Top Banner
Your Most Valuable QA & Dosimetry Tools Product Solutions 2014
28

Your Most Valuable QA & Dosimetry Toolswork.sunnuclear.net/documents/2014-Sun_Nuclear-Product-Solutions... · Your Partner for QA & Dosimetry ... SRS Film QATM ... colleagues in both

Feb 05, 2018

Download

Documents

buimien
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Your Most Valuable QA & Dosimetry Toolswork.sunnuclear.net/documents/2014-Sun_Nuclear-Product-Solutions... · Your Partner for QA & Dosimetry ... SRS Film QATM ... colleagues in both

Your Most Valuable QA & Dosimetry Tools

Product Solutions 2014

Page 2: Your Most Valuable QA & Dosimetry Toolswork.sunnuclear.net/documents/2014-Sun_Nuclear-Product-Solutions... · Your Partner for QA & Dosimetry ... SRS Film QATM ... colleagues in both

2

Page 3: Your Most Valuable QA & Dosimetry Toolswork.sunnuclear.net/documents/2014-Sun_Nuclear-Product-Solutions... · Your Partner for QA & Dosimetry ... SRS Film QATM ... colleagues in both

3

Contents

Better Outcomes Through Technology .........................4

Your Partner for QA & Dosimetry .................................5

The Sun Nuclear Advantage - Different by Design .......6

Total Solution Partner .................................................8

ArcCHECK® ..............................................................12

MultiPlugTM ...............................................................12

CavityPlugTM .............................................................12

3DVH® ......................................................................13

Respiratory MotionSimTM ...........................................13

EPIDoseTM ................................................................14

RadCALC® ................................................................14

MapCHECK® 2 ..........................................................15

MapPHANTM ..............................................................15

MotionSIMTM .............................................................15

IVDTM 2 / rf-IVDTM 2 ...................................................16

ISORADTM .................................................................16

QEDTM ......................................................................16

SRS Film QATM ..........................................................17

EasyCubeTM ..............................................................17

ImageProTM Solution .................................................20

ImageProTM Phantoms ..............................................20

StereoPHANTM ..........................................................21

TomoDoseTM .............................................................21

PROFILERTM 2 ..........................................................22

IC PROFILERTM ........................................................22

SRS PFOFILERTM .....................................................23

IMFTM / GMFTM ..........................................................23

ATLASTM ...................................................................18

Daily QATM 3 ..............................................................18

CHECKMATETM 2 ......................................................19

3D SCANNERTM ........................................................24

3D MiniLiftTM .............................................................25

3D ReservoirTM ..........................................................25

3D TPRTM ..................................................................25

EDGE DetectorTM ......................................................25

1D SCANNERTM ........................................................26

PC ElectrometerTM ....................................................26

Page 4: Your Most Valuable QA & Dosimetry Toolswork.sunnuclear.net/documents/2014-Sun_Nuclear-Product-Solutions... · Your Partner for QA & Dosimetry ... SRS Film QATM ... colleagues in both

4

Better Outcomes Through Technology

Independent and relevant quality assurance is more important

than ever. Choosing a partner with a record of clinically-

proven solutions and a history of outstanding quality and

service is just as important.

Sun Nuclear deals with the challenges confronting every

Radiation Oncology program: safe and effective delivery of

the intended prescriptions and doses to patients. These

challenges have implications including:

► Patient Safety

► Patient Clinical Outcomes

► Facility Continual Improvement

► Facility Employee Satisfaction

► Facility Reputation for Quality

► Facility and Employee Accreditations/Certifications

► Facility Financial Health

In choosing Sun Nuclear, you gain a partner with more than

25 years of proven experience and you become part of a

world-wide network of cancer treatment facilities committed

to improving QA and Dosimetry outcomes.

Resource Depth

► Headquarter facilities and manufacturing are in Melbourne,

Florida, USA

► >60,000 total square feet (5575 square meters)

► 2 linacs with photons, electrons, MLC, and EPID

► 1 Cobalt 60 teletherapy system

► 5 treatment planning and oncology information systems

Page 5: Your Most Valuable QA & Dosimetry Toolswork.sunnuclear.net/documents/2014-Sun_Nuclear-Product-Solutions... · Your Partner for QA & Dosimetry ... SRS Film QATM ... colleagues in both

5

Your Partner for QA & Dosimetry

Worldwide Leader in QA & Dosimetry

Built-In Expertise

► >20 experienced members of customer support team

including clinical physicists

► >30% of Sun Nuclear is dedicated to Research and

Development

► >15 AAPM Task Group memberships including TG-142,

TG-106, and TG-100

Cooperative Partnerships

► >10 industry partnerships including all major OEMs: Varian, Elekta, Accuray®, Philips, ViewRay, Mevion, RaySearch

► #1 worldwide market share leader

► >4,000 cancer treatment facilities worldwide use

Sun Nuclear solutions

► >60 countries use Sun Nuclear solutions

Quality and Support

► ISO 13485 | ISO 9001 certifications

► 9 FDA 510(k) clearances

► 4 continents with full-time Sun Nuclear Support presence:

Europe, Asia, South America, North America

Page 6: Your Most Valuable QA & Dosimetry Toolswork.sunnuclear.net/documents/2014-Sun_Nuclear-Product-Solutions... · Your Partner for QA & Dosimetry ... SRS Film QATM ... colleagues in both

6

The Sun Nuclear Advantage

Quick and Easy Setup

► Patented user calibration

(Wide Field Calibration)

► USB communication with PC

► Single unit, lightweight designs

► Single cable for power and data

► No warm-up or pre-irradiation required

► Intuitive software for short learning curve

Independent

► Checks and balances using diodes, ion

chambers, EPIDs, and log files

► Independent from delivery and

planning systems

► Measurement, and measurement-

based calculations

► No 100% reliance on calculation-only

patient QA

Data Integrity and Transparency

► Original raw and processed data available

to user

► Data export to spreadsheets

► Independently analyze any

measurement data

► All algorithms documented in online help

The Right Detector for the Right Application

► Sun Nuclear manufactures both diode and ion chamber

detectors and arrays.

► Sun Nuclear’s SunPoint Diode Detectors offer the benefits

of smaller active area, better sensitivity, less drift and

increased lifetime.

► SunPoint Diode Detectors are used for small field/patient-

specific applications due to smaller size and better

performance characteristics for small and modulated fields.

► Ion chambers are used for large field/low-modulation

applications where small size is not necessary.

► Sun Nuclear software also supports EPID, Gafchromic film,

and CR detectors.

Many companies claim to develop innovative products,

but their products are not better or different from any

other product. At Sun Nuclear, innovation has a name. Our

solutions are designed for you and the way you work, not

the other way around. The result? Solutions that look and

perform differently in a way that improves outcomes while

saving you time.

Page 7: Your Most Valuable QA & Dosimetry Toolswork.sunnuclear.net/documents/2014-Sun_Nuclear-Product-Solutions... · Your Partner for QA & Dosimetry ... SRS Film QATM ... colleagues in both

7

ImageProTM – Solution for TG-142

► No need to manually name files, or to identify the machine or tolerance. ImagePro’s advanced

algorithms do the work for you.

► ImagePro is delivered pre-installed on Sun Nuclear’s Integrated Data Device (IDD). Simply

connect the IDD to your network and ImagePro is ready to be used on as many computers as

you wish.

► The ImagePro user interface (or client) does not need to be installed because it relies on a

standard browser. It offers security, flexibility and less overhead for you.

ArcCHECK®

► The ArcCHECK has a patented cylindrical detector geometry that measures beams upon entry

and again upon exit. The detector geometry is consistent for all gantry angles, is highly sensitive

to errors, and enables dose calculation throughout the entire phantom.

► The ArcCHECK has a Virtual Inclinometer to ensure measurement at the correct angle. There are

no additional cables or equipment to attach to the linear accelerator.

► The ArcCHECK geometry allows for a hollow central cavity which can be filled with inserts

holding ion chambers, diodes, and film, and can be arranged in many different configurations.

► Reversing the trend towards ever heavier and awkward-to-use phantoms, ArcCHECK weighs only

16kg and uses only one cable for all power and data needs.

3D SCANNERTM

► The 3D SCANNER uses a patented design that allows consistent in-line, cross-line, and diagonal

scans with no detector setup changes or tank shifts. The system improves data quality and

reduces the risk of error.

► AutoSetupTM automatically and reproducibly sets the detector position at the water surface, and

levels and aligns the tank, eliminating subjectivity.

► Save all scans in a searchable database, and preserve original raw scan data after applying

processing layers to provide more data flexibility and easier inter-comparisons.

Industry firsts... > First Daily QA trending database c. 1996, Daily QA Check

> First Electronic 2D Array for patient specific IMRT QA c. 2002, MapCHECK

> First Patient-specific respiratory and gating simulation system c. 2005, MotionSim

> First EPID-based dose analysis for patient-specific IMRT QA c. 2007, EPIDose

> First Dose on patient geometry without a TPS dose calculation c. 2010, 3DVH

Different by Design

Page 8: Your Most Valuable QA & Dosimetry Toolswork.sunnuclear.net/documents/2014-Sun_Nuclear-Product-Solutions... · Your Partner for QA & Dosimetry ... SRS Film QATM ... colleagues in both

8

Total Solution Partner

TrainingModern radiation oncology techniques are complex, and their safety and effectiveness rely on the proper use of QA tools. Many

groups, including ASTRO and AAPM, recommend that meaningful training solutions be incorporated into routine practices.

Sun Nuclear opened a new customer training center, the

Sun Nuclear Training Center, in 2013. The training center is

a purpose-built training center with its own linear accelerator

and demonstrates Sun Nuclear’s commitment to patient

safety and the effective use of QA tools.

Users can experience in-depth exploration of Sun Nuclear

product functionality on a linac, as well as application-

specific coursework such as IMRT QA, VMAT QA, Annual

QA, and Commissioning. The SNTC offers this training in

an environment that closely mimics an actual clinical setting

because a realistic experience ensures the best possible

training outcome.

Training courses on a variety of subjects are offered year

round and are delivered by Sun Nuclear subject matter

experts and experienced clinical users. Details on upcoming

training courses are posted on the Sun Nuclear

Support website.

Page 9: Your Most Valuable QA & Dosimetry Toolswork.sunnuclear.net/documents/2014-Sun_Nuclear-Product-Solutions... · Your Partner for QA & Dosimetry ... SRS Film QATM ... colleagues in both

9

On-site/Online Product Training

All Sun Nuclear products are designed for easy installation and setup. As an added service, Sun Nuclear offers professional

product installation and training for all products. Customers have two options to choose from:

Option 1: On-site Installation and Training

Option 2: Online Product Training

On-site/Online Training Comparison On-site Online

Trained Sun Nuclear professional on site Y -

Comprehensive syllabus Y Y

Provides users with in-depth and efficient product use experience Y Y

Training on advanced features Y Y

Training on real-time operation of the software Y Y

Secure Internet connection with unlimited customer attendees Y Y

Oversee and physically assist with installation of devices and software Y -

Remote assistance with installation of software Y Y

Assist with acquisition of sample measurements Y -

Train users on comprehensive product functionality Y Y

Question and answer (Q&A) session Y Y

Maximize benefits of the product(s) purchased Y Y

Shorten learning curve and begin using product(s) sooner Y Y

Page 10: Your Most Valuable QA & Dosimetry Toolswork.sunnuclear.net/documents/2014-Sun_Nuclear-Product-Solutions... · Your Partner for QA & Dosimetry ... SRS Film QATM ... colleagues in both

10

SupportAt Sun Nuclear we know the value of good support, and are committed to providing customers with the best possible support

experience. Sun Nuclear’s Support team includes over twenty professionals with diverse physics, teaching, engineering,

customer support, and linguistics backgrounds.

As always, Sun Nuclear customers receive complimentary standard technical support, and there is never a limit to the number

of times you may contact us with a question.

Telephone

► Worldwide: Standard Sun Nuclear telephone support is

offered from 8am to 8pm ET, Monday through Friday.

► Regional: Standard extended regional telephone support

hours are available.

► After hours/on-call telephone support options

are available.

Secure Remote Support

► Sun Nuclear users who have a current maintenance

contract may request live online assistance.

► Online assistance allows trained Sun Nuclear support

professionals to collaborate directly in the user’s PC

environment to evaluate and resolve issues, relieving the

user of this task.

Sun Nuclear Support Site

► All Sun Nuclear users have access to an exclusive,

user-only Sun Nuclear Support Website which provides

the following benefits:

> Access to software, manuals, and software license

keys in real time for your products.

> Exclusive instructional material including videos

and FAQs.

> Sun Nuclear account management (inventory of

products and product details).

> Ability to request information and support.

Visit www.sunnuclear.com/Support

Total Solution Partner

I have to say the 3D SCANNER is a step above anything I’ve used in the past. Even with my lack of hands

on experience I was able to meet all of my deadlines and found that the system was exceptionally easy to

pickup and use right out of the box. Sun Nuclear’s trainer was also exceptional and by the end of the first

afternoon I was navigating the software, processing scans and even exporting scans for outside analysis.

The tank setup is by far and away the easiest setup of a scanning tank I have ever used and a testament

to the ground up approach to the engineering of this scanning system. From my experience with 3D

SCANNER, I’ll be a Sun Nuclear tank user for all of my future sites.

Mike Tallhamer, M.Sc St. Anthony’s Hospital

Denver, USA

Page 11: Your Most Valuable QA & Dosimetry Toolswork.sunnuclear.net/documents/2014-Sun_Nuclear-Product-Solutions... · Your Partner for QA & Dosimetry ... SRS Film QATM ... colleagues in both

11

Education

Exploring the Future of QA and Dosimetry

QA & Dosimetry Symposium (QADS) is the only event for

Medical Physicists and Dosimetrists that brings leading

clinical innovators together to discuss the latest practices and

issues related to QA and Dosimetry in Radiation Oncology.

The symposium provides attendees with practical guidance

from industry experts on new and relevant products,

processes, and techniques in today’s evolving Radiation

Oncology environment. The talk schedule is supplemented

by opportunities to interact with leading industry vendors and

colleagues in both professional and social settings.

Benefits

► Continuing education credits (CAMPEP & MDCB)

► Presentations from the industry’s leading authorities on

the latest practices and future direction of Radiation

Oncology QA & Dosimetry

► Major vendor participation

► Leading resort venues

► Free registration for students and residents

► Web access to presentations after the symposium

► Networking with peers

Register at www.qasymposium.com

Weekly Product Webinars

Sun Nuclear hosts weekly educational Product Webinars as

part of its educational program. The webinars are hosted

online, and are presented by clinical users and product

experts. All webinars last approximately 30 minutes, with a

Q&A session following the webinar. Visit us online to register

for our upcoming webinars. Webinar recordings are also

available upon request.

Register at www.sunnuclear.com

QA & DosimetrySYMPOSIUMExploring the future of QA and Dosimetry

Ongoing education is an important component of a

successful continual improvement program, which ultimately

helps to improve outcomes and patient safety. Sun Nuclear

is a leader in the sharing of information and ideas from both

industry and clinical perspectives. Our educational activities

focus on both product and task specific subjects as well as

broader issues in the radiation oncology industry.

Page 12: Your Most Valuable QA & Dosimetry Toolswork.sunnuclear.net/documents/2014-Sun_Nuclear-Product-Solutions... · Your Partner for QA & Dosimetry ... SRS Film QATM ... colleagues in both

12

Applications Patient plan QA

- RapidArc®/VMAT, FFF, SmartArc and TomoTherapy®

- Routine IMRT and 3D conformal Machine and setup QA

Highlights The world’s first true 4D detector array The world’s first cylindrical detector array The world’s first electronic array designed specifically for

rotational dosimetry

Key Benefits Utilizes SunPoint® Diode Detectors

Consistent BEV regardless of gantry angle High detector density in BEV 221 detectors in central 10x10cm Easy setup and lightweight (16kg) Measure both composite and per control point Real-time updates (50ms)

Detector type: SunPoint Diode Detectors

Detector quantity: 1386

Detector spacing: 1.0cm

Array diameter: 21.0cm

Inherent buildup: 3.3 +/- 0.1g/cm2

Inherent backscatter: 3.3 +/- 0.1g/cm2

Array geometry: Helical Grid, 1cm offset

Phantom Material: PMMA (Acrylic)

Weight: 16kg

Cavity Options

Applications IMRT, Arc, SRS, SBRT dose distributions in cavity Supports 3DVH

Highlights Precision fitted for ArcCHECK central cavity Accepts ion chambers, diodes, film detectors Position single detectors in any of 25 unique locations Position film sizes up to 13 x 21cm in X or Y axis Angle indicator allows rotation to any angle within

ArcCHECK cavity

Applications Measure dose at isocenter

Supports 3DVH

ArcCHECK®

MultiPlug™

CavityPlug™

Page 13: Your Most Valuable QA & Dosimetry Toolswork.sunnuclear.net/documents/2014-Sun_Nuclear-Product-Solutions... · Your Partner for QA & Dosimetry ... SRS Film QATM ... colleagues in both

13

Applications DVH QA - Estimate impact on patient dose and DVH - Works with ArcCHECK, MapCHECK, EPIDose, and film

Highlights Dose to phantom planar QA is converted to a heterogeneous dose to patient for DVH QA Use with arc delivery or IMRT Overlay dose on patient CT Does not require a secondary dose algorithm Virtual Gel: High density, high resolution dose reconstruction in ArcCHECK volume Quick Stat Templates: Flexible, easy plans for data analysis

Key Benefits Clinically relevant patient QA Eliminates dose to phantom guesswork No commissioning required

Supports coplanar and non-coplanar beams: Yes

Estimation of absolute patient dose: Yes

Analysis by Patient DVH: Yes

Accounts for TPS error: Yes

Accounts for beam delivery error: Yes

Statistics per anatomical structure: Yes

DICOM RT compatibility: Full integration

VMAT and RapidArc® support: Yes

Analysis of beam by beam errors: Yes

Applications Respiratory/breathing dosimetric effects simulation Prostate motion simulation (including non-cyclical) Gating dosimetric effects simulation

Highlights Patient-specific respiratory patterns imported from 4DCT Interplay effect simulation - averages possible interplay effects over multiple fractions Track individual voxels as a function of time Movie mode features 4D playback of dose accumulation, motion, and time Track individual voxels as a function of time Customize moving target, margin, direction of motion, one-time/repeating motion, etc.

Key Benefits Allows quantification of patient-specific motion effects on 3D dose distribution in patient anatomy No additional setup or phantom required Motion effects are fully patient specific Dose distributions in patient anatomy and geometry Up to 5 dose reconstructions per second No commissioning

Supports coplanar and non-coplanar beams: Yes

Estimation of absolute patient dose: Yes

Analysis by Patient DVH: Yes

Accounts for TPS error: Yes

Accounts for beam delivery error: Yes

Statistics per anatomical structure: Yes

DICOM RT compatibility: Full integration

VMAT and RapidArc® support: Yes

Analysis of beam by beam errors: Yes

3DVH® 3DVH Options

Respiratory MotionSim™

Page 14: Your Most Valuable QA & Dosimetry Toolswork.sunnuclear.net/documents/2014-Sun_Nuclear-Product-Solutions... · Your Partner for QA & Dosimetry ... SRS Film QATM ... colleagues in both

14

Applications EPID based patient plan QA

- MV EPID images to absolute dose - Comparison, subtraction, overlay, profiles, more - Rotational delivery support Flatness and symmetry

Highlights Independent of TPS and linac - Dose QA audits the TPS and linac - Catch errors image QA would miss - Independent algorithms and software

Tests output of TPS dose algorithms

Key Benefits Easy to use - No buildup on EPID required - Combine split fields into one - Maximize the benefits of MV EPID

Supports any EPID distance and dose plane setup Use SNC PatientTM software for complete analysis

Independent dose comparisons: Yes

Increased data density for QA: Yes

Audit the TPS and delivery system: Yes

No buildup on EPID: Yes

Measure at any gantry angle: Yes

One model works with each:

Linac/EPIDPhoton energySource to EPIDSource to Dose planeDose plane depth

Applications Performs conventional 2D and 3D, IMRT and compensator

based IMRT treatment plans Robust VMAT support using Regions of Interest (ROI) BrachyTherapy support for HDR, HDR (Xoft), LDR, and

Permanent Implant treatment plans

Highlights Intuitive and customizable user interface Flexible installation architecture - local or server based Flexible site licensing Capable of importing from RTP or R&V Systems and

Exporting to R&V or RTP Systems

Key Benefits Fast, easy, and accurate dosimetric verification Rapid commissioning (configuration of up to two (2)

machines is included with every package) Professional calculation printouts and powerful

administrative and physics data reporting tools Increases throughput, relieves workload pressure, and

reduces costs department wide

RTP import and export: Yes

Verify and record export: Yes

Independent IMRT verification: Yes

Electron and photon calculations: Yes

Rapid commissioning: Yes

VMAT support with ROI: Yes

PC requirements:

Windows NT4, 2000, XP , Vista, and 7. 133MHz processor or faster128MB RAM

EPIDose™ RadCALC®

Page 15: Your Most Valuable QA & Dosimetry Toolswork.sunnuclear.net/documents/2014-Sun_Nuclear-Product-Solutions... · Your Partner for QA & Dosimetry ... SRS Film QATM ... colleagues in both

15

Applications Patient plan QA

- IMRT QA - RapidArc®/VMAT, FFF and TomoTherapy®

(with accessories) Machine, MLC, and setup QA

Highlights Most detectors of any 2D array Highest detector density of any 2D array Largest field size of any 2D array

MapCHECK 2 is a turn-key solution

Key Benefits Proven solution for film-less rotational and IMRT delivery Utilizes SunPoint Diode Detectors Easy to use and efficient software Single click comparisons Rotational capable when used with MapPHAN option Real-time updates (50ms) Interfaces with the IMF™

Detector type: SunPoint Diode Detectors

Detector quantity: 1527

Detector spacing: 7.07mm

Array size: 32.0 x 26.0cm

Inherent buildup: 2.0 ± 0.1g/cm2

Inherent backscatter: 2.7 ± 0.1g/cm2

Array geometry:• Detector spacing parallel

to X and Y axes: 1.0cm• Row spacing: 0.5cm

Phantom Material: PMMA (Acrylic)

Weight: 7.1kg

Applications Patient plan QA for rotational delivery

- RapidArc®/VMAT, FFF and TomoTherapy®

- Coronal and sagittal orientation

Construction: Virtual WaterTM

Available depths: 5.0cm, 10.0cm

Area: 35.0 x 38.0cm2

Weight: 8.0kg (5cm), 21.0kg (10cm)

Applications Dose QA for moving targets Assess interplay effects in MLC-based IMRT QA Compare gated versus non-gated delivery Gating system commissioning and QA

Platform: Carbon Fiber

Maximum travel: X-Axis: 4in (10.2cm); Y-Axis: 4in (10.2cm)

Motors: 2 phase bipolar steppers

Max velocity: 2in/second (5.08 cm/second)

Resolution: 0.13μm

Repeatability: 0.13μm bidirectional

Communication: RS-232, 19,200kbps

Power: 100-240V, single-phase, 47-63Hz, 4A

Weight: 15kg

MapCHECK® 2 MapPHAN™

MotionSIM™

Page 16: Your Most Valuable QA & Dosimetry Toolswork.sunnuclear.net/documents/2014-Sun_Nuclear-Product-Solutions... · Your Partner for QA & Dosimetry ... SRS Film QATM ... colleagues in both

16

Applications Dose verification during treatment Final check of complete system Total Body Irradiation (TBI)

Highlights Real-time wireless option available (rf-IVD 2) Use with control module or PC software Expandable to 52 channels, 4 channels standard Record and verify interface

Key Benefits Simple to install and easy to use Automatic patient temperature compensation 12 calibration keys Patient database Correction factor tools Accuracy and reliability proven worldwide Use with industry standard re-usable detector options - QED - Flat for easy patient placement - ISORAD - Cylindrical for isotropic response

Channels:Standard: 4 (IVD 2) / 8 (rf-IVD2)Maximum: 52

Temperature compensation:

Automatic-within 0.10C (Negative polarity only)

Repeatability: ± 0.2% or ± 0.1cGy

Polarity:Bipolar (negative or positive polarity detectors)

Leakage: Automatic compensation

Warm-up time: < 30 sec

Power:rf-IVD 2 : Rechargeable NiMH battery (12 hr)IVD 2 : Power supply

Weight: 0.34kg

Highlights Flat design for easy placement on patient Accurate and stable SunPoint Diode Detector

Key Benefits Real-time dose reading

Highlights Cylindrical design for isotropic response Accurate and stable SunPoint Diode Detector

Key Benefits Real-time dose reading

15-2

5 MV

1-4 M

V

6-12

MV

1-4MV

6-12

MV

15-25

MV

ELECTR

ON

SKIN

Detector area: 1.54mm2

Active dimension: 1.4mm diameter

Dose rate dependence: ± 1%, 75 - 250cm SSD

Detector sensitivity: 27nC/Gy

Detector stability: 0.5%/kGy at 6MV

Detector area: 0.64mm2

Active dimension: 0.8 x 0.8mm

Dose rate dependence: ± 1%, 75 - 250cm SSD

Detector sensitivity: 32nC/Gy

Detector stability: 0.5%/kGy at 6MV

IVD™ 2 / rf-IVD™ 2 IsoRad™ Detector

QED™ Detector

Page 17: Your Most Valuable QA & Dosimetry Toolswork.sunnuclear.net/documents/2014-Sun_Nuclear-Product-Solutions... · Your Partner for QA & Dosimetry ... SRS Film QATM ... colleagues in both

17

Applications Comprehensive SRS Patient Plan QA and analysis

Convert scanned film image data to dose on any stereotactic modality

Compare film dose information to an imported patient plan

Highlights Co-developed with Accuray® Analyze from within the SNC Patient™ software

Analyze film as if it were MapCHECK measured data Compatible with all SRS modalities

Key Benefits Extract a 3D dose plane at any arbitrary angle and save

Extract from a CT data set or dose volume Use as a reference for film alignment and analysis Vidar and EPSON flatbed scanner support Scan film and save as RAW or TIFF images

Scanner/film requirements:

Epson® flatbed scanner for EBT® film

Compatible with:DICOM RT, IMRT, VMAT and Stereotactic

Regional settings: All Windows supported regions

Operating system:Windows XP or Windows 7 (32 or 64 bit)

Processor: Intel Dual Core 2.4GHz or higher

RAM: 4GB RAM

Drive space: 5GB

Video card:Open GL, 32MB RAM, 1280x1024 resolution

Applications Dose measurements - Ionization chambers, TLDs and Film Simulation of heterogeneities - Bones/adipose/muscle/lung/titanium CT scanner QA - Calibration of Hounsfield scale Stereotactic QA

Highlights RW3 construction - 15 slabs Options available - Body Phantom Package - provides head and neck phantoms and body shaped scattering blocks that can hold up to 15 films - Cylindrical Phantom Package - provides four cylindrical extensions to hold dosimeters and heterogeneities - Oval Phantom Package - provides two oval extensions to hold dosimeters and heterogeneities

Density (g/cm3): 1.045

EasyCube dimensions (cm): 18.0 x 18.0 x 18.0

EasyCube internal dimensions (cm): 16.0 x 16.0 x 16.0

Wall thickness (cm): 1.0

Film distance plate quantity and thickness (mm):

(16) 10.0(1) 5.0(2) 2.0(1) 1.0

Film size (cm): 16.0 x 16.0

Number of films: 16

SRS Film QA™ EasyCube™

Page 18: Your Most Valuable QA & Dosimetry Toolswork.sunnuclear.net/documents/2014-Sun_Nuclear-Product-Solutions... · Your Partner for QA & Dosimetry ... SRS Film QATM ... colleagues in both

18

Applications Radiotherapy center QA management Scheduling, equipment, and messaging management

Highlights A front end for data acquisition A back end for reporting and analysis Suitable for both small and large facilities Upload documents to the Attachment Center Audit Log shows overdue/failed test - Searches Attachment Center for documents that have expired or are about to expire Tests can be cloned either as a group or individually to a

new machine

Key Benefits Integrated test libraries Customizable tests Resource tracking management Facility and user management Attachments TG-142 and TG-51 libraries

Organization structure:Daily, monthly, annual, dosimetry, mechanical, safety, and imaging

Test libraries:TG-40, TG-142, TG-51Manufacturer recommended

Customize tests: Yes

Resource tracking: Yes

User management: Yes

Communication center: Yes

Data Management: Yes

Networkable: Yes

Applications Daily linac monitoring and trending

- Output, flatness, symmetry, energy, light-radiation field coincidence

Rotational delivery supported

Highlights Utilizes both ion chambers and diodes for optimal results Measure output, flatness, symmetry, field size, energy in a

single exposure Trend analysis database software included No warm-up or pre-irradiation required Real-time wireless available (rf-Daily QA 3)

Key Benefits No buildup or flipping required Automatic temperature and pressure correction Real-time measurements Simple two-step operation

– Start and then Record Interfaces with the IMF™

Detector type:Sun Point® Diode DetectorsVented Ion Chambers

Detector quantity:Diodes: 12 totalChambers: 13 total

Detector spacing: Diodes: 5.0

Chamber active area: Electron: 0.6cm3; Photon: 0.3cm3

Inherent buildup: Chambers: 1.0 ± 0.1g/cm2

Inherent backscatter: 2.3g/cm2

Field Size: 20.0 x 20.0cm

Weight: 5.7kg

ATLAS™ Daily QA™ 3

Page 19: Your Most Valuable QA & Dosimetry Toolswork.sunnuclear.net/documents/2014-Sun_Nuclear-Product-Solutions... · Your Partner for QA & Dosimetry ... SRS Film QATM ... colleagues in both

19

Applications Daily central axis output Repeatability Monitor chamber linearity

Highlights Built-in rechargeable batteries Precision alignment template User calibration - Stores up to 15 calibration values - Automatic memory position advance

Key Benefits Cost effective solution Simple to use Automatic temperature and pressure correction Large, easy to read display

Detector type: Vented Ion Chamber

Detector quantity: 1 - CAX

Detector spacing: 5.0mm

Field sizes: 10 x 10 and 20.0 x 20.0cm

Detector volume: 0.6cm3

Inherent buildup: 1.0 ± 0.1g/cm2

Inherent backscatter: Acrylic, 3.6cm

Weight: 4.9kg

CHECKMATE™ 2

Page 20: Your Most Valuable QA & Dosimetry Toolswork.sunnuclear.net/documents/2014-Sun_Nuclear-Product-Solutions... · Your Partner for QA & Dosimetry ... SRS Film QATM ... colleagues in both

20

Applications Cone Beam CT QA MV and kV QA Winston Lutz using film and EPID MLC log file analysis Radiation-isocenter Congruence Light Field Congruence Starshot Beam Flatness and Symmetry

Highlights Comprehensive and focused TG-142 QA Tests output of TPS dose algorithms Original software built from the ground up Intuitive, Secure User Interface Includes Integrated Data Device (IDD) with pre-installed ImagePro software, database and web server.

Key Benefits Transparent and documented algorithms

No software installation No equipment requirement One-click batch reports Future trend analysis

MV-QA Easy setup, alignment clearly marked on overlay and image Automatic QA: image scaling, positioning offset, spatial resolution, contrast, uniformity and noise

ImagePro™

Software

SOLUTION FOR TG-142

Phantoms

MV Line pairs/mm: 0.1, 0.2, 0.4, 0.7

MV ROI: 9 (4 spatial, 4 contrast, 1 center)

MV Dimensions (cm): 12.7 x 10.2 x 1.9

kV-QA Easy setup, alignment clearly marked on overlay and image Automatic QA: image scaling, positioning offset, spatial resolution, contrast, uniformity and noise

kV Line pairs/mm: 0.6, 1.2, 1.8, 2.4

kV ROI: 28 (4 spatial, 23 contrast, 1 center)

kV Dimensions (cm): 12.7 x 10.8 x 1.6

Page 21: Your Most Valuable QA & Dosimetry Toolswork.sunnuclear.net/documents/2014-Sun_Nuclear-Product-Solutions... · Your Partner for QA & Dosimetry ... SRS Film QATM ... colleagues in both

21

FS-QA: Field Size QA and Constancy Symmetric and asymmetric light field/radiation field coincidence and jaw positioning Flatness and symmetry for photon and electron beam profile constancy

WL-QA: Isocenter QA and Constancy Winston-Lutz measurements Imaging and radiation field isocenter coincidence Cone-beam CT positioning/repositioning End-to-end IGRT positioning tests 3D isocenter offset results

StereoPHAN™

Dimensions (cm): 6.0 x 6.0 x 6.0

Sphere size (mm): 7.0

Sphere center accuracy (mm): =/-0.2

Field sizes (cm): 10 x 10; 15 x 15

Markers (±0.1mm): 56 - Field size (7 per field edge)

Dimensions (cm): 17.8L x 17.8W x 0.6D

Applications Stereotactic (SRS/SRT/SBRT) end-to-end testing Quality assurance of image fusion algorithms for CT and MRI imaging modalities Absolute, relative and point dose dosimetry QA

measurements at isocenter with ion chambers; relative dose distribution using film

Highlights Design allows for interference-free dose measurements

Target volumes in CT/MRI cube eliminate need for CT/MRI markers Ion chamber cube geometry enables easy ion chamber cross-calibration to water Support for stereotactic frame and frameless systems

Key Benefits No tools required for assembly

Phantom stand holds the inserts, making them easily accessible during testing

Multi-functional cube inserts increase efficiency All components fit into a durable rolling case suitable for

storage and air travel Stand base can be mounted to a couch that uses the Lok-Bar™ system

Material: PMMA (Acrylic)

Weight (including all inserts):

11.1kg (24.6 lbs)

Measurement cubes: 85mm x 85mm x 85mm

StereoPHAN dimensions:

518mm x 276mm x 323mm

Page 22: Your Most Valuable QA & Dosimetry Toolswork.sunnuclear.net/documents/2014-Sun_Nuclear-Product-Solutions... · Your Partner for QA & Dosimetry ... SRS Film QATM ... colleagues in both

22

Applications Use for Machine QA on TomoTherapy® Hi-Art®

Post component replacement verification Daily QA tests

Highlights Only array designed for TomoTherapy Hi-Art Measure entire Hi-Art beam in a single measurement Accessibility to raw data Import water tank data and compare to TomoDose files

Key Benefits Fast and efficient setup Ten measurement axes for both X and Y direction

Measure output, symmetry, flatness, field size, penumbra, beam center

Analyze and compare up to 15 profiles at once Instantaneous dose rate plot

Detector type: SunPoint Diode Detectors

Detector quantity: 223 total on X and Y

Detector spacing: X: 5.0, Y: 4.0, Y Off-axis: 8.0mm

Array size: 53.0 x 9.8cm

Array geometry: X: 53.0 Y: 9.8cm (8.0cm at ±19.0)

Inherent buildup: 1.0g/cm2

Inherent backscatter: 2.8g/cm2

Phantom Material: PMMA (Acrylic)

Weight: 5.0kg

TomoDose™

Applications Alternative to a water phantom for many tests Flatness, symmetry, field size, beam center, penumbra width, light-radiation field coincidence

Real-time beam constancy, steering, diagnostics Evaluate collimator and rotational sag Profile comparison, subtraction, and overlay Identify startup/time dependent anomalies

Highlights Quick and precise beam QA measurement

Sets up in minutes versus hours for a typical water tank Multiple real-time parameter calculations No warm-up or pre-irradiation required Import water tank data and compare to PROFILER files

Key Benefits Real-time beam measurements Fast and efficient setup One measurement checks multiple parameters

Interfaces with the IMF™

Detector type: SunPoint® Diode Detectors

Detector quantity: 139 Total: X Axis: 57 / Y Axis: 83

Detector spacing: 4.0mm

Array size: 22.4 x 32.8cm

Inherent buildup: 1.0g/cm2

Inherent backscatter: 2.0g/cm2

Phantom Material: Virtual Water™ top plate

Weight: 5.0kg

PROFILER™ 2

Page 23: Your Most Valuable QA & Dosimetry Toolswork.sunnuclear.net/documents/2014-Sun_Nuclear-Product-Solutions... · Your Partner for QA & Dosimetry ... SRS Film QATM ... colleagues in both

23

Applications All PROFILER 2 applications plus: Linac manufacturing and field service

Annual QA Comprehensive linac QA and benchmarking against linac manufacturer measurements

Highlights Accuracy within 0.5% of a water tank Unique curved ion chamber design maximizes signal

and minimizes volume averaging Very low noise to signal ratio (0.15%)

Key Benefits 45cm length diagonals

40 x 40cm field measurement at 75cm SDD Interfaces with the GMF™

Detector type: Parallel plate Ion Chamber

Detector quantity:251 Total: X Axis: 63 / Y Axis: 65-Diagonal: 63 / +Diagonal: 63

Detector spacing: 5.0mm

Array size: 32.0 x 32.0cm

Detector volume: 0.046cm3

Detector sensitivity: 1.4nC/Gy

Inherent buildup: 0.9g/cm2

Inherent backscatter: 2.3g/cm2

Phantom Material: PMMA (Acrylic)

Weight: 7.0kg

IC PROFILER™

Applications SRS cone based radiation therapy

- On Accuray® Recommended QA Equipment list Daily, monthly, and annual machine QA

including commissioning

Highlights Only array designed for cone based SRS Accepted and proven for clinical and factory settings Import water tank data and compare to

PROFILER files Real-time analysis

Key Benefits Fast and efficient setup Identify startup/time dependent anomalies

One measurement checks multiple parameters Attaches to the CyberKnife® birdcage

Detector type: SunPoint Diode Detectors

Detector quantity: 125 total on four axes

Detector spacing: 4mm with radial offset

Radial field size: 10, 20, 40, 60, 80, 100, and 120mm

Axes length: 143 (X), 127 (Y), 124mm (Diagonals)

Inherent buildup: 0.5g/cm2

Phantom Material: PMMA (Acrylic)

Weight: 1.5kg

SRS PROFILER™

Page 24: Your Most Valuable QA & Dosimetry Toolswork.sunnuclear.net/documents/2014-Sun_Nuclear-Product-Solutions... · Your Partner for QA & Dosimetry ... SRS Film QATM ... colleagues in both

24

Applications Measure output, symmetry, flatness, field size, penumbra, beam center, and more at any angle. Determine gravity-induced dose constancy, symmetry,

flatness, MLC and jaw shift changes

Highlights Hold device measurement planes at machine isocenter:

MapCHECK, MapCHECK 2, PROFILER 2, Daily QA3, rf-Daily QA3, and SRS PROFILER with adapter plate

Key Benefit Fast, efficient, reproducible setup Measure at 100cm SDD +/- 1.0mm Supports electron cone (GMF)

Construction: Aluminum Alloy

Buildup clamps: Adjustable

Compatible linacs: (IMF) Elekta, Siemens, Varian

Weight:(IMF-E) 7.9kg(IMF-S) 8.1kg(IMF-V) 7.4kg

Maximum buildup: (IMF) 5cm

Dimensions:(IMF-E) 39.4cm x 28.7cm x 48.8cm(IMF-S) 39.4cm x 31.5cm x 48.0cm(IMF-V) 39.4cm x 31.5cm x 39.1cm

IMF™ GMF™

Construction: Aluminum Alloy

Buildup clamps: Adjustable

Compatible linacs: (GMF) Elekta, Varian

Weight:(GMF-E) 16.8kg(GMF-V) 3.7kg

Maximum buildup:(GMF-E, ICP only) 3cm(GMF-V, ICP only) 1.5cm

Dimensions:(GMF-E) 56.5cm x 56.5cm x 56.0cm(GMF-V) 40.0cm x 47.1cm x 42.2cm

Applications Measure output, symmetry, flatness, field size, penumbra, beam center, and more at any angle. Determine gravity-induced dose constancy, symmetry,

flatness, MLC and jaw shift changes

Highlights Hold device measurement planes at machine isocenter:

MapCHECK 2, PROFILER 2 and Daily QA 3 (adapter plate required)

Key Benefit Fast, efficient, reproducible setup Measure at 100cm SDD +/- 1.0mm Supports electron cone (GMF)

Page 25: Your Most Valuable QA & Dosimetry Toolswork.sunnuclear.net/documents/2014-Sun_Nuclear-Product-Solutions... · Your Partner for QA & Dosimetry ... SRS Film QATM ... colleagues in both

25

Applications Detector positioning in water

Output factors and dose calibrations Annual and routine QA

Highlights Suitable for TG-51 Automatic water level detection

- Places the detector at the water surface Pendant and PC software control 30cm scan depth Advanced scanning software option available

Key Benefits Fast and efficient detector positioning Accommodates all common electron cones

Multi-function pendant Horizontal ruler for off-axis detector positioning Farmer type, universal, and reference detector

holders included Leveling platform available

Inner dimensions: 35.0 x 39.0 x 36.2cm

Inner volume: 48 liters at 34.5cm

Arm positioning increments: 0.01, 0.1, 1.0 & 10.0cm

Arm positioning accuracy: ± 0.01cm

Scan depth maximum: 30.0cm

Tank material: PMMA (Acrylic)

Weight empty with arm: 10.9kg

Charge: 2pC – 10mC, 15fC resolution

Rate:Continuous: 2pA – 50nA 0-105 PC/Pulse

Leakage: < 10fA

A/D Converter: 16 bit

Non-linearity: +/- 0.1%

Long term stability: <+/- 0.5%

Repeatability: <+/- 0.25%

Connector: TNC or BNC

Weight: 0.4kg

1D SCANNER™ PC Electrometer™

Applications Reference class dosimetry for absolute dose calibration

Mobility intensive assignments

Highlights World’s smallest reference class electrometer USB powered — no batteries or power cord

Quick to setup and easy to use Fully configurable software interface

Key Benefits Pocket size, lightweight design Two independent measurement channels

Interfaces with the 1D SCANNER Less than 1-minute warm-up time Temperature probe input Single USB cable connection Fast sampling interval (100-1,000 μs) Detector libraryConstruction: Aluminum Alloy

Buildup clamps: Adjustable

Compatible linacs: (GMF) Elekta, Varian

Weight:(GMF-E) 16.8kg(GMF-V) 3.7kg

Maximum buildup:(GMF-E, ICP only) 3cm(GMF-V, ICP only) 1.5cm

Dimensions:(GMF-E) 56.5cm x 56.5cm x 56.0cm(GMF-V) 40.0cm x 47.1cm x 42.2cm

Page 26: Your Most Valuable QA & Dosimetry Toolswork.sunnuclear.net/documents/2014-Sun_Nuclear-Product-Solutions... · Your Partner for QA & Dosimetry ... SRS Film QATM ... colleagues in both

26

Applications TPS commissioning Linac acceptance Annual QA

Highlights Patented cylindrical design scans on any axis Ring drive positions detector in any relevant 3D location Automated leveling and beam center positioning Built-in control box/electrometer Works with both diodes and ion chambers

Key Benefits 0.1mm positioning accuracy

Reproducible setup Less than 20 minute setup 65cm scan range, no need for tank shift

Vertical (mm): 400

Diameter (mm): 650

Ring (degrees): 330

Motors: Encoded stepper/servo

Scanning modes: Continuous and step

Scanning speed range (mm/sec): Variable up to 8

Scanning accuracy (mm): 0.1

Water Tank:

Diameter inner, Height / Width (mm): 676 / 673 / 875

Water capacity (L): 166

Weight empty / full (kg): 44 / 210

The Cylindrical Advantage

Maintains ideal orientation of the detector during in-line, cross-line and diagonal scans Amount of cable being irradiated remains constant during scan, diminishing cable effect More scanning area— up to 65cm in any direction Less volume than traditional tanks with smaller scanning area Less volume means faster fill and drain time No need to shift the tank and risk setup inconsistencies during shifting Water ripples settle faster because of curved walls Circular design makes 3D SCANNER a stronger shape than a traditional tank with 8 seams Better design for acquiring numerous and various diagonal scans as with Cyberknife®

Detector travels the same path for multiple measurement angles, allowing for self diagnostics

AutoSetup™

No more guesswork, highly reproducible setup results; 3D SCANNER setup results from users with no experience compared to 3D SCANNER setup results from experts yield maximum variation of only 0.3mm. Automatically levels tank by sensing water level relative to vertical Z-axis. Automatically aligns ring center to beam axis with motor-driven tank shift +/- 3cm Automatically establishes in-line and cross-line home positions for ring drive using a series of beam measurements Ready to scan in less than 20 minutes

3D SCANNER™

3D SCANNER:

Page 27: Your Most Valuable QA & Dosimetry Toolswork.sunnuclear.net/documents/2014-Sun_Nuclear-Product-Solutions... · Your Partner for QA & Dosimetry ... SRS Film QATM ... colleagues in both

27

Highlights

Straddles linac couch ring for stability Fits through standard doorways Legs fold for compact storage Quick and easy disassembly for transport

Highlights

187L (49.5gal) capacity 6 minute fill time 3 minute drain time Easily maneuverable

Applications SRS and IMRT beam modeling TPS commissioning

Highlights Accurate penumbra characterization for all fields Ideal for steep dose gradients and small fields On Accuray® Recommended QA Equipment list 842 times smaller in volume than micro ion chambers

100 times more signal than micro ion chambers

Key Benefits No dose volume averaging Accuracy for critical penumbra measurements Works with all common water phantoms

Detector type: SunPoint® Diode Detector

Diode die location:

0.3mm from top4.3mm from end2.7mm from sideIndicated by a target on top of housing

Water depth equivalent: 0.5mm

Housing wall thickness: 0.13mm, brass

External dimensions: 3.8 x 5.5 x 38mm

Impedance: >200Mohm at 10mV reverse bias

Output polarity: Negative

Cable: 3.4mm dia. x 1.8m long, triax

Connector: TNC or BNC

Highlights

Supports Varian®, Elekta®, Siemens® and CyberKnife® delivery systems

Less than 5-minute installation with no additional tools

20cm TPR drain measurement: 1.5 minutes 20cm TPR fill measurement: 3.5 minutes

3D MiniLift™ EDGE Detector™

3D Reservoir™

3D TPR™

Page 28: Your Most Valuable QA & Dosimetry Toolswork.sunnuclear.net/documents/2014-Sun_Nuclear-Product-Solutions... · Your Partner for QA & Dosimetry ... SRS Film QATM ... colleagues in both

Corporate Headquarters 3275 Suntree Boulevard Melbourne, FL 32940 USA