OlympusFluoview FV500

Training

Welcome to...

Cancer Center Confocal Core

Page 2: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

1)Tomakestateoftheartimagingtechnologiesavailableto1)Tomakestateoftheartimagingtechnologiesavailabletothelargerscientificcommunityofusersthelargerscientificcommunityofusers

2)Topromotemoreinteractionbetweentechnology2)Topromotemoreinteractionbetweentechnologydevelopersandusersdevelopersandusers

3)Tobringexpertiseinnewimagingtechnologiestostudents3)TobringexpertiseinnewimagingtechnologiestostudentstheUMNtheUMN

Page 3: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

Contentofthelecture

Principlesofconfocalimaging.Differentimplementations/modes.

Primeronusage

Imageprocessingandrestoration

Page 4: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

LaserScanningConfocalMicroscope

Page 5: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

•High quality, multi-wavelengthfluorescence images•3-Dimensional information

•Optical sections through thick sample•Microscope Techniques:

•Colocalization•Fluorescence Recovery afterPhotobleaching (FRAP)•FRET•Time Course

Why use Why use ConfocalConfocal??????

Page 6: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

ConfocalMicroscopesAvailable

•• OlympusFV500OlympusFV500 MCRB506MCRB506

•• OlympusFV1000uprightOlympusFV1000upright 1‐205NHH1‐205NHH

•• OlympusFV1000inverted1‐205NHHOlympusFV1000inverted1‐205NHH

•• ZeissSpinningDiskZeissSpinningDisk 1‐205NHH1‐205NHH(comingsoon(comingsoon––livecellcapable)livecellcapable)

•• NikonC1sispectralNikonC1sispectral 13SnyderHall13SnyderHall

Page 7: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

Confocalmicroscope.Design

www.olympusconfocal.com

•Inputpinhole

•Exitpinhole

•Scanninghead

•Detector–PMT

Page 8: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

WidefieldvsConfocalImages

http://www.meyerinst.com/confocals/tcs-spe/side-by-side.jpg

Page 9: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

Confocal ImagingConfocal Imaging The key feature of confocal microscopy is that the “in focus”

region of a specimen is the region detected. There is controlledrejection of stray light.

This gives the advantage of observing specimens whose “out-of-focus” illumination contributes to degradation of the detectedimage resolution.

The disadvantage is that the confocal microscope doesn'tproduce a complete optical image of the specimen. So theoptical sections must be reconstructed to generate a 3D view.

Acquisition speed Quantum Efficiency

Page 10: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

Confocalmicroscopevs.widefield.

www.olympusconfocal.com

Page 11: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

PointscanningConfocalmicroscopevs.widefield.

Thewholepictureistakenatonce

Eyepieceimage

Potentiallyfastimaging

Highphotodamage

Highbackgroundnoise(secondaryfluorescence)

“Cheap”

Thinsections(.5‐1.5μm)

Maxthickness~50μm

Highcontrastanddefinition

Reducedphoto‐damaged

Scanning

“slow” “‐”eyepieces

digitalzooming

Limitednumberoflasercolors

expensive

Widefield Confocal

Page 12: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

Depth of FieldDepth of Field

4x 10x

40x20x

Page 13: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

PrincipleofconfocalmicroscopyInconfocalmicroscopytwopinholesaretypicallyused:

– Apinholeisplacedinfrontoftheilluminationsourcetoallowtransmissiononlythroughasmallarea

– Thisilluminationpinholeisimagedontothefocalplaneofthespecimen,i.e.onlyapointofthespecimenisilluminatedatonetime

– Fluorescenceexcitedinthismanneratthefocalplaneisimagedontoaconfocalpinholeplacedrightinfrontofthedetector

– Onlyfluorescenceexcitedwithinthefocalplaneofthespecimenwillgothroughthedetectorpinhole

– Needtoscanpointontothesample

CONDENSER LENS

OBJECTIVE LENSBIOLOGICAL

SAMPLE

OUT-OF-FOCUS PLANE

OUT-OF-FOCUS PLANE

"POINT" SOURCE OF LIGHT "POINT"

DETECTOR APERTURE

IN-FOCUS (OBJECT) PLANE CONTAINING ILLUMINATED SPOT

Page 14: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

In‐focusLightDetected

Page 15: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

Out‐of‐FocusLightBlocked

Page 16: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and
Page 17: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and
Page 18: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

Before Using the Confocal

• Fill in access form with user information, proof of

HazardousChemicalWastetrainingandIDcardnumber.

• Turnintotheconfocaladministrator(Margaret,470CCRB,

322CRorMMC806).

• Readmanualandpolicydocuments.Handbooksoon.

• CompleteadvancedtrainingcoursegivenbyMarkSanders

or otherwise demonstrate familiarity with the FV500

system.You Are Here

Page 19: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

The goals of a shared instrument grant areto provide access to state-of-the-arttechnology that will facilitate interactionand enhance scientific productivity.

A confocal microscope typifies thattechnology, which because of cost is usuallybeyond the scope of a single laboratory toeither purchase or maintain.

NCRRSHAREDINSTRUMENTATIONGRANT(#1NCRRSHAREDINSTRUMENTATIONGRANT(#1S10RR16851)S10RR16851)

Page 20: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

UNIVERSITY OF MINNESOTA

BONE TUMOR BIOLOGY LABORATORY

DEPARTMENT OF ORTHOPAEDIC SURGERY

AND NCI DESIGNATED COMPREHENSIVE

CANCER CENTER

Room 506 CCRB

Page 21: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

System Components:1. BX61 Upright Motorized Research Microscope2. FV500 Laser Scan Head3. Laser Combiner and Argon, Gr HeNe and Red HeNe lasers4. Fiber Optic Delivery System5. Transmitted Light Detector6. BX61 Handswitch, Prior Motorized Stage Control Joystick,

Transmitted Light Source7. Microscope Control Unit8. Argon, Green HeNe and Red HeNe Laser Power Supplies9. Prior Motorized Stage Controller10. Mercury Burner Power Supply11. FV500 Control Unit and Power Supply12. FV500 Computer13. Computer Monitor and Surge Protected Power Outlet

Page 22: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

PrincipleandAdvantagesofConfocalImaging

Confocalmicroscopyoffersseveraladvantagesoverconventionalopticalmicroscopy,includingcontrollabledepthoffield,theeliminationofimagedegradingout‐of‐focusinformation,andtheabilitytocollectserialopticalsectionsfromthickspecimens.Thekeytotheconfocalapproachistheuseofspatialfilteringtoeliminateout‐of‐focuslightflareinspecimensthatarethickerthantheplaneoffocus.

Page 23: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

A confocal imaging system achieves out‐of‐focus rejection by twostrategies: a) by illuminating a single point of the specimen at any onetimewithafocusedbeam,sothat illumination intensitydropsoffrapidlyaboveandbelowtheplaneoffocusandb)bytheuseofblockingapinholeapertureinaconjugatefocalplanetothespecimensothatlightemittedaway from the point in the specimen being illuminated is blocked fromreachingthedetector.

Page 24: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

FV550VisiblelasersBlueDiode[3mW,405nm]yieldsgoodexcitationforthefollowingfluorochromes:

• DAPI–nuclearstain

Argon[10mW,488nm]yieldsgoodexcitationforthefollowingfluorochromes:

• GFP–greenfluorescentprotein

• FITC–antibodylabeling

• YOYO‐1–DNA

• Calcein–bonegrowth

GreenHeliumNeon (HeNe) [1mW,543nm]yields goodexcitation for the followingfluorochromes:

• PropidiumIodide(PI)–DNA,RNA

• TRITC–antibodylabeling

• Cy3–antibodylabeling

• Rhodamine‐Phalloidin–actinfibers

RedHeNe[10mW,633nm]yieldsgoodexcitationforthefollowingfluorochromes:

• Cy5–antibodylabeling

• AlexaFluor633–antibodylabeling

Page 25: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

DYE [excitation/emission] Laser DYE [excitation/emission] Laser

Ca++ Crimson [590/615] Red HeNe eGFP [488/509] ArgonCa++ Green-1 [506/531] Argon FITC [494/518] ArgonCa++ Green-2 [503/536] Argon Fluo-3 [506/526] ArgonCa++ Green-5N [506/532] Gr HeNe Fura Red [472/657] Red HeNeCa++ Orange [549/576] Gr HeNe MitoTracker [490/516] ArgonCa++ Orange-5N [___/580] Gr HeNe PI [536/617] Gr HeNeCy 3 [550/565] Gr HeNe Rhodamine-Phalliodin [542/565] Gr HeNeCy 3.5 [581/596] Gr HeNe SNARF-1 (pH !) [488/530] Red HeNeCy 5 [650/670] Red HeNe Texas Red [595/615] Red HeNeCy 5.5 [675/695] Red HeNe TRITC [496/520] Gr HeNeDil [549/565] Gr HeNe YFP [514/527] Gr HeNeDiO [484/501] ArgonNOT on list:

AlexaFluor-1 [632/647] Red HeNe Calcein [494/517] ArgonAPC [650/660] Red HeNe DS Red [558/583] Red HeNeAPC/Cy7 [650;755/767] Red HeNe YoYo-1 [491/509] Argon

TableListingtheDyesinthe<AvailableDyes>ListandSomePopularDyesNotontheList

Page 26: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

System Start-up [1]

TurnontheSurgeProtectedPowerOutlet(Computer,monitor,FV500ControlUnit&LG‐PS2).PressthePOWERbuttononthetowertoturnonthecomputer.

Page 27: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

System Start-up [2]

TurnonLaserPowerSuppliesONLYasNEEDED(Argon,GreenHeNe,RedHeNe)and“ON”fortheArgonlaser.Therecommendedwarm‐upis10minfortheArgonand30minfortheGrHeNe.

RedHeNe

GreenHeNe

Argon

TurnontheMercuryBurner

Page 28: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

System Start-up [3]TurnontheBX‐UCB(MicroscopeControlUnit)

TurnonthePriorStageControllerPowerSupply

Turntheredleveronthenitrogentank90°counterclockwisetopowertheairtable

Page 29: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

Waitat~2minforthemicroscopesystemstoinitializethendouble‐clicktheFluoviewIcontolaunchtheconfocalprogram.Note:ittakesabout2minutestolaunchafterclickingontheFluoviewIcon

LogontothecomputerusingUSERNAMEandPASSWORD

You will see the Main Fluoview Software Window:

Page 30: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

DONOT moveobjectivesbyhandonthemicroscope,asthiswillcauseasoftwareerror

UsingtheAutomatedBX61Microscope

MotorizedFeaturesoftheMicroscope:

1. ObjectiveNosepiece

2. FluorescentMirrorCubeTurret

3. FluorescentLightShutter

4. CondenserTurret

5. CondenserTopLens

6. ReflectedLightAperture

7. ReflectedLightFieldDiaphragm

Page 31: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

ConfiguringtheMicroscope

Themicroscopeandscanunit(FV500)canbeconfiguredontheFLUOVIEWsoftware.

Usethefollowingproceduretoconfigurethemicroscope:

Fromthepanelpagetabsshownonthebottomrightofthe[acquire]panel,selectthe[settings]sub‐panel.

Selectthe<scopecontrol>buttonatthebottomofthepanel.Thewindowasshownwillappear.

Page 32: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

GraphicUserInterface(GUI)[Light Path] box

selects the light path <BI> is for directobservation, <LSM> is for scanning,<TV> not used.

[Mirror Unit] box

clicking the cube automaticallyswitches the turret.

[Shutter] box

click to open (9:00) / close (12:00) theEPI shutter.

[Nosepiece] box

click to choose the objective.

[Top Lens] box

clicking the engages (centered icon) ordisengages (to side) the lens into thelight path.

[Condenser] box

selects the condenser used.

[Aperture Iris] box

changes the AS value.

Page 33: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

Whythe20xobjectivewasnotherefor9months

EscapethestageBEFOREBEFOREchangingobjectives

Page 34: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

Thissoftwareusespanel‐typewindows.Functionsareexecutedbyselectingthepanelpagetab(attheedgeofthepanel)ofthefunctiontobeexecutedandthepanelwillmovetothefront.

TheFLUOVIEWsoftwareisorganizedbytwokindsofpanels,thefunctionpanels(left,downside)anddisplaypanel(bigoneonright)

UsingtheFLUOVIEWSoftware

Page 35: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

Thepanelstructureofthesoftwareisset‐upaccordingtothetreebelow

Page 36: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

FocusingontheSpecimen–CombinationwithBX61

1. Fromthepanelpagetabsshownonthebottomrightofthe[acquire]panel,selectthe[settings]sub‐panel.

2. Selectthe<BI>buttoninthe[lightpath]groupbox.The<BI>buttonlookspushedintoindicatethatitisselected.

3. Pushthecubebutton(3)ofthehandswitch(2)toengagetheoptimumcubeforspecimendyd.Inthecubedisplaywindow(1),theselectedcubeisdisplayed.

4. Focusonthespecimenbylookingintotheeyepiece.

Page 37: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

Switchingtoscanmode:

Selectthe<LSM>buttoninthe[lightpath]groupbox.The<LSM>buttonlookspushedintoindicatethatitisselected.

NOTE: whenscanningisstartedwhilethe<BI>buttonisselected,theLSMlightpathisselectedautomatically.Itisswitchedbacktothevisual observationautomaticallywhenscanningcompletes.

Page 38: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

SelectingtheDyingMethod

Page 39: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

SettingtheChannels

Page 40: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

SettingtheScanSpeed

SettingtheXYObservationMode

Page 41: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

SettingtheCross‐sectiontobeObserved

Page 42: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

StoppingRepeatedScanning

AcquiringImage

Page 43: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

SavingImage[1]

Page 44: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

SavingImage[2]

Page 45: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and
Page 46: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

Saving, Opening and Shredding Images Saving,OpeningandShreddingImages

Page 47: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

ShreddingImages

Page 48: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

DisplayinganImageinSimulatedColors

Page 49: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and
Page 50: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and
Page 51: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

ImageProcessing

Page 52: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

ImageAnnotationImagescanbeannotatedbyclickingtheAnnotationbuttoninthetoolbar

Page 53: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

ImageAnalysis

Page 54: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

TransferringAnalysisDatatoAnotherApplication[1]

Page 55: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

TransferringAnalysisDatatoAnotherApplication[2]

Page 56: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

Viewing3DImage

Page 57: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

SystemShut‐down[1]

• Cleanobjectivesofanyresidualoilusinglenspaper

• Returnmicroscopestagetonon‐escapedposition(UP)

• Returnobjectivesto4x• ExitfromtheFVprogram.Thismaytakeawhile,howeverifyouturnoffthemicroscopebeforeexiting,asystemerrorwilloccur

• Transferyourfilestoremovablemedia.Only1.5GBofstorageisallowed

• TurnoffthePriorStageControllerPowerSupply• TurnofftheBX‐UCB(MicroscopeControlUnit)

• TurnofftheMercuryBurner.

Page 58: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

• TurnoffLaserPowerSupplies

• Select<shutdown>onthecomputerandselectLOGOFFfromthedrop‐downmenu.Thecomputerwilllogyournameoutandreturntotheloginmenubox.

• Selecttheshutdownbuttontoshutdownthecomputer.Note:powerwillgooff.

• TurnofftheSurgeProtectedPowerOutlet.Donotturnthemonitor,FV500ControlUnitorLG‐PS2off.

• Turnredleverontank90°clockwisetostopflowofgastotable.

• Signoutonthelogsheet.

• Reportanyproblems.

SystemShut‐down[2]

Page 59: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

On-line Reservations for Confocal Microscope [1]

Signupusingtheon‐linereservationscheduleat:

<http://www.umncc.calendarhost.com/confocal/cgi-bin/calweb/calweb.cgi>

Reservationsareonafirst‐come,first‐servebasis.

Ifyoumustcancel,dosoinadvance.

LoginwithIDandpassword,bothcasesensitive.Thesameonesareusedforlog‐intotheFV500.

Page 60: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

On-line Reservations for Confocal Microscope [2]Selectdateandapproximatetimebyclickingneartime

Page 61: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

On-line Reservations for Confocal Microscope [3]Fill‐inuserinformation.NOTE:theUSERIDwillfillinthePIfieldandappearontheweeklycalendarintheappropriatetimeslot

Page 62: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

Huh?! What?!

Hands-on session to follow...

Page 63: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

• Youmaywanttouseourlaserscanningconfocalmicroscopes

• Toreservetime:– ContactMark([email protected])toscheduleatrainingsession.

– 612‐624‐3454

• ImagingCenterhours9am‐5pmM‐F

Page 64: Welcome to Olympus Fluoview FV500 Training · •Nikon C1si spectral 13 Snyder Hall. Confocal ... C ON D EN SER L EN S OBJEC TIVE ... Read manual and

UseoftheUniversity‐wideImagingUseoftheUniversity‐wideImagingCentersCenters

• Youarewelcometoconsultwithourstaffanduseouropticalimagingcapabilitiesandexpertise.– www.cbs.umn.edu/ic– www.bipl.umn.edu

– www.uic.umn.edu

• Togetstarted:– ContactMarkSanders([email protected])

– 612‐624‐3454

• Imagingfacilitiesarestaffed9am‐5pmM‐F