HOW DID OUR ANTIBODY PERFORM? ANTIBODY CUSTOMER REVIEW: Phospho-ErbB-3 (Y1197) Polyclonal Antibody (STJ90642 )
HOW DID OUR ANTIBODY PERFORM?
ANTIBODY CUSTOMER REVIEW: Phospho-ErbB-3 (Y1197) Polyclonal Antibody (STJ90642 )
What was used?Primary antibody: STJ90642 Phospho-ErbB-3 (Y1197)
Provider: St John's Laboratory
Dilution ratio: 1:1000
Application: Western Blot
Materials for validation: MCF7 cell line
Click for product data sheet PDF
What Was The Protocol?Treatment of Materials When cells were cultured to 70%-80% plating density, lysis buffer with protease inhibitors
(Sigma) and equal volume of 2X loading buffer were added for 10 minutes at 95˚C.
Gel Electrophoresis 10% Polyacrylamide gel with a constant voltage of 160V for 60 minutes.
Transfer 0.45 µm PVDF membrane, 1 piece of gel (wet transfer), constant voltage 100V for 75 minutes.
Blocking 1X TBST with 5% BSA for 1 hour.
Primary Antibody Probing Primary antibody was added to the solution of 1X TBST with 5% BSA, and left at 4˚C overnight.
Membrane Wash Membrane was washed with 1X TBST 3 times.
Secondary Antibody Probing Secondary antibody was added to the solution of 1X TBST with 5% BSA with a 1:2500 dilution ratio, for 1 hour.
Membrane Wash Membrane was washed with 1X TBST 3 times.
Visualization ECL visualization for 3 min (the visualization time was determined by actual result. It was ensured that the visualization time was identical for the same antigen).
What was the protocol?Gel electrophoresis information: 10% polyacrylamide gel, constant voltage of 160V for 60 minutes.
Transfer information: 0.45 µm PVDF membrane, 1 piece of gel (wet transfer), constant voltage 100V for 75 minutes.
No. Antigen Loading amount
Primary antibody
Primary antibody
dilution ratio
Secondaryantibody
dilution ratioTarget band KD Visualization
time
1 MCF7 15µg
STJ96177Phospho-ErbB-3
(Y1197) antibody
1:1000 1:2500 ~170KD 40S
What were the results?
What did the customer think?
Antibody Specificity:
Antibody RPPA Rating:
Testing results were provided independently by R. Bernardo at the Royal College of Surgeons in Ireland. They stated “In MCF7 cell line, there were target bands detected by Phospho-ErbB-3 (Y1197) STJ90642 Antibody.”