Top Banner
Beam Test of BGO Crystal Yunlong Zhang 2009.12.8
20

Beam Test of BGO Crystal Yunlong Zhang 2009.12.8.

Dec 21, 2015

Download

Documents

Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Beam Test of BGO Crystal Yunlong Zhang 2009.12.8.

Beam Test of BGO Crystal

Yunlong Zhang 2009.12.8

Page 2: Beam Test of BGO Crystal Yunlong Zhang 2009.12.8.

CSR

Page 3: Beam Test of BGO Crystal Yunlong Zhang 2009.12.8.

RIBLL

Test counter here

Page 4: Beam Test of BGO Crystal Yunlong Zhang 2009.12.8.

The configuration of beam elements and test counter

beam

TofT2-Si BGO counter

Page 5: Beam Test of BGO Crystal Yunlong Zhang 2009.12.8.

In order to extend the dynamic range of readout, we use multi-readout from the dynodes.

10k 10k10k 10k

300k 300k 300k 300k

Dy8 Dy9 Dy10 A

1R=300K, I=0.256mA at HV=1000V; 0.256W power dissipation

Readout of PMT

Page 6: Beam Test of BGO Crystal Yunlong Zhang 2009.12.8.

Data collection

Ady10Dy9dy8

Enengy signals

Time signals ACFD GG

A

dy8

Peak ADC

22Na gamma calibration:

CAEN568LC

Philip7164

Page 7: Beam Test of BGO Crystal Yunlong Zhang 2009.12.8.

The spectrum of 22Na under -1100V.

Page 8: Beam Test of BGO Crystal Yunlong Zhang 2009.12.8.
Page 9: Beam Test of BGO Crystal Yunlong Zhang 2009.12.8.

1000V Na22

Page 10: Beam Test of BGO Crystal Yunlong Zhang 2009.12.8.

1100V Na22

Page 11: Beam Test of BGO Crystal Yunlong Zhang 2009.12.8.

Fit function:

Page 12: Beam Test of BGO Crystal Yunlong Zhang 2009.12.8.

proton kinetic enengies(MeV)

(Bρ)D3=(Bρ)D4 Tm

expected

(Bρ)D3=(Bρ)D4 Tm

experiment sets

10 0.4399 0.4399

12 0.5043 0.5044

14 0.5450 0.5452

16 0.5829 0.5827

18 0.61858 0.6186

20 0.6524 0.6526

24 0.7154 0.7156

28 0.7735 0.7738

32 0.8278 0.8283

mixtue beam 1.6232 1.6625

Beam particles selection

Test with proton beam

Page 13: Beam Test of BGO Crystal Yunlong Zhang 2009.12.8.

Ady10Dy9dy8

Enengy signals

Time signals ACFD GG

A

dy8

Peak ADC

T2-SI

ORT142

ORT450

CAEN568LC

ORTlogic

enengy signals to Peak ADC

Philip744

Philip7164

Page 14: Beam Test of BGO Crystal Yunlong Zhang 2009.12.8.

Delta E vs E

Page 15: Beam Test of BGO Crystal Yunlong Zhang 2009.12.8.

Tp=14MeV

Page 16: Beam Test of BGO Crystal Yunlong Zhang 2009.12.8.

Without Si-∆E correction

T2-Si 280μm

7μm-Al foil

TOF

10μm

Page 17: Beam Test of BGO Crystal Yunlong Zhang 2009.12.8.

T2-Si 280μm

7μm-Al foil

With T2-Si ∆E correction

We have to understand why the ratio between the -odes is energydependence?

T2-Si 280μm

7μm-Al foil

TOF

10μm

Page 18: Beam Test of BGO Crystal Yunlong Zhang 2009.12.8.

Fit function:Fit function:

L = a+b*x

Page 19: Beam Test of BGO Crystal Yunlong Zhang 2009.12.8.

Saturation of fluorescence?

Page 20: Beam Test of BGO Crystal Yunlong Zhang 2009.12.8.

Thanks