    Tech Note TN-513-1

    MICRO-MEASUREMENTS

    Measurement of Thermal ExpansionCoefficient Using Strain Gages

    TEC

    H

    N

    O

    TE

    Strain Gages and Instruments

    For technical support, [email protected]

    www.micro-measurements.com119

    Document Number: 11063Revision: 01-Nov-2010

    The thermal expansion coefficient is a very basic physicalproperty which can be of considerable importance inmechan ical and structural design applications of amaterial. Although there are many published tabulationsof expansion coefficients for the common metals andstandard alloys, the need occasionally arises to measurethis property for a specific material over a particular

    temperature range. In some cases (e.g., new or specialalloys, composites, etc.), there is apt to be no publisheddata whatsoever on expansion coefficients. In others, datamay exist (and eventually be found), but may encompassthe wrong temperature range, apply to somewhat differentmaterial, or be otherwise unsuited to the application.

    Historically, the classical means for measuring expansioncoefficients has been the dilatometer. In this type ofinstrument, the difference in expansion between a rodmade from the test material and a matching length ofquartz or vitreous silica is compared1,2. Their differentialexpansion is measured with a sensitive dial indicator,or with an electrical displacement transducer. When

    necessary, the expansion properties of the quartz or silicacan be calibrated against the accurately known expansionof pure platinum or copper. The instrument is normallyinserted in a special tubular furnace or liquid bath to obtainthe required temperatures. Making measurements with thedilatometer is a delicate, demanding task, however, andis better suited to the materials science laboratory thanto the typical experimental stress analysis facility. ThisTech Note provides an alternate method for easily andquite accurately measuring the expansion coefficient of atest material with respect to that of any reference materialhaving known expansion characteristics.

    The technique described here uses two well-matched strain

    gages, with one bonded to a specimen of the referencematerial, and the second to a specimen of the test material.The specimens can be of any size or shape compatible withthe available equipment for heating and refrigeration (butspecimens of uniform cross section will minimize potentialproblems with temperature gradients). Under stress-freeconditions, the differential output between the gages onthe two specimens, at any common temperature, is equal tothe differential unit expansion (in/in, or m/m). Aside fromthe basic simplicity and relative ease of making thermalexpansion measurements by this method, it has the distinctadvantage of requiring no specialized instruments beyondthose normally found in a stress analysis laboratory. This

    technique can also be applied to the otherwise difficult

    task of determining directional expansion coefficients ofmaterials with anisotropic thermal properties.

    Because typical expansion coefficients are measuredin terms of a few parts per million, close attention toprocedural detail is required with any measurementmethod to obtain accurate results; and the strain gage

    method is not an exception to the rule. This Tech Note hasbeen prepared as an aid to the gage user in util izing the fullprecision of the modern foil strain gage for determiningexpansion coefficients. Given in the first of the followingsections is an explanation of the technical principlesunderlying the method. The next section describes,in some detail, the strain-gage-related materials andprocedures in making the measurement. Basically, thelatter consists of essentially the same techniques requiredfor any high-precision strain measurement in a variablethermal environment. Suggested refinements for achievingmaximum accuracy are then given in the following section;after which, the principal limitations of the method aredescribed.

    Principle of The Measurement Method

    When a resistance strain gage is installed on a stress-freespecimen of any test material, and the temperature ofthe material is changed, the output of the gage changescorrespondingly. This effect, present in al l resistance straingages, was formerly referred to as temperature-inducedapparent strain, but is currently defined as thermal output3.It is caused by a combination of two factors. To begin with,in common with the behavior of most conductors, theresistivity of the grid alloy changes with temperature. Anadditional resistance change occurs because the thermalexpansion coefficient of the grid alloy is usually differentfrom that of the test material to which it is bonded.Thus, with temperature change, the grid is mechanicallystrained by an amount equal to the difference in expansioncoefficients. Since the gage grid is made from a strain-sensitive alloy, it produces a resistance change proportionalto the thermally induced strain. The thermal output of thegage is due to the combined resistance changes from bothsources. The net resistance change can be expressed asthe sum of resistivity and differential expansion effects asfollows:

    = + ( )

    R

    RF TG s G G

    (1)

    mailto:micro-measurements%40vishaypg.com?subject=mailto:micro-measurements%40vishaypg.com?subject=
  • 8/12/2019 tn5131tn

    2/12

    TEC

    H

    N

    O

    TE

    For technical questions, [email protected]

    TN-513-1

    Micro-Measurements

    Document Number: 11063Revision: 01-Nov-2010

    www.micro-measurements.com120

    Measurement of Thermal Expansion Coefficient Using Strain Gages

    where:

    R/R= unit resistance changeG= thermal coefficient of resistivity of

    grid material

    s G= difference in thermal expansion coefficientsbetween specimen and grid, respectively

    FG= gage factor of the strain gage

    T= temperature change from arbitrary initialreference temperature

    The indicated strain due to a resistance change in the gageis:

    iI

    R R

    F= /

    (2)

    where: FI= instrument gage factor setting

    Then, the thermal output in strain units can be expressedas:

    T O G S G S G G

    I

    F T

    F/ ( / )=

    + ( )

    (3)

    where:

    T/O(G/S) = thermal output for grid alloy Gonspecimen material S

    Or, in the usual case, with the instrument gage factor setequal to that of the strain gage, so that FI= FG,

    T O G S G

    GS G

    FT/ /( )= + ( )

    (4)

    It should not be assumed from the form of Equation (4)that the thermal output is linear with temperature, sinceall of the coefficients within the brackets are themselvesfunctions of temperature. As an example, typical thermaloutput characteristics for a Micro-Measurements A-alloygage (self-temperature-compensated constantan grid),bonded to steel, are represented by the solid curve inFigure 1. The lot of foil identified in the upper right cornerof the graph was specially processed to minimize thethermal output over the temperature range from about 50to +300F [45 to +150C]. Strain gages fabricated fromthis lot of foil are intended for use only on material suchas steel with a coefficient of expansion of approximately6 x 10-6/F [11 x 10-6/C]. If the gages are installed on some

    other material with a dif ferent coefficient of expansion, theresult is to effectively rotate the curve in Figure 1 about itsreference point at +75F [+24C]. Installation on a materialwith a higher coefficient of expansion than steel will rotatethe curve counterclockwise, while a material with a lowerexpansion coefficient than steel will cause clockwiserotation. For example, the broken curve labeled A in thefigure illustrates the general effect of installing a gage fromthe subject lot on a beryllium alloy having an expansion

    coefficient of about 9 x 10-6

    /F [16 x 10-6

    /C]. Similarly,if a gage from this lot were bonded to a titanium alloywith a somewhat lower expansion coefficient than steel,the thermal output would be shifted in the manner of thebroken curve labeled B.

    The principle of measuring expansion coefficients withstrain gages then becomes evident from Figure 1, sincethe rotation from one thermal output curve to the other isdue only to the difference in thermal expansion propertiesbetween the materials represented by the two curves. Analgebraic demonstration of the principle can be obtainedby rewriting Equation (4) twice; once for the gage installedon a specimen of the test material of unknown expansion

    coefficient S, and again for the same type of gage installedon a standard reference material with a known expansioncoefficient R:

    T O G S G

    GS G

    FT/ /( )= + ( )

    (5a)

    T O G RG

    GR G

    FT/ /( )= + ( )

    (5b)

    Figure 1 Rotation of the thermal output froma strain gage when installed on materials with

    differing thermal expansion coefficients.

    TEMPERATURE IN CELSIUS

    TEMPERATURE IN FARENHEIT

    +400

    -50 0 +50

    24C

    A

    B

    STD

    75F

    +100 +150 -200 +250

    +300

    +200

    +100

    0

    -100

    -100 0 +100 +200 +300 +400 +500

    -200

    -300

    -400

    -500

    APPARENTMICROSTRAIN

    (BasedonInstrumentG.F.

    of2.0

    0)

    Lot No. A38AD497

    http://www.micro-measurements.com/http://www.micro-measurements.com/
  • 8/12/2019 tn5131tn

    3/12

    TEC

    H

    N

    O

    TE

    For technical questions, [email protected]

    TN-513-1

    Micro-Measurements

    Document Number: 11063Revision: 01-Nov-2010

    www.micro-measurements.com121

    Measurement of Thermal Expansion Coefficient Using Strain Gages

    Subtracting Equation (5b) from (5a), and rearranging,

    S R

    T O G S T O G R

    T =

    ( )/ ( / ) / ( / )

    (6)

    Thus, the difference in expansion coefficients, referredto a particular temperature range, is equal to the unitdifference in thermal output for the same change intemperature. Although this technique for measuringexpansion coefficients is widely applicable, and oftenthe most practical approach, there is relatively littleinformation about it in the technical literature. Repre-sentative applications are described in the bibliography tothis Tech Note4,5.

    Measurement Procedures

    Reference Material

    Selection of the material to be used as a reference standardis naturally an important factor in the accuracy ofthe method, as it is for any other form of differentialdilatometry. In principle, the reference material couldbe any substance for which the expansion properties areaccurately known over the temperature range of interest.In practice, however, it is often advantageous to selecta material with expansion properties as close to zero aspossible. Doing this will provide an output signal thatclosely corresponds to the absolute expansion coefficientof the test material, and permits a more straightforwardtest procedure. The thermal expansion of the referencematerial should also be highly repeatable, and stable withtime at any constant temperature. In addition, the elasticmodulus of the material should be great enough thatmechanical reinforcement by the strain gage is negligible.

    An excellent reference material with these and the other

    desirable properties is ULE Titanium Silicate Code 7972,available from Corning Glass Company, Corning, NY14831.* As illustrated in Figure 2, this special glass has anextremely low thermal expansion coefficient, particularlyover the temperature range from about 50 to +350F[45 to +175C)]. It should be noted, however, that thematerial has a low coefficient of thermal conductivity,making it slow to reach thermal equilibrium. For optimumresults, a dwell time of at least 45 minutes should be used ateach new temperature point before taking data. Anotherpotential disadvantage of titanium silicate as a referencematerial is its brittleness, since it will fracture readily ifdropped on a hard surface. Because of the foregoing, alow-expansion metal (such as Invar or a similar alloy) may

    offer a preferable alternative if the alloy has repeatableand accurately known expansion properties over thetemperature range of interest.

    Strain Gage Selection

    The type of strain gage selected for use in measuringexpansion coefficients is also an important consideration,just as it is for stress analysis and transducer applications.Gage selection usually requires weighing a variety offactors which can directly or indirectly affect the suitabilityof a particular gage type to a specified measurement task.To assist gage users in this process, our Tech Note TN-505

    provides extensive background data for gage selection,along with procedures, recommendations, and applicationexamples6. The subject Tech Note should serve as theprimary reference on gage selection, supplemented here byspecial considerations applicable to the measurement ofexpansion coefficients.

    For good accuracy, combined with ease of installation, agage from Micro-Measurements CEA Series is ordinarilya suitable choice. This assumes that the temperatureextremes for the measurements fall within the range of

    Thermal Expansion: 80 to +150C

    Temperature (C)

    L/L(ppm)

    8

    6

    4

    2

    0

    -2

    80 60 40 20 0 +20 +40 +60 +80 +100+120+140

    Thermal Expansion: 0 to 150C

    Temperature (C)

    Code 7972

    L/L(ppm)

    6

    4

    2

    0

    0 10 30 50 70 90 110 130 150

    -2

    Figure 2 Thermal expansion characteristics of the titanium silicate reference material (data source: Corning Glass Company).

    * Also available from Micro-Measurements as Part No. TSB-1.

    See Appendix for specimen dimensions.

  • 8/12/2019 tn5131tn

    4/12

    TEC

    H

    N

    O

    TE

    For technical questions, [email protected]

    TN-513-1

    Micro-Measurements

    Document Number: 11063Revision: 01-Nov-2010

    www.micro-measurements.com122

    Measurement of Thermal Expansion Coefficient Using Strain Gages

    greatest stability and precision for the constantan foil in

    this type of gage [about 50 to +150F (45 to +65C)].If a wider temperature range is involved, a gage from theWK Series becomes the preferred choice. The latter gagetype is somewhat stiffer, however, and consideration ofreinforcement effects may be necessary if the test materialhas a low modulus of elasticity, or the test specimen is thinand narrow.

    In each of the foregoing cases, a 350 gage is preferable inorder to minimize self-heating by the excitation current.The 350 gage is also advantageous in reducing the effectsof small imbalances which may occur due to unsymmetricresistance changes in the leadwires with temperature. Inaddition, it is good practice, when feasible, to employ a

    medium gage length say, 1/8 in [3 mm] or larger formore stable operation and improved heat transfer to thesubstrate.

    Another gage parameter to be specified is the self-temperature-compensation (S-T-C) number. In principle,as indicated by Equation (6), it should not matter whatS-T-C number is selected. Only the difference in thermaloutput, for the same gage type on two different materials,is involved in the expansion calculations. Practically,however, there are two considerations which may influencethe choice. One of these is the availability of the selectedgage in the desired series, gage pattern, and resistance.

    As a rule, the greatest selection of gages is available in the

    06 and 13 S-T-C groups, since these are the most widelyused compensations for stress analysis and transducerapplications. It will often be expedient, therefore, tospecify one of the above for the S-T-C number.

    When expansion measurements must be made over anextended temperature range, or at high or low temperatureextremes, the S-T-C number should be carefully selectedto obtain the best measurement accuracy. It is evidentfrom Figure 1 that, with excessive mismatch between theS-T-C number of the gage and the expansion coefficientof the specimen, the slope of the thermal output curve canbecome very steep at one or both extreme temperatures.Under such circumstances, a small error in temperature(or temperature deviation between the reference and test

    materials) can produce a large error in the thermal output

    signal. Judicious selection of the S-T-C mismatch can beused to simultaneously keep the slopes of the thermaloutput curves for both the test and reference materialsunder reasonably good control in the temperature rangeof interest.

    Almost any single-element linear grid pattern canbe employed for measuring expansion coefficients. Asindicated earlier, however, the two gages one on thereference specimen, and one on the test material must always be well-matched. That is, the gages mustbe identically the same type, and must be from the samemanufacturing lot to assure closely related thermal outputcharacteristic. Both requirements can be met by simply

    using a pair of gages taken from the same package. Gagesof the identical type taken from different packages, buthaving the same lot number, will be equally close in theirthermal outputs. When a still closer relationship is desiredfor greater measurement accuracy, a dual-grid gage patternsuch as the 125MG (Figure 3) can be selected, and the gridscut apart to form two individual gages. The resulting gagesare, in effect, identical twins, and will provide the closestpossible match in thermal output characteristics (as in allother properties).

    Gage Installation

    As noted, one of the advantages of this method is that

    the specimens of the reference and test materials can beof any convenient size or configuration suitable to theavailable heating or refrigeration equipment. In fact, thetwo specimens can even be different in size or shape ifthere is a reason to have them so. In general, however,specimens should be uniform in cross section to minimizetemperature gradients induced during heating or cooling;and the use of flat specimens will make for easier andhigher-quality gage and temperature sensor installations.The specimens should also be large enough in cross sectionso that the strain gage stiffness is negligible comparedto the overall section stiffness. Beyond the foregoing,selection of the specimen dimensions for about the same

    thermal inertia will be helpful in most quickly achievingthe same temperature when both specimens are heated orcooled together.

    Specimen surfaces should be thoroughly cleaned andprepared for bonding as described in Micro-MeasurementsInstruction Bulletin B-129, which includes specific step-by-step procedures for a wide variety of materials7. For bestaccuracy, bonding should be done with a high performanceadhesive such as M-Bond 600 or 610. Both adhesives arecapable of forming thin, hard gluelines for maximumfidelity in transmitting strains from the specimen surface tothe gage. These adhesives are intended for use on relativelysmooth, nonporous surfaces, and should not be used where

    Figure 3 Micro-Measurements

    type 125MG dual-grid strain gage pattern.

    ~3.5x actual size

    http://www.micro-measurements.com/http://www.micro-measurements.com/
  • 8/12/2019 tn5131tn

    5/12

    TEC

    H

    N

    O

    TE

    For technical questions, [email protected]

    TN-513-1

    Micro-Measurements

    Document Number: 11063Revision: 01-Nov-2010

    www.micro-measurements.com123

    Measurement of Thermal Expansion Coefficient Using Strain Gages

    the adhesive is required to fill surface irregularities or to

    seal pores. For the latter conditions, the recommendedadhesive is M-Bond AE-10 or AE-15. In all cases, completeinstructions for applying and curing the adhesive areincluded in the package with the material.

    Extra care is required in the selection of leadwires andtheir attachment to the gages, in order to obtain themost accurate results. Thermally produced resistancechanges in the leadwires will generate circuit outputswhich are indistinguishable from the thermal outputsbeing measured. If these differ in any way between thereference and test specimens, the indicated differentialexpansion data will be in error accordingly. To minimizesuch effects, leadwire resistance should be kept as low as

    possible by employing a generous wire size, and by keepingthe leads short. The wiring should also be the same for bothspecimens in size, length, and routing. If measurementsare to be made on both specimens in the same chamber orliquid bath at the same time, the leadwires should be keptphysically together throughout as much of their lengthas practical. Leadwire insulation must be selected, ofcourse, for compatibility with the temperature range andenvironment encountered in the measurements.

    In attaching leadwires to the gage solder tabs or to solderterminals, the solder joints should be smooth, bright, andfree of spikes or excess solder. The joints should also be asuniform as possible; and the leadwires should be dressed

    the same on both specimens. After lead attachment, thegage installations must be thoroughly cleaned with rosinsolvent to remove all traces of soldering flux and residues.

    The final step in the installation is to apply a protectivecoating system which is appropriate to the expectedtest environment. Since these tests are normallyconducted under short-term laboratory conditions, acoating is selected for basic protection against moisture,dew point condensation in cold tests and minimum/maximum operating temperature range. The coatingrecommendations in the following table also take intoconsideration low reinforcement of the specimen. Fur-ther details on these and other coatings can be found in

    Micro-Measurements Strain Gage Accessories DataBook.

    The process of gage installation has been summarized

    very briefly here, since detailed instructions are suppliedelsewhere in our technical publications. It should beappreciated, however, that proper gage installation is abasic requirement for accurate measurement of expansioncoefficients. In general, gage installations should be of thehighest quality comparable to those found in precisionstrain gage transducers. Care should also be taken that thetwo gage installations, on the reference and test specimens,are as uniform as possible to minimize small physicaldifferences which could affect the differential thermalresponse. If installation questions or problems arise, theuser should consult the our Applications EngineeringDepartment for assistance. Figure 4 is a photograph ofa properly installed strain gage on a metal specimen for

    thermal expansion measurements. A bondable resistancetemperature sensor (see Figure 6) is installed adjacentto the gage to monitor the specimen temperature. Thisphotograph shows the installation just prior to applicationof the protective coating over the gage and temperaturesensor.

    Strain and Temperature Instrumentation

    Basically, any stable precision strain indicator can be usedfor the strain measurements needed in this procedure.Satisfactory instruments for this purpose include theModel P3 and Model 3800 Strain Indicators produced bythe Instruments Division of Micro-Measurements. Beyondthe necessity for instrument precision and stability, it isimportant that the gage excitation voltage be kept lowenough to avoid the effects of self-heating in the gage.Both the Models P3 and 3800 are high-gain instrumentswith low excitation voltages. Using these strain indicators,there is ordinarily no self-heating problem with a gagesuch as the 125MG pattern installed on a metal specimen

    PROTECTIVE COATING

    Operating

    F

    Temperature

    RangeC

    Coating

    +60 to +250

    0 to +150

    100 to +500

    452 to +400

    [+15 to +120]

    [20 to +65]

    [75 to +260]

    [269 to+200]

    M-Coat A or C

    W-1 Wax

    3140 or 3145 RTV

    Two coats M-Bond 43B

    Figure 4 Strain gage (half of the 125MG dual-gage

    pattern, at top) and resistance temperature sensor,

    installed side-by-side on a specimen of test material

  • 8/12/2019 tn5131tn

    6/12

    TEC

    H

    N

    O

    TE

    For technical questions, [email protected]

    TN-513-1

    Micro-Measurements

    Document Number: 11063Revision: 01-Nov-2010

    www.micro-measurements.com124

    Measurement of Thermal Expansion Coefficient Using Strain Gages

    with reasonably good heat-dissipating characteristics.

    When measurements are made with other instrumentshaving higher excitation voltages, or with gages installedon specimens of low thermal conductivity, self-heatingmay be excessive, and the voltage applied to the gage mustbe reduced. Comprehensive background information andguidelines for setting excitation voltages are provided inTech Note TN-5028.

    Either of two basic circuit arrangements can be used inmeasuring expansion coefficients. One of these, shownin Figure 5a, employs separate, three-wire, quarter-bridge circuits for the gages on the reference and testspecimens. With this arrangement, the gage outputsare read individually, and subsequently subtracted to

    determine the differential strain for use with Equation (6).Since the separate circuits permit monitoring the gagesindependently, it is relatively simple to identify the causeof any improper or anomalous strain readings which mayoccur when conducting the test. A disadvantage of thisapproach is that it requires a switch-and-balance unit(when used with a single-channel strain indicator) or atwo-channel instrument.

    The second arrangement (Figure 5b) uses the propertiesof the half-bridge circuit to perform the subtractionelectrically. When the two gages are connected as adjacentarms of the bridge circuit, the instrument output is equalto the difference in the individual thermal outputs. The

    circuit is obviously simpler in terms of both wiring and

    instrumentation, and is direct-reading. Its primarydisadvantage lies in the difficulty of isolating the gagewhich may be malfunctioning when improper operation issuspected.

    In both of the foregoing circuit arrangements, the leadwiresto the gages should be as short as possible, and should be ofthe same wire size and length. Since leadwires #1 and #3 arealways in adjacent arms of the bridge circuit, they shouldbe particularly well-matched and maintained physicallytogether throughout their lengths, to minimize differentialresistance changes which could appear in the instrumentoutput. With a half-bridge circuit such as shown in Figure5b, it is also necessary that leadwire #2 be connected at the

    midpoint of the jumper between the gages. This is done toplace half of the jumper resistance in series with each gagein its respective bridge arm, and thus avoid a false outputsignal due to the thermally induced resistance change inthe jumper wire. It is worth noting that a 6-in [~150-mm]dissymmetry in the wiring whether in leadwires #1 and#3, or in the jumper in AWG 30 [0.25 mm] wire size willcause a false output of about 17per 100F [per 55C].

    Temperature measurement also requires care andconsideration to obtain accurate expansion data. Typically,a temperature-sensing probe is placed immediately adjacentto the gage, and in intimate contact with the specimensurface, to indicate the specimen/gage temperature. This

    procedure assumes that previous verification has been made,by multiple temperature measurements on the specimen asnecessary, to assure uniform specimen temperature underconditions of thermal equilibrium in the test chamber.Since the materials in the reference and test specimensnormally differ in their thermal conductivity and specificheat, it is necessary that the temperatures at both gage sitesbe measured. The temperature must be the same, of course,whenever paired strain readings are made.

    Depending primarily on personal preference andinstrumentation availability, temperatures can be measuredeither with thermocouples or with resistance temperaturesensors. If a thermocouple is employed on each specimen,

    type J (iron-constantan) is preferred, assuming that thetest temperature range is compatible with this type. Thesensing junction should be small, as should the leadwires(in the range of AWG 30 to AWG 26 [0.25 to 0.4 mm]), andpremium grade thermocouple wire should be selected.Heat transfer from the specimen to the junction can beimproved by taping the first 2 to 3 in [50 to 75 mm] of theextension wires to the specimen surface.

    An alternate approach is to use resistance temperaturesensors such as Micro-Measurements TG-Series (Figure6). The temperature sensor looks like a strain gage, andhas essentially the same construction except that thegrid is made from high-purity nickel foil. It is installed

    Figure 5 Strain gage circuits for measuring

    thermal expansion coefficients: (a) separate

    quarter-bridge circuits; (b) half-bridge circuit.

    Test Material

    Reference Matl

    Test Material

    Reference Matl

    1

    2

    3

    1

    2

    3

    1

    2

    3

    eo

    eo

    1

    2

    3

    http://www.micro-measurements.com/http://www.micro-measurements.com/
  • 8/12/2019 tn5131tn

    7/12

    TEC

    H

    N

    O

    TE

    For technical questions, [email protected]

    TN-513-1

    Micro-Measurements

    Document Number: 11063Revision: 01-Nov-2010

    www.micro-measurements.com125

    Measurement of Thermal Expansion Coefficient Using Strain Gages

    with standard strain gage installation procedures, andshould be mounted side-by-side with the strain gage on thespecimen surface. Because it is physically like the straingage, and is attached to the specimen in the same way,the temperature sensor has about the same heat-transfercharacteristics and thermal time constant as the straingage. When used in conjunction with a specially designedpassive resistance network for linearization and signalscaling (Micro-Measurements Type LST), it permits directmeasurement of temperature with any conventional strainindicator. The small size and low stiffness of the TG-Seriestemperature sensor present minimum mechanical restraint

    to the free thermal expansion and contraction of thespecimen.

    Making Expansion Measurements

    For any method of dilatometry, it is always necessarythat the reference and test specimens be exposed to atleast two different temperatures in measuring theexpansion coefficient. The actual means of achieving thedesired temperatures in a particular case depends on thetemperatures involved, and on the available facilities.These may consist, for instance, of ovens, or liquid baths,or various other forms of environmental chamber. Thestrain gage method imposes no special restrictions on the

    nature or design of the chamber. On the contrary, the sizeand shape of the specimen can usually be adapted to suitthe existing facilities. Since the available equipment varieswidely from one laboratory to the next, the followingremarks are limited to the general requirements for anydilatometric temperature chamber.

    Two of the most desirable features of a chamber formeasuring expansion coefficients are uniformity andstability of temperature. To avoid errors due to thedevelopment of thermal stresses in the specimen, thetemperature should be uniform throughout the specimenat the time of measurement. This condition can beestablished only if the chamber temperature at equilibrium

    is essentially uniform at least in the region containing

    the specimens. Temperature stability in the chamber is alsonecessary to permit measuring specimen temperatures andstrains under static, nonvarying conditions.

    Thermal equilibrium in the specimen can be achieved ina chamber equipped with a forced convection system tovigorously circulate the heat-transfer medium past thespecimen surfaces. Heating and cooling rates should also bekept low to minimize temperature gradients perpendicularto the specimen surface. The required condition of uniformtemperature throughout the specimen is difficult to judge,however, and is not necessarily assured by observingequal temperature readings at different points on thesurface. One of the most effective ways to test for control

    over the uniformity of specimen temperature is to make acontinuous plot of strain gage output versus temperatureover the working temperature range in both the heatingand cooling directions. In this process, the temperatureis changed incrementally; and, at each test temperature,after the specimen is evidently in thermal equilibrium,the temperature and thermal output are recorded andplotted. If uniformity of specimen temperature is actuallyachieved, the heating and cooling legs of the plotted curveshould very nearly coincide. If, on the other hand, the twoportions of the curve are significantly separated to form ahysteresis loop, a likely cause is nonuniform temperaturedistribution through the thickness of the specimen. In thelatter case, the heating and cooling rates must be lowered,

    or thermal stabilization times increased, or other measurestaken to essentially eliminate the temperature gradients.

    Means must be provided for supporting the specimens inthe chamber so that friction cannot impede expansion orcontraction. In some cases, a simple way to accomplishthis is to suspend the specimens from one end. Althoughthe specimen may be strained slightly by its own weight,the strain is constant (as long as the elastic modulus isessentially constant), and does not affect the change inthermal output with temperature. If the elastic modulusof the test material changes significantly over the rangeof temperatures to be encountered, the error due to thiseffect must be evaluated to determine the suitability of

    the method. Another approach is to lay the specimens onthe floor of the chamber or compartment, supported by alayer of fiberglass cloth or some other low-friction medium.When this method is used, its effectiveness should beverified by observing the behavior of the thermal outputas the specimen is cycled through the working temperaturerange. Erratic output, hysteresis, or lack of repeatabilitymay indicate excessive friction.

    Before performing actual measurements to determine thecoefficient of expansion, the entire system, including bothspecimens (with gages installed and power applied), shouldbe stabilized by cycling several times to temperatures atleast 10F [5C] above the highest, and below the lowest,

    Figure 6 Micro-Measurements TG-Series

    ETG-50B/W bondable temperature sensor.

  • 8/12/2019 tn5131tn

    8/12

    TEC

    H

    N

    O

    TE

    For technical questions, [email protected]

    TN-513-1

    Micro-Measurements

    Document Number: 11063Revision: 01-Nov-2010

    www.micro-measurements.com126

    Measurement of Thermal Expansion Coefficient Using Strain Gages

    test temperatures. One of the reasons for this procedure

    is that residual stresses are generally present in all of thecomponents the reference and test specimens, the gagesas manufactured and installed, the leadwires, etc. Thermalcycling is intended to relax and/or redistribute any residualstresses which might otherwise change during the test andcause the data to be nonrepeatable. The cycling procedureshould be performed at low enough rates of temperaturechange to minimize thermal stresses in the specimens dueto temperature gradients. Otherwise, the thermal stress,superimposed on the residual stress, may cause yielding,and thus defeat the purpose of the cycling.

    Normally, after the second or third stabilizing cycle,the thermal output at any given temperature should be

    highly repeatable. If not, and if the lack of repeatabilityis significant compared to the accuracy required from thetest, the sources of the variability must be found. In suchcases, the problem may be associated with the temperature,or the strain, or both. Careful re-reading of this Tech Notemay provide the clue for finding and correcting the trouble.Further assistance, if needed, can be obtained from ourApplications Engineering Department.

    Following stabilization, verified by reproducible strainindications throughout the temperature range, the user isready to perform the final measurements for determiningthe thermal expansion properties of the test material.When the oven or other chamber is such that only a single

    specimen can be accommodated, the two specimens aretested one-at-a-time, using the circuit of Figure 5a. Theresulting two sets of thermal output data are subtracted(and the difference divided by the temperature change) asindicated by Equation (6) to give the differential thermalexpansion coefficient. With the preferable arrangement,having both specimens together in the chamber, themeasurements can be made separately as in Figure. 5a,or the differential thermal output can be read directly asshown in Figure 5b.

    Special Precautions and Refinementsfor Improving Accuracy

    When attempting to achieve greater and greater accuracywith the strain gage method (or with any method), it isnecessary to examine ever smaller effects which mayintroduce errors. In some instances, these second-ordererrors are well-defined, systematic in nature, and responsiveto routine procedures for correction or elimination. Inothers, the cause-and-effect relationship is more nebulous,and error reduction is accomplished primarily by techniquerefinement i.e., by removing or minimizing all of theknownpossiblesources of error.

    An example of a readily correctable inaccuracy (in certaincases) is the error due to transverse sensitivity. This error

    arises because the strain field induced in the gage grid bythe difference in thermal expansion between the specimenand grid [Equation (1)] is generally different from thatemployed in gage factor calibration9. When both thereference and test materials are isotropic in their thermalexpansion properties, the transverse-sensitivity error,which is ordinarily quite small, can be corrected for rathereasily. Although not derived here, correction can be madeby multiplying the differencein thermal outputs [Equation

    (6)] by the factor (1 0.285 Kt)/(1 + Kt), where Kt is thedecimalized transverse sensitivity of the gage in use. Thiscorrection factor is not applicable to orthotropic materials,for which case differential thermal outputs between areference gage and two perpendicularly oriented specimengages are required to correct for transverse sensitivity.

    Another minor error source is the variation of gage factorwith temperature. The gage factor specified for Micro-Measurements strain gages is measured at +75F [+24C].At any other temperature it is slightly different. Withconstantan gages, for example, the gage factor variesdirectly with temperature, at a rate of about 0.5% per 100F[0.9% per 100C]. In contrast, the gage factor of K-alloy

    (modified Karma) gages varies inversely with temperature.The rate of change depends on the S-T-C number of thegage, but is generally in the range from 0.5 to 1.0% per100F [0.9 to 1.8% per 100C]. Representative plots ofgage factor variation with temperature are illustrated inFigure 7 for both types of gages. The technical data sheetcontained in each gage package includes data for the gagefactor variation applicable to that gage type.

    Complete elimination of the small error introduced bygage factor variation is not always feasible, but first-ordercorrection, to remove most of the error, is relatively simple.When expansion measurements are made incrementally

    Figure 7 Gage factor variation with temperature

    (typical) for A- and K-alloy strain gages.

    TEMPERATURE IN CELSIUS

    TEMPERATURE IN FARENHEIT

    +2.0%

    0 +50

    24C

    03

    09

    A-alloy

    K-alloy

    75F

    +100 +150 -200 +250

    +1.0%

    0

    0 +100 +200 +300 +400 +500

    1.0%

    2.0%

    GAGE

    FACTORV

    ARIATION

    (Typical)

    http://www.micro-measurements.com/http://www.micro-measurements.com/
  • 8/12/2019 tn5131tn

    9/12

    TEC

    H

    N

    O

    TE

    For technical questions, [email protected]

    TN-513-1

    Micro-Measurements

    Document Number: 11063Revision: 01-Nov-2010

    www.micro-measurements.com127

    Measurement of Thermal Expansion Coefficient Using Strain Gages

    across the working temperature range, the differential

    thermal output for each increment in temperature can becorrected individually. This is done by multiplying thedifference in indicated thermal outputs from the specimenand reference gages by the factor 1/(1 + FG). The term FGin the foregoing is the decimalized change in gage factor(with sign) corresponding to the middle temperature ofeach measurement increment.

    Sometimes, the average differential expansion coefficientis to be determined over the full temperature range bymaking only two sets of measurements, at the temperatureextremes. The same correction procedure can be applied,using the FG for the mid-range temperature, but it willbe much less effective because the thermal output is a

    nonlinear function of temperature.When the leadwire resistance can be kept very low,as recommended in the preceding section, the signalattenuation (desensitization) caused by the inertresistance in series with the gage should be negligible. If, onthe other hand, the series resistance is greater than about 1percent of the gage resistance, the user who is striving formaximum accuracy may wish to perform a correction. Forthis purpose, the indicated thermal outputs are multipliedby the factor (RG+ RL)/RG, where RGis the gage resistance,and RLis the leadwire resistance in series with the gage inthe same arm of the bridge circuit. An alternative, fordirect reading of corrected strains, is to set the gage factor

    control of the instrument at FGx RG/(RG+ RL), where FGisthe specified gage factor of the gages in use.

    The supposition is made, in the strain gage method ofmeasuring expansion coefficients, that if the two gages(and gage circuits) behave identically, then any difference intheir outputs can be due only to the difference in expansionproperties between the reference and test specimens. It

    is obvious, therefore, that the highest accuracy will be

    achieved by minimizing all differences in gage behavior.For this reason, as noted earlier, the thermal outputcharacteristics of the gages should be as nearly the same aspossible. However, two nominally identical gages from thesame manufacturing lot do not especially have identicalthermal outputs. Instead, as shown in Figure 8, thereis a tolerance on the thermal output.* Almost all of thetolerance can be removed by splitting a dual-element gage(such as the 125MG pattern) to make a pair of twin gages,and this procedure is always recommended when highaccuracy is the goal. The same reasoning underlies therepeated emphasis in this Tech Note on the uniformity ofgage installations. Identical instal lation procedures shouldbe used for both gages; and, ideally, there should be no

    visible differences in the completed installations.

    The remaining areas of possible refinement for improvedaccuracy are primarily associated with the measurementsprocedures. Each of the items in the following checklist canbe considered, and steps taken as necessary to satisfy thedesired conditions:

    a) stable, accurate instrumentation, for both temperatureand strain.

    b) high-quality, stable gage installations, exhibitingnegligible drift over the operating temperature range.

    c) gage excitation at a level low enough to avoid self-

    heating effects.d) thermal stabilization of specimens, gages, and wiring

    prior to making expansion measurements.

    e) assurance of thermal equilibrium in the specimenswhen measurements are made.

    f) avoidance of significant thermal stresses duringheating and cooling.

    g) elimination of frictional effects preventing freeexpansion and contraction.

    Except for the absolute accuracy of the instrumentation,the degree to which the foregoing conditions have been metcan be judged quite well by the repeatability of the data.

    Highly reproducible data generally indicate that the systemis functioning properly, and that random error sources arewell-controlled.

    After it has been demonstrated that the measurementsystem and procedures are suitable for obtaining closelyreproducible data from a single specimen, considerationshould be given to the question of variation in thermalproperties from specimen to specimen. The usual purposeof expansion-coefficient measurements is to determinethe nominal value which is representative of a particularmaterial. But the thermal and other physical propertiesof any material tend to vary randomly from specimen to

    TEMPERATURE IN CELSIUS

    TEMPERATURE IN FARENHEIT

    +400

    -50 0 +50

    24C

    A

    75F

    +100 +150 -200 +250

    +300

    +200

    +100

    0

    -100

    -100 0 +100 +200 +300 +400 +500

    -200

    -300

    -400

    -500

    THERMALOUPUT

    m/m

    (BasedonInstrumentG

    .F.of2.00)

    Uncertainty:0.27(m

    /m)C-1

    * See Tech Note TN-504.

    Figure 8 Tolerance band for the thermal output of randomly

    selected A-alloy strain gages from the same manufacturing lot.

  • 8/12/2019 tn5131tn

    10/12

    TEC

    H

    N

    O

    TE

    For technical questions, [email protected]

    TN-513-1

    Micro-Measurements

    Document Number: 11063Revision: 01-Nov-2010

    www.micro-measurements.com128

    Measurement of Thermal Expansion Coefficient Using Strain Gages

    specimen within a lot, and still more widely from lot to

    lot. Since such variation is not subject to the control ofthe user, it becomes necessary to use statistical samplingtechniques, with a sample size large enough to provide anadequate estimate of the mean and standard deviation.Variability in thermal properties is apt to be particularlygreat in materials such as plastics and composites.

    The mechanical and thermal properties of some materials(e.g., graphite, titanium 6A14V, composites with orientedfiber reinforcement, etc.) are highly directional. In suchcases, orientation of the strain gage on the specimen (withrespect to the natural axes of the material, as determinedby the rolling direction, fiber orientation or otherwise)is critical if the directional expansion coefficient is to

    be measured. When it is impossible to determine thedirections of the natural material axes, it may be necessaryto make measurements over a wide range of angles todefine the distribution of the expansion coefficient, or toobtain a rough, integrated average value.

    Limitations

    The strain gage method of differential dilatometry hasvery few special l imitations. Of these, the principal one forsome types of studies may be the allowable temperaturerange. Constantan gages, for instance, should be used forhigh-accuracy measurements only within a temperaturerange from about 50 to +150F (45 to +65C). Higher

    temperatures normally require the use of K-alloy gages,which can provide accurate strain measurements fromapproximately 50 to +400F (45 to +205C). Withspecial techniques, these temperature ranges can sometimesbe extended, depending on the circumstances. Usersshould consult with the Micro-Measurements ApplicationsEngineering Department for recommendations.

    Mechanical reinforcement of the specimen by the straingage can also be a limitation in some instances. Whenthe test specimen is made from a material such as plastic,with a very low modulus of elasticity, the stiffness of thegage may perturb the local strain field and introduce asizeable error. With metal specimens, the reinforcement

    effect is ordinarily negligible unless the specimen is so thinand narrow that the gage stiffness represents a significantfraction of the overall section stiffness.

    Other limitations are generally those common to allmethods of differential dilatometry. For example, theexpansion coefficient of the test material can never bedetermined to greater accuracy than that of the referencematerial. Similarly, the measurements can be no moreaccurate than the instrumentation used to indicate thetemperatures and strains.

    Summary

    This Tech Note has described a simple, straightforwardmeans of measuring the expansion coefficient of a testmaterial relative to that of any reference material havingknown expansion properties. The method is particularlywell-suited to the stress analysis laboratory, since itusually requires no special instrumentation, techniques,or materials not already available in such a facility.Considerable attention has been given here to proceduraldetails aimed at extracting the utmost accuracy fromthe method. Most of the recommended procedures,however, should represent standard practices for a stress

    laboratory which is accustomed to making precisionstrain measurements in a variable thermal environment.Even when expedience dictates somewhat less rigorousprocedures, the method can be used to quickly and easilymeasure thermal expansion coefficients with sufficientaccuracy for many engineering purposes.

    References

    1. American Society for Testing and Materials, StandardTest Method for Linear Expansion of Metals, ASTMStandard No. B95-39.

    2. American Society for Testing and Materials, Linear

    Thermal Expansion of Rigid Solids with a VitreousSilica Dilatometer, ASTM Standard No. E228-71.

    3. Micro-Measurements, Tech Note TN-504, StrainGage Thermal Output and Gage Factor Variation withTemperature, 1989.

    4. Finke, T. E., and T. G. Heberling, Determination ofThermal Expansion Characteristics of Metals UsingStrain Gages, Proceedings, SESA (now, SEM), Vol.XXV, No. 1, 1978, pp. 155-158.

    5. Poore, M. W., and K. F. Kesterson, Measuring theThermal Expansion of Solids with Strain Gages,Journal of Testing and Evaluation, ASTM, Vol. 6, No. 2

    (March 1978), pp. 98-102.

    6. Micro-Measurements, Tech Note TN-505, Strain GageSelection Criteria, Procedures, Recommendations,1989.

    7. Micro-Measurements, Bulletin B-129, Surface Pre-paration for Strain Gage Bonding, 1976.

    8. Micro-Measurements, Tech Note TN-502, OptimizingStrain Gage Excitation Levels, 1979.

    9. Micro-Measurements, Tech Note TN-509, Errors Dueto Transverse Sensitivity in Strain Gages, 1982.

    http://www.micro-measurements.com/http://www.micro-measurements.com/
  • 8/12/2019 tn5131tn

    11/12

    TEC

    H

    N

    O

    TE

    For technical questions, [email protected]

    TN-513-1

    Micro-Measurements

    Document Number: 11063Revision: 01-Nov-2010

    www.micro-measurements.com129

    Measurement of Thermal Expansion Coefficient Using Strain Gages

    APPENDIX

    REFERENCE INFORMATIONI. Specification for CORNING GLASS WORKS Titanium Silicate, Code 7972 ULE thermal expansion coeffecient

    Control limit: +40 to +95F [+5 to +35] 0.00 0.017 x 10-6/F [0.00 0.03 x 10-6/C]

    Typical values: +32 to +390F [0 to +200C] 0.017 0.017 x 10-6/F [0.03 0.03 x 10-6/C]

    150 to +390F [100 to +200C] 0.017 0.017 x 10-6/F [0.0c 0.03 x 10-6/C]

    Tolerance within one specimen purchased from Micro-Measurements (Part No. TSB-1):

    +40 to +95F [+5 to +35C] 0.000.008 x 10-6/F [0.00 0.015 x 10 -6/C]

    This tolerance also applies to typical values noted above.

    Micro-Measurements specimen size: 6 x 1 x 0.25 in [155 x 30 x 6.5 mm]Micro-Measurements specimen finish: 80 grit

    II. Thermal Output Scatter of Micro-Measurements Strain Gages

    All data are based on a 2or 95% confidence level over the temperature range of +32 to +350F [0 to +175C]

    Single-element A-alloy gages: 0.15in/in/F [0.27m/m/C]

    Single-element K-alloy gages: 0.25in/in/F [0.45m/m/C] EA-XX-125MG-120 with one grid on Code 7971 and other on unknown material: 0.03in/in/F [0.05m/m/C]

    WK-XX-125MG-350 used as described for the EA gage: 0.06in/in/F [0.10m/m/C]

    III. Correction for Transverse Sensitivity

    With Kt, in decimal form, multiply the parenthetic expression [T/O(G/S) T/O(G/R)] in Equation (6), by(1 0.285 Kt ) /(1 + Kt) for isotropic materials only.

    IV. Correction for Gage Factor vs. Temperature

    For any temperature increment, multiply the parenthetic expression [T/O(G/S) T/O(G/R)] in Equation (6), by1/(1 + FG). The termFGin decimal form, corresponds to the midpoint of the temperature increment over whichthermal output measurements are made.

    V. Correction for Leadwire Resistance (RL) for a Single Gage in a Three-Wire Configuration

    RLis the resistance of a single leadwire in the three-wire connection to the instrument. To avoid the tedious task ofcorrecting all individual readings by the factor (RG+ RL)/RG, it is much simpler to adjust the gage factor setting ofthe instrument to F1= FGxRG/(RG+ RL).

    To evaluate the need for this correction, the approximate lead resistances for typical Micro-Measurementscables are:

    326-DFV, 326-DTV: 0.043 ohms/ft [0.141 ohms/m]

    330-DFV, 330-FFE, 330-FJT, 330-FTE: 0.108 ohms/ft [0.354 ohms/m]