> #87 1 Abstract—Omnisys Instrument AB has developed a Water Vapor Radiometer for the ALMA project from scratch in 18 months and will deliver 50 units within the next 18 months. The radiometer includes an Optical Relay, Mounting Frame as well as the Radiometer core module. The radiometer core include optics and calibration system as well as a schottky mixer based front- end and a filterbank back-end. These subsystems are supported by a thermal management subsystem, control and communication subsystem as well as a power subsystem. The system and subsystems will be described in the paper. Index Terms—Radiometer I. INTRODUCTION Omnisys Instrument AB has developed a Water Vapor Radiometer for the ALMA project from scratch in 18 months and will deliver 50 units within the next 18 months. The radiometer includes an Optical Relay, Mounting Frame as well as the Radiometer core module. Fig 1. The Water Vapor Radiometer including Mounting Frame and Optical Relay, mounted on the FESS on an ALMA antenna. The size and mass of the system is about a factor three lower than allocated and facilitates manual handing by one man. The radiometer core include optics and calibration system as well as a schottky mixer based front-end and a filterbank back-end. These subsystems are supported by a thermal management subsystem, control and communication subsystem as well as a power subsystem. II. RADIOMETER CORE The optics in the radiometer core is shown below. The beam enters through a window (implemented in a tilted PPA30 sheet) and is reflected at two active mirrors (M2&M3) before the receiver horn and mixer. In between a chopper wheel is operated deflecting the beam on either the cold load or the hot load with the cycle: SKY-COLD LOAD-SKY-HOT LOAD. Fig 2. Optics in the radiometer core. In the RF chain, the signal is received by a corrugated horn, followed by a subharmonic schottky mixer and an LNA. The mixer is provided by Radiometer Physics Gmhb and the LNA Water Vapor Radiometer for ALMA Dr. A. Emrich, S. Andersson, M. Wannerbratt, P. Sobis, Serguei Cherednichenko, D. Runesson, T. Ekebrand, M. Krus, C. Tegnader, U. Krus. Omnisys Instruments AB Gruvgatan 8 SE-42136 V.Frolunda Sweden Email: [email protected] 174
4

Water Vapor Radiometer for ALMA · Water Vapor Radiometer for ALMA Dr. A. Emrich, S. Andersson, ... One is used for bias electronics and thermal regulation and one, as shown below,

Aug 25, 2018

Download

Documents

ngodien
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Water Vapor Radiometer for ALMA · Water Vapor Radiometer for ALMA Dr. A. Emrich, S. Andersson, ... One is used for bias electronics and thermal regulation and one, as shown below,

> #87 1

Abstract—Omnisys Instrument AB has developed a Water

Vapor Radiometer for the ALMA project from scratch in 18 months and will deliver 50 units within the next 18 months. The radiometer includes an Optical Relay, Mounting Frame as well as the Radiometer core module. The radiometer core include optics and calibration system as well as a schottky mixer based front-end and a filterbank back-end. These subsystems are supported by a thermal management subsystem, control and communication subsystem as well as a power subsystem. The system and subsystems will be described in the paper.

Index Terms—Radiometer

I. INTRODUCTION Omnisys Instrument AB has developed a Water Vapor Radiometer for the ALMA project from scratch in 18 months and will deliver 50 units within the next 18 months. The radiometer includes an Optical Relay, Mounting Frame as well as the Radiometer core module.

Fig 1. The Water Vapor Radiometer including Mounting Frame and Optical Relay, mounted on the FESS on an ALMA antenna. The size and mass of the system is about a factor three lower than allocated and facilitates manual handing by one man. The radiometer core include optics and calibration system as well as a schottky mixer based front-end and a filterbank back-end. These subsystems are supported by a thermal management subsystem, control and communication subsystem as well as a power subsystem.

II. RADIOMETER CORE The optics in the radiometer core is shown below. The beam enters through a window (implemented in a tilted PPA30 sheet) and is reflected at two active mirrors (M2&M3) before the receiver horn and mixer. In between a chopper wheel is operated deflecting the beam on either the cold load or the hot load with the cycle: SKY-COLD LOAD-SKY-HOT LOAD.

Fig 2. Optics in the radiometer core. In the RF chain, the signal is received by a corrugated horn, followed by a subharmonic schottky mixer and an LNA. The mixer is provided by Radiometer Physics Gmhb and the LNA

Water Vapor Radiometer for ALMA Dr. A. Emrich, S. Andersson, M. Wannerbratt, P. Sobis, Serguei Cherednichenko, D. Runesson,

T. Ekebrand, M. Krus, C. Tegnader, U. Krus.

Omnisys Instruments AB Gruvgatan 8 SE-42136 V.Frolunda

Sweden Email: [email protected]

174

A
Text Box
W3B
akerr
Text Box
20th International Symposium on Space Terahertz Technology, Charlottesville, 20-22 April 2009
Page 2: Water Vapor Radiometer for ALMA · Water Vapor Radiometer for ALMA Dr. A. Emrich, S. Andersson, ... One is used for bias electronics and thermal regulation and one, as shown below,

> #87 2

is provided by Miteq (custom implementation).

 

x2

x3

DDS

15 .275 GHz

12 5 MHz

HMC53 5LP4

V to F

V to F

V to F

V to F

LVDS

LVDS

LVDS

LVDS

0.5 – 2 GHz

2 – 4.5 GHz

4.5 – 6.5 GHz

6.5 – 8 GHz

HMC 460LC5

HMC 460LC5

HMC 460LC5

HM C460 LC5

HMC46 0LC5HMC 46 0LC5

NBB-310

NBB-310

NBB-310

NBB-310

LO source

MixerLNA

X6Mul

Filterbank back-end

Fig 3. RF chain block diagram. The LNA is followed by a filterbank. The local oscillator (LO) for the mixer is created by a 15.275 GHz synthesizer followed by an active x6 multiplier, both designed and produced by Omnisys. The active multiplier includes a high precision power output control function, enabling accurate LO power setting for each individual schottky mixer. The optimum power level varies with about a 2-3 dB.

HornMixer

LNA

Multiplier

Fig 4. The M3 mirror with the front-end electronics.

III. CALIBRATION LOADS The Cold Load use a Peltier cooler and the Hot load use resistive heaters for accurate temperature control. The nominal operating temperatures are 283 and 353 K respectively and the dew point at the ALMA site is avoided.

Fig 5. The Hot Load with isolation.

The Cold Load is 3 reflection design, while the Hot Load is a 7 reflection design, both implemented in aluminium coated with 1.2 mm CR-110 Eccosorb. The SWR and back scatter has been measured both on lab bench and in the system. Both loads has been simulated in 3D in Solidworks simulation module CosmosWorks, as shown below. The internal thermal differences are within 50 mK.

Fig 6. The Hot Load thermal simulation in CosmosWorks. Several temperature sensors were used in the evaluation of the loads and two are used in each in the production model, to allow for a independent safety temperature monitoring function. A typical temperature variation is shown below and the variation is well below 10 mK.

175

akerr
Text Box
20th International Symposium on Space Terahertz Technology, Charlottesville, 20-22 April 2009
Page 3: Water Vapor Radiometer for ALMA · Water Vapor Radiometer for ALMA Dr. A. Emrich, S. Andersson, ... One is used for bias electronics and thermal regulation and one, as shown below,

> #87 3

‐0.02

‐0.015

‐0.01

‐0.005

0

0.005

0.01

0.015

0.02

0 100 200 300 400 500 600

K diference

Seconds

Temperture  sensor difference

Fig 6. Thermal stability of the Hot Load.

IV. FILTERBANK The back-end consist of a four channel filterbank covering 0.5-8 GHz. It is implemented in one box but in two compartments. One is used for bias electronics and thermal regulation and one, as shown below, for the microwave electronics.

Fig 7. Microwave compartement of the Filterbank back-end.

Several types of microstrip, lumped and commercial filters are used in different configurations in the four channels to allow for sharp response as well as out of band rejection. Differential schottky diode detectors are use with voltage to frequency converters for the final detection. A typical frequency response is shown below for both sidebands.

Response at RF

25

27

29

31

33

35

37

39

41

43

45

47

49

173300 175300 177300 179300 181300 183300 185300 187300 189300 191300 193300

MHz

dB

Ch0Ch1Ch2Ch3

Fig 8. Frequency response of complete WVR.

V. THERMAL REGULATION There has been extensive work in the design, test and optimization of the thermal control and regulation system. The Central Base plate in the radiometer core is regulated by the

fan speed control in the cooling channel, while the calibration loads and the back-end have separate thermal regulation systems.

Fig 9. Simulated and tested step response of a thermal regulation function.

Fig 10. Long term (14 h) thermal stability of the Hot and Cold load.

Typical regulation accuracies are well below 10 mK and the

system is compliant to operating between 16-21 C ambient temperature at 5000 m elevation. Start up from cold is also within specification.

Fig 11. The radiometer installed on an ALMA antenna.

176

akerr
Text Box
20th International Symposium on Space Terahertz Technology, Charlottesville, 20-22 April 2009
Page 4: Water Vapor Radiometer for ALMA · Water Vapor Radiometer for ALMA Dr. A. Emrich, S. Andersson, ... One is used for bias electronics and thermal regulation and one, as shown below,

> #87 4

VI. SYSTEM TESTS The first WVR has been tested at the site at 3000 m

elevation and test data is shown below. Typical mounting and dismounting of the WVR at the antenna is below 30 minutes, due to a clever arrangement of mechanical/optical interface.

0

50

100

150

200

250

300

4100 4200 4300 4400 4500 4600 4700 4800 4900

K

Seconds

WVR first light at ALMA

Tsrc0

Tsrc1

Tsrc2

Tsrc3

Fig 11. First light data from WVR #102 at the ALMA site.

VII. CONCLUSION A new Water Vapor Radiometer has been developed by

Omnisys Instruments for the ALMA project in 18 months. It is compliant to all requirements and adapted for long lifetime operations, limited and simple maintance and reasonable per unit cost. More than 50 units will be produced over the next 18 months.

177

akerr
Text Box
20th International Symposium on Space Terahertz Technology, Charlottesville, 20-22 April 2009