The coupling of HPTLC with DART mass spectrometry Elena S. Chernetsova, Gertrud E. Morlock Intitute of Food Chemistry, University of Hohenheim (Stuttgart, Germany) e-mail: [email protected]
17

The coupling of HPTLC with DART mass spectrometry · DART-MS does not consume the complete analyte quantity from the plate possible to perform HPTLC/DART-MS and HPTLC/ESI -MS from

Mar 22, 2020

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: The coupling of HPTLC with DART mass spectrometry · DART-MS does not consume the complete analyte quantity from the plate possible to perform HPTLC/DART-MS and HPTLC/ESI -MS from

The coupling of HPTLC with DART mass spectrometry

Elena S. Chernetsova, Gertrud E. Morlock

Intitute of Food Chemistry,

University of Hohenheim (Stuttgart, Germany)

e-mail: [email protected]

Page 2: The coupling of HPTLC with DART mass spectrometry · DART-MS does not consume the complete analyte quantity from the plate possible to perform HPTLC/DART-MS and HPTLC/ESI -MS from

DART = Direct Analysis in Real Time(R.B. Cody et al., 2005)

DART is an “open-air” ionization source for mass spectrometry, enabling the ultrafast analysis of solid or liquid samples in the gaseous stream WITHOUT a sample preparation

2

Russian Chemical Reviews 80 (3) 235-255 (2011)

Page 3: The coupling of HPTLC with DART mass spectrometry · DART-MS does not consume the complete analyte quantity from the plate possible to perform HPTLC/DART-MS and HPTLC/ESI -MS from

DART: „open-air“ ionisation source

Volatilization / Desorption / Ionization

3

reproduced from www.ionsense.com

Direct Analysis in Real Time mass spectrometry (DART-MS)

Page 4: The coupling of HPTLC with DART mass spectrometry · DART-MS does not consume the complete analyte quantity from the plate possible to perform HPTLC/DART-MS and HPTLC/ESI -MS from

4

2006-2010• Horizontal gas supply• Manual plate introduction• Low reproducibility!

Coupling TLC/HPTLC with DART-MS

Page 5: The coupling of HPTLC with DART mass spectrometry · DART-MS does not consume the complete analyte quantity from the plate possible to perform HPTLC/DART-MS and HPTLC/ESI -MS from

5

2011 Desorption at an angle Ability for scanning

across the TLC plates Positioning

at the focus of the gas beam is critical

Coupling TLC/HPTLC with DART-MS

Page 6: The coupling of HPTLC with DART mass spectrometry · DART-MS does not consume the complete analyte quantity from the plate possible to perform HPTLC/DART-MS and HPTLC/ESI -MS from

6

Attractive:scanning in the ‚substance window‘

Coupling TLC/HPTLC with DART-MS

Page 7: The coupling of HPTLC with DART mass spectrometry · DART-MS does not consume the complete analyte quantity from the plate possible to perform HPTLC/DART-MS and HPTLC/ESI -MS from

7

visualisation of the gas beambased on colour change during a chemical reaction

for optimal positioning of a TLC plate in DART-MS

Visualisation for the HPTLC plate, treated with sugar solution + β-naphthol reagent

E.S. Chernetsova et al. Some new features of Direct Analysis in Real Time mass spectrometry utilizing the desorption at an angle option // Rapid Commun. Mass Spectrom., 2011. In press.

Coupling TLC/HPTLC with DART-MS

Page 8: The coupling of HPTLC with DART mass spectrometry · DART-MS does not consume the complete analyte quantity from the plate possible to perform HPTLC/DART-MS and HPTLC/ESI -MS from

even the non-alignment of 1 mm is critical

8

Problem: non-ideal alignment of the home-built x-table

Coupling TLC/HPTLC with DART-MS

100 mm

Page 9: The coupling of HPTLC with DART mass spectrometry · DART-MS does not consume the complete analyte quantity from the plate possible to perform HPTLC/DART-MS and HPTLC/ESI -MS from

9

AVAILABLE NOW: x-y-z-table for DART

Coupling TLC/HPTLC with DART-MS

Page 10: The coupling of HPTLC with DART mass spectrometry · DART-MS does not consume the complete analyte quantity from the plate possible to perform HPTLC/DART-MS and HPTLC/ESI -MS from

10

AVAILABLE NOW: x-y-z-table for DART

Coupling TLC/HPTLC with DART-MS

Y-movement

Page 11: The coupling of HPTLC with DART mass spectrometry · DART-MS does not consume the complete analyte quantity from the plate possible to perform HPTLC/DART-MS and HPTLC/ESI -MS from

11

Coupling TLC/HPTLC with DART-MS

1 µg HMF per band

HMF – 5-hydroxymethylfurfural

horizontal scanning by DART-MS

10 µg HMF per band

RSD of 15% and better

Page 12: The coupling of HPTLC with DART mass spectrometry · DART-MS does not consume the complete analyte quantity from the plate possible to perform HPTLC/DART-MS and HPTLC/ESI -MS from

12

Coupling TLC/HPTLC with DART-MS

10000 ng HMF per band

DART-MS spectrumas the analyte identity proof

[HMF+NH4]+

HMF – 5-hydroxymethylfurfural

1 µg HMF per band

horizontal scanning by DART-MSRSD of 15% and better

Page 13: The coupling of HPTLC with DART mass spectrometry · DART-MS does not consume the complete analyte quantity from the plate possible to perform HPTLC/DART-MS and HPTLC/ESI -MS from

Coupling TLC/HPTLC with DART-MS

Movement of the table „left to the right“

1st scan, 0.2 mm/shorizontal scanning

by DA

RT-MS

„right to the left“

2nd scan, 1.0 mm/s

5 zones of 10 µg HMF per band

Page 14: The coupling of HPTLC with DART mass spectrometry · DART-MS does not consume the complete analyte quantity from the plate possible to perform HPTLC/DART-MS and HPTLC/ESI -MS from

HPTLC-ESI-MS:

Elution-Head-based HPTLC-MS

14

DART-MS does not consume the complete analyte quantity from the plate

possible to perform HPTLC/DART-MS and HPTLC/ESI-MS from the same plate, successively.s

cannin

g D

ART

-M

S,SIM

at m

/z179 (c

affe

ic a

cid

)

scannin

g D

ART

-M

S, S

IM a

t m/z

253 (c

hry

sin

e)

propolis extract,

HPTLC for flavonoids

caffeic acid

chrysine

chrysine

HPTLC-DART-MS:

Desorption-based HPTLC-MS

Chrysine

Identification of flavonoids in propolis

poster # 8j

Page 15: The coupling of HPTLC with DART mass spectrometry · DART-MS does not consume the complete analyte quantity from the plate possible to perform HPTLC/DART-MS and HPTLC/ESI -MS from

Conclusions

HPTLC-DART-MS

Scanning analysis across the plate section. No solvents are used, ionization takes place in the flow of gas

Semi-desctructive: possible to repeat the analysis from the same zone

Mass spectra: additional information due to the new mode of ionization

Quantitation capabilities are now higher due to the x-y-z-table

Sensitivity lacking

HPTLC-ESI-MS

Elution of selected zones by the solvent from HPLC pump

Destructive: cuts the zone of interest

Mass spectra contain sodium adducts and multiply charged ions

Quantitation is highly reliable, but accurate zone marking is essential

Sensitivity: generally 2-3 orders of magnitude better than for HPTLC-DART-MS

15

Page 16: The coupling of HPTLC with DART mass spectrometry · DART-MS does not consume the complete analyte quantity from the plate possible to perform HPTLC/DART-MS and HPTLC/ESI -MS from

Our acknowledgements

16

Professor Dr. Schwack (University of Hohenheim, Stuttgart, Germany)

Dr. Dogan (Merck)

Rolf Rolli and Dr. Natsias (CAMAG)

Pete Ryan and Sue Kennerley (KR Analytical, UK)

Brian Musselman (IonSense, USA)

Organizers of the'Chromatographic Methods of Investigating the Organic Compounds'

Page 17: The coupling of HPTLC with DART mass spectrometry · DART-MS does not consume the complete analyte quantity from the plate possible to perform HPTLC/DART-MS and HPTLC/ESI -MS from

Thank you for your attention!