Prague, 11 September 2019r Magdalena Marcińska, Agnieszka Parys-Proszek, Maria Wróbel, Tomasz Kupiec Internal validation study of the Applied Biosystems TM NGM Detect TM kit
38

Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

Apr 21, 2020

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

Prague, 11 September 2019rMagdalena Marcińska, Agnieszka Parys-Proszek, Maria Wróbel, Tomasz Kupiec

Internal validation study of the Applied Biosystems TM NGM DetectTM kit

Page 2: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

Workflow

Magnetic extraction

Organic extraction(bones,teeth)

Quantifiler Trio

NGMDetectGlobalFilerNGM

Electrophoresis3500xL

X2

Page 3: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

SGM, Identifiler - 140 ’

Identifiler Plus, NGM, NGMSelect - 117’

GlobalFiler – 59’

NGMDetect –34’

Amplification time

Page 4: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

Main characteristics

• Higher sample input volume:

NGMDetect: 15µl 500pg

NGMSelect: 10µl 1000 pg

GlobalFiler: 15µl 1000 pg

• Six color approach• 16 autosomal STRs, Amelogenin, additional Yindel• 7 mini-STRs below 230 bp• Internal quality control system

• Newly designed primers except for D12S391, THO1 primers – share with GlobalFiler and NGMSelect kits

• Rearrangement of the loci - Double typing procedure – complementary marker positions

Page 5: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

Rearrangement of loci in NGM-Detectand NGM – SElect kits

Optimized positioning of SE33 with lower molecular weight range

www.thermofisher.com

Page 6: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

New features: IQC, Yindel

IQC system - two synthetic DNA fragment: LMW (IQCS) and HMW (IQCL)- quality control of PCR and sample’s quality check

HMW (IQCL)LMW (IQCS)

Y INDEL

Page 7: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

Internal validation experiments

- Sensitivity,- Reproducibility,- Accuracy,- Mixture study,- Casework samples analysis – bones samples,

Threshold set-up:LOD, LOQ, Stochastic threshold, Stutter ratio,

Procedure validation performed according to SWGDAM guidelines

Page 8: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

Sensitivity – 500 pg

-Three replicates of control DNA 007 in serial dilutions

Page 9: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

Sensitivity - 250 pg

Page 10: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

Sensitivity – 125 pg

Page 11: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

Sensitivity - 63 pg

Page 12: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

Sensitivity - 31 pg

Page 13: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

Sensitivity - 16 pg

Page 14: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

Sensitivity - 7 pg

Page 15: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

Sensitivity - stochastic effect (31 pg)

Page 16: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

Sensitivity - stochastic effect (31 pg)

Page 17: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

Sensitivity - stochastic effect (31 pg) Sensitivity - stochastic effect (31 pg)

Page 18: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

Sensitivity - summary

PHR > 60%

No results

PHR < 60%

Page 19: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

Mixture proportion

Control 007 Control 9947

1 1

1 2

1 5

1 8

1 10

1 20

1 30

1 50

1 1

2 1

5 1

8 1

10 1

20 1

30 1

50 1

NGM Detect Control DNAXY 007

Control DNAX 9947A

D2S1338 20,23 19,23

SE33 17,25.2 19,29.2

D16S539 9,10 11,12

D18S51 12,15 15,19

TH01 7,9.3 8,9.3

D12S391 18,19 18,20

D3S1358 15,16 14,15

FGA 24,26 23,24

Y-indel 2

Amelogenina X,Y X

VWA 14,16 17,18

D21S11 28,31 30

D18S51 12,15 15,19

D2S441 14,15 10,14

D8S1179 12,13 13

D19S433 14,15 14,15

D22S1045 11,16 11,14

D10S1248 12,15 13,15

Mixture analysisMixture analysis

Page 20: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

Mixture analysis

1:1

Mixture analysis Mixture analysis

Page 21: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

Mixture analysis

Page 22: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

Mixture analysis

Page 23: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

Mixture analysis

STR n-4 n-2 n+4 n-3 n+3D2S1338 0,1481 - - - -

SE33 0,1928 0,0798 0,085 - -D16S539 0,1461 - - - -D18S51 0,1593 - 0,055 - -TH01 0,0507 - - - -

D12S391 0,1963 - - - -D3S1358 0,1533 - - - -

FGA 0,1375 - 0,0476 - -VWA 0,1744 - - - -

D21S11 0,1408 - 0,0408 - -D1S1656 0,1751 0,0357 0,0509 - -D2S441 0,0889 - - - -D8S1179 0,1276 - 0,0303 - -D19S433 0,1271 - - - -D22S1045 - - - 0,1645 0,0919D10S1248 0,1588 - - - -

NGM DetectInj. 20”

NGM DetectInj. 10”

LOQ600 (B)

500 (G,Y, P)1000 (R)

250 (B, G,Y, P)450 (R)

LOD 250 (B, G,Y, P)500 (R)

120 (B, G,Y, P)200 (R)

Stochastic 2196 1014

Stutter:Thresholds LOD, LOQ:

Page 24: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

Efficiency comparison of GlobalFiler/NGM and NGM Detect kits in different types of

casework samples.

Page 25: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

Casework sample analysis – bone, teeth samples

12 bone samples and 5 teeth samples with different degradation level

0

20

40

60

80

100

1.00 1.25 1.30 1.50 2.59 3.84 3.97 4.40 5.95 7.99 9.46 10.80 15.61 18.90 22.70 27.00 125.00

% t

yped

mar

kers

Degradation Index

samples with DNA input below 0,5ng

NGM Detect

Page 26: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

Casework samples analysis – bone sampleGlobalFiler

NGM-Detect

Page 27: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

GlobalFiler

Casework samples analysis – tooth DI=8

D16S539SE33D2S1338

D2S441D10S1248

Page 28: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

NGM Detect

Casework samples analysis – tooth DI=8

D2S441D10S1248

D16S539SE33

D2S1338

Page 29: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

• NGM Detect = 11 STRs

• GlobalFiler = 13 STRs

• NGM Detect & GlobalFiler = 18 STRs

NN & mother

LR=263

LR=85

LR=12135

Casework samples analysis – tooth DI=8

Statistical parameteres in double amplification

Page 30: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

GlobalFiler

Casework samples analysis – petrous bone DI=19

D2S441D10S1248

D16S539SE33

D2S1338

Page 31: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

NGM Detect

Casework samples analysis – petrous bone DI=19

D2S441D10S1248

D16S539SE33

D2S1338

Page 32: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

• NGM Detect = 11 STRs

• GlobalFiler = 14 STRs

• NGM Detect & GlobalFiler = 18 STRs

Casework samples analysis – petrous bone DI=19

Increased number of positively typed markersin double amplification strategy

Page 33: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

Background artifact peaks in analyses of old bones

Page 34: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

Background artifact peaks in analyses of old bones

Page 35: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

New formulation: NGM – Detectchanges to the IQC large and THO1 primers

NGM-Detect14 STRs

Global Filer12 STRs

NGM-Detect& GF

14 STRs

Page 36: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

New formulation: NGM – Detect –changes to the IQC large and THO1 primers

NGM-Detect10 STRs

Global Filer4 STRs

NGM-Detect& GF

12 STRs

Page 37: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

1. Short amplification time2. Higher input volume3. Rearrangement of the loci (e.g. D2S1338, SE33)4. Quality control markers (IQCS, IQCL) - a good control

of amplification proccess

• Comparable results of NGM Detect and GlobalFiler,NGM with good quality samples

• NGM Detect kit is particularly useful for the analysis ofchallenging samples with incomplete profiles generatedwith GlobalFiler/NGM

• NGM Detect kit as a supplemental kit – dualamplification strategy

Summary – main advantages of NGM-Detect

Page 38: Thermo Fisher Scientific - Internal validation study …...Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo

Thank you for attention!

Acknowledge

mgr Agnieszka Parys-Proszekmgr Maria Wróbeldr Tomasz Kupiec

When used for purposes other than Human Identification or Paternity Testing the instruments and software modules cited are forResearch Use Only. Not for use in diagnostic procedures.Thermo Fisher Scientific and its affiliates are not endorsing, recommending, or promoting any use or application of Thermo FisherScientific products presented by third parties during this seminar.Information and materials presented or provided by third parties are provided as-is and without warranty of any kind, includingregarding intellectual property rights and reported results.Parties presenting images, text and material represent they have the rights