PRIKAZ DOPLEROVOG EFEKTA U MEDICINSKOJ ULTRAZVUČNOJ DIJAGNOSTICI dipl. inž. NEGOSLAV ĐAJA HAPEL, Beograd
Nov 07, 2014
PRIKAZDOPLEROVOG EFEKTA
U MEDICINSKOJ ULTRAZVUČNOJ
DIJAGNOSTICI
dipl. inž. NEGOSLAV ĐAJA
HAPEL, Beograd
POKRETNIIZVOR
POKRETNIIZVOR
POSMATRAČ
POSMATRAČ
PRODUŽENA TALASNA DUŽINA - SMANJENA UČESTANOST
SABIJENA TALASNA DUŽINA - POVEĆANA UČESTANOST
.POKRETNI IZVOR ZVUKA
POVIŠENAUČESTANOST
SNIŽENAUČESTANOST
DOPLEROV EFEKAT
PROMENA UČESTANOSTI (FREKVENCIJE) ZVUKA, SVETLOSTI
ILI DRUGIH TALASA, PROUZROKOVANA KRETANJEM
IZVORA ILI POSMATRAČA
PROMENA UČESTANOSTI SE ZOVE DOPLEROV ILI FREKVENTNI
POMAK
PROMENA UČESTANOSTI SE ZOVE
FREKVENTNI POMAK
DOPLEROV POMAK
IZRAČUNAVANJE FREKVENTNOG POMAKA
V . cos = C . ( ± F d )
2 Fo.
V = BRZINA KRVI
cos KOSINUS UGLA IZMEĐU ULTRAZVUČNOG TALASAI VEKTORA PROTOKA KRVI
=
C = BRZINA ULTRAZVUKA U TKIVU (OKO 1540 m/s)
F d = DOPLEROV (FREKVENTNI) POMAK
F O = FREKVENCIJA ULTRAZVUČNOG TRANSDJUSERA
DOPLEROVA JEDNAČINA
PRAVAC PROTOKA KRVI BRZINOM V
PRAVAC TRANSDJUSERA I
IZMERENA BRZINA V
.COS
FREKVENTNI POMAK
DOPLEROVA JEDNAČINA
BRZINA PROTOKA KRVI
IZRAČUNAVANJE BRZINE PROTOKA KRVI
FFT = BRZA FURIJEOVA TRANSFORMACIJA
POMOĆU FFT ANALIZE SE IZ SLOŽENOG DOPLEROVOG
SIGNALA IZDVAJAJU INDIVIDUALNE FREKVENCIJE
UPROŠĆENI PRIKAZ RAZLIČITIH BRZINA DOBIJENIH SPEKTRALNOM ANALIZOM (Bin)
INTENZITET RAZLIČITIH OPSEGA BRZINA CRVENIH KRVNIH ZRNACA
TRODIMENZIONALNI SPEKTRALNI PRIKAZ BRZINA CRVENIH KRVNIH ZRNACA
ODREĐIVANJE PRAVCA I SMERA PROTOKA KRVI
BOJENJE PRAVCA I SMERA PROTOKA KRVI
BART - BLUE AWAY RED TOWARDS
“SLEPA” SONDA ZA KONTINUALNI DOPLER
SONDA ZA PULSNI DOPLER
PULSNI DOPLER
KONTINUALNI DOPLER
PULSNI DOPLER
0
CW
PW
FREKVENTNO PREBACIVANJE (ALIASING EFEKAT)
+Vmax
-Vmax
t
FREKVENTNO PREBACIVANJE
ALIASING EFEKAT
PULSNI DOPLER (PW)
1. MOGUĆNOST ODREĐIVANJA PROTOKA PO
PRAVCU, PO SMERU I PO DUBINI.
2. MOŽE DA SE MERI SAMO OGRANIČENA
PROMENA DOPLEROVE FREKVENCIJE,
ODNOSNO, MALA BRZINA.
1. NEMA ODREĐIVANJA PROTOKA PO DUBINI, VEĆ SE REGISTRUJU SVE BRZINE
DUŽ CELOG KRVNOG SUDA KROZ KOJI ULTRAZVUČNI SNOP PROLAZI.
2. MOGU DA SE MERE SVE BRZINE PROTOKA
U KARDIOLOGIJI ( do 7,0 m/s, što odgovara
gradijentu pritiska od 196 mmHg).
KONTINUALNI DOPLER (CW)
VIŠA UČESTANOST
TRANSDJUSERA
1. KVALITETNIJA M I 2D SLIKA –
BOLJA REZOLUCIJA
2. MANJA BRZINA KOD DOPLERA
NIŽA UČESTANOST
TRANSDJUSERA
1. SLABIJA M I 2D SLIKA –
MANJA REZOLUCIJA
2. VEĆA BRZINA KOD DOPLERA