Ultra-Wide £ System MANUFACTURER #507-8, Geoje3-dong, Yeonje-gu Busan, Korea. Zip_ 611-804 tel_ 051 850 2500 fax_ 051 861 4693 HEAD OFFICE 8th FL. WorldMeridian Venture Center, 426-5, Gasan-dong, Geumcheon-gu, Seoul, Korea tel_ 02 2016 7000, fax_ 02 2016 7001 homepage_ http://www.osstem.com Edition 12 / 2010
13

ultra wide system-en - osstem.cz · abutment 16 Surgical Procedures 17 Drilling sequence 20 Insertion depth guide Indication of Ultra-Wide System 4 GS Ultra-Wide System 5 GS Ultra-Wide

Aug 18, 2019

Download

Documents

trinhxuyen
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: ultra wide system-en - osstem.cz · abutment 16 Surgical Procedures 17 Drilling sequence 20 Insertion depth guide Indication of Ultra-Wide System 4 GS Ultra-Wide System 5 GS Ultra-Wide

1 Total Dental Solution Provider 2

Ultra-Wide£

System

MANUFACTURER#507-8, Geoje3-dong, Yeonje-guBusan, Korea. Zip_ 611-804tel_ 051 850 2500fax_ 051 861 4693

HEAD OFFICE8th FL. WorldMeridian Venture CenterⅡ, 426-5,Gasan-dong, Geumcheon-gu, Seoul, Koreatel_ 02 2016 7000, fax_ 02 2016 7001homepage_ http://www.osstem.com

Edition 12 / 2010

Page 2: ultra wide system-en - osstem.cz · abutment 16 Surgical Procedures 17 Drilling sequence 20 Insertion depth guide Indication of Ultra-Wide System 4 GS Ultra-Wide System 5 GS Ultra-Wide

Indication of Ultra-Wide�

SystemImmediate placement at the extraction socket

Immediate replacement of the failed implant

Delayed placement in the healed mature bone

12 US Ultra-Wide� System

13 US Ultra-Wide� fixture

15 US Ultra-Wide� abutment

16 Surgical Procedures

17 Drilling sequence

20 Insertion depth guide

Indication of Ultra-Wide� System

4 GS Ultra-Wide� System

5 GS Ultra-Wide� fixture

7 GS Ultra-Wide� abutment

8 SS Ultra-Wide� System

9 SS Ultra-Wide� fixture

1 SS Ultra-Wide� abutment

21 Surgical KIT

21 Ultra KIT

22 Instrument dimension indication

24 Instruction for Use

Table of Contents

SS Ultra-Wide�

1-Stage Internal connection

GS Ultra-Wide�

Internal connection

US Ultra-Wide�

External connection

Caution!!! When inserting Ultra-Wide fixture, maintain more than 1.5mm survival alveolar bone in order to prevent bone loss.

Page 3: ultra wide system-en - osstem.cz · abutment 16 Surgical Procedures 17 Drilling sequence 20 Insertion depth guide Indication of Ultra-Wide System 4 GS Ultra-Wide System 5 GS Ultra-Wide

GS Ultra-Wide�

System

05

Features of GS Ultra-Wide£ Fixture

Ultra-Wide£ Systemwww.osstem.com

Connection

GS Ultra-Wide fixtures that have a 11 internal connection are

compatible with GS standard abutments.

Neck

Bevel neck design of fixture provides more bone volume which is

beneficial to minimizing bone resorption and support for

aesthetic gingival contour.

Body & Thread

0.8 pitch, 45° angled and 0.45 thread depth provides an

enhanced self-tapping ability.

Surface

RBM surface has good biocompatibility which induces secure

osseointegration. (Surface roughness Ra: 1.2~1.8㎛)

Apex

The thread which is existing until apex of fixture and 4 rounded cutting

edges provide progressive initial installation ability and good initial

stability in immediate placement case.

Hex 2.5mm

M2.0 Screw

∅3.35

11。

GS Ultra-Wide£ Fixture

Ultra-Wide£

System

�GS Ultra-Wide� fixture

�GS Ultra-Wide� abutment

Page 4: ultra wide system-en - osstem.cz · abutment 16 Surgical Procedures 17 Drilling sequence 20 Insertion depth guide Indication of Ultra-Wide System 4 GS Ultra-Wide System 5 GS Ultra-Wide

06 Total Dental Solution Provider 07

Connection Standard

L D ∅6.0 ∅7.0

7 GS2W6007R02 GS2W7007R02

8.5 GS2W6008R02 GS2W7008R02

10 GS2W6010R02 GS2W7010R02

11.5 GS2W6011R02 GS2W7011R02

13 GS2W6013R02 GS2W7013R02

∅4.25

∅4.2

∅6.0

∅4.25

∅7.0

4

3

2

1

∅5.2

L

Order Code of GS Ultra - Wide£ Fixture

Fixture Only

- Fixture : Product Code (ex : GS2W6010R02)

Pre-Mounted Fixture

- Fixture + Simple Mount + Cover Screw : B + Product Code (ex : BGS2W6010R02)

∅6.0 fixture ∅7.0 fixture Pre-Mounted fixture

Rigid Transfer Angled Freeform ST

Freeform Convertible Combination

cylinder

Combination Angled

cylinder

GoldCast Convertible GoldCast

cylinder

Plastic cylinder

1. GS Ultra-Wide fixtures are wholly compatible with GS standard abutment components.

2. Prosthetic procedure is the same as GS prosthetic procedure.

(Refer to " Prosthetic procedure for GS Implant System")

The actual length of GS Ultra-Wide fixture is shorter by 0.5mm than the labeled length for easy control of

fixture's insertion depth. (Except for 7mm)

Important

Line-up

Ultra-Wide£ Systemwww.osstem.com

GS Ultra-Wide£ Fixture GS Ultra-Wide£ Abutment

AbutmentsType

Cement

retained

restorations

Screw

retained

restorations

Page 5: ultra wide system-en - osstem.cz · abutment 16 Surgical Procedures 17 Drilling sequence 20 Insertion depth guide Indication of Ultra-Wide System 4 GS Ultra-Wide System 5 GS Ultra-Wide

09

SS Ultra-Wide�

System

Ultra-Wide£

System

�SS Ultra-Wide� fixture

�SS Ultra-Wide� abutment

Features of SS Ultra-Wide£ Fixture

Ultra-Wide£ Systemwww.osstem.com

Connection

SS Ultra-Wide fixtures that have a 8�internal connection are

compatible with SS wide abutments.

Trans-gingival section

Machined surface facilitates soft tissue management and plaque

control.

Neck

Expanded thread of fixture provides an enhanced initial stability.

Body & Thread

0.8 pitch, 45�angle and 0.45 thread depth provides an enhanced

self-tapping ability.

Surface

RBM surface has good biocompatibility which induces secure

osseointegration. (Surface roughness Ra: 1.2~1.8㎛)

Apex

The thread which is existing until apex of fixture and 4 rounded

cutting edges provide progressive initial installation ability and

good initial stability in immediate placement case.

SS Ultra-Wide£ Fixture

Octa 2.9mm

M2.0 Screw

Page 6: ultra wide system-en - osstem.cz · abutment 16 Surgical Procedures 17 Drilling sequence 20 Insertion depth guide Indication of Ultra-Wide System 4 GS Ultra-Wide System 5 GS Ultra-Wide

10 Total Dental Solution Provider 11

Platform Wide

G/H 2.0

L D ∅6.0 ∅7.0

7 SS2WB62007A SS2WB72007A

8.5 SS2WB62008A SS2WB72008A

10 SS2WB62010A SS2WB72010A

11.5 SS2WB62011A SS2WB72011A

13 SS2WB62013A SS2WB72013A

Order Code of SS Ultra - Wide£ Fixture

Fixture Only

- Fixture : Product Code (ex: SS2WB62007A)

Pre-Mounted Fixture

- Fixture + Simple Mount + Cover Screw : A + Product Code (ex : ASS2WB62007A)

1. SS Ultra-Wide fixtures are wholly compatible with SS wide abutment components.

2. Prosthetic procedure is the same as SS prosthetic procedure.

(Refer to " Prosthetic procedure for SS Implant System")

Important

Line-up

Ultra-Wide£ Systemwww.osstem.com

SS Ultra-Wide£ Fixture SS Ultra-Wide£ Abutment

∅6.0 fixture ∅7.0 fixture Pre-Mounted fixture

Solid Excellent

Solid

Com Octa

Com Octa

Plus

Hanaro Com Octa

Angled

Com Octa

Gold

Octa Octa gold

cylinder

Octa plastic

cylinder

AbutmentsType

Cement

retained

restorations

Screw

retained

restorations

∅4.2

2.0

∅6.0

L

∅6.0

∅5.2

∅7.0

∅6.0

Page 7: ultra wide system-en - osstem.cz · abutment 16 Surgical Procedures 17 Drilling sequence 20 Insertion depth guide Indication of Ultra-Wide System 4 GS Ultra-Wide System 5 GS Ultra-Wide

12 Total Dental Solution Provider 13

US Ultra-Wide�

System

Ultra-Wide£

System

�US Ultra-Wide� fixture

�US Ultra-Wide� abutment

Ultra-Wide£ Systemwww.osstem.com

US Ultra-Wide£ Fixture

Features of US Ultra-Wide£ fixture

Connection

US Ultra-Wide fixtures that have a 3.4mm external hex

connection are compatible with US wide abutments.

Neck

Platform switching effect of fixture minimizes bone resorption

and offers an aesthetic result.

Body & Thread

0.8 pitch, 45�angle and 0.45 thread depth provides a good self

tapping ability. Expanded thread of fixture provides an enhanced

initial stability.

Surface

RBM surface has good biocompatibility which induces secure

osseointegration. (Surface roughness Ra: 1.2~1.8㎛)

Apex

The thread which is existing until apex of fixture and 4 rounded

cutting edges provide progressive initial installation ability and

good initial stability in immediate placement case.

Hex 3.4mm

M2.5 Screw

Page 8: ultra wide system-en - osstem.cz · abutment 16 Surgical Procedures 17 Drilling sequence 20 Insertion depth guide Indication of Ultra-Wide System 4 GS Ultra-Wide System 5 GS Ultra-Wide

14 Total Dental Solution Provider 15

Ultra-Wide£ Systemwww.osstem.com

US Ultra-Wide£ Fixture US Ultra-Wide£ Abutment

Line-up

Platform Wide

L D ∅6.0 ∅7.0

7 BFW6007A BFW7007A

8.5 BFW6008A BFW7008A

10 BFW6010A BFW7010A

11.5 BFW6011A BFW7011A

13 BFW6013A BFW7013A

∅5.1

∅4.2

∅6.0

∅6.0

∅7.0

∅5.2

L

Order Code of US Ultra - Wide£ Fixture

Fixture Only

- Fixture : Product Code (ex : BFW6007A)

Pre-Mounted Fixture

- Fixture + Simple Mount + Cover Screw : A + Product Code (ex : ABFW6007A)

∅6.0 fixture ∅7.0 fixture Pre-Mounted fixture

1. US Ultra-Wide fixtures are wholly compatible with US wide abutment components.

2. Prosthetic procedure is the same as US prosthetic procedure.

(Refer to "Prosthetic procedure for US Implant System")

Important

Cement Angled Safe

UCLA Esthetic-low Esthetic-low

gold cylinder

Esthetic-low

plastic cylinder

ER

AbutmentsType

Cement

retained

restorations

Combination

retained

restorations

Screw

retained

restorations

Page 9: ultra wide system-en - osstem.cz · abutment 16 Surgical Procedures 17 Drilling sequence 20 Insertion depth guide Indication of Ultra-Wide System 4 GS Ultra-Wide System 5 GS Ultra-Wide

16 Total Dental Solution Provider 17

SurgicalProcedures

Ultra-Wide£

System

GS, SS and US Ultra-Wide fixture use the same drilling sequence

and surgical instruments which is included in Ultra Kit, and it

provides a convenient implant placement.

Ultra-Wide£ Systemwww.osstem.com

Drilling sequence for Ultra-Wide£ Fixture

Drilling sequence with Trephine drill

Conventional drilling sequence

∅4.2/∅5.0 ∅4.6/∅5.5

∅4.6 ∅4.6/∅5.5

Tap∅6.0

Tap∅6.0 ∅6.0

∅6.0

∅7.0

Tap∅7.0∅5.5/∅6.5 ∅7.0

∅5.2/∅6.0 ∅5.5/∅6.5 Tap∅7.0

Existing surgical

procedure application

※ When bone is especially hard(D1), the Surgical Tap must be used.

Page 10: ultra wide system-en - osstem.cz · abutment 16 Surgical Procedures 17 Drilling sequence 20 Insertion depth guide Indication of Ultra-Wide System 4 GS Ultra-Wide System 5 GS Ultra-Wide

18 Total Dental Solution Provider 19

Ultra-Wide£ Systemwww.osstem.com

Surgical procedure with Trephine drill Surgical procedure with Trephine drill

∅2.0 drill

2.0/3.0 Pilot drill

∅3.0 drill

∅4.1(∅4.3) drill

∅4.6 drill

∅4.6/∅5.5

Direct drill

∅5.5 Trialpin

∅6.0 Ultra-Wide fixture

3.0/4.1 Pilot drill

10mm

1mm

∅4.2/∅5.0

Trephine drill

∅4.6/∅5.5

Direct drill

∅5.5 Trial pin

∅6.0 Ultra-Wide fixture

10mm

1mm

� Conventional drilling sequence

¤“6.0 Ultra-Wide£

fixture, Length : 10mm

� Delayed placement in the healed mature bone

¤“6.0 Ultra-Wide£

fixture, Length : 10mm

�Surgical tap is recommended in case of hard bone.

�Verify the depth with Trial pin after final drilling.

�Recommended rpm of Trephine drill : 800~1000rpm

Recommended rpm of drilling : 800~1500rpm

Recommended rpm of tapping : below 20rpm

∅4.2/∅5.0

Trephine drill

∅4.6/∅5.5

Direct drill ∅6.0 Ultra-Wide fixture

US GS SS

US GS SS

US GS SS

∅6.0 Ultra-Wide fixture

US GS SS

∅4.2/∅5.0

Trephine drill

∅4.0 failed

fixture

∅4.6/∅5.5

Direct drill

� Immediate replacement of the failed implant

¤“6.0 Ultra-Wide£

fixture, Length : 10mm

� Immediate placement at the extraction socket

¤“6.0 Ultra-Wide£

fixture, Length : 10mm

Removing the septal bone

with trephine drill

Removing failed fixture

with trephine drill

Page 11: ultra wide system-en - osstem.cz · abutment 16 Surgical Procedures 17 Drilling sequence 20 Insertion depth guide Indication of Ultra-Wide System 4 GS Ultra-Wide System 5 GS Ultra-Wide

20 Total Dental Solution Provider 21

Ultra-Wide£ Systemwww.osstem.com

Insertion Depth Guide Surgical KIT

0.5~1.0

GS Ultra-Wide£ SS Ultra-Wide£ US Ultra-Wide£

Length = base on 10mm

10

0

-1

�GS Ultra-Wide£ ; 0.5~1mm Sub-Crestal

�SS Ultra-Wide£ ; Correspond with Bone Level

�US Ultra-Wide£ ; 0.25mm Supra-Crestal

� Ultra KIT(OUK)

* Lower case component - Open wrench(ASOW) for Simple mount

6.5 Direct drill : 3D6513FNLC SideCut drill :OSLMS20M

Mount driver :ASMDS1.2 Hex hand driver:

AHD12SH5.5 Direct drill : 3D5513FNLC

3 cutter twist drill : 3D4613FNLC

Surgical Tap : OAST60LC5.5 Trial pin :UWFTP55

Surgical Tap : OAST70LC

4.2/5.0 Trephine drill :TD42S

5.2/6.0 Trephine drill :TD52S6.5 Trial pin :UWFTP65

0.25

Page 12: ultra wide system-en - osstem.cz · abutment 16 Surgical Procedures 17 Drilling sequence 20 Insertion depth guide Indication of Ultra-Wide System 4 GS Ultra-Wide System 5 GS Ultra-Wide

2322 Total Dental Solution Provider

� Drill & Surgical tap � Trephine drill

� Trial pin

Drill ¤“4.6

∅4

.6×

13

∅5

.5×

13

∅6

.5×

13

U/W

∅6

.0

U/W

∅7

.0

Direct drill ¤“5.5 Direct drill ¤“6.5 Tap ¤“6.0 Tap ¤“7.0

� Direct drill - 2-step drill has two functions - pilot drill & twist drill

Enables final drilling directly

without pilot drilling

Enhancement of initial stability in the

extract socket by decreasing the dead

space at the apex area

13

10

7

-1∅4.6

∅5.5

∅5.5

6

0

33.3

8.5

11.511.5

∅4.6 ∅5.5

∅5.5 ∅6.5

∅6.5

�Twin window provides clear view and copious

irrigation at target site and facilitates the

removal of bone in Trephine drill.

�You can experience a sufficient cutting

capability which is induced by 10 cutting edges

and chip groove design without spattering.

�Use for the removal of failed fixture

�Use for the removal of septal bone

�Use for the initial drilling

�It is simple to confirm 3mm drilling depth under the

extraction socket.

�Measurement of the socket's internal diameter after

removal of failed fixture.

�Use for purpose of parallel pin(¤“2.0, ¤“3.0).

Drilling Trial pin

11.5

8.5

7

10

13

6

0

21 8

6

4

2

0

11.5

8.5

6

3

0

7

10

13

31

∅4

.2 /

∅5

.0

∅5

.2 /

∅6

.0

Trial pin

¤“5.5

Trial pin

¤“6.5

∅3.0

Inner Dia. ∅4.2 ∅5.2

Outer Dia. ∅5.0 ∅6.0

5mm 3mm

Ultra-Wide£ Systemwww.osstem.com

Instrument Dimension Indication Instrument Dimension Indication

∅2.0

Page 13: ultra wide system-en - osstem.cz · abutment 16 Surgical Procedures 17 Drilling sequence 20 Insertion depth guide Indication of Ultra-Wide System 4 GS Ultra-Wide System 5 GS Ultra-Wide

24 Total Dental Solution Provider 25

Surgical Procedure How to Maintain Surgical KIT

Cautions Cautions

The operator must check the following items before starting the surgery.

Patient's condition

�Sufficient bone volume and quality prior to treatment �Detailed health status

�Smoking and/or drinking habits �Masticatory pattern and habit

�Status of oral hygiene �Psychological state

�Patient's understanding of implant surgery

Treatment Plan

Discuss treatment options with your patient. Include the pros and cons of each treatment as well as the

procedure involved.

Medical Diagnosis

Check the patient’s health history before implant surgery. Discuss the treatment plan and health history with the

patient’s primary care physician before performing surgery.

Pay Particular Attention to the Following during an Implant Procedure

�Improper upper/lower posterior height �Improper lower anterior width

�Extremely poor bone quality �Congenital or acquired heart conditions

�Ischemic heart patient (angina, myocardial infarction) �High blood pressure

�Patient s distrust of implant treatment

①During surgery, be sure to keep used tools in saline or distilled

water.

② Rinse tools with distilled or running water to remove all blood and

debris.

③ Alcohol wash all tools used in surgery

Caution: Do NOT use hydrogen peroxide

*Exposure to hydrogen peroxide may cause discoloration to laser

markings and/or TiN coating

④ Remove all moisture with a dry cloth or warming fan

⑤ Replace dried tools into the Kit case

(Refer to the color-coding for easy placement.)

⑥ Autoclave Kit for 15 minutes at 132, then store at room

temperature.

Caution

Precautions Immediately after surgery separate, wash and store all tools.

Re-sterilize Kits prior to surgery (132 for 15 minutes)

Hanaro Kits have a 1 year product warranty (after opening); all drills and drivers are recommended for up to 50 uses.