UTRASONIC SCALPEL Set-up and operation Model No: ABC-123-45-678 Serial No. 1234-56-879 Throubleshooting
17
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Untrasonic Scalpel

UTRASONIC SCALPELSet-up and operation

Model No: ABC-123-45-678

Serial No. 1234-56-879

Throubleshooting

Page 2: Untrasonic Scalpel

Place generator in a flat and stable surface

Page 3: Untrasonic Scalpel

Check if the power cord in the back is securely attach.

Page 4: Untrasonic Scalpel

Set your hand piece and the ultrasonic cable in the mayo table.

Page 5: Untrasonic Scalpel

Insert the silver end of the blue cable into the bottom opening

of the hand piece.

Page 6: Untrasonic Scalpel

Holding the silver end of the cable, turn the hand piece clockwise

Until its tight.

Page 7: Untrasonic Scalpel

Insert the torque wrench on the tip on the hand piece.

Page 8: Untrasonic Scalpel

Turn the torque wrench clockwise until you hear 2 clicks

Indicating that the hand piece is secure.

Page 9: Untrasonic Scalpel

Turn on the generator by pressing the power button located

Located in the right lower corner.

Page 10: Untrasonic Scalpel

Pass off the other end of the cable to be plug in this port

located in the left lower corner.

Page 11: Untrasonic Scalpel

The circulator will then insert and connect the hand piece

into the generator.

Page 12: Untrasonic Scalpel

Your circulator can now start test process by pressing test

button in the generator.

Page 13: Untrasonic Scalpel

You can also perform a test by pressing any of the

button in the hand piece.

Page 14: Untrasonic Scalpel

You will hear 3 beeps when the test is done,

now the ultrasonic scalpel Is ready to use.

Page 15: Untrasonic Scalpel

Throubleshooting

• No Power

-Check if power cord is securely attach and the power button is turned on

• Hand piece not working.

- Check cable connecting to the generator

- Make sure hand piece is tightly connected to the cable

- Restart generator by turning it off and on again

- Perform a retest

• Power is not enough or to much.

- Increase or decrease power by pressing the power level button up

and down as needed.

• Generator is to loud or to low

- Turn volume knob down or up located in the left lower corner next to

the power button.

Report Issue

Page 16: Untrasonic Scalpel

Request for Repair

Ultrasonic Scalpel Generator

Model No: ABC-123-45-678

Serial No. 1234-56-879

Reported Issue: (Choose only one)

No Power

Electrical (e.i. Frayed cable)

Mechanical (e.i. Broken parts)

Other

Choose Room no. 1

2

3

4

5 Send to Bio-med

Page 17: Untrasonic Scalpel

END

John Springer CST