Top Banner
The UE directive 2004/40/EC and Magnetic Resonance Imaging a debate Thierry Metens Ph D Université Libre de Bruxelles, Hôpital Erasme ain awareness week at the European Parliame
11

The UE directive 2004/40/EC and Magnetic Resonance Imaging a debate Thierry Metens Ph D Université Libre de Bruxelles, Hôpital Erasme Brain awareness week.

Dec 24, 2015

Download

Documents

Edward Preston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: The UE directive 2004/40/EC and Magnetic Resonance Imaging a debate Thierry Metens Ph D Université Libre de Bruxelles, Hôpital Erasme Brain awareness week.

The UE directive 2004/40/EC and Magnetic Resonance Imaging

a debate

Thierry Metens Ph D Université Libre de Bruxelles, Hôpital Erasme

The UE directive 2004/40/EC and Magnetic Resonance Imaging

a debate

Thierry Metens Ph D Université Libre de Bruxelles, Hôpital Erasme

Brain awareness week at the European Parliament

Page 2: The UE directive 2004/40/EC and Magnetic Resonance Imaging a debate Thierry Metens Ph D Université Libre de Bruxelles, Hôpital Erasme Brain awareness week.

Magnetic resonance Imaging MRIMagnetic resonance Imaging MRI

Clinical MRI: strong C° evolution since early 1980's =>unique medical imaging progress

worldwide > 20000 scanners & > 60M exams/year

Brain MR:unequalled tissue resolution (T1, T2, MTC, ...) Emergency brain Imaging acute stroke, Diffusion MRI

Brain structural connectivity: Nerve Tractography

Brain activation: fMRIBrain metabolism: MRS

WIP: Interventional procedures, brain peroperative MRI ,MRI&FUS thermotherapy/drug delivery, ...

MRI Both clinical and scientific unique data

Page 3: The UE directive 2004/40/EC and Magnetic Resonance Imaging a debate Thierry Metens Ph D Université Libre de Bruxelles, Hôpital Erasme Brain awareness week.

MRI: the technical ingredients MRI: the technical ingredients

Main Magnetic field (static) B0: 0.2 - 7.0T (earth ~ 50T)Clinical Brain: optimum @ 3.0T

Pulsed Magnetic field Gradients:1KHz switching rate up to 60mT/m slew rate 200mT/m/ms

Gradient field shielding

Pulsed Radio frequency B1 field: B1 ~ 10-30T @ 10-500MHz Non ionizing radiation

Faraday Shielding

0Bz

Page 4: The UE directive 2004/40/EC and Magnetic Resonance Imaging a debate Thierry Metens Ph D Université Libre de Bruxelles, Hôpital Erasme Brain awareness week.

MRI: the technical ingredients MRI: the technical ingredients

Main Magnetic field: B0 field shielding => Strong B0 gradient

Ferromagnetic objects attraction MAIN DANGER Transient Vertigo

Pulsed Magnetic field Gradients:dB/dt Nerve stimulation, Induced currents in patient and in the close vicinity IEC Limits

Pulsed Radio frequency B1 field:Patient tissue heating SAR IEC Limits

Page 5: The UE directive 2004/40/EC and Magnetic Resonance Imaging a debate Thierry Metens Ph D Université Libre de Bruxelles, Hôpital Erasme Brain awareness week.

Includes any EM device: portable phone, Oven, MRI, ...

Page 6: The UE directive 2004/40/EC and Magnetic Resonance Imaging a debate Thierry Metens Ph D Université Libre de Bruxelles, Hôpital Erasme Brain awareness week.

Includes any EM device: portable phone, Oven, MRI, ...

Exposure limit values & Action values Conflict with MRI

Exposure limit values & Action values Conflict with MRI

exposure limit values: limits on exposure to electromagnetic fields which are based directly on established health effects and biological considerations. Compliance with these limits will ensure that workers exposed to electromagnetic fields are protected against all known adverse health effects

action values: the magnitude of directly measurable parameters, provided in terms of electric field strength (E), magnetic field strength (H), magnetic flux density (B) and power density (S), at which one or more of the specified measures in this Directive must be undertaken.Compliance with these values will ensure compliance with the relevant exposure limit values.

Page 7: The UE directive 2004/40/EC and Magnetic Resonance Imaging a debate Thierry Metens Ph D Université Libre de Bruxelles, Hôpital Erasme Brain awareness week.

Exposure limit values & Action values Conflict with MRI

Exposure limit values & Action values Conflict with MRI

Main Magnetic field: No ELV, AV 200mT clinical MRI 3 T Limit reached < 50 cm from magnet edge, Patient installation impossible (3T)

Movement in B0: 1Hz, 40mA/m2 head & trunk Exceeded <1m from magnet when v = 1m/s (Estimation < 500mA/m2 , factor 10)

Patient installation impossible

Pulsed Magnetic field Gradients: 1KHz10mA/m2 head & Trunk During Scan Exceeded <1m from magnet edgePatient critical monitoring (Anaesthesia, paediatric or IC patient) impossible

interventional procedure impossible ( factor x 50)

Pulsed Radio frequency B1 field: 10-100MHzELV Whole Body 0.4W/Kg, Head&Trunk 10W/Kg, Limbs 20W/Kg,

averaged over 6minutes Almost Never exceeded by Workers

Page 8: The UE directive 2004/40/EC and Magnetic Resonance Imaging a debate Thierry Metens Ph D Université Libre de Bruxelles, Hôpital Erasme Brain awareness week.

The paradoxThe paradox

MRI would have to be replaced in some instances by X ray procedures that have proven deleterious health effects on workers!

The development of MRI could slow down, especially: High field ( >7T), Brain studies, Interventional & therapeutic Targeted drug delivery

Alliance for MRI (UE MP, ESR, Scientific societies: statement & petition)

The directive transposition has been postponed up to April 2012 ( 2008/46/EC)

Page 9: The UE directive 2004/40/EC and Magnetic Resonance Imaging a debate Thierry Metens Ph D Université Libre de Bruxelles, Hôpital Erasme Brain awareness week.

New ICNIRP recommendations for MRI (2009)New ICNIRP recommendations for MRI (2009)

ICNIRP recommends also to take ferromagnetic object risk into account

(nb: yet in the directive obligations of employers)

Page 10: The UE directive 2004/40/EC and Magnetic Resonance Imaging a debate Thierry Metens Ph D Université Libre de Bruxelles, Hôpital Erasme Brain awareness week.

The debateThe debate

The Directive is needed: Protection of Workers requires a legal frame

The limits must be based on safe values from demonstrated adverse health effects, not on vague extrapolations or extreme precautionary values

Studies should be undertaken in order to develop effective measurement devices to ensure limits are not exceeded in real life and to estimate possible MRI long term effects

Propositions: Alliance = exemption/IEC limits? Modification of ELV/AV for MRI?

Page 11: The UE directive 2004/40/EC and Magnetic Resonance Imaging a debate Thierry Metens Ph D Université Libre de Bruxelles, Hôpital Erasme Brain awareness week.

Thank you for your attention and for your essential work

Thank you for your attention and for your essential work