Use of Handheld XRF Bruker Tracer III-SD Research Laboratories of Archaeology The University of North Carolina at Chapel Hill Lindsay Bloch Last updated June 2015
21

Use of Handheld XRF may2015 - Archaeologyarchaeology.sites.unc.edu/files/2017/03/UNC-Handheld-XRF-Manual.pdf · All documents referenced in this guide are located in folder Basic

May 01, 2018

Download

Documents

hoangxuyen
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Use of Handheld XRF may2015 - Archaeologyarchaeology.sites.unc.edu/files/2017/03/UNC-Handheld-XRF-Manual.pdf · All documents referenced in this guide are located in folder Basic

UseofHandheldXRFBrukerTracerIII-SD

ResearchLaboratoriesofArchaeologyTheUniversityofNorthCarolinaatChapelHill

LindsayBloch

LastupdatedJune2015

Page 2: Use of Handheld XRF may2015 - Archaeologyarchaeology.sites.unc.edu/files/2017/03/UNC-Handheld-XRF-Manual.pdf · All documents referenced in this guide are located in folder Basic

ThisguideisintendedtoorientuserstotheproperoperationofthehandheldXRFanalyzer,andtodetailmethodsforanalyzingarchaeologicalmaterials,especiallyceramics.TheinformationinthisguidehasbeencompiledfrommultiplesourcessuppliedbyBruker,aswellasinformationimpartedbyBrukerrepresentativesintrainingsessions,andDr.JeffSpeakman,DirectoroftheCenterforAppliedIsotopeStudies,UniversityofGeorgia.AlldocumentsreferencedinthisguidearelocatedinfolderBasicInformationontheDesktopofXRFlaptop.

Page 3: Use of Handheld XRF may2015 - Archaeologyarchaeology.sites.unc.edu/files/2017/03/UNC-Handheld-XRF-Manual.pdf · All documents referenced in this guide are located in folder Basic

TABLEOFCONTENTS

I. ANALYZERSPECIFICATIONS…………………………………………………….……………………………………….1II. SAFETY……………………………………………………….……………………………………………………………………1III. INSTRUMENTCOMPONENTSANDINVENTORY……………………………………….……………………….1 A.TheBrukerTracerIII-SDHandheldXRFAnalyzer…………………………….……………………………….1 B.VacuumPump……………………………………………………………………….…………………………..……..…….2 C.Computer…………………………..………………………………….………………………………………………....…….2 D.AuxiliaryEquipment…………………………………………………………………………………………………………2IV. BASICOPERATION………………………………………….………………………………………………….…………….3

A.Connecting……………………………………………………………………………………………………………………..3 B.S1PXRFProgram………………………………………………………………………………………..……………………4

i)Settings………………………………………………………………………………………………………..…..…………6ii)XRayOpsProgram………………………………………………………………………………………….……………7iii)AnalyzingSamples……..……………………………….……………………………………………..……………...7iv)InterpretingSpectra……………………………………………………………….………………….………………8v)SpectrumOverlay…………………………………………………….……………………………………………….10

C.ArtaxProgram……………………………………………………………………………………………….……….………10i)Installation…………………………………………………………….…………………………………..……………..10ii)CreatingProjects…………………………………………………….……………………………….……………….11iii)ExtractingNetPhotonCounts………………………………………………………………………………….11iii)ExportingData…………………………………………………………………………………………….…………..13iv)Visuals……………………………………………………………………………………………………………….…….13

IV. QUANTIFICATION…………………………………………………………….…………….…………….…………….….13A. UseofCalibrations…………………………………………………….………………………………………………..13B. UsinganExistingCalibration………………………….……………………………………………………........14

i)S1PXRFCalibration………………………………………….…………………………………………………………15 ii)S1CalProcessCalibration…………………………………………………………………………………………..15

C. AddingorRemovingCalibrationReferenceMaterial………………………..…………………………15D. TestingYourCalibration………………………………………………………………….……………………….….16

V. SELECTINGAPPROPRIATESAMPLES…………………………………………………..…………………………..16 Appendix.SettingsforCeramicsCalibration…………………………..…………….…………….………….18

Page 4: Use of Handheld XRF may2015 - Archaeologyarchaeology.sites.unc.edu/files/2017/03/UNC-Handheld-XRF-Manual.pdf · All documents referenced in this guide are located in folder Basic

I. ANALYZERSPECIFICATIONS

TheBrukerTracerIII-SDhasanX-raytubewitharhodiumtargetandasilicondriftdetector(SiPINdiode).Itcontainsnoradioactivematerial.TheTracerII-SDoperateswitharesolutionof147eVFWHMfor5.9keV,at200,000countspersecond.TheX-raybeamiscollimatedwithdimensionsof4mmx3mm.Spotsizeofanalysisisapproximately10mmindiameter.Themaximumvoltageis40keV,whichisinthesoftX-rayrange,anddoesnotpenetrateasdeeplyintoobjectsashardX-raysusedformedicalapplications.

II. SAFETY

X-raysareharmfultothehumanbody.Inordertominimizeexposureduringtheoperationoftheinstrument,allsafetyproceduresmustbefollowedatalltimes.ThereareanumberofsafetyresourcesfortheBrukerTracerIII-SD.Beforeoperatingthisinstrument,allUNCusersmust:

• ReadtheBrukerRadiationSafetyguide(onlaptop),• CompleteabriefonlinecourseinanalyticalradiationsafetythroughUNC

EnvironmentalHealthandSafety:http://ehs.unc.edu/training/self-study/analytical-x-ray-training/

• ReadUNCRadiationProtectionProgramguidefortheuseofX-raydevicesandsigntheincludedoperatorspage(inbinderwithlaptop).

III. INSTRUMENTCOMPONENTSANDINVENTORY

ThefullhandheldXRFrigconsistsofthreemainelements:theXRFanalyzer,thevacuumpump,andthelaptopcomputerorPDA.Eachuserisresponsibleforinventoryingallcomponentsbeforeandafterinstrumentuse.Anydiscrepanciesordamagesshouldbereportedimmediately.

A. TheBrukerTracerIII-SDHandheldXRFAnalyzer

Inadditiontotheinstrument,thecasefortheanalyzershouldalsocontain:

2rechargeablebatterypacks,Powercable-fordirectlypoweringtheanalyzerorrechargingthebatteries,Beigecommunicationcable,Boxof4filterslides,Tweezersforinsertingandremovingfilters,Metalnoseprotector-tobeplacedontheinstrumentbeforestorageandremovedpriortouse,

Page 5: Use of Handheld XRF may2015 - Archaeologyarchaeology.sites.unc.edu/files/2017/03/UNC-Handheld-XRF-Manual.pdf · All documents referenced in this guide are located in folder Basic

2

Sampletable,Safetycap,Foldingmirror,Plastictabletopstand,Replacementfilmwindowsfornose,InstrumentkeyBehindthefoamonthelidofthebox:UserGuide,QuickStartGuide,ChartofPeriodicTableandX-rayenergiesSoftwarediscs.

B. VacuumPumpThevacuumpumpmaybepluggedintothewallorrunonbatterypower.Thecaseforthevacuumshouldalsocontain:

Powercable,Replacementbattery,Batterycharger,Connectingtubing.

C. ComputerThemajorityofusersprefertousetheinstrumentsoftwareprovidedonthelaptop.NotethatthiscomputerhasnotbeennetworkedtoconnecttotheUNCnetwork.Thisallowsustooperatethesystemwithouttheuseofpasswords,andmaintainsdatasecurity.Thecomputerbagwillcontain: HPlaptop, Powercable, Mouseandmousepad, X-raysafetybinder. D.AuxiliaryEquipmentInadditiontotheessentialcomponents,referencematerialsandotherequipmentmaybecheckedoutasneeded.Theseinclude: PDAandassortedPDAaccessories, Fullyadjustabletripodfortheanalysisoflargeorstationaryobjects, Standardreferencematerialsaspressedpowdersorglass, Additionalfilmwindows, Samplecupsandfilm, Referencebooksandreadings, Externalharddrivefordatabackup.

Page 6: Use of Handheld XRF may2015 - Archaeologyarchaeology.sites.unc.edu/files/2017/03/UNC-Handheld-XRF-Manual.pdf · All documents referenced in this guide are located in folder Basic

3

IV. BASICOPERATIONA. Connecting

TheQuickStartGuidemaybefoundintheanalyzerboxorinBasicInformationFolder.

ConnectingtoComputer:Thepinswithinthecommunicationscableareveryfragile.Withtheinstrumentrestingsecurelyonthetable,lifttheflapofthecommunicationsportontheinstrument.Lineupthereddotonbaseoftheportwiththereddotonthecable,andgentlyinsertuntilyouhearitclick.InserttheUSBendofthecableintotheportlabeledCOMM3.Thesoftwarewillnotrecognizetheinstrumentifitispluggedintoanyotherport.PoweringtheInstrument:Plugtheinstrumentintoapowersourceorinsertachargedbattery.Turnthekeytotheright.Theyellowstatuslightshouldilluminate.Thislightindicatesthattheinstrumentispoweredon,butisNOTcreatingX-rays.Oncetheinstrumenthasbeenturnedonforadayofuse,leaveiton.Theinstrumentworksbestwhenturnedonregularly(approximately1Xweek),butthelifeoftheX-raysourcewillbeshortenedifturnedonandoffrepeatedly.Placementandsafety:Considerthelayoutofyourspaceandthemovementofpeoplewhensettinguptheinstrumentforlabuse.YouwanttominimizethepotentialexposuretoX-rays.ThebeamoftheX-rayshouldalwaysfaceawayfromyourselfandothers.Maintainthelaptopandothertoolsatasafedistance(seeUserGuidepp.7;1footisideal)fromthebaseoftheinstrumentwhileX-raysarebeingproduced.Ifpossible,setupinstrumentsothatthebeamisdirectedatawall(Figure1).Usetheinstrumentwiththeincludedsampletableandsafetyshieldforsmallobjects.Thesampletableisplacedonthenoseoftheinstrumenttoprovideastablesurfaceforsamples.Smallsamplesmayfitunderthesafetycap.Thiscapblocksmost,butnotallX-raysfromescaping.Itshouldbeplacedsothatthebrassflangeisdirectlyovertheinfraredproximitysensor.Ifthesampleistoolargetofitunderthesafetycap,thesamplemustbeplaceddirectlyoverboththedetectorwindowandtheinfraredproximitysensor.Whileitispossibletoblocktheinfraredproximitysensorwithapieceofopaquetapeorpaperinsteadofthecap,thisisnotrecommended,asitallowsforincreaseddispersionofX-rays.Operatorswhochoosetocircumventthissafetymeasuredosoattheirownrisk.

Page 7: Use of Handheld XRF may2015 - Archaeologyarchaeology.sites.unc.edu/files/2017/03/UNC-Handheld-XRF-Manual.pdf · All documents referenced in this guide are located in folder Basic

4

Figure1.Typicalanalyzersetup.TheX-raybeamisdirectedtowardsasolidwall.ThemirrortotheleftallowstheoperatortochecktheX-raylight.Laptopislocatedonefootfromthebaseoftheanalyzer.

B. S1PXRF

S1PXRFistheprimarysoftwareforcollectingandanalyzingXRFdata.Thisprogramisusedtoestablishinstrumentsettings,viewandexportcollecteddata,andanumberofotheractions,detailedbelow.Oncetheanalyzerhasbeenconnectedtothecomputerandpoweredon,openS1PXRF.Itwillautomaticallyopendisplayingthelastspectrumthatwasgenerated.ThecontentsoftheS1PXRFwindowareshownbelow(Figure2).

a)Start/Stopindicator.OncetheTracerinstrumentisconnectedtothecomputer,andkeyhasbeenturnedtoon,clickontheredbutton.Ifasuccessfulconnectionismade,thelightwillturngreenandtheStartbuttonwillappear.b)Thenameofthecurrentlyopenfileappearsinthefilefield.Whenfirstopened,thewindowwillshowthelastsavedoropenedfile.TheseriesofbuttonstotherightoftheStart/Stopindicatorexpands,shrinks,compresses,andbroadensthespectrumforeasierviewing.Thehouseiconrestoresthespectrumtoitsoriginaldimensions.Thedropdownwindowremovesoraddsthebackgroundorforegroundspectra,ifoverlayhasbeenenabled(seeSpectrumOverlaybelow).c)WhentheTracerinstrumentisonandconnectedtoS1PXRF,theinstrument’scurrenttubesettingswillappear.d)Thetubesettingsusedtorecordthefileonscreen(seebabove)willappearinparentheses.

Page 8: Use of Handheld XRF may2015 - Archaeologyarchaeology.sites.unc.edu/files/2017/03/UNC-Handheld-XRF-Manual.pdf · All documents referenced in this guide are located in folder Basic

5

Figure2.S1PXRFwindow.

e)Whenafileisopen,orasampleisbeinganalyzed,spectralpeaksappear.f)Thedatabarprovidesinformationaboutthespectrum.Thereadingsaresometimesdependentuponwhereonthespectrumthecursorhasbeenplaced.g)Thecountrateshowstheamplitudeatthepointwherethecursorhasbeenplaced.h)Thedetectortemperature(inCelsius)isdisplayed.Thissettingshouldnotchange.Ifitdoeschange,theinstrumentmaybedamaged.TurnofftheTracerandcontactBrukerimmediately.i)Ambienttemperature(inFahrenheit)showsthetemperatureinternaltotheTracer.Althoughvariable,thetemperaturegenerallyshouldbeinarangebetween70°Fto115°F.j)KeVisthevoltageatcursorlocation.Thereadingwillchangedependinguponthelocationofthecursor.k)LSecisthelengthoftimethesamplewasanalyzed.Ifshootingasample,thereadingistherealtimelengthofanalysis.Whencollectingatimedassay(seepage7),thelabelwillchangetoRSecandthetimerwillcountdowntozero.l)Thex-axisrecordsenergy,measuredinKeV.m)VacshowsthepressureinTorrwithinthenoseoftheinstrument(betweenwindowanddetector).Whenundervacuum,thevalueshouldbeunder10.n)Rawandvalidcountratesarerecorded.Generally,theratioshouldbelessthan3:2(raw:valid).o)Cursorlocation.Placingthearrowinthespectraldisplayandclickingtheleftmousebuttonwillmovethecursortothatlocation.Inaddition,usingthe←→keysonthekeyboard,thelocationmaybemovedonechannelatatime.

Page 9: Use of Handheld XRF may2015 - Archaeologyarchaeology.sites.unc.edu/files/2017/03/UNC-Handheld-XRF-Manual.pdf · All documents referenced in this guide are located in folder Basic

6

i) Settings

FollowtheinstructionsintheUserGuideorQuickStartGuidetosetupforanalysis:

1. Placesampletableandsafetycaponinstrumentnose.ThiscontainstheX-rays thatareproducedatseveralpointsduringsetup. 2. OpenDownloadTab.EnsureBaudrateissetto115200, EnsurePortissettoCOMM3.

3. Clickonreddotinupperleft,toturnitgreen. 4. OpenSetupTab.EnsureBackScatterboxisunchecked. EnsurePCTriggerboxischecked.

Theguidetoinstrumentsettingsforanalysisofspecificmatricesand/orelementsiscalledFilterandVoltageandCurrentSelection.Yourresearchdesignandsamplematrixwillaffectwhichsettingsyouchoose.Ifyouplantoquantifyyourdatausinganexistingcalibration,makesurethatyoursettingsmatchthecalibrationsettings,asdetailedinSectionIV.Ifusingafilter,unscrewthebrassnutontheanalyzernoseandinsertfilter,withtheovalsidefacingintotheinstrument.Theinstrumentdoesnotneedtobeturnedoffforthisprocedure.BEFOREUSE,ALWAYSCHECKFORTHEPRESENCEOFAFILTER.

Ifusingavacuum,setthevacuumupright.Connectthetubingtothepumpandinstrument,payingattentiontothelabeledends.Switchpumpon,andensurethatthepressurereachesbelow10Torr.Youwillseethewindowonthenosesuckdownintotheinstrument.Thepressuremayfluctuateabitoverthecourseoftheday,butshouldalwayshoverbelow10Torr.Monitorthroughouttheday,andreplacethenosewindowifnecessary.

1. InS1PXRF,OpenTubeTab. 2. SelectKTITube. 3. SelectRead. 4. Selecttheappropriatevoltageandcurrentasindicatedintheguideabove. Pulselengthandfilter#donotneedtobeadjusted.Voltage:theenergyoftheX-raysproducedbytheinstrumentisgivenaskiloelectronvolts(kVorkeV).Adjustthevoltagesettingstofocusonparticularpartsofthespectrum.Ingeneral,asatomicmassincreases,sodoesthevoltageneededtoexcitethatelement.Forexample,uselowvoltagewhenfocusingonmajor(low-Z)elements.Current:theflowofelectronchargeisgivenasmicroamps(µa).Ageneralruleofthumbisthathighdensitymaterials,suchasmetals,uselowcurrent.Lowdensitymatrices,suchasceramicsandglass,usehighcurrent.

Page 10: Use of Handheld XRF may2015 - Archaeologyarchaeology.sites.unc.edu/files/2017/03/UNC-Handheld-XRF-Manual.pdf · All documents referenced in this guide are located in folder Basic

7

Inmostcases,thepresetoptionsinS1PXRFwillmeetyourneedsforvoltageandcurrent.Ifyoumustcreatecustomsettings,thisisdonethroughXRayOps(seebelow).Ensurethatsafetycapisonthenose.ChecktheboxforPCtrigger.Youshouldheartheclickoftheanalyzerandseetheredlight,indicationsthatX-raysarebeingproduced.WhenyouclosetheKTITubewindow,X-raysshouldstopbeingproducedandtheredlightwillgooff.Theinstrumentisnowreadytocollectdata.

ii) XRayOps:

ThisprogramallowsyoutomanuallycontrolallthesettingsintheS1PXRF.Inmostcases,thepresetoptionsinS1PXRFwillmeetyourneedsforvoltageandcurrent.OperatethisprogramwithEXTREMECAUTION,asitwilloverwriteexistingpresetsinS1PXRF.

1. BeforeopeningXRayOps,ensurethatS1PXRFisclosed. 2. SelectCOMMPort3. 3. Checkboxforhighspeedcommunications. 4. CheckboxforOpenX-rayCommPort. Theprogramisnowabletoreadthesettingsfromthemotherboardinthe analyzeritself. 5. Takeascreenshotofthedefaultsettingsandsavewithdescriptivetitle. 6. Overwriteasettingrowwiththedesiredsettings.Savesettingsundernewfile name. 7. CloseXRayOpsandreopenS1PXRF. 8. Selectyourcustomvoltageandcurrentsettings. 9. WHENYOURANALYSISISCOMPLETE,RETURNTOXRAYOPSANDRESTORE DEFAULTSETTINGS.

Thisinstrumentisequippedwitha5-wattbulb.Voltagetimescurrent=watts(ex:40keVx30µa=1.2watts).Thecurrentandvoltageproductmustnotexceed5watts.Ingeneral,ifusinghighvoltage,uselowcurrent;lowvoltage,highcurrent.

iii) AnalyzingSamples

Placeyoursampleoverthewindow.Makeasclosecontactasyoucan.EnsurethatIRsensorisblockedbythesampleorthesafetycap.SeeSectionV.forselectingappropriatesamples.Inmostinstances,youwillwanttodoatimedassay.Thiswillensurethatallofyourmeasurementsinastudyareofaconsistentlength.Ifyouwishtoquantifyyourresultsintermsorparts-per-millionorweight%,ensurethatyourtimingandallothersettingsmatchthatofthestandardsusedtocreatethecalibration.FulldetailsoncreatingandusingcalibrationsmaybefoundinSectionIV.Ingeneral,asthe

Page 11: Use of Handheld XRF may2015 - Archaeologyarchaeology.sites.unc.edu/files/2017/03/UNC-Handheld-XRF-Manual.pdf · All documents referenced in this guide are located in folder Basic

8

heterogeneityofsamplesincreases,intermsofelementalcompositionandmatrixconsistency,soshouldruntime.AlwaysselectAutosave.Otherwise,youmayinadvertentlyoverwritefiles.Alwayscreate.PDZfiles.Dependingonthetypeofanalysisyouwishtodo,youmayalsocreate.TXTand.CSVfilessimultaneously.Notethatyoucanalsocreatethesefiletypeslater,byopeningaspectrumfile(.PDZ)andselectingFileàSaveAs.Chooseyourfilenamescarefully.Theyshouldincludeacodeforyoursettings,sothattheyareimmediatelydistinguishablefromotheranalyses.Forexample,ifyouarerunningseparateanalysesformajorelementsandtraceelementsonthesamesamples,ensurethatthefilenamesaredistinct.Additionally,itisgoodpracticetoputfilesforeachtypeofanalysisinaseparatefolder.Theinstrumentsettingsarestoredasmetadataineach.PDZfile;however,notethatthefilterinformationmaynotbecorrect.Theinstrumentcan’ttellwhichfilteryou’reusing-thatfieldisautomaticallypopulatedbasedontheKTITubesettings.Onceyouselectsave,thecountdownwillbeginandX-rayswillbeproduced.Ifyouhaveselectedmultipleruns,theywillautomaticallyrun-withfilenamesindexed,e.g.,Sample1-1,Sample1-2,Sample1-3.Whilesamplesarebeingrun,itisimportanttomonitorrawcountratesandvalidcountrates.Thesevaluesappearinthedatabaratthebottomofthewindow.RawCountRate-thecountofphotonsdetectedpersecond.Forthisinstrument,thisvalueshouldnotexceed200,000.ValidCount-theupperlimitofdetection.Thisvalueshouldnotexceed100,000.Ifeithercountrateexceedstherecommendedthresholds,thisisanindicationthatthedetectorisbeing“blinded”bytoomanyphotons.Adjustthecountratebyloweringthecurrentsettings.Whenusingafilter,thecountrateswilloftenbeinthelowthousands.

iv) InterpretingSpectra

S1PXRFdisplaystheresultsofasaspectrum.Thex-axisisenergy,representedaskiloelectronvolts(keV).They-axisisphotoncounts,orintensity.Thespectrumlinethereforedisplaysthecountsofphotonsateachenergylevel.WhenexcitedbyX-rays,eachelementproducescharacteristicsecondaryX-rays.TheirenergyisnotedonthePeriodicTableandX-rayEnergieschart.Identifyingelementalpeaks:ThecharacteristicX-raysforeachelementwillproducepeaksatseveralspecificenergylevels.Thesedependonwhichelectronshellejectsanelectron,andwhichshellgivesupanelectrontofillthespace(Figure3).ElementsuptoandincludingbariumwillshowK-alpha(denotedasKa1)andK-betapeaks.AK-alphapeakrepresentsphotonsproducedwhenelectronsmovefromtheL-shellintotheK-shell.AK-betapeakrepresentsphotonsproducedwhenelectronsmovefromtheM-

Page 12: Use of Handheld XRF may2015 - Archaeologyarchaeology.sites.unc.edu/files/2017/03/UNC-Handheld-XRF-Manual.pdf · All documents referenced in this guide are located in folder Basic

9

shellintotheK-shell.ElementsheavierthanbariumwillshowL-shellpeaks,whiletheheaviestelementswillshowM-shellpeaks.BariumandsomemetalswillfluorescefromboththeirK-shellandL-shell.Wherethereisspectralinterference(peakoverlap),thissoftwarecannotdifferentiatebetweentheoverlappingelementalpeaks.Theusermustfamiliarizethemselveswiththeexpectedelementsintheirsamples,orusetheArtaxprogramoutlinedbelowtorunprobabilisticassessmentsofwhichelementsarepresent.

Figure3.Bohrmodelofsiliconatomwithlabeledelectronshells.

1. Inmaintoolbar,SelectID. 2. SelectElem,whichwillpopupaperiodictable. 3. Selecttheshellyouwishtosearch(K,L,M). 4. Selectanelementtoseethelocationofitsx-rayenergy. 5. Toleaveamarkerforthatelement,selectAdd.Itwilllabelonlythealphapeak. 6. Z+andZ-buttonswillmoveelementbyelementupordowntheperiodictable.

Dependingonyoursample,theremaybeseveralpeaksthatareunabletobeidentifiedbyS1PXRF.TheseincludeComptonPeaks,SumPeaks,andEscapePeaks.Itisimportanttorecognizethesepeaks,sothatyoudonotmisidentifythemasanearbyelementalpeak.ComptonPeak-Aspectralpeakthatreflectsthepresenceoflightelementswithinasample.Whileelementssuchascarbonandoxygencannotbemeasureddirectlybytheinstrument,theComptonpeakcanbeviewedasaproxyforcarbondensitywithinasample.Thehigherthepeak,themorecarbon.ThelocationoftheComptonpeakwillvary,dependingonthefiltersettings.Withafilterforlighterelements(low-Z),thepeakwillbebetween2.5-3keV.Underunfilteredconditions,orfilteringforhigh-zelements,theComptonpeakwillfallbetween18.5-19.5.Thispeaktypicallylooksdistinctfromelementalpeaks,beingwiderandflatter,sometimeswithaslopingshoulder.Sumpeaksformwhenanelementisabundantwithinasample.Whentwophotonsfromthatelementreachthedetectorattheexactsametime,thedetectorgivesthemacountofone,andanenergylevelthatisthesumoftheirenergies.Thiswillcauseapeaktoforminthespectrumthatislocatedat2xthecharacteristicenergyoftheabundantelement.Forexample,inasamplewithabundantiron,theK-alphapeakforironwillbelocatedat6.405KeV,andthesumpeakwillbelocatedat12.91KeV.

K" L" M"

Page 13: Use of Handheld XRF may2015 - Archaeologyarchaeology.sites.unc.edu/files/2017/03/UNC-Handheld-XRF-Manual.pdf · All documents referenced in this guide are located in folder Basic

10

Escapepeaksalsoformwhenanelementisabundantwithinasample.Thesepeaksformduetotheinterferenceofsiliconfluorescenceatthesurfaceofthedetector.Thepeaksarelocatedonthespectrumat:elementkeV-siliconkeV(1.74keV).Forexample,inasamplehighincopper,theK-alphapeakforcopperwillbelocatedat8.046keV,andthecopperescapepeakwillbelocatedat6.306keV(8.046-1.74).Thiscopperescapepeakisespeciallyimportanttoremember,becausethepeakoverlapswiththeironK-alphapeak,andcouldbemisidentified. v)SpectrumOverlayInS1PXRFitispossibletoviewtwospectraatthesametime.

1. Openorcollectdataononeofthespectrayouwishtooverlay. 2. Fromthemaintoolbar,selectSetup. 3. SelectSpectrumoverlay.Anewboxwillpopup. 4. MoveAtoB.Thecurrentspectrumwillnowbeoutlinedinblue. 5. Openorcollectdataonnewspectrum.Itwillappearoverlaidinredoutline. 6. Thefilewindowwillnowshowthenamesofbothfiles. 7. ClearSpectrumBorMoveAtoBasrequiredforsubsequentanalyses. 8. Fromthedropdownmenutotheleftofthefilewindowyoucanchangewhich fileorfilesyouareviewing.

C. ARTAX

Artaxisaprogramthatallowsforthevisualizationofmultiplespectra,manipulationsofspectra,andreturnscorrectedphotoncounts,whichcanbeusedforquantitativeanalyses.

i) Installation

ArtaxisinstalledontheXRFlaptop.However,ifyouwishtoconductanalysisonyourownPC,copythefolderArtaxfromfolderBasicInformation,andfollowthesedirections.

1. Copythefolder‘Artax’directlyintoyourC:drive. 2. Makeyourcomputershowinvisiblefolders.Thiswillvaryaccordingtoyour operatingsystem.ForWindows7:StartMenuàControlPanelà AppearanceàFolderOptionsàShowHiddenfilesandfolders. 3. CopytheBruker-AXSfolderdirectlyintotheProgramDatafolderonC:drive. 4. Within‘Artax’folder,select‘ArtaxBest’andthendouble-clickonArtax.exeto opentheprogram.Youmaywishtopinthisprogramtothetaskbarorcreatea shortcutontheDesktop.

Page 14: Use of Handheld XRF may2015 - Archaeologyarchaeology.sites.unc.edu/files/2017/03/UNC-Handheld-XRF-Manual.pdf · All documents referenced in this guide are located in folder Basic

11

ii) CreatingaProjectinArtax

Ifyouhavealreadycreatedandsavedaprojectwithamethodandelementalpeaksidentified,youcanskipthefollowingsteps,andsimplyreopenyourexistingfile.

1. OpenArtax.Thereisnopassword.Clickthroughtheinitialerrormessages. 2. UnderProjectàNewProject.Adirectorywillthenopenontheleftsideof screen. 3. RightclickonfolderlabeledObjectsàAddNode. 4. Namethisnode:Points.ThenameforthisnodemuststartwithPoints,but anythingelsemayfollow. 5. FileàAddSpectrum.Navigatetoyourspectraandselectalltoimport. 6. RightclickonPointsfolderàAddSpectra 7. Inmainwindow,clickonSpectrumtab.Allofyourspectrashouldappear. 8. FileàSaveProjectAs.Selectanameanddestinationforyourproject.

iii) ExtractingNetPhotonCounts

ThespectrathataregeneratedbytheXRFaredirectlybasedonthecountsofphotonsatparticularenergylevels,asmeasuredbytheX-raydetector.WithinS1PXRFitispossibletoretrievethesecountsateachenergylevel.However,theseinitialvaluesdonotaccountforinter-elementaleffects(whenspectralpeaksoverlap).Forexample,ananalysisofrawphotoncountswouldlikelymisidentifyanumberofironandcobaltpeaksiftheyarebothpresentinthesample,asthepeaksforironandcobaltoverlap.UsingARTAX,itispossibletoovercomesomeoftheseconfoundingeffects,withtheBayesianDeconvolutionprocess.TheresultswillbegivenasNetIntensitiesorNetPhotons,whicharecountsoftheareaundereachspectralpeak,minusbackground,minusinter-elementaleffects.Dependingupontheelementspresent,netphotoncountsshouldbelinearlycorrelatedtoparts-per-millionorweight%values.BayesianDeconvolution-aprobabilisticmethodforassessingthelikelihoodthataparticularelementispresentwithinthesample.WhilethisisanautomatedfunctionofArtax,itisaniterativeprocedurethatreliesupontheuser’sknowledgeandjudgmentoftheirsamples. 1. FindtheComptonPeak.Thispeakisnotrelatedtoasingleelement,ratheritis thereflectionofmultiplelightelements.Ittypicallydoesnotlooklikeother peaksinthespectrum.Itmaybewider,orconsistofmorethanonepeak.For spectrathatrangebetween1-40KeV,theComptonPeakwillbepresentat around19KeV. 2. ClickontheapproximatecenteroftheComptonPeak.Averticallinewill appear.ThenselectSpectrumàNormalizefromthetopmenu.Thespectrawill becomelessfuzzy. 3. Clickthe‘X’totherightofspectrumnameintoolbar.Allwilldisappear.Reopen onespectrumbyselectingitfromtheleftmenu.

Page 15: Use of Handheld XRF may2015 - Archaeologyarchaeology.sites.unc.edu/files/2017/03/UNC-Handheld-XRF-Manual.pdf · All documents referenced in this guide are located in folder Basic

12

4. OpentheMethodEditor.Itislocatedtotheleftofthetextboxintoolbar.It lookslikealittleboxwithredandgreendots.Youmayalsofinditbynavigating toMeasurementàMethod. 5. Typeanewnameintothetoptextbox. 6. DonotchangetheMeasurementTab. 7. InCorrections,ensurethatEscapeandBackgroundarechecked.Cycles=10.For mostspectra,Min/Maxwillrangebetween1-40KeV.Ifyou’reunsure.Lookat yourspectraandseetheminandmaxontheX-axis. 8. ClickAddonthetop,andthenOkonthebottomofthiswindow. 9. NowthenameofyourmethodshouldappeartotherightoftheMethodEditor iconintoolbar. 10. OpenthePeriodicTable.ItisjusttotherightoftheMethodnameintoolbar. Thisisaniconwithseriesofverticallines. 11. Clickinthecenterofalargepeakonyourspectrum.Youwillseethatatleast oneoptionappearstotherightoftheperiodictable.Choosetheappropriate element.Ingeneral,thiswillbethemorecommonelementinthelist. 12. ClickontheBayesianDeconvolutionbutton,locatedtotherightofthePeriodic Tablebutton.Abluefitlinewillnowappear.Ensurethatthelinematchesup fairlywellwithyouridentifiedelement.TheBayesiandeconvolutionisa probabilisticmeasureofelementalidentification. 13. Goalongyourspectrumandidentifyeachpeak/setofpeaks.RunBayesian deconvolutionafteraddingeachelement-ensurethatthefitline improves/matches. 14. Onceyouthinkyouhavefoundalloftheelements,reopenMethodEditor. 15. UnderIdentificationtab,selectPresetList.ThePeriodicTablewillappear,as wellasabutton‘GetElements’.Clickthisbuttonandthetablewillpopulate withtheelementsyou’veselected. 16. SelectReplaceandOKinMethodEditorMenu. 17. Closethisspectrumandopenanother.Rundeconvolution.Theelementsshould auto-identify,andthefitlineshouldfitwell.Ifitdoesnot,lookforelementsin thisspectrumthatwerenotidentifiedintheprevious.FollowSteps14-16to addelementstoMethod. 18. RepeatStep17severaltimes,untilyou’reconfidentthatyourmethodaccounts forallelementsinthesamples.

Page 16: Use of Handheld XRF may2015 - Archaeologyarchaeology.sites.unc.edu/files/2017/03/UNC-Handheld-XRF-Manual.pdf · All documents referenced in this guide are located in folder Basic

13

iv) ExportingData

1. Toexportyourdataasnetphotoncounts(aka‘netintensities’),clickonPoints folder,thenAnalyzeàEvaluateResults. 2. WhenEvaluationiscomplete,clickonPointsfolder,thenExportàExport ResultstoExcel.Chooseafilenameanddestination. 3. TherewillbeanerrorwhenyougotoopenthisfileinExcel,becausefilewas notcreatedinExcel.Ignorethiserror.Theworkbookwillhavetwosheets:The firstwilllistyourinstrumentsettingsandthesecondwillhavenetintensitiesfor eachelement. 4. Youmayusethesenumbersinsimilarwaystoppm/wt%.Asageneralrule, countsbelow100photonsareconsideredzero.TheBayesiandeconvolutionhas removedinter-elementaleffects(i.e.,accountedforoverlappingpeaksof certainelements)

v) ArtaxVisuals

1. Tozoomintoyourspectrum:holddownCTRL,clickandscrollinmainwindow. 2. Toundozoom,double-clickinmainwindow. 3. Topanacrossspectrum,movecursorbelowX-axis,clickandscroll. 4. OptionsàDisplayPropertieswillallowyoutochangethebackground,remove axes,etc.Tocreateunderfill,OptionsàFilledSpectrum. 5. Toremovethelinesmarkingelementalpeaks:PeriodicTableàUncheck‘Lines.’ 6. TochangethecolorofanElementlabel,openperiodictable,dragthedesired coloroverthedesiredelement. 7. Tochangethecolorofspectrumlineorfill,selectthespectrumfromthe dropdownlistintaskbar.Clickonthecolorboxtoleftofname,andselect desiredcolor. 8. TocopythespectrumintoadocumentExport->CopyGraphic(Ctrl-C).

V. QUANTIFICATION

A. Calibrations

Whilevisualcomparisonsornetphotoncountsmaybesufficientinavarietyofinstances,obtainingquantitativedataasparts-per-millionorweightpercent,isoftendesired.Brukerhasdevelopedanumberofcalibrationsspecifictocertaintypesofmatrices:obsidian,copperalloys,goldalloys,mudrock,andothers.Thesecalibrationswerecreatedbyrunninganumberofstandards,areferencedataset,onthisinstrument,andcalculatingbest-fitmodelsthatcanbeappliedtosamplesofunknowncomposition.

Page 17: Use of Handheld XRF may2015 - Archaeologyarchaeology.sites.unc.edu/files/2017/03/UNC-Handheld-XRF-Manual.pdf · All documents referenced in this guide are located in folder Basic

14

Allcalibrationshavebeenbuiltfromanalysescompletedonthisparticularinstrument(T3S2448).Theinternalgeometryofeachinstrumentisslightlydifferent.YoucannotreliablyuseacalibrationdevelopedonanotherBrukeranalyzer.

Foreachtypeofmatrixthereisaseparatereferencedataset,calibrationworksheetandcalibrationfile.Onemayuseaworksheetas-is,oraddandsubtractstandardsinthedatasetinordertoobtainacalibrationthatbestmatchesthesamplesyouintendtoquantify.Itisfundamentaltothecalibrationprocessthatthecalibrationdatasethaselementsinthesamerangeofconcentrationsasthesamplesyouintendtorun.Themodelisnotabletoaccuratelypredicttheconcentrationofelementsthatexceedtheknownquantitiesinthereferencedataset.TheBrukercalibrationdataandworksheetsarefoundinfolderT3S2448onthedesktop.NotethattheExcelfilesusedtocreatethecalibrationsandtoproducequantitativedatahaveanExcelmacrocalledS1CalProcess,whichmustbeenabledinordertofunction.ToinstalltheExcelmacro,seetheinstructionsinBrukerCalProcessServiceBulletininS1CalProcessfolder.Forceramics,usethecalibrationcreatedbyJeffSpeakmanandAliceHuntattheCenterforIsotopeStudies,UniversityofGeorgia(HuntandSpeakman2015).Thiscalibrationreplacesthefactory-installedmudrockcalibrationadvocatedbyBruker.Ifusingthesedatainpublication,propercitationofthecalibrationmustbeincluded.ThecalibrationfilesmaybefoundintheUGACeramicsCalibrationfolderontheDesktop.TheyarebuiltfromtheBrukerS1CalProcessmacro.SpeakmanandHunthavetweakedthestandardsettingsusedtocollectdataforceramics.Whenusingthisceramicscalibration,followthesettingsgiveninAppendix.

B. UsinganExistingCalibration 1. OpentheExcelspreadsheetforthecalibration.Resaveitwithanewnamein yourdatafolder.NEVERedittheoriginalfile!Eachfilehasworksheetsthat createcalibrationsformajorelementsaswellastrace.

2. Checkthesettingsusedtocollectvaluesforthereferencematerial,foundinthe PDZFilestab.Inordertouseacalibration,everythingmustmatch:Timeof Assay,Voltage,Current,Vacuum,Filter.Ifcollectingdataonmajorandtrace elements,youwillneedtoruneachsampletwice,matchingthemajorandtrace settingsforthecalibrationspectra. 3. Collectdataonyoursamples. 4. OpenthereferencestandardsfilesinArtax.Identifyallpeaks,usingBayesian deconvolutiontoconfirm.Makealistofallelements-besuretospecifywhich peakisidentified(K-alpha,L-alpha) 5. OpenspectraforyoursamplesinArtax.Identifyallpeaks,usingBayesian deconvolutiontoconfirm.Makealistofallelements.Ideallyalloftheelements forreferenceandsampleswillalign.Thisisthegoalofreferenceset.Notethat

Page 18: Use of Handheld XRF may2015 - Archaeologyarchaeology.sites.unc.edu/files/2017/03/UNC-Handheld-XRF-Manual.pdf · All documents referenced in this guide are located in folder Basic

15

ifthereisanelementpresentinyoursamplesthatisnotinyourreferenceset, youwillnotbeabletomeasurethatelement.Ifyouneedtoaddadditional standards,theprocedureisfoundinthefollowingsection. 6. Ifthereferencesetandyoursamplesareinagreement,you’rereadyto quantify.ThiscanbedonefilebyfileinS1PXRF,orbatchedwithinthe calibrationspreadsheet.

i)S1PXRFcalibration:

1. OpendesiredspectruminS1PXRF.2. SetupàSelectCoef(coefficient)3. Findtheappropriatecalibrationfile.Factory-installedcalibrationsarefoundin

thefolderT3S2448onthedesktop.Thecalibrationwillhavethefileextension .CFZ.Youwillnowseethefilenameappearsingrayonthemenubar,nexttoa CONCbutton.

4. HitCONC.Anewwindowwillappearwiththevaluesofyoursampleinthesame formatastheywereenteredforthecalibrationspectraintheExcelfile.For example,amajorelementcalibrationwilllikelypresentmajorelementsasoxide weightpercent,whichisgeologystandardpractice.Youcancheckontheformat ofthecalibrationspectraonthefarrightoftheMaChem(orTrChem)tabofthe calibrationworksheet.

5. SelectCopyinordertopasteyourresultsintoanewsheet.6. Ifyouwishtolookatbothtraceandmajorelements,youwillneedtochange

theCOEFfileundersetupandrepeatwithyournewspectrum.

ii)S1CalProcesscalibration:

APDFentitledMassCalibrationwithstep-by-stepinstructionsandscreenshotsisinCalibrationsfolderinBasicInformation,aswellasaproselistofinstructionsentitledS1CalProcessQuantification.

C.AddingorRemovingCalibrationReferenceMaterial

1. Collectdataonyournewreferencematerial.Thedatamustbecollectedbythe sameinstrumentonwhichyoursampleswillberun.Youcannotuseanother instrument’scalibrationspectra,asthegeometryofeachdetectorisdifferent andwillreturndifferentvalues.2. Openthebasecalibrationfileforyourmatrixtypeandresavewithanewname.3. OnPDZtab,UnderAdd-InsàS1CalProcessàInsertPDZ.AllyourPDZfilesmust beinthesamefolder.5. ChooseeitherMajorElement(MAChem)orTraceElement(TRchem) worksheet,dependingonwhatsettingyouused.

Page 19: Use of Handheld XRF may2015 - Archaeologyarchaeology.sites.unc.edu/files/2017/03/UNC-Handheld-XRF-Manual.pdf · All documents referenced in this guide are located in folder Basic

16

6. InAdd-InsàS1CalProcessàSelect/Removeelementsfromcalibration.Addall elementsthatwereidentifiedinArtaxFORBOTHREFERENCEMATERIALAND SAMPLES.Removeanypreexistingelementsinthelistthatarenotpresentin yourmaterials.Notethatnotallelementscanbeadded(e.g.,gallium).Keep rhodiuminthecalibrationatthispoint.7. InAdd-InsàS1CalProcessàAdd/Removechemfiles.ADDFILESFIRST,anddo notclick“AddNewChemistry”.Thenremoveallunwantedchemfiles.8. YourChemsheetwillnowhavealistofthereferencematerials,butthevalues willbeidentical.THISISOKAY-IGNOREALLVALUESBUTTHOSEONFARRIGHT OFSHEET.Pasteinthereferencevaluesforyourreferencematerialhere.You canuseeitherppmorwt%.9. InAdd-InsàS1CalProcessàCreateChemFile.MakesureComptonis19.5-22if inatracemode.DefaultforMAchemis2.5-3kev.10. Keeptheslopecorrectionsandincludedelements.Gothroughandremovea referencematerialfromanelementifithasazerovalue,oraveryhigh difference.Avg.differenceshouldbelessthan5,butideally0foreachelement.

C. TestingyourCalibration

1. FollowthedirectionsabovetobatchcreateconcentrationsinS1CalProcess,or individuallythroughS1PXF.Addthesevaluestoanewspreadsheet.2. Checkfornegativenumbers.Inmostcases,converttozero.3. Throwoutanyelementswithnomeasuredresults.Forinstance,galliumwas visibleinthespectrainArtax,buttherearenoreferencevaluesinthereference set.Thecalibrationreturned0gallium.4. Ifyouhavezerosinyourdata,addaconstant.Inthiscase-sincetherange betweenhighandlowvalueswasverygreat,Iused0.0001wt%(1ppm).5. Log-transformthedata.6. Plottheexpectedvaluesagainstreferencevaluesforeachelement.Ifthereis referencematerialthatwasnotincludedinthecalibrationforaparticular element,youwanttoremoveitfromtheplotdata.Eachelementshouldshowa positivelinearcorrelation,ideallywithr2valueof1orcloseto1.Elementsthat showpoorcorrelationshouldnotbeincludedinconcentrationsofunknowns- theyshouldbeleftinthecalibration,butvaluesnotrecorded.

VI. SELECTINGAPPROPRIATESAMPLES

WhendevelopingaresearchprojectwiththehandheldXRFanalyzer,therearemanyconsiderationstokeepinmindregardingyoursamples.

Page 20: Use of Handheld XRF may2015 - Archaeologyarchaeology.sites.unc.edu/files/2017/03/UNC-Handheld-XRF-Manual.pdf · All documents referenced in this guide are located in folder Basic

17

1. Yoursampleshouldbeflatenoughtorestfirmlyagainsttheinstrument aperture.Anyairspacebetweenasampleandthedetectorwillattenuatethe signal.ThisisespeciallyimportantformeasuringthelighterZelementsusinga vacuumorhelium.2. YoursampleshouldbeInfinitelyThick.IntermsofX-rayfluorescence,infinitely thickmeansthatasampleisthickenoughthattheX-rayspenetratewithout passingallthewaythroughthesample.Thedepthofpenetrationisreferredto astheMassAttenuationCoefficient.Thedepthwillvarydependingonthe matrix,andtheinstrumentsettings.Foranalysisoftraceelementsinsilicates, approximately3.5mmorgreaterisconsideredinfinitelythick.Foramore completediscussionofmassattenuation,seePhysicsfolderinBasic Information.3. Surfaceconditionmatters.WhiletheX-raysdopenetrateintothesample, surfacecontaminantswillmuddythesignal.Thisisespeciallyaconcernfor porousitemslikeceramics.Ifpossible,useafreshlybrokensurfaceorabradea surfacewithatungstencarbideorsiliconcarbidebittorevealafreshsurface withoutsurfacecontaminants,paints,glazes,etc.Rinsewithdeionizedwater andallowtoairdry.4. Bemindfulofsharppointsoredges.Thewindowcoveringthenoseofthe instrumentisverythin,andtearseasily.Whileabrokenwindowcancontinueto beusedwithoutavacuum,itwillhavetobereplacedforvacuumorhelium flow.Windowsareexpensivetoreplace.5. Powderedsamplesshouldbeashomogenousanddenselypackedaspossible. Ideallyallpowdersshouldbepressedintopellets.Samplecupswithovendried andfirmlypackedpowdermaybesubstituted,butthesignalwillbe substantiallyweakerduetotheinterferenceofthesamplecupfilmandthe largerporespaceofthesample.

Page 21: Use of Handheld XRF may2015 - Archaeologyarchaeology.sites.unc.edu/files/2017/03/UNC-Handheld-XRF-Manual.pdf · All documents referenced in this guide are located in folder Basic

18

Appendix.UniversityofGeorgiaCeramicCalibrationSettingsForTraceElements(Mn-Nb): GreenFilter NoVacuum Voltage:40keV Current:30μa SampleRunTime:200seconds SampleNamePrefix:TRForMajorElements(Al-Zn): NoFilter Vacuum Voltage:15keV Current:25μa SampleRunTime:200seconds SampleNamePrefix:VForMajorElements(Al-Zn)+Mg: NoFilter HeliumFlow* Voltage:15keV Current:25μa SampleRunTime:200seconds SampleNamePrefix:HeCreateaheliumflowformeasuringmagnesiumbyattachingaheliumtank. 1. Carefullypuncturethewindowfilmonthenosewithoutdamagingthemetal screen. 2. Attachtygontubing(orotherflexible,non-contaminatingtubing)tothehelium tank. 3. Openheliumflow.Keepflowprettylow.Youshouldnotbeabletohearit,and barelyfeelit. 4. Plugtheotherendofthetubingdirectlyontothevacuumportoninstrument. 5. Operatenormally.