APPLICATION BRIEF ICP-Mass Spectrometry Introduction Titanium (Ti) is commonly used in metallic hip replacement joints. Owing to wear and tear or different types of electrochemical processes that could take place in the body, there is a serious potential of leakage of Ti into the surrounding tissues. As such, it is important to monitor the Ti serum levels in patients that have undergone or are due to undergo a hip replacement surgery. The analysis of Ti in serum by inductively coupled plasma mass spectrometry (ICP-MS) faces many challenges due to significant spectroscopic interferences from Ca on the major isotope of Ti (m/z 48) and other molecular ions such as PO + , SO + , CO 2 + , ArC + and NO 2 + . The PerkinElmer NexION ® 300D ICP-MS is a robust, highly innovative cell-based system benefitting from three quadrupoles with the middle one acting as a universal cell. This cell can be used both as a collision cell using inert gases, such as He, and as a true dynamic reaction cell with the benefits of using the most appropriate reactive gas for the efficient targeting of any spectral interference. In this work, we describe a method to shift the Ti ions away from the interferences using Reaction mode with ammonia as the reaction gas to determine the concentration of Ti in serum. The Advantages of the NexION 300D ICP-MS for the Determination of Titanium in Serum Authors: Fadi Abou-Shakra David Price PerkinElmer, Inc. Shelton, CT
2

The advantages of the NexION 300D ICP-MS for the determination of titanium in serum

Jun 15, 2015

Download

Health & Medicine

PerkinElmer describes a method to shift the Ti ions away from the interferences using Reaction mode with ammonia as the reaction gas to determine the concentration of Ti in serum. 011159_01
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: The advantages of the NexION 300D ICP-MS for the determination of titanium in serum

A P P L I C A T I O N B R I E F

ICP-Mass Spectrometry

Introduction

Titanium (Ti) is commonly used in metallic hip replacement joints. Owing to wear and tear or different types of electrochemical processes that could take place in

the body, there is a serious potential of leakage of Ti into the surrounding tissues. As such, it is important to monitor the Ti serum levels in patients that have undergone or are due to undergo a hip replacement surgery. The analysis of Ti in serum by inductively coupled plasma mass spectrometry (ICP-MS) faces many challenges due to significant spectroscopic interferences from Ca on the major isotope of Ti (m/z 48) and other molecular ions such as PO+, SO+, CO2

+, ArC+ and NO2+.

The PerkinElmer NexION® 300D ICP-MS is a robust, highly innovative cell-based system benefitting from three quadrupoles with the middle one acting as a universal cell. This cell can be used both as a collision cell using inert gases, such as He, and as a true dynamic reaction cell with the benefits of using the most appropriate reactive gas for the efficient targeting of any spectral interference. In this work, we describe a method to shift the Ti ions away from the interferences using Reaction mode with ammonia as the reaction gas to determine the concentration of Ti in serum.

The Advantages of the NexION 300D ICP-MS for the Determination of Titanium in Serum

Authors:Fadi Abou-Shakra

David Price

PerkinElmer, Inc. Shelton, CT

Page 2: The advantages of the NexION 300D ICP-MS for the determination of titanium in serum

For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs

Copyright ©2013, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners. 011159_01

PerkinElmer, Inc. 940 Winter Street Waltham, MA 02451 USA P: (800) 762-4000 or (+1) 203-925-4602www.perkinelmer.com

Analytical methodology

Figure 1 shows a scan for 10 ppb Ti in Reaction mode (NH3 gas flow 0.9 mL/min, rpq 0.2) at m/z 129-134. The m/z of the observed peak suggest that the Ti complex observed is TiNH(NH3)4, as evidenced by the peaks reflecting the Ti isotopic abundance. The formation of this complex was stable and reproducible. Figure 2 shows a calibration curve for Ti at mass 131. Although Xe+ has an isotope at m/z 131, it will not interfere with the Ti measurements since Xe reacts rapidly with NH3:

Xe+ + NH3 → Xe + NH3+ Rate Constant: ≈10-10

Using matrix-matched calibration, Ti was measured in a Seronorm™ Trace Elements Serum L-1 reference material (LOT0903106) diluted 10x with 0.1% HNO3. The results obtained (10.6 ± 0.3 ng/mL) compared favorably with the recommended value for Ti in this sample (11.2 ng/mL).

In order to gauge the detection capability of the method, 10 repeat analyses of a blank containing 10 ppm Ca (in order to simulate the Ca levels in 10x diluted serum) were performed. The observed concentration in the matrix was 0.031 ± 0.00122 ng/mL. This data suggests a detection limit (3 sigma) of 0.004 ng/mL in solution (0.04 ng/mL in serum) and a limit of quantification (10 sigma) of 0.0122 ng/mL in solution (0.12 ng/mL in serum).

Summary

This application brief highlights the potential of the NexION 300D ICP-MS for the analysis of Ti in serum. Using NH3 in Reaction mode, we successfully overcame the variety of interferences from serum on Ti. As a result, we were able to determine the levels of Ti in serum both accurately and reproducibly.

Figure 1. Spectrum for 10 ppb Ti in Reaction mode.

Figure 2. Calibration for Ti in Reaction mode.