Vinyl Chloride CH2:CHCI No.131L 0.2 0.5 1.0 1.5 2.0 2.5 3.0 ppm n=2 131L 0.02 ppm4 pump strokesYellow Reddish brown Temperature 0 to 40 (32 to 104 ) correction used Relative humidity 0 to 90 % correction not used 10 % for 0.2 to 1 ppm, 5 % for 1 to 3 ppm10 tubes for 5 tests 30 monthsin the refrigeratorDetecting limit : Colour change : Operating conditions : Relative standard deviation : Tube quantity and number of tests per box : Shelf life : Performance 0.1 to 0.2 ppm 0.2 to 3.0 ppm 3.0 to 6.9 ppm 4400 mL2200 mL1100 mL1/2 1 2.3 6 min 3 min 1.5 min Measuring range Number of pump strokes Correction factor Sampling time CH2:CHCl Cr 6 H2SO4 HCI HCl Base Chloride Reaction principle Other substances measurable with this detector tube Allyl chloride Factor : 16 2 3.2 to 48 ppm 1,1,2,2-Tetrachloroethane Factor : 10 2 2 to 30 ppm Substance Correction No. of pump strokes Measuring range Permeation tube method Calibration gas generation When used, these tubes are to be connected. See page 2-3. Possible coexisting substances and their interferences Tetrachloroethylene 1/3 Reddish brown Trichloroethylene 1/5 Benzene, Toluene 200 ppm No Ethylene 200 ppm Substance Concentration Interference Changes colour by itself to Water vapour is trapped in the white layer of the pretreatment tube. Scale
1

Vinyl Chloride CH :CHCI No

May 05, 2022

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Vinyl Chloride CH :CHCI No

Vinyl Chloride CH2:CHCI No.131L

0.2

0.5

1.0

1.5

2.0

2.5

3.0

ppm

n=2

131L

0.02 ppm(4 pump strokes)Yellow → Reddish brownTemperature 0 to 40 ℃ (32 to 104 ℉) correction usedRelative humidity 0 to 90 % correction not used10 %(for 0.2 to 1 ppm), 5 %(for 1 to 3 ppm)10 tubes for 5 tests30 months(in the refrigerator)

Detecting limit :Colour change :Operating conditions :

Relative standard deviation :Tube quantity and number of tests per box :Shelf life :

Performance

0.1 to 0.2 ppm 0.2 to 3.0 ppm 3.0 to 6.9 ppm 4(400 mL) 2(200 mL) 1(100 mL) 1/2 1 2.3 6 min 3 min 1.5 min

Measuring rangeNumber of pump strokesCorrection factorSampling time

CH2:CHCl + Cr6 + + H2SO4 → HCIHCl + Base → Chloride

Reaction principle

Other substances measurable with this detector tube

Allyl chloride Factor : 16 2 3.2 to 48 ppm1,1,2,2-Tetrachloroethane Factor : 10 2 2 to 30 ppm

Substance Correction No. of pump strokes Measuring range

Permeation tube methodCalibration gas generation

When used, these tubes are to be connected. See page 2-3.

Possible coexisting substances and their interferences

Tetrachloroethylene ≧ 1/3 + Reddish brownTrichloroethylene ≧ 1/5 + Benzene, Toluene ≧ 200 ppm − NoEthylene ≧ 200 ppm −

Substance Concentration Interference Changes colour by itself to

Water vapour is trapped in the white layer of the pretreatment tube.

Scale