Flomeko2019 International Flow Measurement Conference Uncertainty of SO 2 measurements in dryers due to water droplet and water film condensation Stanislav Knotek 1 Jan Geršl 1 Aleš Paták 2 1 Czech Metrology Institute 2 Institute of Scientific Instruments of the Czech Academy of Sciences Czech Republic 25-28th June 2019, Lisbon, Portugal
20

UncertaintyofSO measurementsin dryers due to water droplet ...flomeko2019.lnec.pt/proceedings/1094 Stanislav Knotek.pdf · Condensation a) Filmwise condensation – on hydrophilic

Jan 06, 2020

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: UncertaintyofSO measurementsin dryers due to water droplet ...flomeko2019.lnec.pt/proceedings/1094 Stanislav Knotek.pdf · Condensation a) Filmwise condensation – on hydrophilic

Flomeko2019International Flow Measurement Conference

Uncertainty of SO2 measurements in dryers due to water droplet and water film condensation

Stanislav Knotek 1

Jan Geršl 1

Aleš Paták 2

1Czech Metrology Institute2Institute of Scientific Instruments of the Czech Academy of Sciences

Czech Republic

25-28th June 2019, Lisbon, Portugal

Page 2: UncertaintyofSO measurementsin dryers due to water droplet ...flomeko2019.lnec.pt/proceedings/1094 Stanislav Knotek.pdf · Condensation a) Filmwise condensation – on hydrophilic

SulfNorm – EMPIR project

Intention of the project

� Prenormative work for new Standard Reference Method (SRM)

� SRM today – unconditioned sampling - based on analysis of SO2 dissolved in water

� New SRM – conditioned sampling – using P-AMS (Portable Automated Measuring

System) – cold and dry gas is needed for SO2 measurement

Motivation

� SO2 pollution has significant impact on health of population

� Most of SO2 is produced by industrial processes

� Accurate measurement is needed

Main issue

� SO2 bias due to physical and chemical processes during cooling and drying ?

Page 3: UncertaintyofSO measurementsin dryers due to water droplet ...flomeko2019.lnec.pt/proceedings/1094 Stanislav Knotek.pdf · Condensation a) Filmwise condensation – on hydrophilic

Tasks of CMI in the project

A.2.2.7

“… create a mathematical model of SO2 losses in dryers based on cooling and condensation of water.”

A.2.2.11

“… to obtain estimations of the SO2 losses in the selected dryer.”

Key part of the dryer in P-AMS

� Gas cooler (ECP .000)

� Borosilicate glass heat exchanger

Type of the measuring system

� Portable Automated Measuring System

(P-AMS, M&C® PSS-5)

Page 4: UncertaintyofSO measurementsin dryers due to water droplet ...flomeko2019.lnec.pt/proceedings/1094 Stanislav Knotek.pdf · Condensation a) Filmwise condensation – on hydrophilic

Scheme of P-AMS

Scheme of P-AMS

Page 5: UncertaintyofSO measurementsin dryers due to water droplet ...flomeko2019.lnec.pt/proceedings/1094 Stanislav Knotek.pdf · Condensation a) Filmwise condensation – on hydrophilic

Scheme of heat exchanger

Gas cooler with jet-stream heat exchanger Scheme of glass heat exchanger

Page 6: UncertaintyofSO measurementsin dryers due to water droplet ...flomeko2019.lnec.pt/proceedings/1094 Stanislav Knotek.pdf · Condensation a) Filmwise condensation – on hydrophilic

Scheme of the processes in the dryer

Scheme of glass heat exchanger

Key processes

� Condensation (droplets, water film)

� Mass transfer through gas-liquid interface

� Chemical processes in liquid

� Diffusion

Page 7: UncertaintyofSO measurementsin dryers due to water droplet ...flomeko2019.lnec.pt/proceedings/1094 Stanislav Knotek.pdf · Condensation a) Filmwise condensation – on hydrophilic

Condensation

a) Filmwise condensation – on hydrophilic surface (borosilicate glass)

b) Dropwise condensation on wall – on hydrophobic surface

c) Droplets condensation in air – for temperature under dew point

Page 8: UncertaintyofSO measurementsin dryers due to water droplet ...flomeko2019.lnec.pt/proceedings/1094 Stanislav Knotek.pdf · Condensation a) Filmwise condensation – on hydrophilic

Condensation of droplets

Droplet growth by condensation

Fluids

properties

Vapor partial pressure (Kelvin eq.)

Droplet temperature

Ambient

conditions

Droplet

properties

SR … relative humidity

Initial droplet

diameter

ps … saturation vapor

pressure

Page 9: UncertaintyofSO measurementsin dryers due to water droplet ...flomeko2019.lnec.pt/proceedings/1094 Stanislav Knotek.pdf · Condensation a) Filmwise condensation – on hydrophilic

Liquid film thickness by Nusselt theory

• Vertical wall

• Stagnant vapor

• Laminar flow regime

• Smooth film surface

Condensation of liquid film

Page 10: UncertaintyofSO measurementsin dryers due to water droplet ...flomeko2019.lnec.pt/proceedings/1094 Stanislav Knotek.pdf · Condensation a) Filmwise condensation – on hydrophilic

Liquid film thickness by Nusselt theory

� Vertical wall

� Stagnant vapor

� Laminar flow regime

� Smooth film surface ?

� Our case - countercurrent vapor motion

– relatively complex models

� Our assumption – Nusselt theory gives acceptable estimate of the film thickness

Condensation of liquid film

Page 11: UncertaintyofSO measurementsin dryers due to water droplet ...flomeko2019.lnec.pt/proceedings/1094 Stanislav Knotek.pdf · Condensation a) Filmwise condensation – on hydrophilic

Henry’s Law

Mass transfer through gas-liquid interface

�� � �������� ���

��� � ��exp �∆�

1

��

1

��

Mass accommodation coefficient, α

The ratio of molecules absorbed through the gas-liquid interface

to the number of molecules which hit the liquid surface.

� � 5.4 � 0.6 %at 295 K

�� � �������

Page 12: UncertaintyofSO measurementsin dryers due to water droplet ...flomeko2019.lnec.pt/proceedings/1094 Stanislav Knotek.pdf · Condensation a) Filmwise condensation – on hydrophilic

Chemical reactions

Chemical processes of SO2 in water

Resulting concentration

!" ∙ �"!

� !$%

!$"%

!"

Dissociation const. KH , K1 , K2 = f(T)

(maximal concentration for given

temperature and partial pressure of SO2)

Page 13: UncertaintyofSO measurementsin dryers due to water droplet ...flomeko2019.lnec.pt/proceedings/1094 Stanislav Knotek.pdf · Condensation a) Filmwise condensation – on hydrophilic

Governing equations

Diffusion processes

&�� ', )

&)� *�

&"�� ', )

&'"

&���', )

&)� *�

&"���', )

&'"

Boundary conditions on gas-liquid interface

�*�&

&'�� '� , ) �

�+̅

4�� '� , ) �

�� '� , )

���

�*�

&

&'�� '� , ) �

�+̅

4�� '� , ) �

�� '� , )

���

��� … Henry’s Law constant

� … mass accommodation

coefficient

*�, *� … diffusion coefficients

+̅ … mean thermal velocity of SO2

'� … location of the interface

�� ��

���*�*�

Page 14: UncertaintyofSO measurementsin dryers due to water droplet ...flomeko2019.lnec.pt/proceedings/1094 Stanislav Knotek.pdf · Condensation a) Filmwise condensation – on hydrophilic

Dependence on dimensionality

Simulations of diffusion into growing droplet

Ambient conditions

� 0.01 ppm SO2 in air

� T=293 K

� Relative humidity 101%

Conclusion

� 3D simulation needed

Page 15: UncertaintyofSO measurementsin dryers due to water droplet ...flomeko2019.lnec.pt/proceedings/1094 Stanislav Knotek.pdf · Condensation a) Filmwise condensation – on hydrophilic

Dependence on initial concentration of SO2 in stack gas

Simulations of diffusion into growing droplet

Ambient conditions

� 0.01 – 1 ppm SO2 in air

� T=293 K

� Relative humidity 101%

Conclusions

� Maximal concentration is in

agreement with Henry’s Law

� The time to equilibrium state

decreases with increasing ppm

Page 16: UncertaintyofSO measurementsin dryers due to water droplet ...flomeko2019.lnec.pt/proceedings/1094 Stanislav Knotek.pdf · Condensation a) Filmwise condensation – on hydrophilic

Simulations on simplified geometry

Simulations of diffusion into water film

δ= const.

0.1 mm

Ug (0.1,0.2,0.4 m/s)

�./"�0 (1, 10, 100 ppm)

Convection

+

Diffusion

nSO2 =0

�./"123=?

15cm

6 mm

Page 17: UncertaintyofSO measurementsin dryers due to water droplet ...flomeko2019.lnec.pt/proceedings/1094 Stanislav Knotek.pdf · Condensation a) Filmwise condensation – on hydrophilic

Resulting concentration of SO2 in gas using COMSOL Multiphysics®

Simulations of diffusion into water film

Page 18: UncertaintyofSO measurementsin dryers due to water droplet ...flomeko2019.lnec.pt/proceedings/1094 Stanislav Knotek.pdf · Condensation a) Filmwise condensation – on hydrophilic

Resulting concentration of SO2 in gas after passing through the cooler

Simulations of diffusion into water film

Initial conditions

� 1 – 100 ppm SO2 in gas

� T∞= 20 °C

� Tw=5 °C

� Gas relative humidity = 101%

Conclusion

� The SO2 losses increase slightly with decreasing concentration

� The SO2 losses increase strongly with decreasing gas flow rate

Page 19: UncertaintyofSO measurementsin dryers due to water droplet ...flomeko2019.lnec.pt/proceedings/1094 Stanislav Knotek.pdf · Condensation a) Filmwise condensation – on hydrophilic

Summary of simplifications and challenges

Used simplifications

� The 2D simplified geometry with given dimensions (D=6mm, L=15cm)

� Constant water film thickness along the whole tube (t=0.1 mm)

� Water film thickness estimated using Nusselt theory (stagnant vapor)

� Constant temperature of the gas and water film (T∞= 20 °C, Tw=5 °C)

� Relative humidity 101 %

� …

Remaining challenges

� Better modeling of the film thickness (countercurrent vapor flow)

� Initial gas properties (temperature, humidity) on inlet

� Validation

Page 20: UncertaintyofSO measurementsin dryers due to water droplet ...flomeko2019.lnec.pt/proceedings/1094 Stanislav Knotek.pdf · Condensation a) Filmwise condensation – on hydrophilic

Conclusions

� First attempt using several simplifications – qualitative results

� Although the maximal concentration of SO2 in water droplet can be reached

in time less than 0.1s, the droplet growth is slow

� Filmwise condensation is dominating in dryers

� The SO2 losses increase slightly with decreasing concentration

� The SO2 losses increase strongly with decreasing gas flow rate

Thank you for your attention

Czech Metrology InstituteStanislav Knotek