10/29/2012 1 Clinical Applications of Corneal Optical Coherence Frontiers in Optics/Laser Science XXVIII Rochester, NY, Oct 14-18, 2012 Corneal Optical Coherence Tomography David Huang, MD, PhD Weeks Professor of Ophthalmic Research Professor of Ophthalmology & Biomedical Engineering Casey Eye Institute, Oregon Health & Science University Portland, Oregon Financial Interests: Dr. D. Huang has a significiant financial interest in Optovue, a company that may have a commercial interest in the results of this research and technology. This potential individual conflict of interest has been reviewed and managed by OHSU. Optovue, Inc.: stock options, patent royalty, grants, speaker honorarium & travel support Carl Zeiss Meditec, Inc.: patent royalty New OCT Technology Lead to An Explosion of New Products for Anterior Segment Imaging Zeiss Cirrus Optovue RTVue Bioptigen Envisu Zeiss Visante Optovue iVue Heidelberg Spectralis Optopol Copernicus Heidelberg SL-OCT Tomey CASIA David Huang, MD, PhD www.COOLLab.net
30
Embed
Clinical Applications of Corneal Optical Coherence … · Clinical Applications of Corneal Optical Coherence ... Dr. D. Huang has a significiant financial interest in Optovue, ...
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
10/29/2012
1
Clinical Applications of Corneal Optical Coherence
Frontiers in Optics/Laser Science XXVIIIRochester, NY, Oct 14-18, 2012
Corneal Optical Coherence Tomography
David Huang, MD, PhDWeeks Professor of Ophthalmic Research
Professor of Ophthalmology & Biomedical Engineering
Casey Eye Institute, Oregon Health & Science University
Portland, Oregon
Financial Interests:Dr. D. Huang has a significiant financial interest in Optovue, a company that may have a commercial interest in the results of this research and technology. This potential individual conflict of interest has been reviewed and managed by OHSU.Optovue, Inc.: stock options, patent royalty, grants, speaker honorarium & travel supportCarl Zeiss Meditec, Inc.: patent royalty
New OCT Technology Lead to An Explosion of New Products for Anterior Segment Imaging
Zeiss Cirrus Optovue RTVue
Bioptigen Envisu
Zeiss Visante Optovue iVue
Heidelberg Spectralis
Optopol Copernicus
Heidelberg SL-OCT
Tomey CASIA
David Huang, MD, PhD www.COOLLab.net
10/29/2012
2
Anterior Segment OCT Systems
Dedicated systems scan wider (12-16 mm)
Systems converted from retinal scanners have
and deeper (6-7 mm)
Zeiss Visante
Heidelberg SL-OCT
Tomey CASIA
limited scan ranges (~6mm width & 2mm depth)
Optovue RTVue
Optovue iVue
Zeiss Cirrus
H id lb S t li Heidelberg Spectralis
Bioptigen Envisu
Optopol Copernicus
David Huang, MD, PhD www.COOLLab.net
Wavelength Matters
1310 nm gives deeper penetration.
830 nm systems have higher axial resolution.
(Axial resolution: 17-18 µm)
Zeiss Visante
Heidelberg SL-OCT
Tomey CASIA
(Axial resolution: 3-5 µm)
Optovue RTVue
Optovue iVue
Zeiss Cirrus
Heidelberg Spectralis
Bioptigen Envisu Bioptigen Envisu
Optopol Copernicus
David Huang, MD, PhD www.COOLLab.net
10/29/2012
3
Better penetration at 1310 nm Greater resolution achieved w/ 830 nm systems
830 nm1310 nm
Schwalbe’s line
Scleral spur
Epithelium
Endothelium
Angle recess
David Huang, MD, PhD www.COOLLab.net
Visante TD-OCT RTVue FD-OCT
Fourier vs. Time Domain
Time domain systems are slower
Fourier domain systems are much faster
Zeiss Visante (2 kHz)
Heidelberg SL-OCT (200 Hz)
Optovue RTVue (26 kHz)
Optovue iVue (26 kHz)
Zeiss Cirrus (27 kHz)
Heidelberg Spectralis (40 kHz)
Bi ti E i (17 kH ) Bioptigen Envisu (17 kHz)
Optopol Copernicus (20 kHz)
Tomey CASIA (30 kHz)
David Huang, MD, PhD www.COOLLab.net
10/29/2012
4
Clinical Applications of Corneal OCT
Keratoconus DetectionLASIK planning FS Laser AK
Guiding PTK IOL Power Calculation Intacs
David Huang, MD, PhD www.COOLLab.net
Keratoconus Diagnosis with OCT
Bing Qin, MD, PhD
Yan Li, PhD
David Huang, MD, PhD www.COOLLab.net
10/29/2012
5
Forme fruste keratoconus is the most important risk factor for post-LASIK corneal ectasiaBinder P, Analysis of ectasia after LASIK: risk factors, J Cat Refract Surg 33:1530, 2007
T h N l S i i Ab l T t l
Topography does not screen out all eyes at risk
Randleman JB et al. Risk assessment for ectasia after corneal refractive surgery. Ophthalmol 2007; 115:37-50
Topography Normal Suspicious Abnormal Total
# Ectasia Cases
25 27 33 93
Percent 27% 29% 41% 100%
OCT could provide more accurately map corneal thickness, epithelial thickness, and corneal cuvatures
1.5mm 1.5mm
Rp Ra
Dn0 = 1
n1 = 1.376
n2 = 1.336
aa R
nnK 01
pp R
nnK 12
David Huang, MD, PhD www.COOLLab.net
RTVue (Optovue, Inc.)
10/29/2012
6
Corneal Mapping over 8 Meridians(8 x 1019 A-scans) in 0.31 Second
RTVue FD-OCT(Optovue, Inc.)
David Huang, MD, PhD www.COOLLab.net
Detecting Keratoconic Thinning with OCT Pachymetric Indices
General thinning Min Min
Focal thinning Minimum - median
Asymmetric thinning I-S IT-SN
Minimum = 404 µmY = -710 µm
Y location of the Minµ
Li Y, Meisler M, Tang M, Lu A, Thakrar V, Reiser B, Huang D. Keratoconus diagnosis with optical coherence tomography pachymetry mapping. Ophthalmology 2008;115:2159-2166.
Visante Prototype (Carl Zeiss Meditec, Inc.)
10/29/2012
7
OCT Keratoconus Pachymetric VariablesParameter Explanation Keratoconus Cutoff
(1 percentile)IT-SN Average thickness of the IT octant minus that of the SN
octant< -51 μm
I-S Average thickness of the inferior (I) octant minus that of the superior (S) octant
< -49 μm
Min Minimum corneal thickness < 455 μm
Min - Med Minimum corneal thickness-median corneal thickness <-29 μm
Y Min Y coordinate of minimum corneal thickness Y < -1.35 mm
Based on 66 normal and 52 keratoconus subjects from the
Yan Li, PhD; David Huang, MD, PhD www.COOLLab.net
Based on 66 normal and 52 keratoconus subjects from the following:
David Huang, MD, PhD, Doheny Eye Institute, Los AngelesQinmei Wang, MD, Wenzhou Medical College, WenzhouRobert Brass, MD, Brass Eye Center, New York
Y location of the min >-734 -734 ~ -1069 -1070 ~ -1353 <-1353
Sum
mat
ion
Keratoconus Risk Score
Each variable will be assigned a score of 1, 2, 3 if it exceeds 20, 5, 1 percentile thresholds.The keratoconus risk score of the eye is the summation of all scores.
Magnified corneal OCT image Epithelial thickness map
Central
D=2 5 mm
Superior
Inferior
2 group of subjects: normal and keratoconus Exclusion criteria: eyes with late keratoconic
changes such as scar or hydrops
D=2,5 mm
Yan Li, PhD; David Huang, MD, PhD www.COOLLab.net
Epithelial Map Repeatability
N Central Superior Inferior
Normal 103 0.7 µm 0.8 µm 0.7 µm
Keratoconus 35 1.0 µm 1.0 µm 1.0 µm
pooled standard deviation (SD)
Li Y, Tan O, Brass R, Weiss JL, Huang D. Corneal epithelial thickness mapping by Fourier-domain optical coherence tomography. Ophthalmology; 2012 Aug 20, Epub ahead of print.
10/29/2012
10
Average epithelial thickness maps show inferotemporal thinning
Normal
thinning over apexthinning over apex
Keratoconus
Yan Li, PhD; David Huang, MD, PhD www.COOLLab.net
Epithelial Map Variables Minimum epithelial thickness in µm
Minimum – maximum (MIN-MAX) focal thinning in µm
Map standard deviation (MSD) in µm Map standard deviation (MSD) in µm
CCT = Central corneal thickness (mm)Pp = posterior corneal power (D)
ALadj = sqrt(AL) if AL < 24.4mm= sqrt(AL+0 8*(AL 24 4)) if AL > 24 4mmELP = effective lens position (mm)
CCT = central corneal thickness (mm)
ECP, effective corneal power (D)
= 1.0208 * net corneal power – 1.6622 (for myopic laser vision correction)2
= 1.1 * net corneal power – 5.736 (for hyperopic laser vision correction)3
1. Tang M, Li Y, Huang D. An intraocular lens power calculation formula based on optical coherence tomography: a pilot study. J Refract Surg 2010; 26:430-437 2. Tang M, Wang L, Koch DD, Li Y, Huang D, Intraocular Lens Power Calculation after Previous Myopic Laser Vision Correction Based on Corneal Power Measured by
Fourier-Domain Optical Coherence Tomography, J Cataract Refract Surg. In press.3. Tang M, Wang L, Koch DD, Li Y, Huang D, Intraocular Lens Power Calculation after Myopic and Hyperopic Laser Vision Correction Using Optical Coherence Tomography.
Saudi Journal of Ophthalmology. 2012 Feb; 26: 19-24
= sqrt(AL+0.8 (AL-24.4)), if AL > 24.4mmpACD = ACD-constant
= [(A-constant*0.5663)-65.6+3.595]/0.9704
Input VariablesPartial coherence biometer
(e.g. IOL-Master, Carl Zeiss Meditec) or ultrasound A-scan
1) ACD – anterior chamber depth
2) AL – axial eye length
3) Net corneal power
4) Anterior corneal power
5) Posterior corneal power
6) Central corneal thicknessFourier-domain OCT
(RTVue, Optovue, Inc.)
Maolong Tang, PhD; David Huang, MD, PhD www.COOLLab.net
10/29/2012
22
Prospective clinical study on post-laser vision correction cataract surgery
Clinical SiteMyopic LVC
(eyes)Hyperopic LVC
(eyes)(eyes) (eyes)
Cullen Eye Institute 12 6
Doheny Eye Institute 10 3
Total 22 9Total 22 9
LVC = laser vision correction
Postoperative manifest refraction spherical equivalent was evaluated at 1 month
Tang M, Wang L, Koch DD, Li Y, Huang D, Intraocular Lens Power Calculation after Myopic and Hyperopic Laser Vision Correction Using Optical Coherence Tomography. Saudi Journal of Ophthalmology. 2012 Feb; 26: 19-24
Accuracy of OCT-based IOL power calculation compared with Haigis-L after myopic LVC
Keratometry Method
IOL Formula
Prediction Error (D)
Range (D)
MAE (D)
Within 1D
IOL-Master Haigis-L -0 35 ± 0 94 (-2 32 1 30) 0 73 73%O aste a g s -0.35 ± 0.94 (-2.32, 1.30) 0.73 73%
OCT OCT 0.07 ± 0.80 (-1.24, 2.15) 0.57 82%
p = 0.19. n = 22 eyes of 16 subjects.
Maolong Tang, PhDDavid Huang, MD, PhD
www.COOLLab.net
10/29/2012
23
Accuracy of OCT-based IOL power calculation compared with other methods after myopic LVC at the Doheny Eye Institute
Keratometry IOL Formula
Prediction Range MAEMethod
IOL FormulaError (D) (D) (D)
Clinical History Hoffer-Q 0.31 ± 2.52 -1.47 to 2.09 1.78*
CL Overrefraction Holladay II 0.01 ± 2.08 -3.09 to 3.19 1.46*
IOL-Master Haigis-L -0.28 ± 1.19 -2.32 to 0.98 0.88
Orbscan II Holladay II 0.17 ± 1.61 -2.72 to 1.51 1.28
OCT OCT-based 0.49 ± 0.64 -0.33 to 1.58 0.60
*p<0.05 compared to OCT. N = 10 eyes.The standard IOL formula that provided the smallest MAE was used for all keratometry methods except OCT.
OCT-based IOL calculation was significantly better than clinical history and contact lens over-refraction methods.
Maolong Tang, PhD, David Huang, MD, PhD www.COOLLab.net
Conclusion
The predictive accuracy of OCT-based IOL power calculation is equal to or better than p qcurrent standards for post-LVC eyes.
More improvement are possible Aberration measurement or ray tracing
Better prediction of IOL position
Prediction of astigmatism outcomeg
Maolong Tang, PhD, David Huang, MD, PhD www.COOLLab.net
10/29/2012
24
Grant & Material Supports
NIH R01 EY018184: Guiding The gTreatment of Anterior Eye Diseases with Optical Coherence Tomography
Unrestricted grant from Research to Prevent BlindnessPrevent Blindness
Grant & material support from Optovue, Inc.
www.COOLLab.net
David Huang, MD, PhD
Maolong Tang, PhD
Ou Tan, PhD
Yimin Wang, PhD
Yan Li, PhD
Xinbo Zhang, PhD
Jason Tokayer, MS
Janice Van Norman, COT
Yali Jia, PhD
Kathleen S. Torok, MA
10/29/2012
25
New Technologies from The Laboratory
HIGHER SPEED GREATER DEPTH RANGE
Anterior Segment Imaging and Extraction of Biometric Parameters
• James G. Fujimoto, Massachusetts Institute of Technology
Volumetric OCT data
OCT registration
3D refraction correction
3D OCT biometry
10/29/2012
26
Pachymetry Map
I. Grulkowski et al., Volumetric anterior segment biometry using OCT registration and refraction correction techniques, manuscript in preparation
Topography
I. Grulkowski et al., Volumetric anterior segment biometry using OCT registration and refraction correction techniques, manuscript in preparation
10/29/2012
27
360° Anterior Chamber Angle Survey
I. Grulkowski et al., Volumetric anterior segment biometry using OCT registration and refraction correction techniques, manuscript in preparation
Anterior Segment Imaging Using VCSEL
• VCSEL technology
• Long coherence length
1 mm
I. Grulkowski et al., Retinal, anterior segment and full eye imaging using ultrahigh speed swept source OCT with vertical-cavity surface emitting lasers, Biomed. Opt. Express 3, 2733 (2012).
10/29/2012
28
Full Eye Imaging for Non-Contact Ocular Biometry
Cornea Intraocular DistancesCentral Corneal
0 527±0 003mm
50ms
Crystalline Lens
Thickness0.527±0.003mm
Aqueous Depth 3.305±0.030mm
Anterior Chamber Depth
3.831±0.029mm
Lens Thickness 3.880±0.030mm
Vitreous Depth 18.674±0.018mm
Axial Eye
5 mm
Retina
Axial Eye Length
26.384±0.016mm
IOL Master – 20 umImmersion Ultrasound – 100 um
Speckle free imaging – cornea in-vivo
a.
a. Epithelium & Bowman’s Layer
b. b. Endothelium
c. c.Epithelium & Bowman’s Layer
d.d.
Endothelium
Courtesy of Maciej Wojtkowski, Nicolaus Copernicus University, Torun, Poland
M. Szkulmowski,et al. "Efficient reduction of speckle noise in Optical Coherence Tomography," Opt. Express 20, 1337, 2012
10/29/2012
29
Speckle free imaging – eye lens in-vivo
Segmented tomogramSegments: 1000Averaging: 1Tomo. lines: 1000Acq. time: 250 ms
Segmented tomogramSegments: 1000Averaging: 16Tomo. lines: 1000Acq. time: 250 ms
Capsule
Epithelium
Courtesy of Maciej Wojtkowski, Nicolaus Copernicus University, Torun, Poland
M. Szkulmowski,et al. "Efficient reduction of speckle noise in Optical Coherence Tomography," Opt. Express 20, 1337, 2012
CornealTopography
Placido Tomography
Scheimpflugcamera
SS‐OCT
Keratoplasty
Corneal Penetrating Keratoplasty
Anterior surfaceK1 (ax)K2 (ax)
38.38 D (165°)43.88 D (80°)
9.64 mm 35.0 D (146.5°)7.98 mm 42.3 D (56.5°)
9.44 mm 45.8 D (146°)6.82 mm 49.5 D (56°)
Posterior surfaceK1 (ax)
N.A.( )
K2 (ax) 8.39 mm -4.8 D (146.5°)5.73 mm -7.0 (56.5°)
7.03 mm -5.7 D (154.0°)5.29 mm -7.6 D (64.0°)
Pachymetrymin. Thickness
519 µm 572 µm
Courtesy of Maciej Wojtkowski, Nicolaus Copernicus University, Torun, Poland
K. Karnowski, B. J. Kaluzny, M. Szkulmowski, M. Gora, and M. Wojtkowski, Biomedical Opt Express, 2 (9), 2709-2720, 2011
10/29/2012
30
Large scale 3-D imaging
M. Gora et al. Optics Express 17(17), 14880-14894, 2009, I. Grulkowski et al..Opt Express, 17 (6), 4842-4858, 2009.
S. Ortiz et al. Biomed. Opt. Express 2, 3232-3247 (2011) , S. Ortiz et al., Opt Express, 18 (3), 2782-2796, 2010.
K. Karnowski et al., Biomedical Opt Express, 2 (9), 2709-2720, 2011
Courtesy of Maciej Wojtkowski, Nicolaus Copernicus University, Torun, Poland
Conclusions
Future commercial OCT systems will be faster
Accurate topography of optical surfaces Accurate topography of optical surfaces Optical modeling and ray tracing