Investigation about Log Reduction ofBlood Contamination byWasher-Disinfectors
• We tried to figure out the log reduction of blood contamination by washer-disinfectors with practical cleaning cycles, using the test pieces contaminated quantitatively by sheep blood.
3
Hierarchic Structure of Working Group
• JAPANESE SOCIETY OF MEDICAL INSTRUMENTATION ( JSMI )
–Accrediting Committee of Reprocessing Technicians
–Working Group of Cleaning Evaluation
The WG was set up in April 2006 for cleanliness evaluation with 11 group members assigned.
4
Members of Working GroupRyo Fushimi Osaka Univ. Hosp.
Measurement Result in LaboratorySetting Parameters of Main Wash Residual
ProteinAverage
(μg)℃ min. Detergent (type, v%, pH)
90 5 Alkaline 0.3 11.3 30 30.845
36
23
20
60 10 Alkaline 0.3 11.3 37 31.621
34
29
37
50 10 Enzymatic 0.5 8.2 32 42.458
33
56
33
16
Measurement Result in HospitalsSetting Parameters of Main Wash Residual
ProteinAverage
(μg)℃ min. Detergent (type, v%, pH)
93 10 Alkaline 0.3 11.3 97 68.75851
93 10 Alkaline 0.3 11.3 20 29.73732
60 8 Alkaline 0.5 12.0 21 16.01413
40 5 Alkaline 0.5 11.5 28 28.32631
pretreatment: immersing into enzymatic detergent solution (0.8 v%) for 5 min. prior to processing by WD
17
Conclusion
• The measurements of the residual protein in the laboratory varied between 20 and 58 μg and the rate of protein decrease showed 1/100,000 down to 1/34,000.
• The measurements of the residual protein in the hospitals varied between 13 and 97 μg and the rate of protein decrease showed 1/153,000 down to 1/20,000.
• Washer-disinfectors have power 5 of the log reduction factor of sheep blood.