IMPORTANT PRODUCT UPGRADE NOTICE January 2014 Dear Valued Customer, The FIT4Safety Canada Recommendations for Best Practice in the Safe Use of Diabetes Sharps 1 advocate that all safety engineered devices (SEDs) have the following features: An easy one hand activation A clear view of the tip of the sharp at the entry site A clear indication of the dose delivery A clear indication of the activation of the safety feature No exposure to the sharp following the dose administration Protection from exposure at both ends of the the needle to prevent NSI before and after use As a result of the recommendations we are phasing out of the production of our BD Autoshield™ 5mm and 8mm pen needles in preparation for discontinuing. At this time we are upgrading our current customers to our newest generation safety pen needle, the BD Autoshield™ Duo. BD Autoshield™ Duo offers: Passive protection after use the only pen needle that automatically covers the front and back needles after use Singlehanded injection technique shortest SED that helps reduce the risk of intramuscular injection without the use of a skin lift Flexibility to work with all the injectable pens your facility uses To assist you with the transition from BD Autoshield™ 5mm and 8mm pen needles to BD Autoshield™ Duo, enclosed you will find the instruction for use, product information and useful links. Should you have any questions or concerns please contact your BD Diabetes Care Sales Consultant or BD Customer Service at 18002685430. Please note that this letter is intended for healthcare professionals. Thank you for your continued support. Sincerely, Sarah Christilaw Sarah Christilaw Product Manager Diabetes Care BD Medical 2100 Derry Road West, Suite 100 Mississauga, ON L5N 0B3 Canada tel: 800.268.5430 fax: 905.288.6006 www.bd.com Becton, Dickinson and Company
4

Upgrade Notice 329505 Customer FINAL EN - …images.medimart.com/portal/Advertisement_2014-02-10.pdf · IMPORTANTPRODUCT)UPGRADE)NOTICE January’2014’ ... Becton, Dickinson and

Sep 07, 2018

Download

Documents

lamkhanh
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Upgrade Notice 329505 Customer FINAL EN - …images.medimart.com/portal/Advertisement_2014-02-10.pdf · IMPORTANTPRODUCT)UPGRADE)NOTICE January’2014’ ... Becton, Dickinson and

IMPORTANT  PRODUCT  UPGRADE  NOTICE January  2014    Dear  Valued  Customer,    The  FIT4Safety  Canada  Recommendations  for  Best  Practice  in  the  Safe  Use  of  Diabetes  Sharps1  advocate  that  all  safety  engineered  devices  (SEDs)  have  the  following  features:    

• An  easy  one  hand  activation  • A  clear  view  of  the  tip  of  the  sharp  at  the  entry  site  • A  clear  indication  of  the  dose  delivery  • A  clear  indication  of  the  activation  of  the  safety  feature  • No  exposure  to  the  sharp  following  the  dose  administration      • Protection  from  exposure  at  both  ends  of  the  the  needle  to  prevent  NSI  before  and  after  use  

As  a  result  of  the  recommendations  we  are  phasing  out  of  the  production  of  our  BD  Autoshield™  5mm  and  8mm  pen  needles  in  preparation  for  discontinuing.  At  this  time  we  are  upgrading  our  current  customers  to  our  newest  generation  safety  pen  needle,  the  BD  Autoshield™  Duo.      BD  Autoshield™  Duo  offers:  

• Passive  protection  after  use    -­‐  the  only  pen  needle  that  automatically  covers  the  front  and  back  needles  after  use  

• Single-­‐handed  injection  technique    -­‐  shortest  SED  that  helps  reduce  the  risk  of  intramuscular  injection  without  the  use  of  a  skin  lift  

• Flexibility  to  work  with  all  the  injectable  pens  your  facility  uses    To  assist  you  with  the  transition  from  BD  Autoshield™  5mm  and  8mm  pen  needles  to  BD  Autoshield™  Duo,  enclosed  you  will  find  the  instruction  for  use,  product  information  and  useful  links.    Should  you  have  any  questions  or  concerns  please  contact  your  BD  Diabetes  Care  Sales  Consultant  or  BD  Customer  Service  at  1-­‐800-­‐268-­‐5430.    Please  note  that  this  letter  is  intended  for  healthcare  professionals.  Thank  you  for  your  continued  support.    Sincerely,  

Sarah Christilaw  Sarah  Christilaw  Product  Manager  Diabetes  Care  

BD Medical 2100 Derry Road West, Suite 100 Mississauga, ON L5N 0B3 Canada tel: 800.268.5430 fax: 905.288.6006 www.bd.com

Becton, Dickinson and Company

Page 2: Upgrade Notice 329505 Customer FINAL EN - …images.medimart.com/portal/Advertisement_2014-02-10.pdf · IMPORTANTPRODUCT)UPGRADE)NOTICE January’2014’ ... Becton, Dickinson and

BD  SAFETY  PEN  NEEDLES  

Catalog  No.   UPC  Code   Product  Name   Needle  Size   Quantity  

308-­‐329505   3  82903  29505  1   BD  Autoshield™  Duo  

30G  x  5mm  (3/16  in.)   100/Box  

308-­‐329305   3  82903  29305  6   BD  Autoshield™  5mm   29G  x  5mm  (3/16  in.)  

200/Box  

308-­‐329308   3  82903  29308  7   BD  Autoshield™  8mm   29G  x  8mm  (5/16  in.)  

200/Box  

    Useful  Links   1. FIT4  Safety  Canada  Recommendations  for  Best  Practice  in  the  Safe  Use  of  Diabetes  Sharps   http://www.fit4diabetes.com/files/7813/7215/5610/FIT_Safety_Recommendations_Canada_June_13.pdf   2. BD  Company  Website    www.bd.com/autoshieldduo    

Page 3: Upgrade Notice 329505 Customer FINAL EN - …images.medimart.com/portal/Advertisement_2014-02-10.pdf · IMPORTANTPRODUCT)UPGRADE)NOTICE January’2014’ ... Becton, Dickinson and

† List of pens and dosers on package.

BD

AutoShield™

Duo

0,30mm

x 5mm

30G x 3/16"

YYYY-MM

XXXXXXX

Outer CoverClear Outer

ShieldNeedle

White SleevePeel Tab

Inner ShieldProduct

SizeBD AutoShield™ Duo Safety Pen Needle with Dual Automatic Protective Shields

Instructions for Use for 30G (0.30mm) x 5mm

Outer Shield Covers Needle

Before Use

After Use

Pen Connection End

Activated Shield with Red Indicator Band Protected Pen Connection End

1. Attaching the Needle to pen† for injection

2. Performing the injection

3. Removing the Pen Needle

1.1 Verify the integrity of each unit pack before use. Clean the rubber seal on the pen per your usual technique or protocol.Take the Peel Tab off the Pen Needle. Holding the Outer Cover, push and twist the Pen Needle onto the pen in a clockwise direction until it meets resistance.

1.2 Pull ONLY the Outer Cover straight off.

1.3 Always check the flow in the Pen Needle before each injection by priming the device with an airshot. Dial 2 units, point the pen up and press the button. A drop or stream of liquid should appear at the Needle tip. IF NOT, repeat as recommended by the pen’s instructions. If the pen still does not prime, change the Needle and repeat the priming steps.

Note: A drop of liquid may fall back into the Shield, which may result in excess fluid being left on the skin after injection.

Dial the prescribed dose on the pen (per manufacturers instruction). Hold the pen as shown. Your thumb should not be on the button during this step.

Insert the Needle straight into the skin in one continuous motion until the Clear Outer Shield retracts and the White Sleeve is flush with the skin.

Refer to figures 2.1a, 2.2 and 2.3.

Maintain constant pressure against the skin and deliver the dose by depressing the button with your thumb. Do not remove the Needle until the dose has been completely delivered. Refer to the instructions of the pen manufacturer for specific hold time.

If there is a concern for potential intramuscular injection, use a wide pinch, allowing approximately 1 inch of skin between the fingers after the pinch is completed.

Refer to figure 2.1b.

X Incorrect Technique

DO NOT pinch less than 1 inch of skin between the fingers.

AVOID holding the insulin pen angled as you would hold an insulin syringe.

DO NOT inject with the Needle pointing at an angle towards your fingers; this may result in a needle-stick injury.

3.1 Once the entire dose has been given, lift the pen away from the skin. The Inner Shield will automatically deploy and lock in place. A Red Indicator Band will appear confirming Shield is locked in place and that the Pen Needle has been used.

3.2 Always hold the Pen Needle by the White Sleeve when removing. Twist the pen counterclockwise.

3.3 Pen Connection End is protected. Protection is confirmed when orange shield deploys and covers Needle.

DO NOT place your fingers on the activated shields.

Discard the used Pen Needle into a Sharps Collector.

Note: Dispose of all plastic packaging materials.

Do not reuse

000

0

0

2.1

0

X

3.1 3.2 3.3

1.11.2

1.3

b

2.1 2.2 2.3a

The BD AutoShield™ Duo Pen Needle is intended for use with pen injector devices for the injection of drugs.

The product has two safety shields, which lock in place after use (patient-end) and upon removal of the Needle from the pen (pen connection-end).The locked shields help reduce the occurrence of needle sticks from both ends of the Needle.

The abdomen is the preferred injection site

BD, BD Logo and BD AutoShield are trademarks of Becton, Dickinson and Company. © 2011 BD. Patent Pending ”Type A” EN ISO 11608-2 2011-04 11D06912559 DUOENFRC

Becton, Dickinson and Company1 Becton Drive Franklin Lakes, NJ 07417 USAwww.bd.com/us

Becton Dickinson Canada Inc.2100 Derry Road West, Suite 100 Mississauga ON. L5N 0B3 CANADAwww.bd.com/ca

Page 4: Upgrade Notice 329505 Customer FINAL EN - …images.medimart.com/portal/Advertisement_2014-02-10.pdf · IMPORTANTPRODUCT)UPGRADE)NOTICE January’2014’ ... Becton, Dickinson and

BD

AutoShield™

Duo

0,30mm

x 5mm

30G x 3/16"

YYYY-MM

XXXXXXX

Enveloppe externe

Embout protecteur

externe clair Aiguille

Manchon blanc

Languette métallique

Embout protecteur

interne Format du

Produit BD AutoShieldMC Duo Aiguille de sécurité avec double système d’auto-verrouillage pour stylo injecteur

Instructions d’utilisation Pour calibre 30 (0.30mm) x 5mm

Enveloppe externe couvre l’aiguille

Avant usage

Après usage

Bout connecteur du stylo injecteur

Embout protecteur activé avec bande indicatrice rouge Bout connecteur

du stylo injecteur protégé

000

0

0

2.1

0

X

3.1 3.2 3.3

1.11.2

1.3

b

2.1 2.2 2.3a

1. Fixer l’aiguille au stylo injecteur† pour effectuer l’injection

2. Effectuer l’injection

3. Retrait de l’aiguille à stylo injecteur

1.1 Vérifier l’intégrité de chaque sachet avant l’utilisation. Nettoyer le sceau de caoutchouc du stylo selon votre technique habituelle ou le protocole proposé. Retirer la languette métallique de l’aiguille. En maintenant l’enveloppe externe, pousser et visser l’aiguille dans le sens des aiguilles d’une montre sur le stylo injecteur, jusqu’à ce qu’il y ait une résistance.

1.2 Retirer UNIQUEMENT l’enveloppe externe.

1.3 Toujours vérifier l’écoulement de l’insuline dans l’aiguille avant chaque injection en amorçant le dispositif selon les étapes suivantes. Prélevez 2 unités, pointez le stylo vers le haut et appuyez sur le bouton injecteur. Une goutte ou un jet de liquide devrait apparaître au bout de l’aiguille. Sinon, répétez le processus tel que recommandé selon les instructions du stylo injecteur. Si le stylo injecteur ne s’amorce toujours pas, changer l’aiguille et répéter les étapes d’amorçage.Note: Une goutte de liquide pourrait retomber dans l’embout protecteur, ce qui pourrait entraîner un excédant de fluide laissé sur la peau suite à l’injection.

Régler la dose prescrite sur le stylo injecteur (selon les instructions du fabriquant). Tenir le stylo injecteur tel qu’illustré. Votre pouce ne devrait pas être sur le bouton d’injection à cette étape-ci.

Insérer l’aiguille directement dans la peau en un mouvement continu jusqu’à ce que l’embout externe clair et le manchon blanc soient au niveau avec la peau.

Référez-vous aux figures 2.1a, 2.2 et 2.3.

Maintenir une pression constante contre la peau et injecter la dose en appuyant sur le bouton d’injection à l’aide de votre pouce. NE PAS retirer l’aiguille avant que la dose soit complètement injectée. Se référer aux directives du fabriquant du stylo injecteur pour obtenir les temps d’injections recommandés.

S’il y a un risque potentiel d’injection intramusculaire, faire usage d’un pli cutané large de façon à tenir 2,5cm de peau entre les doigts lorsque le pli est complété.

Référez-vous à la figure 2.1b.

X Technique inexacte

NE PAS pincer moins de 2,5cm de peau entre les doigts.

ÉVITEZ de tenir le stylo injecteur dans un angle, de la même façon que l’on tiendrait une seringue à insuline.

NE PAS injecter avec l’aiguille pointant à un angle dirigé vers les doigts, ceci pourrait causer une piqûre accidentelle.

3.1 Dès que la dose complète a été injectée, retirer le stylo injecteur de la peau. L’embout protecteur interne se déploiera automatiquement et se verrouillera. Une bande indicatrice de couleur rouge apparaîtra, confirmant que l’embout protecteur est verrouillé et que l’aiguille a été utilisée.

3.2 Toujours tenir l’aiguille à stylo par le manchon blanc lors de son retrait. Tourner le stylo-injecteur dans le sens contraire des aiguilles d’une montre.

3.3 Le bout connecteur du stylo injecteur est protégé. La protection est assurée lorsque l’embout orange se déploie et couvre l’aiguille.

NE PAS placer vos doigts sur l’embout protecteur activé.

Disposer de l’aiguille utilisée dans un contenant prévu à cet effet. Note: Disposer adéquatement de tout emballage de plastique.

L’aiguille pour stylo-injecteur BD AutoShieldmc Duo est conçue pour être utilisée avec des dispositifs d’injection de médication de type stylo.

Ce produit offre deux embouts sécuritaires qui se bloquent automatiquement après usage (côté patient) et lors du retrait de l’aiguille du stylo injecteur (embout connecteur du stylo). Les embouts ainsi bloqués aident à diminuer la possibilité de blessures causées par les deux extrémités de l’aiguille.

L’abdomen est le site d’injection privilégié

Becton, Dickinson and Company1 Becton Drive Franklin Lakes, NJ 07417 USAwww.bd.com/us

Becton Dickinson Canada Inc.2100 Derry Road West, Suite 100 Mississauga ON. L5N 0B3 CANADAwww.bd.com/ca

† Liste des stylos injecteurs et doseurs sur l’emballage.

Ne pas réutillser

BD, le logotype BD Logo et BD AutoShield sont des marques de commerce de Becton, Dickinson and Company. © 2011 BD. Brevet en cours d’enregistrement. ”Type A” EN ISO 11608-2