PCR Instrumentation 9 9 Thermal Cyclers 01 Our performance, accuracy and simplicity define our status as the reliable choice for modern reseachers Thermo currently offer one of the most extensive ranges of authorised PCR machines. From the low through- put PCR Sprint to our sophisticated robot compatible MultiBlock System, our range is designed to provide value for money features that meet the demands of modern researchers. PCR Instrumentation Designed to offer performance, accuracy and simplicity User-friendly software interfaces Outstanding block performance Robot compatibility Unique Active Tube Temperature control Flexible sample format Licensed thermal cycling Thermo has been manufacturing thermal cyclers since 1986. During this period, we have cont- inually improved every aspect of thermal cycling from speed to temperature uniformity, reliability to flexibility. With this experience and our program of reviewing and imple- menting innovative design features, we have become a leading supplier of thermal cycling equipment world- wide.
13

Thermal Cyclers - PCR Instrumentation

Jan 12, 2023

Download

Documents

Khang Minh
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Thermal Cyclers - PCR Instrumentation

PCR Instrumentation 9 9

Thermal Cyclers 01 Our performance, accuracy and simplicity define our status as thereliable choice for modern reseachers

Thermo currently offer one of themost extensive ranges of authorisedPCR machines. From the low through-put PCR Sprint to our sophisticatedrobot compatible MultiBlock System,our range is designed to providevalue for money features that meetthe demands of modern researchers.

PCR Instrumentation

• Designed to offer performance, accuracyand simplicity

• User-friendly software interfaces• Outstanding block performance• Robot compatibility• Unique Active Tube Temperature control• Flexible sample format• Licensed thermal cycling

Thermo has been manufacturingthermal cyclers since 1986. During this period, we have cont-inually improved every aspect ofthermal cycling from speed to temperature uniformity, reliability to flexibility. With this experience andour program of reviewing and imple-menting innovative design features,we have become a leading supplierof thermal cycling equipment world-wide.

CAT03 A4_Innenseiten 1 16.12.2003 11:53 Uhr Seite 9

Page 2: Thermal Cyclers - PCR Instrumentation

• • • • •

• • •• • •

• • • •• • ••

• •• •• •

PCR Instrumentation 09 10

Px2 Thermal Cycler• Advanced personal thermal cycler

capable of performing anything fromsimple to the most sophisticated PCR protocols

• Accurate temperature control• Multiple interchangeable block

options: 0.5ml and 0.2ml standard and gradient 96 well blocks, 384 well and in situ blocks

PCR Sprint Thermal Cycler • Affordable, low throughput personal

thermal cycler• Small footprint• Two interchangeable block options:

24 x 0.2ml tubes and 20 x 0.5ml tubes

Instrument Selection Chart

MBS Satellite Thermal Cycler• Flexible, computer-controlled PCR

system that allows you to perform anything from a few to thousands ofPCRs on the one system

• Start-up kits available• Full traceability of achieved PCR

parameters• Multiple satellite block options:

0.5ml and 0.2ml standard and gradient 96 well blocks, as wellas a 384 well block

MBSR Robotic Satellite Thermal Cycler • Robot compatible MBS satellite

thermal cycler• Unique CD drawer mechanism

provides superior robot compatibility• Two block options: Gradient 96 well

and standard 384 well

1 2 3 4 5

PCR Sprint PxE Px2 MBS MBSR

Active tube control

Gradient block option

96 well block option

384 well block option

in situ flat block option

Interchangeable blocks

Computer controlled user interface

Traceable PCR runs

Robot compatible

Price

PxE Thermal Cycler NEW The new PxE Thermal Cycler PCR – nothing more, nothing lessAll the essential features required for accurate thermal cycling

• Robust unit providing many of thefeatures evident on more sophistica-ted machines at a lower cost

• Two fixed block options: 0.2ml and 0.5ml standard 96 well blocks

NEW

CAT03 A4_Innenseiten 1 16.12.2003 11:53 Uhr Seite 10

Page 3: Thermal Cyclers - PCR Instrumentation

Temperature Control (Fig.1)The key to the design of any thermalcycler is uniform and accurate temperature control. Thermo's thermalcyclers can use up to five controlmodes for monitoring temperatureduring PCR reactions: Active TubeControl, Simulated Tube Control,Simulated Plate Control, SimulatedSlide Control and Block Control.

Temperature Uniformity (Fig.2)It is paramount for reproducibleresults to attain a uniform temperatureacross the block. In Thermo thermalcyclers this is monitored across theentire block during each stage of thethermal cycling protocol. Every instrument is calibrated usingequipment traceable to the UKNational Physical Laboratory stand-ards. A dynamic uniformity test isconducted as part of Thermo's QualityAssurance procedure to ensure theymeet this specification prior todispatch.

Temperature Control Options

Temperature Gradients (Fig.3)By using our gradient blocks, it is quickand easy to optimise primer to template annealing temperatures.Multiple annealing temperatures canbe assessed in one experiment bydefining a temperature gradient of upto 15°C across the block. The optimisedtemperature can then be transferreddirectly to standard cycling due to ourinnovative software and block design.

Active Tube Control (Fig.4)Thermo's unique method of choice formonitoring and controlling any thermalcycling reaction is to measure theactual temperature of the reaction mix during PCR. When this method isemployed, the machine actively com-pensates for differences in the programmed temperature and theactual temperature achieved in thereaction mix by adjusting the blocktemperature. This method of tempera-ture control ensures that the program-med temperatures and dwell times areaccurately achieved, resulting in optimal PCR performance.

Simulated Tube/Plate and Slide Control(Fig.5)When there is a requirement to useevery position within a block, werecommend that Simulated Control beused. We have developed advancedsoftware algorithms (Simulated Tube,Plate and Slide Controls) that achievetemperature accuracy similar to ActiveTube Control for tube, 96 / 384 wellplate and slide formats. The softwareconstantly calculates the sample tem-perature, factors in the sample volumeand then adjusts the block temperatureto ensure that the sample reaches theprogrammed temperature.

Block Control (Fig.6)Our thermal cyclers can also use BlockControl, which is the sole controlmethod used on some thermal cyclers.Block Control is best used for prolon-ged static incubations. If Block Controlis used during short incubation peri-ods, such as during a PCR reaction, thereaction mix is at the set temperaturefor a considerably shorter time thanthat programmed, which could lead toa reduced yield of PCR product.

Flexible Network of Thermal CyclingBlocksThe MultiBlock System (MBS) compris-es a central computer with sophisti-cated software, which can be net-worked to up to 30 individual satelliteblocks. The choice of the number andtype of each block is completelyflexible, giving you the freedom to customise your PCR system. Satellite blocks are available in 0.2ml(gradient and standard), 0.5ml (gradi-ent and standard) and 384 well for-mats, all of which can be added singularly providing further flexibility.

Accurate Temperature ControlOptimal PCR performance can be ob-tained in our blocks because of theiraccurate and reproducible temperaturecontrol. Our Active or Simulated tem-perature control methods account for

Benefits• Computer-controlled network of

thermal cycling blocks for flexible system approach to PCR

• Capacity for 1 to 30 blocks networkedto one PC for flexible throughput

• Multiple block options for flexibilityof consumable format and easyprotocol optimisation

• Accurate temperature control for optimal PCR performance and reproducibility

• Software interface for enhanced graphics and user friendliness

• Software features for traceability in line with GLP

• Robot-compatible version (MBSR) for automated sample handling

• Satellite blocks available separatelyfor flexible addition to system

• Start-up kits available for low costaccess to the technology

MultiBlock System (MBS)

CAT03 A4_Innenseiten 1 16.12.2003 11:53 Uhr Seite 11

Page 4: Thermal Cyclers - PCR Instrumentation

PCR Instrumentation 11 12

Fig.1

Fig.5

Fig.6

Fig.4

Fig.3

Fig.2

the thermal lag that occurs when heatis transferred from our eight peltier de-vices to the sample mix. Both of thesetemperature control options monitorand efficiently adjust the heating orcooling of the peltiers in the PCR block.While Simulated Control is adequatefor many applications, Active TubeControl is superior because the changein block temperature is based on thetemperature of the sample itself. Thisensures that both set temperaturesand desired dwell times are accuratelyachieved.

User-Friendly SoftwareThe MBS controlling software hasbeen specially designed for ease ofuse. This is most evident when thegraphical options for editing and running protocols are viewed. It isfurther demonstrated when control

capabilities. The MBS software per-mits efficient management of resultsdata much more effectively than ther-mal cyclers with onboard software.With computer control, you can store unlimited numbers of programs in thefamiliar Windows® file structure, aswell as retrospectively view programresults data (run logs) in graphicaland text-based format. Furthermore,both programs and run logs can be printed for hard copy storage in alab book. These features combined provide traceability that is in line with many GLP procedures.

Gradient CapabilityIf transferring a PCR protocols from analternative instrument manufacturer,or just starting with a new primer set,our gradient blocks easily facilitateprotocol optimisation.

of the blocks is considered. With a few simple mouse clicks, you can program a block and start a protocol.Furthermore, you can either controland program all blocks in unison orprogram each block independently.

Following the start of a run, the gra-phical display provides clear second-by-second information about blocktemperature and experimental parame-ters. In addition, information about allblocks networked to the system isavailable in graphical and tabular for-mat, so you can immediately identifywhich blocks are active and what pro-tocols are running on those blocks.

Printable Programs and Run LogsComputer control provides users withmore benefits than simply a graphicalinterface and large data storage

CAT03 A4_Innenseiten 1 16.12.2003 11:53 Uhr Seite 12

Page 5: Thermal Cyclers - PCR Instrumentation

These blocks can be programmed witha protocol containing a simultaneousrange of annealing temperatures. The optimal annealing temperaturecan subsequently be determined byonboard gradient calculation software.Once optimised, our outstanding well-to-well uniformity (±0.3°C within30 sec) ensures reproducible resultswhen conditions are transferred tostandard protocols. Thermal gradientblocks are available in 0.2ml and0.5ml formats.

Space Saving DesignMBS satellite blocks can be positionedin any arrangement: next to each otheror at opposite ends of the laboratory.This gives you the freedom to design anetwork that makes the most of yourlaboratory space. The MBS does notrequire any bulky power supply unitsto operate. Slots in the lid and blocksurround maintain the airflow in theMBS blocks. This means that theinstruments can be placed up againstone another and still be adequatelycooled. In addition, the block orienta-tion (portrait) makes the MBS anextremely narrow unit, making themost of limited bench space.

Outstanding UniformityThe MBS exhibits an excellent blockuniformity, of ±0.3°C across the blockwithin 30 seconds of reaching the tar-get temperature (measured at 72°C).This block uniformity ensures that you

obtain reproducible and consistentresults, irrespective of the position ofthe samples within the block.

Adjustable Heated Lid DesignOur heated lid design is simple tooperate. In one action, the lid is closedand the heated plate lowered to thecorrect height. Increased pressure andtemperature have been incorporatedto ensure ultimate flexibility in choiceof sample format and sealing options.

Reliability AssuredEach networked MBS block is indepen-dently controlled and has individualMains power supply. If the MBS soft-ware is switched off or if the computercrashes or is switched off, the satelliteunits continue the existing run to completion. The built-in diagnosticsand calibration check facility ensurethe continued reliability of the MBS.

Start-Up Kits AvailableStart-up kits are available to users intwo formats: a one-block option and a two-block option. Both formatsinclude a PC with PCR authorised soft-ware and full configuration. For userswho require more blocks or for thosewho have a system and whose labhas exceeded capacity, the ability toadd single economically priced satellite blocks further enhances thesystem's value. Whether running a few samples or thousands, the MBSsystem offers value for money.

MultiBlock System (MBS)

Edit a program: Editing, saving and downloa-ding programs is easy with the MBS software

In order to reduce the cost of entry into thisadvanced technology we are pleased to intro-duce special start up kits – please enquire

View active block: Both real time temperatureplots from active blocks and archive loggedtemperature plot data can be viewed

CAT03 A4_Innenseiten 1 16.12.2003 11:53 Uhr Seite 13

Page 6: Thermal Cyclers - PCR Instrumentation

PCR Instrumentation 14 14PCR Instrumentation 13 14

MultiBlock System – RoboticCompatible Version (MBSR)

MultiBlock System Specification

The MBS Robotic Thermal CyclingSystem – MBSR

A robot compatible solution for highthroughput PCR applications

The MBSR offers many of the featuresevident in the standard MBS plus theadditional benefit of being robotcompatible.

This compatibility is achieved throughthe unique CD drawer block ejectionmechanism, which makes integrationwith plate handling robots simple andfast. These features, combined withthe stackable design, position theMBSR as the preferred choice for themodern high throughput laboratory.

* MultiBlock performance measurements are traceable to the UK National Physical Laboratory, the equivalent ofU.S.Nist and comparable standards laboratories worldwide.

Benefits• Robot-compatible version of the MBS

for automated sample handling• Networkable with standard MBS

units for flexible system approach• Stackable design for maximum use

of available space

Performance* MBS MBSR

Heated lid temperature 95°C - 120°C 105°C

Dimension of modules (w x d x h) 200 x 300 x 290mm 342 x 545 x 246mm

Power 550W 550W

Weight 8.4kg 20.6kg

Block temperature range 4°C - 99°C 15°C - 99°C

Block heating rate Up to 3°C/sec Up to 3°C/sec

Block cooling rate Up to 2°C/sec Up to 2°C/sec

Block uniformity ±0.3°C within 30sec ±0.3°C within 30sec

Precise control 0.1°C 0.1°C

Software

Number of programs Unlimited

Maximum number of program stages 35

Maximum number of steps per stage 44

Gradient software Yes

Maximum programmed dwell time 18 hours

Pause facility Yes

Temperature ramping Yes

Time increment/decrement Yes

Temp increment/decrement Yes

Auto restart facility Yes

Run "end time" calculation Yes

File protection Yes

Program naming Yes

Ability to edit during cycling Yes

See section 11 for Thermo'sautomated workstations

The MBSR integratedwith the Thermo CRSplate handling robot.

CAT03 A4_Innenseiten 1 16.12.2003 11:53 Uhr Seite 14

Page 7: Thermal Cyclers - PCR Instrumentation

MultiBlock System – RoboticCompatible Version (MBSR)

Product Block Capacity Gradient Facility Temperature Control Available Catalogue Number

MBS

0.5ml standard block 48 x 0.5ml tubes N/A Active tube control HBMBS0596 x 0.3ml tubes Simulated tube control1 plate 0.3ml x 96 well Simulated plate control

Block control

0.2ml standard block 96 x 0.2ml tubes N/A Active tube control HBMBS021 plate 0.2ml x 96 well Simulated tube control

Simulated plate controlBlock control

0.5ml gradient block 48 x 0.5ml tubes Up to 15°C spread Active tube control HBMBSG0596 x 0.3ml tubes Simulated tube control1 plate 0.3ml x 96 well Simulated plate control

Block control

0.2ml gradient block 96 x 0.2ml tubes Up to 15°C spread Active tube control HBMBSG021 plate 0.2ml x 96 well Simulated tube control

Simulated plate controlBlock control

384 well block 1 plate 0.04ml x 384 well N/A Simulated plate control HBMBS384Block control

Robotic MBSR

0.2ml gradient block MBSR 96 x 0.2ml tubes Up to 15°C spread Simulated plate control HBMBS96R1 plate 0.2ml x 96 well Block control

384 well block MBSR 1 plate 0.04ml x 384 well N/A Simulated plate control HBMBS384RBlock control

Accessories

MBS computer N/A N/A N/A HBMBSCOM

MBS software N/A N/A N/A HBMBSSW

Tube thermistor 0.2ml N/A N/A HBPXTTM02

Tube thermistor 0.5ml N/A N/A HBPXTTM05

MultiBlock System Ordering Information

CAT03 A4_Innenseiten 1 16.12.2003 11:53 Uhr Seite 15

Page 8: Thermal Cyclers - PCR Instrumentation

PCR Instrumentation 15 16

Advanced Personal Thermal CyclerDesigned to offer performance, accu-racy and simplicity in thermal cycling,the Px2 incorporates a number ofcustomer driven features built uponthe foundation of previous modelssuch as the successful PCR Express.

The combination of quality enginee-ring, accurate temperature control

Px2 Thermal Cycler

Benefits• Accurate temperature control for

optimal PCR performance• Gradient capability for rapid

protocol optimisation• Outstanding well-to-well uniformity

for reproducible PCR performance• High definition 1/4 VGA display for

enhanced user-friendliness• Keypad with functional softkeys for

improved software navigation• Software designed for easy editing

and protocol manipulation• Multiple interchangeable block

options for flexible choice ofconsumable type

• Adjustable heated lid for flexible choice of consumable type

containing a simultaneous range ofannealing temperatures. The optimalannealing temperature can subse-quently be determined by onboardgradient calculation software. Onceoptimised, our outstanding well-to-well uniformity (±0.3°C within 30sec) ensures reproducible resultswhen conditions are transferred to standard protocols. Thermal gradientblocks are available in 0.2ml and0.5ml formats.

High Definition 1/4 VGA DisplayControl of the Px2 centres on the large,high definition, 1/4 VGA display. Thescreen is designed to improve user-friendliness by providing detailedinformation about the status of thesystem, in graphical or text format,with a minimal amount of user intervention.

and user-friendly software ensuresthat the Px2 is an excellent choice forscientists using PCR as an integraltechnique in their laboratories.

Accurate Temperature ControlOptimal PCR performance can beobtained in our blocks because of theiraccurate and reproducible temperaturecontrol. Our Active or Simulated temperature control methods accountfor the thermal lag that occurs whenheat is transferred from our eightpeltier devices to the sample mix.Both of these temperature controloptions monitor and efficiently adjustthe heating or cooling of the peltiersin the PCR block. While SimulatedControl is adequate for many applica-tions, Active Tube Control is superiorbecause the change in block tempe-rature is based on the the temperatureof the sample itself. This ensures thatboth set temperatures and desireddwell times are accurately achieved.

Gradient CapabilityIf transferring a PCR protocol from analternative instrument manufacturer,or just starting with a new primer set,our gradient blocks easily facilitateprotocol optimisation. These blockscan be programmed with a protocol

CAT03 A4_Innenseiten 1 16.12.2003 11:54 Uhr Seite 16

Page 9: Thermal Cyclers - PCR Instrumentation

Px2

Keypad with Functional SoftkeysProgramming and navigation throughthe various screens of the Px2 ismade easy with the intuitive keypaddesign and softkeys.

User-Friendly SoftwareDay-to-day use of the Px2 is madeeven easier by the integration of user-friendly software. Running, editingand general protocol manipulationcan be performed with limited keystrokes and minimal scrolling between screens.

Multiple Interchangeable BlockOptionsThe Px2 provides excellent flexibilityin sample format for multi-user labs.Six interchangeable block options,including a flat block for in situ appli-cations and a 384 well option forhigh sample throughput, ensure thatyou have a range of choices. Shouldyour requirements change, it is possi-ble to purchase additional blocks rat-her than a complete thermal cycler.

Adjustable Heated Lid DesignOur heated lid design is simple tooperate. In one action, the lid is closedand the heated plate lowered to the

correct height. Increased pressure andtemperature have been incorporatedto ensure ultimate flexibility in choice of consumable format andsealing options.

Variable Speed FanTo cut down on lab noise, we haveintroduced a variable speed fan. Thefan is designed to efficiently dissipateexcess heat but only activates whenneeded.

Px2 Thermal Cycler Specification

Performance*

Temperature control available Active Tube Control

Stimulated Plate Control

Stimulated Tube Control

Simulated Slide Control

Block Control

Block temperature range 4°C - 99°C

Block heating rate Up to 3°C/sec

Block cooling rate UP to 2°C/sec

Block uniformity ±0.3°C within 30 sec

Display resolution 1/4 VGA (320 x 240 resolution)

Block formats available 0.2ml (gradient and standard)

0.5ml (gradient and standard)

384 well

in situ flat block

Dimensions (w x d x h) 240mm x 390mm x 280mm

Power 550W

Weight 9kg

Programming (all unit variants)

Number of programs 99

Number of directories 10

Max number of program stages 10

Max number of steps per stage 10

Maximum programmed dwell time 9hrs, 59min, 59sec

Pause facility Yes

Temperature ramping Yes

Time increment/decrement Yes

Temperature increment/decrement Yes

Auto restart facility Yes

Run "end time" calculation Yes

Program naming Yes

* Px2 performance measurements are traceable to the UKNational Physical Laboratory, the equipment of US NISTand comparable standards labratories worldwide.

Product Catalogue Number

Px2 chassis HBPX2

0.2ml gradient block* HBPXBG02**

0.5ml gradient block* HBPXBG05**

Flat block* HBPXBFB**

384 block* HBPXB384**

Tube thermistor 0.2ml HBPXTTM02

Tube thermistor 0.5ml HBPXTTM05

Px2 Ordering Information

* Block only, requires a chassis** Compatible with the popular PCR Express

CAT03 A4_Innenseiten 1 16.12.2003 11:54 Uhr Seite 17

Page 10: Thermal Cyclers - PCR Instrumentation

PCR Instrumentation 17 18

Benefits• Accurate temperature control for

excellent PCR performance• 96 well format for flexible sample

throughput• Custom designed software for simple

running of PCR protocols• Low cost for affordable personal

cycling• Compact design for economy of

space• Automatic self-checks for guaranteed

instrument performance• Backed by Thermo's quality

guarantee for peace of mind

Essential PCR MachineThe PxE thermal cycler has beendeveloped to provide all of the fundamental features for value formoney thermal cycling. Focusing onlyon those features essential for PCR,the unit retains a high level ofaccuracy and performance that issynonymous with other products inthe range. The combination ofrobustness, affordability and featuresevident in more sophisticated machines defines its status as a coreindispensable tool for the researcherand student alike.

PxE Thermal CyclerNEW

Accurate Temperature ControlTwo versions of simulated temperaturecontrol are available on the PxE, onefor use with plates and a second foruse with tubes. These methods ensureexcellent PCR performance byadjusting heating and cooling of thePCR block. Controlling thermal transferin this way allows you to accuratelyachieve set protocol parametersbecause variations in consumabletype and sample load are accountedfor. If you have optimised protocolson old thermal cyclers, block controlis also available.

96 Well FormatThe PxE thermal cycler is available in a choice of two block formats, providing great flexibility in the choiceof consumable format and through-put. The 0.2ml block has a 96 tubecapacity and accommodates tubes,plates or strips. The 0.5ml holds48 samples in 0.5ml tubes and 96 samples in 0.3ml tubes, platesor strips.

Custom Designed SoftwareSimple running and editing of PCRprotocols is a key consideration inthe PxE. The unit's custom designedsoftware ensures that even researcherswith no experience of our user interface find the unit easy to program.

Low CostFocusing on all that is essential in athermal cycler has allowed Thermo toproduce an affordable instrumentwithout compromising quality or PCRperformance.

Compact DesignWith bench space at a premium inmany laboratories, the small footprintof the PxE is desirable for manyusers. The compact design is alsofunctionally relevant because itfacilitates airflow around the unitto improve heat dissipation, thusenhancing protocol run times.Additionally, as the airflow is ventedthrough the lid, the clearance requirement around the unit isreduced.

Automatic Self-ChecksConsistent performance of the PxEis assured every time the unit ispowered up. Self-checks are runautomatically to stringently test theperformance of the unit prior to use with valuable samples. Thisaction ensures that the instrumentadheres strictly to the published specification.

Quality GuaranteedEach PxE instrument undergoes astrict test procedure and is subse-quently awarded a certificate ofconformity and quality guarantee.This document provides assurancethat the instrument meets or exceeds its published specificationprior to dispatch from the factory.

CAT03 A4_Innenseiten 1 16.12.2003 11:54 Uhr Seite 18

Page 11: Thermal Cyclers - PCR Instrumentation

Description Catalogue Number

0.2ml fixed block 96 well thermal cycler (96 well block – holds 96 tubes or 1 x 96 well plate) HBPXE02

0.5ml fixed block 96 well thermal cycler (96 well block – holds 48 x 0.3ml tubes, 1 x 96 well plate) HBPXE05

PxE Ordering Information

PxE Thermal CyclerNEW

PxE Thermal Cycler Specification

Performance*

Temperature control available Stimulated Tube Control

Stimulated Plate Control

Block control

Block temperature range 4°C - 99°C

Block heating range Up to 3°C/sec

Block cooling rate Up to 2°C/sec

Block uniformity ±0.5°C within 30 sec

Block options 96 x 0.5ml (standard)

96 x 0.2ml (standard)

Dimensions (w x d x h) 244mm x 405mm x 285mm

Weight 7.2kg

Programming (all unit variants)

Number of programs 99

Number of directories 6

Max number of program stages 10

Max number of steps per stage 10

Ability to edit during cycling Yes

Auto restart during power failure Yes

Pause facility Yes

Temperature ramping Yes

Time increment/decrement Yes

Temperature increment/decrement Yes

Auto restart facility Yes

Run "end time" calculation Yes

Program naming Yes

* PxE performance measurements are traceable to the UK National Physical Laboratory, the equivalent of U.S.Nist and comparable standards laboratories worldwide.

CAT03 A4_Innenseiten 1 16.12.2003 11:54 Uhr Seite 19

Page 12: Thermal Cyclers - PCR Instrumentation

PCR Instrumentation 19 20

Accurate Temperature ControlOptimal PCR performance can beobtained in our blocks because of theiraccurate and reproducible temperaturecontrol. Our Active or Simulated temperature control methods accountfor the thermal lag that occurs whenheat is transferred from our peltierdevices to the sample mix. Both ofthese temperature control optionsmonitor and efficiently adjust theheating or cooling of the peltiers inthe PCR block. While Simulated Controlis adequate for many applications,Active Tube Control is superior becau-se the change in block temperature isbased on the the temperature of thesample itself. This ensures that bothset temperatures and desired dwelltimes are accurately achieved.

Auto-Adjustable Heated LidThe auto-adjustable heated lid ena-bles you to undertake oil free thermalcycling without having to set the lidheight to suit the tube type in use.

Multiple Block OptionsThe PCR Sprint has two block options:a 0.2ml block, which has space for24 tubes, and a 0.5ml block, whichhas space for 20 tubes. The blockcan be interchanged if necessary byusing the thumbscrews evident onthe block surround.

User-Friendly SoftwareThe proven user interface allowseven the most sophisticated programsto be created and stored with ease.Advanced edit features enable timeand temperature increments anddecrements to be programmed easily.Up to 60 programs can be storedunder user-defined names within six sub-directories. The sub-directorystructure allows multiple users tostore programs within their own storage space.

Small FootprintThe PCR Sprint is compact bydesign, which makes the unit anideal personal thermal cycler.

Personal Thermal Cycler for Userswith Low Throughput RequirementsThe PCR Sprint thermal cycler is a lowthroughput personal cycler ideal forusers who perform small amounts ofPCRs on a regular basis. The unitoffers two main choices of sampleblock format. Both formats performsimilarly to more sophisticated andexpensive machines, making the PCRSprint an affordable choice for manyusers.

Benefits• 20 or 24 well capacity for users with

a low throughput requirement• Small footprint for maximum use

of available space• 0.2ml or 0.5ml block formats for

flexible choice of sample format• Auto-adjustable heated lid for ease

of use

PCR SprintThermal Cycler

CAT03 A4_Innenseiten 1 16.12.2003 11:54 Uhr Seite 20

Page 13: Thermal Cyclers - PCR Instrumentation

Product Capacity Catalogue Number

0.5ml PCR Sprint 20 x 0.5ml tubes HBSP05

0.2ml PCR Sprint 24 x 0.2ml tubes HBSP02

0.5ml replacement block 20 x 0.5ml tubes HBSPB05

0.2ml replacement block 24 x 0.2ml tubes HBSPB02

PCR Sprint Ordering Information

PCR Sprint Specification

PCR SprintThermal Cycler

Performance*

Temperature control available Active Tube Control

Simulated Tube Control

Block Control

Block temperature range 4°C - 99°C

Block heating rate Up to 3°C/ sec

Block cooling rate Up to 2°C/ sec

Block uniformity ±0.5°C within 15sec

Heated lid temperature range 95°C - 120°C

Standard accessories Tube thermistor for 0.2ml or 0.5ml block

Interchangeable block modules 0.5ml

0.2ml

Dimensions (w x d x h) 180mm x 300mm x 230mm

Power 250W

Weight 7.5kg

Programming (all unit variants)

Number of programs 60

Number of directories 6

Maximum number of program stages 5

Maximum number of steps per stage 5

Maximum programmed dwell time 9 hr 59 min 59 sec

Pause facility Yes

Temperature ramping Yes

Time increment/decrement Yes

Temperature increment/decrement Yes

Auto restart facility Yes

Run "end time" calculation Yes

Program naming Yes

* PCR Sprint performance measurements are traceable to the UK National PhysicalLaboratory, the equivalent of U.S.Nist and comparable standards laboratories worldwide.

CAT03 A4_Innenseiten 1 16.12.2003 11:54 Uhr Seite 21