Basic EEG

Dr FL Chow

Page 2: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Electroencephalography (EEG)

• Measure spatial distribution of voltage fields and variation over time

• Sum of excitatory and inhibitory postsynaptic potentials from apical dendrites of pyramidal cells in outer layer of cerebral cortex

• Modified by input from subcortical structures, e.g. thalamus, ascending reticular activating system

Page 3: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

↓Transmembrane potential (depolarization) results in excitatory post-synaptic potentials typically on dendrites

Local hyperpolarization leads to inhibitory post-synaptic potentials typically on neuron cell body

Combination of EPSP and IPSP induces currents within and around the neuron and potential field recordable on the scalp. Duration of PSP 100ms ≈ average alpha wave

Page 4: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Polarity

• The dendritic generators have two poles (dipole: + and -) and are oriented vertically

• Scalp electrode signal detection requires synchronous discharge in approximately 10cm2 of cortex

• Differential amplifier records potential difference between two scalp electrodes

• Upward deflection when input 1 is relatively negative compared to input 2

Page 5: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Polarity on single derivation

Input 1 Input 2 Difference Deflection

+ 50 + 80 − 30 Up

+ 50 + 30 + 20 Down

+ 50 + 50 0 −

+ 50 -50 + 100 Down

Page 6: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Montage

• Montage is a collection of derivations for multiple channels recorded simultaneously and arranged in a specific order

• Scalp electrodes applied according to the International 10-20 System

• Fp (frontopolar), F (frontal), C (central), P (parietal), O (occipital) and T (temporal)

• Odd numbers (left), even (right), A (ear)

Page 7: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

The 10-20 International System of Electrode Placement

Page 8: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

The 10-20 International System of Electrode Placement

Page 9: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Referential recording

• Common reference (e.g. ear, vertex), second input is always the reference

• Locate the site of maximal involvement when the reference is inactive

• Reference electrode may not be inactive

Bipolar recording

• Measure difference between nearby points (e.g. anterior-posterior, transverse)

• Localization by phase reversal

• Amplitude means potential difference, not necessarily the most active site

Page 10: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Bipolar recording

Page 11: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Normal EEG

• Maturational changes, wakefulness, drowsiness and sleep, background rhythm, symmetry

• Delta – frequencies below 4 Hz, during sleep in adults, in temporal region during wakefulness and generalized maximal anterior during drowsiness in normal elderly

• Theta – 4 Hz to less than 8 Hz, in children and young adults during wakefulness, during drowsiness in adults

• Alpha – 8 to 13Hz, posterior dominant rhythm, relaxed and eye closed during wakefulness, sinusoidal, right side higher

• Beta – above 13 Hz, most prominent anteriorly, increased during drowsiness, barbiturates, benzodiazepines

• Normal sleep activity (Vertex waves, spindles, positive occipital sharp transients of sleep, K complexes, delta)

Page 12: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

EEG analysis

• Frequency • Voltage • Location • Morphology • Polarity • State • Reactivity • Symmetry • Artifact

Page 13: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Alpha rhythm

Page 14: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Beta rhythm

Page 15: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Vertex wave and sleep spindle

Page 16: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Posterior occipital sharp transients of sleep (POSTS)

Page 17: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

K complex and POSTS

Page 18: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Lambda waves (occipital positive potentials in visual scanning)

Page 19: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Mu rhythm (7 – 11 Hz) from sensorimotor cortex, in wakefulness and drowsiness, asynchronus, asymmetric

Page 20: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Lateral eye movement in drowsiness

Page 21: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Encephalopathy

• Slowing of alpha rhythm, excess slowing during wakefulness, loss of alpha rhythm, loss of normal sleep transients, abnormal arousal pattern, frontal intermittent rhythmic delta (FIRDA), loss of normal variability, loss of reactivity to stimuli, burst suppression, electrocerebral inactivity

• Generalized periodic discharges 1Hz in dementia and myoclonus (CJD), triphasic waves, periodic lateralized epileptiform discharge (PLED) in HSV and stroke, generalized periodic epileptiform discharges (GPED) in post anoxia, SE and drugs; burst suppression (drug)

Page 22: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Breach rhythm (left side skull defect)

Page 23: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Attenuation of background activity (cerebral contusion)

Page 24: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Diffuse slowing post head trauma

Page 25: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Reactivity to sound (attenuation)

Page 26: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Reactivity (attenuation followed by delta activity)

Page 27: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Frontal intermittent rhythmic delta activity

Page 28: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Focal slowing and sharp waves due to cerebral tumor

Page 29: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Triphasic waves in hepatic encephalopathy

Page 30: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Right cerebral infarction in acute renal failure

Page 31: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Generalized polyspikes in myoclonic status epilepticus

Page 32: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Burst suppression pattern (post cardiac arrest)

Page 33: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Alpha coma (frontal, non-reactive)

Page 34: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Electrocerebral inactivity

Page 35: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Seizures in ICU

• Prevalence higher in those after convulsive status epilepticus and in neurological ICU

• Seizures in encephalopathy tend have slower frequencies, less defined onset and offset

• Diagnosis of seizure activity requires evolution in frequency, morphology or location of EEG pattern, possibly also clinical condition of the patient

• Evolution may be subtle, differentiation between ictal and interictal can be difficult

Page 36: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Seizure activity begins on left side

Page 37: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Clinical (convulsive) and electrographic seizure

Page 38: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Seizure ends

Page 39: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Nonconvulsive status epilepticus (focal)

Page 40: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Nonconvulsive status epilepticus (generalized)

Page 41: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Burst suppression on barbiturate

Page 42: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Diffuse slow activity

Page 43: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Generalized spikes and polyspikes

Page 44: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Generalized nonconvulsive status epilepticus

Page 45: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Periodic lateralized epileptiform discharge (PLED)

Page 46: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

PLED evolved into electrographic seizure

Page 47: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Seizure activity ends

Page 48: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Artifacts that mimic seizures

Physiological

• Ocular – Movement, electroretinogram

• Muscle

• Sweat

• Tongue and mouth – Glossokinetic potential

• Vascular – ECG, pulse

• Movement – tremor, respiration

• Skull defect

Non-physiological

• Electrodes – disc, wire

• External source – 50Hz, ICU equipment

• Machine – Amplifier, setting

• Patient care – percussion, suctioning

Page 49: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Muscle artifact (relative sparing in mid line derivations)

Page 50: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Muscle artifact (shivering)

Page 51: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Filtered muscle artifact

Page 52: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Muscle artifact (chewing movement)

Page 53: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Snore artifact (groups of fast activity, mid line sparing)

Page 54: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Eye movement (fast down and slow up)

Page 55: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Sweat artifact (irregular delta activty slower than 1Hz)

Page 56: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Electrode artifact (mirror image)

Page 57: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Pulse artifact and electrode artifact

Page 58: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Rhythmic artifact (on hemodialysis)

Page 59: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

Respirator artifact

Page 60: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

F/29. Admitted for repeated convulsion. There was asymmetric limb twitching, pelvic thrusting and difficulty in open the eyes for examination. What did the EEG show? What may help to confirm the diagnosis?

Proposed answer: Normal posterior dominant alpha activity

Page 61: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

M/35. Known epilepsy post head trauma and neurosurgery. Brought to A&E for convulsion. The patient was paralyzed and intubated in A&E, transferred to ICU and connected to ventilator. What did the 16-channel EEG show?

Proposed answer: Generalized electrographic seizure activity

Page 62: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

F/26. Admitted ICU for status epilepticus. Convulsion stopped after put on propofol iv infusion. What did the EEG show and what would you do?

Proposed answer: Generalized electrographic seizure activity, increase antiepileptic

Page 63: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

F/56. Known SLE and cerebral lupus on gabapentin. Admitted ICU for low BP and neutropenic fever. Gabapentin was withhold. Noted mental dullness. The following EEG disappeared after midazolam 2mg iv. What was the diagnosis?

Proposed answer: Muscle artifacts

Page 64: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

M/61. Admitted for fever and confusion. CSF showed lymphocyte pleocytosis, mild elevation in CSF protein and normal blood glucose. What was the clinical possibility and what did the EEG show?

Proposed answer: Herpes simplex encephalitis, periodic lateralized epileptiform discharge (left)

Page 65: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

F/62. End stage renal failure. Admitted for CAPD peritonitis. On cefepime for E.Coli in PD fluid. Noted mental dullness and mild limb twitching. What did the EEG show and what was the possible diagnosis?

Proposed answer: Clinical and electrographic seizure, triphasic waves, possibly cefepime related

Page 66: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

F/47. Presented with weight loss, poor memory, delusional disorder, ataxia and myoclonus. What did the EEG show? What was the possible diagnosis? What differential diagnosis has to be excluded?

Proposed answer: Creuztfeldt-Jacob disease, Hashimoto’s encephalopathy

Page 67: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

M/57. Admitted for acute ischemic stroke. Noted impaired conscious state and recurrent right side focal convulsion. What did EEG show?

Proposed answer: Attenuation and slow rhythm on left side

Page 68: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

F/50. Post renal transplant on Tacrolimus. Admitted ICU for repeated tonic-clonic convulsion. What did the EEG show and what was the likely diagnosis?

Proposed answer: Sharp and slow waves, consider posterior leucoencephalopathy syndrome

Page 69: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

M/42. Admitted ICU post ROSC after 20min of CPR for sudden cardiac arrest due to VF. Remained comatose. EEG monitoring on therapeutic hypothermia. What did the EEG show and what would be the prognosis of this patient?

Proposed answer: Reactivity, favorable

Page 70: The 10-20 International System of Electrode Placementksiresearchorg.ipage.com/tdr/references/BasicEEG.pdf · The 10-20 International System of Electrode Placement . Referential recording

References

• Primer of EEG with a Mini-Atlas. AJ Rowan and E Tolusky. Butterworth Heinemann 2003

• Atlas of EEG in Critical Care. LJ Hirsch and RP Brenner. Wiley Blackwell 2010

• Prognostic value of continuous EEG monitoring during therapeutic hypothermia after cardiac arrest. AO Rossetti et al. Critical Care 2010 (14): R173

• EEG recordings of Electrodiagnostic Unit and ICU of Caritas Medical Centre for construction of scenarios