What is Spectroscopy? • The study of the interaction of electro-magnetic radiation with matter. • We use spectroscopy as a tool to look indirectly at molecules. Spectroscopy
Jan 19, 2016
What is Spectroscopy?
• The study of the interaction of electro-magnetic radiation with matter.
• We use spectroscopy as a tool to look indirectly at molecules.
Spectroscopy
Electromagnetic Radiation
E = Electric field
H
E
H = Magnetic field
y
x
z
Direction ofpropagation
Wavelength
• Transports energy.
• Electric and magnetic fields oscillate: that’s the “wave”.
• Moves at speed of light.
• Wavelength, frequency, energy all related.
WavelengthDistance between crests
AmplitudeHalf the height of trough to
crest
FrequencyNumber of crests that pass a point in space every second
E = h c / = c / = c /
c = speed of light (3 x 108 m/s)E = energy = wavelength = frequency h = Planck’s constant (4 x 10-15 eV sec)
Spectroscopy
Wavele
ng
th
(nm
)
En
erg
y a
s w
aven
um
bers
(c
m-1)
Microwave
Infrared
Visible
Ultraviolet
X-rays
Vibrational
transitions
104
105
103
102
10
1
103
102
104
105
106
107
Rotationaltransition
s
Electronictransition
sInner shell
Electronictransition
sOuter shell
Each region
Differentspectroscopic technique
The spectroscopic technique used to study a particular transition will depend on the energy difference between the ground and excited states (E).
Different frequencies of light (electromagnetic radiation) interact with different aspects molecular motion (potential
and kinetic energy).
IR: molecular vibrations
UV-Vis: valance electroni
c excitatio
n
X-ray: core
electron excitation
Radio waves:Nuclear spin states(in a magnetic field)
Transitions between energy levels which involve electrons, electronic transitions, between
molecular orbitals.
Ultraviolet-Visible Spectroscopy
In UV/Vis spectroscopy we talk in terms of the
wavelength () of the transition.
Nuc
Level 1
Level 2
Level 3
Gain energyMove to level 2 or
3
Lose energyDrop to level 2 or
1
Energy levels of electrons
Transitions occur between200-400 nm in the UV
region400-750 nm in the visible
regionFor a transition to occur the incident radiation must have the correct energy to excite the molecule from the ground state to a higher, ‘excited’ state
Molecular Geometry
En
erg
y
S0 (Ground Estate)
S1 (Excited Estate)
rotational levels
vibrational levels
UV-Vis Absorption Spectroscopy
h
Important Terms and Symbols in UV-Vis Absorption
SpectroscopyTerm and Symbol
Radiant Power (P, P0)
Absorbance (A)
Transmittance (T)
Path length (b)
Absorptivity (a)
Molar absorptivity ()
Definition
Energy of radiation
Log (P0/P)
P/P0
-
A/lc
A/lc
Alternative Name and Symbol
Radiation Intensity (I, I0)
Optical density (OD); Extinction (E)
Transmission (T)
l, d
Extintion coefficient (k)
Molar Extintion coefficient
Beer-Lambert Law
A c l=
= absorbance (no units).
= Molar extintion coefficient (M-1 cm-1).
= Concentration (M).
= pathlength (cm).
A
c
l
Beer-Lambert Law
I0
I
S
l
dx
Block of absorbing matter
A beam of incident intensity I0 passes through length l of an absorbing medium, which decreases the transmitted intensity to I.
Considering a cross-section of area S and infinitesimal thickness dx containing dn molecules, each of which has a cross-sectional surface of photon capture of a.
Based on the probability of photon capture, we can imagine the total area of the molecular photon capturing surface to be dS, and thus the ratio of photon capture surface area to total surface area is dS/S.
This ratio represents the probability of photon capture.
Beer-Lambert Law
I0
I
S
ldx
The intensity entering the section is Ix, proportional to number of photons per square centimeter. Loss of intensity through absorption is given by dIx, which is an absolute quantity; thus the relative loss of power is:
= probability of photon capture.
dS
S
Now consider the light intensity passing through distance dx.
dIx
Ix(minus sign indicates loss)
dIx
Ix
dS
S=
The relative loss of power must be equal to the relative probability for capture.
Beer-Lambert Law
I0
I
S
ldx
Note that dS is the sum of all capture areas for all molecules, and must therefore be proportional to the number of molecules
dS = adn (concept of absorptivity)
where a is the photon capture cross section (molecular absorptivity co-efficient).Integrating over all of the molecules within the entire block (to obtain the total absorbance over an infinite number of slices of thickness dx), we get:
dIxIx
=Io
I adnS0
n
which gives:II0
=anS
-ln
Now convert to base 10 and invert the fraction to change the sign:
I0
I=
an2.303S
log
Beer-Lambert Law
I0
I
S
ldx
The cross-sectional area S can be expressed as a volume (cm3 or mL) divided by a length (pathlength b in cm), thus:
Finally, since n/V has units of concentration (molecules per cm3), we can convert n/V to moles per liter according to:
Which leads to:I0
I=
anl2.303V
log
=Vl
S in cm2
Combining everything together, we get:
1000 cm3/L
V(cm3)=c
n6.02 x 1023
mol x
6.02 x 1023 alc2.303 x 1000
I0
I=log
Beer-Lambert Law
I0
I
S
ldx
6.02 x 1023 acl2.303 x 1000
I0
I=log
Finally, collecting all of the
constants into a single term, , we get:
A c lI0
I=log =
A c l=
= absorbance (no units).
= Molar extintion coefficient (M-1 cm-1). (wavelength dependent)
= Concentration (M).
= pathlength (cm).
A
c
l
1) Real Limitations• Beer-Lambert law is derived on the basis of having a
dilute solution. At high concentrations, (>0.01 M) intermolecular interactions can occur between molecules (dipole-dipole interactions), directly altering
• Intermolecular interactions can also occur when secondary species (i.e., salts) interact with charged chromophores, thus altering
• The value of is also dependent on refractive index of the medium, thus solvent effects on absorbance can be observed (correct using n/(n2 + 2)2 instead of ).
Limitations / Beer-Lambert Law
2) Apparent Chemical Deviations• Arise as a result of chemical reactions involving the
absorbing sample (association, dissociation, oxidation, reduction, addition, elimination, etc…).
Limitations / Beer-Lambert Law
Classic example is pH sensitive chromophores (indicator dyes). Figure shows the absorbance at 430 nm and 570 nm for a dye that undergoes a protic equilibrium that depends on total dye concentration.
0.00 4.00 8.00 12.00 16.000.000
0.200
0.400
0.600
0.800
1.000
Ab
sorb
ance
Indicator (M x 105)
= 430 nm
= 570 nm
Limitations / Beer-Lambert Law3) Instrumental Deviations due to
Polychromatic Radiation• Strict adherence to Beer-Lambert Law requires
perfectly monochromatic light.
• In cases where the spectral band width of the incoming light is many nanometers, then the sample can have a range of values.
In practice:• Narrower spectral
bandwidth will lead to better linearity.
• Taking measurements at the maximum wavelength
(at max) will lead to better linearity.
Ab
so
rba
nc
e
Ab
so
rba
nc
e
Wavelength Concentration
Band B
Band A
Ba
nd
B
Ba
nd
A
Limitations / Beer-Lambert Law4) Instrumental Deviations due to Stray
Light• Imperfections in monochromator gratings often lead to
0.005 – 0.2% stray light.
• Differs greatly in wavelength from the value of the principal radiation, and may not even pass through sample (i.e., depends on how light-tight the instrument is).
• Observed absorbance in presence of stray light is given by:
A’I0 + Is
I + Is
log=
Where Is is the intensity of non-absorbed stray radiation. This leads to a negative deviation in A vs C, as shown in the Figure0 2.5 5.0 7.5 10.0
1.0
2.0
Ab
sorb
ance
Concentration (M x 103)
5 %
1 %
0.2 %0.0 %
Is
I0
x 100 %
• Measurement of absorbance by molecular samples requires that the sample be placed in some kind of recipient.
• Reflection of light occurs at four different interfaces, leading to ~8.5% loss in transmitted light.
• Attenuation of light can also occur via scattering or absorbance by species other than the solution.
Reflection losses at interfaces
Incident beam, I0
Emergent beam, I
Scattering losses in solution
Reflection losses at interfaces
Instrument components: UV-VIS
Instruments for absorption measurements
signal processo
r
optical
source
h sample
h
detect
or
selector
SourcesA source must:
• Generate a beam of radiation with sufficient power for easy detection and measurement.
• Provide output power that is both stable and intense.
Instrument components: UV-VIS
Types of spectroscopic sources
1. continuous sources
Wavelength
Rela
tive e
nerg
y
H2 and D2 lamp / Xe arc lamp
Tungsten filament lamps
2. lines sources
Wavelength
Inte
nsit
y
Hollow cathode lampHg vapor lamp / Laser
Instrument components: UV-VIS
Sources
6000 K
4000 K3000 K
2000 K
Xenon arcCarbon
arcTungsten lampNerst
glower
Rela
tive
en
erg
y
Wavelength (nm)
500 1000
1500
2000
2500
3000
1
10
102
103
104
Sample containers
Instrument components: UV-VIS
Glass - 340 - 2500 nmQuartz - 200 - 2500 nm
Polystyrene - VisibleMetacrylate - UV
1. Filters
1.1 Interference filters
1.2 Absorption filters
Wavelength selectors
Instrument components: UV-VIS
The simplest method for isolating a narrow band of radiation.
Work by selectively absorbing radiation from a narrow region.
Use constructive and destructive interference to isolate a narrow range of wavelengths.
2. Monochromators
2.1 Prism
Wavelength selectors
Instrument components: UV-VIS
2.2 GratingsFinely grooved highly reflective surface (Diffraction).
Quartz or glass cut at an angle (Refraction).
Collimating mirrors
Entrance slit Exit slit
Grating surface
>
Wavelength selectors
Instrument components: UV-VIS
0.5
1.0
Ab
sorb
an
ce
Wavelength
Nominal wavelength
Effective bandwidt
h
1/2 Peak heigh
t
- +h--
e-
DetectorsInstrument components: UV-VIS
The detector changes the radiation transmitted from the UV-Vis Spectrophotometer into a current or voltage.
1. Phototube
3. Photodiode array
2. Photomultiplier
Light strikes photocathode (-) that emits electrons.Electrons are accelerated towards the anode (+).Current proportional to photons.
Electrons that are accelerated towards a series of increasingly positive anodes (+).Electrons are directed towards the collecting anode.
Multi-element detector composed of an array of solid-state detectors.
Signal processors
Instrument components: UV-VIS
Amplifies the voltage produced by the detector.
Converts the voltage into absorbance units.
Instruments: UV-VISSingle beam UV-Vis spectrophotometer
Instruments: UV-VISDouble beam UV-Vis spectrophotometer
Instruments: UV-VIS
Ultraviolet-Visible SpectroscopyBonding electrons appear in and molecular
orbitalsnonbonding in n
En
erg
y
n
n
n
Antibonding
Antibonding
Nonbonding
Bonding
Bonding
Electronic transitions can occur between various states.
The energy of the transitions increases in the following order:
(n *) < ( *) < (n *) < ( *)
Electronic Transitions
• Most absorption spectroscopy of organic compounds is based on these transitions.
• The absorption peaks for these transitions fall in an experimentally convenient region of the spectrum (200 - 700 nm).
• These transitions need an unsaturated group in the molecule to provide the p electrons.
• Excitation from the ground state to the excited state requires EM radiation with a wavelength of 150 nm.
• Not useful for routine spectroscopy
• The number of organic functional groups with n * peaks in the UV region is small.
• Excitation requires EM radiation with a wavelength in the range 150 - 250 nm.
* bandn * band
n * band
* band
ChromophoreGroup on molecule responsible for electronic transition
• These molecules generally contain conjugated systems.
• Adjacent double bonds are conjugated.
• The net effect is to bring the ground state and excited states closer in energy.
• The longer the chain of conjugation the longer the wavelength of the absorption band.
Bathochromic shift
Hypsochromic shift
Hyperchromic effect
Hypochromic effect
Shift in absorption to a longer wavelength.
Shift in absorption to a shorter wavelength.
An increase in the intensity of absorption.
An decrease in the intensity of absorption.
Some Spectroscopic Terminology
• When one absorbing species reacts to form another absorbing species, there usually is a point(s) at which the spectra cross. This is the isobestic point.
• Example, acid/base indicators.
Isosbestic Point
There are 3 classes of protein chromophores peptide bonds, amino acid side chains and prosthetic
groups.
Most proteins and nucleic acids are colorless in the visible region, however they absorb in the near-UV region. Chromophores in proteins and nucleic acid absorb light only at wavelengths lower than 300 nm. On the other hand, measurements have to be done in wavelengths longer than 170 nm, below which the absorbance of water becomes too high.
Absorption Spectra of Biopolymers
UV absorption spectra of proteins
For a typical protein there is a distinctive absorption peak at 280 nm (due to the * transitions in aromatic amino acids), a stronger one at 190 (due to the * transition in the amide group), and a shoulder at 210-220 nm (due to the weaker n π* transition in the amide group).
Ab
sorb
an
ce
Wavelength (nm)
2.0
1.5
1.0
0.5
0
0.4
0.3
0.2
0.1
250 300200
4.5 x 10-7 M
2.25 x 10-6 M
Chromophores in proteins
The peptide bond
Amino acid side chains
• A number of amino acids – Asp, Glu, Asn, Gln, Arg & His have electronic transitions around 210 nm. Usually these cannot be observed in proteins because they are masked by Intense peptide bond absorption.
• The most useful range for proteins is above 230nm where there are absorptions from the aromatic side chains of Phe, Tyr and Trp.
• The absorption of Phe is low and if Tyr & Trp are present in a protein it contributes little to the absorption above 230nm. The disulphide group of Cys also has a weak absorption ~250 nm which can be important in protein optical activity.
• The absorption spectra of Tyr & Trp contain contributions from at least three electronic transitions. Assignment of individual transitions in proteins containing many Tyr and Trp in a variety of environments is not possible.
Amino acid side chainsThe aromatic Trp, Tyr, and Phe are the only residues absorbing significantly at wavelengths higher than 230 nm. The maximum at 280 nm is due to Trp and Tyr.
-
Mola
r ab
sorp
tivit
y (
M-1 c
m-1)
Tyrosine
Phenylalanine
Tryptophan
200 220 240 260 280 300 320
100005000
2000
1000
10
20
50
100
200
500
4000020000
4000
3000
2000
1000
0190 200 210 220 230
Alanine
Lysine HCl
Methionine
Wavelength (nm)
Absorption Spectra of Biopolymers
The local environment of the amino acids in a native protein depends on the electric properties of the peptide chain near by and the associated solvent. The resulting change in spectra is sufficient to diagnose whether a protein is folded or not, but CD is more informative.
SECONDARY STRUCTURE
UV spectra of poly-L-lysine hydrochloride (random coil) changes upon formation of secondary structure (raising pH induces helix formation - reducing the net positive charge of the lysine side chain -; raising the temperature induces beta sheet).
Many proteins include prosthetic groups such as iron porphyrins in hemeproteins, flavin in flavoproteins, or phosphate groups, that have strong absorption in the visible and near UV. Because these bands are sensitive to the environment, they can be used in the study of enzymatic reactions.
Absorption Spectra of Biopolymers
x 1
0-4
Chromophores in genetic material
The strong near-UV absorption of nucleic acids is due to the purine and pyrimidine bases. The electronic states of these bases are much more complex than those of protein chromophores, and apparent gaussians in the spectra are really composites of several electronic transitions.
The nucleotides all have similar absorption spectra, dominated by an absorption peak at 260 nm and a shoulder at 200 nm.The free base, the nucleoside (attached to sugar), the nucleotide (attached to a sugar phosphate), and the denatured polynucleotide all have similar spectrum.
In general polynucleotides and nucleic acids absorb less per nucleotide that their constituent nucleotides. Also, native double-stranded DNA absorbs less per nucleotide than denatured (melted strands). This results from the interactions between adjacent bases. Stacked base pairs in a double helix absorb less than partially stacked bases in a single strand, which absorb less than the mononucleotides.
Absorption Spectra of BiopolymersUV absorption spectra of nucleic acids
180
200
220
240
260
280
300
0
0.2
0.4
0.6
0.8
1.0E. coli DNA 82°CEnzymatic digest 25°CNative 25°C
(nm)
Ab
sorb
an
ce
What determines absorbance?• The absorbance spectrum for a chromophore under standard
conditions is only partly determined by its chemical structure.
• The environment of the chromophore also affects the precise spectrum obtained.
• pH• Solvent polarity• Orientation effects
• Chromophores can act as reporter molecules which can give information about their immediate environment.
• Defined conditions (pH and solvent).
• Protonation/deprotonation effects resulting from pH changes or oxidation/reduction effect electron distribution in chromophores.
• Solvent polarity –Altered spectra in different solvents (DMSO, dioxane, ethylene glycol, glycerol and sucrose) compared to water = Solvent perturbation.
• Orientation effects result from the relative geometry of neighboring chromophore molecules. E.g. hyperchromicity of nucleic acids.
Applications of absorption spectra
Quantitative
Reaction rate calculation
Standard curve
Structural studies
Enzymatic
Identification
Beer-Lambert law to determine concentration.
Continuous or stopped.
measure concentration.
folding, assembly, denaturation, ligand binding
pathways, reaction intermediates
of compounds (vitamins, hormones) not much use for proteins/nucleic acids, unless they contain groups with absorption in the visible region of spectrum.
Immunodetection commercial detection kits.