8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
1/84
Chapter 40 Radiation
Protection Procedures ALARA and Occupational
Exposure
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
2/84
ALARA
ALARA stands for As Low As Reasonably Achievable. It is the basic principle of radiationprotection procedures.
There is much that we can do to keep exposureto the patient and the operator as low aspossible.
The chiropractor is unique in the fact that youcan perform radiography and refer your patientsfor other types of examinations.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
3/84
Occupational Exposures
In radiologic technology, 95% of theoccupational exposure comes fromfluoroscopy and mobile radiography.
Neither would be used in your office so theworst case scenario is that you wouldreceive 5% of the exposure that atechnologist would receive.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
4/84
Occupational Exposures
During radiography, the operator should bebehind a protective barrier. These barriers are usually considered as secondary
barriers so protection would be from tube leakage andscatter from the patient. The tube should never bepointed toward this type barrier.
If the barrier can have the tube angled toward the
barrier. It must be a primary barrier. Staying behind the barrier effectively eliminates thesource of occupational exposure if the shielding isadequate.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
5/84
Occupational Exposures
Medical Imaging Exposures Fluoroscopy: All personnel will wear protective apron.
If extremities get into the beam lead gloves can beworn.
The radiologist will usually be close to the machine duringfluoroscopy so their exposure will be higher than that thetechnologist. Aprons between the Image intensifier andBucky Slot covers reduce radiologist exposure.
The technologist should stand as far away from the table as
possible during the exam and move closed only whennecessary. The radiologist will use short burst of exposure and keep the
exposure time as short as possible. The 5 minute clock timerwill alarm when 5 minutes of fluoroscopy has been used.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
6/84
Occupational Exposures
Medical Imaging Exposures Mobile radiography:
The technologist must wear a lead apron duringmobile plain film or fluoroscopy examinations.
An apron must be assigned to each portablemachine.
The exposure cord for portable radiographicmachines must be 2 meters long to maximizedistance from the tube during exposures.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
7/84
Occupational Exposures
Radiology Ancillary Staff Assuming the rooms are adequately shielded,
the receptionist, file room and darkroom staffshould not receive any occupationalexposure.
Radiology ancillary staff should not be used to
hold patients during radiography.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
8/84
Occupational Radiation Monitoring
Occupational Radiation Monitoring is required if there isany likelihood that an individual will receive more than1/10 of the recommended dose.
With just plain film radiography, monitoring may not be
required as long as the operator stays in the controlbooth during all exposures. There are some exams such as stress views of the ankle
where the operator would be in the room with the patient.If this is done, monitoring would be necessary.
If the operator ever holds a patient monitoring would benecessary.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
9/84
Occupational Radiation Monitoring
Occupational radiation monitoring offersno protection against exposure. It merelyrecords the exposure received.
If needed, find a certified laboratory toprocess the dosimeters.
Types of monitors
Film badges TLD OSL
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
10/84
Film Badges
Film badges have been used since the 1940sand are still used today.
Exposures below 10 mR are not measured on
the film. Filters along with the window in the badge allowestimation of the energy of the exposure.
The must be worn with the proper side to the
front. They are typically worn on the collar so they
would remain outside the lead apron.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
11/84
Film Badges
Advantages Inexpensive Easy to handle and process Reasonably accurate
Disadvantages Can not be reused Sensitive to heat and humidity Must be changes monthly
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
12/84
TLD
TLD has several advantages over filmbadges. Not sensitive to heat or humidity
Measure exposures to 5mR More sensitiveand accurate.
Can be changed quarterly instead of monthly
Disadvantages Cost but changing badges less frequentlythan monthly eliminates cost problem.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
13/84
Optically Stimulated Luminescence
All of the advantages of the TLD over filmbadges plus: Can be re-read to confirm exposure More accurate than TLD
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
14/84
Where to wear the monitor
The whole body badge is typically worn atcollar level so it can be outside the leadapron.
Fetal monitoring badges used duringpregnancy are worn at waist level underthe apron.
Hand or finger TLDs are worn on theextremity.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
15/84
Occupational Radiation Monitoring Reports
State and federal regulations require that theresults of the occupational radiation monitoringprogram be recorded in a precise fashion andmaintained for review.
Specific information is required to be on thereport including current and cumulativeexposure.
Each site of monitoring must be identified
separately. There will also be a control monitor to measurethe background exposure during transport,handling and storage.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
16/84
Occupational Radiation Monitoring Reports
The supplier of the badges must know the typeof radiation for proper calibration of theequipment.
The badges are control are shipped back to thesupplier in a timely manner.
For lost or damaged badges, a health physicistwill estimate the exposure.
The annual exposure is discussed with each
worker and receipt of the information isdocumented. Monthly reports may be posted butcare must be taken with sensitive information.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
17/84
Lead apron used for operator or patientprotection must be the equivalent of 0.5mm oflead.
They must be worn when in a room during theexposure or during fluoroscopy.
Half aprons are effective means to providegonad protection during radiography.
0.25 mm of lead aprons should be avoided asthey only attenuate 66% of the beam at 76 kVp.
Protective Apparel
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
18/84
Aprons used in interventional radiology shouldbe a wrap around type. Thyroid shields may alsobe worn.
Lead gloves can be worn when the hands are inthe beam. Aprons must be stored on specially designed
racks or laid flat on the floor. They are never
folded. Aprons are tested annually for cracks or holes inthe lead, usually by fluoroscopy.
Protective Apparel
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
19/84
During fluoroscopy radiologic technologistshould stand as far as possible from themachine. Standing behind the radiologist offers added
protection. If you must be in the room, position your body
as far away from the primary beam aspossible.
Position
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
20/84
Patient holding
Many patients will find the x-ray examination tobe physically demanding. Some may not becapable of staying in position.
This is a particular challenge for weight-bearingradiography. Mechanical supportive devices arelimited for erect studies.
If you have a radiographic table, the patient maybe examined recumbent. Sponges may be usedas supportive devices.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
21/84
Patient holding
Radiology or office staff should never hold apatient. Family or friends may be called upon toassist the patient.
The person assisting the patient must wear alead apron and if their hands will be in the beamlead gloves.
Position the person as far away from the primary
beam as possible. Since the person holding the patient may be aparent, make sure they are not pregnant.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
22/84
Reducing Unnecessary PatientDose
As a doctor, you have the responsibility todetermine if the radiography is necessary and
justified.
There are more practice guidelines availableevery year to assist in determining if theexamination will yield necessary diagnosticinformation.
There are many examinations that areperformed knowing that they will yield littlehelpful information so they in no way justify thepatient radiation dose.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
23/84
Reducing Unnecessary PatientDose
Check to see if the patient has previousexaminations that may make the newexamination not necessary.
You may be sued if you dont take filmsand the treatment plan fails because youmissed something the films would show.
The yield of information must be greaterthan the risk of radiation exposure.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
24/84
Reducing Unnecessary PatientDose
Routine x-ray examinations should not beperformed.
Used the most accurate tests to confirmor rule out your working diagnosis.
Consider using MR instead of CT
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
25/84
Repeat Examinations
One area of unnecessary patient exposure isrepeated x-ray examinations. Past estimates offrequency has been as high as 10% but theyshould normally not exceed 5%.
Most of the retakes are of the lumbar spine,abdomen and thoracic spine.
Most retake are due to the exposure factors
being incorrect resulting in an over exposed orunder exposed film. Proper measurement areimportant.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
26/84
Repeat Examinations
Positioning errors account for about 25% ofretakes. Proper training and practice is importantto fine tune positioning skills.
Motion causes about 11% of retakes so properpatient communication during the exam isimportant.
But do not be afraid to retake a poor quality
film. If you can not see a problem makes itlikely you will miss it. Poor quality exams arenever justified.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
27/84
Radiographic Technique
Use as high kVp as possible to getadequate contrast and reduce patientexposure.
Collimate the beam to slightly smaller thanfilm size or the area of interest, whicheveris smaller.
Use the fastest-speed screen-filmcombination consistent with the nature ofthe examination.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
28/84
Positioning
When taking films with the patient seated,do not allow the gonad to be in the primarybeam. Position the patient lateral to thebeam.
For female patients turn the patient PA toreduce breast and gonad exposures whenpossible.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
29/84
Patient shielding
Some form of patient shielding should beused on all patients able to reproduce.
All children should have shielding. Pre-menopausal women should be
shielded except when the shield wouldinterfere with the examination.
Men should be shielded beyond 50 years.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
30/84
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
31/84
Patient shielding
Shielding must be used when the gonadslie in or near the useful beam and when itdoes not interfere with obtaining the
required diagnostic information. Accurate placement is extremely important.
Repeated examinations can result formimproper placement of the shield.
Proper patient positioning and collimationshould not be relaxed when gonad shieldsare in use.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
32/84
Ten Commandments of ALARA
1. Understand and apply the cardinal principles ofradiation control: time, distance and shielding.
2. Do not allow familiarity to result in a false
security.3. Never stand in the primary beam.4. Always wear protective apparel when not
behind a protective barrier.5. Always wear a radiation monitor and position itoutside the protective apron at collar level.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
33/84
Ten Commandments of ALARA
6. Never hold a patient during radiographicexaminations. Use mechanical restrainingdevices when possible. Otherwise, use
patients or friends to hold the patient.7. The person holding the patient must wear
protective apron and if possible, gloves.8. Use gonadal protective on all people of
childbearing age when it will not interfere withthe examination.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
34/84
Ten Commandments of ALARA
9. Examinations of the pelvis or lowerabdomen of a pregnant patient should beavoided whenever possible, especiallyduring the first trimester.
10.Always collimate to the smallest field sizeappropriate to the examination.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
35/84
Chapter 31 Quality Control
Two areas of activity are designed toensure the best possible image qualitywith the lowest possible exposure andminimum costs. Quality Assurance deals with people Quality Control deals with instrumentation
and equipment.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
36/84
Chapter 31 Quality Control
Two areas of activity are designed toensure the best possible image qualitywith the lowest possible exposure andminimum costs. Quality Assurance deals with people Quality Control deals with instrumentation
and equipment.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
37/84
Ten Step Approach to Quality Assurance
1. Assign responsibility2. Delineate scope of care
3. Identify aspects of care4. Identify outcomes that effect the aspectsof care.
5. Establish limits of the scope ofassessment.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
38/84
Ten Step Approach to Quality Assurance
6. Collect and organize data.7. Evaluate care when outcomes are
reached.8. Take action to improve care9. Assess and document actions
10.Communicate information toorganization-wide QA Program
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
39/84
QA Projects
Things that QA can evaluate includes Scheduling of patients Instructions given to patients Wait times in the office Interpretation of films Retake analysis Record accuracy
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
40/84
QA Program
Quality Assurance deals with people andprocesses used to complete tasks.
QA involves training and record keeping.
As the owner of the equipment, you will beresponsible for your radiology services.
The State of California Department of
Radiologic Health established theStandards of Good Practice that is thefoundation of QA and QC in radiography.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
41/84
QA and QC Requirements
Degree of requirements vary by state.California and New York have very tightstandards for quality control of theradiographic and processing equipment.
We are required by statue to teach QAand QC in the radiology program. It iscovered in detail in 9 th Quarter. Mytextboook covers QC in detail.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
42/84
Quality Control
An acceptable QC program has threesteps:
1. Acceptance Testing2. Routine performance monitoring3. Maintenance
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
43/84
Acceptance Testing
The x-ray machine, cassettes and filmprocessor or digital system are the largestcapital expense you may experience.
It makes economic sense to make surethat the equipment meets the performancestandards.
It is recommended that a third party suchas a health physicist do the testing.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
44/84
Acceptance Testing
Areas that should be tested include on the x-raymachine: Shielding of Room Focal spot size Calibration of mA, timer or mAs Calibration of kVp Linearity of exposure Beam alignment Grid centering Collimation Filtration (HVL)
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
45/84
Acceptance Testing
Areas that should be tested on the x-raycassettes:
Screen contact Screen speed Light leaks
Light spectrum matching
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
46/84
Acceptance Testing
Areas that should be tested on the x-rayfilm processor:
Developer temperature Replenishment rates Travel time
Water flow Hypo retention
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
47/84
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
48/84
Quality Control
After a major repair, the machine should beretested to ensure that it was repaired properly.
When the testing shows that the machine is not
performing properly, service or preventivemaintenance is required.
Manufactures establish recommendedpreventive service schedules. When these arefollowed many repairs become unnecessary.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
49/84
Radiographic Quality Control
Areas of concern in x-ray machine Focal Spot Size will impact spatial resolution Filtration will impact patient exposure Collimation will impact patient exposure kVp calibration will impact image quality and
exposure.
Exposure timer accuracy will impact imagequality and exposure
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
50/84
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
51/84
Focal Spot Testing
When the machine is installed or the tubeis replaced, the focal spot size should bemeasured. Then annually thereafter.
A pin hole camera, star test pattern or linepair test tool.
As the tube ages, the focal spot tends togrow and spatial resolution is lost.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
52/84
Filtration
The filtration is measured but determiningthe half value layer of the beam at specificexposure levels. Minimum filtration is
2.5mm aluminum. As a tube ages, tungsten will plate the x-ray port and increase filtration. This cancause technique problems. Inadequatefiltration will significantly increase patientexposure.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
53/84
Collimation
If the collimation is misaligned, intendedanatomy can be missed.
It can be tested in many ways from usingpennies to using test patterns.
Misalignment can not exceed 2% of theSID.
It is tested semiannually and after thereplacement of the collimator lamp.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
54/84
kVp Calibration
In diagnostic radiology, any change willimpact patient exposure. A variation ofabout 3% will impact contrast and imagedensity.
Can be tested with filtered ion chambers,filtered photodiodes or even a cassette
with calibrated filters. Tested annually.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
55/84
Exposure Timer Accuracy
The exposure time is the responsibility ofthe operator. It will impact the density ofthe image and spatial resolution.
Tested with an ion chamber, multi-meterinternally or even a spinning top.
Exposure time must be within 5% forexposure times greater than 10 ms and20% less than 10 ms.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
56/84
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
57/84
Exposure Reproducibility
Any exposure using the same factorsshould produce the same level of densityand contrast on the image.
Sequential exposure should bereproducible to within 5%
Can be tested with a rate meter or stepwedge and densitometer.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
58/84
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
59/84
Darkroom and Processing
The development of the image isdependent upon the temperature of thedeveloper, its concentration and how long
the film is in the developer. The film is sensitive to variations in theenvironment and processing from the timeit is manufactured until it is processed.
Darkroom and Processor QC is the keyprocess of Quality Control.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
60/84
Processing
Processor densitometry is performed dailybefore the first patient is exposed.
A sensitometer is used to produce a step
wedge image on the film that is evaluatedwith a densitometer. The densitometer reads the optical
density of the processed image. A digital thermometer is used to test the
chemical temperatures in the processor.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
61/84
Processing
Key densities on the processed film aremeasured and then graphed.
Base plus Fog is measured on an area ofunexposed film to check the darkroomenvironment.
Speed is tested at the level of exposure thatproduces a density of 1.25OD
Contrast is tested at the level that produced adensity of 0.40 OD and one that produced adensity of 2.20.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
62/84
Processing
By monitoring these densities, problemswith film processing can be detectedbefore image quality deteriorates.
In 9 th Quarter we will cover how to performprocessor QC and problem solve.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
63/84
Waste Records
Since used fixer is classified as ahazardous waste material, it is importantto maintain accurate records of usage and
disposal. The extent of records vary by city, countyand state. You are responsible for theproper disposal of the waste. Someregions include developer as hazardouswaste.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
64/84
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
65/84
Silver Recovery
Old radiographic films and repeated filmsare retained for silver recovery. X-rayimages can not be disposed of in normal
trash. They also can not be used to clean theprocessor rollers.
Waste recovery companies will either burnor chemically remove the silver from thefilm.
Performance standards for film
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
66/84
Performance standards for filmprocessor and darkroom equipmentMeasurement Frequency Tolerance CA Tolerance US
Sensitometry Daily B+F: 0.05MD: 0.10 ODContrast 0.10OD
B+F: 0.08 ODMD 0.15 ODContrast 0.15 OD
Safelight Semiannual < 0.05 OD in 2minutes n/a
Darkroom temp Monthly 70F 5 n/a
Darkroomhumidity
Monthly 50% 10% n/a
Developer temp Daily 0.5F 2-3
Replenishment Daily 5%
Transport Annual 3% 3%
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
67/84
Accessory QC
The cassettes and screens are the area ofchief concern. Problems with either willresult in artifacts on the images and
increased retakes. The screens need to be properly cleaned
frequently. Mammography screens are
cleaned daily. California recommendsmonthly cleaning.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
68/84
Accessory QC
Dirty screens produce white artifacts onthe image.
Multiple white artifacts indicate the need to
replace the screens. Screen contact is tested semiannually. A
problem with screens contact will cause aloss of resolution.
As screens age, they loose speed so thisis also tested.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
69/84
Accessory QC
The other important accessory is thegonad protection devices and lead aprons.
Improper care of the apron can result in
cracks and holes in the lead that reducestheir effectiveness. Aprons and shields are tested
semiannually. The easiest was to testthem is with video-fluoroscopy but film canbe used.
Performance Standards for
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
70/84
Performance Standards for Accessories
Measurement Frequency Tolerance CA Tolerance US
Screen contact Semi annual No problemsdetected
No problemsdetected
Aprons/shields Annually No holes No holes
Screen
matching
Annually 0.05 OD for all
cassettes usedScreen cleaning Monthly Semimonthly
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
71/84
Record keeping
All electromechanical devices needperiodic service.
There are three types of maintenance. Scheduled maintenance such as processor
monthly or weekly service. It includesobserving moving parts and lubrication.
Preventive maintenance is planned serviceand replacement of parts at regular intervalsbefore they fail at inopportune times.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
72/84
Record keeping
Non-scheduled maintenance is the worsttype of service because it impacts patientservice. It may also be very expensive.
With proper scheduled and preventivemaintenance, non-scheduled service can beminimized.
All service schedules should meetmanufacture recommendations.
All service should be documented as partof the quality assurance program.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
73/84
Retake Analysis
Required part of a QA program inCalifornia.
Evaluation includes
View repeated Cause of the repeat
Rate of retakes should be less than 5%.
Information can be gathered from the logthat the state mandates for patients beingexposed to radiation.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
74/84
Retake Analysis
Done every three months using arelatively large sample of data to seetrends in: Type of examination being repeated. Reasons for the repeated films. Determine if additional training or review is
needed. Determine if equipment service might be
required.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
75/84
Repeat Analysis
Your patient log can be designed tocapture both films usage and repeatedfilms.
Data is gathered from log for analysis Repeated films can be put into two main
categories: X-ray Personnel Errors Equipment malfunctions
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
76/84
X-ray Personnel Errors
Failure to measure patient. Use of improper technical factors (mAs,
kVp or distance) Incorrect positioning Improper Collimation Improper use of accessories such as
cassettes, grids or filters
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
77/84
X-ray Personnel Errors
Improper handling of exposed orunexposed films.
Failure to clearly communicate to thepatient breathing instructions and toremain still.
Failure to observe patient duringexposure.
X-ray Equipment and Accessory
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
78/84
X ray Equipment and AccessoryFailure or Malfunctions
Inaccurate calibration of kVp and mA. Inaccurate timer calibration Dirty or damaged cassettes Improperly labeled or damaged grids Malfunctioning collimator
Improper film storage or processorfunction.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
79/84
Reasons for Retake Films
Over or under exposure accounts for over50% of retakes nationally.
Errors in positioning (25%) Patient motion 11% Processing errors 6% Wrong view, beam alignment,cassette
screen or grid errors and artifacts.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
80/84
Retake Films by Region
Cervical Spine Exams = 7% Thoracic Spine Exams = 17% Lumbar and Abdominal X-rays = 40% Skull, Chest,Lower Extremities = 15% Majority or retake results in
unnecessary exposure to gonads orblood forming organs of the body.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
81/84
Daily Log Design
Most regulatory agencieswill require a log ofpatients having radiationexposure.
Columns can be added tocapture film usage,repeats, views repeatedand reason.
This data can begathered and analyzed.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
82/84
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
83/84
Retake Analysis
The reason of eachretake is recordedand percentages arecomputed todetermine the overallrate and rate byreason.
Less than 5% is ideal.
8/11/2019 Week 9 C Chapter 40 Radiation Protection, Chapter 31 QC 84
84/84
End of Lecture
Return to Lecture Index
http://localhost/var/www/apps/conversion/tmp/scratch_10/new_page_1.htmhttp://localhost/var/www/apps/conversion/tmp/scratch_10/new_page_1.htm