1 Martin Stuhlin ger, ESAC XMM EPIC MOS Monitoring
8

XMM 1 EPIC MOS Monitoring. XMM-Newton 2 Martin Stuhlinger, ESAC MOS Bad Pixel Monitoring.

Dec 22, 2015

Download

Documents

Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: XMM 1 EPIC MOS Monitoring. XMM-Newton 2 Martin Stuhlinger, ESAC MOS Bad Pixel Monitoring.

1Martin Stuhlinger, ESAC

XMM

EPIC MOS Monitoring

Page 2: XMM 1 EPIC MOS Monitoring. XMM-Newton 2 Martin Stuhlinger, ESAC MOS Bad Pixel Monitoring.

XMM-Newton

2

Martin Stuhlinger, ESAC

MOS Bad Pixel Monitoring

Page 3: XMM 1 EPIC MOS Monitoring. XMM-Newton 2 Martin Stuhlinger, ESAC MOS Bad Pixel Monitoring.

XMM-Newton

3

Martin Stuhlinger, ESAC

MOS Background Monitoring

Page 4: XMM 1 EPIC MOS Monitoring. XMM-Newton 2 Martin Stuhlinger, ESAC MOS Bad Pixel Monitoring.

XMM-Newton

4

Martin Stuhlinger, ESAC

MOS Serial CTI (Al)

Page 5: XMM 1 EPIC MOS Monitoring. XMM-Newton 2 Martin Stuhlinger, ESAC MOS Bad Pixel Monitoring.

XMM-Newton

5

Martin Stuhlinger, ESAC

MOS Parallel CTI (Al)

Page 6: XMM 1 EPIC MOS Monitoring. XMM-Newton 2 Martin Stuhlinger, ESAC MOS Bad Pixel Monitoring.

XMM-Newton

6

Martin Stuhlinger, ESAC

MOS1 Parallel CTI CCD1

Page 7: XMM 1 EPIC MOS Monitoring. XMM-Newton 2 Martin Stuhlinger, ESAC MOS Bad Pixel Monitoring.

XMM-Newton

7

Martin Stuhlinger, ESAC

MOS2 Parallel CTI CCD1

Page 8: XMM 1 EPIC MOS Monitoring. XMM-Newton 2 Martin Stuhlinger, ESAC MOS Bad Pixel Monitoring.

XMM-Newton

8

Martin Stuhlinger, ESAC

Conclusions

Bad pixel situation is still stable on low level.

• Possible adjustment of background: decrease fixed-offset table of several CCDs by 1 ADU to compensate long-term drift.

• Serial CTI stable since cooling except MOS2 CCD7.

• Parallel CTI stable since cooling. Science ~rev. 700 parallelCTI nearly constant.

• Current CTI model still valid.