Page 1
Vein-On-A-Chip: A Microfluidic Platform for Functional
Assessments and Staining of Intact Veins
by
Zhamak Abdi Dezfooli
A thesis submitted in conformity with the requirements
for the degree of Master of Applied Science
Institute of Biomaterials and Biomedical Engineering
University of Toronto
© Copyright by Zhamak Abdi Dezfooli (2015)
Page 2
ii
Vein-On-A-Chip: A Microfluidic Platform for Functional Assessment
and Staining of Intact Veins
Zhamak Abdi Dezfooli
Master of Applied Science
Institute of Biomaterials and Biomedical Engineering
University of Toronto
2015
Abstract
The ability to perform systematic studies on vein function and structure is of key importance for
understanding many vascular diseases and the underlying mechanisms governing transport of
cells and macromolecules through the blood vessel wall, both of which are beneficial for drug
discovery. Available experimental methods are limited by poor optical access and limited
microenvironmental control for in vivo studies, and a partial representation of blood vessel
structure in the case of in vitro approaches. In this thesis, I present a microfluidic platform for the
in vitro investigation of intact mouse mesenteric vein segment (200-300 µm diameter and 1.5-2
mm length) function under near-physiological conditions. The lumen of the pressurized vein
segments are stained on-chip for CD31 to highlight the capacity of our platform for in-situ
immunofluorescence staining without the otherwise required tissue processing and sectioning.
The selected approach is automatable and reduces the time and staining solution consumption by
an order of magnitude compared to traditional methods.
Page 3
iii
Acknowledgments
First and foremost, I would like to express my gratitude to Prof. Axel Guenther for introducing
me to this exciting subject, guiding me towards meeting my goals, and supervising me
throughout the project.
I would also like to thank our collaborator, Prof. Steffen-Sebastian Bolz, for his guidance and
valuable inputs. I am thankful for all the support I received from the members of Bolz lab
particularly, Dr. Firhan Malik for helping me in the development of the on-chip
immunofluorescence staining protocol.
I would like to express my appreciation to my committee members, Prof. Aaron Wheeler, Prof.
Jonathan Rocheleau, Prof. Dan Dumont, and Prof. Myron Cybulsky for their invaluable advice
and scientific guidance and contribution, as well as to Dr. Paul Van Slyke for his helpful
experimental recommendations with regards to working with leukocytes.
I would also like to thank the supervisory committee at The Centre for Microfluidic Systems,
specifically Ms. Lindsey Fiddes for providing technical support in regards to device fabrication
and imaging.
Moreover, my special thanks to all the members of Guenther lab for their suggestions and
support, specifically Dr. Ali Oskooei for helping me in the design of the temperature control
system.
Last but not least, I am very much thankful to my family. To my parents and my sister who have
always believed in me and for their support in every sense of the word throughout my whole life.
I am indebted to my two amazing children for being so patient with me and for supporting me
throughout the years of my studies.
Page 4
iv
Table of Contents
Acknowledgments .......................................................................................................................... iii
Table of Contents ........................................................................................................................... iv
List of Tables ................................................................................................................................. vi
List of Abbreviations .................................................................................................................... vii
List of Figures ................................................................................................................................ ix
List of Appendices .......................................................................................................................... x
Chapter 1 : Introduction .................................................................................................................. 1
1.1 Hypothesis ........................................................................................................................... 2
1.2 Specific Aims ...................................................................................................................... 3
1.3 Background ......................................................................................................................... 3
1.3.1 Immune Function and Selective Barrier Function of Venous System .................... 3
1.3.2 Microfluidic Studies in Blood Vessels ................................................................... 5
1.4 Thesis Structure .................................................................................................................. 7
Chapter 2 : Device Study .............................................................................................................. 8
2.1. Device Design and Fabrication ........................................................................................... 8
2.2. Physical Characterization .................................................................................................... 9
Chapter 3 : Experimental Methods ............................................................................................. 16
3.1. Microfluidic platform ........................................................................................................ 16
3.2. Vein isolation and functional response assessment .......................................................... 18
3.3. Immunofluorescence Staining Procedure ......................................................................... 20
Chapter 4 : Results ...................................................................................................................... 22
4.1. Functional Responses ........................................................................................................ 22
4.2. Immunofluorescence Staining .......................................................................................... 24
Chapter 5 : Discussion ................................................................................................................ 27
Page 5
v
Chapter 6 : Conclusion ................................................................................................................ 31
Chapter 7 : Future Directions ...................................................................................................... 32
References ..................................................................................................................................... 34
Appendices .................................................................................................................................... 40
Page 6
vi
List of Tables
Table 1: Young’s modulus measurement of chip-hosted mouse mesenteric veins at different
pressure ...…………………………………………………………………………………….. 15
Table 2: Protocol for on-chip immunofluorescence staining of intact vein endothelium ……... 21
Page 7
vii
List of Abbreviations
Endothelial cell EC
Smooth muscle cell SMC
Acetylcholine Ach
Phenylephrine PE
Bovine serum albumin BSA
3- (N-morpholino) propanesulfonic acid MOPS
Norepinephrine NE
Leukotrienes LTs
polymorphonuclear leukocytes PMNLs
cluster of differentiation 31 CD31
Tumor necrosis factor TNF-α
Interleukin 8 IL-8
Intercellular adhesion molecule 1 ICAM1
Junctional adhesion molecule JAM
Endothelial cell-selective adhesion molecule ESAM
Platelet/endothelial-cell adhesion molecule 1 PECAM-1
Vascular cell-adhesion molecule 1 VCAM1
Immunoglobulin superfamily Igs
Poly-(dimethylsiloxane) PDMS
Particle Image Velocimetry PIV
Paraformaldehyde PFA
Phosphate Buffered Saline PBS
Page 8
viii
TRITC Tetramethylrhodamine
TxRed Texas Red
Nitric Oxide NO
Lipopolysaccharides LPS
Page 9
ix
List of Figures
Figure 1: A schematic illustration of the leukocyte migration cascade in venules …………… 4
Figure 2: Microfluidic device design …………………………………………………………. 9
Figure 3: Flow characterization within the microfluidic device ……………………………… 11
Figure 4: Demonstration of sheath flow in the microchannel .………………………………... 13
Figure 5: Fixation channel design …………………………………………………………….. 15
Figure 6: Illustration of the experimental setup ………………………………………………. 17
Figure 7: Vein isolation and functional response assessment ………………………………… 19
Figure 8: On-chip functional assessment of mouse mesenteric vein …………………………. 23
Figure 9: Immunofluorescence staining of CD3 ……………………………………………… 25
Figure A1: Experimental setup- photograph and side view illustration ……………………… 40
Figure A2: Illustration of temperature control setup …………………………………………. 42
Figure A3: Dose response data ……………………………………………………………….. 43
Figure A4: Staining data ……………………………………………………………………… 45
Figure A5: Bubble Formation Test ………………………………………………………….... 48
Page 10
x
List of Appendices
A1. Experimental setup ………………………………………………………………………... 40
A2. Temperature control setup…………………………………………………………………. 41
A3. Dose Response Data ………………………………………………………………………. 42
A4. Staining Data ……………………………………………………………………………… 44
A5. Bubble Formation Test ……………………………………………………………………. 47
Page 11
1
Chapter 1 : Introduction
Veins are the blood vessels of the vascular system that are responsible returning blood flow to
the heart. Vein wall consists of three main layers: an outer layer that is called tunica adventitia
and is composed of connective tissue, nutrient vessels and autonomic nerves; a middle layer
called the tunica media is made up of smooth muscle cells (SMCs), elastic fibers and connective
tissue and is the thickest layer that provides structural support, vasoreactivity and elasticity; an
inner layer called the tunica intima with endothelial cells (ECs) and a small amount of sub-
endothelial connective tissue. The diameter of the lumen and the thickness of the wall of venous
blood vessels vary between 3 cm and 1.5 mm (i.e. vena cava) to 20 µm and 2 µm (i.e. venule),
respectively, depending on the position of the vascular bed. Therefore, compared to arteries,
veins have thinner wall and larger inner diameter due to significantly less SMCs and elastic
tissue in the structure of the middle layer- the tunica media.1-3 Veins play significant roles in
various physiological processes, respond to chemical and physical stimuli, synthesize bioactive
substances within the vascular wall, and either promote or prevent the transport of cells and
biomolecules. The soft wall of veins makes them vulnerable to irregular dilation, compression
and penetration by tumour and inflammatory processes. Diapedesis of leukocytes and plasma
macromolecule extravasation during inflammation are two examples of the transport processes
through the vein wall during inflammation.2, 4-6 However, the level of systematic research in the
field remains limited, in part because of experimental challenges related to the work with intact
organs. This is due to the very fragile and fine structure of the vein tissue and susceptibility of
the blood vessel to different external stimuli.
Common in vivo methods such as non-invasive, invasive and in situ cannulation techniques for
studying veins are challenging because of different technical and physiologic related limitations.
The complex interplay between different parameters e.g. signals from surrounding tissues and
the inability to independently alter chemical cues in the vessel microenvironment are limiting
factors concerning in vivo platforms. Also, limited spatial resolution due to high signal-to-noise
ratio when imaging at a distance and through centimetres of tissue in the body, as well as lack of
optical access due to limitation in imaging depth are other technical challenges of in vivo
approaches. For example, acquiring precise images by imaging modalities such as single- and
two-photon techniques usually can only be achieved to a depth of 100 µm from the surface of the
Page 12
2
tissue, while much deeper penetration is needed for in vivo imaging. Another important
limitation is physiologic movement from breathing, cardiac contractions and the pulsatile flow of
blood, which negatively affect the sharpness of images compared to in vitro methods. However,
in vivo techniques are valuable because of the ability to maintain physiological conditions.7-11 On
the other hand, existing assays for in vitro studies of blood vessels for evaluating physiological
functions or drug development often use a confluent layer of isolated endothelial cells (ECs).
These models, to a limited extent, recapitulate important pathophysiological features experienced
by veins in vivo, e.g., correct cell alignment or perfusion at a defined shear rate. Such assays
include perfusion chambers, microfluidic devices and Boyden chambers.12, 13
Current in vitro technologies to study intact vein function and structure, either mounting the
vessel segment on two wires (isometric approach) or cannulation and perfusion with glass
micropipettes (isobaric approach),14-16 require skilled personnel and are not scalable; i.e.,
increasing the throughput proportionally consumes more time and workload. Techniques
available for whole tissue staining require the expertise to perform complex, multi-step
procedures, which consume both time and staining solution. The time and solution consumption
for whole mount fluorescent immunohistochemistry are on average 3-5 days for sample
preparation and on the order of milliliters, respectively, including 2 hours to one day fixation and
1-4 days incubation with the primary antibody.17, 18 Moreover, in vitro staining of the intact
blood vessel lumen is very challenging and difficult to perform using existing methods due to
inaccessibility of the cells. Operating an experiment with conventional methods that involves
functional assessment, staining, and imaging of an intact vein segment requires manipulation of
the specimens between three to four different instruments: the pressure myograph, tissue
processor, microtome (if tissue sectioning is required), and confocal microscope. Hence, an in
vitro approach that is capable of maintaining an intact and functional vein under physiological
conditions and which is also scalable i.e., increasing the throughput does not equivalently
increase the reagent, time and workload consumption, is a promising tool for the investigation of
blood vessel function at the cellular level.
1.1 Hypothesis
I hypothesize that a microfluidic-based approach to pressure myography assures the maintenance
of an isolated intact small vein under physiological environmental conditions, and makes in situ
Page 13
3
functional assessment and immunofluorescence staining of the blood vessel lumen possible with
1 to 5 days reduction in time, 10 fold reduction in solution consumption, and reduced manual
operation than that associated with conventional techniques, and controlled physical and
chemical microenvironment.
1.2 Specific Aims
The following experimental strategy was carried out in order to probe the hypothesis:
i. The design, fabrication and testing of the microfluidic device for hosting and reversibly
fixating an isolated mouse mesenteric vein.
ii. The development of a protocol for the functional assessment of the chip-hosted vein to
apply abluminal flow of vasoactive substances under physiological conditions.
iii. The development of a protocol for immunofluorescence staining of the lumen in a chip-
hosted vein segment.
1.3 Background
1.3.1 Immune Function and Selective Barrier Function of Venous System
The vein wall functions as a barrier between the intra- and extravasal spaces; yet, it is rather
locally selective, and regulated. It is well known that the post-capillary venules involve the
exchange processes of the tissues and contribute in the resorption of interstitial fluid and
transportation of large molecules.19, 20 These vessels are composed of endothelial cells, pericytes
and a basement membrane.21 Studies show that the vascular endothelium permeability to
macromolecules increases and is dependent on the activation of a receptor-mediated
physiological mechanism.22 Inflammatory mediators such as histamine, serotonin, TNF and
bradykinin promote the extravasation of macromolecules from venules.23-25 The flux of
macromolecules through the blood vessel wall is associated with the formation of leakage sites in
the post-capillary venules.4
Furthermore, venules and post-capillary venules regulate the diapedesis of polymorphonuclear
leukocytes (PMNLs) and are the primary sites for leukocytes extravasation.26 Studies have
shown that there are specific locations in the basement membrane of venules that are closely
associated with gaps between pericytes and contain low extracellular matrix protein. These sites
are preferred pathway for transmigrating leukocytes during their passage through venules.21, 26, 27
Page 14
4
Leukocyte migrate through a multiple cascade paths, mediated by fluid shear stress along with
chemical cues, and consists of the following steps (Fig. 1): The first step is referred to as
margination. Interaction between leukocytes and the large number of disk-shaped erythrocytes
that are contained in the complex fluid blood traveling through post-capillary venules initiates
the circulating leukocytes to migrate from the venule centre towards the wall, thereby initiating
surface interaction between leukocytes and endothelial cells. The interactions initially take place
via selectins (i.e., E- selectin, P-selectin and L-selectin) and their counter-ligands and are strong
enough to counter lift forces but too weak to firmly attach the leucocytes to the endothelial layer
in the presence of fluid shear. Leukocyte rolling is then followed by firm adhesion to the
endothelium. The final stage is the transendothelial migration which is either paracellular or
transcellular. The paracellular migration is through intraendothelial junctions. Transcellular
migration is through transcytotic pathways in the continuous endothelium. These pathways are
responsible for the transport of cells and macromolecules across the endothelial barrier by a
vectorial transport mechanism into the inflamed tissue.28-31
Figure 1: A schematic illustration of the leukocyte migration cascade in venules. The migration
process consists of margination, rolling, adhesion and transendothelial migration. The adhesion
comprises of attachment and intravascular crawling and transendothelial migration is done
through the venule wall and their respective molecular components involved.
Page 15
5
1.3.2 Microfluidic Studies in Blood Vessels
In the past decade, several microfluidic tools have been developed for the study of different
aspects of blood vessel physiological function.32-34 In comparison to conventional cell and tissue
culture methods, these platforms and assays have a higher throughput with a smaller sample and
reagent consumption. These methods also provide precise control over the physical and chemical
characteristics that are relevant to vascular function, such as flow, shear stress, and reagent
concentration of the microenvironment. They assure better optical access, as well as more
physiological and pathological relevant conditions.
Most of the reported approaches are based on the mono- or co-culture of EC on either a stiff
microchannel wall or an engineered gel within a microfluidic device. Khan et al. studied the
inflammatory state of endothelium by assessing the effects of disturbed flow on ECs. To
generate disturbed flow, they designed a microfluidic chip which consisted of patterned channels
lined with ECs. Under constant flow, altering the geometry of channels will change the shear
stress applied on the ECs, which resulted in a change in activation and expression of cell
adhesion molecules such as VCAM-1 and ICAM-1 on the endothelial cell layer. They also
quantified the adhesiveness of the ECs to leukocytes by measuring the number of adherent
cells.35 Han et al. reported a platform that simulated the 3D configuration of neutrophil
transmigration during inflammation by placing an extracellular matrix with collagen type I in a
microfluidic device. Using a chemo-attractant gradient in this microfluidic device, they
quantitatively visualized the effect of chemoattractants e.g., IL-8 and fMLP on neutrophil
transmigration through a layer of cultured ECs. The researchers noted that the number of
migrated neutrophils and the distance they travelled through the collagen gel decreased in the
absence of intact endothelium, despite the presence of an optimal fMLP concentration gradient.
They also observed that the number of migrated neutrophils showed a reverse correlation with
the stiffness of the type I collagen.36 Schaff et al. studied the effect of shear stress on the
interaction of neutrophils and cultured ECs in a microfluidic platform, and measured neutrophil
rolling velocity and the adhesion rate under a defined shear stresses. Using their platform, they
confirmed that the rate of the neutrophil recruitment to the substrate was very sensitive to
disturbances in flow streamlines as it was increased at the entrance of the microchannels. They
also showed that dose response of stimulation of the ECs by the chemokine IL-8 can trigger β2
integrin activation.37 Zhang et al. developed an in vitro vascular system that recapitulated the
Page 16
6
geometry of native vasculature with a circular microchannel. This was used to investigate the
relationship between shear stress applied on the ECs cultured in the channels and dose of the
applied pharmacological agent. They validated their platform for drug screening by examining
the effects of four vasoactive drugs on the secretion of nitric oxide from the endothelial cells
under physiological conditions: acetylcholine, phenylephrine, atorvastatin, and sildenafil. This
study demonstrated the importance of physiologically-realistic endothelial cell geometry and
flow rates in drug validation applications. The researchers also demonstrated a monocyte
adhesion assay, which could be used for the assessment of efficiency of anti-inflammatory
drugs.38 Namdee et al. explored how particle size, along with hemodynamics (blood shear rate
and vessel size) and hemorheology (blood hematocrit) affect the capability of spheres to
marginate (localize and adhere) to inflamed endothelium in a microfluidic model of human
microvessels. They used cell free plasma from whole blood and showed that microspheres
present excessively higher margination than nanospheres in all hemodynamic conditions.39 Tsai
et al. reported a microfluidic vasculature model that recapitulate platelets, leukocytes, and
endothelial cells activation under controlled flow conditions. The microfluidic device enabled
quantitative study on how biophysical alteration, for instance shear stress, in hematologic
diseases such as sickle cell disease (SCD) and hemolytic uremic syndrome (HUS), cause
vasculature occlusion and thrombosis. It consisted of bifurcated microchannels that were
luminally covered with monolayer of endothelial cells that resembled a microvasculature system.
The researchers used the model for studying how the drug hydroxyurea quantitatively affects
microvascular obstruction in SCD by applying blood sample from patient with SCD to the in
vitro microvasculature and investigating the effect of an antiplatelet drug on aggregation of
platelets with changing shear. In addition, they used their microsystem as an in vitro model of
HUS and showed that shear stress influences microvascular thrombosis/obstruction and assessed
the effectiveness of the drug eptifibatide, which decreases platelet aggregation in patients with
HUS.40
All of the preceding studies were based on the culture of EC monolayers on either a stiff
microchannel wall, or a gel surface within a device in the absence of other components of the
blood vessel wall i.e. perivascular basement membrane and the connective tissues. Hence, the
cultured monolayers cannot form physiologically relevant levels of junctional complexes.
Page 17
7
Compared to cell culture-based platforms, tissue and organ-based microfluidic approaches have
also been reported that have better representation of in vivo conditions.
Our research group previously developed a microfluidic platform for the assessment of isolated
resistance artery structure and function under physiological conditions, 37 °C temperature and 45
mmHg transmural pressure.41 Small arteries were isolated from a mouse mesentery, loaded onto
the microfluidic device, and immobilized by applying sub-atmospheric pressure on the blood
vessel wall through microfluidic channels. These arteries were kept intact in well-defined
microenvironments that reflected the in vivo conditions of terminal arteries. This platform
isolated the luminal and abluminal surfaces of the blood vessel, which allowed for localized
stimulation of the blood vessel. This capability was validated by dose-response measurements to
the abluminal application of phenylephrine and acetylcholine, which matched the results from
conventional myography techniques. An automated microfluidic platform has further been
developed by our research group that is capable of on-chip staining of smooth muscle cells and
functional assessment of cerebral resistance arteries under physiological environment
conditions.42 Using this platform, the abluminal structure of the blood vessel was visualised by
imaging the position of the smooth muscle cell nuclei, actin filaments and voltage gated calcium
channels. The smooth muscle cell function and intracellular calcium were assessed
simultaneously by staining the SMCs with FURA-2AM during exposure to varying
concentration of PE (a vasoactive mediator) with and without a calcium blocker incubation,
which provided a better understanding of vasomotor dynamics and calcium response.
1.4 Thesis Structure
This thesis is organized as follows: parametric study on the microfluidic device is carried out in
the Chapter 2. Chapter 3 conveys the experimental methods employed, including the chip design
and experimental setup, the device characterization and the protocols for functional assessment
and immunofluorescence staining. Chapter 4 presents the obtained results from the experimental
strategy. Chapter 5 communicates the discussion on the approach and the results. Finally,
Chapter 6 and 7 summarize the findings and suggest potential applications of the described
approach, respectively.
Page 18
8
Chapter 2 : Device Study
2.1. Device Design and Fabrication
The design of the microfluidic device is a modified version of the devices previously introduced
by our group for the functional assessment of mouse mesenteric and cerebral resistance
arteries.41, 42 The design was prepared using a computer aided design program (Autocad 2011,
San Rafael, CA, U.S.A), which consisted of a loading well and four microchannel networks as
follows (Fig. 1): a network of fixation channels (indicated in blue color) which held the vein
segment in the desired position within the inspection region. The second channel network
(indicated in yellow and green and color) comprised two lines, including one core channel and
two side channels that merged upstream of the microchannel and connected to the blood vessel
luminal side. The inlet of the core-channel of the perfusion line (yellow) was connected to a
single reservoir to facilitate the blood vessel loading procedure. The side-channels inlet (green)
was connected via a selector valve to one of six individual reservoirs and allowed for the
controlled luminal perfusion of different solutions. Besides, the two channels were designed to
produce co-flow of two different solutions using flow focusing technique. This facilitates further
application of the platform in investigation of the leukocyte-ECs interaction by flowing the cells
through the side channels and positioning them at the vicinity of the endothelium. The third
microchannel network, superfusion channels (indicated in red color), defined the vein
microenvironment on the abluminal side and allowed physiological buffer, vasoactive drugs and
inflammatory mediators to be superfused on two opposing sides, for application of different
concentrations. Finally, a vacuum line was connected to two previously reported bubble traps43
that were located in the superfusion channels which prevented bubbles from getting through the
inspection area and damaging the blood vessel segment. All channels were 150 µm deep and
smaller than the diameter of the pressurized and unconstricted vein to prevent the blood vessel
segment from passing through the inspection area during loading. The width of the different
microchannels varied between 30 µm at the four fixation locations and 500 µm at the fixation
and superfusion channels. The microfluidic device was fabricated in PDMS
(polythimethysiloxane), using standard single-layer soft lithography techniques, and permanently
sealed on a microscope glass slide (thickness 1mm).
Page 19
9
Figure 2: Microfluidic device design. The device comprised a loading well and four sets of
microchannel networks: fixation channels presented in blue for holding the vein segment in the
inspection area, perfusion channels with two inlets indicated in yellow and green merge upstream
of the microchannel that connect to the blood vessel luminal side and the loading well,
superfusion channels (colored in red) that had two separate pairs of inlets and are connected to
one outlet with two bubble traps on each line, and one vacuum line for the two bubble traps.
2.2. Physical Characterization
To experimentally examine the shear stress exerted on the walls in our system, flow velocity in
different regions of the inspection area was measured by micro- particle image velocimetry
(micro-PIV). A PDMS device was designed and fabricated to mimic the geometry of the original
microfluidic device with a feature of pressurized blood vessel on the perfusion line and two
inlets for the side flow and one inlet for the core flow (Fig. 2A). The micro-PIV measurement
was done by flowing 1.1 µm fluorescent particles through the core channel of the perfusion line
from a reservoir connected to the inlet. The two side channels were blocked and the inlet
pressure was maintained at 4.5 mmHg, using hydrostatic pressure to represent the transmural
pressure in the vein-on-chip microfluidic platform. The device outlet was connected to a syringe
pump in withdrawal mode with a constant flow rate of 1 mL/hr (3×10-4 cm3/s). Fluorescent
particles were illuminated by a laser (λ=550 nm) and visualized by an inverted microscope with
20× objective (field of view: 435µm×325µm). Images were recorded at a rate of 5 frames per
second.
Page 20
10
In order to determine the flow parameters in the perfusion channel, we assumed that the flow
was fully developed. To justify this assumption, the hydrodynamic entrance length Ln for the
flow through the perfusion channel was calculated according to the equation (1):44
(1)
In the part of the channel before the expanded section with hydraulic diameter (Dh) =168 µm and
the Reynold’s number (Re) = 2.5, Lh was equal to 414 µm. This shows that the flow was fully
developed when it passed through the 22.5 mm long channel. In addition, according to equation
(1) hydrodynamic entrance length for the expanded section was equal to 373 µm, while the
length of the section was 550 µm. This number could vary between 420 µm (i.e. Dh = 150 µm)
and 383 µm (i.e. Dh = 194 µm) for flow through the chip-hosted vein, depending on the size of
the blood vessel which influences the width of the expansion in the blood vessel due to the
applied transmural pressure. Therefore the flow gradually became fully developed after half way
through the section. For simplicity, this divergence was not considered in the analytical solution
and calculation of the shear stress. Assuming fully developed flow, i.e., a parabolic velocity
distribution, the cross sectionally averaged velocity obtained from micro-PIV measurements in
the channel at the expanded section which was representative of a pressurized blood vessel 300
µm wide was recorded to be 6.5 mm/s (Fig. 2B). This number is very close to the result from
analytical calculations which was 6.2 mm/s. However, the difference between the two parabolas
in the figure, is as a result of the channel width extension from 200 µm to 300 µm and the
variation in the entrance length, as discussed above.
The shear stress applied to the walls of a micro-channel with rectangular cross-section due to a
fluid flow was calculated by .45 In this equation, (dyne/cm2) is the fluid viscosity, Q
(cm3/s) is the flow rate, and w (cm) and h (cm) are the width and height of the blood vessel in the
inspection area, respectively. Considering a fluid viscosity of 0.007 dyne s/cm2 for the buffer, the
mentioned velocity caused shear stress of about 1.6 dyne/cm2. This number falls in the minimum
range of shear stress applied on the wall of blood vessel in an inflamed tissue.46, 47
Page 21
11
Figure 3: Flow characterization. (A) Schematic of the experimental setup for micro-PIV
measurement. The pressurized blood vessel-feature is magnified. (B) Analytical and
experimental velocity profile resulted from the micro-PIV measurement.
Moreover, the physical parameters of the flow-focusing geometry employed in the design of the
perfusion channel were characterized as follows: the same microfluidic device used for flow
characterization in the previous experimental setup was used. Pure 3- (N-morpholino)
propanesulfonic acid (MOPS) was perfused through the core channel of the perfusion inlet at a
flow rate of QB via a syringe pump. Fluorescent buffer (rhodamine in MOPS) was flowed
through the two side channels of perfusion inlet from a reservoir. Because the two channels were
identical, the flow rate in each channel was half the flow rate of the reservoir, QL. Outlet of the
perfusion channel was connected to a syringe pump in withdrawal mode and the flow rate was
set at Q, which was also equal to the total rate of QL+QB. Therefore, altering the core channel
flow rate, QB, resulted in a change in the side channels flow rates, QL. A relationship between the
relative flow rates, QL/Q, and relative thickness, δ/w (where δ and w are the side flow width and
the channel width at the expanded area, respectively, as are shown in Fig 3A) can be derived
from the parabolic velocity profile:
32
46
wwQ
QL (2)
Equation (2) indicates that altering the core channel flow rate (QB) results in a change in the
thickness of the side flow solution, δ. Fluorescent images were taken while increasing the flow
rate QB to experimentally approve the control over thickness of the side flows, and the resulting
Page 22
12
data was plotted (Fig. 3A and 3B). Experimental results from varying relative thickness (δ/w) vs.
relative flow rate (QL/Q) was in agreement with theory, as shown in Fig. 3C. This characteristic
gives rise to the ability of controlling the concentration of the side flow solutions. Also,
thicknesses of the fluid in the side channels (δ) were measured for different core channel flow
rates (QB) which confirmed the consistency between the top layer and the bottom layer (Fig. 3D).
As shown in Fig 3B, the velocity gradient close to the wall of the microchannel is similar to the
one in the dimension of small venules, thus the shear stress applied to the wall by the fluid flow
is the same as in vivo condition in small venules. All of the above mentioned flow characteristics
of the device support further application of the platform as a tool for studying leukocyte
migration.
Page 24
14
Figure 4: Sheath flow. (A) Fluorescence imaging of the channel with a feature of pressurised
blood vessel in the middle that shows the control of the thickness of cell populated side regions.
(B) Velocity profile in pressurized blood vessel shows the velocity gradient in the vicinity of the
vessel wall is the same in both vein and postcapillary venule. (C) Experimental results of
changes in relative thickness (δ/w) vs. relative flow rate (QL/Q) was in agreement with the
analytical prediction. (D) Thickness of the fluid in side channels (δ) were measured for different
core channel flow rates (QB) and showed consistency between the top layer and the bottom layer.
Furthermore, the design of the fixation channels was justified for veins by calculating the width
of the channels from Young’s modulus ( E ) based on half-space model as follows:48
(3)
where hD is hydraulic diameter of the fixation channels, P is the difference between the luminal
and the fixation pressures, AL is the length of the blood vessel wall that penetrates into the
fixation channel after applying pressure to the channel and is a dimensionless constant
parameter (wall function) which is set to 2.1.48
A vein segment was loaded onto the chip device (loading procedure will be described in the next
chapter) and the pressure applied to the fixation channels was incrementally changed and the
bright field images were recorded. In Fig. 5, the images labelled (a) to (d) show the increased
aspiration length (i.e. length of the blood vessel wall entered into the fixation channels) by
increasing the fixation pressure from 0 mmHg to 22 mmHg. The images were processed and the
Young’s modulus was calculated according to equation (3); The results shown in Table 1 are in
agreement with the data obtained literature.49 Then the fixation pressure was set at 22 mmHg and
the transmural pressure was incrementally changed by increasing the height of the hydrostatic
head (Fig 5, images e and f). Circumferential elastic modulus (Eθ) of the blood vessel wall was
calculated by measuring the inner and outer diameter according to the following equation:50
)(
)1(222
00
2
0
2
i
ii
RRR
RRPE
(4)
where ΔPi is the transmural pressure increment, R0 and Ri are the external and internal,
respectively, ΔR0 is the change in external diameter due to the change in transmural pressure,
and δ is the Poisson ratio which is 0.45 for isotropic elastic material. The circumferential elastic
modulus of the vein segment was calculated based on equation (4) and was 3.7 kPa and 5 kPa for
4.5 mmHg and 11 mmHg change in transmural pressure, respectively.
Page 25
15
Figure 5: Fixation channel design. Images labelled as (a) to (d) show unpressurized vein
segment for fixation pressure from 0 mmHg to 22 mmHg, respectively. The aspired section of
the blood vessel wall is marked in red. Images (e) and (f) indicate pressurized vessel at 4.5
mmHg and 11 mmHg transmural pressure, respectively. Scale bar represents 200 µm.
Table 2: Young’s modulus measurement of chip-hosted mouse mesenteric veins at different
pressure (n=3)
The angle of the fixation channels applied a tension to the wall of the vein segments to mimic the
in vivo tension on the blood vessel wall. Moreover, veins are less stiff than arteries because their
walls are covered with less SMCs, therefore they tend to collapse easily and it is challenging to
keep the lumen open. The direction of the angles in the design of the fixation channels helped to
overcome this problem by pulling the tissue towards the two ends of the vessel segment at each
side.
Ffix
(mmHg)
E (kPa) Ref. #6
E
(kPa)
7.4 3.9±0.6 4.8±0.7
14.7 11.1±0.8
22.1 18.1±0.9
Page 26
16
Chapter 3 : Experimental Methods
3.1. Microfluidic platform
The experimental platform for on-chip functional assessment and immunofluorescence staining
of vein comprises of the previousely described microfluidic device which was connected to five
computer controled syringe pumps (Harvard Apparatus, Holliston, MA) four of which were
connected to the superfusion inlets and run in perfusion mode for delivery of different
concentration of drugs and physiological buffers to the abluminal side of the blood vessel and the
remaining one was connected to the perfusion outlet and operated in withdrawal mode. The inlet
of the perfusion side channels was connected to a computer controlled selector valve (VICI
Valco Instruments Canada Corporation, H-C25Z-3186EUHA) and the other perfusion inlet (the
core channel) was connected to a reservoir. The superfusion channel outlet and the fixation
channel inlets were connected to three open, MOPS filled reservoirs. The height difference
between the two reservoirs would define the fixation pressure for holding the blood vessel. The
two bubble traps in the superfusion lines were connected to a miniaturized vacuum pump
(Parker’s CTS Micro Diaphragm Pump, USA) via the manifold. A stereo microscope (model:
Nikon) and an inverted fluorescence microscope (model Ti Eclipse, Nikon, Japan) were
connected to a CCD camera (model EXi Blue, QImaging, Surrey, BC, Canada) for blood vessel
loading and imaging, respectively. An automated temperature control system maintained the
desired temperature during different stages. A peristaltic pump (Fisher Scientific 13-875-410
DIG. FH100 PUMP 14-4000ML/MIN EA) was employed as a heat sink for the cooling system
(details in Appendices A1). A commercially available manifold with ten fluidic inputs/outputs to
interface with the PDMS device by means of luerlock connectors (Idex Health & Science, Oak
Harbor, WA) was employed to connect the fluidic inlets and outlets with tubing.
Page 27
17
Figure 6: Experimental setup. Schematic illustration of the vein-on-a chip platform consisting
of a microfluidic device inserted into the manifold (1), five computer controlled syringe pump
four of which are connected to the superfusion inlets and one is connected to the perfusion outlet
(2), a computer controlled selector valve connected to the perfusion inlet (3), an automated
temperature control system (4) and a vacuum system (5).
The physiological pressure and temperature were provided as previously described.41 Briefly, the
temperature was preserved using a thermoelectric element (model TE-35-0.6-1.0, TE
Technology) attached to a sapphire glass which was placed on top of the device, where the blood
vessel segment was positioned. A thermistor (model MP-2444, TE Technology) was embedded
into the microfluidic device at the close vicinity of the vein segment during the PDMS moulding
that allowed for on-chip control of the temperature in the inspection area. In order to achieve a
cold temperature (4˚C) for the purpose of immunofluorescent staining, an aluminium cooling
jacket was designed and attached to the TE element to operate as a heat sink to cool down the
warm side of the TE element. Details of the set up for maintaining a target temperature are
illustrated in Appendix A2. Physiological transmural pressure of 4.5 mmHg for small veins51
was maintained by applying hydrostatic head at the both perfusion inlets (Fig. 5).
Page 28
18
3.2. Vein isolation and functional response assessment
For functional assessment, vein segments were isolated from 4-8 month old C57BL/6 mice. A
segment (approximately 1-2mm in length) of mouse mesenteric vein (diameter approximately
200-300 μm) was isolated by removing the adhering fat and connective tissue (Fig. 6A and 6B).
The order of the mesenteric veins refers to the bifurcating branches inside the mesentery. The
major vein supplying the mesentery is the first-order vein and subsequent branches accordingly
increase in their order.52 The isolated vein segment was reversibly mounted onto the microfluidic
device as previously described by Guenther et al.41 In summary, the vessel was placed in the
loading well and was drawn through the perfusion channel to the inspection area where it was
reversibly immobilized at eight points using negative hydrostatic head, which were set at 22
mmHg. While maintaining physiological temperature (37C) and transmural pressure of 4.5
mmHg assumed for small vein, viability of the blood vessel’s structural component was
confirmed by dose response measurements performed with vasoactive substrates (Fig 6C).
Intactness of the smooth muscle cells (SMCs) was confirmed through contractile responses in the
presence of increasing concentration (0.5, 2.5, 5, 10µM) of phenylephrine (PE) that is superfused
at a constant flow rate of 1 mL/hr. Subsequently, the endothelial cells (ECs) were assessed
through the dilatory response of the pre-constricted vessel in the presence of increasing
concentrations (0.25- 1 µM) of acetylcholine (Ach). Bright field images of the vein segments
were obtained using a CCD camera (1392 × 1040 pixel resolution, model EXi Blue, QImaging,
Surrey, BC, Canada).
Page 29
19
Figure 7: Vein isolation and functional response assessment. (A) Illustration of the order of
branching microvessels through schematic and bright-field micrograph of mouse mesentery.
Roman Numbers in the right image indicate the blood vessel order. Scale bar 1 mm (B) Bright-
Page 30
20
field micrograph illustrating mesenteric vein isolation. Scale bar 250 µm (C) Schematic of the
experimental setup for dose response assessment. (D) Schematically illustration of the top view
(middle image) and the vertical cross section (bottom) and the horizontal cross section (right) of
the chip hosted, pressurized vessel.
3.3. Immunofluorescence Staining Procedure
To examine the microfluidic device functionality for luminal staining, immunofluorescence
staining for CD31 on endothelium was performed. In this set of experiments, six reservoirs were
connected to the perfusion channel. Via a selector valve, the reservoirs opened to the perfusion
channel one at a time, for the purpose of executing immunofluorescence staining protocols on
the vein segments. After loading the vein segments onto the chip, the vessels were perfused and
superfused with MOPS at a constant flow rate of 4 µL/min for 2 hours at 37 °C to avoid any
probable stimulation related to isolation or loading. Then the chip was cooled down to 4 °C and
vein segments were fixed by perfusion of 4% paraformaldehyde (PFA) for 30 minutes, blocked
with 3% bovine serum albumin (BSA) in phosphate-buffered saline (PBS) for 1 hour and
incubated with 1:100 rat anti-mouse CD31 antibody (BD Biosciences) over night. All the steps
were performed with a flow rate of 4μl/min, except for incubation with primary antibodies,
where the flow was stopped after 5 minutes perfusion of the antibody. The temperature was
maintained at 4 °C during all the steps. Then the veins were washed with perfusion of PBS for 30
minutes and incubated with 1:200 goat anti-rat TRITC (in some cases TxRed) for one hour at
room temperature. Finally the veins were individually imaged using a confocal microscope
(model Ti Eclipse, Nikon, Japan) with magnification 10× (working distance: 4 mm, numerical
aperture: 0.45); and magnification 40× (working distance: 3.6 mm, numerical aperture: 0.6). To
obtain control data, a blood vessel segment was perfused with the secondary antibody for one
hour right after the blocking and washed with PBS for 10 min before imaging. Summary of on-
chip staining procedure is presented through table 2.
Page 31
21
Table 2: Protocol for on-chip immunofluorescence staining of intact vein endothelium
Step Solution Time Temperature
(° C)
Perfusion
Flow rate
(µl/min)
Total
Volume
(µl)
Fixing 4% PFA in
PBS
30 min 4 4 137
Blocking 3% BSA 60 min 4 4 257
Primary antibody 1:100 rat anti-
mouse CD31
antibody
12 hrs 4 4 and 0 100
Wash PBS 30 min 25 4 137
Secondary antibody 1:200 goat
anti-rat
TRITC
60 min 25 2 137
Wash PBS 30 min 25 4 137
Total time 15.5 hrs
Page 32
22
Chapter 4 : Results
4.1. Functional Responses
After loading the vein segment onto the microfluidic chip and heating the vessel to 37 °C and
pressurizing to 4 mmHg, the functional response of the SMC and EC layers was verified through
superfusion of varying concentration of PE and ACh respectively.
Figure 8 A and C, respectively, represent images from chip-hosted vein segments, abluminally
exposed to PE and Ach through the superfusion channels at the top and the bottom. Summary of
the results from the vein functional response to PE (n=5) and Ach (n=4) are shown in Fig. 8 B
and D. Captured images were then manually processed using ImageJ (ImageJ 1.48v, U.S.A.) to
quantify the contractile responses as explained in the discussion section.
Page 33
23
Figure 8: On-chip functional assessment of mouse mesenteric vein. Representative images of
vein functional response to (A) PE and (C) Ach. (B) and (D) Dose response plots to PE (n=5)
and Ach (n=4), respectively. Scale bars are 200 µm. Error bars are standard error.
Page 34
24
4.2. Immunofluorescence Staining
After completing the staining procedure, the manifold was detached from the fluidic connections
and transferred to the confocal microscope to picture the expression of CD31 on the ECs of the
blood vessel walls. Figure 9A shows a representative confocal image of intact whole vein
segment luminally stained for CD31. The intensity of the “top” and “bottom” walls of the vein
segment was measured from the orthographic projection of the confocal and plotted for four
stained sample and one control sample (Fig 9 B and C) for quantifying the data.
Page 35
25
Figure 9: Immunofluorescence staining of CD31. (A) Confocal image of an intact vein
segment stained luminally for CD31 with 10× (bottom) and 40× (top, inset), respectively.
Arrows indicate the endothelial cell boundaries. Scale bars are 50μm (top) and 100μm (bottom).
Page 36
26
(B) and (C) Orthographic projection and the respective intensity profiles of the confocal images
of stained (scale bar is 50 µm) and control (scale bar is 100 µm) samples obtained with 10×
objective. (D) Scattered plot of the two maxima obtained from the intensity profiles.
Page 37
27
Chapter 5 : Discussion
The main advancement achieved with the presented platform is to keep an isolated small mouse
mesenteric vein segment alive and intact under well-controlled physical (including the
temperature, transmural pressure and perfusion flow rate) and chemical microenvironment
conditions, for functional assessment and immunofluorescence staining of the lumen.
Figures 8A-D show the dose response results from the chip-hosted vein segments, abluminally
exposed to phenylephrine (PE) and acetylcholine (Ach) through the superfusion channels at the
top and the bottom, respectively. Bright field images of the samples (Fig A3) indicated that the
contractile responses of mouse mesenteric veins to vasoactive drugs were different from the
contractile responses of arteries from the same vascular bed. Unlike the sample of arteries,41 in
our results vein wall did not thicken after exposure to PE, but collapsed. This is probably because
of the differences in the wall structure of the two types of blood vessel, i.e. the amount of the
SMC layers around the entire vein is much less than that of the arteries which affects the
vasoconstriction.1, 53 Therefore, it was not possible to differentiate between the inner and outer
borders of the blood vessel wall and only the outer lines of the vein wall were counted for
quantifying the functional responses. Also, as described in the experimental method section for
functional assessment, the isolated vein segments (diameter approximately 200-300 µm) were
loaded onto the microfluidic chip and sandwiched within the PDMS channel with 150 µm height
and the glass slide and the cross-section of the samples was imaged from bottom as
schematically illustrated in the Fig 7 D. Moreover, confocal images of the constricted vein
segments showed that not the entire vascular wall but the sides that were exposed to the
vasoactive drugs exhibit functional responses. Therefore, change in the width of the blood vessel
segments which represented the distance between the side walls (denoted by D), in response to
different concentrations of the vasoactive drugs, was used as a measure of percentage
constriction and dilation. In order to measure D (or D0) the images were post processed in
ImageJ by drawing a line that connected the top and bottom side of the vessel, as illustrated in
the figure. The constriction and dilation percentage of vessels were calculated by normalizing D
to the maximum and minimum vessel width (denoted by D0) i.e. before applying the constrictor
and dilatory mediators, respectively. Therefore the contraction percentage is estimated as (D0-
Page 38
28
D)/D0. In our experiments, all the functional veins showed contractile response to PE and Ach
with the highest constriction at 10 µM and highest dilation at 1 µM, respectively.
Phenylephrine is known for its effect on SMCs to induce vessel contraction by binding to
adrenergic receptors on the SMCs. This results in the buildup of internal calcium which causes
increased tension within the cells around the intact blood vessel, leading to the entire vessel
constricting.54, 55 The results from on-chip measurements showed an increasing trend in the
contraction of the vein walls in response to PE. The maximum constriction was about 72% which
was achieved by superfusion of 10 µM PE. The maximum constriction is commensurate with
other reported data for rat mesenteric small veins, which were produced using pressure
myography technique, in response to 10-4 M norepinephrine at 2 mmHg and 6 mmHg transmural
pressure, i.e. ~85% and ~70%, respectively.52, 56
Acetylcholine has effects on stimulating intracellular nitric oxide (NO) production and release in
endothelial cells during blood vessel relaxation.57 Chip-hosted veins showed weak responses to
low concentrations of Ach and the maximum dilation was about 40% in response to 1 µM. This
number is less than the reported results from on-chip dose response measurements for arteries
(i.e. ~75%).41 This difference might be due to more endothelium-derived NO in arteries than in
veins. Other studies on arterial and venous grafts in patients after coronary bypass operations
seem to support our rationale.58
One important feature of our approach is the capability of in situ luminal staining of the intact
veins right after applying chemical (such as inflammatory mediators and drugs) or physical (such
as shear stress) stimuli on the blood vessel. To confirm this, endothelium of the chip-hosted vein
segments were stained for CD31. CD31 is a glycoprotein commonly used as an endothelial cell
marker. It forms a large fraction of endothelial cell intracellular junctions and is localized in the
lumen surface of blood vessels.59, 60 Fig. 3A and B demonstrates representative confocal images
of a luminally stained vein segment. Arrows in the top image of Fig. 3A indicate cell boundaries
which were fluorescently labelled. Configuration and alignment of the labelled cells validate that
the endothelial cells of the vein segment were stained. The intensity histogram from the
orthographic projection of the confocal images peaked at the region with low intensity and the
second peak fell in the region with higher intensity. The second peak of the counts fell in the
region with approximately 4 times higher intensity in the stained vessels compared to the control.
Page 39
29
Quantitative data was obtained by measuring the mean intensity of “top” and “bottom” walls of
the blood vessels through orthographic projection of the confocal images. In order to quantify the
data from fluorescent intensity measurement, orthographic projection of the confocal images
were taken with the best visualized cross-section in the XZ direction and the intensity profiles
were created along arbitrary lines across the images as is shown in Fig. 3C and 3B. The maxima
in the intensity profiles represent the “top” and “bottom” walls of the vein segments. The top and
bottom maxima were normalized by subtracting and dividing by the intensity at the mid-point
between them, i.e. the lumen of the blood vessel. Fig. 3D shows the data obtained from the
maximum intensities of intensity profiles for four stained samples and one negative control.
Stained blood vessels demonstrated on average four times higher intensity in the regions where
CD31 was expressed compared to the negative control. Nonetheless, the inconsistency in the
intensity of the fluorescent images among the samples is mostly because of the fragility of the
vein wall, which at times resulted in the blood vessel segment collapsing in the microchannel.
This effect is due to small changes in the transmural pressure (e.g. Fig. A4 B), which needs to be
addressed in future experiments in order to improve the yield of the platform.
Finally, the described approach is advantageous over conventional whole tissue
immunofluorescence staining methods because it provides the possibility of in situ luminal
staining of the intact vein right after applying chemical (such as inflammatory mediators and
drugs) or physical (such as altering the shear stress) stimuli on the blood vessel. In contrast,
procedures for conventional whole tissue staining need primary preparation including sectioning
and paraffin embedding of the blood vessels. Moreover in our approach the consumed solution is
ten times less than in the conventional methods. Finally, our method is faster since on average
there is a three days reduction in the work process compared to whole tissue staining methods.
The introduced platform has a number of challenges which some have been resolved and some
need further investigation and effort in order to improve the yield. One important experimental
problem was the appearance of bubbles in the inspection area, despite the presence of the bubble
traps in the superfusion lines. The bubbles appeared either inside or outside of the blood vessel
during the experiments that required long time i.e. > 12 hours perfusion through the blood vessel
which often damaged the endothelium of the blood vessel or resulted in collapsing the lumen as
is shown in Fig A5. A and B. Four main reasons were identified through a set of tests that could
cause the appearance of bubble: a) Volume of the bubble trapes was 0.5 µl which would fill up
Page 40
30
by large size bubbles and result in non-functioning. Implementing multi and parallel bubble traps
in the channels could be one approach to overcome this issue which has not been done yet. b)
Surface profilometry showed that the surface roughness of the sections of the manifold that were
in contact with the PDMS around the inlet holes was one source of the bubble formation.
Polishing the surface and preventing any damage to the surface improved the system in that
regards. c) Gas permeability of the fluidic tubing and syringes was another cause for bubble
formation which was resolved by changing the materials to less permeable substances, i.e. using
PEEK tubing (permeability: < 1 cm2/s) instead of Tygon tubing (permeability: 12- 30 cm2/s), and
gas-tight glass syringes instead of plastic syringes. d) Air diffusion through the manifold inlet
and the tubing connectors was another reason for the formation of bubbles in the channels, which
was temporary resolved by filling the gaps with epoxy glue. Details of the experimental setup
and conditions are explained in the Appendix A5.
Another problem regarding the experimental approach was linked to transport of the chip
containing a pressurized vein segment between different imaging modalities. As it was
mentioned earlier in this section, sudden change in the transmural pressure often resulted in
collapsed veins due to the very thin and limp wall of the blood vessel. The fluctuation in the
transmural pressure was mostly because of the change in the hydrostatic pressure applied to the
inlet of the perfusion channel which was maintained by holding a reservoir at a defined height.
One way of resolving this problem is to attach the perfusion reservoir to the manifold, thus
reduce the risk of any sudden fluctuation to the inlet pressure. Also, different approach for the
maintenance of the trasmural pressure, such as attaching a pressure transducer to the inlet of the
perfusion channel could potentially be helpful to overcome this problem.
Chip cracking was the other experimental difficulties which affected the yield of the platform.
The main reason behind this problem was the uneven force applied to the microscope slide
which was bonded to the PDMS device from the manifold. Although, this problem was less
notable when the thickness of the PDMS device was homogeneous along the chip which was an
issue relating the PDMS fabrication.
Page 41
31
Chapter 6 : Conclusion
We have developed an automated microfluidic approach for studying the intact small veins that
enables functional investigations while maintaining the near-physiological conditions as well as
immunofluorescence staining of the lumen surface. Parametric study of the platform is
performed and the shear stress applied to the wall of the blood vessel is measured for a defined
flow rate identical to the flow rate perfused through the vein segment in blood vessel
experiments. Moreover, capability of the platform for generating sheath flow and control over
the thickness of the solutions within the on-chip pressurized blood vessel is confirmed. Also, the
design of the fixation channels for immobilizing a vein segment is justified through the
measurement of the elasticity of the veins in the chip. Furthermore, performance of the vein-on-
a-chip is evaluated through the following procedure: the mouse mesenteric veins are isolated and
manually loaded onto the microfluidic platform. Viability of the vein segments is assessed
through superfusion of varying concentrations of phenylephrine and acetylcholine. Functional
assessments of the endothelium and the smooth muscle cells show 72% and 40% constriction
and dilation of the veins, respectively. Finally, a protocol for in situ immunofluorescence
staining of the lumen of the chip-hosted veins is developed. Using this protocol, pressurized
veins are stained on-chip for CD31 markers confirming the capacity of our platform for in-situ
immunofluorescence staining, without the need for tissue processing and sectioning. The
selected approach is automatable and reduces the time for staining procedures by on average 3
times compared to traditional methods. Moreover, it requires ten times less staining solution and
antibodies, i.e., 100-300 µL for the vein-on-a-chip vs. 1 mL, for the current conventional
techniques, and is capable of performing all the experimental phases from functional assessment
to immunofluorescence and imaging in less than two days on each intact vein segment.
Page 42
32
Chapter 7 : Future Directions
The established platform in this work can be used in further applications such as inflammation
studies at a molecular level. It is known that during inflammation, the endothelium of small veins
is activated due to the expression of selectins, their counter ligand, intracellular adhesion
molecule-1 (ICAM-1) and vascular cell adhesion molecule-1(VCAM-1) in response to chemical
and physical cues. Subsequently, recruitment of leukocytes from the blood stream into the
inflamed tissue takes place.5, 30, 61A proposed strategy could be to use the platform for
inflammatory activation of intact veins to quantify the expression of cell adhesion molecules
which could be exploited for detailed study of inflammation in cellular and molecular level.
One approach is superfusion of inflammatory mediators and chemical stimulation of the blood
vessel by the drugs from abluminal side. Because there are separate superfusion channels on
each side of the vein segment, the blood vessel can be activated by exposure to an inflammatory
mediator (e.g. TNF-α, LPS, fMLP) 62-65 on one side while the other side is superfused with a pure
buffer (e.g. MOPS). Then in situ immunofluorescence staining protocol can be performed for
visualizing the expression of respective cell adhesion molecules. Therefore, the activation and
control experiment can be achieved on one blood vessel in a homogeneous physical condition.
Another proposed application is to use the platform for creating controlled perfusion flowrates
and as a way to precisely regulate the applied shear stress to the blood vessel wall. This makes
the platform an ideal tool for investigating shear-dependent endothelium activation in molecular
level, and the related vascular disorders. This will be of relevance specifically in situations with
chronic venous diseases such as vasculitis and varicose veins.8, 66, 67
As we have previously mentioned an important characteristic of this platform is the possibility of
generating sheath flow within the blood vessel. This feature of the platform supports the
margination of leukocytes near the blood vessel wall which mimics in vivo leukocyte
margination during inflammation that promotes leukocyte- EC interaction.68, 69 Thereby, it
triggers the initiation of leukocyte extravasation from luminal flow to the abluminal
environment. Therefore another application of this platform is to quantitatively visualize
leukocyte extravasation cascade in vitro on an intact vein portion, and the effects of chemokines
on leukocyte rolling and adhesion can be studied. In order to achieve high-resolution cell-level
Page 43
33
imaging with this platform, a microfluidic device designed on a thinner substrate than the current
microscope slide (i.e. ~1.5 mm) such as coverslip glass is preferred. This would decrease the
working distance, by increasing magnification and numerical aperture.
As it was mentioned before, one important feature of this platform is the ability to apply
controlled perfusion luminal flow. This can be simultaneously done by probing the functional
responses of the blood vessel as well as the expression of molecules and genes on the wall which
are caused by abluminal perfusion of specific drugs (through the superfusion channels). That
feature when coupled with the micro-scale volume of reagent consumption suggests that the
proposed platform can be a suitable candidate for applications used in drug developments.
Page 44
34
References
1. Burton, A.C. Relation of structure to function of the tissues of the wall of blood vessels.
Physiol. Rev 34 (1954).
2. Hirakawa, S. Veins: Their functional role in the circulation. (Springer-Verlag, Tokyo;
1993).
3. Wiedeman, M.P. Dimensions of Blood Vessels from Distributing Artery to Collecting
Vein. Circulation Research 12, 375-378 (1963).
4. Berczi, V., Monos, E. & Nadasy, G. in Physiological Reviews, Vol. 75 861995).
5. Ley, K., Laudanna, C., Cybulsky, M.I. & Nourshargh, S. Getting to the site of
inflammation: the leukocyte adhesion cascade updated. Nature reviews. Immunology 7,
678-689 (2007).
6. Rao, R.M., Yang, L., Garcia-Cardena, G. & Luscinskas, F.W. Endothelial-Dependent
Mechanisms of Leukocyte Recruitment to the Vascular Wall. Circulation Research 101,
234-247 (2007).
7. Zamboni, P. et al. In vitro versus in vivo assessment of vein wall properties. Annals of
Vascular Surgery 12, 324-329 (1998).
8. Eriksson, E.E.K., Eva; Lundmark, Karin; Rotzius, Pierre; Hedin, Ulf; Xie, Xun Powerful
Inflammatory Properties of Large Vein Endothelium In Vivo. Arterioscler Thromb Vasc
Biology 25, 6 (2005).
9. Asbeutah, A.M. et al. In Vivo and In Vitro Assessment of Human Saphenous Vein Wall
Changes. The open cardiovascular medicine journal 1, 15-21 (2007).
10. Khandoga, A.G. et al. In Vivo Imaging and Quantitative Analysis of Leukocyte
Directional Migration and Polarization in Inflamed Tissue. PLoS ONE 4, e4693 (2009).
11. Aikawa, E. in Cardiovascular Imaging: Arterial and Aortic Valve Inflammation and
Calcification
377 (Springer, Boston, USA; 2015).
12. Yang, L. et al. ICAM-1 regulates neutrophil adhesion and transcellular migration of
TNF-α-activated vascular endothelium under flow. Blood 106, 584-592 (2005).
13. Furie, M.B., Naprstek, B.L. & Silverstein, S.C. Migration of neutrophils across
monolayers of cultured microvascular endothelial cells. An in vitro model of leucocyte
extravasation. Journal of Cell Science 88, 161-175 (1987).
14. Bevan, J.A. & Osher, J.V. A direct method for recording tension changes in the wall of
small blood vessels in vitro. Agents and Actions 2, 257-260 (1972).
Page 45
35
15. Mulvany, M.J. & Halpern, W. Mechanical properties of vascular smooth muscle cells in
situ. Nature 260, 617-619 (1976).
16. Bridges, L.E., Williams, C.L., Pointer, M.A. & Awumey, E.M. Mesenteric Artery
Contraction and Relaxation Studies Using Automated Wire Myography. Journal of
Visualized Experiments : JoVE, 3119 (2011).
17. Yokomizo, T. et al. Whole-mount three-dimensional imaging of internally localized
immunostained cells within mouse embryos. Nature protocols 7, 421-431 (2012).
18. Graham, A.V., M. in IHC staining protocol for whole mount samples, Vol. 2015 (abcam
discover more, US; 2015).
19. Kovalenko, N. & Chernukh, A. [Role of the venule and lymphatic links of
microcirculation in transcapillary exchange at the level of the functional unit of an
organ]. Biulleten'eksperimental'noi biologii i meditsiny 91, 367-369 (1981).
20. Yuan, Y., Chilian, W.M., Granger, H.J. & Zawieja, D.C. Permeability to albumin in
isolated coronary venules. American Journal of Physiology-Heart and Circulatory
Physiology 265, H543-H552 (1993).
21. Wang, S. et al. Venular basement membranes contain specific matrix protein low
expression regions that act as exit points for emigrating neutrophils. The Journal of
experimental medicine 203, 1519-1532 (2006).
22. Grega, G. & Adamski, S. The role of venular endothelial cells in the regulation of
macromolecular permeability. Microcirculation, endothelium, and lymphatics 4, 143-167
(1988).
23. Kilzer, P., Chang, K., Marvel, J., Kilo, C. & Williamson, J.R. Tissue differences in
vascular permeability changes induced by histamine. Microvascular research 30, 270-
285 (1985).
24. Michel, C.C. & Curry, F.E. Microvascular Permeability. Physiological Reviews 79, 703-
761 (1999).
25. Medzhitov, R. Inflammation 2010: New Adventures of an Old Flame. Cell 140, 771-776
(2010).
26. Lewis, R.E. & Granger, H.J. Diapedesis and the permeability of venous microvessels to
protein macromolecules: The impact of leukotriene B4 (LTB4). Microvascular research
35, 27-47 (1988).
27. Feng, D., Nagy, J.A., Pyne, K., Dvorak, H.F. & Dvorak, A.M. Neutrophils emigrate from
venules by a transendothelial cell pathway in response to FMLP. The Journal of
experimental medicine 187, 903-915 (1998).
Page 46
36
28. Ley, K. & Gaehtgens, P. Endothelial, not hemodynamic, differences are responsible for
preferential leukocyte rolling in rat mesenteric venules. Circulation Research 69, 1034-
1041 (1991).
29. Geng, J.G. Directional migration of leukocytes: their pathological roles in inflammation
and strategies for development of anti-inflammatory therapies. Cell Res 11, 85-88 (2001).
30. Muller, W.A. Leukocyte–endothelial-cell interactions in leukocyte transmigration and the
inflammatory response. Trends in Immunology 24, 326-333 (2003).
31. Mamdouh, Z., Mikhailov, A. & Muller, W.A. Transcellular migration of leukocytes is
mediated by the endothelial lateral border recycling compartment. The Journal of
experimental medicine 206, 2795-2808 (2009).
32. van der Meer, A.D., Poot, A.A., Duits, M.H.G., Feijen, J. & Vermes, I. Microfluidic
Technology in Vascular Research. Journal of Biomedicine and Biotechnology 2009
(2009).
33. Wong Kh Fau - Chan, J.M., Chan Jm Fau - Kamm, R.D., Kamm Rd Fau - Tien, J. &
Tien, J. Microfluidic models of vascular functions.
34. Prabhakarpandian, B., Shen, M.-C., Pant, K. & Kiani, M.F. Microfluidic devices for
modeling cell-cell and particle-cell interactions in the microvasculature. Microvascular
research 82, 210-220 (2011).
35. Khan, O.F. & Sefton, M.V. Endothelial cell behaviour within a microfluidic mimic of the
flow channels of a modular tissue engineered construct. Biomedical microdevices 13, 69-
87 (2011).
36. Han, S. et al. A versatile assay for monitoring in vivo-like transendothelial migration of
neutrophils. Lab on a Chip 12, 3861-3865 (2012).
37. Schaff, U.Y. et al. Vascular mimetics based on microfluidics for imaging the leukocyte--
endothelial inflammatory response. Lab on a chip 7, 448-456 (2007).
38. Zhang, B., Peticone, C., Murthy, S.K. & Radisic, M. A standalone perfusion platform for
drug testing and target validation in micro-vessel networks. Biomicrofluidics 7, 044125
(2013).
39. Namdee, K., Thompson, A.J., Charoenphol, P. & Eniola-Adefeso, O. Margination
Propensity of Vascular-Targeted Spheres from Blood Flow in a Microfluidic Model of
Human Microvessels. Langmuir 29, 2530-2535 (2013).
40. Tsai, M. et al. In vitro modeling of the microvascular occlusion and thrombosis that occur
in hematologic diseases using microfluidic technology. The Journal of Clinical
Investigation 122, 408-418 (2012).
41. Gunther, A. et al. A microfluidic platform for probing small artery structure and
function(). Lab on a chip 10, 2341-2349 (2010).
Page 47
37
42. Yasotharan, S., Pinto, S., Sled, J.G., Bolz, S.S. & Gunther, A. Artery-on-a-chip platform
for automated, multimodal assessment of cerebral blood vessel structure and function.
Lab on a chip 15, 2660-2669 (2015).
43. Lochovsky, C., Yasotharan, S. & Gunther, A. Bubbles no more: in-plane trapping and
removal of bubbles in microfluidic devices. Lab on a chip 12, 595-601 (2012).
44. S. G. Kandlikar, S.G., D. Li, S. Colin and M. R. King Heat Transfer and Fluid Flow in
Minichannels and Microchannels
Edn. 2. (Elsevier, United States; 2014).
45. Morris, C. & Forster, F. Oscillatory flow in microchannels. Experiments in Fluids 36,
928-937 (2004).
46. Lawrence, M.B., Kansas, G.S., Kunkel, E.J. & Ley, K. Threshold Levels of Fluid Shear
Promote Leukocyte Adhesion through Selectins (CD62L,P,E). The Journal of Cell
Biology 136, 717-727 (1997).
47. Lawrence, M.B. & Springer, T.A. Leukocytes roll on a selectin at physiologic flow rates:
Distinction from and prerequisite for adhesion through integrins. Cell 65, 859-873
(1991).
48. Theret, D.P., Levesque, M.J., Sato, M., Nerem, R.M. & Wheeler, L.T. The Application of
a Homogeneous Half-Space Model in the Analysis of Endothelial Cell Micropipette
Measurements. Journal of Biomechanical Engineering-transactions of The Asme 110
(1988).
49. Wesly, R.L., Vaishnav, R.N., Fuchs, J.C., Patel, D.J. & Greenfield, J.C. Static linear and
nonlinear elastic properties of normal and arterialized venous tissue in dog and man.
Circulation Research 37, 509-520 (1975).
50. Black, J. & Hastings, G. Handbook of biomaterial properties. (Springer Science &
Business Media, 2013).
51. Haddy, F.J. & Gilbert, R.P. The Relation of a Venous-Arteriolar Reflex to Transmural
Pressure and Resistance in Small and Large Systemic Vessels. Circulation Research 4,
25-32 (1956).
52. Enouri, S., Monteith, G. & Johnson, R. Characteristics of myogenic reactivity in isolated
rat mesenteric veins. American Journal of Physiology - Regulatory, Integrative and
Comparative Physiology 300, R470-R478 (2011).
53. Zocalo, Y. et al. Structural and Functional Properties of Venous Wall: Relationship
between Elastin, Collagen, and Smooth Muscle Components and Viscoelastic Properties.
ISRN Physiology 2013, 9 (2013).
Page 48
38
54. Lee, C.-H. et al. The mechanism of phenylephrine-mediated [Ca2+]i oscillations
underlying tonic contraction in the rabbit inferior vena cava. The Journal of Physiology
534, 641-650 (2001).
55. Collins, E.M., Walsh, M.P. & Morgan, K.G. Contraction of single vascular smooth
muscle cells by phenylephrine at constant [Ca2+]i. American Journal of Physiology -
Heart and Circulatory Physiology 262, H754-H762 (1992).
56. Enouri, S., Monteith, G. & Johnson, R. Effects of Acute Transmural Pressure Elevation
on Endothelium-Dependent Vasodilation in Isolated Rat Mesenteric Veins. Journal of
Vascular Research 51, 27-36 (2014).
57. Kellogg, D.L., Zhao, J.L., Coey, U. & Green, J.V. Acetylcholine-induced vasodilation is
mediated by nitric oxide and prostaglandins in human skin. Journal of Applied
Physiology 98, 629-632 (2005).
58. Nishioka, H. et al. Difference in acetylcholine-induced nitric oxide release of arterial and
venous grafts in patients after coronary bypass operations. The Journal of Thoracic and
Cardiovascular Surgery 116, 454-459.
59. Ilan, N., Cheung, L., Pinter, E. & Madri, J.A. Platelet-Endothelial Cell Adhesion
Molecule-1 (CD31), a Scaffolding Molecule for Selected Catenin Family Members
Whose Binding Is Mediated by Different Tyrosine and Serine/Threonine
Phosphorylation. Journal of Biological Chemistry 275, 21435-21443 (2000).
60. Liu, L. & Shi, G.-P. CD31: beyond a marker for endothelial cells. Cardiovascular
research 94, 3-5 (2012).
61. Yadav, R., Larbi, K.Y., Young, R.E. & Nourshargh, S. Migration of leukocytes through
the vessel wall and beyond. Thrombosis and Haemostasis (2003).
62. Davies, M.J. et al. The expression of the adhesion molecules ICAM-1, VCAM-1,
PECAM, and E-selectin in human atherosclerosis. The Journal of Pathology 171, 223-
229 (1993).
63. Szmitko, P.E. et al. New Markers of Inflammation and Endothelial Cell Activation: Part
I. Circulation 108, 1917-1923 (2003).
64. Dauphinee, S.M. & Karsan, A. Lipopolysaccharide signaling in endothelial cells. Lab
Invest 86, 9-22 (2005).
65. Newton, K. & Dixit, V.M. Signaling in Innate Immunity and Inflammation. Cold Spring
Harbor Perspectives in Biology 4 (2012).
66. Smith, M.L., Smith, M.J., Lawrence, M.B. & Ley, K. Viscosity-Independent Velocity of
Neutrophils Rolling on P-Selectin In Vitro or In Vivo. Microcirculation 9, 523-536
(2002).
Page 49
39
67. Sprague, A.H. & Khalil, R.A. Inflammatory cytokines in vascular dysfunction and
vascular disease. Biochemical Pharmacology 78, 539-552 (2009).
68. Goldsmith H. L., S.S. Margination of Leukocytes in Blood Flow through Small Tubes.
Microvascular research 27, 18 (1984).
69. Pearson, M.J. & Lipowsky, H.H. Effect of Fibrinogen on Leukocyte Margination and
Adhesion in Postcapillary Venules. Microcirculation 11, 295-306 (2004).
Page 50
40
Appendices
A1. Experimental setup
Figure A1: Experimental setup. (A) Photograph and (B) schematic illustration of the vein-on-
a-chip platform consisted of: (1)Microfluidic device with manifold (2) Selector valve and
reservoirs (3) Syringe pumps (4) Temperature control system (5) Peristaltic pump (6) Vaccum
Page 51
41
system for operating the bubble raps (7) Sterio microscope (8) Inverted microscope (9) CCD
camera (10) Computer
A2. Temperature control setup
The desired temperatures during different stages of experiment are achieved and controlled at the
inspection area of the chip as illustrated in Fig. A1. thermoelectric element (model TE-35- 0.6-
1.0, TE Technology) connected to a computer-controlled single mode temperature controller is
used to generate warm and cold temperatures of interest at the organ bath. A sapphire disk
(diameter: 25mm, thickness: 1mm, thermal conductivity: 35 W/m. K, model WSR-251, UQG
Optics, Cambridge, UK) is placed in between the TE element and the PDMS device using a
thermally conductive glue (thermal conductivity: 7.5 W/m. K) to produce a horizontally uniform
and constant temperature across the inspection area. To accurately generate and control
temperature at the inspection section, a thermistor (model MP-2444, TE Technology) was
embedded into the chip at the vicinity of vein segment location during the molding process of
PDMS. Heating (37 °C) and cooling (4 °C) is feedback controlled from thermistor. Temperature
control is automated in a custom software program (LabView, National Instruments, Austin, TX,
USA). In order to dissipate heat from the side of the TE element that is not in contact with the
PDMS device, a custom made cooling element is machined in aluminium and attached. The
cooling jacket is placed on top of the TE element and perfused with water at a rate of 30 mL/min,
using a peristaltic pump (Fisher Scientific 13-875- DIG. FH100 PUMP 14-4000ML/MIN EA).
Water perfusion produces an on-chip temperature change from 37 °C to 4 °C at a rate of 2
°C/min.
Page 52
42
Figure A2: Temperature control setup. Schematic illustration of the automated temperature
control setup consisted of: a TE element glued to a sapphire dick which was placed on top of the
organ bath, a thermistor embedded in the PDMS device at the vicinity of the vein segment
(distance: 1.5 mm), a temperature controller connected to the TE element and the thermistor, a
cooling jacket which was placed on top of the TE element to dissipate heat produced during the
cooling mode operation, and a peristaltic pump connected to the cooling jacket for the water
perfusion.
A3. Dose Response Data
Figure A2 (A) and (B) represent chip-hosted vein segments, abluminally exposed to
phenylephrine (PE) and acetylcholine (Ach) through the superfusion channels at the top and the
bottom, respectively.
Page 54
44
Figure A3: Dose response data. Chip-hosted mouse mesenteric vein segments (A) constricted
in response to incrementally increasing phenylephrine (PE) concentration, and (B) dilated in
response to incrementally increasing acetylcholine (Ach) concentration. Contrast is adjusted to
keep consistency in the images. Scale bars are 200μm.
A4. Staining Data
This section shows the images and respective intensity plots obtained from four stained samples.
Page 56
46
Figure A4: Staining data. Orthographic projections and the respective intensity profiles of the
confocal images of four samples stained for CD31. Scale bars represent 50 µm for images A and
B, and 100 µm for images C and D.
Page 57
47
A5. Bubble Formation Test
In order to investigate the source of bubble formation in the inspection area (i.e. where the vein segment
was held) during long run experiments (i.e. >12hours) a set of tests were performed. The experimental
platform was set up as follows: a PDMS device that contained of all the channel networks of the original
microfluidic device, except for the fixation channels, was used within the manifold. All of the fluidic
connections were implemented similar to the original vein-on-a-chip platform. Perfusion and superfusion
flow rates were set at 1 mL/hr. The bubble traps were connected to a miniature vacuum pump which was
operating throughout the experiment. The chip was incrementally heated up to 37 °C and images were
taken at a rate of 1fps for 16 hours from underneath using a CCD camera (model EXi Aqua,
QImaging, Surrey, BC, Canada) attached to a telecentric lens (zoom 7000, magnification: 18-108
mm, working distance: 127 mm, Navitar, Japan). Initially five main sources for the presence of
bubbles were assumed: a) heating the chip, b) roughness of the rings on the manifold which were used to
seal the fluidic inlets and outlets c) the small capacity of the bubble traps, d) gas diffusion through the
tubing and the syringes, and e) infusion of gas to the channels through the manifold inlets and connectors.
The assumptions were tested by running stepwise experiments. First, we performed the experiments at
room temperature and 37 °C, which rejected the first assumption as variation in the temperature did not
affect the bubble creation. In the next step, we polished the surface of the manifold that were in contact to
the PDMS for the sealing purposes. Then we changed the tubing material from plastic to PEEK and used
gas tight glass syringes instead of plastic ones. Next we sealed the manifold inlets by filling the gaps
between the tubing. After each step we repeated the experiment. Fig. A5 shows image data obtained from
the experiments. The results showed that the tubing and syringe material influence the bubble formation.
As well, polished surface and proper sealing between the manifold, the PDMS, and the tubing had effects
on the presence of bubbles in the microchannel. Fig. A5. F shows the number of bubbles presented in the
microchannels in each experiment plotted over time. The presence and absence of bubbles were indicated
as 1 and 0, respectively.
Page 58
48
Figure A5: Bubble formation test. A and B) Images showing the presence of bubbles in the
superfusion and perfusion microchannels, respectively. Scale bar: 200 µm. C) Image of the chip
without bubble. Scale bar: 2 mm. D) Image of chip with bubbles in the superfusion inlet and the
bubble trap. Circles indicate the location of the bubbles inside the microchannels. Scale bar: 2
Page 59
49
mm. E) Schematic of the manifold (left) and bottom surface of the connection inset (right). One
of the rings on the manifold which support the seal between the manifold and the PDMS device
is denoted by a circle. F) Bubble presence plotted over time for four experimental conditions as
follows: Test 1: before polishing, tygon tubing, plastic syringe, room temperature. Test 2: before
polishing, tygon tubing, plastic syringe, 37 °C. Test 3: before polishing, peek tubing, plastic
syringe. Test 4: after polishing, peek tubing, gastight syringe.