MMIICC TTRRAACC MMTT--33000000 TTFF--TTII TTeesstt IInnddiiccaattoorr FFiixxttuurree

OOppeerraattiioonn MMaannuuaall

Page 2: TF-TI_ops

TTeesstt IInnddiiccaattoorr OOppeerraattiioonn MMaannuuaall

MIC TRAC MT-3000 2 2008 Gagemaker, LP.

RCTIMT30009-99

Page 3: TF-TI_ops

TTeesstt IInnddiiccaattoorr OOppeerraattiioonn MMaannuuaall

MIC TRAC MT-3000 3 2008 Gagemaker, LP.

RCTIMT30009-99

Materials Needed: • MIC TRAC MT-3000 base unit and CPU readout • 5/32" hex wrench • Test Indicator • 50 in/lb torque wrench • Test indicator fixture (TF-TI Block) and cap screw • Cloth • Flat face anvil (TF-1F Block) cap screws and washers • ZEP I.D. Red cleaner • Test indicator accessory block (TF-SS)

NOTE: Procedure valid for test indicators with a 0.001" graduation only.

1. Inspect the test indicator visually as follows: • Check for damage and excessive wear. • Check for clear bezel, legible dial face. • Inspect for proper function of the bezel

adjustment and lock. • Inspect for missing or lose screws or worn

contact point. • Check for smoothness of travel by depressing

the indicator shaft throughout the entire travel range. Repair any indicators with restricted movement prior to calibration.

2. Clean both of the receiver pads and the

mounting surfaces of the test indicator fixture using the cloth and ZEP I.D. Red cleaner.

3. Locate the test indicator fixture. Place the

bottom of the fixture on an angle on top of the left receiver pad.

Setup

Page 4: TF-TI_ops

TTeesstt IInnddiiccaattoorr OOppeerraattiioonn MMaannuuaall

MIC TRAC MT-3000 4 2008 Gagemaker, LP.

RCTIMT30009-99

4. While holding the test indicator fixture down

against the top of the receiver, insert the cap screw into the hole.

5. Use a 5/32" hex wrench to tighten the screw.

6. Clean the flat face anvil and place the shoulder

side against the right receiver pad shoulder.

7. While holding the anvil against the receiver pad

shoulder, insert the two cap screws with washers into the holes on either side of the anvil.

8. While applying pressure toward the receiver

pad shoulder, use a 5/32" hex wrench to slightly tighten the screws.

9. Use a 30 in/lb torque wrench to secure the cap

screws.

Setup

Page 5: TF-TI_ops

TTeesstt IInnddiiccaattoorr OOppeerraattiioonn MMaannuuaall

MIC TRAC MT-3000 5 2008 Gagemaker, LP.

RCTIMT30009-99

10. Locate the test indicator accessory block. With

the thumbscrew facing forward, slide the test indicator accessory block over the right flat face.

11. Position the test indicator accessory block so

there is a 1/8" gap between the accessory block and the flat face anvil. Tighten the thumb screw to secure in place.

12. Loosen the screw on the test indicator

accessory block and mount the test indicator in the dovetail slot.

Setup

Page 6: TF-TI_ops

TTeesstt IInnddiiccaattoorr OOppeerraattiioonn MMaannuuaall

MIC TRAC MT-3000 6 2008 Gagemaker, LP.

RCTIMT30009-99

13. Turn the thumb screw on the test indicator

accessory block and adjust the height of the test indicator. Position the contact point into the space between the flat face anvil and the test indicator accessory block.

Note: Loosen the test indicator fixture and adjust

the angle as necessary to position the contact point.

14. Tighten the thumb screw to secure in place.

15. Turn the coarse adjust knob counterclockwise

to center the contact point of the test indicator in the space between the test indicator accessory block and the flat face anvil.

16. Secure the coarse adjust lock.

17. Turn the fine adjust knob clockwise until the

contact point touches the flat face anvil.

Setup

Page 7: TF-TI_ops

TTeesstt IInnddiiccaattoorr OOppeerraattiioonn MMaannuuaall

MIC TRAC MT-3000 7 2008 Gagemaker, LP.

RCTIMT30009-99

18. Turn the test indicator dial to align the needle

with zero.

Setup

Page 8: TF-TI_ops

TTeesstt IInnddiiccaattoorr OOppeerraattiioonn MMaannuuaall

MIC TRAC MT-3000 8 2008 Gagemaker, LP.

RCTIMT30009-99

Materials Needed: • MIC TRAC MT-3000 base unit and CPU readout • 5/32" hex wrench • Seiko printer (optional) • Lightweight gage oil • Test indicator • Gage Calibration Record 1. Press the key pad on the CPU panel.

The readout displays 0.0000.

2. Turn the fine adjust knob clockwise to the first

calibration value. Note: If you pass the calibration value, turn the

dial back and approach the value again, from the same direction. This practice will increase the accuracy of the calibration.

Calibration If using a computer with the MT-3000, go to page 10.

Page 9: TF-TI_ops

TTeesstt IInnddiiccaattoorr OOppeerraattiioonn MMaannuuaall

MIC TRAC MT-3000 9 2008 Gagemaker, LP.

RCTIMT30009-99

3. If you have a Seiko Printer attached to your

MIC TRAC for printing a permanent record of calibration measurements, press the PRINT pad on the CPU panel.

4. Record any deviations on the Gage Calibration

Record or in-house calibration report. Continue with remaining measurements.

5. Remove the test indicator and continue with the

same calibration process for the next test indicator.

6. After calibrating all test indicators, be sure to

remove the test indicator fixture, flat face anvil, and test indicator accessory block from the MT-3000. Oil the fixtures and anvil and return them to the storage case.

Calibration

Page 10: TF-TI_ops

TTeesstt IInnddiiccaattoorr OOppeerraattiioonn MMaannuuaall

MIC TRAC MT-3000 10 2008 Gagemaker, LP.

RCTIMT30009-99

Materials Needed: • MIC TRAC MT-3000 base unit and CPU readout • Test indicator • Computer • 5/32" hex wrench • CERTIFI software • Lightweight gage oil • MT-4-USB Digital Data Acquisition Card • Brother P-Touch Label Printer (optional) 1. Start Certifi by double clicking the Certifi icon.

CERTIFI Icon

2. Click the New Calibration Report icon.

New Report Icon

Calibration If using a computer with the MT-3000, begin here.

Page 11: TF-TI_ops

TTeesstt IInnddiiccaattoorr OOppeerraattiioonn MMaannuuaall

MIC TRAC MT-3000 11 2008 Gagemaker, LP.

RCTIMT30009-99

3. With the cursor in the Gage Template field,

press the Enter key. 4. Highlight the Caliper 6” line and click on the OK

button.

5. Complete the following Client information:

• Company name and address. • OEM # • Contact name. • PO/Account - purchase order or account

number. • Calibration Date - automatically displays,

but can be changed by typing over the displayed date.

• Print # • Serial #

6. Enter the following information about the gage

you are calibrating: • Serial Number • Location • Gage Manufacturer • Model • Description

Calibration

Page 12: TF-TI_ops

TTeesstt IInnddiiccaattoorr OOppeerraattiioonn MMaannuuaall

MIC TRAC MT-3000 12 2008 Gagemaker, LP.

RCTIMT30009-99

7. You may change the tolerances for the caliper

if you wish. 8. You can enter additional master values by right

clicking your mouse over an existing master value. Select Insert to insert a new line. Then type in the new master value in that line.

9. Click on the CERTIFI Monitor icon to open the

CERTIFI Monitor Window. 10. Choose the “Y” axis from the pull down box.

11. Bring the MIC TRAC jaws together and click

the Zero button twice.

Calibration

Page 13: TF-TI_ops

TTeesstt IInnddiiccaattoorr OOppeerraattiioonn MMaannuuaall

MIC TRAC MT-3000 13 2008 Gagemaker, LP.

RCTIMT30009-99

12. Enter the gage measurements in the Measured

column as follows:

• Turn the Y axis adjust knob to open the caliper jaws until the caliper indicator reads the first calibration value.

Note: If you pass the calibration value, turn the

dial back and approach the value again, from the same direction. This practice will increase the accuracy of the calibration.

• Click the Send To Table bar. Certifi records the value and displays a green box in the In/Out column if the value is within tolerance. A red box displays if the value is not within tolerance. The Deviation column shows the deviation of the measured value from the master value.

13. Change the Next Calibration Date, if

necessary, by typing over the displayed date.

Calibration

Page 14: TF-TI_ops

TTeesstt IInnddiiccaattoorr OOppeerraattiioonn MMaannuuaall

MIC TRAC MT-3000 14 2008 Gagemaker, LP.

RCTIMT30009-99

14. Enter NIST information as necessary. With the

cursor in the single box press the Enter key to bring up a list of current NIST information. For new NIST information select Edit from the menu and press the Preferences box. Select the NIST tab to enter new information.

15. Click the Save button. Note: CERTIFI automatically assigns a filename

to each calibration. DO NOT change the filename to avoid problems with the CERTIFI database.

16. When the confirm window displays, click the

Yes button if you want to print the calibration report.

Calibration

Page 15: TF-TI_ops

TTeesstt IInnddiiccaattoorr OOppeerraattiioonn MMaannuuaall

MIC TRAC MT-3000 15 2008 Gagemaker, LP.

RCTIMT30009-99

Note: If you have a Brother P-Touch Label Printer for printing calibration stickers, the Confirm window for printing a calibration tag displays. 17. Click the Yes button to print the calibration tag.

Affix the Calibration Tag to the caliper. 18. Remove the test indicator and continue with the

same calibration process for the next test indicator.

19. After calibrating all test indicators, be sure to

remove the indicator fixture, flat face anvil, and test indicator accessory block from the MT-3000. Oil the fixtures and anvil and return them to the storage case.

Calibration