Page 1: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

The world leader in serving science

Methods for LTQ OrbitrapA Guided Tour with Examples

Dr. Michaela Scigelova

Page 2: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

2

Outline

Data acquisition strategies Method setup and examples

• OT_LTQ_Big6• MS in Orbi + 6 MS/MS in ion trap

• OT_3OT• MS in Orbi + 3 MS/MS in Orbi

• NL_MS3_AccMass• Phosphopeptides with accurate NL • MS in Orbi, MS2 in Orbi, and MS3 in ion trap

• OT_MSA• Phosphopeptides• ‘composite’ MS2 and MS3 spectrum = MSA• MS in Orbi, MSA in LTQ

Page 3: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

3

What is the objective of the analysis?

One or more of these:• Maximise the protein ID• Boost the sequence coverage• Phosphorylation• De novo sequencing• Quantitation

Page 4: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

4

Data Acquisition Strategies

Parallel mode• Survey MS in Orbitrap• MS/MS in LTQ

For highest protein ID and coverage

assuming very complex mixtures

Serial mode• Survey MS in Orbitrap • MS/MS in Orbitrap

For de novo sequencing, PTM discovery

Page 5: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

5

Mass Accuracy and Resolution Settings

Resolution Mass Accuracy

Survey MS (OT) 15,000 3 ppm

MS/MS (OT) 7,500 3 ppm

Resolution Mass Accuracy

Survey MS (OT) 60,000 3 ppm

MS/MS (LTQ) Unit ~200 ppm

Parallel Mode

Serial Mode

0.9 s

0.1 s

0.35 s

0.2 s

Page 6: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

6

AGC settings LTQ

Full MS 3.00e+04 IT 50ms SIM 1.00e+04 IT 100ms MSn 1.00e+04 IT 100ms

Page 7: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

7

AGC settings orbitrap

Full MS 5.00e+05 - 1.00e+06 IT 500ms SIM 1.00e+05 - 2.00e+05 IT 500ms MSn 1.00e+05 - 2.0e+05 IT 500ms

Page 8: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

8

Settings for HCD, Lock mass

Act.Q 0.13-0.14 Normalized Collision Energy 60-70

specify the FT lock mass abundance (%) Diagnostics/Set Device Default value 10%

Page 9: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

The world leader in serving science

OT_LTQ_Big6

MS in Orbi + 6 MS/MS in ion trap

Page 10: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

10

Parallel Mode – Method Setup

Step-by-step guide Method BIG 6 Tune settings

• Max fill time 100 ms• Number of microscans 1

Page 11: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

11

Parallel Method – starting the method

Page 12: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

12

Parallel Method – Survey MS

Page 13: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

13

Parallel Method – Add another 6 scan events

Page 14: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

14

Parallel Method – BIG 6

Save yourself the typing: use copy and paste!

Page 15: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

15

Parallel Method – DDA Scan event 2

Can be used for directing MS/MS to a preferred region – e.g. 800-1300 in glycopeptide analysis

Page 16: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

16

Parallel Method – DDA Scan event 2

Page 17: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

17

Parallel Method – DDA Scan event 2

Ignore

Page 18: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

18

Parallel Method – DDA Scan event 2

This is the key to PARALLEL mode

Page 19: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

19

Parallel Method – DDA Scan event 2

Page 20: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

20

Parallel Method – DDA Scan event 2

Import lists of target parent ions (parent list) or contaminants (reject list)

Page 21: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

21

Parallel Method – DDA Scan event 2

Reject ‘1+’ and ‘unassigned’ for max number of PROTEIN ID

Page 22: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

22

Parallel Method – DDA Scan event 2

Leave unticked if going for max sequence COVERAGE

Page 23: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

23

Parallel Method – DDA Scan event 2

Even better COVERAGE: if you inject sample 3 times and set the following:

Page 24: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

24

Parallel Method – DDA Scan event 2

Page 25: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

25

Parallel Method – DDA Scan event 3

Default settings:

WRONG!!!!

!

WRONG!!!!

!

Page 26: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

26

Parallel Method – DDA Scan event 3

Page 27: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

27

Parallel Method – DDA Scan event 4

Page 28: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

28

Parallel Method – DDA Scan event 5

Page 29: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

29

Parallel Method – DDA Scan event 6

Page 30: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

30

Parallel Method – DDA Scan event 7

Page 31: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

31

When you have built your method.. Check it!

Page 32: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

32

Who Is Working this Way?

John Yates (Scribbs Inst.) Ruedi Aebersold (ETH Basel) David Goodlet (ISB Washington)

Yates et al., AC 2006, 78, 493-450

Page 33: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

The world leader in serving science

OT_3OT

MS in Orbi + 3 MS/MS fragmented in LTQ and measured in OT

Page 34: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

34

Serial Method Setup

Full scan in Orbi + 3 MS/MS in Orbi Tune settings

• Max fill time 100 ms• Number of microscans 2 to improve S/N in MSn

Resolution Mass Accuracy

Survey MS (OT) 15,000 3 ppm

MS/MS (OT) 7,500 3 ppm

Can resolve 4+ ions

0.35 s

0.2 s

Page 35: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

35

Serial Method – Full scan MS

Page 36: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

36

Serial Method – MS + 3 MS/MS

Save yourself the typing: use copy and paste!

Page 37: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

37

Serial Method – Major change in ‘Current Segment’

There is NO box ticked!

Page 38: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

38

Serial Method – DDA Scan event 2

Page 39: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

39

Serial Method – DDA Scan event 3

Page 40: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

40

Serial Method – DDA Scan event 4

Page 41: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

41

Who Is Doing It This Way?

Matthias Mann (MPI Martinsreid)• Lock mass ‘trick’• Divide the mass range is sections

Roman Zubarev (Uppsala)

Olsen et al., MPC 2005, 4, 2010-2021

Page 42: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

42

Want ppb Mass Accuracy?

Use Lock mass in the method

Olsen et al., MPC 2005, 4, 2010-2021

Page 43: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

43

Lock Mass - Caveats

Lock mass should be present in your analysis all the time • Can be a background ion• Or spike in a lock mass solution via a Y-connector (P-773,17 nL

swept volume) with fused silica tubing attached to a syringe pump Use multiple lock masses For calibrating MS/MS spectra the lock mass closest to your

parent mass of interest is taken (good for peptides) Once you fill in the lock mass dialog box, you can export it, save

it, and import it to other methods

Page 44: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

The world leader in serving science

NL_MS3_AccMass

MS in Orbi + MS/MS in Orbi + MS3 in LTQ

Page 45: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

45

Strategy for Phosphopeptides

Enrichment!!! Use prominent neutral loss of phosphate in MS2 to trigger MS3 Accurately defined neutral loss

• Use Orbi for MS and MS2• Cuts down on false positive MS3 triggers

MS3 is highly informative for peptide sequence and PTM position Data review is fast and easy – search MS3 scans only Thorough search (MS2 + MS3) provides high confidence ID

Olsen and Mann, PNAS 2004, 101, 13417-22

Page 46: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

46

NL MS3 Method – starting the method

Page 47: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

47

NL MS3 Method – Survey MS

Page 48: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

48

NL MS3 Method – MS2

Page 49: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

49

NOTE: phosphopeptide analysis

NL MS3 Method – MS2

Accurate NL determination avoids false triggers of MS3 more efficient analysis

Page 50: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

50

NOTE: phosphopeptide analysis

NL MS3 Method – MS2

Page 51: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

51

NOTE: phosphopeptide analysis

NL MS3 Method – MS2

Page 52: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

52

NOTE: phosphopeptide analysis

The system recalculates NL for all relevant charge states

NL MS3 Method – MS2

Page 53: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

53

NOTE: phosphopeptide analysis

NL MS3 Method – MS2

Page 54: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

54

NL MS3 Method – MS3

Page 55: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

55

NL MS3 with accurate mass support – MS3

Page 56: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

The world leader in serving science

NL_MSA

MS in Orbi + MSA in LTQ

Page 57: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

57

Multistage Activation - MSA

To improve the signal, it is advantageous to lump together the fragments from both MS2 and MS3 scans and search them as one ‘strong’ scan

Page 58: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

58

CID MS² CID MS³ [M+2H-98]2+

Phosphopeptide analysis with MSA

APPDNLPSPGGpSR 5 fmol

MSA

Data courtesy of Karl Mechtler

Page 59: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

59

MSA Method – starting the method

Page 60: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

60

MSA Method – Survey MS

Page 61: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

61

MSA – data dependent MS2

Page 62: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

62

Dynamic exclusion

Page 63: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

63

Neutral Loss – measured in LTQ

Page 64: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

64

Current Segment

Page 65: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

65

MSn Settings

Page 66: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

66

Monoisotopic precursor selection

Page 67: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

67

Neutral Loss – no need to spell it our for different charge states

Page 68: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

68

Multistage Activation - ON

Page 69: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

69

MSA - Conclusions

MSA enables to activate ions from MS2 that show neutral loss from parent, while still keeping the rest of the MS2 fragments in the trap. The MS3 fragments from activated NL candidates are ‘added’ to the rest of the MS2 fragments

Data can be processed by SEQUEST/BioWorks or Mascot

Page 70: The world leader in serving science Methods for LTQ Orbitrap A Guided Tour with Examples Dr. Michaela Scigelova.

70

Who is Doing it this Way

Donald Hunt (Uni Virginia) Karl Mechtler (IMP, Vienna) Andy West, Peter Francis (GSK, Stevenage)

Schoeder et al, AC2004, 76, 3590-3598.