Improving the Thomson scattering Diagnostic Improving the Thomson scattering Diagnostic installed on the Large Helical Device installed on the Large Helical Device (changed) (changed) Validity Check of Thomson Scattering Validity Check of Thomson Scattering Data of Very High Density Plasmas in Data of Very High Density Plasmas in LHD LHD K. K. Narihara Narihara , I. Yamada, H. Hayashi and H. , I. Yamada, H. Hayashi and H. Funaba Funaba National Institute for Fusion Science National Institute for Fusion Science 509 509 - - 5292 Toki, Japan 5292 Toki, Japan P8-24
20

Validity Check of Thomson Scattering Data of Very High ... · for for VbVb=0.9Vr; and ~1.6V for =0.9Vr; and ~1.6V for VbVb=0.5 =0.5 VrVr.. 3. The newly installed 133. The newly installed

Oct 17, 2020

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Validity Check of Thomson Scattering Data of Very High ... · for for VbVb=0.9Vr; and ~1.6V for =0.9Vr; and ~1.6V for VbVb=0.5 =0.5 VrVr.. 3. The newly installed 133. The newly installed

Improving the Thomson scattering Diagnostic Improving the Thomson scattering Diagnostic installed on the Large Helical Deviceinstalled on the Large Helical Device

(changed)(changed)

Validity Check of Thomson Scattering Validity Check of Thomson Scattering Data of Very High Density Plasmas in Data of Very High Density Plasmas in

LHDLHD

K.K. NariharaNarihara, I. Yamada, H. Hayashi and H. , I. Yamada, H. Hayashi and H. FunabaFunaba

National Institute for Fusion ScienceNational Institute for Fusion Science509509--5292 Toki, Japan5292 Toki, Japan

  

P8-24

Page 2: Validity Check of Thomson Scattering Data of Very High ... · for for VbVb=0.9Vr; and ~1.6V for =0.9Vr; and ~1.6V for VbVb=0.5 =0.5 VrVr.. 3. The newly installed 133. The newly installed

MotivationsMotivationsHow confident is the Thomson scattering data for a How confident is the Thomson scattering data for a very high density plasma, which accompanies very very high density plasma, which accompanies very intense back/foreintense back/fore--ground plasma light?ground plasma light?We examined the effect of back/foreWe examined the effect of back/fore--ground plasma ground plasma light on the performance of an avalanche photodiode light on the performance of an avalanche photodiode (APD) used for the detecting the Thomson(APD) used for the detecting the Thomson--scattered scattered light, and established a criterion for the validity.light, and established a criterion for the validity.We newly installed thirteen 80We newly installed thirteen 80--channel scanning ADCs channel scanning ADCs to monitor the DC level of to monitor the DC level of APDsAPDs, which is a measure , which is a measure of the intensity of the plasma light.of the intensity of the plasma light.

Page 3: Validity Check of Thomson Scattering Data of Very High ... · for for VbVb=0.9Vr; and ~1.6V for =0.9Vr; and ~1.6V for VbVb=0.5 =0.5 VrVr.. 3. The newly installed 133. The newly installed

Scattering Configuration of LHD TSScattering Configuration of LHD TS

Highly oblique back scattering

Whole plasma region along a major radius on the mid plane

200-spatial channels (the plasma is covered by 130 channels)

Segmented mirror

Tips of fiber array

Page 4: Validity Check of Thomson Scattering Data of Very High ... · for for VbVb=0.9Vr; and ~1.6V for =0.9Vr; and ~1.6V for VbVb=0.5 =0.5 VrVr.. 3. The newly installed 133. The newly installed

Scattering Parameters as a Function of the Scattering Position

Page 5: Validity Check of Thomson Scattering Data of Very High ... · for for VbVb=0.9Vr; and ~1.6V for =0.9Vr; and ~1.6V for VbVb=0.5 =0.5 VrVr.. 3. The newly installed 133. The newly installed

PolychromatorPolychromator

LHD_polychromator#36

400 600 800 1000 1200WAVELENGTH(nm)

0.0

0.2

0.4

0.6

0.8

1.0 YAGH_aH_b HeI

1keV

2keV

4keV

8keV

16keV

Expected Plasma light

Thomson scattered light

ch5

ch4ch3

ch2ch1

Page 6: Validity Check of Thomson Scattering Data of Very High ... · for for VbVb=0.9Vr; and ~1.6V for =0.9Vr; and ~1.6V for VbVb=0.5 =0.5 VrVr.. 3. The newly installed 133. The newly installed

_+

R2

R1100kΩ

1nF

APD

Vb

47Ω

10kΩ

1881M FASTBUS ADC(1024ch)

NI6225 scanning ADC(1040ch)

-6C1=10nF

Circuit Diagram of APD and Data Acquisition System

1 of 1000

EG& C30950 CD1161

Newly installed

Page 7: Validity Check of Thomson Scattering Data of Very High ... · for for VbVb=0.9Vr; and ~1.6V for =0.9Vr; and ~1.6V for VbVb=0.5 =0.5 VrVr.. 3. The newly installed 133. The newly installed

Setup for Test MeasurementSetup for Test Measurement

Pulse generator

DC-powerYHP-LED5082-4850 APD

HighVoltage Vb

50ns pulse light

APD: EG&G C30950-CD1161 S/NC2087DOperating Voltage Vr=307V @ R=675kV/WDark current Id= 140 nABand Width BW=25 MHz

Page 8: Validity Check of Thomson Scattering Data of Very High ... · for for VbVb=0.9Vr; and ~1.6V for =0.9Vr; and ~1.6V for VbVb=0.5 =0.5 VrVr.. 3. The newly installed 133. The newly installed

Output Voltage as a Function of Bias VoltageOutput Voltage as a Function of Bias Voltage

APD S/N=C2087-D

100 150 200 250 300 350Bias Voltage (V)

0

50

100

150

200

250

Out

put(m

V)

Vr

Page 9: Validity Check of Thomson Scattering Data of Very High ... · for for VbVb=0.9Vr; and ~1.6V for =0.9Vr; and ~1.6V for VbVb=0.5 =0.5 VrVr.. 3. The newly installed 133. The newly installed

Vb=0.9Vr

0 20 40 60LED current (mA)

024

DC

Out

(V)

Vb=0.5Vr

0 20 40 60LED current (mA)

024

DC

Out

(V)

LED-CURRENT vs DC-OUTPUT for two different high voltage

Page 10: Validity Check of Thomson Scattering Data of Very High ... · for for VbVb=0.9Vr; and ~1.6V for =0.9Vr; and ~1.6V for VbVb=0.5 =0.5 VrVr.. 3. The newly installed 133. The newly installed

Pulse Output as a Function of DCPulse Output as a Function of DC--level for a Fixed level for a Fixed Pulse Light Intensity:Pulse Light Intensity:

10%

Guaranteed range

APD S/N=C2087-D

0 1 2 3DC Voltage (V)

0

150

300

Out

put(m

V)

Vb=0.9Vr

Vb=0.5Vr

10%

Lower Vb is favorable for measuring denser plasma:

But too low Vb (eg. 0.4 Vr) causes the frequency response of the APD slow (the depression layer width becomes narrower and the APD has a higher capacity.)

Page 11: Validity Check of Thomson Scattering Data of Very High ... · for for VbVb=0.9Vr; and ~1.6V for =0.9Vr; and ~1.6V for VbVb=0.5 =0.5 VrVr.. 3. The newly installed 133. The newly installed

FASTBUS1881M

pulseAmp

NIM modules

pulseAmp

pulseAmp

Pxi-6225

pulseAmp pulseAmp

Page 12: Validity Check of Thomson Scattering Data of Very High ... · for for VbVb=0.9Vr; and ~1.6V for =0.9Vr; and ~1.6V for VbVb=0.5 =0.5 VrVr.. 3. The newly installed 133. The newly installed
Page 13: Validity Check of Thomson Scattering Data of Very High ... · for for VbVb=0.9Vr; and ~1.6V for =0.9Vr; and ~1.6V for VbVb=0.5 =0.5 VrVr.. 3. The newly installed 133. The newly installed

Example #69360:Example #69360:

#69360time=1.03355s

3.20 3.38 3.56 3.74 3.92 4.10 4.28 4.46 4.64 4.82 5.00 MAJOR RADIUS (m)

0.0

0.2

0.4

0.6

0.8

1.0

Te(

keV

), ne

(10^

21 m

^-3)

Te

ne

0.0 0.5 1.0 1.5 2.0 2.5TIME(s)

-0.05

0.00

0.05

0.100.15

volt

0 200 400 600 800channel#

-0.10-0.050.000.050.100.150.20

volt

DC

OU

T(V

)D

C O

UT

(V)

ch404

Time evolution of DC-level

1.03s

We assumed that all APD’s gain depend on Vb/Vr similarly, which is not yet checked.Density calibration is not yet completed.

DC-level of all APDs are low enough toguarantee the linearity.

VbVb=0.5 =0.5 VrVr; Window fully opened.; Window fully opened.

Page 14: Validity Check of Thomson Scattering Data of Very High ... · for for VbVb=0.9Vr; and ~1.6V for =0.9Vr; and ~1.6V for VbVb=0.5 =0.5 VrVr.. 3. The newly installed 133. The newly installed
Page 15: Validity Check of Thomson Scattering Data of Very High ... · for for VbVb=0.9Vr; and ~1.6V for =0.9Vr; and ~1.6V for VbVb=0.5 =0.5 VrVr.. 3. The newly installed 133. The newly installed

Comment on the 9Comment on the 9thth Campaign DataCampaign Data

APD S/N=C2087-D

0 1 2 3DC Voltage (V)

0

150

300

Out

put(m

V)

Vb=0.9Vr

Vb=0.5Vr

Vb=0.9Vr; window 10% open

We have no information on interferometer and VDC

Suppositions:

1: if VDC<0.8V for all APD: error on ne & Te would be less than 30% ::OK

2. If VDC > 2.5V for most APD: Te profile would be irregular and unnaturalWe would disregard it:: OK (We would not be deceived).

3. If VDC < 2 V for all APDs: signal error <30%; ne & Te profile may be smooth in shape but have error 60~90%; highly deformed profile. We should be much careful not to be deceived.

Page 16: Validity Check of Thomson Scattering Data of Very High ... · for for VbVb=0.9Vr; and ~1.6V for =0.9Vr; and ~1.6V for VbVb=0.5 =0.5 VrVr.. 3. The newly installed 133. The newly installed

ConclusionsConclusions1. 1. We evaluated the effect of the plasma light on the We evaluated the effect of the plasma light on the pulsepulse--response of response of APDs APDs used for detecting the used for detecting the Thomson scattered light.Thomson scattered light.2. For accuracy better than 10%, the DC2. For accuracy better than 10%, the DC--outputs outputs induced by plasma light should be less than ~0.8 V, induced by plasma light should be less than ~0.8 V, for for VbVb=0.9Vr; and ~1.6V for =0.9Vr; and ~1.6V for VbVb=0.5 =0.5 VrVr..3. The newly installed 133. The newly installed 13--80ch scanning ADCs 80ch scanning ADCs routinely monitor the DCroutinely monitor the DC--levels of all APD outputs, levels of all APD outputs, thus enabling the linearitythus enabling the linearity--check on the acquired data.check on the acquired data.4. For measurement under super bright radiation, 4. For measurement under super bright radiation, combined optimization ofcombined optimization of VbVb and the windowsand the windows’’aperture will be needed. aperture will be needed.

Page 17: Validity Check of Thomson Scattering Data of Very High ... · for for VbVb=0.9Vr; and ~1.6V for =0.9Vr; and ~1.6V for VbVb=0.5 =0.5 VrVr.. 3. The newly installed 133. The newly installed
Page 18: Validity Check of Thomson Scattering Data of Very High ... · for for VbVb=0.9Vr; and ~1.6V for =0.9Vr; and ~1.6V for VbVb=0.5 =0.5 VrVr.. 3. The newly installed 133. The newly installed
Page 19: Validity Check of Thomson Scattering Data of Very High ... · for for VbVb=0.9Vr; and ~1.6V for =0.9Vr; and ~1.6V for VbVb=0.5 =0.5 VrVr.. 3. The newly installed 133. The newly installed
Page 20: Validity Check of Thomson Scattering Data of Very High ... · for for VbVb=0.9Vr; and ~1.6V for =0.9Vr; and ~1.6V for VbVb=0.5 =0.5 VrVr.. 3. The newly installed 133. The newly installed