Www.GasBadge.com. What is the It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor .

Dec 24, 2015

Download

Documents

Jocelyn Hall
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Www.GasBadge.com. What is the  It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor .

www.GasBadge.com

Page 2: Www.GasBadge.com. What is the  It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor .

What is the

www.GasBadge.com

Page 3: Www.GasBadge.com. What is the  It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor .

It is the Next Generation, Calibration Station for the GasBadge®

Plus Personal Monitor

www.GasBadge.com

Page 4: Www.GasBadge.com. What is the  It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor .

Optional PrinterLCD Display

Instrument Bay

Bump TestButton

CalibrateButton

www.GasBadge.com

Page 5: Www.GasBadge.com. What is the  It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor .

Physical Specifications

• Desktop unit• Can mount directly on a wall • Universal power input• Fresh air port• One gas inlet port • Serial printer interface• USB interface for PC• Dual push-buttons

www.GasBadge.com

Fresh Air Gas inlet

Page 6: Www.GasBadge.com. What is the  It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor .

Functionality

• Automatic bump tests• Automatic calibration• Automatic report generation• USB interface for connection to PC• Download event logger• Stores last 200 reports (non-

volatile memory)• External serial printer interface• Optional built-in printer• Supports GasBadge Plus

www.GasBadge.com

Page 7: Www.GasBadge.com. What is the  It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor .

Bump Test

• Push button activated• Measures sensor response• 50% response = pass (toxic)

– 19.6% for O2• Bump test report printed

automatically– To serial port / printer– To internal printer– Can be downloaded to PC

www.GasBadge.com

Page 8: Www.GasBadge.com. What is the  It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor .

Bump Test Report

Prints automatically if external printeris connected, or if internal printer is installed

www.GasBadge.com

Industrial Scientific Corp.Cal Plus v1.0.025 Dec 04 Time: 14:45Gas Badge (or Gas Badge Pro)Serial Number: 1234567890Software v1.0Hardware v2.0Bay Number 1 (for multiple bay version)Bump Test: PASSSensor: COGas Reading: 61 ppmCal Gas Conc.: 100 ppmHigh Alarm: 70 ppmLow Alarm: 35 ppmCylinder Lot No.:_____________________________Technician Initials:_____________________________________________________________________

Page 9: Www.GasBadge.com. What is the  It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor .

Calibrations

• Push button activated• Adjusts sensors accuracy• Calibration report printed

automatically– To serial port / printer– To internal printer– Can be downloaded to PC

• Will check to see if instrument was previously calibrated, and question user.

www.GasBadge.com

CalAgain?

Page 10: Www.GasBadge.com. What is the  It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor .

Calibration Certificate

Prints automatically if external printeris connected, or if internal printer is installed

www.GasBadge.com

Industrial Scientific Corp.Cal Plus v1.0.025 Dec 04 Time: 14:45Gas Badge (or Gas Badge Pro)Serial Number: 1234567890Software v1.0Hardware v2.0Bay Number 1 (for multiple bay version)Zero: PASSCalibration: PASSSensor: COSpan Reserve: 110 ppmCal Gas Conc.: 100 ppmHigh Alarm: 70 ppmLow Alarm: 35 ppmNext Cal Due: 24 Jan 05Cylinder Lot No.:_____________________________Technician Initials:_____________________________

Page 11: Www.GasBadge.com. What is the  It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor .

Display

• Graphic LCD• Displays actions of CalPlus• Backlight • Multiple languages – English, French,

German, and Spanish• LED’s show status (pass/fail)

Ready Zero inProgress

www.GasBadge.com

Page 12: Www.GasBadge.com. What is the  It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor .

Modes of operation

• Print Event Log• Print Cal and Bump memory buffer• Clear memory buffer• Select calibration frequency (days)• Set real time clock• Select language• Perform system check

– Checks pump, board, power, printer, solenoid, and EEPROM

www.GasBadge.com

Page 13: Www.GasBadge.com. What is the  It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor .

Modes of operation

• Enter by pressing and holding both buttons for 3 seconds

• “Bump” button scrolls through different modes of operation

• “Calibrate” button enters that mode of operation

www.GasBadge.com

Page 14: Www.GasBadge.com. What is the  It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor .

Modes of operation

• Print Events– Prints event log from

instrument currently connected– Press “Calibrate” to print– Press “Bump” to go to next

mode– If “Calibrate” is pressed, Cal

Plus checks to ensure this is OK

www.GasBadge.com

PrintingEvents

If the calibrate button is pressed while the ‘Print Events’ screen is displayed, the

‘Print OK’ screen is shown, verifying the users selection to download and print the

saved data from the instrument. If the user presses the bump button, No is

selected and the user is returned to the ‘Print Events’ screen. If the user presses calibrate, Yes is selected and the data is downloaded and sent out the printer port, ‘Printing Events’ is displayed. When the printing is finished, the user is taken to

the ‘Print Buffer’ screen.

PrintEvents

Print OKN Y

Page 15: Www.GasBadge.com. What is the  It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor .

Modes of operation

• Print Buffer– Prints entire buffer of reports– Press “Calibrate” to print– Press “Bump” to go to next

mode– If “Calibrate” is pressed, Cal

Plus checks to ensure this is OK

www.GasBadge.com

Printing...

If the calibrate button is pressed while the ‘Print Buffer’ screen is displayed, the

‘Print OK’ screen is shown, verifying the users selection to print every saved bump

and cal report. If the user presses the bump button, No is selected and the user is returned to the ‘Print Buffer’ screen. If

the user presses calibrate, Yes is selected and every saved bump and cal

report are sent out the printer port, Printing’ is displayed. The buffer is NOT erased. When the printing is finished, the user is taken to the ‘Clear Buffer’ screen.

PrintBuffer

Print OKN Y

Page 16: Www.GasBadge.com. What is the  It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor .

Modes of operation

• Clear Buffer– Clears entire buffer of reports

• Buffer only clears when forced

• Buffer over-writes when full FIFO (first in first out)

– Press “Calibrate” to clear– Press “Bump” to go to next mode– If “Calibrate” is pressed, Cal Plus

checks to ensure this is OK

www.GasBadge.com

ClearBuffer

Clear OKN Y

If the calibrate button is pressed while the ‘Clear Buffer’ screen is displayed, the

‘Clear OK’ screen is shown, verifying the users selection to clear every saved

bump and cal report. Pressing the bump button will select No and return the user

to the ‘Clear Buffer’ screen. While pressing the calibrate button will select

Yes, and clear the records and then take the user to the ‘Cal Days’ screen.

Page 17: Www.GasBadge.com. What is the  It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor .

Modes of operation

• Set Calibration Frequency– Printed on calibration certificates

• Set in days (0-365)• Added to date programmed in

station– Press “Calibrate” to set # of days– Press “Bump” to go to next mode– If “Calibrate” is pressed,

• “Bump” increases value• “Calibrate” accepts value

www.GasBadge.com

Cal Days30

30 ^ OK

If the cal button is pressed while the ‘Cal Days’ screen is displayed, the ’30 Days’ screen is

shown, allowing the user to increase the number of days until the next calibration by

pressing the bump button. The number 30 will increase slowly (by 1 at a time), then faster

while the bump key is continually pressed. The rollover point is set at 365 days, then the counter will start back at 0. Pressing the

calibrate button will save the new number of days (that is displayed) in memory and take the

user to the ‘Set Date’ screen.

Page 18: Www.GasBadge.com. What is the  It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor .

Modes of operation

• Set Date / Time– Printed on calibration / bump

certificates • International format dd/mm/yy• 24 hour clock format

– Press “Calibrate” to set date– Press “Bump” to go to next mode– If “Calibrate” is pressed,

• “Bump” increases value• “Calibrate” accepts value

www.GasBadge.com

Set Date

01 Month ^ OK

If the cal button is pressed while the ‘Set Date’ screen is displayed, the ’01 Month’ screen is

shown, allowing the user to increase the month to set the correct date. The month number will

increase by 1 at a time. The rollover point is set at 12 months, then the counter will start back at

1. Pressing the calibrate button will save the new month to the clock, and bring up the ’01 Day’ screen, which will cycle up to 31, then start back at 1. Pressing the calibrate button will save the new day and bring up the ’05

Year’ screen. The year (last 2 digits only) will cycle up to 99 and then start back at 00.

Pressing the calibrate button will save the new year to the clock and bring up the ’12 Hour’

screen. The clock is set using a 24-hour clock, the hours will cycle up to 23, then start back at 00. Pressing the calibrate button will save the

new hour to the clock and bring up the ’00 Min’ screen. This screen will cycle up to 60 and then start back at 00, by pressing the bump button.

Pressing the calibrate button will save the minutes to the clock and return the user to the

‘Select Language’ screen.

01 Day ^ OK

05 Year ^ OK

12 Hour ^ OK

00 Min ^ OK

Page 19: Www.GasBadge.com. What is the  It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor .

Modes of operation

• Select Language

– English, Spanish, French and German

• Appears on display

• Prints certificates in this language

– Press “Calibrate” to change language

– Press “Bump” to go to next mode

– If “Calibrate” is pressed,

• “Bump” scrolls to next language

• “Calibrate” accepts language

www.GasBadge.com

SelectLanguage

English^ OK

Espanol^ OK

Francais^ OK

Deutsch^ OK

If the cal button is pressed while the ‘Select Language’ screen is displayed,

the ’English’ screen is shown. This allows the user to select English as the

default language by pressing the calibrate button, and then the user will taken to the ‘Exit’ screen. Pressing the bump button will bring up the ‘Espanol’ selection. While ‘Espanol’ is displayed,

pressing the calibrate will select Spanish as the default language and

take the user to the ‘Exit’ screen. Pressing the bump button will bring up the ‘Francais’ screen, which allows the

user to select French as the default language. By pressing the calibrate

button, French is selected and the user is taken to the ‘Exit’ screen. Pressing

the bump button will bring up the ‘Deutsch’ screen, allowing German to be selected as the default language.

Pressing the bump button will return the user to the ‘Select Language’ screen.

Pressing the calibrate button will select German as the default language and

take the user to the ‘Exit’ screen.

Page 20: Www.GasBadge.com. What is the  It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor .

Modes of operation

• System Check

– Tests current draw, pump, solenoids, and printer

– Press “Calibrate” to run system check

– Press “Bump” to go to next mode (Exit)

– If “Calibrate” is pressed, test starts

• If tests fail, error messages displayed

www.GasBadge.com

SystemCheck

RunningCheck...

If the bump button is pressed while the ‘System Check’ screen is displayed, the user is taken to the ‘Exit’ screen. If the cal button is pressed while the ‘System

Check’ screen is displayed, the calibrator starts a check of all the

systems. Any errors will be reported as shown in figures 2 and 3, when the

check is finished. These errors will be displayed for 30 seconds, then the calibrator will return to the ‘System

Check’ screen. Once the user exits the setup mode, the errors will be visible

again on the display.

Page 21: Www.GasBadge.com. What is the  It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor .

Modes of operation

• Exit– Press “Calibrate” to exit– Press “Bump” to go to next mode

(Print Buffer)

www.GasBadge.com

Exit

Ready

If the bump button is pressed while the ‘Exit’ screen is displayed, the user is

taken back to the ‘Print Events’ screen. If the cal button is pressed while the ‘Exit’ screen is displayed, the set up mode is exited and the calibration

station is returned to normal operation. If there is no instrument installed, the

station will display ‘Ready’. If there is an instrument installed, and it has been

through a calibration or bump test, the result will be displayed, i.e. ‘Cal Pass’.

Page 22: Www.GasBadge.com. What is the  It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor .

Modes of Operation

PrintBuffer

ClearBuffer

Cal Days30

SetDate

SelectLanguage

Exit

PrintEvents

SystemCheck

www.GasBadge.com

First Mode

“Bump” button selects mode“Calibrate” button enters mode

Page 23: Www.GasBadge.com. What is the  It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor .

PC interface

• Connects to PC via USB• View and save:

– Bump reports– Calibration reports

• Download event log from instrument• Downloads calibration and bump

reports• Setup instrument

www.GasBadge.com

Page 24: Www.GasBadge.com. What is the  It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor .

www.GasBadge.com

Page 25: Www.GasBadge.com. What is the  It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor .

www.GasBadge.com

Sample Calibration Certificate

Page 26: Www.GasBadge.com. What is the  It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor .

www.GasBadge.com

Sample Bump Test Certificate

Page 27: Www.GasBadge.com. What is the  It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor .

Sample Event log

Page 28: Www.GasBadge.com. What is the  It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor .

Accessory Items

• Calibration gas• Calibration gas cylinder holder (wall

mount)• Internal printer• External serial data printer (120

VAC)• External battery powered printer

(battery only)

www.GasBadge.com

Page 29: Www.GasBadge.com. What is the  It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor .

Pricing

• Non-printer version: $795.00 USD• Printer version: $1095.00 USD

www.GasBadge.com

Page 30: Www.GasBadge.com. What is the  It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor .

Availability

• Taking orders September 1, 2005

www.GasBadge.com

Page 31: Www.GasBadge.com. What is the  It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor .

Any Questions?

www.GasBadge.com