Top Banner
NIR Analyses in fishmeal An Icelandic Network 2014-03-27 Stefan Lundgren Perten Instruments AB Perten Instruments Group
36

Trials on nir perten 2014

Apr 11, 2017

Download

Business

FIFIsland
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Trials on nir perten 2014

NIR Analyses in fishmeal An Icelandic Network

2014-03-27

Stefan Lundgren Perten Instruments AB

Perten Instruments Group

Page 2: Trials on nir perten 2014

AGENDA

• Briefly about Perten, really briefly...

• NIR Analyzers from Perten Instruments

• Parameters analyzed in Fishmeal by NIR

• Results from trial at Matis 2013

• NIR Networks

Page 4: Trials on nir perten 2014

Analytical solutions for Agri Food Industries

Grain

Flour

Food

Feed

Page 5: Trials on nir perten 2014

A long history in NIR

In 1980 Perten Instruments developed its first NIR instrument.

Page 6: Trials on nir perten 2014

Global – Local

Icepack Ehf

Reykjavik

Page 7: Trials on nir perten 2014

DA 7250 NIR Analyzer

• 6-second analysis

• Grains, slurries, powders and more

• No sample preparation

• SEE IT OUTSIDE!

Page 8: Trials on nir perten 2014

DA 7300 Online NIR Analyzer

• Inline analysis

• Grains, pastes, powders and more

• Built-in camera

Page 9: Trials on nir perten 2014

DA 7300 Online NIR Analyzer

• Over-belt analysis

• Grains, meals, powders and more

Page 10: Trials on nir perten 2014

Plus Platform overview

NetPlus

server

NetPlus - Reports

Reports on e.g. production

Production manager

NetPlus - Remote

Instrument performance

and configuration

Instrument manager

NetPlus - Application

Organize calibrations,

find data, export/import

Application specialist

Results Plus

Sample analysis, with GUI

Process Plus

Process analysis,

No GUI

Results Plus?

Future additions

Instrument software Browser software

Page 11: Trials on nir perten 2014

NIR in Fishmeal – Analysis parameters

Parameter Unit Reference method

Moisture % by weight Drying Oven

Protein % by weight, as is

Kjeldahl

Fat % by weight, as is Soxhlet

Ash % by weight, as is Ashing Oven

Salt % by weight, as is Titration

TVN

Cadaverine

mg/100g

g/kg

Steam distillation

HPLC

Page 12: Trials on nir perten 2014

NIR in Fishmeal – Trial at Matis

• Some 200 samples were provided by SVN and Matis

• All samples were scanned on DA7250 • Analyzed for Moisture, Protein, Fat and Ash, using Perten’s global

calibration models

• Results were good but not optimal

• Matis provided samples with Salt and TVN values

• New calibration models were developed by Perten

• Feasibility study carried out on Cadaverine in Fishmeal

Page 13: Trials on nir perten 2014

Moisture

Parameter SEC SEP R2 Range C-set V-set

Moisture 0.43 0.35 0.93 2.1-9.3 184 38

Page 14: Trials on nir perten 2014

Protein

Parameter SEC SEP R2 Range C-set V-set

Protein 0.87 0.83 0.87 65.9-74.3 184 38

Page 15: Trials on nir perten 2014

Fat

Parameter SEC SEP R2 Range C-set V-set

Fat 0.52 0.51 0.95 6.3-15.4 182 38

Page 16: Trials on nir perten 2014

Ash

Parameter SEC SEP R2 Range C-set V-set

Ash 0.39 0.40 0.97 9.5-17.8 184 29

Page 17: Trials on nir perten 2014

Salt

Parameter SEC SECV R2 Range C-set

Salt 0.31 0.55 0.97 2.2-8.4 53

Page 18: Trials on nir perten 2014

TVN

Parameter SEC SEP R2 Range C-set V-set

TVN 1.9 1.9 0.92 0.7-30.0 123 24

Page 19: Trials on nir perten 2014

Cadaverine

Parameter SEC SEP R2 Range C-set V-set

Cadaverine 109 218 0.96 100-1800 35 8

Page 20: Trials on nir perten 2014

NIR Networks

•Basic Ideas

•Advantages

•An example

•Organisation

•Calibration Monitoring

• Standardisation

• Instrument Monitoring

•Costs

Page 21: Trials on nir perten 2014

NIR Networks – Basic ideas

• Ensuring a uniform, fair and cost effective form of quality assurance and basis for payment

• A common accredited reference laboratory

• Rapid analyses based on international standard methods

• Easy-to-use analytical instruments with identical specifications

• Minimized operator influence on results

• ’Approved’ calibration models • All instruments are compared to a

’Master Instrument’

Page 22: Trials on nir perten 2014

NIR Networks - Advantages

• Centrally controlled calibration models and reference analyses

• No local influence on results gives a uniform and fair classification of products

• Simplified quality assurance for the users • Calibration monitoring through the network using

web based software

• Daily check sample

• Powerful tool for implementing an Icelandic standard

• Harmonized quality control on produced fishmeal

Page 23: Trials on nir perten 2014

NIR Networks - Advantages

• Building a common sample database • Better and more robust calibration models gives

reliable analysis results

• Few reference samples per network member -> Lower analysis costs!

• Yearly updates

• Reduced development costs for new analysis parameters and products, eg

• Cadaverine

• Raw materials

• Same calibration models used for both process control and out-loading

Page 24: Trials on nir perten 2014

NIR Network - Organisation

Calibration Center

Perten Applications

Group

Administration Center

NetPlus Network

Manager

Networked

Local instruments

Reference Laboratory

Master instrument

Page 25: Trials on nir perten 2014

NIR Network - Organisation

• Steering Committee • Representatives from users, reference laboratory and instrument

supplier

• Reference Laboratory • Matis Laboratories • Accreditations: ISO 17025, SWEDAC • Certification in acc. With ISO 9001 och ISO 14001 • Controls:

• Ring tests in Iceland and Europe • Third party revisons by customers

• Calibration Center • Perten Applications Group – Stockholm / Matis Reykjavik

• Administration Center • Matis Neskaupstadur

• Technical Support • Perten Technical Group – Stockholm & Icepack - Reykjavik

Page 26: Trials on nir perten 2014

NIR Network – Calibration Monitoring

• The Master Instrument is compared to reference analyses at the accredited laboratory

• Samples are sent in by network members on a regular basis

• Some 100 control samples collected from all participating factories each year

• Results are available on the internet

• NetPlus Reports – Web based reports software

• Updated regularly with new samples

Page 27: Trials on nir perten 2014

NIR Network - Standardisation

• The Master is validated and adjusted against reference analyses

• The Reference Laboratory sends out reference samples once a year for instrument control

• Typically 10 samples per product type

• Scanned on every instrument in the network

• Scans collected via internet by the Admin. Center

• Predicted results are compared to the Master

• Individual instruments are adjusted via internet if needed

Page 28: Trials on nir perten 2014

NIR Network - Standardisation

• The Master is validated and adjusted against reference analyses

9.8

9.9

10

10.1

10.2

10.3

10.4

10.5

Reference lab DA7200 Master

Page 29: Trials on nir perten 2014

NIR Network - Standardisation

• Individual instruments are adjusted against the Master via internet if needed

8,5

10,0

11,5

maste

r

slave

1

slave

2

slave

3

slave

4

slave

5

slave

6

slave

7

slave

8

slave

9

slave

10

Page 30: Trials on nir perten 2014

NIR Network– Instrument monitoring

Continuous monitoring of instrument stability

• Instrument Self-Diagnostics

• Daily Check Samples

• Relevant sample material

• Synthetic sample (Polystyrene)

• Analyzed daily to confirm stability

• Statistics in Results Plus instrument software

Page 31: Trials on nir perten 2014

NIR Network – Outlier handling

• ’Outliers’ during analysis

• Samples that deviate from samples in the calibration sample database

• Spectral difference or difference in content (range)

• Sent to Reference Lab for chemical analysis

• Added to calibration sample database for updating and improving calibration models

Page 32: Trials on nir perten 2014

NetPlus Reports

• Information on all reference samples is available for all participants in the network via internet

Page 33: Trials on nir perten 2014

NIR Network – Costs

Network price for the instrument

• DA7250 NIR Analyzer: € 48 000

Annual fee

• Xx xxx ISK per instrument

Includes

• Yearly standardisation against Master instrument

• Specified number of reference analyses per year

• Updated calibration models

• Annual maintenance service of instruments

• Remote support via internet

Page 34: Trials on nir perten 2014

NIR Network – Summary

Participating in an NIR Network cooperation

• Simplified and cost-effective quality assurance

• Increased credibility for NIR analyses • Quality certified by accredited laboratory

• Reduced time spent and costs for upkeep of NIR analyses for product controls

• Lower costs for developing new NIR analyses

Page 35: Trials on nir perten 2014

Questions?

Page 36: Trials on nir perten 2014

Your partner in Agri Food Quality Control