
Pixel Size [1/1000 mm] 50.0 x 50.0 No
Film Sheet Formats,
Portrait
Number of Pixels
[columns x rows]
14" x 17"
6922 x 8368
14" x 14"
6882 x 6882
11" x 14"
5376 x 6922
10" x 12"
4880 x 5860
8" x 10"
3852 x 4880
In define new
Do not modify/add
Select only
Delete only
Film Sheet Formats,
Landscape
Number of Pixels
[columns x rows]
14" x 17"
8368 x 6922
14" x 14"
6882 x 6882
11" x 14"
6922 x 5376
10" x 12"
5860 x 4880
8" x 10"
4880 x 3852
Medium type Blue film Yes
Film destination Processor No
Color appearance Grayscale 8 bit No
8/12/2019 CT00-000.814.32.02.02.pdf
34/246
34 Specific Camera Information - Agfa
SOMATOM Definition Flash CT00-000.814.32.02.02 Siemens, 200910.09 H WS CS SD CV
Page 34 of 246
NOTE Do not modify and/or add film sizes.
Do not modify pixel sizes.
Delete film sizes not available at the customer site.
SCP / Hardcopy Camera Configuration (provided by OEM) 0
Identification
Printer settings
Background Black Yes
Transformation No magnification No
Filming properties Display Changes allowed
Device type: DICOM printer
Device name: DRYSTAR 5503
Manufacturer: AGFA
SW Revision: R 5.0.0 B2
Physical Interface: Ethernet 10/100 Base-Tx
Network protocol: TCP/IP DICOM
DICOM Conformance Statement: AGFA Medical Imaging DICOM Conform-ance Statement, DrayStar 5503, SW ver-sion 5.0
Image memory: 2 * 256 Mbyte (RAM memory)
Hard disk: min. 18 Gbyte
Selected Variable name Setting
n.a. Nickname CT hostname
n.a. Use this profile only for incoming call-ing AE Title
PR_hostname
n.a. Use this profile only for local SCPserver
DEFAULT_SCP
n.a. Host profile type GENERIC-GENERIC (customized)
No Number of Copies 1
No Print priority Low
8/12/2019 CT00-000.814.32.02.02.pdf
35/246
Siemens, 2009 CT00-000.814.32.02.02 SOMATOM Definition Flash10.09 H WS CS SD CV
Specific Camera Information - Agfa 35
Page 35 of 246
No Polarity Normal
No Film destination output tray n.a.
n.a. Tray 1 Selected
n.a. Tray 2 Deselected
n.a. Tray 3 Deselected
n.a. Tray 4 Deselected
No Film orientation Portrait
No Trim No
No Film Size ID (empty)
No Medium type empty
n.a. Print even if Film Size ID / MediumType is not supported
No
n.a. Print even if Application Type is notsupported
No
No Initially allocated memory size perprint job (Mbytes):
100
No Kernel Site, Edit Custom:
Interpolation type;Cubic high res. -smoothing factor -0.7
No Magnification Magn. none
No Pixel Size ratio 1.0 / 1.0
No Perception LUT Linear
n.a. Kanamori-like value 0
n.a. Custom Value empty
No Illumination (cd/m) 2000
No Reflected ambient light (cd/m2) 10
No Border density (OD x 100) Black
No Empty image density (OD x 100) Black
No Minimum density empty
No Maximum density Site 320
n.a. Header annotation Contents: Edit custom:
None
Courier 12
Selected Variable name Setting
8/12/2019 CT00-000.814.32.02.02.pdf
36/246
36 Specific Camera Information - Agfa
SOMATOM Definition Flash CT00-000.814.32.02.02 Siemens, 200910.09 H WS CS SD CV
Page 36 of 246
n.a. Footer annotation Contents: Edit custom:
None
Courier 12
n.a. Window (%) empty
n.a. Level (%) empty
n.a. Image orientation r0
n.a. Conformance level 1
n.a. N-event enabled Off
n.a. Mammo modality site, No
n.a. Association time-out(s) 0
n.a. Image time-out(s) 0
n.a. ALLOW IMPLICIT VR LITTLEENDIAN
Selected
n.a. ALLOW EXPLICIT VR LITTLEENDIAN
Selected
n.a. ALLOW EXPLICIT BIG ENDIAN Selected
n.a. Resolution 508
n.a. Supported abstract syntax Presentation LUT SOP class
Print job SOP class
Printer SOP class
Basic annotation box SOP class
Verification service class
Basic film box SOP class
Basic film session SOP class
Basic color PM Meta SOP class
Print Queue Management SOPINSTANCE
Basic Grey Image Box SOP Class
Basic Color Image Box SOP Class
Basic gray PM Meta SOP class
Print queue management SOP class
n.a. Early processing allowed Yes
n.a. LR like off
Selected Variable name Setting
8/12/2019 CT00-000.814.32.02.02.pdf
37/246
Siemens, 2009 CT00-000.814.32.02.02 SOMATOM Definition Flash10.09 H WS CS SD CV
Specific Camera Information - Agfa 37
Page 37 of 246
Error / Warning Messages (examples)
Image Quality and Optical Density 0
No specific parameter available
Real Size Scaling
Camera warning orerror messages
Printer message Message displayedon SCU (*2)
Defined DCS (*1) Sent by SCP
Open front door COVER OPEN COVER OPEN Camera cover,drawer or door open
Tray open or frontdoor open
BAD SUPPLY MGZ BADSUPPLY MGZ Problem with supplymagazine
Sent print job to cam-era with empty inputtray
SUPPLY EMPTY SUPPLY EMPTY Camera out of film
Camera is warmingup
PRINTER INIT PRINTER INIT Camera initializing
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
Film size / Direction Desired (mm)Measured
(mm)
Deviationobtained fromdesired value
in %
14" x 17"
Portrait
X 330.0 330.0 0.00
Y 390.0 389.5 0.13
14" x 17"
Landscape
X 410.0 409.5 0.12
Y 330.0 330.0 0.00
14" x 14"
Portrait
X 330.0 330.0 0.00
Y 320.0 319.5 0.16
14" x 14"
Landscape
X 330.0 329.5 0.15
Y 330.0 330.0 0.00
11" x 14"
Portrait
X 250.0 250.0 0.00
Y 320.0 320.0 0.00
11" x 14"
Landscape
X 330.0 330.0 0.00
Y 250.0 250.0 0.00
8/12/2019 CT00-000.814.32.02.02.pdf
38/246
38 Specific Camera Information - Agfa
SOMATOM Definition Flash CT00-000.814.32.02.02 Siemens, 200910.09 H WS CS SD CV
Page 38 of 246
10 x 12
Portrait
X 230.0 230.0 0.00
Y 270.0 270.0 0.00
10 x 12
Landscape
X 280.0 280.0 0.00
Y 230.0 230.0 0.00
8" x 10"
Portrait
X 180.0 180.0 0.00
Y 230.0 230.0 0.00
8" x 10"
Landscape
X 230.0 230.0 0.00
Y 170.0 170.0 0.00
Film size / Direction Desired (mm)Measured
(mm)
Deviationobtained fromdesired value
in %
8/12/2019 CT00-000.814.32.02.02.pdf
39/246
Siemens, 2009 CT00-000.814.32.02.02 SOMATOM Definition Flash10.09 H WS CS SD CV
Specific Camera Information - Agfa 39
Page 39 of 246
Agfa Drystar 5500 2.2
General Information 0
Block Diagram
Fig. 5:
Available Film Sizes
Performance and Connections
Associations
Maximum number of associations supported: 10
Format Release status Film type Medium type
14" x 17"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
14" x 14"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
11" x 14"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
10 x 12Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
8" x 10"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
Other n.a. n.a. n.a.
Released: Yes - Film size is available and released
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
8/12/2019 CT00-000.814.32.02.02.pdf
40/246
40 Specific Camera Information - Agfa
SOMATOM Definition Flash CT00-000.814.32.02.02 Siemens, 200910.09 H WS CS SD CV
Page 40 of 246
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image,4x5 format, Copies=1, default center/width, magnification type none.
The optimum connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
NOTE The background command does not work properly. The back-ground is always white.
NOTE All four edges of the diagnostic area are cut (approx 1 mm)for 8x10, 10x12, 11x14,14x17.
NOTE Typical line structure of the thermo head is visible.
Known effects
NOTE The built in densitometer has to be set to X-Rite 301x toachieve the required image quality.
SCU / Modality Software Configuration 0
Time 69 sec
Application Entity
PropertiesDisplay Changes allowed
AE Title DRY5500 YesPort number 104 No
HC Device /
General settingsDisplay Changes allowed
Type DICOM printer No
8/12/2019 CT00-000.814.32.02.02.pdf
41/246
Siemens, 2009 CT00-000.814.32.02.02 SOMATOM Definition Flash10.09 H WS CS SD CV
Specific Camera Information - Agfa 41
Page 41 of 246
Class
(Printer spec file name)
agfadrystar5500 No
DICOM Node DRY5500 Yes
Filming properties Display Changes allowed
Hold printed film jobs 10 Yes
Min. density 20 Yes
Pixel Size [1/1000 mm] 50.0 x 50.0 No
Film sheet formats,Portrait
Number of pixels[columns x rows]
14" x 17"6922 x 8368
14" x 14"
6882 x 6882
11" x 14"
5376 x 6922
10" x 12"
4880 x 5860
8" x 10"
3852 x 4880In define newDo not modify/add
Select only
Delete only
Film sheet formats,
Landscape
Number of pixels
[columns x rows]
14" x 17"
8368 x 6922
14" x 14"
6882 x 6882
11" x 14"
6922 x 5376
10" x 12"
5860 x 4880s
8" x 10"
4880 x 3852
Medium type Blue film Yes
Film destination Processor No
Color appearance Gray scale 8 bit No
Background Black Yes
Transformation No magnification No
HC Device /
General settingsDisplay Changes allowed
8/12/2019 CT00-000.814.32.02.02.pdf
42/246
42 Specific Camera Information - Agfa
SOMATOM Definition Flash CT00-000.814.32.02.02 Siemens, 200910.09 H WS CS SD CV
Page 42 of 246
NOTE Do not modify and/or add film sizes.
Do not modify pixel sizes.
Delete film sizes not available at the customer site.
SCP / Hardcopy Camera Configuration (provided by OEM) 0
Identification
Printer Settings
Device type: DICOM printer
Device name: DRYSTAR 5500
Manufacturer: AGFA
SW Revision: V3.0.0 C1
Physical interface: Ethernet (RJ45)
Network protocol: TCP/IP
DICOM conformance statement: AGFA Medical Imaging DICOM Conform-ance Statement, DrayStar 5500, DocumentNo.: 737
Image memory: 2 * 256 Mbyte
Hard disk: 18 GByte
Selected Variable name Setting
n.a. Nickname CT hostname
n.a. Use this profile only for incoming call-ing AE Title
PR_hostname
n.a. Use this profile only for local SCPserver
DEFAULT_SCP
n.a. Host profile type GENERIC-GENERIC (customized)
No Number of Copies 1
No Print priority Low
No Polarity Normal
No Film destination output tray n.a.
n.a. Tray 1 Selected
n.a. Tray 2 Deselected
8/12/2019 CT00-000.814.32.02.02.pdf
43/246
Siemens, 2009 CT00-000.814.32.02.02 SOMATOM Definition Flash10.09 H WS CS SD CV
Specific Camera Information - Agfa 43
Page 43 of 246
n.a. Tray 3 Deselected
n.a. Tray 4 Deselected
No Film orientation Portrait
No Trim No
No Film Size ID (empty)
No Medium type (empty)
n.a. Print even if Film Size ID / MediumType is not supported
No
n.a. Print even if Application Type is notsupported
No
No Initial memory size allocated per printjob (Mbytes):
100
No Kernel Site, Edit Custom:
Interpolation type;
Cubic high res. -smoothing factor -0.7
No Magnification Magn. none
No Pixel Size ratio 1.0 / 1.0
No Perception LUT Linearn.a. Kanamori-like value 0
n.a. Custom Value empty
No Illumination (cd/m) 2000
No Reflected ambient light (cd/m2) 10
No Border density (OD x 100) Black
No Empty image density (OD x 100) Black
No Minimum density empty
No Maximum density Site 320
n.a. Header annotation Contents: Edit custom:
None
Courier 12
n.a. Footer annotation Contents: Edit custom:
None
Courier 12
n.a. Window (%) Empty
Selected Variable name Setting
8/12/2019 CT00-000.814.32.02.02.pdf
44/246
44 Specific Camera Information - Agfa
SOMATOM Definition Flash CT00-000.814.32.02.02 Siemens, 200910.09 H WS CS SD CV
Page 44 of 246
n.a. Level (%) Empty
n.a. Image orientation r0
n.a. Conformance level 1
n.a. N-event enabled Off
n.a. Mammo modality site, No
n.a. Association time-out(s) 0
n.a. Image time-out(s) 0
n.a. ALLOW IMPLICIT VR LITTLEENDIAN
Selected
n.a. ALLOW EXPLICIT VR LITTLEENDIAN
Selected
n.a. ALLOW EXPLICIT BIG ENDIAN Selected
n.a. Resolution 508
n.a. Supported abstract syntax Presentation LUT SOP class
Print job SOP class
Printer SOP class
Basic annotation box SOP class
Verification service classBasic film box SOP class
Basic film session SOP class
Basic color PM Meta SOP class
Print Queue Management SOPINSTANCE
Basic Grey Image Box SOP Class
Basic Color Image Box SOP Class
Basic gray PM Meta SOP class
Print queue management SOP class
n.a. Early processing allowed Yes
n.a. LR like off
Selected Variable name Setting
8/12/2019 CT00-000.814.32.02.02.pdf
45/246
Siemens, 2009 CT00-000.814.32.02.02 SOMATOM Definition Flash10.09 H WS CS SD CV
Specific Camera Information - Agfa 45
Page 45 of 246
Error / Warning Messages (examples)
Image Quality and Optical Density 0
No specific parameter available
Real Size Scaling
Camera warning orerror messages
Printer message Message displayedon SCU (*2)
Defined DCS (*1) Sent by SCP
Sent print job to cam-era with front dooropen
COVER OPEN COVER OPEN Camera cover,drawer or door open
Tray open or frontdoor open
BAD SUPPLY MGZ BADSUPPLY MGZ Problem with supplymagazine
Sent print job to cam-era with empty inputtray
SUPPLY EMPTY SUPPLY EMPTY Camera out of film
Printer warming up PRINTER INIT PRINTER INIT Camera initializing
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
Film size / Direction Desired (mm)Measured
(mm)
Deviationobtained fromdesired value
in %
14" x 17"
Portrait
X 330.0 330.0 0.00
Y 400.0 399.0 0.25
14" x 17"
Landscape
X 410.0 409.0 0.24
Y 330.0 330.0 0.00
14" x 14"
Portrait
X 330.0 329.5 0.15
Y 330.0 330.0 0.00
14" x 14"
Landscape
X 330.0 330.0 0.00
Y 330.0 329.5 0.15
11" x 14"
Portrait
X 250.0 249.5 0.20
Y 330.0 330.0 0.00
11" x 14"
Landscape
X 330.0 330.0 0.00
Y 250.0 249.5 0.00
8/12/2019 CT00-000.814.32.02.02.pdf
46/246
46 Specific Camera Information - Agfa
SOMATOM Definition Flash CT00-000.814.32.02.02 Siemens, 200910.09 H WS CS SD CV
Page 46 of 246
10 x 12
Portrait
X 230.0 230.0 0.00
Y 270.0 269.5 0.19
10 x 12
Landscape
X 280.0 279.5 0.18
Y 230.0 230.0 0.00
8" x 10"
Portrait
X 180.0 179.5 0.28
Y 230.0 230.0 0.00
8" x 10"
Landscape
X 230.0 230.0 0.00
Y 170.0 169.5 0.29
Film size / Direction Desired (mm)Measured
(mm)
Deviationobtained fromdesired value
in %
8/12/2019 CT00-000.814.32.02.02.pdf
47/246
Siemens, 2009 CT00-000.814.32.02.02 SOMATOM Definition Flash10.09 H WS CS SD CV
Specific Camera Information - Agfa 47
Page 47 of 246
Agfa Drystar 5302 2.3
General information 0
Block diagram
Fig. 6:
Available film sizes
Performance and Connections
Associations
Maximum number of associations supported: 11
Format Release status Film type Medium type
14" x 17"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
14" x 14"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
11" x 14"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
10 x 12Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
8" x 10"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
Other n.a. n.a. n.a.
Released: Yes - Film size is available and released
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
8/12/2019 CT00-000.814.32.02.02.pdf
48/246
48 Specific Camera Information - Agfa
SOMATOM Definition Flash CT00-000.814.32.02.02 Siemens, 200910.09 H WS CS SD CV
Page 48 of 246
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image,format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
NOTE The background command does not work properly. The back-ground is always white.
NOTE If the film size was not included in the configuration of thecamera, the connection to the camera is not available.
Known effects
NOTE The built in densitometer has to be set to X-Rite 301x toachieve the required image quality.
SCU / Modality Software Configuration 0
Time 67 sec
Application Entity
PropertiesDisplay Changes allowed
AE Title DRY5302 Yes
Port number 104 No
HC Device /
General settingsDisplay Changes allowed
Type DICOMprinter No
Class
(Printer spec. file name)agfadrystar5302 No
DICOM Node Agfa5302 Yes
8/12/2019 CT00-000.814.32.02.02.pdf
49/246
Siemens, 2009 CT00-000.814.32.02.02 SOMATOM Definition Flash10.09 H WS CS SD CV
Specific Camera Information - Agfa 49
Page 49 of 246
NOTE Do not modify and/or add film sizes.
Do not modify pixel sizes.
Delete film sizes that are not available at the customer site.
Filming properties Display Changes allowed
Hold printed film jobs 10 Yes
Min. density 20 Yes
Pixel Size [1/1000 mm] 79.375 x 79.375 No
Film Sheet Formats,
Portrait
Number of Pixels
[columns x rows]
14" x 17"
4358 x 5232
14" x 14"
4358 x 4303
11" x 14"
3348 x 4358
10 x 123070 x 3653
8" x 10"
2376 x 3070
In Define new
Do not modify/add
Select only
Delete only
Film Sheet Formats,
Landscape
Number of Pixels
[columns x rows]
14" x 17"
5232 x 4358
14" x 14"
4303 x 4358
11" x 14"
4358 x 3348
10 x 12
3653 x 3070
8" x 10"
3070 x 2376
Medium type Blue film Yes
Film destination Processor No
Color appearance Gray scale 8 bit No
Background Black Yes
Transformation No magnification No
8/12/2019 CT00-000.814.32.02.02.pdf
50/246
50 Specific Camera Information - Agfa
SOMATOM Definition Flash CT00-000.814.32.02.02 Siemens, 200910.09 H WS CS SD CV
Page 50 of 246
SCP / Hardcopy Camera Configuration (provided by OEM) 0
Identification
Printer settings
Device type: DICOM printer
Device name: DRYSTAR 5302
Manufacturer: AGFA
SW Revision: V1.80 c1
Physical Interface: Ethernet (RJ45)
Network protocol: TCP/IP
DICOM Conformance Statement: Revision 1.1, June 2005
Image memory: 512 Mbyte (internal memory)
Hard disk: 35 GB
Selected Variable name Setting
n.a. Nickname CT hostname
n.a. Use this profile only for incoming "Calling AETitle":
PR_hostname
n.a. Use this profile only for local SCP server DEFAULT_SCP
n.a. Host Profile Type GENERIC-GENERIC (custom-ized)
No Number of Copies 1
No Print priority Low
No Polarity Normal
No Film orientation Portrait
No Trim No
No Film Size ID Deselected
No Medium type Deselected
n.a. Print even if Film Size ID / Medium Type isnot supported
No
n.a. Print even if Application Type is not sup-ported:
No
No Initial allocated memory size per print job
(Mbytes):
100
8/12/2019 CT00-000.814.32.02.02.pdf
51/246
Siemens, 2009 CT00-000.814.32.02.02 SOMATOM Definition Flash10.09 H WS CS SD CV
Specific Camera Information - Agfa 51
Page 51 of 246
No Kernel SITE (*) Edit Custom:
Interpolation type: Cubic high
res. -smoothing factor -0.7No Magnification Magn: Max. - custom scale fac-
tor 1.0
No Pixel Size Ratio: 1.0 / 1.0
No Perception LUT Linear
n.a. Kanamori-like value 0
n.a. Custom Value Deselected
No Illumination (cd/m2) 2000
No Reflected ambient light (cd/m2) 10
No Border density (OD x 100) Black
No Empty image density (OD x 100) Black
No Minimum density Empty
No Maximum density Site 300
n.a. Header annotation Contents: Edit custom:
None
Courier 12
n.a. Footer annotation Contents: Edit custom:
None
Courier 12
n.a. Window (%) Empty
n.a. Level (%) Empty
n.a. Image orientation R0
n.a. Conformance level 1
n.a. N-event enabled Off
n.a. Mammo modality Site, No.
n.a. Association time-out(s) 0
n.a. Image time-out(s) 0
n.a. ALLOW IMPLICIT VR LITTLE ENDIAN Selected
n.a. ALLOW EXPLICIT VR LITTLE ENDIAN Selected
n.a. ALLOW EXPLICIT BIG ENDIAN Selected
Selected Variable name Setting
8/12/2019 CT00-000.814.32.02.02.pdf
52/246
52 Specific Camera Information - Agfa
SOMATOM Definition Flash CT00-000.814.32.02.02 Siemens, 200910.09 H WS CS SD CV
Page 52 of 246
Error / Warning Messages (examples)
n.a. Resolution 320
n.a. Supported SOP classes: Edit
n.a. Supported abstract syntax Presentation LUT SOP class
Print job SOP class
Printer SOP class
Basic annotation box SOPclass
Verification service class
Basic film box SOP class
Basic film session SOP class
Basic color PM Meta SOPclass
Print Queue ManagementSOP INSTANCE
Basic Grey Image Box SOPClass
Basic Color Image Box SOPClass
Basic gray PM Meta SOP class
Print queue management SOP
class
n.a. LR-like Off
Action that results incamera error mes-sages
Printer messagesMessage displayed on
SCUDefined DCS(*1)
Sent by SCP (*2)
Top cover open COVER OPEN COVER OPEN Camera cover, drawer ordoor open
Open film supply tray BAD SUPPLYMGZ
BAD SUPPLYMGZ
Problem with supplymagazine
Empty input tray SUPPLY EMPTY SUPPLY EMPTY Camera out of film
Sent wrong film size n.a. n.a. Status -> aborted
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
Selected Variable name Setting
8/12/2019 CT00-000.814.32.02.02.pdf
53/246
Siemens, 2009 CT00-000.814.32.02.02 SOMATOM Definition Flash10.09 H WS CS SD CV
Specific Camera Information - Agfa 53
Page 53 of 246
NOTE If film size is not inserted in configuration of camera, connec-tion to camera is not available.
Image Quality and Optical Density 0
No specific parameter available.
Real size scaling
Film size / Direction Desired (mm)Measured
(mm)
Deviationobtained fromdesired value
in %
14" x 17"
Portrait
X 330.0 330.0 0.00
Y 390.0 391.0 0.26
14" x 17"
Landscape
X 410.0 411.0 0.24
Y 330.0 330.0 0.00
14" x 14"
Portrait
X 330.0 331.0 0.30
Y 330.0 330.0 0.00
14" x 14"
Landscape
X 330.0 330.0 0.00
Y 330.0 331.0 0.30
11" x 14"
Portrait
X 250.0 250.5 0.20
Y 330.0 330.0 0.00
11" x 14"
Landscape
X 330.0 330.0 0.00
Y 250.0 251.0 0.40
10 x 12
Portrait
X 230.0 230.0 0.00
Y 270.0 271.0 0.37
10 x 12
Landscape
X 280.0 281.0 0.36
Y 230.0 230.0 0.00
8" x 10"
Portrait
X 170.0 170.5 0.29
Y 230.0 230.0 0.00
8" x 10"
Landscape
X 230.0 230.0 0.00
Y 170.0 170.0 0.00
8/12/2019 CT00-000.814.32.02.02.pdf
54/246
54 Specific Camera Information - Agfa
SOMATOM Definition Flash CT00-000.814.32.02.02 Siemens, 200910.09 H WS CS SD CV
Page 54 of 246
Agfa Drystar 5300 2.4
General information 0
Block diagram
Fig. 7:
Available film sizes
Performance and Connections
Associations
Maximum number of associations supported: 11
Format Release status Film Type Medium type
14" x 17"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
14" x 14"Portrait n.a. N/A n.a.
Landscape n.a. N/A n.a.
11" x 14"Portrait Yes DT2B Blue-based
Landscape Yes DT2B Blue-based
10 x 12Portrait n.a. N/A n.a.
Landscape n.a. N/A n.a.
8" x 10"Portrait n.a. N/A n.a.
Landscape n.a. N/A n.a.
Other n.a. N/A n.a.
Released: Yes - Film size is available and released
Released: No - Film size is available but not released
Released: n.a. - Film size is not available
8/12/2019 CT00-000.814.32.02.02.pdf
55/246
Siemens, 2009 CT00-000.814.32.02.02 SOMATOM Definition Flash10.09 H WS CS SD CV
Specific Camera Information - Agfa 55
Page 55 of 246
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image,format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
Restrictions
NOTE For HCC-independent restrictions, refer to (Restrictions andknown effects / p. 8).
NOTE The background command does not work properly. The back-ground is always white.
Known effects
NOTE The built in densitometer has to be set to X-Rite 301x toachieve the required image quality.
SCU / Modality Software Configuration 0
Time 67 sec
Application Entity
PropertiesDisplay Changes allowed
AE Title DRY5300 Yes
Port number 104 No
HC Device /
General settingsDisplay Changes allowed
Type DICOMprinter No
Class
(Printer spec file name)agfadrystar5300 No
DICOM Node DS5300 Yes
Filming properties Display Changes allowed
Hold printed Film Jobs 10 Yes
Min. density 24 Yes
8/12/2019 CT00-000.814.32.02.02.pdf
56/246
56 Specific Camera Information - Agfa
SOMATOM Definition Flash CT00-000.814.32.02.02 Siemens, 200910.09 H WS CS SD CV
Page 56 of 246
NOTE Do not modify and/or add film sizes.
Do not modify pixel sizes.
Delete film sizes not available at the customer site.
SCP / Hardcopy Camera Configuration (provided by OEM) 0
Identification
Pixel Size [1/1000 mm] 79.375 x 79.375 No
Film Sheet Formats,
Portrait
Number of Pixels
[columns x rows]
14" x 17"
4256 x 5174
11" x 14"
3300 x 4256
In define new
Do not modify/add
Select only
Delete only
Film Sheet Formats,
Landscape
Number of Pixels
[columns x rows]
14" x 17"
5174 x 4256
11" x 14"
4256 x 3300
Medium type Blue film Yes
Film destination Processor No
Color appearance Grayscale 8 bit No
Background Black Yes
Transformation No magnification No
Filming properties Display Changes allowed
Device type: DICOM print server/printer
Device name: DRYSTAR 5300
Manufacturer: AGFA
SW Revision: 1.82 C1
Physical Interface: 10/100 Base - T
Network protocol: TCP/IP
DICOM Conformance Statement: Rev.1.2, October 24, 2005
Image memory: 512 Mbyte
Hard disk: None
8/12/2019 CT00-000.814.32.02.02.pdf
57/246
Siemens, 2009 CT00-000.814.32.02.02 SOMATOM Definition Flash10.09 H WS CS SD CV
Specific Camera Information - Agfa 57
Page 57 of 246
Printer settings
Selected Variable name Setting
n.a. Nickname CT hostname
n.a. Use this profile only for incomingCalling AE title
PR_hostname
n.a. Use this profile only for local SCPserver
All
n.a. Host Profile Type GENERIC-GENERIC (customized)
No Number of Copies 1
No Print priority Low
No Polarity Normal
No Film orientation Portrait
No Trim No
No Film Size ID Deselected
No Medium type Deselected
n.a. Print even if Film Size ID / MediumType is not supported
No
No Kernel SITE
Interpolation type: Cubic High Res -Smooth Factor -2.5